EVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY L. Lança 1, * and A. Silva 2

Similar documents
Radiation Protection Dosimetry Advance Access published April 22, 2008

Shielding Calculation: Radiographic Room. Jerry Williams

Evaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray

A Standardised Approach to Optimisation

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

The effect of breast shielding during lumbar spine radiography

Patient dose in routine X-ray examinations in Yazd state

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

Radiation Protection Dosimetry Advance Access published September 13, 2006

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

Retrospective evaluation of exposure index (EI) values from plain radiographs reveals important considerations for quality improvement

Student Perception of Optimisation compared with the Evidence of Optimisation in Paediatric Radiography Case Studies

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY

THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES

Research Article. Corresponding author H. Osman

NEW YORK CITY COLLEGE OF TECHNOLOGY THE CITY UNIVERSITY OF NEW YORK. Department of Radiologic Technology & Medical Imaging

Proteus XR/f Patient positioning guide

Electronic collimation and radiation protection in paediatric digital radiography: revival of the silver lining

Chiropractic. Table of Contents SCHEDULE OF FEES. Schedule EFFECTIVE MARCH 11, 2013

The print quality of this copy is not an accurate representation of the original.

Doses to Patients arising from Dental X-ray Examinations in the UK, A Review of Dental X-ray Protection Service Data

CT Quality Control Manual FAQs

Unit 1 Exploring and Understanding Data

JMSCR Vol 06 Issue 12 Page December 2018

X-ray (Radiography) - Bone

Evaluation of radiation dose and image quality following changes to tube potential (kvp) in conventional paediatric chest radiography

Assessing the standard dose to standard patients for x-ray investigations

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

ABSTRACT I. INTRODUCTION

THE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS

Jaydev K. Dave a) * Department of Radiology, Thomas Jefferson University, Philadelphia, PA 19107, USA

Radiography. 1. Introduction. 2. Documentation of Compliance. 3. Didactic Competency Requirements. 4. Clinical Competency Requirements

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

CHAPTER 3 Describing Relationships

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

IRMER Ionising Radiation Medical Exposure Regulations

ORTHOSCAN MOBILE DI POSITIONING GUIDE

The impact of SID and collimation on backscatter in radiography

Functional Movement Test. Deep Squat

HEALTH SCIENCES AND ATHLETICS Institutional (ILO), Program (PLO), and Course (SLO) Alignment

Montgomery County Community College RAD 224 Clinical Education V

Appendix I. List of stakeholders consulted with on the Patient Radiation Protection Manual and members of the Medical Exposure Radiation Unit

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

Radiological Technology (Radiography): Emergence of a Profession

QUICK START MANUAL 1

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

Trends in Patients Exposure Doses during Radiographic Examination in Japan

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Radiation Protection education of Radiology trainees

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

IAEA-TECDOC-796. Radiation doses in diagnostic radiology

Bontrager, Kenneth: Radiographic Positioning and Related Anatomy 8 th edition, C.V. Mosby, 2010, ISBN# #882 Scan-trons and pencils

BRS BASIC RADIOLOGICAL SYSTEM USER MANUAL

Table of Contents. Plots. Essential Statistics for Nursing Research 1/12/2017

Medica Health Plans. Minnesota Fee Schedule Revised 5/1/2016 NEW PATIENT EXAMS: MN Medicaid. Medicare

Frank: Merrill's Atlas of Radiographic Positioning & Procedures, 12th Edition

DENTAL HMO EMPLOYER SPONSORED or VOLUNTARY

Dosimetry in digital mammography

Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program

Body Bar FLEX. Exercises for the Core and Abdominals. by Gordon L. Brown, Jr. for Body Bar, Inc.

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Establishment of local diagnostic reference levels for quality control in intraoral radiography

Institutional review board approval was obtained prior to the start of this study.

Code of Practice for Radiation Protection in Dentistry. Code of Practice For Radiation Protection in Dentistry

How to interpret scientific & statistical graphs

CT Dose Reduction in Pediatric Patients

Iranian Journal of Medical Physics

Chapter 1: The Human Organism

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

1 Course Syllabus + Study Guide for Lecture and Laboratory

ESI Wellness Program The BioSynchronistics Design. Industrial Stretching Guide

Do the same as above, but turn your head TOWARDS the side that you re holding on to the chair.

ATHLETIC CONDITIONING ON THE ARC BARREL

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

DRIVE CLEAR OF PAIN A COMFORTABLE RIDE WITH ERGONOMIC ADVICE THE KEY TO YOU WILL BE SITTING PRETTY SIMPLE TIPS FOR

DENTAL HMO EMPLOYER SPONSORED or VOLUNTARY

4.3 Measures of Variation

The radiologist and the raiders of the lost image

Measurement of Absorbed Doses in Anatomical Phantoms

Audit Report. National Audit of Paediatric Radiology Services in Hospitals

I. INTRODUCTION III. OVERALL DESIGN

Radiation Safety in the Catheterization Lab

Bony Thorax. Anatomy and Procedures of the Bony Thorax Edited by M. Rhodes

Managing Patient Dose in Computed Tomography (CT)

Skyscan 1076 in vivo scanning: X-ray dosimetry

TRAINING THE CORE BEGIN WITH ONE SET OF ALL 17 EXERCISES FOR A TOTAL OF 250 REPS. NEXT, MOVE TO TWO SETS FOR A TOTAL OF 500 REPS.

Identify two variables. Classify them as explanatory or response and quantitative or explanatory.

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations

Undertaking statistical analysis of

Scientific Exhibit Authors:

International Journal of Radiation Research, January 2017

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Radiographic Procedures 1

Payment Policy. Chiropractic Care. Policy Specific Section: September 10, 2012 November 10, 2012

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations

PERSONAL COMPUTER WORKSTATION CHECKLIST

Transcription:

Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 7 doi:10.1093/rpd/ncn143 EVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY L. Lança 1, * and A. Silva 2 1 School of Health Technology, Lisbon Polytechnics, Lisbon, Portugal 2 Department of Electronic, Telecommunications and Informatics, Aveiro University, Aveiro, Portugal The exposure index (lgm) obtained from a radiographic image may be a useful feedback indicator to the radiographer about the appropriate exposure level in routine clinical practice. This study aims to evaluate lgm in orthopaedic radiography performed in the standard clinical environment. We analysed the lgm of 267 exposures performed with an AGFA CR system. The mean value of lgm in our sample is 2.14. A significant difference (P 5 0.000 0.05) from 1.96 lgm reference is shown. Data show that 72% of exposures are above the 1.96 lgm and 42% are above the limit of 2.26. Median values of lgm are above 1.96 and below 2.26 for Speed class (SC) 200 (2.16) and SC400 (2.13). The interquartile range is lower in SC400 than in SC200. Data seem to indicate that lgm values are above the manufacturer s reference of 1.96. Departmental exposure charts should be optimised to reduce the dose given to patients. INTRODUCTION The purpose of radiation protection is to maintain radiation exposure at the lowest practicable level. The radiation exposure risk should be minimised and this must be guided by two aspects of radiation protection: radiation protection actions and radiation protection principles (1). Radiation protection actions point towards the use of time, shielding and distance to protect patients, personnel and the public. Radiation protection principles deal with the concepts of justification or positive net benefit, optimisation and dose limitation (this concept is applied only for public and occupational exposures). Exposure optimisation should contribute to protect patients from unnecessary exposures during medical diagnosis and the ALARP (As Low as Reasonably Practicable) principle should always be applied in clinical practice. For diagnostic purposes, the optimisation of exposure involves the relationship between three core aspects of the imaging process (2) : (i) choice of radiographic technique; (ii) radiation dose to the patient and (iii) diagnostic quality of the radiographic image. These three aspects are determinants of the diagnostic quality of the radiographic image and depend on the radiographer s options for each individual patient examination. The choice of the most appropriate radiographic technique involves the management of exposure parameters: the patient s radiation exposure and the exposure on the imaging detector to produce the most accurate diagnosis. This means that a correct exposure at the detector should provide optimum image contrast of the radiographic image. *Corresponding author: luis.lanca@estesl.ipl.pt In digital imaging systems, the dose delivered to the patient could be an over or underexposure (3) because of their inherent dynamic range. Overexposure might still provide good image quality, but may cause unnecessary dose being delivered to the patient. Manufacturers provide a wide variation of different exposure index (lgm, log of median exposure) scales to measure the radiation exposure at the detector (4). LgM is in relation to the absorbed dose at the phosphor plate and is determined by the pixel values (3). For AGFA CR systems, the lgm provides the dose feedback indicator (5). Exposure index The lgm value obtained from a radiographic image could be a useful feedback indicator to the radiographer about the appropriate exposure level in routine clinical practice. The AGFA lgm is labelled as lgm and it indicates how close the actual detector dose is to the expected dose. The lgm value is related to detector exposure and it does not replace patient s dose-related parameters such as dose-area product (DAP) or entrance skin exposure (ESE). LgM is the logarithm of the median value of the pixel histogram of the segmented image. The analysis of the segmented histogram produces the AGFA dose feedback number that indicates how close the average detector dose in some region of interest behind the patient was to the average detector dose expected from the speed class (SC) used for acquiring that image (5). The relationship between pixel value and exposure must be known: the expected lgm value for any SC, according to vendor specifications is about 1.96 and should be consistent to a 2.5 mgy exposure measured at the detector (5). Each change of 0.3 (log) # The Author 2008. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org

L. LANÇA AND A. SILVA in lgm corresponds to doubling or halving of dose RESULTS because of its logarithmic nature. For example, if the Table 1 summarises the lgm results in orthopaedic lgm value for a given image is calculated as 2.26, it radiological examinations considered in this study. indicates that the dose was about twice that expected The table shows that in the five most requested for the selected SC. The dose level at the detector is examinations (knee, foot, pelvis, shoulder and determined as the median of the logarithmic pixel lumbar vertebrae), lgm mean value is higher than values in the main histogram lobe. 2.04. As an example, knee is the most requested LgM is related to the X-ray exposure for each examination (73) where lgm mean is higher (2.35) radiological projection that is carried out. The ranging from 1.77 to 2.62 (range 0.85). A wide range patient s exposure must be in order to get a constant of lgm for most of the examinations can be found. dose on the imaging plate (IP) and this varies as a There is evidence that the lgm range varies considerably among the same examination (e.g. knee, 0.85; function of patient attributes (e.g. sex, weight) radiographic technique and exposure parameters. LgM pelvis, 0.95; shoulder, 1.11). value will also vary with the specified SC setting of The mean value of lgm in our sample is 2.14. A the digitiser. This means that doubling the exposure sample t-test at a significance level of 5% shows a dose at the same SC will cause an increase of the significant difference (P ¼ 0.000 0.05) from the LgM value up to 0.3 (log). If the SC is doubled 1.96 lgm reference value. without modifying the mas setting, then the LgM value will decrease by 0.3 (log) (6) Figure 1 shows that median values of lgm are. above 1.96 and below 2.26 for SC200 (2.16) and Exposures in CR systems may cause unnecessary SC400 (2.13). The interquartile range is lower in patient dose due to overexposure. This problem SC400 (0.30) than in SC200 (0.46). The range between should be avoided in routine clinical practice. If dose upper and lower quartiles is smaller in SC400. is maintained at a relatively constant IP, at a value Figure 2 compares lgm values in two groups of that is considered to be appropriate for the exam or patients (female and male). The lgm median is patient type, then dose consistency in the CR environment could be achieved (5) highest for females both at 200 (2.36) and 400 SC. (2.30). However, in this group of patients, there is no The aim of our study was to evaluate lgm in important variability, with an interquartile range of orthopaedic radiography performed in a routine clinical environment. Table 1. Exposure index. MATERIALS AND METHODS Examination Count Mean SD Range In this study, we analysed the lgm of 267 exposures (max min) performed for a period of 1 month. Radiographs were obtained in the routine clinical environment Wrist (with 3 2.69 (2.76 2.62) 0.10 0.14 by three experienced radiographers. For statistical cast) reasons, the exposures obtained at an SC of 100 were Leg 3 2.39 (2.46 2.33) 0.07 0.13 excluded (n ¼ 2) and 265 exposures (125 female and Knee 73 2.35 (2.62 1.77) 0.17 0.85 140 male) obtained at 200 SC and 400 SC were considered. Foot 19 2.25 (2.59 1.87) 0.21 0.72 All the radiological projections were acquired Elbow 2 2.24 (2.26 2.23) 0.02 0.03 using an AGFA CR system (MD-30 image plate). Wrist 12 2.21 (2.53 1.84) 0.25 0.69 Pelvis 17 2.13 (2.56 1.61) 0.23 0.95 Exposure information such as patient-related data, Shoulder 28 2.09 (2.49 1.38) 0.28 1.11 examination data, radiographic technique, exposure Pelvis 6 2.07 (2.36 1.90) 0.16 0.46 parameters and post-processing information were Orthostatic recorded in a spreadsheet. Calcaneus 3 2.05 (2.30 1.87) 0.22 0.43 Minitabw 15 statistical software was used to Lumbosacral 4 2.04 (2.30 1.67) 0.31 0.63 perform the statistical analysis. This software provides the Anderson Darling (AD) test which was junction a Lumbar 24 2.04 (2.62 1.44) 0.29 1.18 used to test the hypotheses that the data (lgm) vertebrae a follow a normal (Gaussian) distribution. Ankle 8 2.03 (2.43 1.77) 0.20 0.66 Hip 6 2.02 (2.30 1.51) 0.32 0.79 The lgm reference value used for this study is 1.96 (5) and a reference standard deviation of 0.2 (6) Hand 4 2.00 (2.10 1.87) 0.12 0.23 Cervical 14 1.93 (2.33 1.67) 0.19 0.66 was selected for the upper and lower limits for statistical analysis. A significance level of P 0.05 was Hand digits 8 1.86 (2.20 1.38) 0.28 0.82 vertebrae used for statistical tests. Two patient gender groups Patella 25 1.86 (2.98 1.02) 0.43 1.96 (female and male) and three patient weight groups (overweight, normal weight and underweight) were Foot digits 6 1.76 (1.87 1.61) 0.12 0.26 considered. a Includes AP and lateral projections. Page 2 of 7

EVALUATION OF EXPOSURE INDEX Figure 1. Boxplot of lgm at 200 and 400 SC. The box contains 50% of the data. The line inside the box indicates the median value of lgm. The upper edge of the box indicates the 75th percentile of lgm, and the lower edge indicates the 25th percentile. The range of the middle two quartiles is the interquartile range. The ends of the vertical lines indicate the minimum and maximum data values. The points outside the ends are outliers. Figure 2. Boxplot of lgm at 200 and 400 SC comparing female to male patients. Page 3 of 7

L. LANÇA AND A. SILVA Page 4 of 7 Figure 3. Probability plot of lgm comparing female to male patients at 1.96 lgm reference. The plot points represent the proportion of lgm failures: note that they do not follow the straight line closely. The fitted line, which is a graphical representation of the percentiles, indicates the lgm 1.96 reference. The lateral lines represent the 95% confidence intervals (CI).

EVALUATION OF EXPOSURE INDEX Page 5 of 7 Figure 4. Probability plot of lgm comparing female to male patients at 1.96 lgm reference in SC200 and SC400.

0.43 (SC200) and 0.42 (SC400). In addition, the lgm variation between minimum and maximum data values in SC200 (1.54 2.98) is wider than in SC400 (2.03 2.62). Male patients have similar median lgm both for SC200 and SC400 (2.07 and 2.05, respectively). SC400 also exhibits the least variability, with an interquartile range of 0.28 and the least difference between minimum and maximum data values (1.74 2.33). SC200 shows wider interquartile range distribution (0.39) and wider lgm values for lower and upper boundaries (1.31 and 2.59). A probability plot of lgm with the reference value of 1.96 and 0.2 SD is shown in Figures 3 and 4. If the data fit the normal distribution AD value will be closer to zero, and the associated P-value will be larger than the chosen P-level (0.05). The AD statistic test shows that both for female and male patients the data do not follow the specified distribution (Figure 3). Female patients shows an AD value of 140.180 (P, 0.001) and male patients show an AD value of 19.870 (P, 0.001). Despite the fact that results from both groups show a significant difference from the normal distribution, male group of patients had a smaller AD value which indicates that the results from this group are closer to the normal distribution. Figure 4 shows the lgm probability plot and compares female with male patients at 1.96 lgm reference in SC200 and SC400, respectively. The results for SC200 show a significant difference (P, 0.001) from the normal distribution for both groups of patients. However, lower AD value is found in male patients (19.771) when compared with female patients (131.586). This means that for SC200 male patient s lgm is closer to the reference values than the female group. SC400 shows that the data follow the fitted distribution line fairly closely for male patients. AD statistical value is 0.824 and should be considered as normal (P. 0.250). Female patients in SC400 do not fit the reference line. The AD statistical test (9.459) is not normal (P 0.001) showing significant statistical difference from the reference value. Concerning the patient s weight, a significant statistical difference is found in the normal weight group (female: n ¼ 98; AD ¼ 127.496; P, 0.001; male: n ¼ 116; AD ¼ 13.499; P, 0.001). In addition, overweight females (n ¼ 21; AD ¼ 20.295; P, 0.001) show a significant difference from 1.96 reference. Although other groups results do not fit very well with lgm reference values, no statistically significant differences were found in underweight female (n ¼ 6), underweight male (n ¼ 8) and overweight male (n ¼ 16) groups. L. LANÇA AND A. SILVA 42% of evaluated exposures were above the limit of 2.26, which indicates that the IP receives at least double the exposure necessary to produce an adequate image. Findings also show that lgm is higher in female patients than in male patients. This may present a real clinical problem because an lgm higher than the manufacturer s recommendation could be construed as the routine practice of overexposure. This may be the result of an inadequate exposure chart, particularly for female patients. Exposure parameters and the choice of the most appropriate SC for each examination should be carefully studied in order to obtain the desired image quality at the lowest exposure dose. The development of an adequate radiographic technique involves the management of exposure parameters, the patient s radiation exposure and the exposure on the imaging detector to produce the most accurate diagnosis. This should be accomplished with an optimisation of exposures and image quality. Further investigations and modifications to exposure charts could lead to a decrease of dose at the detector and a decrease of patient s exposure. Studies performed on a different CR system shows that it is possible to obtain lower exposure indices than those recommended by the manufacturer (7). The establishment of recommended exposure indices remains unclear and the exposure indices used in current clinical practice could be significantly higher than the optimum level. The lgm is also sensitive to a number of other factors, the most critical one being segmentation. Any errors in the segmentation algorithm (including background that does not belong to the body part or excluding portions of the body part) can cause variations in lgm. In a related effect, collimation can also affect lgm (5). Reliable lgm feedback occurs only when the system has been calibrated properly. This study was performed in a routine clinical environment, and data were collected from a CR system that has a normal maintenance programme. CONCLUSIONS Results found in this study seem to indicate that lgm values are higher than the manufacturer s reference level of 1.96. Departmental exposure charts should be optimised in order to provide a significant reduction of dose at the detector. This action, along with further studies for exposure optimisation should result in a substantial reduction of lgm and consequently contribute to the reduction of patient radiation exposure. DISCUSSION The analysis of the results shows that lgm is far above the recommended target of 1.96 (5). At least FUNDING This study was supported by AGFA Healthcare Portugal. Page 6 of 7

REFERENCES 1. Seeram, E. and Brennan, P. Diagnostic reference levels in radiology. Radiol. Technol. 77, 373 384 (2006). 2. European Commission (EC). Optimisation of Protection in the Medical Uses of Radiation. EUR 19793 (Luxembourg: Office for Official Publications of the European Communities) (2002). 3. International Commission on Radiological Protection. Managing patient dose in digital radiology. ICRP Publication 93. Ann. ICRP 34(1) (Oxford: Elsevier) (2004). EVALUATION OF EXPOSURE INDEX 4. Willis, C. E. Strategies for dose reduction in ordinary radiographic examinations using CR and DR. Pediatr. Radiol. 34 (Suppl 3), S196 S200 (2004). 5. Schaetzing, R. Management of paediatric radiation dose using Agfa computed radiography. Pediatr. Radiol. 34 (Suppl 3), S207 S214 (2004). 6. Dose monitoring software user manual (Mortsel, Belgium: Agfa Gevaert) (2003). 7. Peters, S. and Brennan, P. Digital radiography: are the manufacturers settings too high? Optimisation of the Kodak digital radiography system with aid of the computed radiography dose index. Eur. Radiol. 12, 2381 2387 (2002). Page 7 of 7