High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS)

Similar documents
A Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS

Finally, a Solid Phase Extraction product for Drugs of Abuse

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

Comparison of Different Whole Blood Sample Pretreatment Methods for Targeted Analysis of Basic Drugs

Overcoming the Challenges of Sample Preparation and LC/MS/MS Method Development for Clinical Applications

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Strata -X Polymeric SPE

Application Note. Abstract. Authors. Pharmaceutical

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Extraction of 11-nor-9-carboxy-tetrahydrocannabinol from Hydrolyzed Urine by ISOLUTE. SLE+ Prior to GC/MS Analysis

Rapid and Robust Detection of THC and Its Metabolites in Blood

Validation of an Automated Method to Remove

Strata-X- NEW! Drug B. Drug B Plus for In-Well Hydrolysis. Simplify Your Solid Phase Extraction.

Rapid Analysis of Synthetic Cannabinoids and their Metabolites in Urine using Solid Phase Extraction and LC/MS/MS. JWH018-4-Hydroxypentyl metabolite

UCT STYRE SCREEN INNOVATION THROUGH CHEMISTRY

PAIN PANEL. Research and Solutions

A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using CLEAN UP CCX2 and LC-MS/MS

Fast and easy separation of 23 drugs of abuse. including high, stable resolution of isobaric opioids from human urine by UHPLC-MS/MS

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Keep It Clean. Solid Phase Extraction Solutions for Drugs of Abuse Analysis. Neutral Drugs of Abuse Basic Drugs of Abuse

Application. Detection of Cannabinoids in Oral Fluid Using Inert Source GC/MS. Introduction. Authors. Abstract. Forensic Toxicology

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using 500 MG CLEAN UP CCX2 and LC-MS/MS

E XCEL LENCE JUST GOT BETTER UCT FORENSICS CLEAN SCREEN XCEL SPE COLUMNS

Performance of an ultra low elution volume 96-well plate

High-Throughput GC/MS Confirmation and Quantitation of Codeine and Morphine in Urine Using the DSQ II

Direct Analysis of Urinary Opioids and Metabolites by Mixed-Mode µelution SPE Combined with UPLC/MS/MS for Forensic Toxicology

Modernizing the Forensic Lab with LC-MS/MS Technology

Simultaneous Determination of Prescription and Designer Benzodiazepines in Urine and Blood by SPE and LC-MS/MS

Matrix Factor Determination with the Waters Regulated Bioanalysis System Solution

Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes

The Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis

Protein Precipitation for Biological Fluid Samples Using Agilent Captiva EMR Lipid 96-Well Plates

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

Determination of Pharmaceutical Residues in Bovine Milk via LC MS/MS Following Solid Phase Extraction

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

Extraction of Synthetic and Naturally Occurring Cannabinoids in Urine Using SPE and LC-MS/MS

Comprehensive Study of SLE as a Sample. Preparation Tool for Bioanalysis

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

A Fast, Accurate, and Sensitive Quantitiation of Vitamin C in Plasma by Impact Protein Precipitation and Kinetex 5 µm Core-Shell XB-C18 HPLC Columns

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

Title Revision n date

A Novel Platform of On-line Sample Pre-treatment and LC/MS/MS Analysis for Screening and Quantitation of Illicit Drugs in Urine

Determination of Amantadine Residues in Chicken by LCMS-8040

Application Note. Author. Abstract. Introduction. Food Safety

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis

Screening by immunoassay and confirmation & quantitation by GC-MS of buprenorphine and norbuprenorphine in urine, whole blood and serum

PAIN PANEL. Application Guide. Screen 41 compounds in under 4 minutes

Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE. SLE+ Prior to LC/MS-MS Analysis

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

Phenomenex roq QuEChERS Kits. Overview. Spinach Facts

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

Core E Analysis of Neutral Lipids from Human Plasma June 4, 2010 Thomas J. Leiker and Robert M. Barkley

Fast quantitative Forensic Analysis of THC and its Metabolites in Biological Samples using Captiva EMR- Lipid and LC/MSMS

Sample Preparation Techniques for Biological Matrices: Finding the right balance to achieve optimal results

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research

Authors. Abstract. Forensic Toxicology. Irina Dioumaeva, John M. Hughes Agilent Technologies, Inc.

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Integration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation

Fig. 1: Chemical structure of arachidonic acid COOH CH 3

APPLICATIONS TN A Comparison of Various Kinetex C18 Phases USP: L1

Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Extraction of 25-hydroxy Vitamin D from Serum Using ISOLUTE. PLD+ Prior to LC-MS/MS Analysis

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

Analysis of HMF by HPLC

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

The challenging extraction of non-polar contaminants out of a non-polar vegetable oil sample

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Supporting Information

Enrichment of Phospholipids from Biological Matrices with Zirconium Oxide-Modified Silica Sorbents

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

Improved Isolation and Analysis of Mycotoxins from Cereals, Beer and Wine

Determination of Patulin in Apple Juice Using SPE and UHPLC-MS/MS Analysis

Quantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Determination of Ractopamine and Clenbuterol in Beef Samples

Automated Hydrolysis, Extraction and Determination of Opioids in Urine using a Novel Robotic Autosampler and LC-MS/MS Platform

The Development of LC/MS Methods for Determination of Polar Drugs of Abuse in Biological Samples

Run 1 Run 2 TDP PLP TDP PLP. 3.0e5 2.8e5 2.2e5. 2.6e5 2.4e5. 2.2e5 2.0e5. Intensity, cps. 1.8e5 1.6e5 1.4e5 1.2e5 1.0e5 4.0e4. 2.0e e5 1.

All stocks and calibration levels were prepared in water: methanol (50:50) v/v to cover range of all steroid concentrations (refer Table 1).

Comparison of Different Whole Blood Sample Pretreatment Methods for Targeted Analysis of Basic Drugs. Table 1. List of pain panel drugs

Interested in conducting your own webinar?

Author. Introduction. Abstract

Transcription:

High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS) Michael Rummel, Matthew Trass, Michael Campognone, and Sky Countryman Phenomenex, Inc., 411 Madrid Avenue, Torrance, CA 90501 USA PO66690109_L_1

Objective The prescription of opiate drugs is common for the treatment of chronic pain. Due to the addictive nature of these drugs, many opiates are regulated under Schedule I and routinely tested as part of pre-employment screening or workplace performance monitoring. The confirmation of these analytes by GC/MS can be complicated by two factors, 1) removal of endogenous compounds from urine that can co-elute with target analytes and 2) the presence of non-target opiates which have similar mass ions. The objective of this work was to develop a high throughput assay for opiates in urine utilizing solid phase extraction (SPE) for sample clean up and analysis by either GC/MS or LC/MS/MS. Method Urine samples were spiked at 2,000 ng/ml with morphine, codeine, and other commonly encountered opiate related interferences. A 2 ml aliquot of urine was then spiked with internal standards and 400 µl concentrated HCI. Hydrolysis was performed by heating the sample at 90 C for 40 minutes in a heating block, or in an autoclave for 15 minutes on a liquid cycle. After samples were cooled, they were centrifuged for 10 minutes at 2000 rpm and the pellet was discarded. The hydrolyzed samples were then buffered and extracted using a 30 mg/3 ml SPE tube containing a polymeric cationic exchange sorbent under the conditions specified in Table 1. Wash conditions were optimized to remove polar cationic contaminants that interfered with the GC/MS analysis. Separation by GC/MS of morphine and codeine from other opiate related inferences was evaluated using two different derivatization procedures. Samples were simultaneously analyzed by LC/MS/MS on an Applied Biosystems API 3000 without the need for derivatization.

Table 1: SPE Protocol Strata -X-C 30 mg/3 ml SPE Tubes (Part Number: 8B-S029-TBJ) Condition: - 1 ml of Methanol - 1 ml of Water Load: - To 1 ml of Urine, add 1 ml of 100 mm Sodium Acetate Buffer at ph 5.8-5.9 Wash: - 1 ml of 100 mm Sodium Acetate Buffer at ph 5.8-5.9-1 ml of Methanol Dry: - 5 minutes at 10 inches of Mercury Elute: - 1 ml 7:2:1 Ethyl Acetate / Isopropanol / Ammonium Hydroxide - Evaporate to dry at < 40 C. Derivatization: - Reconstitute sample in 50 µl of Pyridine and 50 µl of Propionic Anhydride, vortex - React mixture at 60 ºC for 30 minutes - Allow sample to cool and then dry - Reconstitute in 50 µl of Ethyl Acetate

Objective Due to the increase in sensitivity provided by newer MS instrumentation, especially LC/MS/MS, the volume of urine needed for extraction in order to meet SAMHSA detection limits can be reduced. We were able to reduce the volume to 1 ml of urine, which was then diluted 1:1 with a buffer solution to a total volume of 2 ml. This volume allowed us to use a new optimized SPE format with 30 mg of polymeric cation exchange sorbent packed into a 3 ml tube. Polymeric sorbents offer several important advantages over silica based sorbents, including improvements in tube to tube reproducibility due to metal content in silica, the reduced susceptibility to de-conditioning if the sorbent bed dries out between processing steps, and the improved flow characteristics. By using a 30 mg bed mass, we were able to significantly reduce the solvent volume required for conditioning, wash, and elution steps. This reduced solvent volume saves both time and money significantly for subsequent blow down steps prior to derivatization. The recoveries for morphine and codeine were 97 % and 98 % respectively, with %CVs < 4 % for replicate injections over three different concentrations, indicating that the sorbent mass was sufficient to reproducibly perform the extraction. When developing the method, several different wash and elution options were explored that provided visually different results. The opiate species are cationic and were retained via ion-exchange mechanisms on the Strata - X-C sorbent. Washing with acidified water solutions helps to remove polar acidic and neutral components. In this case, the use of buffer significantly reduced the presence of a yellowish impurity (Figure 1). However, even the use of a buffer solution was not sufficient to completely remove this impurity. Additional experiments suggested that this impurity was insoluble in strong organic solutions. By changing the elution solvent system from basified methanol to a mixture of 7:2:1 ethyl acetate / isopropanol / ammonium hydroxide, we were able to almost completely remove this impurity from the final extracts. GC/MS remains the golden standard for SAMHSA opiate testing. The large number of routinely prescribed opiate related compounds that cause positive hits in the ELISA screening assay requires a GC/MS methodology that is able to accurately confirm the presence of morphine or codeine. To improve this separation, the use of a newly developed GC stationary phase specifically designed for the analysis of drugs of abuse was evaluated. To further improve resolution, we investigated the difference in chromatographic performance between BSTFA and propionic acid anhydride derivatized samples. The elution order of morphine and codeine was reversed on the ZB-Drug-1 vs. standard 5 and 5ms phases. BST- FA derivatives were well resolved from all common coeluting impurities, though problems were observed with norcodeine (Figure 2). The use of BSTFA also required the formation of an oxime derivative to prevent multiple reaction products. By forming a propionic acid anhydride derivative, resolution of morphine and codeine from common interferences was greatly improved (Figure 3). Furthermore, we found that we were able to eliminate the oximation step, further reducing processing time. The SPE protocol presented here can also be used to prepare samples for LC/MS/MS. Samples prepared for analysis by LC/MS/MS did not require derivatization (Figure 4). LC/MS/MS has been reported to be at least one order of magnitude more sensitive than GC/MS for drugs of abuse. This would suggest that the volume of urine needed for SPE could be further reduced. If volumes of diluted urine could be reduced to about 1.5 ml, it would allow for the use of a 96-well plate SPE format, which can be easily adapted for automation. Automation could be used to streamline workflows and reduce error in sample processing.

Figure 1: Optimization of Wash & Elutions Conditions for SPE Original Urine Matrix Wash: Buffer Elute: EA/IPA/NH 4 OH Wash: HCl Elute: EA/IPA/NH 4 OH Wash: Buffer Elute: Methanol Wash: HCl Elute: Methanol

Figure 2: GC/MS Analysis of Samples Derivatized Using BSTFA 8 11 2,3 7 1 4 5 6 9 10 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60 min Column: Zebron ZB-Drug-1 Dimensions: 10 meters x 0.18 mm x 0.18 µm Part No.: 7CD-G023-08 Injection: Split 10:1 @ 260 C, 1 µl Carrier Gas: Helium @ 0.7 ml/min (constant flow) Oven Program: 180 C to 340 C @ 20 C/min Detection: MS @ 320 C Sample: 1. Morphine-d3 2. Morphine 3. Norcodeine 4. Codeine-d6 5. Codeine 6. Oxymorphone 7. Hydromorphone-d6 8. Hydromorphone 9. Oxycodone-d6 10. Oxycodone 11. Hydrocodone

Figure 3: GC/MS Analysis of Samples Derivatized Using Propionic Acid Anhydride 1 6,7 2 3 5 9 4 8 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 min Column: Zebron ZB-Drug-1 Dimensions: 10 meters x 0.18 mm x 0.18 µm Part No.: 7CD-G023-08 Injection: Split 10:1 @ 260 C, 1 μl Carrier Gas: Helium @ 0.7 ml/min (constant flow) Oven Program: 185 ºC to 320 ºC @ 30 ºC/min for 3 min Detection: MS @ 320 C Sample: 1. Hydrocodone 2. Codeine-d6 3. Codeine 4. Oxycodone 5. Hydromorphone 6. Morphine-d3 7. Morphine 8. Oxymorphone 9. Norcodeine

Figure 4: LC/MS/MS Analysis Without Derivatization 7 1,2,4 3 4 5 6 8 9 App ID 18846 0 0.5 1 1.5 2 min s Column: Kinetex 2.6 µm C18 100 Å Dimensions: 50 x 2.1 mm Part No.: 00B-4462-AN Mobile Phase: A: 0.1 % Formic acid in H 2 O B: 0.1 % Formic acid in Acetonitrile Gradient: 95:5 (A/B) for 0.24 min then linear to 70:30 over 1 min, hold for 1.25 min, then to 5:95 for 0.3 min, then to 95:5 for 1.8 min Flow Rate: 450 µl/min Temperature: 25 C Detection: Mass Spectrometer (MS) Sample: 1. Morphine-d3 (IS) 2. Morphine 3. Oxymorphone 4. Hydromorphone 5. Norcodeine 6. Codeine-d6 7. Oxycodone 8. Hydrocodone 9. Codeine

Table 2: Mass Ions for Chromatographic Analysis GC/MS BSTFA GC/MS Propyl LC/MS/MS Codeine-d6 *377, 349 *361,288,235 MRM 306.4>*152.0 Codeine *371, 343, 234 *355,282,229 MRM 300.4>*152.0 Hydrocodone *386, 297, 371 *299,242,214 MRM 300.4>*199.1 Hydromorphone-d6 *361, 450 *291,347 Hydromorphone *355, 444, 429 *285,341,228 MRM 286.2>*185.2 Morphine-d3 *432, 417 *344,271,400 MRM 289.2>*152.2 Morphine *429, 414, 401 *341,268,397 MRM 286.2>*152.2 Norcodiene *429, 414, 356 *223, 397, 101 MRM 286.1>*152.2 Oxycodone-d6 *480, 465 *377,320 Oxycodone *474, 459, 460 *371,314,298 MRM 316.4>*298.5 Oxymorphone *532, 517, 287 *357,300,413 MRM 302.3>*227.0

Conclusion A new procedure for the extraction and analysis of opiates from urine was developed that dramatically reduced overall sample processing time and improved chromatographic results by either GC/MS or LC/MS/MS. Trademarks Kinetex, Strata-X, and Zebron are trademarks of Phenomenex, Inc. Applied Biosystems is a registered trademark, and API 3000 is a trademark of Life Technologies Corporation and its affiliated company Applied Biosystems. 2009 Phenomenex, Inc. All rights reserved.