A comparison of accommodation amplitudes in pseudophakic eyes measured with three different methods

Similar documents
Comparison of ultrasound and optic biometry with respect to eye refractive errors after phacoemulsification

Review Article Real and Pseudoaccommodation in Accommodative Lenses

Postoperative refraction changes in phacoemulsification cataract surgery with implantation of different types of intraocular lens

Accuracy of Biometry for Intraocular Lens Implantation Using the New Partial Coherence Interferometer, AL-scan

Comparative Efficacy of the New Optical Biometer on Intraocular Lens Power Calculation (AL-Scan versus IOLMaster)

Predictability and accuracy of IOL formulas in high myopia

Visual performance of Acrysof ReSTOR compared with a monofocal intraocular lens following implantation in cataract surgery

Anterior movement of the crystalline lens in the process of accommodation in children

Comparison of the Effects of Bright Light, Phenylephrine 2.5%, Tropicamide 1%, and Pilocarpine 2% on Anterior Chamber Depth and IOL Power

Structural changes of the anterior chamber following cataract surgery during infancy

New supplementary intraocular lens for refractive enhancement in pseudophakic patients

Comparison of the ocular wavefront aberration between pharmacologicallyinduced and stimulus-driven accommodation

1 Cathedral Eye Clinic, University of Ulster, Academy Street, Belfast, UK.

Anterior Chamber Depth Change Following Cataract Surgery in Pseudoexfoliation Syndrome; a Preliminary Study

Prospective Study of the New Diffractive Bifocal Intraocular Lens

go the distance NEW AcrySof IQ ReSTOR +2.5 D It s a wide world. Help your patients and everywhere in between.

OPTOMETRY ORIGINAL PAPER

Immersion Vs Contact Biometery for Axial Length Measurement before Phacoemulsification

Appendix Table 1. Ophthalmic drugs approved by the US Food and Drug Administration,

INTRODUCTION J. DAWCZYNSKI, E. KOENIGSDOERFFER, R. AUGSTEN, J. STROBEL. Department of Ophthalmology, University Hospital Jena, Jena - Germany

Partial Coherence Interferometry as a Technique to Measure the Axial Length of the Eye Archived Medical Policy

Bringing astigmatism AND presbyopia into focus.

Refractive Changes after Removal of Anterior IOLs in Temporary Piggyback IOL Implantation for Congenital Cataracts

Refractive Dilemma. Challenging Case

Factors Affecting the Accuracy of Intraocular Lens Power Calculation with Lenstar

Accommodation is a dioptric increase in power of the eye

Managing residual postoperative error

Biometry and intraocular lens power calculation Alexander C. Lee, Mujtaba A. Qazi and Jay S. Pepose

Clinical Pearls. A Quick Guide to Crystalens AO Accommodating Lens

Premium Implant Options for the Cataract Patient. Justin Schweitzer, OD, FAAO Vance Thompson Vision Sioux Falls, South Dakota

Comparison of Intraocular Lens Power Calculation Methods Following Myopic Laser Refractive Surgery: New Options Using a Rotating Scheimpflug Camera

Comparison of posterior capsule opacification between fellow eyes with two types of acrylic intraocular lens

Original Article Influences of anterior capsule polishing on effective lens position after cataract surgery: a randomized controlled trial

Start with ME. LEAVE A LEGACY OF EXCELLENT OUTCOMES FOR PATIENTS WITH ASTIGMATISM. TECNIS TORIC 1-PIECE IOL

Comparison of the Long-term Clinical Results of Hydrophilic and Hydrophobic Acrylic Intraocular Lenses

Research Article Reproducibility of the Optical Biometer OA-1000 (Tomey)

Patient Selection IOL Power Calculation. Patient Selection. Biometry IOL-Power calculation using Vericalc 2.0. AC-Depth > 3.0 mm (FDA 3.

Intrascleral-fixated intraocular lenses for aphakic correction in the absence of capsular support

Efficacy and Safety of Cataract Extraction with Negative Power Intraocular Lens Implantation

AcrySof ReSTOR Multifocal versus AcrySof SA60AT Monofocal Intraocular Lenses: A Comparison of Visual Acuity and Contrast Sensitivity

Trabeculectomy is an effective method for lowering

Issue 15 The following key clinical peer reviewed journals will be reviewed: MONTHLY RESEARCH UPDATE 151(3) American Journal of Ophthalmology 129(5)

CONTROLLED PRECISION.

Visual Outcome in Senile Cataract with Multifocal and Unifocal Intraocular Lens

Electronic poster presentations

Visual Performance Results after Tetraflex Accommodating Intraocular Lens Implantation

The impact of multifocal intraocular lens in retinal imaging with optical coherence tomography

Intraocular Lens Power Estimation in Combined Phacoemulsification and Pars Plana Vitrectomy in Eyes with Epiretinal Membranes: A Case-Control Study

Advanced Eyecare of Orange County/ Kim T. Doan, M.D.

Reliability of Cycloplegic Autorefractor Measurements to determine Spherical and Astigmatic Refractive Errors in Young Children

Megalocornea is a non-progressive, uniformly

Inaccuracy of Intraocular Lens Power Prediction for Cataract Surgery in Angle-Closure Glaucoma

INFORMED CONSENT FOR CATARACT SURGERY

Accuracy of swept-source optical coherence tomography based biometry for intraocular lens power calculation: a retrospective cross sectional study

Intraoperative biometry for intraocular lens (IOL) power calculation at silicone oil removal

THE PENTACAM AXL. Improving Cataract Surgery Outcomes. Optical biometry and anterior segment tomography in one device

RayOne Trifocal IOL. The preloaded platform that performs again and again MADE IN UK

Comparison of Newer IOL Power Calculation Methods for Eyes With Previous Radial Keratotomy PATIENTS AND METHODS. Patients

Nature and Science 2016;14(9)

Myopic Shift after Implantation of a Novel Diffractive Trifocal Intraocular Lens in Korean Eyes

Laser in situ keratomileusis (LASIK) has proven to be

Eye Elongation during Accommodation in Humans: Differences between Emmetropes and Myopes

Cataract Surgery: What You Must Know Before Having It Done

Influence of fogging lenses and cycloplegia on open-field automatic refraction

The aim of refractive surgery is to modify the refractive

Accuracy of Intraocular Lens Calculation Formulas

Learn Connect Succeed. JCAHPO Regional Meetings 2017

Cataract Surgery in the Patient with a History of LASIK or PRK

CLINICAL SCIENCES. The Effect of Texturing the Intraocular Lens Edge on Postoperative Glare Symptoms

Until recently, the treatment of presbyopia

WHAT SETS ACTIVEFOCUS DESIGN APART? THE DIFFERENCE IS IN THE DISTANCE.

Learn Connect Succeed. JCAHPO Regional Meetings 2017

Changes in Corneal Astigmatism and High Order Aberrations After Clear Corneal Tunnel Phacoemulsification Guided by Corneal Topography

Comparison of Corneal Power and Intraocular Lens Power Calculation Methods after LASIK for Myopia

The visual outcome after implantation of the Multifocal Intra Ocular Lens. Dr.Bhargav Dave National Institute of Ophthalmology Pune

Xiaojian Zhong, Erping Long, Wan Chen, Wu Xiang, Zhaochuan Liu, Hui Chen, Jingjing Chen, Zhuoling Lin, Haotian Lin * and Weirong Chen *

The Change in Ocular Refractive Components After Cycloplegia in Children

Complication and Visual Outcome after Peadiatric Cataract Surgery with or Without Intra Ocular Lens Implantation

Author s Affiliation. Original Article. Comparison of Biometry in Phakic and Dense Modes. Muhammad Suhail Sarwar. Unaiza Mariam

Predicting the Postoperative Intraocular Lens Position Using Continuous Intraoperative Optical Coherence Tomography Measurements

Getting Started GUIDE. TECNIS Symfony IOL and TECNIS Symfony TORIC IOL. TECNIS SYMFONY IOL pg 1

Axial Length Shortening After Cataract Surgery: New Approach to Solve the Question

Surgical induced astigmatism in femtosecond laser assisted cataract surgery

Informed Consent For Cataract Surgery. And/Or Implantation of an Intraocular Lens INTRODUCTION

2/26/2017. Sameh Galal. M.D, FRCS Glasgow. Lecturer of Ophthalmology Research Institute of Ophthalmology

Multifocal Toric Swiss Army Knife For the Cataract and Refractive Surgeon

IOL Power Calculation for Children

Synchrony AIOL Key Features

Presbyopia is the age-related loss of accommodation the

Original article. Amon M Pediatric micro-incision cataract surgery Nepal J Ophthalmol 2011;3(5):3-8

Cataract Surgery in Patients with a Previous History of KAMRA Inlay Implantation: A Case Series

Cataract and Refractive Surgery Patients: Still Two Different Breeds?

Evaluation of Toric Intraocular Lenses in Patients With Low Degrees of Astigmatism

FEP Medical Policy Manual

Subnormal Vision in Uneventful Cataract Surgery after 6 Weeks Hospital Based Study

Biometric parameters, corneal astigmatism and ocular comorbidity in cataract surgery patients

Introduction. Nikolaos Trakos 1 Elli Ioachim 2 Elena Tsanou 2 Miltiadis Aspiotis 1 Konstantinos Psilas 1 Chris Kalogeropoulos 1 ORIGINAL RESEARCH

Comparison of two multifocal IOL types short and medium-term visual outcomes

Artiflex Toric Phakic Intraocular Lens Implantation in Congenital Nystagmus

Transcription:

(2008) 22, 65 69 & 2008 Nature Publishing Group All rights reserved 0950-222X/08 $30.00 www.nature.com/eye A comparison of accommodation amplitudes in pseudophakic eyes measured with three different methods G Nemeth 1, A Tsorbatzoglou 1, P Vamosi 2, Z Sohajda 3 and A Berta 1 CLINICAL STUDY Abstract Purpose To compare the accommodative amplitudes with three different methods in pseudophakic eyes with different types of intraocular lenses (IOLs). Methods Fifty-one pseudophakic eyes of 44 patients (age: 72.0278.53 years) were studied. One of two different types of IOL were implanted (N ¼ 30, three-piece Alcon s Acrysof s MA60AC and N ¼ 21, one-piece Alcon s Acrysof s SA60AT) in-the-bag after standard phacoemulsification. The time of the examinations was 13.8577.35 months postoperatively. We measured the amplitude of accommodation with three different methods: (1) subjective minus-lenses-to-blur method; (2) a new optical device (ACMaster s, Carl Zeiss, Jena, Germany) using partial coherence interference (PCI) technique under physiological stimulus; and (3) objective anterior chamber depth measuring with a standard A-scan ultrasonic device (Ultrascan Imaging System s, Alcon Laboratories, Forth Worth, TX, USA) before and after pharmacological relaxation of ciliary muscle. Results We measured 0.8370.63 D amplitude of accommodation with subjective minus-lenses-to-blur method. The IOL position did not change significantly during physiological accommodation effort measured with PCI method ( 0.02670.134 mm). The change in the IOLs position between near fixating and after ciliary muscle relaxation was 0.1870.28 mm measured with ultrasound. There were no significant differences between values of one-piece and three-piece IOL groups. Conclusion The amplitude of accommodation measured by subjective and objective methods are different and are not comparable with each other. We did not observe any difference between values of examined two types of IOLs. (2008) 22, 65 69; doi:10.1038/sj.eye.6702519; published online 14 July 2006 Keywords: pseudophakic accommodation; amplitude of accommodation; partial coherence interfence method Introduction Pseudophakic accommodation is the change in the refractive power of the eye associated with either movement and/or tilt of an intraocular lens (IOL). Pseudophakic pseudoaccommodation refers to factors associated with good near visual acuity in patients implanted with IOLs, and can be explained by one or the combination of the following factors: minus cylinder, higher order corneal multifocality, and/or small pupil size resulting in increased depth of field. 1 5 Forward movement of the implanted IOL induced by accommodative effort of the ciliary muscle has also been reported. 6 10 In the recent study, we measured and assessed the amplitude of accommodation in pseudophakic eyes implanted with one of two different types of IOLs with three different methods, one subjective and two objective methods. Patients and methods Fifty-one pseudophakic eyes of 44 patients (age: 72.0278.53 years, range: 54 84 years) were studied prospectively. Exclusion criteria were more than 1.0 D of corneal astigmatism or any 1 Ophthalmology, Medical Health and Science Center, University of Debrecen, Debrecen, 2 Ophthalmology, Szent Rokus Hospital, Budapest, 3 Ophthalmology, Kenezy Gyula Hospital, Debrecen, Correspondence: G Nemeth, Ophthalmology, University of Debrecen, Nagyerdei Blvd. 98, Debrecen 4012, Tel: þ 36 52415816; Fax: þ 36 52415816. E-mail: nemeth222@ yahoo.com Received: 14 March 2006 Accepted in revised form: 8 June 2006 Published online: 14 July 2006 Authors have no proprietary interests or conflict of interests in this study

66 A comparison of accommodation amplitude in pseudophakic eyes eye pathology affecting complete visual acuity. Standard phacoemulsifications were carried out after clear corneal incision. One of two different types of monofocal intraocular lenses (IOLs) were implanted in-the-bag. The one-piece Acrysof s SA60AT (Alcon Laboratories, Forth Worth, TX, USA) has open-loop design (group 1: N ¼ 21) without haptic angulation, the three-piece Acrysof s MA60AC (Alcon Laboratories, Forth Worth, TX, USA) has haptic angulation of 101 (group 2: N ¼ 30). After a mean of 13.8577.35 months of operations (group 1: 7.0574.71 months; group 2: 17.0176.44 months), we measured the amplitude of accommodation of pseudophakic eyes using three approaches in the same visit. The examinations were performed without knowing the results of the other ones. The chronological order of the three methods are: subjective minus-lensesto-blur method and objective measurement of the anterior chamber depth (ACD) with two techniques detailed below. 1% cyclopentolate hydrochloride, administered three times at 15-min intervals. All patients were seated, with their heads vertical and their eyes level during the measurements to ensure that IOL movements were only produced by ciliary muscle activity. ACD measurements were performed by contact A-scan (Ultrascan Imaging System s, Alcon Laboratories, Forth Worth, TX, USA). The average of 10 10 measurements was presented as the ACD value. ACD shift was calculated by subtracting ACD under cyclopentolate from ACD during near fixation, corresponding to maximal ciliary relaxation and physiological ciliary contraction. Statistical analysis was performed using SPSS 11.0 software, the data were described as mean7sd. The differences in values between methods were recorded with the unpaired test of Wilcoxon and a P-value of 0.05 was considered as the level of significance. The association between values were described with Spearman s correlation of rho (r). Method 1: Subjective minus-lenses-to-blur The patients were asked to look at the illuminated, distant visual acuity chart number of the 20/20 line with the observed eye. With distance correction, minus diopter (D) lenses were added in front of the observed eye in 0.25 D steps. The patients reported when the 20/20 line could no longer be held in sharp focus. The maximum minus lens power added over the distance correction with which the patients can see the number in 20/20 line was recorded as the amplitude of accommodation. Method 2: ACD measurement with a new device working with partial coherence interference technique The ACD was measured with ACMaster s (Carl Zeiss, Jena, Germany) with built-in, movable target for physiologic near and distance fixation. All of the patients were seated and were forced to fixate to the target at near and at distance too. The device measured the fixating eye, so avoiding interference from convergence movement. The mechanism of the partial coherence interference (PCI) method has been described in details elsewhere. 11 13 A mean of five measurements were recorded with distance correction followed by 3.0 D added to distance correction for near fixating. The ACD shift was calculated subtracting the two mean values. Method 3: ACD measuring in near fixation followed by pharmacologically induced maximal ciliary muscle relaxation The ACD was measured during fixation with the contralateral eye at 30 cm and after instillation of Results The best-corrected visual acuity was 0.8870.2 in a decimal scale. The mean of distance correction was 0.9270.91 D. It was 0.9370.99 D in the one-piece group and was 0.7770.82 D in the three-piece group. The difference was not significant between these two groups (P ¼ 0.44). The axial length was 22.5470.89 mm. It was 22.6870.81 mm in the one-piece group and was 22.4670.96 mm in the three-piece group. There were no statistically significant difference in axial length between the two groups (P ¼ 0.68). Subjective minus-lenses-to-blur The largest minus lens with which the 20/20 line was just readable by the patient was 0.8370.63 D. It was 0.9570.67 D (range: 2.5 0.0 D) in the one-piece group and was 0.8170.53 D (range: 2.0 0.0 D) in the threepiece group. The difference between the two groups was not significant (P ¼ 0.4). ACD measurement with a new device working with PCI technique ACD shift measured under physiological conditions was 0.02670.134 mm. It was 0.04370.193 mm (range: 1.038 to 0.298 mm) in the one-piece group and was 0.01470.079 mm (range: 1.063 to 0.385 mm) in the three-piece group. The difference between the one-piece and three-piece group was not significant (P ¼ 0.46). The data are summarized in Table 1.

A comparison of accommodation amplitude in pseudophakic eyes 67 Table 1 Amplitude of accommodation measured with a subjective method and with a PCI device under physiological conditions and the movement of the IOLs measured with A-scan ultrasound before and after cycloplegia ACD measurement with A-scan ultrasound in near fixation followed by pharmacologically induced maximal ciliary muscle relaxation ACD shift was 0.1870.28 mm. It was 0.1670.3 mm (range: 0.51 to 0.2 mm) in the one-piece group and was 0.270.28 mm (range: 0.49 to 0.1 mm) in the threepiece group. The difference between the one-piece and three-piece groups was not significant (P ¼ 0.68). There was no significant statistical correlation between pharmacologically induced ACD shift values and the diopter value of minus-lenses-to-blur method (Spearman s rho (r) ¼ 0.16, P ¼ 0.37). Discussion Minus-lensesto-blur (D) ACD shift measured physiologically with PCI using physiological stimulus (mm) ACD shift measured with ultrasound (between before and after cycloplegia) (mm) One-piece IOL 0.9570.67 0.04370.193 0.1670.3 Three-piece IOL 0.8170.53 0.01470.079 0.270.28 All 0.8370.63 0.02670.134 0.1870.28 ACD ¼ anterior chamber depth; IOL ¼ intraocular lense; PCI ¼ partial coherence interference. Positive values represent forward, and negative values represent backward movement of IOLs. Certain patients with monofocal IOL can achieve good visual acuity for distance and for near with distance correction. This phenomenon is previously known as apparent accommodation. 14 The amplitude of apparent accommodation varies significantly from patient to patient. Previous studies confirmed that the amplitude of apparent accommodation is correlated with depth of focus, 15,16 pupil diameter, 17 20 and degree of corneal astigmatism. 21 25 There are many studies about the forward movement of various types of IOLs during accommodation effort. 6 10 However, the backward movement is possible with accommodative (AT-45 Crystalens) 26 or with other types of IOLs. 9 The phenomenon based on movement of an IOL is referred as pseudophakic accommodation. The common principle of the true pseudophakic accommodation is the ciliary muscle constrictionmediated movement of the IOL along the axis of the eye. Many former studies have measured accommodative amplitude subjectively and objectively in pseudophakic eyes. 27 30 The subjective methods depend on patient cooperation and are a sum of the amplitude of pseudophakic accommodation and pseudoaccommodation. The objective methods with different ciliary muscle medications (pilocarpin or cyclopentolate) can overestimate the actual IOL movement as it is shown possible in the recent study too. Measuring the pseudophakic accommodation and pseudoaccommodation is an interesting and challenging task: what method should be used for measuring accommodative amplitude of the pseudophakic eyes and which method is the most accurate? Langenbucher et al 31 divide the possible techniques for measuring pseudophakic accommodation into groups and subgroups. According to their study, there is static, dynamic, as well as objective and subjective methods. In our study, we chose a dynamic, subjective method, the minus-lenses-to-blur method, and two static, objective, indirect methods, measuring the ACD with standard A-scan ultrasound and with ACMaster s under desaccommodative and accommodative condition with and without eye drop-induced ciliary muscle relaxation. However, subjective methods do not report differences between true pseudophakic accommodation and pseudoaccommodation. Our data about the cycloplegia-induced-acd shift is comparable with the values of literature in case of both types of IOLs. 6,8 10 According to Nawa et al, 32 it is an equivalent of approximately 0.3 0.35 D of pseudophakic accomodation at an axial length of 22.5470.89 mm. It is about a third of that measured with subjective minuslenses-to-blur method. So the detected ACD shifts were too small for itself to explain the actual, subjective accommodation amplitude. Objective methods is based on measuring the ACD thereby the position of the IOL in the two marginal accommodative conditions. The ACD can be measured with different ophthalmological techniques, such as ultrasound biometry, high-resolution magnetic resonance imaging, ultrasound biomicroscopy, Scheimpflug imaging, anterior segment optical coherence tomography, and partial coherence interferometry. 31,33,34 We used objective measurements with a newly developed high accuracy interferometric device. Almost every method for measuring pseudophakic accommodation and pseudoaccommodation has considerable potential for evaluating error. 35 Therefore, measuring amplitude of accommodation in pseudophakic eyes poses some problems. Inducing accommodation with miotic eye drops causes strong pupillary miosis, so measuring through small pupils are difficult with some methods including PCI. Pupil dilation without causing cycloplegia (phenylephrine) induces greater high-order aberrations, 36 with

68 A comparison of accommodation amplitude in pseudophakic eyes cycloplegic drugs (cyclopentolate) measurements are relatively easy, but the ability of the patients to deliver a normal accommodative response may decrease. Besides all of these, the eye drop-induced ciliary muscle constriction or relaxation shows only a maximum possibility of IOL movement, but does not represent the actual, physiologically produced pseudophakic accommodation. This major disparity is well demonstrated in our study. The differences between the values of our three methods measuring pseudophakic accommodation are rather significant. Physiologically (PCI) there was no IOL movement neither in the one-piece nor in the three-piece group, with pharmacologically relaxed ciliary muscle, there was a mean of 0.3 0.35 D enforced pseudophakic accommodation. With subjective method, a mean of 0.83 D was measured as total accommodation amplitude. It is an aggregate of pseudoaccommodation and pseudophakic accommodation amplitudes, and means that approximately 0.5 D is probably achieved by the combination of the above-mentioned and other factors, such as patients motivation and potential of visual perception. Our two IOLs have different configurations: the Alcon s Acrysof s MA60AC IOL has 101 angulation between haptic and optic, the Alcon s Acrysof s SA60AT IOL has an open-loop design without haptic angulation. Despite this fact, we did not observe any difference between diopter values of accommodation amplitude measured with any methods. In conclusion, we demonstrated that different measurement methods of amplitude of accommodation with two types of monofocal IOLs show significant variance and are not comparable with each other. References 1 Huber C. Myopic astigmatism: a substitute for accommodation in pseudophakia. Doc Ophthalmol 1981; 52: 123 178. 2 Trindade F, Oliveira A, Frasson M. Benefit of against-therule astigmatism to uncorrected near acuity. J Cataract Refract Surg 1997; 23: 82 85. 3 Nakazawa M, Ohtsuki K. Apparent accommodation in pseudophakic eyes after implantation posterior chamber intraocular lenses: optical analysis. Invest Ophthalmol Vis Sci 1984; 25: 1458 1460. 4 Fukuyama M, Oshika T, Amano S, Yoshitomi F. Relationship between apparent accommodation and corneal multifocality in pseudophakic eyes. Ophthalmology 1999; 106: 1178 1181. 5 Oshika T, Mimura T, Tanaka S, Amano S, Fukuyama M, Yoshitomi F et al. Apparent accommodation and corneal wavefront aberration in pseudophakic eyes. Invest Ophthalmol Vis Sci 2002; 43: 2882 2886. 6 Hardman Lea SJ, Rubinstein MP, Haworth SM. Pseudophakic accommodation? A study of the stability of capsular bag supported, one piece, rigid tripod, or seft flexible implants. Br J Ophthalmol 1990; 74: 22 25. 7 Gonzalez F, Capeans C, Santos L, Suarez J, Cadarso L. Anteroposterior shift in rigid and soft implants supported by the intraocular capsular bag. Graefe s Arch Clin Exp Ophthalmol 1992; 230: 237 239. 8 Lesiewska-Junk H, Kaluzny J. Intraocular lens movement and accommodation in eyes of young patients. J Cataract Refract Surg 2000; 26: 562 565. 9 Findl O, Kiss B, Petternel V, Menapace R, Georgopoulos M, Rainer G et al. Intraocular lens movement caused by ciliary muscle contraction. J Cataract Refract Surg 2003; 29: 669 676. 10 Findl O, Kriechbaum K, Menapace R, Koeppl C, Sacu S, Wirtitsch M et al. Laserinterferometric assessment of pilocarpine-induced movement of an accommodating intraocular lens. Ophthalmology 2004; 111: 1515 1521. 11 Fercher AF, Mengedoth K, Werner W. length measurement by interferometry with partially coherent light. Opt Lett 1988; 13: 186 188. 12 Hitzenberger CK, Baumgartner A, Drexler W, Fercher AF. Interferometric measurement of corneal thickness with micrometer precision. Am J Ophthalmol 1994; 118: 468 476. 13 Drexler W, Baumgartner A, Findl O, Hitzenberger CK, Sattmann H, Fercher AF. Submicrometer precision biometry of the anterior segment of the human eye. Invest Ophthalmol Vis Sci 1997; 38: 1304 1313. 14 Betman JW. Apparent accommodation in aphakic eyes. Am J Ophthalmol 1950; 33: 921 928. 15 Elder MJ, Murphy C, Sanderson GF. Apparent accommodation and depth of field in pseudophakia. J Cataract Refract Surg 1996; 22: 615 619. 16 Niessen AGJE, de Jong LB, van der Heijde GL. Pseudoaccommodation in pseudophakia. Eur J Implant Refract Surg 1992; 4: 91 94. 17 Nakazawa M, Ohtsuki K. Apparent accommodation in pseudophakic eyes after implantation of posterior chamber intraocular lenses. Am J Ophthalmol 1983; 96: 435 438. 18 Nakazawa M, Ohtsuki K. Apparent accommodation in pseudophakic eyes after implantation of posterior chamber intraocular lenses: optical analysis. Invest Ophthalmol Vis Sci 1984; 25: 1458 1460. 19 Ravalico G, Baccara F. Apparent accommodation in pseudophakic eyes. Acta Ophthalmol 1990; 68: 604 606. 20 Elder MJ, Murphy C, Sanderson GF. Apparent accommodation and depth of field in pseudophakia. J Cataract Refract Surg 1996; 22: 615 619. 21 Huber C. Planned myopic astigmatism as a substitute for accommodation in pseudophakia. J Am Intraoc Implant Soc 1981; 7: 244 249. 22 Sawusch MR, Guyton DL. Optimal astigmatism to enhance depth of focus after cataract surgery. Ophthalmology 1991; 98: 1025 1029. 23 Datiles MB, Gancayco T. Low myopia with low astigmatism correction gives cataract surgery patients good depth of focus. Ophthalmology 1990; 97: 922 926. 24 Hillman JS, Bradbury JA. Apparent accommodation by myopic astigmatism with monofocal intraocular lenses. Eur J Implant Refract Surg 1990; 2: 101 104. 25 Huber C. Myopic astigmatism as a substitute for accommodation in pseudophakia. Dev Ophthalmol 1981; 5: 17 26.

A comparison of accommodation amplitude in pseudophakic eyes 69 26 Koeppl C, Findl O, Menapace R, Kriechbaum K, Wirtitsch M, Buehl W et al. Pilocarpine-induced shift of an accommodating intraocular lens: AT-45. Crystalens J Cataract Refract Surg 2005; 31: 1290 1297. 27 Rosenfield M, Portello JK, Blustein GH, Jang C. Comparison of clinical techniques to assess the near accommodative response. Optom Vis Sci 1996; 73: 382 388. 28 Rosenfield M, Cohen AS. Repeatability of clinical measurements of the amplitude of accommodation. Ophthalmic Physiol Opt 1996; 16: 247 249. 29 Kragha IKOK. Amplitude of accommodation: population and methodological differences. Ophthalmic Physiol Opt 1986; 6: 75 80. 30 Rutstein RP, Fuhr PD, Swiatocha J. Comparing the amplitude of accommodation determined objectively and subjectively. Optom Vis Sci 1993; 70: 496 500. 31 Langenbucher A, Huber S, Nguyen NX, Seitz B, Gusek-Schneider GC, Kuchle M. Measurement of accommodation after implantation of an accommodating posterior chamber intraocular lens. J Cataract Refract Surg 2003; 29: 677 685. 32 Nawa Y, Ueda T, Nakatsuka M, Tsuji H, Marutani H, Hara Y et al. Accommodation obtained per 1.0 mm forward movement of a posterior chamber intraocular lens. J Cataract Refract Surg 2003; 29: 2069 2072. 33 Findl O, Drexler W, Menapace R, Bobr B, Bittermann S, Vass C et al. Accurate determination of effective lens position and lens-capsule distance with 4 intraocular lenses. J Cataract Refract Surg 1998; 24: 1094 1098. 34 Findl O, Drexler W, Menapace R, Hitzenberger CK, Fercher AF. High precision biometry of pseudophakic eyes using partial coherence interferometry. J Cataract Refract Surg 1998; 24: 1087 1093. 35 Whitefoot H, Charman WN. Dynamic retinoscopy and accommodation. Ophthalmic Physiol Opt 1992; 12: 8 17. 36 Carkeet A, Velaedan S, Tan YK, Lee DY, Tan DT. Higher order ocular aberrations after cycloplegic and noncycloplegic pupil dilation. J Refract Surg 2003; 19: 3106 3322.