HPLC Analysis with Fluorescence Detection of Chlorophyll Degradation Products Pheophytins and Pyropheophytin in Virgin Olive Oil

Similar documents
Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Increasing resolution using longer columns while maintaining analysis time Advantages of the wide power range of the Agilent 1290 Infinity LC System

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine-

High-Resolution Analysis of Intact Triglycerides by Reversed Phase HPLC Using the Agilent 1290 Infinity LC UHPLC System

Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Maximizing Efficiency Using Agilent InfinityLab Poroshell 120 Columns

Analysis of Counterfeit Antidiabetic Drugs by UHPLC with the Agilent 1220 Infinity Mobile LC

Performance Characteristics of the Agilent 1290 Infinity II Multicolumn Thermostat

Chiral Screening for SFC and UHPLC with the Agilent 1260 Infinity II SFC/UHPLC Hybrid System

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Application Note. Authors. Abstract. Food

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

Application Note. Agilent Application Solution Analysis of fat-soluble vitamins from food matrix for nutrition labeling. Abstract.

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Analysis of HMF by HPLC

Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system

Authors. Abstract. Introduction. Environmental

Ultrafast analysis of synthetic antioxidants in vegetable oils using the Agilent 1290 Infinity LC system

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Agilent 1260 Infinity Analytical SFC System

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

Qualitative and quantitative determination of phenolic antioxidant compounds in red wine and fruit juice with the Agilent 1290 Infinity 2D-LC Solution

Application Note. Abstract. Authors. Pharmaceutical

HPLC to UHPLC Transfer of USP Method for Amlodipine Besylate Using the Agilent 1290 Infinity II LC

Determination of Adulterants in Olive Oil Analysis of Waxes and Fatty Acid Methyl and Ethyl Esters with Capillary GC and Agilent OpenLAB Software

Application Note. Author. Abstract. Introduction. Food Safety

PAPRIKA EXTRACT SYNONYMS DEFINITION DESCRIPTION FUNCTIONAL USES CHARACTERISTICS

2D-LC as an Automated Desalting Tool for MSD Analysis

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement

Analysis of bisphenol A leaching from baby feeding bottles

Automated Sample Preparation for FAME Analysis in Edible Oils Using an Agilent 7696A Sample Prep WorkBench

Application Note. Abstract. Author. Biotherapeutics & Biosimilars. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Performance characteristics of the Agilent 1100 Series preparative pump. Technical Note. Abstract

Fat-Soluble Vitamins Analysis on an Agilent ZORBAX Eclipse PAH Polymeric C18 Bonded Column

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

using the Agilent 7696A Sample Prep

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

Independent Column Temperature Control Using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Extended Application Note

EASIMIP TM PATULIN Product Code: P250 / P250B

application Natural Food Colorants Analysis of Natural Food Colorants by Electrospray and Atmospheric Pressure Chemical Ionization LC/MS

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

Application Note. Authors. Abstract. Petrochemical

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Confident, lower-pressure analysis of carbohydrates, alcohols, and organic acids. Agilent Hi-Plex Ligand-Exchange HPLC Columns

Analysis of Vitamins Using an SFC/UHPLC Hybrid System with a Triple Quadrupole LC/MS for Quantification

Quality Analysis of Extra Virgin Olive Oils Part 7

Chiral Multicolumn Method Development on the Agilent 1260 Infinity II SFC System

High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column

Ultra High Performance Liquid Chromatograph. Nexera X2 C196-E079

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Author. Introduction. Abstract

A Novel HILIC Column for High Speed N-linked Glycan Analysis

USP purity analysis of pravastatin sodium using the Agilent 1120 Compact LC

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Comparison of a UPLC Method across Multiple UHPLC Systems

High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS)

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography

HPLC Analysis of Sugars

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Quantitative Analysis of Amphetamine-Type Drugs by Extractive Benzoylation and LC/MS/MS. Application. Introduction. Authors. Abstract.

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

Online 2D-LC Analysis of Complex N-Glycans in Biopharmaceuticals Using the Agilent 1290 Infinity 2D-LC Solution

Determination of 44 Pesticides in Foodstuffs by LC/MS/MS Application

Author. Introduction. Small Molecule Pharmaceuticals & Generics

Relative Measurement of Zeaxanthin Stereoisomers by Chiral HPLC

EASI-EXTRACT BIOTIN Product Code: P82 / P82B

ZIDOVUDINE, LAMIVUDINE AND ABACAVIR TABLETS Draft proposal for The International Pharmacopoeia (September 2006)

The challenging extraction of non-polar contaminants out of a non-polar vegetable oil sample

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods

Application Note. Authors: C. Ledesma, M. Gibert, J.R. Gibert Ingenieria Analitica S.L. Extracts from various food products

Determination of Aflatoxins in Food by LC/MS/MS. Application. Authors. Abstract. Experimental. Introduction. Food Safety

STANDARD OPERATING PROTOCOL (SOP)

DRAFT TANZANIA STANDARD

Direct Analysis of Folic Acid in Digestive Juices by LC/TOF-MS Application

Product Guide for LudgerSep TM ur2 UHPLC Column for DMB Sialic Acid Analysis

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

STANDARD OPERATING PROTOCOL (SOP)

Analysis of Organic Acids and Alcohols Using the Agilent J&W DB-624UI Ultra Inert GC Column

Product Guide for LudgerSep TM R1 HPLC Column for DMB labelled Sialic Acid Analysis

Development and Validation of RP-HPLC Method for the Estimation of Gemigliptin

The HPLC Preparative Scale-Up of Soybean Phospholipids Application

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

APPLICATIONS Improving Intact Biogeneric Protein Separations with Aeris WIDEPORE Core-Shell Columns

Fast determination of residual glycerol and glycerides in biodiesel by SFC/MS using the Agilent 1260 Infinity Analytical SFC System

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS

CORESTA Recommended Method No. 78

Transcription:

HPLC Analysis with Fluorescence Detection of Chlorophyll Degradation Products Pheophytins and Pyropheophytin in Virgin Olive Oil Application Note Authors Xueqi Li, Selina Wang UC Davis Olive Center Robert Mondavi Institute for Wine and Food Science Davis, CA 95616 Michael Woodman Agilent Technologies, Inc. Suite 550, 10 N. Martingale Road Schaumburg, IL 60173 Abstract This application note shows a simple and rapid method for the analysis of the chlorophyll thermal degradation products pheophytins (PP) and pyropheophytin a (PPP) in extra virgin olive oil (EVOO) by modifying the photomultiplier (PMT) of a fluorescence detector (FLD). By exchanging the standard PMT with a red-sensitive Hamamatsu R928HA PMT, the raw increase in signal was approximately 120-fold. In addition, an overall improvement of 180-fold was observed during analysis due to reduced baseline noise. A straightforward direct analysis of diluted EVOO was performed without further sample preparation in this method, which provides less solvent consumption and shorter LC cycle times, with good resolution, over the traditional International Standards Organization (ISO) solid phase extraction (SPE)/ultraviolet detector (UV) method.

Introduction EVOO is obtained from olives (Olea europaea L.) through a mechanical procedure. Chlorophylls are the natural green pigments existing in EVOO, and are prone to degrade to pheophytins and pyropheophytin as a result of processing and storage. Chlorophylls, particularly, can be affected by acid conditions and heating during fruit milling and paste beating, thus greater accessibility between components and the generalized release of the acid compounds produce pheophytinization in great extension [1]. This leads to the use of the degradation products of chlorophylls as one of the quality parameters for EVOO. The ISO method of PPP analysis was adopted by Australia and New Zealand as part of olive oil standards in 2010, but has yet to be adopted in any other country. With the obtained values, it can be determined if an EVOO was properly stored or was adulterated with refined olive oil. The traditional SPE/UV method in ISO 29841 includes several steps in sample preparation that are very solvent-consuming and time-consuming. In this modified FLD method, an isopropanol diluted olive oil sample was ready for the HPLC injection. A quaternary solvent gradient method was used to include a fourth strong solvent wash (tetrahydrofuran) on a quaternary gradient pump, which eliminated the need to premix any solvents and also greatly reduced the oil residues on the column from previous analysis. Furthermore, the FLD was enhanced more than 180 by exchanging the standard PMT with one more sensitive in the red region. Experimental Instrument configuration ISO 29841 SPE/UV method An Agilent 1290 Infinity Binary LC system, controlled by ChemStation (Rev. C.01.03 [44]), and consisting of the following modules was used. A diode array detector (DAD) was used as the UV detector in the original ISO method: An Agilent 1260 Infinity Quaternary LC system, controlled by ChemStation (Rev. C.01.03 [44]), and consisting of the following modules was used: Agilent 1260 Infinity Quaternary Pump (G4204A) Agilent 1290 Infinity Autosampler (G4226A) Agilent 1290 Infinity Thermostatted Column Compartment (G1316C) Agilent 1260 Infinity Fluorescence Detector (G1321B), equipped with Hamamatsu R928HA PMT Sample preparation All solvents used were LC grade (same in LC conditions). Fresh Milli-Q water was obtained from a Milli-Q Integral system equipped with a 0.22-µm membrane point-of-use cartridge (Millipak). Approximately 300 mg of olive oil sample was dissolved in 1 ml of petroleum ether and vortexed to mix them well. The sample solution was applied onto the 1,000 mg/6 ml silica cartridge (Phenomenex, Torrance, CA, USA) and rinsed twice with 1 ml portions of petroleum ether. As soon as the solvent had drained to the top of column packing, the nonpolar substances were eluted twice with 5 ml petroleum ether/diethyl ether (90:10, v/v) and discarded. Then, the pheophytin fraction was eluted with 5 ml of acetone and collected in a 10-mL glass tube covered with foil (protected from light). The solvent was evaporated to dryness on a rotary evaporator at max 20 C, and the residue was dissolved in 1 ml acetone for further HPLC analysis [2]. A 200 mg amount of olive oil sample was dissolved in 1 ml of reagent isopropanol by vortex for 15 seconds. Agilent 1290 Infinity Binary Pump (G4220A) Agilent 1290 Infinity Autosampler (G4226A) Agilent 1290 Infinity Thermostatted Column Compartment (G1316C) Agilent 1290 Infinity Diode Array Detector (G4212A) 2

LC conditions The LC chromatographic conditions of the ISO 29841 SPE/UV method are listed in Table 1. The original 4.6 250 mm C18 column was modified for a 3.0 100 mm C18 column in this situation. An isocratic method was used. Table 1. Chromatographic Conditions of ISO SPE/UV Method Stationary phase Agilent ZORBAX Eclipse Plus C18 column, 3.0 100 mm, 1.8 µm Eluting mobile phase Water/methanol/acetone (4:36:60, v/v/v) Flow rate 0.5 ml/min Volume injected 5 µl Detector Diode array detector, 410 nm Results and Discussion Figure 1 shows an overlay chromatogram for the two ISO sample preparation and detection schemes, SPE/UV (in blue) and dilution/fld (in red), using the modified chromatographic conditions newly developed for the dilution/fld variation. The obtained PPP values from both methods were 15% (calculated based on Equation 1). Both sample separation methods showed good resolution on HPLC, and resolution of pheophytin a and pheophytin a is significantly improved over the specified ISO method chromatographic conditions. mau 100 80 Pheophytin a 60 The sample was run with the LC system using the chromatographic conditions listed in Table 2, and following the quaternary solvent gradient shown in Table 3. 40 20 0 Pheophytin a Pyropheophytin a Table 2. Chromatographic Conditions of Modified FLD Method Stationary phase Agilent ZORBAX Eclipse Plus C18 column, 3.0 100 mm, 1.8 µm Mobile phase Gradient shown in Table 3 Volume injected 5 µl Detector A red-sensitive FLD, excitation 430 nm/emission 670 nm Table 3. Time (min) Quaternary Solvent Gradient of Modified FLD Method A Water % B C Tetrahydrofuran % Methanol % D Acetone % 0 4 0 76 20 0.5 4 4 0 36 60 0.5 14 4 0 36 60 0.5 14.2 0 100 0 0 1.0 20.9 0 100 0 0 1.0 21 4 0 76 20 0.5 25 End of test, ready to inject new sample Flow rate (ml/min) 4 Figure 1. 6 8 10 12 14 16 min An overlay chromatogram of both sample preparation methods (SPE and simple dilution) of an extra virgin olive oil with 15% PPP content in chlorophyll a degradation products. Conditions were those used for the modified fluorescence method. The sample preparation time of the is 30 minutes, with an HPLC analysis time of 40 minutes or longer. With the modified UHPLC/FLD method, on the quaternary system with a 3.0 100 mm C18 column, the dilution step is brief and the analysis time is reduced to 25 minutes. Solvent consumption and total analysis time for the UHPLC analysis are 40% or less when compared to the standard ISO conditions. It represents a significant reduction in analysis cost and time, and has demonstrated improved sensitivity for low level samples. % PPP = 100 Appp/(App + Appp) Where % PPP = % pyropheophytin a content in olive oil sample Appp = peak area of pyropheophytin a App = total peak area of pheophytin a and pheophytin a Equation 1. Calculation of the PPP content in olive oil. 3

When a lower PPP content EVOO sample was analyzed, the UV detection was insufficient in quantification as shown in Figure 2. A recently harvested EVOO sample was analyzed under both conditions. The PPP value gained from the ISO SPE/UV method was 0, which was theoretically inaccurate since the decarbomethoxylation from pheophytins to pyropheophytin occurs once olives are processed to oil. However, a small amount of PPP (6%) was still detectable under the modified FLD method. mau 100 80 60 40 20 0 4 Pheophytin a Pheophytin a Pyropheophytin a 6 8 10 12 14 16 min Figure 2. An overlay chromatogram of both methods of an EVOO with 6% PPP content in chlorophyll a degradation products. Figure 3 further demonstrates the high selectivity and sensitivity of modified FLD method. An extra light olive oil was analyzed by both methods. Extra light olive oil is a blend of mostly refined olive oil and virgin olive oil. Thus, it has considerably low PPP content. As shown in Figure 3, the response signal under a DAD method was extremely low, with a high background noise. This prevented the PPP content quantification for the EVOO portion in this extra light olive oil. Conversely, with the modified FLD method, signals as low as 0.15 LU could be detected, and the PPP content was able to be further quantified by peak areas. mau 0.10 0.08 0.06 0.04 0.02 0 A 6 8 10 12 14 16 min LU B 10.107 0.30 Pheophytin a 0.25 14.962 0.20 11.078 Pyropheophytin a Pheophytin a 0.15 6 8 10 12 14 16 min Figure 3. Separation of the degradation products of chlorophyll a under both methods for the same extra light olive oil sample, ISO SPE/UV method (A) and modified FLD method (B). 4

Replicate analyses using the specified gradient and column with two different PMTs are summarized in Table 4. The raw increase in signal with the Hamamatsu R928HA was approximately 120-fold. The further improvement, due to reduced baseline noise, gave an overall improvement of 180-fold. Table 4. Summary of Improved Results after Changing the Photomultiplier Tube to the Suggested Agilent Alternate Pheophytin a, a, and pyropheophytin detection Summary of modified FLD analysis excitation 430 nm/emission 670 nm PMT Standard R928HA Peak area total* 6.8 854.5 Average Signal/Noise 34.8 6847.5 * Combined response of pheophytin a, pheophytin a and pyropheophytin Conclusions The modified FLD method has been shown to be a promising option to the traditional. In this case, low level PPP samples can be measured when the photomultiplier is exchanged from the standard PMT to the red-sensitive R928HA model. This results in a substantial cost savings in time, labor, SPE materials, and solvent consumption in any laboratory needing to analyze large numbers of samples. References 1. D. Hornero-Méndez, B. Gandul-Rojas, and M.I. Mínguez-Mosquera Routine and sensitive SPE- HPLC method for quantitative determination of pheophytin a and pyropheophytin a in olive oils Food Research International 38, 1067-1072, doi:10.1016/j.foodres.2005.02.022 (2005). 2. Vegetable fats and oils - Determination of the degradation products of chlorophylls a and a (pheophytins a, a, and pyropheophytins), ISO 29841:2009. For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. 5

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2014 Printed in the USA April 10, 2014 5991-4384EN