A novel phantom design for brachytherapy quality assurance

Similar documents
Brachytherapy Planning and Quality Assurance

Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w

Applications in brachytherapy

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations

2D linear array device as a quality assurance tool in brachytherapy applications

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

Definitions. Brachytherapy in treatment of cancer. Implantation Techniques and Methods of Dose Specifications. Importance of Brachytherapy in GYN

Developing a Verification and Training Phantom for Gynecological Brachytherapy System

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998)

Advanced film dosimetry for a UK brachytherapy audit

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

Radiochromic film dosimetry in water phantoms

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Material-specific Conversion Factors for Different Solid Phantoms Used in the Dosimetry of Different Brachytherapy Sources

Iranian Journal of Medical Physics

Basics of Cervix Brachytherapy. William Small, Jr., MD Professor and Chairman Loyola University Chicago

Recent proceedings in Brachytherapy Physics

Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors

MAMMOSITE BRACHYTHERAPY DOSIMETRY EFFECT OF CONTRAST AND AIR INTERFACE ON SKIN DOSE. A RESEARCH PAPER SUBMITTED TO THE GRADUATE SCHOOL


Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range

Pulsed Dose Rate for GYN Brachytherapy

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005

Gafchromic film dosimetry of a new HDR 192 Ir brachytherapy source

A B S T R A C T. 1. Introduction

Dosimetric evaluation of a novel high dose rate (HDR) intraluminal / interstitial brachytherapy applicator for gastrointestinal and bladder cancers

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy

3D ANATOMY-BASED PLANNING OPTIMIZATION FOR HDR BRACHYTHERAPY OF CERVIX CANCER

Monte Carlo and experimental relative dose determination for an Iridium-192 source in water phantom

Origin, funding & collaboration

Treatment planning study of the 3D dosimetric differences between Co-60 and Ir-192 sources in high dose rate (HDR) brachytherapy for cervix cancer

Intravascular Brachytherapy Dosimetry Techniques for Gamma systems

Regulatory Guidelines and Computational Methods for Safe Release of Radioactive Patients II. Brachytherapy

Brachytherapy, Radionuclide Therapy Medical Physics in the Clinic. Raymond K. Wu, PhD Chairman AAPM Exchange Scientist Program

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

Brachytherapy an Overview

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Chapter 13: Brachytherapy: Physical and Clinical Aspects

Determining rectal dose through cervical cancer radiotherapy by 9 MV photon beam using TLD and XR type T GAFCHROMIC Film

A commissioning procedure for breast intracavitary electronic brachytherapy systems

A Study on Dosimetry of Gynaecological Cancer and Quality Assurance of HDR Brachytherapy in BPKMCH, Nepal

In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment

Dosimetric Analysis Report

Research Article Dose Distributions of an 192 Ir Brachytherapy Source in Different Media

ADVANCED TECHNOLOGY CONSORTIUM (ATC) CREDENTIALING PROCEDURES FOR LUNG BRACHYTHERAPY IMPLANT PROTOCOLS

Acknowledgements. QA Concerns in MR Brachytherapy. Learning Objectives. Traditional T&O. Traditional Summary. Dose calculation 3/7/2015

Commissioning of brachytherapy Treatment Planning Systems

Verification of Advanced Radiotherapy Techniques

Assessment of dose homogeneity in conformal interstitial breast brachytherapy with special respect to ICRU recommendations

Review of TG-186 recommendations

2015 Radiology Coding Survival Guide

GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

Digital film dosimetry in radiotherapy and the development of analytical applications software

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

Bell in 1903 was first suggested implanting. First Report on Study of Quality Control Program of High Dose Rate (Hdr) Brachytherapy.

International Journal of Radiation Research, July 2014

Ritu Raj Upreti, S. Dayananda, R. L. Bhalawat*, Girish N. Bedre*, D. D. Deshpande

Predicting Maximum Pacemaker/ICD Dose in SAVI HDR Brachytherapy

Gafchromic EBT-XD film: Dosimetry characterization in high-dose, volumetric-modulated arc therapy

Physics. Copyright 2010 Xoft, Inc.

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

Comprehensive and Practical Brachytherapy March 04-8 March 2018, Ljubljana, Slovenia Day 1 Sunday 4 March 2018

Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT

Gdansk, Poland Gdansk, Poland

University of Wollongong. Research Online

Basic Concepts in Image Based Brachytherapy (GEC-ESTRO Target Concept & Contouring)

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Image based Brachytherapy- HDR applications in Gynecological Tumors

Brachytherapy RTT Review Course Laura Doyle, MS

Brachytherapy The use of radioactive sources in close proximity to the target area for radiotherapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Evaluation of a single-scan protocol for radiochromic film dosimetry

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Current Status of Electronic Brachytherapy Dosimetry

Activity report from JCOG physics group

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

Accurate Accurate boost or Simply Accuboost

Radioactive sources in brachytherapy

Evaluation of PRESAGE dosimeters for brachytherapy sources and the 3D dosimetry and characterization of the new AgX I seed model

Dosimetric Characteristics of the Brachytherapy Sources Based on Monte Carlo Method

DOSIMETRIC STUDY OF I-125 SEEDS USED IN PROSTATE BRACHYTHERAPY

Dosimetry benchmark for MBDCA

A Review of UK Guidelines for Prostate Brachytherapy. Sarah Aldridge. Head of Brachytherapy Physics Guy s & St. Thomas NHS Foundation Trust

How ICD-10 Affects Radiation Oncology. Presented by, Lashelle Bolton CPC, COC, CPC-I, CPMA

Basic Fundamentals & Tricks of the Trade for Intracavitary Radiotherapy for Cervix Cancer. Paula A. Berner, B.S., CMD, FAAMD

QA for Clinical Dosimetry with Emphasis on Clinical Trials

Transcription:

Volume 14, No 1 International Journal of Radiation Research, January 2016 A novel phantom design for brachytherapy quality assurance S. Gholami 1, H.R. Mirzaei 2*, A. Jabbary Arfaee 2, R. Jaberi 3, H.A. Nedaie 1,3, S. Rabi Mahdavi 4, E. Rajab Bolookat 2, A.S. Meigooni 5 1Department of Medical Physics and Biomedical Engineering, Faculty of Medicine, Tehran University of Medical Sciences, Tehran, Iran 2Radiation Oncology Department, Shohada e Tajrish Hospital, Cancer Research Center, Shahid Beheshti University of Medical Sciences, Tehran, Iran 3Radiation Oncology Department, Cancer institute, Tehran University of Medical Sciences, Tehran, Iran 4Department of Medical Physics, Iran University of Medical Sciences, Tehran, Iran 5Comprehensive Cancer Centers of Nevada, Las Vegas, Nevada, United States Technical note *Corresponding author: Dr. Hamidreza Mirzaei, E-mail: mirzaei65@yahoo.com Revised: Feb. 2015 Accepted: April 2015 Int. J. Radiat. Res., January 2016; 14(1): 67-71 DOI: 10.18869/acadpub.ijrr.14.1.67 ABSTRACT Background: One major challenge in brachytherapy is to verify the accuracy of dose distribuons calculated by the treatment planning system. In this project, a new phantom design has been introduced for quality assurance of dose distribuons in gynocological (GYN) brachytherapy implants using EBT GafChromic film. Materials and Methods: This phantom has been designed and fabricated from 90 slabs of 18 16 0.2 cm 3 Perspex to accommodate a tandem and ovoids assembly, which is normally used for GYN brachytherapy treatment. In addion, this phantom design is allowing the use EBT GafChromic films for dosimetric verificaon of GYN implants with Cs-137 Selectron LDR system. With this assembly, GafChromic films were exposed using a plan designed to deliver 2.5 Gy dose to point "A" in Manchester system for tandem and ovoids configuraons and to deliver 1.5 Gy of dose to 0.5 cm distance from the lateral surface of ovoids for using ovoid-pair. The measured dose distribuons with GafChromic films were compared with the TPS isodose lines both numerically and spaally. For a quantave analysis of the results, the measured doses values at several points of interest were evaluated with the treatment planning data and values obtained following the TG-43 dose calculaon formalism. Results: The results of these invesgaons have indicated that the new phantom design enables us to measure differences of greater than ±6% for LDR brachytherapy GYN treatments. Conclusion: The new phantom design could be ulized for the QA procedure of the brachytherapy remote acer loading systems to confirm the accuracy of dose distribuon in GYN implants. Keywords: QA brachytherapy phantom, film dose measurements, Selectron LDR system, TG-43. INTRODUCTION In brachytherapy, radioactive sources are placed near or within the tumor volume in order to maximize the dose delivery to the tumor and minimize the dose to the surrounding normal tissues. GYN brachytherapy treatments for cervix, vaginal and endometrial cancers have been commonly used for several decades (1). Accuracy of dose calculation by the treatment planning systems plays a vital role on the success of these brachytherapy treatments. As

brachytherapy techniques become more complex, the quality assurance (QA) techniques have to be modi%ied to provide a sound and practical method of veri%ication of treatment delivery system. The progress is needed for QA of complex brachytherapy treatment techniques, particularly when it involves treatment with complex low dose rate (LDR) or high dose rate (HDR) gynecologic (GYN) systems. The QA tests are required when a new planning system has been introduced for clinical application. For example, recently a new high dose rate brachytherapy system has been introduced based on 60 Co radionuclide (GZP6 machine) (2). This system has its own planning system which may require QA tests before it is implemented in clinical procedures. Experimental veri%ication of the dose accuracy is one possible QA procedures for determination of any over-dose or under-dose area in the brachytherapy planning volume. One important criteria for these experimental setups is to have a 2D or 3D dosimeter with high spatial resolution such as GafChromic TM (also known as Radiochromic TM : previously distributed by International Speciality Product (ISP) but it is presently distributed by the Ashland Re%ining Company) %ilms. Different investigators had demonstrated the usefulness of the EBT GafChromic TM %ilms for brachytherapy source dosimetry (3,4). These %ilms can be cut to the shape of the experimental geometry for the best representation of the dosimetric setup. In this project, a new phantom has been designed and fabricated for dosimetric veri%ication of the GYN implants with low dose rate (LDR) or high dose rate (HDR) systems, using EBT GafChromic TM %ilm. The accuracy and functionality of this phantom design has been examined using a well stablished LDR 137 Cs Selectron remote afterloading system, which is distributed by Nucletron (Nucletron BV, Veenendaal, The Netherlands) (5). This phantom design enable us to verify the accuracy of the dose distributions around various combinations of tandem and ovoid applicators. The results of the measured data with this device were compared with calculated values from treatment planning systems and also with hand calculations based on TG-43 formalism. MATERIALS AND METHODS LDR Cs-137 selectron remote after-loading system In the present study, the reference points for dose prescription were selected based on Manchester system (6) and ICRU38 recommendation (7). A tandem and ovoids combination was selected, with 5 cm long on tandem #1 and large size ovoids (3cm diameter). With this assembly a set of 17 active and dummy pellets were used in tandem and 8 pellets in each ovoid. From the pellets in tandem, numbers 1,4,7,8,9,10,11,12,13 were chosen to be active. In each ovoid, pellets numbers 2 to 7 were considered active. The remainder pellets were non-active, in order to deliver 2.5 Gy of dose to point "A" in Manchester system (6) in 1.08 hours. For using ovoid- pair geometry, large size (3 cm diameter) ovoids were selected. Again, pellet numbers 2 to 7 were considered active and the remainders were non-active to deliver 1.5 Gy of dose to 0.5 cm lateral distance from surface of ovoids in 1.01 hours. This dose represents the same dwell positions and dwell times as the ovoid portion of the tandem and ovoids assembly. The Nucletron (PLATO v.11.4) treatment planning system (TPS) was utilized for calculation of the dose distribution by this system (5). New phantom design and dose measurements technique The new phantom has been designed and fabricated from 90 slabs of 18 16 0.2 cm 3 Perspex to accommodate GYN applicators (tandem and ovoid). This con%iguration enables us to verify the dose distributions around the applicator with high spatial resolution. Figure 1 shows the schematic diagram of this phantom from different perspective views and also the picture of the sample applicator assembly placed in the phantom. The thin layers of the slabs were carefully 68

machined to accommodate the GafChromic TM %ilms in between the slabs for radiation dosimetry. Also, the phantom design could be adjusted with some adaptive inserts to accommodate other ovoid sizes. The points of interests include points "A" and "B" in Manchester system for tandem and ovoids combination and points "D" and "F" for ovoid-pair, as shown in %igure 2, which are located at 0.5 cm and 1.0 cm distances from the lateral surface of ovoid, respectively. Also, the measured isodose lines were compared with the treatment planning data. Figure 3 shows the picture of the irradiated %ilms that were cut in various sizes and shape to match the curvature and location of the applicator. Film dosimetry procedure For a qualitative evaluation of the new phantom design, several layers of GafChromic TM %ilm have been irradiated using the tandem and ovoid combination and ovoid-pair geometry. The irradiated %ilms for any implant geometry and calibration procedures were scanned with the same technique after 24 hours from exposure time. As suggested in the guideline of the GafChromic TM %ilms, the irradiated %ilms for both calibration and measurement process were scanned in the landscape orientation (8). Scanned EBT GafChromic TM %ilms were transformed to IMAGEJ software (Java-based image processing program, National Institute of Health) (9). The dose value of reference point according to its position in the %ilm was calculated. Both the calibrated and experimental %ilm scanning have been performed using 100 dpi resolutions. The pixel value was obtained from IMAGEJ software. The maximum possible reading of the scanner was I0=65535. Net optical density (NOD) is obtained by subtracting the reading for the base fog (OD of un-irradiated %ilm) form the measured optical density (10). Figure 2. Schemac diagram of the tandem-ovoids and ovoid-pair applicators and calculated points relave to the applicators (Graphs are not to the scale). Figure 1. Schemac diagrams of the new phantom design with the lateral view of the GYN applicator (A), Axial View (B), a photograph of the GYN tandem and ovoid applicator placed within the phantom, and schemacs of different layers of the phantom and applicator (D). (A patent for this phantom design has been submifed to the office of Industrial property with registraon number of 73752). 69 Figure 3. Different slices of EBT-1 films, exposed to the radiaon from tandem and ovoid Assembly. These films were cut in different size and shape to match the curvature of the applicators and were placed various locaons in the phantom. The labels on the films are used to idenfy the orientaon and posion of the films within the phantom. (Z=0 represents a surface that transverse axes of film are on ovoids center).

TG-43 calculation For LDR 137 Cs Selectron remote after loading unit %ilm dosimetries with the new phantom have been completed at multiple levels relative to the GYN applicators. TG-43 calculations were considered for some interesting points to verify the agreement between calculated dose value from %ilm dose measurements and dose value in treatment planning. TG-43 formalism was employed depend on dosimetric parameters of 137 Cs source. These parameters consist of air kerma strength, dose rate constant, geometry function, radial dose function and anisotropy function. TG-43 parameters of 137Cs Selectron LDR brachytherapy sources were considered from Sina's Monte Carlo simulation (11). Dose rate constant value for this unit was 1.102 (cgyh -1 U -1 ). Uncertainty analysis The accurate results of GafChoromic TM %ilm dosimetry signi%icantly depend on the position accuracy of the source with respect to the point of interest (12). A special attention was paid to the distances between the phantom s edge and edge of the applicator and the %ilm. The distance between the phantom slabs edge and the applicators is 95 micrometer. Moreover, there is a 0.1 mm gap between edge of the machined %ilm and the phantom s edge. Also, there is a 0.5 millimeter error in determining the position of points of interest. With considering the slope of dose pro%ile at level of point "A" from IMAGEJ software, we can estimate the uncertainty of total dose due to the uncertainty of the coordinate of the points of interests. RESULTS EBT GafChromic (ilm calibration The calibration curve of the EBT GafChromic TM %ilm response has ±5% uncertainty. Moreover, a 2 % of uncertainty is attributed due to the energy difference between 60Co (used for calibration) and 137Cs. The uncertainty of source-to-%ilm positioning is 4.25%. Table 1 shows the uncertainties analysis for dose measurements in this study. Results of selectron LDR brachytherapy measurements The value of reference air kerma rate for TG-43 dose calculation is extracted from treatment planning system which was 87.88 10-6 Gym 2 h -1. The quantitative analysis of the measured data for Selectron 137Cs LDR source and TG-43 dose calculation is shown in table 2. These results show a good agreement (about 6%) between the %ilm dose measurements, TG-43 dose calculation and treatment planning using designed phantom. Table 1. Total uncertainty analysis for film dose measurements. Source of uncertainty Percent of uncertainty Source-to-film posioning (Δa) 4.25% Calibraon curve fit (Δb) 5% Difference between film calibraon source And Brachytherapy source (Δc) 2% Total uncertainty ( ) 6.86% Table 2. Comparison between the measured and calculated dose values at different points around a typical applicators in GYN implant of Selectron 137 Cs LDR source. Points Dosetreatment planning (cgy) Dose-TG- 43 (cgy) Dose- Film (cgy) Difference between planning and TG-43 calcula#on (%) Difference between film dosimetry and TG-43 calcula#on (%) Difference between film dosimetry and treatment planning (%) Point A 250.0 255.4 244.0-2.1% -4.4% -2.4% Point B 70.0 71.1 67.0-1.5% -5.7% -4.3% Point D 150.0 153.2 145.8-2.0% -4.8% -2.8% Point F 100.0 102.0 95.8-1.9% -6.0% -4.2% 70

DISCUSSION AND CONCLUSION A sophisticated device for veri%ication of the accuracy of dose distribution by a 3D treatment delivery system in brachytherapy is greatly desirable. In this project, a novel phantom was designed and fabricated for QA of the dose distributions around GYN brachytherapy treatments, using EBT GafChromic TM %ilm. The accuracy of this device has been examined with LDR selectron remote afterloading system with 137Cs sources. The result of using this phantom for Selectron system shows good agreement between the %ilm-measured dose values, TG-43 based calculated data and treatment planning values (about 6%). Since the accuracy of the measurements was checked at some reference points and also for isodose lines, it is predicted to obtain good results for other points as well. In table 2, despite the fact that same formalism has been used for the hand calculation with TG-43 and treatment planning calculations, the differences may indicate the uncertainties due to the digitization of positions for the points of interests in the treatment planning system. ACKNOWLEDGMENT The Radiotherapy Department of Shahid Beheshti University sponsored the purchase of the phantom materials and 1ilms used in the investigations. The authors would like to present their appreciation for Dr Mehdi Ghorbani for his invaluable contributions in this project. In addition, the editorial suggestions by Dr Courtney Knaup are greatly appreciated. Con(licts of interest: none to declare. REFERENCES 1. Gerbaulet A, PoFer R, Mazeron J, Van Limbergen E, Editors (2002) The GEC ESTRO handbook of brachytherapy. Brussels: ESTRO [cited 2012 Feb1]. Available from :hfp:// estro-educaon.org/publicaons/documents/gec% 20ESTRO%20Handbook%20of%20Brachytherapy. html 2. Mesbahi A (2008) Radial dose funcons of GZP6 intracavitary brachytherapy 60Co sources: treatment planning system versus Monte Carlo calculaons. Iran J Radiat Res, 5 (4): 181-6. 3. Niroomand-Rad A, Blackwell CR, Coursey BM, Gall KP, Galvin JM, McLaughlin WL, Meigooni AS, Nath R, Rodgers J E, Soares CG (1998) Radiochromic film dosimetry: Recommendaons of AAPM Radiaon Therapy CommiFee Task Group 55. Med Phys, 25: 2093-116. 4. Chiu-Tsao ST, Medich D, Munro J (2008) The use of new GAFCHROMIC EBT film for 125 I seed dosimetry in Solid Water phantom. Med Phys, 35(8): 3787-99. 5. Fragoso M, Love PA, Verhaegen F, Nalder C, Bidmead AM, Leach M, Webb S (2004) The dose distribuon of low dose rate Cs-137 in intracavitary brachyther-apy: comparison of Monte Carlo simulaon, treatment planning calculaon and polymer gel measurement. Phys Med Biol, 49(24): 5459 74. 6. Tod M, Meredith WJ (1953) Treatment of cancer of the cervix uteri a revised Manchester method. Br J Radiol, 26: 252 7. 7. Internaonal Commission on Units and Measurement (ICRU) Dose and Volume Specificaon for Reporng Intracavitary Therapy in Gynecology. ICRU Report No.38.1985; Bethesda (MD): ICRU. 8. ISP (Internaonal specialty products). Handling guide of GafChromic films.available from: hfp:// online1.ispcorp.com/_layouts/gafchromic/content/ products/ebt2/pdfs/ GAFCHROMICEBT2ScanHandlingGuide.pdf. 9. ImageJ Documentaon Wiki. Available from: hfp:// imagejdocu.tudor.lu/ 10.Chan M, Zhang Q, Li J, Parhar P, Schupak K, Burman CH (2012) The Verificaon of iplan commissioning by Radiochromic EBT2 Films. IJMPCERO, 1: 1-7. 11.Sina S, Faghihi R, Meigooni AS, Mehdizadeh S, Mosleh Shirazi MA, Zehtabian M (2011) Impact of the vaginal applicator and dummy pellets on the dosimetry parameters of Cs-137 brachytherapy source. J Appl Clin Med Phys, 12(3): 183-93. 12.Aldelaijan S, Mohammad H, Tomic N, Liang L H, Deblois F, Sarfehnia A, Abdel-Rahman W, Seuntjens J, Devic S (2011) Radiochromic film dosimetry of HDR 192 Ir source radiaon fields. Med Phys, 38(11): 6074-83. 13. Lipińska J, Zwierzchowski G (2006) Dosimetric verificaon of the dose distribuon in pulsed dose rate brachytherapy. Rep Pract Oncol Radiother, 11: 223-8. 14. Kohr P and Siebert A (2007) Quality assurance of brachytherapy acerloaders using a mul-slit phantom. Phys Med Biol, 52: N387 N91. 71