International Journal of Radiation Research, July 2014

Similar documents
Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

Monte Carlo and experimental relative dose determination for an Iridium-192 source in water phantom

Dosimetric Characteristics of the Brachytherapy Sources Based on Monte Carlo Method

Gafchromic film dosimetry of a new HDR 192 Ir brachytherapy source

Material-specific Conversion Factors for Different Solid Phantoms Used in the Dosimetry of Different Brachytherapy Sources

Research Article Dose Distributions of an 192 Ir Brachytherapy Source in Different Media

Comparative Study on Homogeneous and Inhomogeneous Photon Dose Distribution from a High Dose Rate 192 Ir Brachytherapy Source using MCNP5

AAPM Administrative Policy: Joint AAPM/IROC Houston Registry of Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites

A COMPARISON BETWEEN GEANT4 AND MCNPX ON THE DOSIMETRY OF THE 192IR MICROSELECTRON V2 HDR BRACHYTHERAPY

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

Monte Carlo Simulation Study on Dose Enhancement by Gold Nanoparticles in Brachytherapy

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Calculation of Dose Distribution Around a Clinical 252. Cf Source for Neutron Therapy Based on AAPM, TG-43 Protocol

Comparison of 60 Co and 192 Ir sources in HDR brachytherapy

HDR Brachytherapy Dosimetry at the NPL

Modeling and dose calculations of a pure beta emitting 32 P coated stent for intracoronary brachytherapy by Monte Carlo code

Optimisation of eye plaque dosimetry using Monte Carlo method. J. Green, D. Cutajar, S. Guatelli, & Rosenfeld, A.B.

Implantable MOSFET dosimeter response to 192 Ir HDR radiation

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Episcleral eye plaque dosimetry comparison for the Eye Physics EP917 using Plaque Simulator and Monte Carlo simulation

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Brachytherapy, Radionuclide Therapy Medical Physics in the Clinic. Raymond K. Wu, PhD Chairman AAPM Exchange Scientist Program

MAMMOSITE BRACHYTHERAPY DOSIMETRY EFFECT OF CONTRAST AND AIR INTERFACE ON SKIN DOSE. A RESEARCH PAPER SUBMITTED TO THE GRADUATE SCHOOL

Calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

Dosimetry benchmark for MBDCA

Intravascular Brachytherapy Dosimetry Techniques for Gamma systems

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

Dose Calculation for Photon-Emitting Brachytherapy Sources with Average Energy Higher than 50 kev: Full Report of the AAPM and ESTRO

DOSIMETRIC STUDY OF I-125 SEEDS USED IN PROSTATE BRACHYTHERAPY

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

Supplement 2 for the 2004 update of the AAPM Task Group No. 43 Report: Joint recommendations by the AAPM and GEC-ESTRO

Applications in brachytherapy

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers

University of Wollongong. Research Online

Disclosure. Societal Guidelines in High Energy and Skin Brachytherapy. Content. Acknowledgments. Jose Perez-Calatayud.

Chapter 13: Brachytherapy: Physical and Clinical Aspects

Metrology Laboratory of Ionizing Radiation

Review of TG-186 recommendations

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Developing a Verification and Training Phantom for Gynecological Brachytherapy System

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

Calibration of two 90 Sr+ 90 Y dermatological applicators

Radioactive sources in brachytherapy

Initial evaluation of Advanced Collapsed cone Engine dose calculations in water medium for I-125 seeds and COMS eye plaques

Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters

Larry A. DeWerd Accredited Dosimetry and Calibration Laboratory, University of Wisconsin, Madison, Wisconsin, 53706

Brachytherapy Planning and Quality Assurance

Study of a spherical phantom for Gamma knife dosimetry

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD

Radiochromic film dosimetry in water phantoms

A Real-Time Monte Carlo System for Internal Dose Evaluation Using an Anthropomorphic Phantom with Different Shapes of Tumors Inserted

Dosimetric parameters of three new solid core I-125 brachytherapy sources

12 Physics and Clinical Aspects of Brachytherapy

Radiation Dosimetry at the BNL High Flux Beam Reactor and Medical Research Reactor

Advanced film dosimetry for a UK brachytherapy audit

Optimum radiation source for radiation therapy of skin cancer

Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

Jeffrey F. Williamson a) Radiation Oncology Center, Washington University, St. Louis, Missouri 63110

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

Eye plaque dosimetry verification using novel solid state devices

Iranian Journal of Medical Physics

Epithermal neutron beams from the 7 Li(p,n) reaction near the threshold for neutron capture therapy

Dosimetry in digital mammography

THERMOLUMINESCENT (TL) DOSIMETRY OF SLOW-NEUTRON FIELDS AT RADIOTHERAPY DOSE LEVEL

Recent proceedings in Brachytherapy Physics

U.S. Nuclear Regulatory Commission

Metrology Laboratory of Ionizing Radiation

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

Current Status of Electronic Brachytherapy Dosimetry

A REANALYSIS OF THE COLLABORATIVE OCULAR MELANOMA STUDY MEDIUM TUMOR TRIAL EYE PLAQUE DOSIMETRY

Seed coordinates of a new COMS-like 24 mm plaque verified using the FARO Edge

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

IPPE Iron shell transmission experiment with 14 MeV neutron source and its analysis by the Monte-Carlo method

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

In vivo dosimetry and seed localization in prostate brachytherapy with permanent implants

Tissue inhomogeneity in proton therapy and investigation of its effects on BRAGG peak by using MCNPX code

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing

Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Commissioning of an In Vivo Dosimetry system for high dose rate gynaecological brachytherapy

A tissue-equivalent phantom series for mammography dosimetry

Sang Hun Shin, Wook Jae Yoo, Kyoung Won Jang, Seunghyun Cho, 1 Kum Bae Kim, 2 and Bongsoo Lee 3*

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters

The Albedo Dosimeter for Personnel Monitoring in Fast-Neutron Radiation Fields

Impact of range straggling and multiple scattering on proton therapy of brain, using a slab head phantom

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Single Photon Emission Tomography Approach for Online Patient Dose Assessment in Boron Neutron Capture Therapy

Transcription:

Volume 12, No 3 International Journal of Radiation Research, July 2014 Dosimetric characterization of a high dose rate 192 I source for brachytherapy application using Monte Carlo simulation and benchmarking with thermoluminescent dosimetry N. Babaei Bidmeshki 1*, M. Sohrabpour 1, S.R. Mahdavi 2 1Department of Energy Engineering, Sharif University of Technology, P.O. Box 1485893169, Tehran, Iran 2Department of Medical Physics, Faculty of Medicine, Tehran University of Medical Sciences P.O. Box 1449614525, Tehran, Iran Technical note * Corresponding author: Dr. Nadia Babaei Bidmeshki, Fax: +98 21 44053066 E mail: nadia_babaei@energy.sharif.ir Received: May 2013 Accepted: Oct. 2013 Int. J. Radiat. Res., July 2014; 12(3): 265-270 ABSTRACT Background: The purpose of this project was to derive the brachytherapy dosimetric func ons described by American Associa on of Physicists in Medicine (AAPM) TG 43 U1 based on high dose rate 192 I sources. Materials and Methods: The method u lized included both simula on of the designed Polymethyl methacrylate (PMMA) phantom using the Monte Carlo of MCNP4C and benchmarking of the simula on with thermoluminescent (TL) dosimeters. Results: The obtained results for the radial dose func on and anisotropy func on showed nominal errors of less than 3% between TL measurements and the MCNP4C results. Conclusion: It may be concluded that due to small observed errors and the large uncertainty associated with the high dose gradients near the source point the simula on results can be used for dose es ma on. Keywords: Brachytherapy, 192 Ir flexisource, MCNP calculation, radiation dosimetry. INTRODUCTION Brachytherapy is the clinical use of small encapsulated radioactive sources at a short distance from the target volume for irradiation of malignant tumors. Permanent implant brachytherapy has long employed radionuclides that emit principally low energy gamma rays that can be produced to provide suitable activity (1). Valid clinical application for treatment planning requires reliable data of the dosimetric characteristics of a new source design. Since currently varieties of 192Ir sources are commercially available with different core and encapsulation dimensions, accurate dosimetry are required in the close vicinity of the HDR iridium sources. Both theoretical and experimental methods have been widely used to characterize two dimensional dose distributions around brachy sources in water mediums (2 4). Monte Carlo simulation, ilm dosimetry and thermoluminescent dosimetry (TLD) are used for determination of dose distribution around sources in different media (5 7). The aim of this work is calculating the dosimetric data of an HDR 192 Ir seed source (Flexisource) and also comparing the results with the obtained data from TL dosimeters within a PMMA phantom. The data is used for dose distribution measurements around the Flexisource in the after loading Flexitron system.

MATERIALS AND METHODS Source description The radiation source used in this investigation was a HDR 192Ir, Flexisource (Veenendaal, The Netherland), with an initial activity of 370 GBq at the time of manufacturing. The geometric design of the Flexisource is shown in igure 1. The active core of the source is a pure iridium cylinder (density 22.42 g/cm 3 ) with an active length of 3.5 mm and a diameter of 0.6 mm. The active core is covered by a stainless steel 304 capsule (density 8 g/cm 3 ) with outer source dimensions of 0.85 mm and 4.6 mm for diameter and length, respectively. It was assumed that the radioactive material was uniformly distributed within the 192Ir active core (8). The decay scheme of 192Ir was available on line in the Nuclear Data Base of the IAEA. The gamma emission spectrum of 192 Ir is shown as a histogram representation in igure 2. The source is normalized to one Bq of 192 Ir. This is achieved by using the sum of the emission probabilities for all photons per decay (=2.363) of 192 Ir (9). It was observed by Mowlavi et al. that the real spectrum rather than monochromatic spectra with energy of 356 kev was important for dose calculation near and far from the source (10). The β spectrum of the 192 Ir source has not been considered since its contribution to the dose rate distribution for distances greater than 1 mm from the source is negligible due to source encapsulation. Phantom design In selection of a calibration or dosimetric phantom a number of key parameters of the phantom, including thickness for production of the full scattering condition and the composition of tissue substitute must be taken into account. For photons the effective atomic number of the material is important. Water, acrylics, polystyrene or similar low atomic number materials are appropriate for such phantoms (11). Polymethyl methacrylate (PMMA) is one such material with low effective atomic number (Zeff=6.5) that is found suitable for calibration and dose measurements. In this work a PMMA slab phantom of dimensions 30 30 20cm 3 was used to measure the dose distribution around the source. This phantom was designed with a central hole for insertion of the source carrier applicator which was perpendicular to the axis of the TLD discs. The TLD discs were arranged in circular forms with a radial distribution in the range of 0.5 to 9 cm and were placed perpendicular to the source axis. Figure 3 shows the picture of the geometric design of the phantom. Figure 2. The real spectrum of 192 Ir. Figure 1. Schema c diagram of the 192 Ir Flexisource (dimensions in millimeter). Figure 3. Geometric design of the PMMA phantom. Int. J. Radiat. Res., Vol. 12 No. 3, July 2014 266

Monte Carlo simulation The Monte Carlo code system MCNP4c was employed for the simulation. MCNP is a general purpose Monte Carlo radiation transport code which can simulate coupled neutron photon electron transport in three dimensions through complex geometries constructed as Boolean combinations of planes, spheres, cones and cylinders. This code transports photon and electrons in the energy range from 1 kev to 100 MeV. The detailed photon physics treatment includes photoelectric absorption; K and L shell luorescence, Auger emission, coherent scattering with electron binding effects accounted for by form factors, and incoherent scattering. A photoelectric event is considered to be absorption and the photon history is terminated in such an interaction. In the case of a Compton interaction, the inal energy and position of the photon are randomly sampled from Klein Nishina relation based angular probability and energy distribution functions (12). The code was run in combined photon electron mode. An energy cut off value of 1 kev was used to terminate tracking of photon and electron histories. MCNP requires the source for a particular problem to be speci ied in a user de ined input ile. The source includes distributions of the positions, energy and angle of the starting particles. The MCNP tally options utilized for the needed parameters was *F8 tally with Mode p e. The cross section library containing data from the ENDF was included with the MCNP code that was utilized for the computations. Where Ecci is the ef iciency correction coef icient for each TLD disc, Ri and R are individual and average readings of the total TLDs respectively. This correction factor is due to differences in TLD s physical properties such as size, mass, etc. For calibration purposes TLD discs were placed under Plexi glass plates for electronic equilibrium and were irradiated from a calibrated Co 60 source. To obtain the absorbed dose calibration factor in Gy/Count, TLD discs in groups of ive were irradiated with different doses from 0.3 Gy to 5 Gy. Then to calculate the corrected reading of each ield dosimeter the followed equation was used (14) : (2) Dosimetric formalism The dose rate at a point (r,θ) in the phantom around the 192Ir source was determined following the protocol introduced by group 43 AAPM using the following equation (15 16): (3) where r denotes the distance from the center of the active source to the point of interest, r0 denotes the reference distance, and θ denotes the polar angle specifying the point of interest relative to the source longitudinal axis, and θ0 is the reference angle. The de initions of the denoted parameters are shown in igure 4. Thermoluminescent dosimeters TLD 100H (Harshaw) discs with diameter length of 4.5 mm 0.9 mm thickness were used in this study. The protocol for using TLD 100H was described in detail by Mckeever et al. and Furetta (13 14). Brie ly, the discs were irst annealed at 240 C for 10 min. To determine the sensitivity of each individual TLD, ef iciency correction coef icients (ECC) were obtained by the following equation: 267 (1) Figure 4. Coordinate system used for the dosimetry calcula on formalism (Update of AAPM TG 43 report). Int. J. Radiat. Res., Vol. 12 No. 3, July 2014

In equation 3, D(r,θ) is the dose rate at point (r,θ), Sk is the source strength (air kerma strength), Λ is dose rate constant, GL(r,θ) is geometry factor, gl(r) is the radial dose function, and F(r,θ) is anisotropy function. Geometry function is used to improve the accuracy with which dose rates can be estimated by interpolation from data tabulated at discrete points. Physically, the geometry function neglects scattering and attenuation, and provides an effective inverse square law correction based on an approximate model of the spatial distribution of radioactivity within the source. Because the geometry function is used only to interpolate between tabulated dose rates values at de ined points, highly simplistic approximations yield suf icient accuracy for treatment planning: (4) for line source approximation, where β is the angle, in radians, subtended by the tips of the hypothetical line source with respect to the point of interest (r,θ). The radial dose function accounts for dose fall off on the transverse plane due to photon scattering and attenuation, i.e., excluding fall off included by the geometry function: (5) RESULTS AND DISCUSSION The radial dose function, g(r), was calculated according to the TG 43 de inition and is depicted in table 1. Figure 5 details both simulated and measured data for the Flexisource. Comparison of the measured and simulated data isthe radial dose function, g(r), was calculated according to the TG 43 de inition and is depicted in table 1. Figure 5 details both simulated and measured data for the Flexisource. Comparison of the measured and simulated data is found to be nominally less than 3%. For this source, the Monte Carlo calculated g(r) function in the range of 0.5 to 9 cm was itted using a ifth order polynomial curve, namely (7) where a0=0.7027, a1=0.4847, a2= 0.2513, a3=0.0574, a4= 0.006, a5=0.0002. Table I. A comparison of the calculated and measured radial dose func on of the Flexisource 192 Ir brachytherapy source. r(cm) g(r), g(r), Rela ve simula on measurement difference percent 0.6 0.91 0.82 10.01 1 1 1 0 2.4 1.03 1.01 2.15 3 1.02 1.02 0 3.7 0.98 0.96 3.7 4.5 0.98 0.9 7.5 5.2 1.01 1.03 1.94 7.5 0.99 0.97 2.02 9 0.84 0.81 3.6 The 2 D anisotropy function describes the variation in dose distribution around the source, including the effects of absorption and scatter in the medium as a function of polar angle relative to the transverse plane due to self iltration, oblique iltration of primary photons through the encapsulating material and scattering of photons in the medium. 2 D anisotropy function is de ined as: Figure 5. Comparison of the measured dose func on, g(r), of (6) the Flexisource with the Monte Carlo simulated data in PMMA phantom. Int. J. Radiat. Res., Vol. 12 No. 3, July 2014 268

The results of the two dimensional measurements of the dose distribution in phantom were analyzed using the geometry factor, G(r,θ), to arrive at an anisotropy function for the Flexisource. Figures 6(a) and 6 (b) detail both simulated and measured data for the Flexisource. Nominal relative differences with respect to Monte Carlo simulation are found to be 10% at angles between 0 to 30 degree and less than 2% at other angles. The greater errors between simulated and measured F(r,θ) for the angles of 90 to 150 degrees is partially due to the transit dose which depicts itself as a higher value of the measured data at these angles. This same transit dose phenomenon also presents itself as an asymmetric dose distribution with higher values occurring at angles 90 to 150. Figure 7 presents a comparison of the radial dose functions for various 192Ir HDR sources. It indicates that radial dose functions are very similar for these HDR sources. The discrepancies between the radial dose functions far from the source (r > 5 cm) arise from the different sizes of the water phantoms (8). In igure 8, the anisotropy functions of various 192 Ir HDR sources are compared for r = 1 cm. It shows that the anisotropy function is very similar for Flexisource, Nucletron mhdr v2, BEBIG HDR, and Gammamed 12i sources due to their similar geometric design. In contrast, signi icant differences exist between the anisotropy function of these sources and that of the Buchler source. This is due to the larger distance from the active edge to the tip of the source, the larger active diameter, and a larger iltration capsule thickness for the former source model compared with the other ones (8). In view of the very high dose gradient near the source point, the actual measurement may have large associated uncertainties. On the other Figure 7. Comparison of the radial dose func ons for various 192Ir HDR sources. Figure 6. Comparison of the measured and Monte Carlo calculated anisotropy func on for the Flexisource in a PMMA phantom at radial distances of (a) r=4.5 cm, and (b) r=5.2 cm. 269 Figure 8. Comparison of the anisotropy func ons for various 192Ir HDR sources for the radial distance of r = 1 cm. Int. J. Radiat. Res., Vol. 12 No. 3, July 2014

hand, simulated dose measurements at any distance near or far away from the source point can be performed with reasonable accuracy, and thus it may be feasible to partially rely on the simulated data in places where dosimetry is dif icult to carry out. CONCLUSIONS The dosimetric characteristic of the Flexisource 192Ir brachytherapy source by AAPM TG 43 formalism has been performed using standard methods employing thermoluminescent dosimeters in a water equivalent phantom. The results of this study indicate that Monte Carlo simulation in brachytherapy is useful to obtain dosimetric parameters and to verify the measurement data. ACKNOWLEDEGEMETS The authors would like to express their appreciation to Mr. Arjang Shahvar for his technical support in this study. REFERENCES 1. Baltas DS, Sakelliou L, Zamboglou N (2007) The Physics of Modern Brachytherapy for Oncology. Taylor & Francis Group, New York. 2. Ghiassi Nejad M, Jafarizadeh M, Ahmadian Pour MR, Ghahramani AR (2001) Dosimetric characteris cs of 192 Ir sources used in inters al brachytherapy. Appl Radiat Isot, 55: 189 195. 3. Duggan DM (2004) Improved radial dose func on es ma on using current version MCNP Monte Carlo simula on: Model6711 and ISC3500 125 I brachytherapy sources. Appl Radiat Isot, 61: 1443 1450. 4. Wang Z and Hernel NE (2005) Determina on of dosimetric characteris cs of op Seed a plas c brachytherapy 103 Pd source. Appl Radiat Isot, 63: 311 321. 5. Sellakumar P, Sathishkumar A, Supe SS, Anand MR, Nithya K, Sajitha JF (2009) Evalua on of dosimetric func ons for 192 Ir source using radiochromic film. Nucl Instrum Methods, 267: 1862 1866. 6. Sowards KT and Meigooni AS (2002) A Monte Carlo evalua on of the dosimetric characteris cs of the Best Model 2301 125 I brachytherapy source. Appl Radiat Isot, 57: 327 333. 7. Wallace RE (2002) Model 3500 125 I brachytherapy source dosimetric characteriza on. Appl Radiat Isot, 56: 581 587. 8. Granero D, Perez Calatayud J, Casal E, Ballester F, Venselaar J (2006) A dosimetric study on the 192 Ir dose rate Flexisource. Med Phys, 33: 4578 4582. 9. Borg J and Rogers DWO (1999) Spectra and air kerma strength for encapsulated 192 Ir sources. Med Phys, 26: 2441 2444. 10. Mowlavi AA, Cupardo F, Severgnini M (2008) Monte Carlo and experimental rela ve dose determina on for an 192 Ir source in water phantom. Iran J Radiat Res, 6: 37 42. 11. ICRU Report 48 (1992) Phantoms and computa onal models in therapy, diagnosis and protec on. ICRU Publica on, Bethesda. 12. Los Alamos Na onal Laboratory (2000) MCNP4c: Monte Carlo N par cle transport code system. Oak Ridge Na onal Laboratory, New Mexico. 13. McKeever SWS, Moscovitch M, Townsend PD (1995) Thermoluminescence Dosimetry Materials, Proper es and Uses. Nuclear Technology Publishing, Ashford. 14. Fure a C (2003) Handbook of the Thermoluminescence. World Scien fic Publishing, Singapour. 15. Rivard MJ, Coursey BM, Deward LA, Hanson WF, Huq MS, Mitch MG, Nath R, Williamson JF (2004) Update of AAPM Task Group No. 43 Report: a revised AAPM protocol for brachytherapy dose calcula ons. Med Phys, 31: 633 674. 16. Yue NJ (2007) Principles and prac ce of brachytherapy dosimetry. Radiat Measur, 41: 522 527. Int. J. Radiat. Res., Vol. 12 No. 3, July 2014 270