DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

Similar documents
PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

A Standardised Approach to Optimisation

Doses Received by Patients during Thorax X-Ray Examinations

Radiologic Units: What You Need to Know

Radiation exposure of the Yazd population from medical conventional X-ray examinations

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

A Survey of Radiation Exposure Techniques Factors Used for Common Diagnostics X- Ray Examinations in Akwa Ibom State, Nigeria

THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals

Dosimetric Consideration in Diagnostic Radiology

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine

RADIATION SAFETY. Junior Radiology Course

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

Patient dose in routine X-ray examinations in Yazd state

International Journal of Radiation Research, January 2017

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology

Laboratory Safety 197/405. Types of Radiation 198/405

PRINCIPLES AND METHODS OF RADIATION PROTECTION

Entrance Surface Dose from Pediatric Diagnostic X-ray Examinations in a Developing World Setting: Are We ALARA Principle Compliant?

Olowookere, C.J. Dept of Physics Ajayi Crowther University, Oyo

Radiation Doses to Patients Undergoing Selected Radiographic Examinations in Three Hospitals South-West, Nigeria

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Skyscan 1076 in vivo scanning: X-ray dosimetry

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Survey of Radiation Dose Levels in Patients in X-Ray Units of Some Selected Hospitals in Jos Metropolis

The x-rays produced penetrate the body which absorbs, refracts, or reflects the x-ray beam energy depending on the tissue. Bone

1 Course Syllabus + Study Guide for Lecture and Laboratory

Course Syllabus + Study Guide for Lecture and Laboratory

Shielding Calculation: Radiographic Room. Jerry Williams

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

Radiation Quantities and Units

Twelfth Annual Warren K. Sinclair Keynote Address

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations

Table of Contents. Introduction 3. Background 4

OPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY

Measurement of Absorbed Doses in Anatomical Phantoms

Patient Management Image Selection Radiation Biology, Dosimetry & Protection

PNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY

The activity. Suggested Answers

Radiation Safety Guide. Analytical X-Ray Equipment

Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria

Code of Practice for Radiation Protection in Dentistry. Code of Practice For Radiation Protection in Dentistry

Equivalent dose to organs and tissues in Hysterosalpingography calculated with the FAX (Female Adult voxel) phantom

Dose Estimates for Nuclear Medicine Procedures: What are they? Where do they come from?

ABSTRACT I. INTRODUCTION

The impact of SID and collimation on backscatter in radiography

Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February Radiation Safety

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review

Mapping the ASRT Objectives for Radiation Protection (47 Objectives) and Radiation Biology (21 Objectives) to this Text

Ionising radiation is EM radiation that causes ionisation of atoms. The minimum energy needed to ionise any atom is 12 ev.

Radiation Safety. - Orientation Session - Environmental Health, Safety & Risk Management Health Physics Services Unit

Establishment of local diagnostic reference levels for quality control in intraoral radiography

Radiation Protection Dosimetry Advance Access published April 22, 2008

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Introduction to Radiation Protection

Quiz True/False: Large amounts of radiation to insects will cause them to mutate!

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

MA 2030 Radiography Skills for Medical Assistants

Radiation Dose Specification

PERSONNEL MONITORING AND DOSIMETRY POLICIES

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

Dental Radiography STANDARDS & GUIDELINES TABLE OF CONTENTS. College of Dental Surgeons of BC

Radiation Safety in the Catheterization Lab

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Radiation Protection in Laboratory work. Mats Isaksson, prof. Department of radiation physics, GU

A Simple Model for Establishing Exemption Limits

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Radiation Safety For Anesthesiologists. R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua

Background Radiation in U.S. ~ msv/yr msv/yr ~0.02 ~0.02 msv msv/day /day (~2 m rem/day) mrem/day) NCRP 4

UQ X-ray Safety Training Module

Radiation Safety Manual

Practice and Risk at Medical Facilities in Agency Operations

Biological Effects of Ionizing Radiation & Commonly Used Radiation Units

Quality Control Status of Radiology Centers of Hospitals Associated with Mashhad University of Medical Sciences

Radiology Review Course Hotel del Coronado Coronado, California

ORAU Team Dose Reconstruction Project for NIOSH

Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran

Radiation Protection- Cath lab

Assessing the standard dose to standard patients for x-ray investigations

Chem 481 Lecture Material 3/11/09

THE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS

Biological Effects of Ionizing Radiation & Commonly Used Radiation Units

IONISING RADIATION REGULATIONS 99

Entrance Surface Air Kerma for Skull, Pelvis and Abdomen X-ray Examinations in Some Diagnostic Radiology Facilities in Akwa Ibom State, Nigeria

Introduction. Chapter 15 Radiation Protection. Advisory bodies. Regulatory bodies. Main Principles of Radiation Protection

Hand Dose in Nuclear Medicine Staff Members

Research Article. Corresponding author H. Osman


in developing institutional policies, procedures, and /or protocols. The Canadian Society of

Iranian Journal of Medical Physics

Variation of Occupational Doses among Subspecialties in Diagnostic Radiology. A.N. Al-Haj, C.S. Lagarde, A.M. Lobriguito

Ionizing Radiation. Michael J. Vala, CHP. Bristol-Myers Squibb

Evaluation of radiation dose and image quality following changes to tube potential (kvp) in conventional paediatric chest radiography

Radiation Safety General Awareness and ALARA Training

RADIATION HAZARDS AND SAFETY

Transcription:

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA Full Length Research Article 1 Ibrahim, U, 3 Daniel, I.H., 3 Ayaninola, O., 4 Ibrahim, A., 5 Hamza, A.M. and 6 Umar, A.M. 1,3,4,5 Nasarawa State University Keffi, Nigeria 2 Kaduna State University Kaduna, Nigeria. 6 Bauchi State University Bauchi Nigeria 2 Corresponding author email: Isaac.daniel@kasu.edu.ng ABSTRACT A study was carried out to establish the trend of dose received by patient during x-ray examination in Federal Medical Centre, Keffi in Nasarawa state, Nigeria. Entrance skin doses (ESDs) for a common type of x-ray procedures, namely chest AP/PA (anterior/posterior) were measured. A total of 200 data were collected from patients who were exposed to diagnostic X-ray during their routine chest X-ray examinations. The age of the patients ranged from 15 to 68 years old while the weight and height of these patients ranged from 37.5kg to 98.5kg and 130.0cm to 175cm, respectively. The patent s skin dose were determined using Edmond s formula, which is based on the X-ray tube and the radiographic exposure parameters of kvp, mas, SSD and the total filtration of the beams. The calculated mean skin dose ranges from 0.013 ± 0.01mGy to 0.851±0.023mGy. In general, the ESDs measured for this type of x-ray procedures were found to be lower than or in agreement with the guidance level set by the Nigerian Basic Ionizing Radiation Regulation (NBIRR, 2003) standard and other international bodies and does not pose any significant health risk to the patience or the workers. Keywords: Entrance skin Dose, X-ray, Anterior-Posterior, Exposure, and Radiation. INTRODUCTION In medicine, ionizing radiation is used for two main purposes; diagnosis and therapy. Consequently, individuals and the populace at large receive significant exposure to radiation. Diagnostic radiology is a leading cause of man-made radiation exposure to the population. It was estimated that diagnostic radiology and nuclear medicine contributed 96% to the collective effective dose from man-made sources in the U.K (National Radiology Protection Board, (NRPB, 1993). Similar estimate showed that this contribution was 88% in the U.S.A (National Council on Radiation Protection and Measurement (NCRP, 1987).The health of the population would decline if ionizing radiation techniques were not available to diagnose disease and detect trauma. Nevertheless, there is no excuse for complacency and it is a basic premise of radiation protection practice that any exposure should be justified by weighing the potential harm against the perceived benefit. In view of the significant benefits from properly conducted medical exposures, the principal concern in radiological protection is the reduction of examinations that are either unlikely to be helpful to patient management or involve those that are not as low as reasonably achievable in order to meet specified clinical objectives. In order to do this, there is a need to optimize x-ray equipment and logical techniques (NRPB 1990). Patient dose measurement is an integral part of this optimization procedure (Faulkner, et al 1999). Such measurements will reveal x-ray facilities with doses after which possible dose reduction measures may be specified. Dose measurement is also necessary so as to: establish dose constraints, determine risk to patient and to justify the examination. The current philosophy of the International Commission on Radiological Protection (ICRP) in medical practice is that any use of radiation should be justified. After justification, it is important to optimize the procedure. In radiography this means using as low a dose as reasonable to obtain an optimum image of diagnostic quality. There are two categories of doses to patient which are important in diagnostic radiology; the effective dose E, which takes into account of dose equivalent to radiosensitive organs and the entrance skin dose. Most interest in diagnostic radiology is concerned with effective dose since this relates to the risk of stochastic effect such as cancer induction. Drek, (2010) suggested that for public health management of radiation emergencies, one of the essential components of integrated risk assessment is to quickly and accurately assess and categorize the exposure.radiation safety, monitoring, and assessment have become issues of great concern, since at high doses, ionizing radiation is carcinogenic and clinical symptoms are known to be associated with low-dose exposure. The injuries and clinical symptoms induced by exposure to ionizing radiation include, direct chromosomal transformation, indirect free-radical formation, radiation catractogenesis, cancer induction etc. (Serro, 1992). Diagnostic X-rays are used for identifying diseases and other problems during medical examinations. The objective of any diagnostic X-ray examination is to produce images of patients 14

with essential details and sufficient image quality so as to guide practitioners for effective and efficient diagnosis and treatment of various disease conditions. Because of the risks associated with the exposure of the patients to X-rays during the diagnostic X-ray examinations, it is suggested that minimum amount X-radiation should be used and the entrance skin dose should be measured and monitored. It has been demonstrated by Edmonds, 1984 that X-radiation to patients depends on the exposure parameters of kvp, mas, SSD and the filtration. Standard chest examination consists of Posterior-Anterior (PA) Anterior Posterior (AP) and lateral chest X-ray. The films are read together in radiology department, a chest radiograph commonly called chest X-ray (CXR) is used. The two fundamental objectives of this research are; first to measure the patient entrance skin dose in order to have a means for setting and checking standards of good practice as an aid to the optimization of patient protection. Secondly, to estimate the absorbed dose to tissues and organs in the patient in order to determine the risk so that diagnostic techniques can be properly justified and cases of accidental over exposure will be properly investigated. CHEST RADIOGRAPHY Chest radiography, commonly called chest X-ray (CXR) is a projection radiograph of the chest used to diagnose condition affecting the chests, its content and nearby structure. Chest radiographs are the most common examinations taken as a therapeutic tool for many clinical conditions. It is used to diagnose many conditions involving the chest walls, bones of the thorax and structures within the thorax including the lungs, heart and great vessel. Pneumonia and congestive heart failure are very common diagnose by chest-radiograph. Chest radiograph are also used to screen for job related lung disease in industry such as mining where workers are expose to dust and like all methods of technique, chest radiograph employs ionizing radiation in the form of X-ray to generate image of the chest. The typical radiation dose to an adult from a chest radiograph is around 0.02mSv for a front (PA) view, or 0.04mSv for a side (lateral) view. Thus different views of the chest can be obtained by changing the relative orientation of the body and the direction of X-ray beams. A survey on 171 patients in three public hospitals in Nigeria undergoing chest, abdomen and lumbar spine x-ray examinations were carried out. The technical parameter used are tube potential, exposure FFD using TLD chips to measure the radiation dose which was found to varied widely. One unit exposed patient to whipping 120m red per x-ray (Ogundare, 2004). Investigation on the effect of exposure factors on the quality of the skull radiograph in Jos university teaching hospital and was found that the exposure factor to obtain a quality skull x-ray radiograph at minimum dose are 60-70kV, 300-50mA and 0.6 1.6s (Nwankwo et al., 2004). In Nigeria, Radiographers, Radiologist and Medical Physics are required to undergo a two weeks radiation protection compulsory training in order to handle or work with x-ray machines. This regulation was introduced by the NBIRR (Nigeria Basic Ionizing Radiation Regulation) under NNRA (Nigeria Nuclear Regulatory Authority) to ensure the safety of patients and workers. Radiation Dose to Tissue or Organ Expressing the size of a radiation dose is most conveniently done by specifying the amount of energy deposited by the incident radiation. The basic measure of radiation dose is called absorbed dose (Sato et al., 1995). Absorbed Dose The interaction of radiation with matter involves a transfer of energy from the radiation to the matter. Ultimately, the energy transferred either to tissue or to a radiation shield is dissipated as heat. The radiation dose depends on the intensity and energy of the radiation, the exposure time, the area exposed and the depth of energy deposition. The Absorbed Dose is given by: de Absorbed Dose (D) = (1) dm The unit is J/kg or Gray (Gy) formally rad. 1Gy = 100 rad. It is possible to calculate the absorbed dose in a material if the exposure is known. D (Gy) = f. x (Ckg -1 ). (2) Where f is the conversion coefficient depending on the medium (Sato et al., 1995) Equivalent Dose The absorbed dose does not give an accurate indication of the damage that radiation can do. An absorbed dose of 0.1Gy of alpha radiation, for example, is more harmful than an absorbed dose of 0.1Gy or beta or gamma radiation. To reflect the damage done in biological systems from different types of radiation, the equivalent dose is used. H = WR. D(Gy) (3) Where H is the equivalent dose in tissue, and WR is the radiation weighting factor (Sato et al., 1995). The unit is sievert (Sv) formally rem. 1Sv = 100rem. MATERIALS AND METHODS Patient Dose Survey The survey method in this work was based on the guidelines established by the NBIRR protocols. Federal Medical Centre, Keffi was used in this survey. In the x- ray machine used, specific data such as type, model, waveform, filtration, film-screen combination and output were recorded. In this paper, we only considered the posterior anterior (PA) view. For each patient and 15

x-ray unit, the following parameters were recorded: sex, age, weight, height, body mass index, focus-to-film distance (FFD), film size, chest thickness, kvp and mas. The diagnostic x-ray machine used in this project work is anode rotated with tungsten target. Chest radiograph examinations of some patient were considered. The x-ray examinations of patients were carried out using posterior-anterior radiographic view using grade. The x-ray exposure to patients was given from a control room where exposure parameters of kvp and mas were selected for each patient. The source to skin distance (SSD) was obtained for each patient from the focal film distance (FFD) and the thickness of the patient s chest. SSD = FFD thickness of the patient s chest (4) The radiograph of the chest involves two major viewings of positioning of patients during medical examinations; the posterior anterior (PA) view and the lateral view but in this investigation, we only considered the posterior anterior (PA) view. A loaded x-ray films cassette (35cm x 42cm) is fixed on an erect, Bucky or chest stand. The patients stand facing the erect Bucky with the chest in contact with the Bucky, the arms placed on both side of the hips with the shoulder rolled forward (this is to displace the scapulae from a lungs field). A horizontal x ray beam 90cm to the films is centered to the lower border of the scapulae. The beam is collimated to the area of interest. The exposure factors are carefully selected with an FFD of 150cm 180cm. The data collected mere based on the exposure parameters and the total filtration of the beam from the machines. Entrance skin dose is the maximum amount of x-radiation absorbed by living tissues during medical examinations. The skin dose to patients was determined by calculation from the x-ray tube parameters and exposure radiographic parameters using Edmonds (1984) skin dose formula which is given as: skin dose(μgy) = 418(KVp). mas. ( )^ Where kvp is the peak voltage responsible for the quality of penetration ma is the tube current, responsible for quantity of electrons from the filament, T is the total filtration of the beams always a constant (2.9mmAl), and SSD = FFD thickness. (5) RESULTS Table 1: Patient information and x-ray machine parameters. NO OF PATIENTS CHEST THICKNESS (cm) AGE RANGE (years) DOSE (mgy) WEIGHT (Kg) HEIGHT (cm) kvp (kv) 6 7.0 15-17 0.013 37.5-38.5 145-147 50.0 32.8 7 8.0 18-19 0.187 39.5-40.5 148-151 50.0 57.6 6 9.0 20-21 0.253 41.5-43.5 152-156 50.0 51.2 10 10.0 22-23 0.363 44.5-47.5 157-158 50.0 70.4 3 11.0 24-26 0.367 48.5-50.5 159-160 50.0 80.0 2 12.0 27-29 0.371 51.5-52.5 161-172 50.0 80.0 6 13.0 30-31 0.422 53.5-54.5 167-170 50.0 96.0 5 14.0 32-33 0.428 55.5-56.5 160-167 50.0 70.4 7 15.0 34-36 0.481 57.5-59.5 160-170 50.0 80.0 40 16.0 37-46 0.536 60.5-98.5 167-172 50.0 70.0 23 17.0 47-50 0.543 68.5-91.5 169-172 50.0 70.4 34 18.0 51-56 0.624 72.5-98.5 167-174 50.0 96.0 22 19.0 57-60 0.632 79.5-82.5 160-171 50.0 80.0 15 20.0 61-62 0.768 69.5-78.5 165-169 50.0 906. 7 21.0 63-64 0.777 62.5-70.5 171-178 50.0 80.0 5 22.0 65-66 0.787 65.5-68.5 165-174 50.0 70.4 2 23.0 67-68 0.851 58.5-67.5 168-172 50.0 80.0 mas (mas) 16

0.9 0.8 0.7 DOSE (mgy) 0.6 0.5 0.4 0.3 0.2 0.1 0 WEIGHT (Kg) Fig. 1: Chart Showing Absorbed Dose against Weight 30 mean chest thickness (cm) 25 20 15 10 5 0 0 1/8 2/8 3/8 3/8 3/8 3/8 3/8 4/8 4/8 4/8 5/8 5/8 6/8 6/8 6/8 7/8 DOSE (mgy) Fig.2: Chest Thickness against Dose Absorbed 17

Science World Journal Vol 8 (No 2) 2013 DISCUSSION In this survey, biographical data such as patient age, weight, height and machine parameters were recorded. This is shown in Table 1. The weight of the patients ranges from 37.5 to 98.5 kg and the average being around 65.0kg. The maximum absorbed dose recorded is 0.851mGy; these may be attributed to the patients chest thickness or ages which were also high. Hospital data were compared with the guidance levels set by the International Atomic Energy Agency (IAEA) and other international bodies. The histogram in figure 1 shows that the mean doses received by the patient increases with increased in weight and body mass. Similarly, figure 2 shows that the mean dose received by the patient varies with the patients chest thickness. Another influencing factor that causes high ESD obtained in this examination is probably due to the aging effect of X-ray tube. CONCLUSIONS The results presented in this research work indicate that the ESDs received by patients undergoing x-ray examination in medical centre Keffi were within the guidance levels of 0.4mGy set by radiation protection bodies. Considering the condition of the machine used for this examination, a reference skin dose of 1mGy can still be compared with the skin doses evaluated in this research. It is possible that the x-ray machine is properly maintained, although all x-ray machines have to pass through quality control evaluations after each three months before it be allowed to operate. Owing to the fact that the maximum absorbed dose recorded in this research is 0.851mGy which is a bit higher than the guidance level, we therefore have to make the following recommendations. RECOMMENDATION Under age patients undergoing radiological examinations should be given special considerations by cutting down doses using high kilovolt technique and lower mas. The ALARA (as low as reasonably achievable) principle should be used when carrying out x-ray activities. Consistency in Quality Assurance program and training of personnel will go a long way in reducing the radiation doses received by the patients. The estimation of stochastic risk to Keffi population as a result of X-ray diagnostic procedures and the establishment of dose reference levels in Keffi Medical Centre Radiological Unit are guide for future monitoring and studies. REFERENCES Drek, H.C.,May, C.C., and Zanat, C.(2010).Global Networking for BiodosimetryLaboratory capacity in Radiation Emergencies. Health Phys., 98(2),168-171. Edmonds,I.R.(1984). Calculation of patient skin dose from diagnostic x-ray procedurebritish journal of radiology, 57(2), 733-734. Faulkner, K.,Broadhead, D.A., and Harrison, R.M.(1999) Patient dosimetry measurement methods. Applied Radiation and Isotopes. 50, 113-123. International Commission on Radiological Protection (1996). Radiological Protection and Safety in Medicine, Publication 73 (Oxford and New York: Pergamon Press) National Council on Radiation Protection and measurement (1987). NCRP report No. 93, Ionizing Radiation Exposure of the population of the United States NCRP, Bethesda Maryland. National Radiological Protection Board (NRPB)/ Royal College of Radiologist (1990). Patient Dose Reduction in Diagnostic Radiology Documents of the NRPB London.HMSO, 1(3). National Radiological Protection Board, (NRPB). (1993) `` Radiation Exposure of the U.K. Population NRPB Report R263-1993 Review. NRPB, Chilton, U.K. NBIRR (Nigeria Basic Ionizing Radiation Regulation), 2003. Ogundare, I.M.,(2004). Survey on patients in hospital undergoing chest, abdomen and lumber spine x-ray examination. NigJourn.,Phys. 89(3), 102-103 Nwankwo, I.S. (2004). Determination of Exposure factors on the quality of skull x-ray radiographs. A Thesis of Physics department university of Jos. Serro, R.(1992). Popoulation Dose Assessment from radiodiagnosis in Portugal. Radiation Protection Dosimetry 43, 65-68. Assessment of Radioactivity Concentration in Soil of some Mining Areas in Central Nasarawa State, Nigeria 18