Quantitation of Protein Phosphorylation Using Multiple Reaction Monitoring

Similar documents
4-Plex itraq Based Quantitative Proteomic Analysis Using an Agilent Accurate -Mass Q-TOF

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Extended Mass Range Triple Quadrupole for Routine Analysis of High Mass-to-charge Peptide Ions

PTM Discovery Method for Automated Identification and Sequencing of Phosphopeptides Using the Q TRAP LC/MS/MS System

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Agilent 6490 Triple Quadrupole LC/MS System with ifunnel Technology. Ultra sensitive quantitative performance

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Determination of Amantadine Residues in Chicken by LCMS-8040

Quantitative Analysis of Amphetamine-Type Drugs by Extractive Benzoylation and LC/MS/MS. Application. Introduction. Authors. Abstract.

Author. Introduction. Small Molecule Pharmaceuticals & Generics

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Direct Analysis of Folic Acid in Digestive Juices by LC/TOF-MS Application

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Application of a new capillary HPLC- ICP-MS interface to the identification of selenium-containing proteins in selenized yeast

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

application Natural Food Colorants Analysis of Natural Food Colorants by Electrospray and Atmospheric Pressure Chemical Ionization LC/MS

Determination of Aflatoxins in Food by LC/MS/MS. Application. Authors. Abstract. Experimental. Introduction. Food Safety

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Agilent 6410 Triple Quadrupole LC/MS. Sensitivity, Reliability, Value

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

The New 6495 Triple Quadrupole LC/MS

High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column

for the Identification of Phosphorylated Peptides

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

Determination of 44 Pesticides in Foodstuffs by LC/MS/MS Application

Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System

Application Note. Author. Abstract. Introduction. Food Safety

Breaking the Sensitivity Barrier in Triple Quad LC/MS Performance

Characterization of Disulfide Linkages in Proteins by 193 nm Ultraviolet Photodissociation (UVPD) Mass Spectrometry. Supporting Information

Simultaneous Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Vitamin D3 and related compounds by ESI and APCI

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer

Application Note. Abstract. Authors. Pharmaceutical

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Comparison of mass spectrometers performances

High Throughput simultaneous analysis of 39 amino acids in various foods without derivatization using LC-MS/MS

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

Author. Introduction. Abstract

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Improving Selectivity in Quantitative Analysis Using MS 3 on a Hybrid Quadrupole-Linear Ion Trap Mass Spectrometer

A FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ

2D-LC as an Automated Desalting Tool for MSD Analysis

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

Modified QuEChERS for HILIC LC/MS/MS Analysis of Nicotine and Its Metabolites in Fish

Authors. Abstract. Introduction. Environmental

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Measuring Lipid Composition LC-MS/MS

Highly Repeatable Ultra Low Detection of Estradiol Using Triple Quadrupole GC/MS in NCI Mode

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Cannabinoid Quantitation Using an Agilent 6430 LC/MS/MS

Designer Cannabinoids

Efficient Quantitative Analysis of THC and Metabolites in Human Plasma Using Agilent Captiva EMR Lipid and LC-MS/MS

Integration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation

Fat-Soluble Vitamins Analysis on an Agilent ZORBAX Eclipse PAH Polymeric C18 Bonded Column

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

Comparison of Relative Quantification of Monoclonal Antibody N-glycans Using Fluorescence and MS Detection

Screening and Speciation of Raw and Processed Meat Products

Analysis of mycotoxins in food matrices using the Agilent Ultivo Triple Quadrupole LC/MS

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Journal of Chemical and Pharmaceutical Research

Determination of Multi-Residue Tetracyclines and their Metabolites in Milk by High Performance Liquid Chromatography - Tandem Mass Spectrometry

Determination of B-vitamins in Energy Drinks by CE/MS/MS

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

Analysis of Vitamins Using an SFC/UHPLC Hybrid System with a Triple Quadrupole LC/MS for Quantification

Comprehensive Two-Dimensional HPLC and Informative Data Processing for Pharmaceuticals and Lipids

Extraction of 25-hydroxy Vitamin D from Serum Using ISOLUTE. PLD+ Prior to LC-MS/MS Analysis

Maria João Mendes Mendonça Barreira Department of Health Promotion and Chronic Deseases, INSA

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Fully Automated Online Sample Preparation AND Quantification of Amiodarone from Whole Blood using CLAM-LC-MS/MS

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System

Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC

Authors. Abstract. Forensic Toxicology. Irina Dioumaeva, John M. Hughes Agilent Technologies, Inc.

LCMS-8030 Application Report. Steroids: Testosterone, Progesterone, Estradiol, Ethinylestradiol

SUPPORTING INFORMATION FOR: CONCENTRATIONS OF POLYBROMINATED DIPHENYL ETHERS, HEXABROMOCYCLODODECANES AND TETRABROMOBISPHENOL-A IN BREAST MILK FROM

Increased Identification Coverage and Throughput for Complex Lipidomes

Transcription:

Quantitation of Protein Phosphorylation Using Multiple Reaction Monitoring Application Note Authors Ning Tang, Christine A. Miller and Keith Waddell Agilent Technologies, Inc. Santa Clara, CA USA This material was presented in a poster at the 56th ASMS Conference on Mass Spectrometry in 2008. Research presented in posters at scientific conferences may include results from instruments or products that are not yet commercially available. Abstract This application note demonstrates quantitation of percentage of phosphorylation of two sites on a single peptide using the HPLC-Chip/QQQ system. 13 C-labeled peptides from the extracellular signal-regulated kinase (ERK) were synthesized and MRM assays were developed to quantify each peptide. The standard curve of each 13 C-labeled peptide was established. The amount of each phosphopeptide from active ERK spiked into human plasma was measured and calculated in reference to the 13 C-labeled standards.

Introduction Peptide quantitation using multiple-reaction monitoring (MRM) has emerged as an important methodology for biomarker validation. MRM on a triple quadruple mass spectrometer provides superior sensitivity and selectivity for targeted compounds in a complex sample. MRM also offers high precision in quantitation and fast scan speed, which makes it an ideal technology for validating biomarkers in a highthroughput fashion. One area of great interest is the quantitation of protein phosphorylation. Reversible protein phosphorylation plays an important role in cell signaling pathways and the percentage of phosphorylation is often very important to signal transduction. The ERK pathway is essential to transmitting signals from many extracellular agents to regulate cellular processes such as proliferation, differentiation and cell cycle progression. Signaling via the ERK cascade is mediated by sequential phosphorylation and activation of protein kinases in the different tiers of the cascade. The recombinant human ERK1 containing an N-terminal GST tag was expressed in E. coli and activated with MEK1. Up to 7 phosphorylation sites were identified from the activated ERK1. The two key regulatory phosphorylation sites have been identified as the neighboring t202 and y204. 1 Phosphorylation of these sites, located upstream of the conserved protein kinase subdomain VIII region implicated in substrate binding, might function by relieving inhibition of substrate access to the active sites of the kinase domain. The peptides that contain either no phosphorylation site, a single phosphorylation site or both phosphorylation sites have been identified. Hence, the degree of phosphorylation of these two sites is very important to the overall function of ERK1 in the signal transduction pathway. Therefore, in this study, we developed an MRM-based assay to quantify the percentage of phosphorylation by reference to 13 C-labeled peptides. Experimental Sample Preparation Figure 1. HPLC-Chip/QQQ system for protein quantitation. Four peptides labeled at the valine residue with 13 C were purchased. The 4 peptides were as follows: 1) without phosphorylation, 2) with single phosphorylation at t202 or 3) with single phosphorylation at y204 and 4) double phosphorylation at both t202 and y204. Active MAP kinase ERK1 was purchased from Millipore (Temecula, CA). It was denatured, reduced, alkylated and digested using trypsin. 13 C-labeled peptide standards were purchased from Cell Signaling Technology (Danvers, MA). The peptides were labeled at the valine residue. Human plasma was purchased from Sigma (St. Louis, MO). The plasma was depleted of 6 highly abundant proteins using an Hu-6 immunoaffinity column (Agilent) following the standard protocol. After depletion, the sample was buffer-exchanged into an ammonium bicarbonate solution, then reduced, alkylated (IAA) and digested with trypsin under denaturing conditions. Internal calibration curves were acquired using different amounts of 13 C-labeled peptides spiked into 1 µg of human plasma digest. ERK digest (35 fmol/µg plasma protein) was spiked into the samples as the internal standard. LC/MS Analysis An Agilent 6410 Triple Quadrupole (QQQ) LC/MS was connected to the HPLC-Chip interface (see Figure 1 above). HPLC-Chip Conditions: Protein ID chip with 150 x 0.075 mm analytical column and 160 nl enrichment column. Sample load: 1 µg of human plasma digest spiked with different amounts of standard protein digests Injection volume: 1 µl Flow: 300 nl/min analytical pump, 3 µl/min loading pump Mobile phases A: 0.1% formic acid (FA) in water, B: 90% acetonitrile, 0.1% FA Gradients: 2% B to 42% B at 20 min, 90% B at 23 25 min, then 2% B at 25.1 min Stop time: 35 min Triple Quadrupole MS Conditions: Drying gas: 6 L/min, 325 C Fragmentor: 100 V Collision energy: optimized for each transition Dwell time: 70 ms Delta EMV: 400 V 2

Results and Discussion Previous work 2 demonstrated that peptide quantitation using the HPLC-Chip Agilent 6410 Triple Quadrupole LC/MS can be achieved down to 10 amol. In this study, we took on the challenge of quantitating the degree of phosphorylation of two sites on a peptide from ERK protein (amino acid residues 190-208). The sequence and the product ions selected for monitoring the synthetic peptides in the MRM assay are illustrated in Figure 2 and Table 1. The selection of the transitions was critical because of the similarity of the sequences. Selection of Transitions for MRM Assay Chromatographic Separation of the 4 Peptide Standards The transition from 753.3 > 979.9 is the same for both the t202 and y204 peptide, thus it was important to have complete chromatographic separation. The four 13 C-labeled peptide standards were mixed at equal molar ratios and 100 fmol was injected on the HPLC-Chip. A gradient method was developed to separate the 4 peptides (Figure 3). b 14 TY: t202: y204: t202y204: y 16 P y 16 y 16 P P P Figure 3. MRM chromatogram of four 13 C-labeled peptides showed complete chromatographic separation. Figure 2. Sequence and product ions selected for monitoring the four 13 C-labeled peptides. Each peptide was named by the phosphorylation state at amino acids 202 and 204. Phosphotyrosine was labeled as lower case y and phosphothreonine was labeled as lower case t. V represents 13 C-labeled Valine. TY t202 y204 t202y204 Precursor ion Product ions 545.3 615.3 782.5 [M+3H] 3+ b 2+ 14 753.3 615.3 979.9 [M+2H] 2+ y 2+ 16 753.3 979.9 695.3 [M+2H] 2+ y 2+ 16 780.0 647.6 695.3 [M+2H] 2+ y 3+ 16 -H 3 PO 4 Table 1. The charge states, type of fragment and m/z values of the precursor ions and product ions monitored for each 13 C-labeled peptide standard. Bold indicates transition shared between two peptides. 3

QC of incoming phosphopeptide standards We also performed quality control of the incoming 13 C-labeled peptides using the MRM assays developed. The four 13 C-labeled peptide standards were analyzed individually (500 fmol on-column). The non-phosphorylated peptide (TY) was pure and does not contain phosphorylation residues (Figure 4). The two singly phosphorylated peptides (t202 and y204) contained a small percentage of non-phosphorylated peptide (TY). The doubly phosphorylated peptide (t202y204) contained both the singly phosphorylated peptide (t202) and the non-phosphorylated peptide (TY). However, overall the percentage of impurities was relatively low (Table 2). The purity of each synthesized peptide was above 97.6%. Quantitation Curves for the Four 13 C-labeled Peptide Standards For each LC/MS analysis, 1 µg of human plasma digest was injected. The four calibration curves all had excellent linearity with correlation coefficients better than 0.995 (Figure 3). Calibration curves were acquired using human plasma digest spiked with both 13 C-labeled peptides and ERK1 tryptic digest. Limit of quantitation (LOQ) was determined to be 500 amol and limit of detection (LOD) was 50 amol. Figure 4. The four 13 C-labeled peptide standards were injected individually at 500 fmol on-column. Trace amounts of incomplete phosphorylation were observed. The percentage of impurities for each synthetic peptide was analyzed by MRM and calculated in Table 2 below. Peptides Percentage of components TY t202 y204 t202y204 TY 100% t202 0.6% 99.4% y204 0.2% 99.8% t202y204 0.4% 2% 97.6% Table 2. The percentage of each component in the synthetic peptides was calculated by the ratio of MRM peak area. All four peptides are greater than 97% pure. Figure 5. The calibration curves of the four 13 C-labeled peptides spiked into human plasma. 4

The Degree of Phosphorylation at t202 and y204 in Active ERK1 Protein The degree of phosphorylation can be measured by the relative percentage of each peptide as shown in Figure 2. The ratio between the abundance of unlabeled peptide from ERK tryptic digest and the labeled peptide in each sample provided the absolute amount of unlabeled peptide. The percentage for each phosphorylation form was derived from these results and listed in Table 3 below. The relative standard deviation was below 15%. % Molar ratio RSD (n=9) TY 20% 0.13 t202 25% 0.15 y204 21% 0.12 t202y204 34% 0.08 Table 3. Relative amount of four different peptides detected in active ERK1. RSDs of all measurements are below 15%. Conclusions An MRM-based method to quantitate the different phosphorylation states on a single peptide was developed. A critical part of the method development was the chromatographic separation of the target peptides. Because of the sequence similarity between these peptides, the selection of transitions was critical. Chromatographic separation of the peptides allowed the same transition to be used for different peptides. The microfluidic-based HPLC-Chip has been demonstrated to be sensitive, reproducible and capable of fast separation, especially in combination with a triple quadrupole MS/MS. References 1. D. L. Charest, G. Mordret, K. W. Harder, F. Jirik, S. L. Pelech, Molecular cloning, expression, and characterization of the human mitogen-activated protein kinase p44erk1, Mol. Cell. Biol. 13:4679-4690, 1993. 2. N. Tang, C. Miller, K. Waddell, Approaches for the Confirmation of Putative Biomarkers Using Multiple Reaction Monitoring, HUPO Conference Proceedings, 2007. 5

www.agilent.com/chem/proteomics This item is intended for Research Use Only. Not for use in diagnostic procedures. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Agilent Technologies, Inc. 2009 Published in the U.S.A., December 23, 2009 5990-4282EN