Gas Chromatography Syringe Selection Guide

Similar documents
More accurate results.

The Measure of Excellence. GC Autosampler Syringes. CTC GC PAL Autosampler Syringes (PAL is a registed trademark of CTC Analytics AG, Switzerland)

Results of Using Re-washed Vials and Closures

Choose for Your Instrument. GC Supplies - Shimadzu Instrument Quick Pick Guide

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Syringe Care and Use Guide. Complete Guide to Maintaining and Using Hamilton Microliter, Gastight and Specialty Syringes

Hamilton Syringes. LT N LTN RN TLL TLL with slots. Hamilton Syringe Types Microliter. Hamilton Point Styles

AUTOMATED LIQUID HANDLING

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

SYRINGE CARE AND USE. Complete Guide to Maintaining and Using Hamilton GASTIGHT, MICROLITER TM, and Specialty Syringes

Achieving Secure Sealing and Instrument Handling in 9 mm Screw Thread Chromatography Vials

AUTOMATED LIQUID HANDLING

cataleg-inl-2007-def6.indd /01/ :27:30

Liquid Chromatography Syringes

Selecting The Right Syringe

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

INSTRUCTIONS FOR PREPARING AND GIVING AN INJECTION OF ENBREL POWDER

Thermo Scientific Finnpipette Manual Pipetting Systems. you can handle. anything

Savillex Technical Note C-Flow 700d PFA Concentric Nebulizer for ICP-OES

#20 Separation of Alcohols Using Gas Chromatography

Ultimate Core Performance to Maximize Your Investment

Needle Gauge. Needle OD (mm) Needle ID (mm)

INSTRUCTIONS FOR USE AND CARE

Savillex Technical Note

Elcometer 138/2. Surface Contamination Kit. Operating Instructions

PLEASE READ THIS USER MANUAL BEFORE USE

Restek PAL SPME Fibers cat.#

Savillex Technical Note

AS Application Note 1602

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

MEDIUM-FLOW PNEUMOTACH TRANSDUCER

For use only with Lilly 3 ml insulin cartridges (100 units/ml). Do not use other brands of insulin cartridges.

For use only with Lilly 3 ml insulin cartridges (100 units/ml). Do not use other brands of insulin cartridges.

Single Use Curlew TM Multiple Biopsy Forceps

ab Lipid Peroxidation (MDA) Assay kit (Colorimetric/ Fluorometric)

Elemental Scientific Semiconductor Catalog for Thermo ELEMENT2

Free and Total Sulfur Dioxide (SO 2. ) in wine (Ripper method) by automatic titration

DO NOT TRANSFER TO A SYRINGE SEVERE OVERDOSE CAN RESULT

Lipidomic Analysis by UPLC-QTOF MS

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

HumaPen LUXURA HD INSULIN DELIVERY DEVICE INSTRUCTIONS FOR USE

Instructions for Use Enbrel (en-brel) (etanercept) for injection, for subcutaneous use Multiple-dose Vial

SelectTech 3.1 Bench Assembly / Owner s Manual

PRO-V IDS Kit. Bisco. Instructions for Use. Provisional Restorative System

WHAT WORKS? A comprehensive guide to safer needle and syringe selection for injecting drug users. tools for harm reduction

HumaPen SAVVIO INSULIN DELIVERY DEVICE INSTRUCTIONS FOR USE

WHAT WORKS? A comprehensive guide to safer needle and syringe selection for injecting drug users. Second edition. Fully revised and updated

autosamplers ASI-100 and ASI-100T Autosamplers

injections. Needle Tip 1 µl Cone 1BR

Total Acid Number in petroleum products by automatic titration

Surgical Technique Guide

Testing robustness: Immunosuppressant drugs in blood with a TSQ Quantis MS for clinical research

Initial placement 24FR Pull PEG kit REORDER NO:

Stratis INSTRUCTIONS FOR USE. Needle-free Injection System. 0.5mL volume (+/-5%)

SPME-LC Fibers for a Variety of Applications

Increased Efficiency of Biomolecule Identification by Optimization of Trypsin Digestion Buffers

Selection and Optimization of the Sample Introduction System

INSTRUCTIONS FOR USE

NEW! Thermo Scientific ClipTip Pipetting Systems. sealed in security. feel the difference

Blood Glucose Monitoring System. User Guide

Instructions for Use. Welcome!

Innovations in Administration and Injection Systems EXPERIENCED CREATIVE FOCUSED

CoaguChek S System. Quick Reference Guide. This is a CLIA-waived System. Coagulation Testing with Fresh Whole Blood

SelectTech 4.1 Bench. Assembly Manual

Effective Date: August 31, 2006 SUBJECT: CARE AND USE OF NEBULIZER AND INTERMITTENT POSITIVE PRESSURE BREATHING DEVICE

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:

Instructions for Use. HUMALOG KwikPen. insulin lispro injection (rdna origin) 100 units/ml, 3 ml pen

Forensic Analysis of Blood Alcohol Concentration

Scope This policy applies to all personnel and departments that clean, prepare and/or sterilize items intended for patient care use.

Measurement, Separations, and More. More than 2,000 Syringes and Accessories

Operational Guideline

Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA

SelectTech 4.1 Bench Assembly / Owner s Manual

Do not re use needles. If you do you might not get your dose (underdosing) or get too much (overdosing) as the needle could block.

1.0 Purpose The purpose of this guideline is to ensure the safe administration of intramuscular injections to Cheshire Ireland service users.

Integra Flowable Wound Matrix

Cap Clip Rubber Seal Plunger Pen Body Dose Window

Fibrlok II 2529 Universal Optical Fiber Splice

Thermo Scientific Finntip Pipette Tips. HANDHELD PIPETTING Finntip Pipette Tips. Details

insafe System - Directions for Use

HUMULIN 70/30 KwikPen

PLEASE READ THESE INSTRUCTIONS BEFORE USE

Instructions for Use. BASAGLAR KwikPen. insulin glargine injection (100 units/ml, 3 ml pen)

Instructions for Use HEMLIBRA (hem-lee-bruh) (emicizumab-kxwh) injection, for subcutaneous use

SafeTview Syringe Shields Models to Latchkey Syringe Shields Models to Nuclear Medicine. Specifications.

Humanity s Vision Is Our Focus. The Ahmed Glaucoma Valve

Syringe Guide. The Measure of Excellence for your application

TJF-160F/VF Cleaning and Disinfection Checklist

Initial placement 20FR Guidewire PEG kit REORDER NO:

Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS

A patient guide to administration of subcutaneous immunoglobulin replacement therapy - using manual push technique

Simple Strategies for GC/MS Success. Rachael Simon Agilent Technologies Product Manager, Gas Chromatography Supplies

Instructions for Use. For use with. 10 mg vial

Product Guide for LudgerSep TM R1 HPLC Column for DMB labelled Sialic Acid Analysis

Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

Instructions for Use Enbrel (en-brel) (etanercept) Single-use Prefilled SureClick Autoinjector

Transcription:

Gas Chromatography Syringe Selection Guide Peter Morgan, Fisher, Runcorn, Cheshire, UK Technical Note 20552 Key Words Syringe selection Abstract Choosing the correct syringe from the wide selection available can be a difficult task, however with a little more information this process can be simplified. Introduction The syringe plays a pivotal role in the GC system as it takes the sample from the vial and introduces it into the inlet. There is a lot of scope for error when selecting the correct syringe due to a wide range of inlets from different manufacturers and also the wide range of autosampling devices. Once the correct syringe for the inlet and the autosampler has been selected, there are further parameters, such as needle gauge and tip style, that can be crucial when selecting a syringe. In this guide we will simplify the syringe selection process, allowing users to confidently select the correct needle for their instrument How to Select the Correct Syringe Given in Table 1 are the details required to select the most appropriate syringe for an application, with an explanation of syringe components from tip to plunger in subsequent sections. Parameter Example Relevance to selection 1. GC type, Agilent Correct needle length 2. Autosampler Type, Agilent, Gerstal Correct syringe plunger attachment 3. Sample type Liquid, headspace Needle tip style 4. Sample volume 1 µl, 1000 µl Total syringe volume Table 1: Parameters required to select the correct GC syringe Once all of this information has been gathered, the most appropriate syringe can be determined by using the following sections:

2 Needle Length Each injection port is designed differently and requires a specific length of syringe to be used. For reproducible sample transfer from liner to column, injection ports require the sample be injected at a specific depth inside the inlet liner. This makes it critical to have the correct needle length installed for your application. For a list of which needle lengths are compatible with injection ports and autosamplers please refer to Table 2. TriPlus AS1310 (for TRACE 1300/1310) TriPlus RSH on TRACE GC TriPlus RSH on TRACE 1300 GC AS2000, AS3000 Agilent CTC (Gerstel MPS) Perkin Elmer Autosampler Needle Length (mm) Manual Injection Shimadzu 42 x x 50 x x x x 51 x x 56 x 57 x x 65 70 x x 75 x 80 x x 85 x Table 2: Needle length by Autosampler Needle Tip Style Selecting the correct needle tip allows the best performance to be achieved by your syringe. supply needles in the following tip styles: Cone (tapered tip) these are the most versatile needle for autosampler use and resist coring of vial and inlet septa Bevel (sharp tip) these are typically used for manual injections. The tip shape helps reduce septa coring Side hole (dome tip with side hole for sample exit) usually used for headspace and large volume injections Blunt End or 90 (flat tip) used for injectors that do not contain an inlet septa such as Merlin Microseal To determine which needle tips are compatible with your instrument refer to Table 3. Gauge Merlin Microseal injector Tip style TRACE SSL TRACE PTV TRACE 1300/1310 SSL/ PTV Agilent 6890 SSL Perkin Elmer Shimadzu Manual Injection 22/22s x 23 x x x x x x x 25 x 26 x x x x x 23-26 x x Cone x x x x x x x Bevel x x x x x Side hole x x x x Blunt x x Table 3: Needle gauge and tip style by instrument type

Needle Gauge Needle gauge is a measure of the thickness of the needle. The higher the number the thinner the syringe (e.g. a 26 gauge needle is narrower than a 23 gauge). The suffix s, e.g. 23s refers to a needle with a narrower internal diameter (and volume) than is standard for that gauge. Selecting the correct gauge of needle is critical for on-column injections. This is because the syringe must fit inside a capillary column (typically 0.53 mm I.D.). If a large (e.g. 23 gauge) syringe is used, the column will break and the sample will not be injected onto the column. Different injection ports are designed to accept specific needle gauges. For a list of which gauge is suitable for which inlet refer to the Table 2. In general, select the smallest gauge (i.e. thickest) needle that is compatible, as thicker needles are more robust, helping to increase needle lifetimes. Needle Attachment Syringes are available in both fixed and removable needle designs, both of which have their merits. Fixed Needle Syringes Advantage Lower purchase price compared to removable needle designs Disadvantage(s) Cannot be repaired, considered a disposable item Contain adhesives to attach the needle that may be damaged by solvents. Refer to maintenance section for details Removable Needle Syringes Advantage(s) Needle can be replaced when damaged, potentially giving longer lifetimes Overall lower cost over the lifetime of the syringe More versatile as can use the same syringe for multiple applications by changing the needle for another of different tip style, length or gauge Disadvantage Higher purchase price Syringe Volume A wide range of syringe volumes are available. In general using a sample volume lower than 10 % of the maximum syringe volume will reduce the precision of the injection. For liquid sampling, volumes are typically between 0.5 3 µl so the most common syringe volumes are 5 and 10 µl. For headspace sampling the typical volumes are 1 3 ml, so larger volume syringes are required. Large syringe volumes are available (>500 µl) these enable large volume injection (LVI), headspace and on-line sample preparation to be performed. Plunger Design Most plungers are made from stainless steel with an option of a PTFE tip on the end. If no PTFE tip is present then the syringe is not classified as gas tight and should only be used for liquid sampling (i.e. not for headspace). If a PTFE tip is present the syringe is suitable for gas and liquid sampling. Gas tight syringes resist plunger freezing (or sticking) due to the inherent low friction properties of PTFE. This may help extend the lifetime of the syringe, especially if dirty samples are being analyzed that are prone to clogging syringes. If no information in the product literature is given about the plunger and the syringe is not stated as gas tight then the plunger can be assumed to be stainless steel. Super-Elastic plungers are made from a more flexible material than standard plungers and may give longer lifetimes, especially when performing manual injections, than stainless steel plungers. This is because the plungers are less prone to permanent bending. Plunger in Needle syringes are designed for extremely low volumes (< 1µL) and means that the needle contains part of the sample volume 3

4 Plunger Attachment Type The correct plunger attachment must be selected to allow the syringe to interface with the autosampler. If the wrong attachment is selected the syringe cannot be used. Syringe Maintenance Routine Syringe Maintenance and Good Care Practices Syringe lifetimes can be significantly improved by adhering to the maintenance procedures and good practice guidelines given below: Wash syringes daily with a solvent the intended samples are soluble in such as methanol, dichloromethane or acetonitrile. Completely fill and expel the syringe 5 to 10 times with the chosen solvent, removing any solid debris accumulating in the barrel or needle. Remove the plunger and wipe with a lint free cloth wetted with solvent. Finally rinse with acetone and allow to air dry. After using a syringe always rinse with 3 to 5 syringe volumes of solvent to remove any material that may block the syringe. Avoid operating the plunger when the syringe is dry - liquid in the barrel acts as a lubricant reducing syringe wear. Never fully submerge the syringe in solvent, as adhesives are used to assemble the syringe and submerging in solvent may cause damage to these. On non gas tight syringes the syringe barrel and plunger are manufactured together so they match each other perfectly. This means that plungers are not replaceable as the accuracy of the syringe will be compromised. If the wrong plunger is inserted the syringe calibration is no longer valid and must be repeated. The plungers on gas tight syringes are replaceable as the PTFE tip on the plunger creates a gas tight seal. It is not common to autoclave syringes used for GC analysis, however if autoclaving is required it is preferable to use a removable needle syringe as adhesives present in fixed needle designs can be damaged by autoclaving. In general syringes should be autoclaved at a maximum temperature of 100 C. Also, note that autoclaving syringes will reduce their lifetime due to stressing of the adhesives used to assemble other parts of the syringe. If autoclaving multiple syringes simultaneously ensure the correct plunger is reinstalled in the correct barrel. Rough edges present on the tip of the needle can cause tearing of the GC inlet septa leading to septa siloxane contamination peaks in the chromatogram. Rough edges may be removed by smoothing with a fine sandpaper. Unblocking Syringes All syringes block occasionally, however if regular blockages are encountered it is likely that there is a problem with the syringe maintenance program, needle suitability or sample preparation procedures. It is recommended to optimize these procedures, as blocked syringes can be costly both in terms of broken syringes and lost instrument time. When blockages do occur the procedures below enables the majority of blockages to be removed: 1. When a syringe has been blocked by the precipitation of compounds due to drying of salts and buffers, the easiest way to unblock the syringe is to soak the needle (do not submerge the entire syringe) in an appropriate solvent in which the blockage is soluble (often distilled water), for no longer than 5 minutes. This may dissolve the precipitate and unblock the syringe. Draw 5 to 10 syringe volumes through the syringe to completely remove any remaining material. 2. If the syringe is blocked by solid particles or vial/inlet septa a different approach is required. Remove the plunger and fill the barrel of the blocked syringe with another syringe. Re-insert the plunger and gently expel the syringe. This should dislodge the blockage. Never force the plunger as this can cause the barrel to crack or the plunger to bend.

3. If step 2 does not remove the blockage try inserting the needle wire supplied with the syringe into the needle. This may help move the blockage allowing step 2 to be repeated successfully. 4. When a particularly stubborn blockage is encountered it may be necessary to stand the syringe in a beaker of solvent, so that only the needle and needle attachment are submerged, and sonicate for a maximum of 10 minutes to try and remove the blockage. Repeat step 2 to remove the blockage. 5. As a last resort it may be necessary to gently heat the needle causing it to expand slightly enabling the blockage to be removed. Extreme care must be exercised when using this approach as the syringe may be damaged irreversibly. It is advisable to check the syringe calibration after performing this procedure to ensure syringe accuracy has not been affected. Conclusion When selecting a syringe, important factors to consider are: Needle length Syringe plunger attachment Needle tip style Total syringe volume Selecting the correct product will ensure the optimum performance will be achieved from your syringe. Technical Note 20552 thermoscientific.com/gcconsumables 2012 Fisher Inc. All rights reserved. All trademarks are the property of Fisher Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. USA and Canada +1 800 332 3331 France +33 (0)1 60 92 48 34 Germany +49 (0) 2423 9431 20 or 21 United Kingdom +44 (0)1928 534110 Japan +81 3 5826 1615 TN20552_E 07/12S China +86 21 68654588 +86 10 84193588 +86 20 83145199 800 810 5118 India +91 22 6742 9494 +91 27 1766 2352 Australia 1 300 735 292 (free call domestic) New Zealand 0800 933 966 (free call domestic) All Other Enquiries +44 (0) 1928 534 050 Technical Support North America +1 800 332 3331 Outside North America +44 (0) 1928 534 440