Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Similar documents
Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

EDITORIAL NOTE A short CV

Improvement of radiation dosimetry: contribution of the IAEA

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

EDITORIAL NOTE. factor, which has been used instead of the chamber factor N D

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

THE IAEA/WHO NETWORK OF SECONDARY STANDARD DOSIMETRY LABORATORIES SSDL NETWORK CHARTER

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

The determination of timer error and its role in the administration of specified doses

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Recent advances in dosimetry in reference conditions for proton and light-ion beams

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

Strengthening Activity Measurement Quality in Radiation Protection from Metrological Science to Reliable End-user Application

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Radiation Protection Services in R.Macedonia

Dosimetry and QA of proton and heavier ion beams

MAY NOT BE SAME AS THE PUBLISHED VERSION

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

SSDL Newsletter. Contents. From the editor. No. 51 September IAEA/WHO Network of Secondary Standards Dosimetry Laboratories

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at METAS,

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

Zakithi Msimang SAAPMB/SARPA 2011

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Application of international standards to diagnostic radiology dosimetry

Calibration of two 90 Sr+ 90 Y dermatological applicators

CCRI(I)/ International Atomic Energy Agency ACTIVITIES OF THE DOSIMETRY AND MEDICAL RADIATON PHYSICS SECTION Activities in 2011 and 2012

IAEA/WHO NETWORK OF SECONDARY STANDARD DOSIMETRY LABORATORIES

Dosimetry of Small Static Fields Used in External Beam Radiotherapy

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

Diode calibration for dose determination in total body irradiation

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

Calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy

Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable Phantom

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

How accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement

Mammography calibration qualities establishment in a Mo- Mo clinical system

Application(s) of Alanine

Dosimetry Standards and Dissemination Systems for Radiation Processing in China

IAEA SUBPROGRAMME ON DOSIMETRY AND MEDICAL RADIATION PHYSICS (DMRP) REPORT ON ACTIVITIES,

Influence of the exposure time in the area monitors at radiodiagnostic

Detector comparison for small field output factor measurements with flattening filter free photon beams

Recent Activities in Measurement Standards and Dosimetry at ARPANSA,

Sterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration

Overview of Radiation Monitoring Laboratory Capabilities in Tunisia. Azza Hammou, CNRP, Tunis RMCC-8 Meeting, Amman Jordan JUNE 2013 CONTENT

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

High-Dose Standardization Study For γ-rays radiation processing at NIM

Catharine Clark NPL, Royal Surrey County Hospital and RTTQA

GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

SSDL Newsletter. Contents. From the editor. No. 50 February IAEA/WHO Network of Secondary Standards Dosimetry Laboratories

GEX Technical Information Report

International Practice Accreditation

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Accreditation of Dosimetry Services

Active Scanning Beam 3 Checking Delivery/Dosimetry

Good manufacturing practices (GMP) utilized on human blood irradiation process

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

Protection of the contralateral breast during radiation therapy for breast cancer

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Dosimetry in digital mammography

CHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE

Radiochromic film dosimetry in water phantoms

2018 Council on Ionizing Radiation Measurements and Standards. Low Energy ( kev) Electron Beam Calibration. Gary Pageau

IAEA Support to Medical Physics in Africa and Latin America: achievements & challenges

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY

Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment

IAEA section of dosimetry and medical radiation physics (DMRP).

Overview IAEA mandate and organization. IAEA activities in medical radiation physics: achievements and challenges. Contents

QA for Clinical Dosimetry with Emphasis on Clinical Trials

How accurately can we measure dose clinically? Dosimetric accuracy requirement. Accurate Clinical dosimetry. Part I: Fundamentals and Issues


Atoms for Peace and Development

Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa

Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters

HDR Brachytherapy Dosimetry at the NPL

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Metrology and quality of radiation therapy dosimetry of electron, photon and epithermal neutron beams

Activity report from JCOG physics group

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

Transcription:

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry and Monitoring group Agricultural, Medical and Industrial Research School (AMIRS), Nuclear Science and Technology Research Institute (NSTRI), Atomic Energy Organization of Iran (AEOI) - Karaj, Iran Background: Absolute dosimetry of external beam radiotherapy is carried out by the use of ionization chambers. These chambers must be calibrated at a standard dosimetry laboratory before any use in clinical dosimetry. The secondary standard dosimetry laboratory of Iran (SSDL) has the duty of calibrating the ionization chambers used in radiotherapy centers in Iran. Materials and Methods: The present work has described traceability of SSDL radiation measurement standards to relevant international standards, and calibration of therapy level ionization chambers in terms of air kerma and absorbed dose to water against 60 Co gamma radiation, as well as uncertainty evaluation of calibration coefficients. Results: The expanded uncertainties in the determination of air kerma and absorbed dose to water are estimated to be 2% and 2.3% at approximately 95% confidence level, respectively. Conclusion: In order to maintain the requirement of ±5% accuracy in the dose delivery, the combined standard uncertainty of the other factors in the dose delivery; i.e., dose measurement set-up, dose calculations, treatment planning, patient set-up, etc, should be less than 2.3%. Iran. J. Radiat. Res., 2010; 8 (3): 195 199 Keywords: Dosimetry, ionization chamber, calibration, uncertainty. INTRODUCTION A properly calibrated ionization chamber is a prerequisite for absorbed dose determination in external beam radiotherapy. The use of an ionization chamber having a calibration coefficient for 60 Co gamma rays is a common characteristic of all dosimetry protocols and codes of practice for high energy photons and electrons. These documents, introduced by national and international organizations, provide medical physicists with a systematic approach to the dosimetry of external radiotherapy beams (1-4). Calibration coefficient of an ionization chamber is expressed usually in terms of exposure X, air kerma Kair or absorbed dose to water Dw, traceable to a standard dosimetry laboratory. The various steps between the calibration of an ionization chamber in terms of air kerma at the standard dosimetry laboratories and determination of absorbed dose to water at hospitals using air-kerma based dosimetry protocols, introduce undesirable uncertainties into the realization of Dw. Many factors are involved in the dosimetric chain that starts with a calibration coefficient in terms of air kerma, NK, measured in air using a 60 Co beam and ends with the absorbed dose to water, measured in water in clinical beams. Uncertainties associated with the conversion of NK to absorbed dose to air chamber calibration coefficient, ND,air (or Ngas) mean that in practice, the starting point of the calibration of clinical beams already involves a considerable uncertainty. The well established procedures to determine absorbed dose to water; i.e., the ionization method, chemical dosimetry, and water and graphite calorimetry, have considerably improved at primary standard dosimetry laboratories (PSDLs) in recent years. These developments lend to provide ionization chambers with calibration *Corresponding author: Dr. Mostafa Ghafoori, Radiation Dosimetry and Monitoring group, Agricultural, Medical and Industrial Research School (AMIRS), Nuclear Science and Technology Research Institute (NSTRI), Atomic Energy Organization of Iran (AEOI), P.O. Box 31485-498 Karaj, Iran. Fax: +98 261 4464058 E-mail: mghafoori@nrcam.org

A. Solimanian and M.Ghafoori coefficients in terms of absorbed dose to water, ND,w, for use in radiotherapy beams. Many PSDLs already provide ND,w calibrations at 60 Co gamma ray beams and some laboratories have extended these calibration procedures to high energy photon and electron beams. Intercomparisons of primary standards of air kerma and absorbed dose to water generally give agreement within 0.2 0.3% at 60 Co gamma ray quality (1). MATERIALS AND METHODS Traceability to International Standards At secondary standard dosimetry laboratories (SSDLs), calibration coefficients from a PSDL or from BIPM are transferred to hospital users. The dosimetry laboratory of the International Atomic Energy Agency (IAEA) is distinguished to act as a link between primary standards and members of the IAEA/WHO network of SSDLs. Indeed, calibration of all reference standard ionization chambers of the SSDL of Iran, as well as many other SSDLs, is traceable to the BIPM or a PSDL via the IAEA. Intercomparisons of secondary standards of absorbed dose to water are organized annually by the IAEA at 60 Co gamma ray and high energy X-rays through TLD postal dose audit program. Excellent results are reported by the IAEA in recent years. The results of the intercomparisons for the SSDL of Iran for the years 2000 to 2009 are shown in figure 1. Therapy level ionization chambers in Iran are calibrated in comparison with the response of the SSDL reference and working standard ionization chambers in the 60 Co gamma ray beam of a teletherapy unit. All of the SSDL ionization chambers used for calibrations are themselves calibrated at the IAEA dosimetry laboratory (5). However, uncertainties associated with the calibration coefficients, together with other sources of uncertainties in the calibration process, usually lead to an expanded uncertainty of more than 2% in the calibration coefficients of hospital dosimeters. Figure 1. The results of IAEA/WHO TLD radiotherapy level audit for the SSDL of Iran for the years 2000 to 2009. Ionization Chambers Six therapy level ionization chambers were used for this investigation. Two of them are of NPL type secondary standard NE 2561 (UK) and NE 2611 ionization chambers, which were used as reference chambers at calibration laboratories. The other four chambers are well known 0.6 cc farmer type chambers (NE Technology (UK) and Physikalisch- Technische Werkstätten (PTW, Germany) ) which were used both as working standards and field ionization chambers. These chambers are all calibrated at the IAEA dosimetry laboratory at 60 Co gamma radiation in terms of air kerma and absorbed dose to water. The calibration coefficients are reported in the IAEA certificates with both IAEA and SSDL (when available) electrometers. However, the differences between them for each one of the chambers are always less than their reported uncertainties. Calibration of ionization chambers at SSDL Calibration of the ionization chambers used in radiotherapy, were performed in comparison with the SSDL reference or working standard ionization chambers using substitution method. Calibrations were carried out in terms of both air kerma and absorbed dose to water, at 60 Co gamma ray beam of a 60 Co Picker V9 teletherapy unit. When calibrating in terms of air kerma, the chamber with its 60 Co buildup 196 Iran. J. Radiat. Res., Vol. 8 No. 3, December 2010

Calibration of ionization chambers used for dosimetry cap, was positioned free in air, so that its reference point was on the central axis of the beam and the chamber axis was perpendicular to it. The distance from the source to the reference point of the chamber was 80 cm. The radiation field size at reference plane was 10 cm 10 cm figure 2 shows the configuration set-up. 60 Co source 1000 mm Ionization chamber with build-up cap Field size 100mm 100mm Figure 2. Reference conditions for calibration in terms of air kerma for 60 Co gamma radiation. When calibrating in terms of absorbed dose to water, the chamber protected by a PMMA sleeve, was positioned in a 30 cm 30 cm 30 cm water phantom so that its reference point was on the central axis of the beam. The chamber axis was perpendicular to the central axis of the beam and the distance from the source to the reference point of the chamber is 80 cm. The reference point of the chamber was at 5 cm water depth and the size of the radiation field (50% isodose level) at the reference plane was 10 cm 10 cm. Figure 3 shows the configuration set-up for calibration in terms absorbed dose to water. 60 Co source Ionization chamber without build-up cap in the PMMA sleeve Field size 100mm 100mm Figure 3. Reference conditions for calibration in terms of absorbed dose to water for 60 Co gamma radiation. RESULTS AND DISCUSSION Absorbed dose to water, which is the quantity of interest in radiotherapy, is preferably determined by direct measurement in a phantom but it also may be calculated by an air kerma measurement in a 60 Co gamma ray beam free in air. The mean value of the absorbed dose rate at a reference point in a water phantom, determined at the same date by ND,w formalism (2), NK formalism (1) and in air measurement formalism (6), using all six calibrated chambers mentioned in materials and methods section, were in agreement within 1.8% at 95% confidence level (figure 4). Figure 4. Results of absorbed dose rate Variation at a reference point in a 60 Co radiotherapy beam as determined by three formalisms using IAEA calibrated chambers. The reference point was located on the central axis of the beam, at 5 cm depth in a 30 cm 30 cm 30 cm water phantom. The distance from the source to the phantom surface was 80 cm and the field size at this distance was 10cm 10cm. Calibration coefficient of a chamber under calibration is obtained from: SSDL SSDL U N M N = U M where N U is the calibration coefficient (to be determined) and M U is corrected electrometer reading of the user chamber. Also, N SSDL is the calibration coefficient and M SSDL is corrected electrometer reading of the SSDL chamber. The uncertainties in the SSDL calibration coefficients corresponding to one standard deviation are reported in the IAEA certificates to be 0.4% to 1%. The electrometer readings M0 SSDL and M0 U should be Iran. J. Radiat. Res., Vol. 8, No. 3, December 2010 197

A. Solimanian and M.Ghafoori corrected for several factors including scale reading, recombination loss, leakage, radiation background, distance, air density, field homogeneity and humidity. The uncertainty contributions of these factors must be evaluated in the determination of N u. Scale reading is the only factor that has statistical contribution to overall uncertainty, the so called type A, uncertainty. Contribution of other factors to the uncertainty may be regarded as systematic, the so called type B uncertainty. The standard deviation of the electrometer reading of well behaved therapy level ionization chambers does not exceed 0.5%. Thus the uncertainty in both readings, M0 SSDL and M0 U, is overestimated to be 0.5%. Recombination loss for secondary standard instruments and most other therapy level dosimeters is small in a 60 Co beam and the uncertainty is considered to be negligible. Leakage currents including radiation induced leakage have been small for secondary standard dosimeters and for field ionization chambers with possible larger volumes may be considered to vary with an uncertainty of not more than 0.2%. Radiation background for therapy level ionization chambers was negligible with negligible uncertainty (7). Ionization chambers were used at the same source-chamber distance, 80 cm, but this value did not enter the uncertainty calculation. The difference between the positions of the chamber centers could be as much as 1 mm. The percent uncertainty must have been doubled owing to the inverse square law. Regarding the air density correction, the exact temperature and pressure do not enter uncertainty calculation. The temperature and pressure usually may differ between the measurement of two chambers by 0.5 C in air (0.2 C in water) and by 1mb respectively. The uncertainty due to air density correction is about 0.2% in air (0.1% in water). Lateral displacement of the chambers during calibration is too small to cause any detectable difference in response owing to field non-uniformity. Relative humidity in the SSDL laboratory is normally in the range of 20% to 50% during a year and is not high enough to cause leakage. The uncertainties in the calibration of a therapy level ionization chamber in comparison with each one of the six calibrated SSDL chambers are summarized (8) : Parameter Uncertainty (%) Uncorrected reading, M0 SSDL 0.5 Uncorrected reading, M0 U 0.5 SSDL calibration coefficient, N SSDL 0.6 (NK), 0.8 (ND,w) Leakage (user chamber) 0.2 Air density 0.2 (NK), 0.1 (ND,w) Distance 0.3 Recombination loss 0.1 (SSDL), 0.1 (user) Combined standard uncertainty 1.02 (NK), 1.15 (ND,w) Expanded uncertainty (k=2) 2.04 (NK), 2.3 (ND,w) CONCLUSION Calibration of ionization chambers is an important first step in the dosimetry of radiation therapy beams. The SSDL of Iran, which is traceable to international standards of air kerma and absorbed dose to water via IAEA dosimetry laboratory, issues calibration certificates for ionization chambers used in radiotherapy centers in Iran. The certificates provide the ionization chambers with air kerma and absorbed dose to water calibration coefficients in 60 Co gamma radiation. Calibration coefficients are determined with an expanded uncertainty of 2% for air kerma and 2.3% for absorbed dose to water with a coverage factor k=2. According to ICRU (8-9), the available evidence for certain types of tumors points to the need for an accuracy of 198 Iran. J. Radiat. Res., Vol. 8 No. 3, December 2010

Calibration of ionization chambers used for dosimetry ±5% at 95% confidence level in the delivery of an absorbed dose to a target volume. This is interpreted as if it corresponds to approximately two standard deviations. Thus the requirement for accuracy of 5% in the delivery of absorbed dose corresponds to a combined standard uncertainty of 2.5% at the level of one standard deviation. Regarding the combined standard uncertainties for NK and ND,w, we may write where UN is the combined standard uncertainty of NK or ND,w and ur is the combined standard uncertainty of the other factors in the dose delivery i.e. dose measurement set-up, dose calculations, treatment planning, patient set-up, etc. In order to maintain the requirement of ±5% accuracy in the dose delivery, ur is estimated to be less than 2.3%. REFERENCES u 2 N 2 + u r 2.5 1. International Atomic Energy Agency(1997) Absorbed Dose Determination in Photon and Electron Beams: An 2 International Code of Practice, Technical Report Series No. 277 IAEA, Vienna. 2. International Atomic Energy Agency (2000) Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water, Technical Report Series No. 398, IAEA, Vienna. 3. American Association of Physicists in Medicine(1999) Task Group 51: Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams. Med. Phys, 26: 1847-1870. 4. American Association of Physicists in Medicine(1983) Task Group 21: A protocol for the determination of absorbed dose from high-energy photon and electron beams. Med Phys, 10: 741-771. 5. Meghzifene A and Short K (2007) The IAEA calibration and Measurement capabilities SSDL Newsletter, 53: 4-8. 6. International Atomic Energy Agency (1994) Calibration of dosimeters used in radiotherapy, technical reports series No. 374, IAEA, Vienna. 7. Faiz M Khan (2003) Measurement of ionizing radiation, The physics of radiation therapy, Lippincott Williams & Wilkins, Philadelphia, USA. 8. International Atomic Energy Agency (2009) Calibration of Reference Dosimeters for External Beam Radiotherapy. Evaluating the uncertainty TRS No. 469: 57-61. 9. International Comission on Radiation Units and Measurements(1976) Determination of absorbed dose in a patient irradiated by beams of X or gamma rays in radiotherapy procedures, ICRU Report 24, ICRU, Bethesda. Iran. J. Radiat. Res., Vol. 8, No. 3, December 2010 199