SIMULATIONS OF A MODEL-BASED FUZZY CONTROL SYSTEM FOR GLYCEMIC CONTROL IN DIABETES

Similar documents
INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY

Active Insulin Infusion Using Fuzzy-Based Closed-loop Control

Non Linear Control of Glycaemia in Type 1 Diabetic Patients

Back Stepping SMC for Blood Glucose Control of Type-1 Diabetes Mellitus Patients

Insulin Control System for Diabetic Patients by Using Adaptive Controller

Optimal Model Based Control for Blood Glucose Insulin System Using Continuous Glucose Monitoring

Achieving Open-loop Insulin Delivery using ITM Designed for T1DM Patients

Causal Modeling of the Glucose-Insulin System in Type-I Diabetic Patients J. Fernandez, N. Aguilar, R. Fernandez de Canete, J. C.

IJESRT INTERNATIONAL JOURNAL OF ENGINEERING SCIENCES & RESEARCH TECHNOLOGY PID CONTROLLER FOR BLOOD GLUCOSE CONTROL Ajmal M M *, C R Srinivasan *

Outline. Model Development GLUCOSIM. Conventional Feedback and Model-Based Control of Blood Glucose Level in Type-I Diabetes Mellitus

[Frida Svendsen and Jennifer Southern] University of Oxford

An event-based point of view on the control of insulin-dependent diabetes

IDENTIFICATION OF LINEAR DYNAMIC MODELS FOR TYPE 1 DIABETES: A SIMULATION STUDY

Report Documentation Page

Insulin Administration for People with Type 1 diabetes

VIRTUAL PATIENTS DERIVED FROM THE

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes. pump. pump

Outline Insulin-Glucose Dynamics a la Deterministic models Biomath Summer School and Workshop 2008 Denmark

Implementation of Inference Engine in Adaptive Neuro Fuzzy Inference System to Predict and Control the Sugar Level in Diabetic Patient

A Composite Model of Glucagon Glucose Dynamics for In Silico Testing of Bihormonal Glucose Controllers

Modeling of Glucose-Insulin System Dynamics in Diabetic Goettingen Minipigs

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes Individuals: With type 1 diabetes

An Improved Fuzzy PI Controller for Type 1 Diabetes

Fayrouz Allam Tabbin Institute for Metallurgical Studies, Helwan, Egypt

Artificial Pancreas Device System (APDS)

Toward Plug and Play Medical Cyber-Physical Systems (Part 2)

Proposed Clinical Application for Tuning Fuzzy Logic Controller of Artificial Pancreas utilizing a Personalization Factor

A Closed-loop Artificial Pancreas based on MPC: human-friendly identification and automatic meal disturbance rejection

Agent-based Simulation for Blood Glucose Control in Diabetic Patients

Impact of sensing and infusion site dependent dynamics on insulin bolus based meal compensation

ADVANCES in NATURAL and APPLIED SCIENCES

Type-2 fuzzy control of a fed-batch fermentation reactor

A mathematical model of glucose-insulin interaction

Developing an Artificial Pancreas The History & Future of Dose Safety

A K Patra 1, R K Samantaray 2, (DR) J K Maharana 3*

Diabetes Care 33: , 2010

A Dynamic Model with Structured Recurrent Neural Network to Predict Glucose-Insulin Regulation of Type 1 Diabetes Mellitus

Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic Clamp Data

1. Continuous Glucose Monitoring

Proceedings of the 5th WSEAS International Conference on Telecommunications and Informatics, Istanbul, Turkey, May 27-29, 2006 (pp )

The Artificial Pancreas

Optimal Blood Glucose Regulation using Single Network Adaptive Critics

The oral meal or oral glucose tolerance test. Original Article Two-Hour Seven-Sample Oral Glucose Tolerance Test and Meal Protocol

The Realities of Technology in Type 1 Diabetes

Supplementary Appendix

Simulation Study on Type I Diabetic Patient

A Practical Approach to Prescribe The Amount of Used Insulin of Diabetic Patients

What is the role of insulin pumps in the modern day care of patients with Type 1 diabetes?

A Closed-Loop Artificial Pancreas Using Model Predictive Control and a Sliding Meal Size Estimator. Abstract. Introduction

Alternative insulin delivery systems: how demanding should the patient be?

Modeling Type One Diabetes with a Linear Response Function Based Model

SYSTEM MODELING AND OPTIMIZATION. Proceedings of the 21st IFIP TC7 Conference July 21st 25th, 2003 Sophia Antipolis, France

Multivariate Statistical Analysis to Detect Insulin Infusion Set Failure

Hypoglycemia a barrier to normoglycemia Are long acting analogues and pumps the answer to the barrier??

Research Article An Improved PID Algorithm Based on Insulin-on-Board Estimate for Blood Glucose Control with Type 1 Diabetes

Model Reference Adaptive Control of Glucose in Type 1 Diabetics: A Simulation Study

Automatic adaptation of basal therapy for Type 1 diabetic patients: a Run-to-Run approach

Pharmacokinetics Overview

Chapter 3: Linear & Non-Linear Interaction Models

The artificial pancreas: the next step in connectivity and digital treatment of type 1 diabetes

Electronic Supplementary Material to the article entitled Altered pattern of the

A PERSONALISED APPROACH TO INSULIN REGULATION USING BRAIN-INSPIRED NEURAL SEMANTIC MEMORY IN DIABETIC GLUCOSE CONTROL

Peter Beck Design of a mobile, safety-critical in-hospital glucose management system

Smith Predictor Sliding Mode Closed-loop Glucose Controller in Type 1 Diabetes

A switching based PID technique for blood glucose control.

Report Documentation Page

Glucose Concentration Simulation for Closed-Loop Treatment in Type 1 Diabetes

DESIGN AND CONTROL OF THE ARTIFICAL PANCREAS

Analysis of glucose-insulin-glucagon interaction models.

DiasNet an Internet Tool for Communication and Education in Diabetes

MOGA-based Multi-drug Optimisation for Cancer Chemotherapy

Population-Based Incremental Learning Algorithm for Identification of Blood Glucose Dynamics Model for Type-1 Diabetic Patients

Modellering av blodsukkerdynamikk

DIAGNOSIS OF DIABETES NOW WHAT?

Simulation Study on Closed Loop Control Algorithm of Type 1 Diabetes Mellitus Patients

Identification, state estimation, and adaptive control of type i diabetic patients

DIABETES - FACT SHEET

Insulin Pump Therapy in children. Prof. Abdulmoein Al-Agha, FRCPCH(UK)

Design of the Glucose Rate Increase Detector: A Meal Detection Module for the Health Monitoring System

Glycemia Prediction in Critically Ill Patients Using an Adaptive Modeling Approach

Counting the Carbs, Fat and Protein in Type 1 Diabetes Translating the Research into Clinical Practice

Control of the Blood Glucose Level in Diabetic Patient Using Predictive Controller and Delay Differential Equation

UvA-DARE (Digital Academic Repository) The artificial pancreas Kropff, J. Link to publication

A Telematic System for Diabetes Management, Reporting and Patient Advice

Symposium Paper. A Critical Assessment of Algorithms and Challenges in the Development of a Closed-Loop Artificial Pancreas. B. WAYNE BEQUETTE, Ph.D.

Linear Quadratic Control Problem in Biomedical Engineering

The Chemostat: Stability at Steady States. Chapter 5: Linear & Non-Linear Interaction Models. So, in dimensional form, α 1 > 1 corresponds to

Subject Index. Breastfeeding, self-monitoring of blood glucose 56

INSPIRED BY GIVE HIM MORE. The MiniMed 670G system is the world s first insulin delivery system that automatically adapts to your child s needs.

Observer-Based State Feedback for Enhanced Insulin Control of Type I Diabetic Patients

INSPIRED BY GIVE HER MORE. The MiniMed 670G system is the world s first insulin delivery system that automatically adapts to your child s needs.

Dynamic Modeling of Exercise Effects on Plasma Glucose and Insulin Levels

Physiological Closed-loop Controllers for MDCPS

Insulin Pumps - External

Comparison of Mamdani and Sugeno Fuzzy Interference Systems for the Breast Cancer Risk

Mean Glucose Slope Principal Component Analysis Classification to Detect Insulin Infusion Set Failure

Agus Kartono, Egha Sabila Putri, Ardian Arif Setiawan, Heriyanto Syafutra and Tony Sumaryada

Tutorial ADAPT Case study 1. Data sampling / error model. Yared Paalvast Yvonne Rozendaal

Transcription:

Bulletin of the Transilvania University of Braşov Vol. 8 (57) No. 2-2015 Series I: Engineering Sciences SIMULATIONS OF A MODEL-BASED FUZZY CONTROL SYSTEM FOR GLYCEMIC CONTROL IN DIABETES C. BOLDIȘOR 1 S. COMAN 1 Abstract: This paper investigates a conventional fuzzy control algorithm to be used in a closed loop system that should return and keep the blood glucose concentration at normal values for diabetic patients. The controller design follows general guidelines in fuzzy control theory. Simulations are done using a model of glucose kinetics under the influence of insulin injection rates and exogenous glucose disturbance. The model is an extension to the well-known Bergman s minimal model by adding the carbohydrates absorption rate from Hovorka s model as the glucose disturbance input. Key words: glycaemia control, fuzzy control of glucose concentration, Bergman minimal model, insulin to glucose model. 1. Introduction Several control algorithms for automatic blood glucose control in type 1 diabetes mellitus (T1DM) patients were tested over the years, in simulations or in live experiments. Detailed surveys of the most used algorithms in this subject during the latest 10 years are presented in [19], [20] and [14]. Also, very good summaries of the most important studies and results published in recent years are presented in [14] and [10], while the newest are presented in details in [6]. Two different approaches can be noted. The first is the model-based system design, which implies the use of a model to describe the glucose dynamics in various cases. The first example would be the modified Bergman minimal model [3], [8], for T1DM patients, which was intensively used in simulations [4]. Many studies presented solutions based on a model, and a good part of them stepped ahead towards testing the simulated control algorithms in live experiments. The second approach focuses on system design methods and algorithms that use real data from live experiments and physicians experience. Researchers argues that usually there is an important variability in physiological parameters from one patient to another and, even worse, for a single patient during a longer period of time. A blood glucose control system is intended to become a routine therapy solution and during 24 hours there are several influences over the parameters: having meals, doing exercises, night changes in metabolism, and others [18]. Physicians experience and actions can be expressed with better relevance in an approximate logic. Treatment decisions are based on evaluating blood glucose levels in 1 Dept. of Automation and Information Technology, Transilvania University of Braşov.

94 Bulletin of the Transilvania University of Braşov Series I Vol. 8 (57) No. 2-2015 terms of too low, good or too high, and not on strict computations. This suggests that a fuzzy logic (FL) approach is worth to be investigated (at leas. Another type of arguments for fuzzy logic control (FLC) comes from its theory: it is often mentioned that fuzzy control is less sensible to small variations of parameters and, if extended to an adaptive system, it can efficiently cope larger variations too. On the other hand, fuzzy control theory doesn t offer strict design methods. In many cases, the controller s rule-base is set by following a set of design guidelines, handbook solutions or other examples that the engineer knows. Fewer solutions are obtained from recorded data or by translating expert s decisions into control rules. Hence, it seems that designing a fuzzy control system based on physicians treatment actions and experience still has enough motivation. Convincing studies presenting FLC systems for glycaemia control are significantly fewer than other control algorithms. The review published by Youssef et al. in 2009 [20] refers to the FLC as being promising, but largely clinically untested. Lunze et al. published a very good state-of-the-art article in 2013 [14] which do not even mention the FLC solution. One year later, Doyle et al. [10] present a summary of clinical trial protocols which shows the FLC as the less preferred solution. However, encouraging studies and results were reported, even if most of them are simulations. Campos-Delgado et al. [5] presented a two-loop FLC for long-time glycaemic control. Their solution uses two Mamdanitype fuzzy controllers (FC): one in the inner loop to adjust the quantities of injected insulin for three shots before meals, and one in the outer loop to adjust the maximum quantities to be provided to patient in a time-scale of days (as a supervisory support system). The system is mostly meant to incorporate medical knowledge about the treatment of T1DM, and not to offer a real control system in the most conventional meaning. A simple solution for direct control of glycaemia is tested in [7], and compared to PID control. The input and output variables (glucose concentration, its rate of change and insulin dosage) are described by four terms each, and the rule base is based mostly on design guidelines. Mauseth et al. [15] present a conventional table-based Mamdani-type FC which respects most guidelines in designing fuzzy controllers. They also introduce a personalization factor proportional to the patient s total daily dose of insulin. This is actually a scaling gain applied to the controller s output, which has strong influence on the system s performance. These solutions are mostly designed based on control engineering experience, especially in fuzzy control theory. In opposite, Atlas, Nimri and their team [1], [16] proposed a system which applies the fuzzy logic theory to imitate lines of reasoning of diabetes caregivers, and so to focus on medical experience. Finally, for the latest studies on glycaemic fuzzy control, please refer to the reviews in [17] and [6], while for an introduction in FLC theory, design guidelines and examples please refer to the book of Jantzen [12]. 2. A Model of the Insulin to Glucose System, with Meal Glucose Disturbance Probably the most used model to describe the effect of insulin injection rate and glucose disturbance input on the blood glucose concentration is the modified Bergman s minimal model. In its initial form, the model described the kinetics of glucose and insulin concentrations in blood during the intravenous glucose tolerance test [3]. A few years later, Cobelli et al. [8] adapted the model to describe the kinetics

Boldişor, C., et al.: Simulations of a Model-Based Fuzzy Control System for Glycemic 95 of glucose concentration under the effects of exogenous insulin and glucose rates, with application in glycaemic control. Over the years, the modification of Bergman s model for T1DM patients became known as the minimal model. The model is a compartment-based 3-rd order nonlinear one, depicted as follows: dg( p1 X ( G( p1g dt dx( p2x ( p3i(, dt di( ni( bu(. dt b ad(, (1) All notations are described in Table 1 (Detailed descriptions of each are presented in [9] and other values than those used in this paper are listed in [13]). The input variables are the exogenous insulin rate, u(, and the disturbance glucose absorption, d(, while the output is the blood glucose concentration, G(. The conversion factors a and b allow input variables to be described as flows (in terms of [mg/min] and [mu/min], rather than [mg/min/dl], respectively [mu/min/l] as would be require in the equations). Note that, by using these factors, the units would be in consistency with clinical convention of insulin delivery rate prescription. The variables and parameters of the model Table 1 Notation Description (see also [9]) Unit Values G( blood glucose concentration at time t [min] mg/dl - I( blood insulin concentration at time t [min] mu/l - X( a variable defining the effect on the insulin-excitable tissues at time t [min] (proportional to the insulin concentration in a remote compartmen 1/min - d( disturbance input of glucose at time t [min] mg/min - U( exogenous insulin deliver rate at time t [min] mu/min - G b subject s basal glucose concentration in blood mg/dl 81 P 1 the insulin-independent rate of tissue glucose uptake (i.e. glucose effectiveness) 1/min 0 P 2 the active insulin decrease rate (decrease of uptake) 1/min 0.025 P 3 insulin-dependent increase of tissue glucose uptake L/(mU min min) 13 10-6 n decay rate of blood insulin 1/min 0.0926 V the assumed distribution volume of the insulin [14], which in some studies is considered the total blood volume [2] L 12 a glucose input conversion factor 1/L 0.0083 b insulin input conversion factor 1/L 0.0833 D g the quantity of carbohydrates (CHO) in the meal mg 50,000 A g t g,max CHO bioavailability - a factor describing how much of the meal s CHO is absorbed in the blood the approximate time delay from the start of the meal when CHO absorption is at the maximum unitless 0.8 min 40

96 Bulletin of the Transilvania University of Braşov Series I Vol. 8 (57) No. 2-2015 In the example presented in [2], both factors depend on the total blood volume, V, and based on the units we considered, they should be a = 0.1/V and b = 1/V. The remote compartment in the model refers to the glucose uptake activity in the insulin-excitable tissues (as the liver and peripheral cells). The reason for this compartment is that the glucose uptake depends on the insulin concentration at effector site, X(, not directly on the insulin concentration in the total blood volume, I(. The variable has more meaning in compartment-based modelling, than in a medical context. The glucose disturbance input may refer to the injection of glucose in intravenous glucose tolerance tests [3], [14], or to the glucose increase due to carbohydrates absorption after a meal [2]. Based on these, two simulation scenarios can be set: A) The patient has an initial high concentration of glucose, G 0. The control objectives are: to reduce the glucose concentration to its basal value, G b, in less than 3 hours, and to avoid any overshoot (which in medical terms means hypoglycaemia). B) Initially, the patient has a safe glucose concentration, G 0 = G b, and he has a meal with 50 g CHO. The control objectives are to return to a normal value (under 160 mg/dl for T1DM patients) in less than 90 minutes, and also to avoid the overshoot. To run more credible simulations with meal glucose disturbance, we considered an extension to the minimal model by including a sub-model to describe the glucose absorption after having a meal. This is presented by Hovorka et al. in [11], and consists in the following non-linear Equation (notations are also described in Table 1): Dg Agt exp( t / tg, max ) d(. (2) t 2 g,max The time variation of the disturbance input according to this model is depicted in Figure 3a and the accumulation of glucose caused by this disturbance is in Figure 3b. 3. The Fuzzy Control Algorithm The fuzzy controller used here is a Mamdani proportional-derivative (PD) type (see the diagram in Figure 1), having the control error and its derivative as the input variables and the insulin rate to be delivered as the output. Note that negative values for the insulin deliver rates are not possible, so the controller s output is limited to zero to avoid simulation errors. The scaling gains for the input variables are not analysed in this paper (g e = g ce = 1) as they have a smaller influence on the control performances. However, to obtain a faster reaction to large error values, the scaling gain of the output variable was set to g u = 2. This value could be personalized as in [15]. Further research on finding the optimal value is still to be done. The control error is defined as: e( [ G ( G( ] G(, r G b with G r ( being the reference value of glucose concentration at time t, which in our case is set to be constant and equal to Fig. 1. The fuzzy controller diagram

Boldişor, C., et al.: Simulations of a Model-Based Fuzzy Control System for Glycemic 97 The numerical rule-base of the fuzzy controller Table 2 Error change in error ce 1 = 0.5 ce 2 = 0 ce 3 = 0.5 e 1 = 0 0.1 0 0 e 2 = 40 0.2 0.1 0 e 3 = 80 0.3 0.2 0.1 e 4 = 120 0.5 0.4 0.3 e 5 = 160 0.7 0.6 0.3 e 6 = 200 0.9 0.8 0.6 e 7 = 240 1 1 0.8 the basal value of glucose concentration in blood: G r ( = G b. The fuzzy inference mechanism is a Mamdani numerical table-based one [12], with the rules described in Table 2. At each iteration of the control algorithm the insulin dose is calculated as a weighted average of u ij values in rule-table: I J u t ) i 1 j 1 ( w ij u ij, (3) with the weights w ij being inversely proportional to the distance between the (e i ; ce j ) point and the currently measured pair (e(); ce(): w ij ( ei 1 ei )( ce j 1 ce j ) 1. ( e( e )( ce( ce ) i When the values of error and change in error are e( = e i or ce( = ce j (for any j), then the weights are calculated as: w ij ( ce j 1 ce 1 ( ce( ce or respectively as: j j ), ) j w ij ( ei 1 ei ) 1. ( e( e ) i When e( = e i and ce( = ce j the fuzzy controller s output is u( = u ij. 4. Simulations Results and Discussion To evaluate the controller s performance, a Matlab-Simulink model and the Matlab s Fuzzy Logic Toolbox were used. The two situations mentioned in previous section (A and B) were simulated. The results are depicted in Figures 2 and 3. For the B case, the glucose accumulation in blood due to the carbohydrates absorption for a diabetic patient is depicted in Figure 4. In order to have a more realistic control, it was considered that the imaginary insulin deliver device adjusts the dosage every 5 minutes (Note the shapes of insulin dosage in Figures 2b and 3b.). It is very unlikely to have a device that could change the injection rate in a time range of seconds. In terms of control engineering, this behaviour is described as a zero-order hold element, and it also allows a reasonable sampling time. In both cases, the glucose concentration returned to its basal value in the expected time range, and without crossing this value.

98 Bulletin of the Transilvania University of Braşov Series I Vol. 8 (57) No. 2-2015 Fig. 2. Glycemic control results: a) the blood glucose concentration in time from a high initial value to normal and b) the insulin injection rate (adjusted every 5 minutes) that produce this result 5. Conclusions When speaking about designing a system to control the blood glucose concentration ( artificial pancreas ), the caregivers experience in T1DM treatment is the first to be considered. In most situations, their actions can be expresses in terms of fuzzy logic, which means that a fuzzy control algorithm gains enough motivation to be investigated. In fuzzy control theory, most applications follow a set of design guidelines that assures certain performance and an easy implementation. In this paper, we proved the feasibility of using a fuzzy logic based simple solution to control the blood glucose concentration in the Bergman s minimal model. Fig. 3. Glycaemia control results: a) the blood glucose concentration in time after having a meal with 50 g of carbohydrates, and b) the insulin injection rate (adjusted every 5 minutes) that produce this result The simulations of the system we presented suggest that fuzzy controller can assure control performances similar to those obtained by using more complex algorithms, as the intensively used model predictive control. The controller assures a realistic return to normal values for the blood glucose concentration. More important, any overshoot value, which in this case means possible hypoglycaemia, was avoided. The design procedure was quite easy and clear, and algorithm is easy to implement on any numerical device. The controller was obtained mainly from the experience in fuzzy control, but also following what it can be considered realistic results, as those described in today s literature.

Boldişor, C., et al.: Simulations of a Model-Based Fuzzy Control System for Glycemic 99 Fig. 4. The glucose disturbance after having a meal: a) CHO absorption rate in time; b) the raise of glucose concentration for a T1DM patient from its initial value Acknowledgements We hereby acknowledge the structural founds project PRO-DD (POS-CCE, 0.2.2.1., ID 123, SMIS 2637, ctr. No. 11/2009) for providing the infrastructure used in this work. References 1. Atlas, E., et al.: MD-Logic Artificial Pancreas System. A Pilot Study in Adults with Type 1 Diabetes. In: Diabetes Care 33 (2010) No. 5, p. 1072-1076. 2. Bequette, B.W.: Process Control: Modeling, Design, and Simulations. Prentice Hall PTR, 2002. 3. Bergman, R.N., Phillips, L.S., Cobelli, C.: Physiologic Evaluation of Factors Controlling Glucose Tolerance in Man. Measurement of Insulin Sensitivity and -Cell Glucose Sensitivity from the Response to Intravenous Glucose. In: J. of Clinical Investigations 68 (1981), p. 1456-1467. 4. Bergman, R.N.: Minimal Model: Perspective from 2005. In: Hormone Research in Paediatrics 64 (2005) Suppl. 3, p. 8-15. 5. Campos-Delgado, D.U., et al.: Fuzzy- Based Controller for Glucose Regulation in Type-1 Diabetic Patients by Subcutaneous Route. In: IEEE Trans. on Biomedical Engineering 53 (2006) No. 11, p. 2201-2210. 6. Caudal, A., et al.: Closing the Loop. In: Diabetes Technology & Therapeutics 17 (2015) No. 1, p. S27- S38. 7. Chen, J., et al.: Continuous Drug Infusion for Diabetes Therapy: A Closed Loop Control System Design. In: EURASIP J. on Wireless Communications and Networking, 2008, Art. ID 495185. 8. Cobelli, C., Toffolo, G., Ferrannini, E.: A Model of Glucose Kinetics and their Control by Insulin, Compartmental and Noncompartmental Approaches. In: Mathematical Biosciences 72 (1984), p. 291-315. 9. De Gaetano, A., Arino, O.: Mathematical Modelling of the Intravenous Glucose Tolerance Test. In: J. of Mathematical Biology 40 (2000), p. 136-168. 10. Doyle, F.J., Huyett, L.M., Lee, J.B., et al.: Closed-Loop Artificial Pancreas Systems: Engineering the Algorhitms. In: Diabetes Care 37 (2014), p. 1191-1197. 11. Hovorka, R., et al.: Nonlinear Model Predictive Control of Glucose Concentration in Subjects with Type 1

100 Bulletin of the Transilvania University of Braşov Series I Vol. 8 (57) No. 2-2015 Diabetes. In: Physiological Measurement 25 (2004), p. 905-920. 12. Jantzen, J.: Foundations of Fuzzy Control. Wiley, UK, 2007. 13. Kaveh, P., Shtessel, Y.: High Order Sliding Mode Control for Blood Control Regulation. In: Proc. of the 2006 Int. Workshop on Variable Structure Systems, Mon A.3, Alghero, Italy, 2006, p. 11-16. 14. Lunze, K., et al.: Blood Glucose Control Algorithms for Type-1 Diabetic Patients: A Methodological Review. In: Biomedical Signal Processing and Control 8 (2012) No. 2, p. 107-119. 15. Mauseth, R., et al.: Use of a "Fuzzy Logic" Controller in a Closed-Loop Artificial Pancreas. In: Diabetes Technology & Therapeutics 15 (2013) No. 8, p. 628-633. 16. Nimri, R., et al.: Night Glucose Control with MD-Logic Artificial Pancreas in Home Setting: A Single Blind, Randomized Crossover Trial- Interim Analysis. In: Pediatric Diabetes 15 (2014) No. 2, p. 91-99. 17. Nimri, R., Phillip, M.: Artificial Pancreas: Fuzzy Logic and Control of Glycemia. In: Current Opinion in Endocrinology, Diabetes and Obesity 21 (2014) No. 4, p. 251-256. 18. Phillip, M., et al.: Nocturnal Glucose Control with an Artificial Pancreas at a Diabetes Camp. In: The New England Journal of Medicine 368 (2013), p. 824-833. 19. Takahashi, D., Yang, X., Hu, F.: A Survey of Insulin-Dependent Diabetes- Part II: Control Methods. In: Int. J. of Telemedicine and Applications (2008), Article ID 739385, 14 pages. 20. Youssef, J. El, Castle, J., Ward, K.W.: A Review of Closed-Loop Algorithms for Glycaemic Control in the Treatment of Type 1 Diabetes. In: Algorithms 2 (2009) No. 1, p. 518-532.