GC/MS Identification of Flavor and Fragrance Allergens in Some Common Snack Foods Using an Agilent J&W DB-17ms Capillary GC Column

Similar documents
Determination of Allergens in Fragrance Products Using Agilent Deconvolution Reporting Software Application Brief

Independent Column Temperature Control Using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds

using the Agilent 7696A Sample Prep

Analysis of Organic Acids and Alcohols Using the Agilent J&W DB-624UI Ultra Inert GC Column

Analyze Drugs of Abuse with Agilent J&W Ultimate Plus Tubing in an Inert Flow Path

GC-µECD Analysis and Confirmation of Contract Laboratory Protocol Pesticides in Olive Oil

Organophosphorus Pesticides Analysis Using an Agilent J&W DB-5ms Ultra Inert Capillary GC Column. Application Note. Environmental. Abstract.

Forensic Analysis of Blood Alcohol Concentration

Analysis of Omega 3 and Omega 6 FAMEs in Fish Oil and Animal Fat Using an Agilent J&W DB-FATWAX Ultra Inert GC Column

Rapid Separation of Fatty Acid Methyl Esters

Rapid Analysis of 37 FAMEs with the Agilent 8860 Gas Chromatograph

Toxicology Screening of Whole Blood Extracts Using GC/Triple Quadrupole/MS

Analyze Dietary Fatty Acids, Sterols, and Lignans with an Agilent J&W DB-5ms UI Column

Application. Authors. Abstract. Introduction. Forensics

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Rapid and Robust Detection of THC and Its Metabolites in Blood

Automated Sample Preparation for FAME Analysis in Edible Oils Using an Agilent 7696A Sample Prep WorkBench

Highly Repeatable Ultra Low Detection of Estradiol Using Triple Quadrupole GC/MS in NCI Mode

Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Organophosphorus Residues in Olive Oil by GC/FPD with Agilent J&W DB-35ms Ultra Inert

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Page 222 Surface analysis of permanent wave processing hair using DART-MS

A comparison study of the analysis of volatile organic acids and fatty acids

Column Selection for the Analysis of Fatty Acid Methyl Esters Application

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Material Safety Data Sheet

Improving the Analysis of 37 Fatty Acid Methyl Esters

Detection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System

METHOD 8316 ACRYLAMIDE, ACRYLONITRILE AND ACROLEIN BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)

Determination of Gamma-Hydroxy-Butyrate (GHB) in Biological Samples

Application. Detection of Cannabinoids in Oral Fluid Using Inert Source GC/MS. Introduction. Authors. Abstract. Forensic Toxicology

High-Resolution Analysis of Intact Triglycerides by Reversed Phase HPLC Using the Agilent 1290 Infinity LC UHPLC System

Modified QuEChERS for HILIC LC/MS/MS Analysis of Nicotine and Its Metabolites in Fish

Material Safety Data Sheet

A Rapid Method for Detection of Drugs of Abuse in Blood Samples Using the Thermal Separation Probe and the 5975T LTM GC/MS

CORESTA Recommended Method No. 84

Application Note. Authors. Abstract. Food

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

Determination of Adulterants in Olive Oil Analysis of Waxes and Fatty Acid Methyl and Ethyl Esters with Capillary GC and Agilent OpenLAB Software

Toxicological dossier

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Authors. Abstract. Introduction. Environmental

Fig.1. Denatonium benzoate (DB) chemical structure

Analysis of short chain organic acids in tobacco and in some flavored e-liquids

Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS

Title Revision n date

Application Note. Author. Abstract. Introduction. Food Safety

Material Safety Data Sheet Black Soap (Hard)

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Rapidly Analyze a Wide Range of Glycol Ethers by GC-MS Using the New Rxi -1301Sil MS Column

Determination of Bath Salts (Pyrovalerone Analogs) in Biological Samples

New GCMS Applications

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

3-Acetyldeoxynivalenol. 15-Acetyldeoxynivalenol

Official Journal of the European Union

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

Toxicological dossier

Improved Characterization of Perfumes with GC GC-TOFMS

Quantitative Determination of Terpenes in Cannabis Using Headspace Solid Phase Microextraction and GC/MS

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

Author. Introduction. Small Molecule Pharmaceuticals & Generics

Analysis of several common. organic acids in tobacco leaf, snus, and moist snuff

Determination of red blood cell fatty acid profiles in clinical research

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

State of knowledge on abiotic hapten formation (hydrolysis) using examples of fragrance ingredients and state of the art on the technical management

S.C. Moldoveanu, A.G. Hudson, A. Harrison. R.J. Reynolds Tobacco Co.

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods

Application Note. Abstract. Authors. Pharmaceutical

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Author. Introduction. Abstract

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

DETERMINATION OF COMPOSITION OF TRIACYLGLYCEROLS AND COMPOSITION AND CONTENT OF DI-ACYLGLYCEROLS BY CAPILLARY GAS CHROMATOGRAPHY, IN VEGETABLE OILS

Scent Perfique Limited

Extraction of 11-nor-9-carboxy-tetrahydrocannabinol from Hydrolyzed Urine by ISOLUTE. SLE+ Prior to GC/MS Analysis

Direct Analysis of Folic Acid in Digestive Juices by LC/TOF-MS Application

The challenging extraction of non-polar contaminants out of a non-polar vegetable oil sample

Lipidomic Analysis by UPLC-QTOF MS

Interested in conducting your own webinar?

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

Time-of-Flight LC/MS Identification and Confirmation of a Kairomone in Daphnia magna Cultured Medium. Application. Authors. Abstract.

Chromatography Vacuum Ultraviolet Spectroscopy

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

EU Priority PAH Analysis in Pumpkin Seed Oil Using Bond Elut EMR Lipid Cleanup by GC/MS/MS

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

A MODIFICATION OF GAS CHROMATOGRAPHY METHOD FOR THE DETERMINATION OF FATTY ACID COMPOSITION OF MILK FAT

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

Analysis of mycotoxins in food matrices using the Agilent Ultivo Triple Quadrupole LC/MS

Supplementary Information

HPLC Analysis with Fluorescence Detection of Chlorophyll Degradation Products Pheophytins and Pyropheophytin in Virgin Olive Oil

Product Guide for LudgerSep TM ur2 UHPLC Column for DMB Sialic Acid Analysis

E17 ETHYLCELLULOSE. Revision 3 Stage 4

Separation of 37 Fatty Acid Methyl Esters Utilizing a High-Efficiency 10 m Capillary GC Column with Optimization in Three Carrier Gases

Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinElmer Altus UPLC System with PDA Detection

CORESTA RECOMMENDED METHOD NÄ 9

Innovations in Large Volume Injection

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Transcription:

GC/MS Identification of Flavor and Fragrance Allergens in Some Common Snack Foods Using an Agilent J&W DB-17ms Capillary GC Column Application Note Food Authors Doris Smith and Ken Lynam Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19809-1610 USA Abstract The European Union (EU) regulates 26 flavor and fragrance allergens. Some of these allergens are flavoring additives in common snack food such as candy and gum. Cinnamon related allergens are of particular interest, encompassing five of the 26 EU listed allergens. Cinnamon flavored gum and candy MTBE extracts show appreciable levels of these listed allergens. All 24 of the GC/MS amenable allergens are resolved and identified on an Agilent J&W DB-17ms column.

Introduction The European Union (EU) regulates 26 flavor and fragrance allergens (1). Some of these allergens are flavoring additives in common snack food such as candy and gum. Cinnamon related allergens are of particular interest, encompassing five of the 26 EU listed allergens. Cinnamon flavoring can approach percent levels in some of these intensely flavored food products. These concentrations are well beyond acceptable levels of rinse off or leave limits for topical products such as shampoos, creams and perfumes. Acceptable levels of the flavoring agents are still under debate. When used as a flavoring, cinnamon is typically included under the broad label of natural or artificial flavor with an upper limit percentage descriptor. The target limit of listed allergens for rinse off products is 100 ppm. The leave on limit is 10 ppm in fragrance and cosmetic products. In this application, these limits were reference points for candy and gum products where ingestion is the chief route of exposure rather than inhalation or adsorption routes typical for fragrance and cosmetic products. Detection at these levels is readily obtainable by single quad GC/MS for the purposes of identification and semiquantitative evaluation. GC/MS is an effective technique for 24 of the 26 compounds listed as EU flavor and fragrance allergens. Oak and tree moss extracts are also listed allergens but are not suitable for GC/MS analysis. Previous studies of this allergen set by GC/MS focused primarily on fragrance and cosmetic products. Several approaches have been described, one using a deconvolution reporting software (DRS) approach on a HP-5ms column and one using a two-dimensional column approach employing a low thermal mass (LTM) device with Deans switch heart cutting [2,3]. The focus of this application is a single column GC/MS approach to separate the allergens in flavored candy and gum products using the selectivity offered by a DB-17ms. The selectivity of the midpolarity DB17ms column is a useful tool in helping to resolve the allergens away from potential matrix interference. Experimental This analysis was done on an Agilent J&W DB-17ms 30 m 0.25 mm, 0.25 µm column on an Agilent 7890/5975C GC/MS system. Details of the chromatographic conditions are shown in Table 1. Details of the flow path supplies necessary for the analysis are listed in Table 2. Table 1. Chromatographic Conditions for EU Allergen Analysis GC/MS: Agilent 7890A/5975C Triple Axis Detector Sampler: Agilent 7683B, 5.0 µl syringe (Agilent p/n 5183-4729) Injection: 1.0 µl Carrier: Helium fixed pressure 13.19 psi Inlet: 50:1 split ratio 250 C, total flow 60.351 ml/min, 3 ml/min septum purge, gas saver on, 50 ml/min after 1 minute Inlet Liner: MS certified liner (Agilent p/n 5188-6568) Column : Agilent J&W DB-17ms 30 m 0.25 mm, 0.25 µm (Agilent p/n 122-4712) Oven: 150 C (0.1 min) to 195 C (7 C/min); 15 C/min to 280 C, 2 min hold MSD: Transfer line 280 C, source 300 C, quadrupole 180 C Table 2 Flow Path Supplies Vials: Amber screw top glass vials (Agilent p/n 5183-2072) Vial caps: Screw caps (Agilent p/n 5182-0723) Vial inserts: 100 µl glass/polymer feet (Agilent p/n 5181-8872) Syringe: 10 µl (Agilent p/n 5183-4729) Septum: Advanced Green (Agilent p/n 5183-4759) Inlet Seal: Gold plated inlet seal (Agilent part p/n 5188-5367) Inlet liners: MS certified liner (Agilent part p/n 5188-6568) Ferrules: 0.4 mm id short; 85/15 Vespel/graphite (Agilent p/n 5181-3323) 20x magnifier: 20x Magnifier loop (Agilent p/n 430-1020) Sample Preparation Individual 1000 ng/µl EU flavor and fragrance standard solutions from AccuStandard were diluted 1 to 10 in three groups for single point calibration at 100 ng/µl. It was necessary to separate the standards into three groups to achieve a 1:10 dilution. Nine aldehydes and ketones composed the first group, nine alcohols the second and the remaining alcohols and neutral molecules were diluted in the third group. The internal standard 1,4 di-bromo benzene was purchased from Sigma Aldrich, prepared at 1000 ng/µl and diluted 1:10 along with each group of calibration standards. A grand mix of 24 standards plus the internal standard was prepared at a concentration of 40 ng/µl; this mix was diluted 1:5 to form a working level standard at 8 ng/µl. 2

Results and Discussion Samples were ground in a stainless steel coffee grinder to obtain a uniform powder consistency. One-gram cinnamon gum and 10-gram candy samples were weighed to the nearest 0.1 mg into 50-mL centrifuge tubes. A 10-mL amount of HPLC water was added by a class A volumetric pipette, and the tube was capped and vortexed for 30 s to wet the samples with water. A 10-mL amount of methyl tert-butyl methyl ether (MTBE, JT Baker high purity grade) was added by a class A volumetric pipette. The samples were extracted for 30 min on an Eberbach reciprocating shaker set on high speed. Samples were spun down 4000 times to separate the MTBE and aqueous layers. Samples were taken from the top MTBE layer for analysis. Repeatability of the extraction procedure was investigated using five replicate sample preparations of a cinnamon gum sample. The large amount of cinnamaldehyde present in this sample required a 30:1 dilution to bring the response into range of the standard. A semiquantitative analysis of the amount of cinnamaldehyde on a weight to weight basis in the sample was determined in the dilute solutions. The weight percent result for cinnamaldehyde in cinnamon gum was 0.22 ± 0.01 % for duplicate injections of five replicate weighting with a relative standard deviation of 4.3 %. This result is encouraging and indicates a high level of repeatability for the extraction procedure. Extraction efficiency and accuracy evaluations of the extraction procedure were beyond the scope of this study. Given the encouraging result found in the repeatability evaluation a more thorough investigation of the extraction in terms of efficiency and accuracy is a logical next step. Figure 1 shows the separation of 24 GC/MS-amenable allergens on a DB-17ms, 30 m 0.25 mm, 0.25 µm capillary GC column at a concentration of 8 ng/µl. This figure clearly shows that all 24 GC/MS-amenable allergens are detectable and identifiable 20 % below the leave on limit using a DB- 17ms column and a 50:1 split injection. Reduction of the split ratio could achieve lower limits of detection at the expense of a higher matrix background effect. Backflusing using a capillary flow device (CFT) of late-eluting components would be one way of counteracting deleterious matrix effects and increasing sample throughput. 8 ppm Standard Injection of 24 EU Allergens on a DB-17ms Column 1 2 3 4 6 5 7 8a 8b 9 10 12 11 13 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 Time Sample: 8 ppm Column: Agilent J &W DB-17 ms 30 m 0.25 mm, 0.25 µm Carrier: Helium, fixed pressure 13.19 psi retention time locked to 1,4 dibromobenzene Oven: 50 C (1.0 min) to 195 C 7 C/min, then 15 C/min to 280 C 2 min hold Injection: 1.0 µl, Split 50:1, 250 C, septum purge 3 ml/min total flow 60.351 ml/min MSD: Transfer line 280 C, Source 300 C, Quad 180 C 14 15 16 17 18b 18a 19 21 22 20b 20a 23 24 25 1 Limonene 2 Linalool 3 Benzyl Alcohol 4 Citronellol 5 Folione 6 Geraniol 7 1,4 Dibromo benzene 8a Citral (neral) 8b Citral (geranial) 9 Hydroxy citronellal 10 Cinnamaldehyde 11 Anisyl alcohol 12 Cinnaminic alcohol 13 Eugenol 14 α-isomethyl ionone 15 Isoeugenol 16 Lilial 17 Coumarin 18a Farnesol 1 19 Amyl cinnamaldehyde 18b Farnesol 2 20a Lyral 1 20b Lyral 2 21 Amyl cinnamyl alcohol 22 Hexyl cinnamaldehyde 23 Benzyl benzoate 24 Benzyl salicilate 25 Benzyl cinnamate Figure 1. The 8 ppm standard injection of 24 EU allergens on an Agilent J&W DB-17ms, 30 m 0.25 mm, 0.25 µm column (p/n 122-4712). GC conditions are in Table 1. 3

Expanded View of Farnesol and Lyral Separation at the 8 ppm level X 19 21 20b 18a Farnesol 1 19 Amyl cinnamaldehyde 18b Farnesol 2 20a Lyral 1 20b Lyral 2 21 Amyl cinnamyl alcohol 18b 18a 20a Figure 2. Expanded view of farnesol and lyral separation at the 8 ppm level on a Agilent J&W DB-17ms, 30 m 0.25 mm, 0.25 µm column (p/n 122-4712). GC conditions used are in Table 1. The x label peak is a contaminant present in the 24-allergen mix and not an analyte of interest. Citral, farnesol and lyral are all isomeric mixtures and under these conditions two peaks were observed for each of these analytes. Figure 2 is an expanded view of the farnesol and lyral isomers from amyl cinnamaldehyde and amyl cinnamyl alcohol in the 8 ppm, 24-mix standard. One isomer for both farensol and lyral analytes is baseline resolved. The second farnesol peak (18a) is partially resolved from the tail of the amyl cinnamaldehyde peak (19). The second lyral peak (20b) is partially resolved from the front of the amyl-cinnamyl alcohol peak (21). Given the unique elution patterns and mass spectral properties of these analytes, effective identification is achievable at these levels. Figure 3 is an example chromatogram of one of the neat cinnamon gum extracts. Benzyl alcohol, cinnamaldehyde, cinnaminic alcohol, and eugenol are present in this sample at detectable and identifiable levels. Cinnamaldehyde is present at a level nearly 30 times the 100-ppm limit. It was necessary to dilute this extract 30:1 to bring the signal for cinnamaldehyde in range for quantification. Conclusions A simple MTBE extraction procedure produced a high degree of repeatability for allergen analysis in a cinnamon flavored gum sample. The extraction procedure shows promise for allergen analyses in difficult matrices such as chewing gum and candy. Perhaps a solid phase extraction approach using MTBE as a solvent would be a useful next step in more fully characterizing the preparation of these samples. This application successfully demonstrates the utility of a single quad GC/MS approach using an Agilent J&W DB-17ms column for analysis of the EU flavor and fragrance allergens in candy and gum samples. Using the GC/MS conditions described, 24 of the EU listed allergens amenable to GC/MS analysis are detectable and identifiable. Identification at 8 ppm, 20 % below the leave on limit set for fragrance and cosmetic products in the EU directive, was easily achievable. Figure 4 is an example chromatogram of a neat cinnamon candy extract in MTBE. Limonene, benzyl alcohol, cinnamaldehyde, eugenol, and benzyl benzoate are present in this sample at detectable and identifiable levels. Cinnamaldehyde is present at a level more than 50 times the 100-ppm standard in this sample. 4

4 Listed EU Allergens Present in Cinnamon Gum MTBE Extract Big Red Gum Extract Cinnamaldehyde Cinnaminic alcohol Eugenol Long chain Alkanes Benzyl alcohol 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 Time Figure 3. Neat MTBE cinnamon gum extract injection on a Agilent J&W DB-17ms, 30 m 0.25 mm, 0.25 µm column (p/n 122-4712). GC conditions used are in Table 1. 5 EU Listed Allergens Present in Cinnamon Candy MTBE Extract Cinnamaldehyde Eugenol Limonene Benzyl alcohol Benzyl Benzoate 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 Time Figure 4. Neat MTBE cinnamon candy extract injection on a Agilent J&W DB-17ms, 30 m 0.25 mm, 0.25 µm column (p/n 122-4712). GC conditions are in Table 1. 5

Reference 1. EU Directive 2003/15/EC, Official Journal of the European Union, 6 66.26, 11.3.2003 2. W. Luan, C. Sandy, and M. Szelewski, "Determination of Allergens in Fragrance Products Using Agilent Deconvolution Reporting Software," Agilent Technologies publication 5989-8723EN, June 2008 3. F. David and M.S. Klee, "Analysis of Suspected Flavor and Fragrance Allergens in Perfumes Using Two-Dimensional GC with Independent Column Temperature Control Using an LTM Oven Module," Agilent Technologies publication 5990-3576EN, February, 2009 For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2009 Printed in the USA October 1, 2009 5990-4784EN