A RAPID AND SENSITIVE UPLC/UV/MS METHOD FOR SIMVASTATIN AND LOVA S TAT IN IN SU P P O RT O F C L E A NING VA L I DAT IO N S T U DIES

Similar documents
Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Matrix Factor Determination with the Waters Regulated Bioanalysis System Solution

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

Analysis of the Non-Ionic Surfactant Triton-X Using UltraPerformance Convergence Chromatography (UPC 2 ) with MS and UV Detection

Profiling Flavonoid Isomers in Highly Complex Citrus Juice Samples Using UPLC Ion Mobility Time-of-Flight Mass Spectrometry

John Haselden 1, Gordon Dear 1, Jennifer H. Granger 2, and Robert S. Plumb 2. 1GlaxoSmithKline, Ware, UK; 2 Waters Corporation, Milford, MA, USA

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

[ application note note ] ]

[ APPLICATION NOTE ] APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

INTRODUCTION CH 3 CH CH 3 3. C 37 H 48 N 6 O 5 S 2, molecular weight Figure 1. The Xevo QTof MS System.

Reducing Sample Volume and Increasing Sensitivity for the Quantification of Human Insulin and 5 Analogs in Human Plasma Using ionkey/ms

Simultaneous Analysis of Intact Human Insulin and Five Analogs in Human Plasma Using μelution SPE and a CORTECS UPLC Column

Method Development for the Analysis of Endogenous Steroids Using Convergence Chromatography with Mass Spectrometric Detection

A High Sensitivity UPLC/MS/MS Method for the Analysis of Clopidogrel and Clopidogrel Carboxylic Acid Metabolite in Human K 2 EDTA Plasma

Identification & Confirmation of Structurally Related Degradation Products of Simvastatin

Meeting Challenging Requirements for the Quantitation of Regulated Growth Promoters Dexamethasone and Betamethasone in Liver and Milk

Comparison of a UPLC Method across Multiple UHPLC Systems

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

USP Method Transfer of Ziprasidone HCl from HPLC to UPLC

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

Challenges in Developing an Ultra-Sensitive Bioanalytical Method for Ethinylestradiol in Human Plasma

Lipid Class Separation Using UPC 2 /MS

Rapid Analysis of Bisphenols A, B, and E in Baby Food and Infant Formula Using ACQUITY UPLC with the Xevo TQD

Performance of an ultra low elution volume 96-well plate

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Analysis of Food Sugars in Various Matrices Using UPLC with Refractive Index (RI) Detection

PosterREPRINT RAPID, SELECTIVE SCREENING OF URINE SAMPLES FOR GLUCURONIDES BY LC/MS/MS INTRODUCTION ABSTRACT

[ APPLICATION NOTE ] Profiling Mono and Disaccharides in Milk and Infant Formula Using the ACQUITY Arc System and ACQUITY QDa Detector

Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS

Fast Separation of Triacylglycerols in Oils using UltraPerformance Convergence Chromatography (UPC 2 )

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

[Application Note] TOXICOLOGY SCREENING BY UPLC/PDA IN COMBINATION WITH AN EXTENSIVE COMPOUND LIBRARY

Impurity Profiling of Carbamazepine by HPLC/UV

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

SEPARATION OF BRANCHED PFOS ISOMERS BY UPLC WITH MS/MS DETECTION

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

The Raptor HILIC-Si Column

Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System

Sample Concentration and Analysis of Human Hormones in Drinking Water

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

[ APPLICATION NOTE ] The Separation of 8 -THC, 9 -THC, and Their Enantiomers by UPC 2 Using Trefoil Chiral Columns INTRODUCTION APPLICATION BENEFITS

Jose Castro-Perez, Henry Shion, Kate Yu, John Shockcor, Emma Marsden-Edwards, Jeff Goshawk Waters Corporation, Milford, MA, U.S. and Manchester, UK

Abstract. Introduction

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

A UPLC/MS Approach for the Diagnosis of Inborn Errors of Metabolism

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis

The Development of LC/MS Methods for Determination of Polar Drugs of Abuse in Biological Samples

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

2D-LC as an Automated Desalting Tool for MSD Analysis

Pelagia Research Library

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

LC-MS/MS Method for the Determination of Tenofovir from Plasma

The Application of UPLC/MS E for the Analysis of Bile Acids in Biological Fluids

Analysis of N-Linked Glycans from Coagulation Factor IX, Recombinant and Plasma Derived, Using HILIC UPLC/FLR/QTof MS

Sanjog Ramdharane 1, Dr. Vinay Gaitonde 2

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT.

Journal of Chemical and Pharmaceutical Research

Quantification of Budesonide Using UPLC and Xevo TQ-S

Impurity Identification using a Quadrupole - Time of Flight Mass Spectrometer QTOF

A Metabolomics Approach to Profile Novel Chemical Markers for Identification and Authentification of Terminalia Species

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract

Analysis of Pesticides (II) Metribuzin & their metabolites in Rice Jun Yonekubo, Nihon Waters, JAPAN

Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

[ APPLICATION NOTE ] UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research INTRODUCTION APPLICATION BENEFITS

SYNAPT G2-S High Definition MS (HDMS) System

Development and Validation of a Polysorbate 20 Assay in a Therapeutic Antibody Formulation by RP-HPLC and Charged Aerosol Detector (CAD)

LCMS-8030 Application Report. Steroids: Testosterone, Progesterone, Estradiol, Ethinylestradiol

10 mg in 1.4 ml. XTerra MS C 18, 4.6 x 50 mm. Volume injected: Injector Column. Waste. Pump. Pump. Splitter

Analysis of drugs of abuse in biological matrix using Time of Flight technology

Journal of Chemical and Pharmaceutical Research

RP-HPLC Method Development and Validation of Abacavir Sulphate in Bulk and Tablet Dosage Form

PosterREPRINT SIMULTANEOUS QUANTIFICATION OF PSYCHOTHERAPEUTIC DRUGS IN HUMAN PLASMA BY TANDEM MASS SPECTROMETRY

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Determination of Clarithromycin in Human Plasma by LC-EI Tandem Mass Spectrometry: Application to Bioequivalence Study

Rapid Quantitative Analysis of Cannabidiol from Consumer Products Using UltraPerformance Convergence Chromatography

The Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Direct Analysis of Urinary Opioids and Metabolites by Mixed-Mode µelution SPE Combined with UPLC/MS/MS for Forensic Toxicology

[APPLICATION NOTE] DETERMINATION OF AMINO ACIDS IN BEERS USING THE UPLC AMINO ACID ANALYSIS SOLUTION

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System

Supporting information

John P. McCauley and Rui Chen Waters Corporation, Milford, MA, USA INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEY WORDS

Automated Lipid Identification Using UPLC/HDMS E in Combination with SimLipid

Transcription:

RPID ND SENSITIVE PLC/V/MS METD FR SIMVSTTIN ND LV S TT IN IN S P P RT F C L E NING V L I DT I N S T DIES Erin Chambers, Diane Diehl, and Jeff Mazzeo Waters Corporation, Milford, M, S BCKGRND tilizing CQITY PLC Technology with V and MS detection enhances the selectivity, sensitivity, and through-put in cleaning validation studies. This results in improved methods, more confidence in results, and significant cost savings. To highlight this utility, we developed an analytical method in support of a cleaning validation study for simvastatin and lovastatin, two commonly prescribed cholesterol-lowering drugs in the statin class. is newly off patent and has been the focus of several generic companies. For this reason, an analytical method is needed to support cleaning validation studies in a manufacturing environment. NLYTICL CLLENGES Pharmaceutical manufacturing equipment is cleaned after production to avoid cross-contamination in subsequent batches of a different product. Effectiveness of the cleaning process needs to be confirmed by analytical measurements. Therefore, a cleaning validation method must be developed which provides evidence that cleaning processes applied to the equipment are sufficient to remove residues of bulk drug to predetermined safety levels. These methods ensure that subsequent batches of other products are not contaminated by previously manufactured products, or by the cleaning process itself. Cleaning validation methods need to achieve limits of detection which are low enough to detect small quantities of residual drug and they must demonstrate adequate resolution of drug product from interferences from solvents and the swabs or wipes used to sample the equipment surfaces. In addition, the method must be shown to be linear over the range required by the assay. 3 C 3 C C 3 C 3 3 C MW 418.27 Figure 1. Structures of simvastatin and lovastatin. and lovastatin, whose structures are shown in Figure 1 are 3-hydroxy-3-methylglutaryl-coenzyme reductase inhibitors, which significantly impact the synthesis of cholesterol precursors. Drugs of this class represent the most efficient treatment of hypercholesterolemia. is a natural product, while simvastatin is semi-synthetic, being manufactured from lovastatin. s a result, lovastatin and simvastatin need to be detected in the same analytical method. EXPERIMENTL Click on Part Numbers for more information screening approach was utilized to choose the best starting chromatographic conditions. Four different column chemistries were screened at high and low p using two organic modifiers. The resulting chromatograms were evaluated for peak shape, retention, and selectivity. From these data, the column chemistry, p, and mobile phase modifier were chosen. The screening gradient was later optimized to decrease run time if possible, and to improve resolution of simvastatin and lovastatin from interferences extracted from swabs or wipes. 3 C 3 C C 3 MW 44.25 C 3

LC Conditions for Screening LC System: CQITY PLC with Column Manager, CQITY PLC PD and SQ Columns: CQITY PLC BE C 18 2.1 x 5 mm, 1.7 µm Column Temp: Flow Rate: Part Number: 186235 CQITY PLC BE Shield RP18 2.1 x 5 mm, 1.7 µm Part Number:1862853 CQITY PLC BE Phenyl 2.1 x 5 mm, 1.7 µm Part Number:1862884 CQITY PLC SS T3 2.1 x 5 mm, 1.8 µm Part Number:1863538 3 C 6 µl/min Mobile Phase 1:.1 C in 2 (~p 2.7) Mobile Phase 2:.1 N 4 in 2 (~p 11) Mobile Phase B1: Mobile Phase B2: Gradient: Wavelength: MS Conditions MS System: Ionization Mode: Capillary Voltage: Cone Voltage: Desolvation Temp: Desolvation Gas: Source Temp: Me CN 5-95 B in 2 min, hold for.5 min; return to initial conditions 248 nm Waters SQ Detector ESI Positive 3 V 43 V 35 C 55 L/r 12 C SIR Channels: m/z 419.25, Extraction Procedure m/z 45.25 RESLTS ND DISCSSIN Chromatographic Screening The chromatograms resulting from p screening with CN as the organic modifier are shown in Figure 2. n initial PD scan was used from 21 to 3 nm in order to determine the optimal wavelength of 248 nm for detection of simvastatin and lovastatin. ll columns were run at low p (). The C 18, Shield RP18, and Phenyl columns were also run at high p (B). (Note: the SS T3 column is a silica particle and cannot be run at high p.) Chromatograms were evaluated with respect to peak shape, resolution of impurities or degradants from the compounds of interest, and selectivity. Retention of these neutral compounds is relatively unaffected by p. owever, at low p we begin to resolve additional drug compound-related peaks from simvastatin and lovastatin. The slightly broader peak shape observed on some columns at high p is due to coelution of these smaller peaks with the compounds of interest. We chose low p mobile phase due to better peak shape and improved resolution of impurities. We next screened all column chemistries using two organic modifiers. The chromatograms resulting from organic modifier screening are shown in Figure 3. ll columns were run at low p with Me () and CN (B). It is clear that Me is not a selective enough elution solvent to resolve these hydrophobic compounds. and lovastatin elute as a single broad peak when Me is used as the organic modifier. In contrast, the two compounds chromatograph as well-resolved, narrow peaks using CN. The final step involved choosing the optimal column chemistry. Examination of the low p, CN chromatograms indicates that only the SS T3 column is able to adequately resolve all of the impurity peaks from simvastatin and lovastatin. We optimized the chromtographic method using the SS T3 column at low p, with CN as the organic modifier. Cotton swabs (TexWipe ) were sonicated for 45 min in 1 ml of 75:25 CN: 2, or 1 CN. The extract was removed and analyzed for the presence of interferences.

. CN, low p B. CN, high p 1.3.36.38 1.94.55 1.8.5 1.83 2.5 2.4 2.15 C 18 SS T3 RP18 Phenyl Figure 2. PD chromatograms at 248 nm resulting from p screening with CN as the organic modifier. 1.e-2-1. e-2 5.e-3-5. e-3 1.e-2-1. e-2.35.69 2.35 2.5.73 2.43 2.5.57 2.18 2.5 4.e-2 4.e-2. B. Me, low p C 18 SS T3 RP18.33 1.26.33.33 1 1.92 Not applicable 1.8 1.12 1.81.3.36.38 1.4 1.76 1.82 1.94.55 CN, low p 1.8 2.2 2.4 2.5 2.15 Extraction Recovery Recoveries for simvastatin and lovastatin in both the 1 CN and 75:25 CN: 2 extracts were > 95. This was determined by spiking the swabs prior to extraction with known quantities of active drug. Peak area in the extract was then compared to peak area in standards prepared in the identical organic solvent. Limit of Detection (LD) and Linear Dynamic Range The LDs for simvastatin and lovastatin were determined under optimized chromatographic conditions with both V and MS detection. The LD obtained from data acquired using the PD operating at 248 nm was shown to be 5 ng/ml for both analytes (Figure 4). The LD for mass spectrometry operating in SIR mode was 1 ng/ml for each analyte (Figure 4B). The dynamic range of the assay was 5 ng/ml to 5 µg/ml by PD at 248 nm and 1 ng/ml to 1 µg/ml using mass spectrometry detection. Standard curves were linear with 1/x 2 weighting over this range with R 2 values from.983 to.998. ll data points met the criteria for accuracy with all non-ld points having CVs of less than ± 15, and all LD points having CV values of less than ± 2. Sample standard curves for simvastatin and lovastatin using both PD and mass spectrometry detection are shown in Figure 5..57 2.5 Figure 3. PD chromatograms at 248 nm resulting from organic modifier screening at low p. Chromatographic ptimization 2.27 Phenyl.5 1.83. 6.5e-3 6.e-3 5.5e-3 5.e-3 4.5e-3 4.e-3 3.5e-3 3.e-3 2.5e-3 2.e-3 1.5e-3 1.e-3.2 5 ng/ml.39 5.e-4 5.1.2.3.4.5.6.7.8.9 1. 1.1 1.2 1.3 1.4 1.5 1.6 1.7.94 1.29 5 ng/ml 1.47 and lovastatin are very hydrophobic compounds as indicated by their late elution time under reversed-phase LC conditions. The generic screening gradient was modified to range from 5 to 95 CN to better reflect the nature of the analytes, and to shorten the chromatographic run time. se of this shallower gradient also allows us to improve resolution from any additional impurities or interferences discovered during analysis of swab samples. In addition, formic acid was added to the CN (mobile phase B) at a concentration of 75 to reduce V baseline drift. B. 1 1.31.46.47.56.57.68.73.77 1. 1.5.91.94 1.12 1.18.1.2.3.4.5.6.7.8.9 1. 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1 5.14.24.17.19.29.34.4.45.53.63.68.7.73.1.2.3.4.5.6.7.8.9 1. 1.1 1.2 1.3 1.4 1.5 1.6 1.7 Figure 4. Limits of detection were determined for V () and mass spectrometry (B.).79 1 ng/ml.84.85 1 ng/ml.85.99 1.5.88 1.15.93 1.2 1.1 1.25 1.25 1.19 1.25 1.33 1.32 1.32 1.37 1.37 1.41 1.39 1.48 1.45 1.5 1.51 1.5 1.54 1.56 1.6 1.63 1.61 1.62 1.67 1.78 1.75 1.73 1.73 1.79

Selectivity The method was shown to be selective for the analytes of interest. Swabs were extracted as above, the extract removed, and then analyzed to confirm lack of interferences from swabs and extraction solvents. Figures 6 and 7 demonstrate that both analytes are resolved from interferences when either 1 CN or 75:25 CN: 2 are used as swab extraction solvents. Figure 6 demonstrates minimal interferences using V detection, while figure 7 demonstrates a much higher degree of selectivity for the analyte using mass spectrometry.. : PD 248 nm Standard curve is linear from 5 ng/ml to 5 µg/ml 1/x 2 weighting, r 2 =.997, CV from -.4 to 8 6 B. : SQ in SIR mode Standard curve is linear, 1 ng/ml to 1 µg/ml 1/x 2 weighting, r 2 =.99, CV from -.2 to -15.7 75 5 25 ng/ml 5 1 15 2 25 3 35 4 45 5 4 2 Conc 5 1 15 2 25 3 35 4 45 5 55 6 65 7 75 8 85 9 951 : PD 248 nm Standard curve is linear from 5 ng/ml to 5 µg/ml 1/x 2 weighting, r 2 =.998, CV from.4 to 6.7 : SQ in SIR mode Standard curve is linear from 1 ng/ml to 1 µg/ml 1/x 2 weighting, r 2 =.983, CV -.6 to 18 1 5 4 2 ng/ml 5 1 15 2 25 3 35 4 45 5 Conc 5 1 15 2 25 3 35 4 45 5 55 6 65 7 75 8 85 9 95 1 Figure 5. Standard curves for simvastatin (top) and lovastatin (bottom) using V detection at 248 nm () or mass spectrometry in SI R mode (B). 7.e-2 Swab extracted with CN 1: SIR of 2 Channels ES+ TIC 8.4e6 6.e-2 5.e-2 5ng/mL (1) lovastin and (2) simvastatin Swab extracted with 75:25 CN: 2 4.e-2 Swab extracted with 75:25 CN: 2 5 ng/ml (1) lovastatin and (2) simvastatin 1.63 3.e-2 Swab extracted with CN Minimial interference No interference 1.e-2 1.29 1.47.2 1.68-1.e-2.2.4.6.8 1. 1.2 1.4 1.6 1.8 2..1.2.3.4.5.6.7.8.9 1. 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2. Figure 6. V swab interferences. Figure 7. Mass spectrometry swab interferences.

Conclusion We have successfully employed a chromatographic screening strategy to facilitate fast development of a rapid, selective and sensitive LC/MS/V analytical method to support cleaning validation studies. The method achieves LDs with V detection that meet the requirements for most cleaning validation protocols. Single quadrupole mass spectrometry yields lower limits of detection, suitable for the testing of manufacturing equipment used to prepare even very potent drugs. The increased resolution afforded by PLC technology allows us to separate the two analytes of interest from swab and solvent interferences in under 3 minutes. dditional selectivity is realized with mass spectrometric analysis. This rapid analysis time enables manufacturing plants to move to the next production batch faster, thus increasing their productivity and efficiency. In addition, increased simplicity and cost savings are achieved by having a single method for multiple analytes. References dditional information relating to the use of CQITY PLC systems for cleaning validation studies can be found in our document entitled CQITY PLC System: Better utilizing assets in a small laboratory environment with PLC, literature code 722328EN. Waters, The Science of What s Possible, CQITY PLC and PLC are trademarks of Waters Corporation. ll other trademarks are the property of their respective owners. 27 Waters Corporation. ctober 27 722337EN VW-PDF Waters Corporation 34 Maple Street Milford, M 1757.S.. T: 1 58 478 2 F: 1 58 872 199 www.waters.com