JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

INTRODUCTION. Material and Methods

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

QA for Clinical Dosimetry with Emphasis on Clinical Trials

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

Quality assurance in external radiotherapy

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

RPC s Credentialing Programs for Clinical Trials

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Anthropomorphic breast phantoms for quality assurance and dose verification

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

Eric E. Klein, Ph.D. Chair of TG-142

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

I. Equipments for external beam radiotherapy

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS

The Journey of Cyberknife Commissioning

Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa

RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*

International Practice Accreditation

Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition

3D Conformal Radiation Therapy for Mucinous Carcinoma of the Breast

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) What is a Small Field? Indra J. Das/ Small Fields (Page-1)

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Two Week Course Curriculum Catalog

The clinical use of OSLD

Major accidents in radiotherapy

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Characteristics Analysis of High Energy External Radiotherapy Beams in Water

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers

MD 4300: Introduction to Medical Dosimetry (3semester hour) Course Syllabus for Fall 2009

Teaching linear accelerator physics using simulation software. Marco Carlone, BC Cancer Agency Nicole Harnett, Princess Margaret Cancer Centre

Spatially Fractionated GRID Therapy: A Case Study Abstract: Introduction: Case Description: Conclusion: Key Words Introduction

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

Three Dimensional Dosimetry. Session Program: Therapy Education Session Organizer: L. John Schreiner

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Follow this and additional works at: Part of the Medicine and Health Sciences Commons

BELdART-2: Moving towards safer radiotherapy. Steven Lelie 1,2, Wouter Schroeyers 1, Brigitte Reniers 1, Sonja Schreurs 1

Expectations of Physics Knowledge for Certification

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy

DANIEL A. JOHNSON. American Board of Radiology (ABR) Therapeutic Radiologic Physics, Year 2010

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Limits of Precision and Accuracy of Radiation Delivery Systems

RTOG DOSIMETRY DATA SUBMISSION

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group.

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

SRS/SBRT Errors and Causes

AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE. Geoffrey S. Ibbott, Ph.D. June 20, 2017

Radiation qualities in carbon-ion radiotherapy at NIRS/HIMAC

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Comparison of Percentage Depth Dose (PDD) and Post Time Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams Using MRI Technique

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

Electron therapy Class 3: Clinical procedures

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

RADI 6033 Physics of Radiotherapy II Spring 2017

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO MP (CLINICAL ONCOLOGY) PART I EXAMINATION - AUGUST Time : p.m p.m.

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

Small field dosimetry, an example of what a Medical Physicist does (& some more examples) A/PROF SCOTT CROWE MEDICAL PHYSICIST

Planning for the Installation of a Compact Superconducting Proton Therapy Unit in a Busy Medical Center Environment


Applications of Particle Accelerators

Transcription:

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Sang Hyun Cho and Geoffrey S. Ibbott Department of Radiation Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas 77030

ABSTRACT Over the years, the Radiological Physics Center (RPC) has acquired the basic dosimetry data for photon and electron beams during its on-site dosimetry review visits. Consequently, the RPC has become a unique place where one can study some standard behavior among institutions machines of the same model and manufacturer, in terms of their basic dosimetry data (e.g., output factor, depth dose, off-axis profile, etc.). An analysis of the data indicated that photon beams from linear accelerators of the same make, model, and energy could be represented by a single reference data set within ±2%. This observation did not always hold for older machines built before 1990, nor for machines that had been altered from the manufacturer s specifications. More investigation was necessary to demonstrate the validity of this hypothesis for machines that were not included in our initial analysis. As the first step for further investigation, we chose the Siemens Primus, one of the recent model linear accelerators. Our preliminary investigation was conducted for two important dosimetry data sets, output factor and depth dose. Four sets of randomly selected RPC-verified data were analyzed for 6 and 18 MV photon beams. It was found that all the data were within ±2% of the average, which strongly supports our hypothesis. This investigation was supported by PHS grant CA 10953 awarded by the NCI, DHHS.

INTRODUCTION Typical basic photon dosimetry data collected and verified by the Radiological Physics Center (RPC) during its on-site dosimetry review include: - Output factor - Percentage depth dose (PDD) / Tissue maximum ratio (TMR) - Off-axis factors (in air or in water) - Wedge factor

INTRODUCTION The RPC has acquired significant amounts of basic photon dosimetry data for most common medical linear accelerators. The RPC s data can be used to investigate the existence of standard behavior (or average nature) in the basic photon dosimetry data for the same make, model, and energy. An analysis of the accumulated data yielded a hypothesis that photon beams from modern linear accelerators 1 of the same make, model, and energy could be represented by a single set of reference data within 2%, in terms of their basic dosimetry data. 1 : Most computer-controlled accelerators built and released after 1990 (e.g., Varian 600C, 2100C, Siemens Primus, etc.)

INTRODUCTION Our observation did not always hold for older machines, nor for machines that had been altered from the manufacturers specifications. Our observation, at least for photon beams, seemed to support some manufacturers claim that they could match the dosimetry data between the machines with the same model and energy. More investigation was necessary to demonstrate the validity of our hypothesis for machines that were not included in our initial analysis.

METHODS As the 1st step for further investigation, the Siemens Primus, one of the recent model linear accelerators, was chosen. There were four sets of Primus data available, all of which were randomly selected and verified by the RPC during its on-site review. Our preliminary investigation was conducted for two important dosimetry data sets, output factor and percentage depth dose. Our investigation is still on-going and will include other basic dosimetry data.

RESULTS Siemens Primus 6 MV Output Factor (OF) Field Size Average OF % SD 6 x 6 0.954 0.31 10 x 10 1.000 0.00 15 x 15 1.034 0.28 20 x 20 1.056 0.32 30 x 30 1.080 0.48 No. of Data Sets = 4

RESULTS Siemens Primus 18 MV Output Factor (OF) Field Size Average OF % SD 6 x 6 0.952 0.26 10 x 10 1.000 0.00 15 x 15 1.036 0.35 20 x 20 1.056 0.34 30 x 30 1.078 0.36 No. of Data Sets = 4

RESULTS Siemens Primus 6 MV Percentage Depth Dose (PDD) 6 cm x 6 cm 10 cm x 10 cm 20 cm x 20 cm 30 cm x 30 cm PDD %SD PDD %SD PDD %SD PDD %SD 1 97.0 1.56 97.4 1.04 98.6 0.70 99.0 0.27 1.5 100.0 0.00 100.0 0.00 100.0 0.00 100.0 0.00 2 99.2 0.35 98.9 0.10 99.0 0.12 98.6 0.21 3 94.9 0.32 94.9 0.12 95.3 0.24 95.6 0.34 4 90.2 0.59 90.8 0.29 91.5 0.25 91.8 0.11 5 85.6 0.28 86.8 0.33 88.1 0.44 88.7 0.37 6 81.1 0.14 82.4 0.31 84.1 0.27 85.0 0.53 7 76.7 0.13 78.5 0.19 80.5 0.50 81.7 0.43 8 72.3 0.42 74.5 0.36 76.9 0.53 78.3 0.64 9 68.3 0.67 70.7 0.37 73.4 0.39 75.0 0.80 10 64.8 0.56 67.3 0.65 70.4 0.63 72.0 0.62 15 48.3 0.92 51.3 0.79 55.3 0.94 57.5 0.70 20 36.1 1.10 39.0 1.11 43.2 1.22 45.4 0.82 25 26.9 0.79 29.3 0.48 33.4 0.35 35.5 0.40 30 20.3 1.05 22.5 0.94 25.8 0.89 27.9 0.51 No. of Data Sets = 4

RESULTS Siemens Primus 18 MV Percentage Depth Dose (PDD) 6 cm x 6 cm 10 cm x 10 cm 20 cm x 20 cm 30 cm x 30 cm PDD %SD PDD %SD PDD %SD PDD %SD 2 95.5 0.30 96.4 1.10 99.1 1.36 100.4 1.13 3.5 100.0 0.00 100.0 0.00 100.0 0.00 100.0 0.00 4 98.9 0.07 98.7 0.00 98.3 0.36 98.0 0.72 5 95.7 0.49 95.5 0.23 95.1 0.34 94.8 0.53 6 91.7 0.15 91.9 0.08 91.7 0.39 91.6 0.70 7 87.9 0.32 88.2 0.16 88.4 0.08 88.4 0.72 8 84.3 0.42 84.8 0.08 85.2 0.08 85.3 0.66 9 80.6 0.35 81.4 0.26 82.0 0.26 82.3 0.60 10 77.2 0.41 78.1 0.17 79.0 0.35 79.4 0.25 15 61.9 0.46 63.3 0.15 65.1 0.26 65.8 0.38 20 49.6 0.45 51.2 0.16 53.6 0.38 54.5 0.38 25 39.7 0.18 41.4 0.00 43.8 0.32 45.1 0.47 30 31.9 0.44 33.6 0.21 35.9 0.00 37.1 0.38 No. of Data Sets = 4

Summary Our preliminary investigation showed that all the data were within 2% of the average. The results presented here strongly support our hypothesis. More investigation is on-going to test the validity of our hypothesis for most modern machines. If a series of well-established reference dosimetry data becomes available, physicists may perform spot-checking of their dosimetry data using reference dosimetry data, instead of performing full commissioning of the beams.