A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

Similar documents
Adopting DAP as a dose metric in CT

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Cardiac CT - Coronary Calcium Basics Workshop II (Basic)

Computed tomography Acceptance testing and dose measurements

The radiation dose in retrospective

Technology Assessment Institute: Summit on CT Dose Cardiac CT - Optimal Use of Evolving Scanner Technologies

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial

Combined Anatomical and Functional Imaging with Revolution * CT

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

Cardiac CTA Prospective Gating Broad Beam

Radiation Dose Reduction Strategies in Coronary CT Angiography

Assessment of effective dose in paediatric CT examinations

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Patient dose assessment of CT perfusion scanning at the RSCH

Dose Reduction Options in Cardiac CT

FDG-18 PET/CT - radiation dose and dose-reduction strategy

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Skyscan 1076 in vivo scanning: X-ray dosimetry

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Radiation Burden in Cardiology. A quick tour! David Sutton

Low Dose Era in Cardiac CT

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

Why is CT Dose of Interest?

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Introduction and Background

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Estimating Patient Radiation Dose from Computed Tomography

Asymmetric breast dose in coronary angiography

Quality Control and Patient Dosimetry on line for Computed Tomography

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

National Cancer Institute

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

Disclosure Information

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review

CT Radiation Risks and Dose Reduction

Estimation of Organ and Effective Doses for Neonate and Infant Diagnostic Cardiac Catheterizations

Low-dose and High-resolution Cardiac Imaging with Revolution CT

4D PET: promises and limitations

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Managing Patient Dose in Computed Tomography (CT)

CT NUMBER ACCURACY ANALYSIS FOR RADIOTHERAPY TREATMENT PLANNING IMAGING

CT Dosimetry in the Clinical Environment: Methods and Analysis

Patient doses from X-ray computed tomography examinations by a single-array detector unit: Axial versus spiral mode

Survey of patients CT radiation dose in Jiangsu Province

Accounting for Imaging Dose

organizations including the International Commission on Radiation Protection exposure dose management system to all the studies and

Multislice CT versus cone beam CT: dosimetric and image quality comparision.

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

Radiation Dosimetry for CT Protocols

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography

Reduced exposure using asymmetric cone beam processing for wide area detector cardiac CT

PACS: ERGONOMIC CONSIDERATIONS 1

State of the art and future development for standardized estimation of organ doses in CT

Application of 3D Printing to Molecular Radiotherapy Phantoms. Nick Calvert Nuclear Medicine Group The Christie NHS Foundation Trust, Manchester

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Exploration the Method of Low Dose Coronary Artery Imaging with Dual-Source CT

GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT

Typical PET Image. Elevated uptake of FDG (related to metabolism) Lung cancer example: But where exactly is it located?

Radiation Dose Optimization in Cardiac CT: A Technical Review. CHENG Wai Kwong. 17 May Contents

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

A multicentric study on patient dose in multislice CT

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens

Cardiac CTA without and with IV Contrast

Validation of CT Perfusion Imaging Against Invasive Angiography and FFR on a 320-MDCT Scanner

Chief Radiographer TEI Clinical Associate 2016

Radiation Protection- Cath lab

Les Outils Cliniques de Demain en Scanner Cardiaque. Cardiaque Status en ECR 2018 From Diagnosis to Prognosis

X-Ray & CT Physics / Clinical CT

SYNTAX III REVOLUTION Trial Press briefing conference. Prof. Patrick W. Serruys MD, PhD Principal Investigator Imperial College of London

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

CT CARDIAC ANGIOGRAPHY. patient information

Bone Densitometry Radiation dose: what you need to know

Electron Beam CT of the Heart

Multisclice CT in combination with functional imaging for CAD. Temporal Resolution. Spatial Resolution. Temporal resolution = ½ of the rotation time

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation

Bottom up cardiac CT for CABG assessment to resolve breathing artefact

Patient Dosimetry in Mammography and Tomosynthesis:

CT Quality Control Manual FAQs

Fetal Dose Calculations and Impact on Patient Care

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology

Coronary CT angiography with prospective ECG-triggering: an effective alternative to invasive coronary angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Cardiac CT saves money and time as first-line heart test 4/1/2008

Transcription:

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: Rob Loader Oliver Gosling

Introduction Approached by Dr Oliver Gosling (Research Physician, RD&E) to help perform comparison of radiation doses from CTCA with conventional cardiac angiographic techniques within cardiac catheterisation labs. Work was strongly supported by Dr Carl Roobottom (Consultant Radiologist) and Dr Gareth Morgan Hughes (Consultant Cardiologist). Aim was to calculate system & technique specific dose conversion factors to apply to a larger patient population to enable an effective dose to be estimated to compare the two complex imaging modalities.

Its all be done before...?? Conversion factors do exist from previous studies but rely on dated, non specific techniques and limited conversion factors range by ~100% Introduction of prospective gating in CTCA was thought to have significantly reduced radiation doses in CT compared to catheter based angiography.

Conventional angiography (Fluoroscopy & Fluorography)

Calculation of radiation dose for Conventional Cardiac angiography Use of computer based Monte Carlo simulator (PCXMC), integration of many projections for high dose runs. Input values specific to Derrifords imaging systems (matched for filtration etc) Involved audit, key dose parameters and patient BMI recorded by cath lab staff.

PCXMC

Conversion factor for IA

CTCA (64 slice gated computed tomography) 4 Beat Cardiac CTA 0.35 sec scan speed 175ms temporal resolution 4.06 sec scan Heart Rate 60bpm

CTCA Protocol (Derriford Hospital) Heart rate of < 65bpm for P-Gating Beta blockers given to achieve if required CTCA 90 ml of visipaque 320 administered via a cannula in the antecubital fossa at a rate of 6mls/second followed by 50mls of N-saline. 64X0.625mm SW (40mm, 35mm interval) X-ray window at 75% R-R interval (no additional padding). Tube current/kv set according to BMI

Prospective gated CT Dosimetry Use of IMPACT Dosimetry calculator (Monte Carlo based program) Averaging of peripheral measurements and assumes a full rotation per scan. Use of updated organ weighting factors (ICRP 103). Comparison of Effective dose with old organ weighting factors (ICRP 60) Effective dose calculated for all clinically used cardiac protocols (prospective gating only) Identified need for partial rotation Monte carlo simulator

Example

GE VCT Partial rotation scanning Cardiac modes, X-rays on for 180 T=0.35s (cardiac modes) Dicom image header states exposure time of 227ms Originally thought this did not make sense, as 227ms/350ms = 0.65 ~234 (not 180 ) Jacky Bye (GE) kindly reminded me that CT uses a fan beam!

GE VCT Irradiation Actual irradiation = 180 +Fan angle Verified using Gafchrome film XCT Start point relative to patient completely governed by ECG

Radiation exposure during CTCA rotation (227ms exposure)... A dosimetry Headache!

Overlapping axial pitch and dosimetry

A different approach to CTDIw Arbitrary start location of X-ray meant we had to rethink the measurement of CTDIw (look at variance of successive measurements!)

Calculation of nctdiw & DLP Filter bpm (set) kv C N S E W Mean Periph CTDIw (mgy/mas) Scanner CTDIv (for the actual nctdiw mas) Our measure Our measure (DLP) Scanner DLP % differenc e DAP 1st Cardiac medium 60 120 2.570 4.410 4.560 5.270 5.050 4.823 0.082 8.155 Method Cardiac Small 60 120 2.289 4.294 4.138 5.097 4.857 4.597 0.077 7.665 10.67 9.57 38.28 37.36-2.41 Cardiac Large 60 120 2.376 5.928 4.625 6.041 5.050 5.411 0.088 8.811 Cardiac medium 60 100 1.478 3.819 2.865 3.344 2.947 3.244 0.053 5.318 7.59 6.64 26.56 26.57 0.05 60 2nd Cardiac medium 60 120 2.379 7.155 2.505 6.589 1.633 4.471 0.083 8.313s.d Method Cardiac medium 60 120 2.379 5.550 6.104 6.637 0.598 4.722 0.087 8.683 Cardiac medium 60 120 2.379 1.502 6.366 2.999 6.644 4.378 0.082 8.177 0.26 Cardiac Large 60 140 3.425 9.698 5.307 9.216 1.635 6.464 0.120 12.009 Cardiac Large 60 140 3.425 3.511 8.834 3.093 9.283 6.180 0.116 11.592 Cardiac Large 60 140 3.425 4.776 8.758 9.160 2.167 6.215 0.116 11.644 Cardiac Large 60 140 3.425 9.997 3.087 3.191 9.308 6.396 0.119 11.909 0.20 Cardiac Small 60 80 0.578 1.989 1.717 0.105 2.014 1.456 0.026 2.564 Cardiac Medium 60 100 1.455 4.347 3.619 0.299 4.434 3.175 0.057 5.731 Cardiac Medium 60 80 0.637 2.479 2.015 0.136 2.477 1.777 0.031 3.077 Cardiac Large 60 120 2.245 7.074 1.125 6.269 3.836 4.576 0.084 8.369 Cardiac Large 60 120 2.245 7.093 1.052 4.057 6.346 4.637 0.085 8.459 Cardiac Large 60 120 2.245 7.094 0.993 4.251 6.313 4.663 0.085 8.497 Cardiac Large 60 120 2.245 6.150 5.390 6.475 0.610 4.656 0.085 8.487 Cardiac Large 60 120 2.245 5.737 5.693 0.591 6.596 4.654 0.085 8.484 0.05 s.d for pixel value (ROI) Cardiac Large 50 120 2.339 6.909 3.010 6.424 1.525 4.467 0.083 8.278 70.3 Cardiac Large 80 120 2.244 6.511 4.847 6.462 0.718 4.635 0.085 8.455 67.6 Cardiac Large 150 120 2.192 6.942 2.608 6.415 1.746 4.428 0.081 8.113 67.9

Chamber cross calibration performed Cardiac Medium 80 KV 500mAs mean s.d 5.421 5.461 5.419 5.434 0.024 5.446 5.465 5.461 5.457 0.010 5.1 5.102 5.106 5.103 0.003 5.147 5.149 5.158 5.151 0.006

Patient population & Results CTCA (PROS G) IA (Catheter Labs) N 84 94 Medin DLP/DAP 159±47.7 (mgycm) 27.2±12.3 (Gy*cm^2) Median (E msv) 5.4 6.3 IQR (3.5-5.9) (4.2-8.2) BMI 27±3.8 27±3.6

Results

Preliminary results

Organ (Breast) dose with PG CTCA The X-ray tube in conventional (catheter) angiography is positioned under the examination table to reduce the absorbed dose to the breast and to limit the scattered radiation dose to staff. In PG CTCA we have shown the x-ray exposure time is a fraction of the rotation time and triggered by ECG (essentially random). This means that the total absorbed dose to the breasts will vary significantly.

Breast dose Doses shown are for a medium size (70Kg) patient (cardiac medium filter, 120KV, 550mA) using a 32cm CTDIw phantom to represent the patient. Doses are for a single axial rotation. For the average size heart, 4 axial rotations are required to cover the heart (~140mm) Larger patients will require higher KV, ma combinations and larger patients/ breasts will be positioned closer to the X-ray tube. For large female patients, requiring higher tube current it is estimated that the absorbed dose to the breast could exceed 100mGy/scan under certain conditions.

Conclusions & Further work Radiation doses for prospectively gated CTCA have dropped significantly and dose audit from this study places effective doses from CTCA similar (if not lower) to that from those performed within the catheter labs. For the average size patient, Effective dose can no longer be a reason to perform coronary angiography over CTCA (prospective gated). More work is needed in order to ensure breast dose in CTCA is as low as possible and this may require the introduction of Bismuth breast shielding.

Limitations Effective dose contribution to CTCA from scout view and unenhanced scan not accounted for Patient populations not identical Different anthropomorphic phantoms (PCMXC vs ImPACT dosimetry calculator) Patient size not accounted for in ImPACT calculator. ImPACT calculator assumes uniform dose distribution in body.

Questions?