Assessing the role of breathing simulators in OIP testing

Similar documents
Characterizing the Performance of Metered Dose Inhalers with Add-On Devices: New Methods For Clinically Relevant Testing

Go With the Flow REGULATORY LANDSCAPE. Mark Copley at Copley Scientific

Caption: The equipment required for testing Fluticasone Propionate (FP) Inhalation Powder in line with a new product-specific monograph (USP36-NF31).

Optimising the application of in vitro test methods for the demonstration of bioequivalence in orally inhaled products

Inhalation Product Research at FDA

Understanding cascade impaction and its importance for inhaler testing

IMPROVING THE REALISM AND RELEVANCE OF MOUTH-THROAT MODELS FOR INHALED PRODUCT TESTING

Testing Inhaled Generics

The Role of EPAG in Standards and Regulatory Guidance Development

Equivalence Evaluation of Valved Holding Chambers (VHCs) with Albuterol Pressurized Metered Dose Inhaler (pmdi)

INTRODUCTION. size and total nozzle area decrease with stage number. Volumetric air flow rate through

Assessing Quality of Inhaled Products And Links to Efficacy and Safety

University of Groningen. Technology in practice Lexmond, Anne

Dry Powder Inhaler. Developing an Efficient. 3M Conix DPI. White Paper / Spring Proven Solutions that Enable Your Success

OPTIMISING ANALYTICAL STRATEGIES FOR THE DEMONSTRATION OF BIOEQUIVALENCE IN A GENERIC NEBULISER

IVIVC in Pediatric OIPs

Product Guide. MSP Corporation Rice Creek Parkway, Suite 300. Shoreview, Minnesota 55126, U.S.A. Phone: Fax:

COMMENTS. Submitted by The International Pharmaceutical Aerosol Consortium

Device Design Similarity

Understanding the links between drug delivery route and in vitro test methods Mark Copley, Sales Director, Copley Scientific

MDI. Recommended Equipment

Implications of the new CHMP Guideline on the Pharmaceutical Quality of Inhalation and Nasal Products

Last updated: July 2011

EPAG Perspective - Regulatory Advances Related to Nasal Spray Pumps. Dr G.Williams Nasal Drug Delivery Management Forum London, 15 Apr 2010

COPLEY SCIENTIFIC. Quality Solutions for Inhaler Testing. metered-dose inhalers dry powder inhalers nebulisers nasal sprays

Draft Guidance on Fluticasone Propionate; Salmeterol Xinafoate. Fluticasone Propionate; Salmeterol Xinafoate. Powder/Inhalation

VARIABILITY IN CASCADE IMPACTION: SOURCES, IMPACT AND STRATEGIES FOR REDUCTION

Current Challenges and Opportunities in Demonstrating Bioequivalence

I. Subject: Medication Delivery by Metered Dose Inhaler (MDI)

The abbreviated impactor measurement concept

DISSOLUTION TESTING - Exploring the Link Between Particle Size & Dissolution Behavior for OINDPs

Patient. Device Clinician. Safety & efficacy

REDUCING VARIABILITY IN TEST RESULTS FOR OINDPS WITH AUTOMATED ACTUATION

Year in review. Vit Perlik Director of Regulatory Science and Clinical Development

A NEW, ADVANCED HIGH- THROUGHPUT SYSTEM FOR AUTOMATED INHALER TESTING

COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP) GUIDELINE ON THE PHARMACEUTICAL QUALITY OF INHALATION AND NASAL PRODUCTS

Misty Max 10 nebulizer

HARD TWO-PIECE HPMC CAPSULES FOR PHARMACEUTICAL APPLICATIONS IN INHALATION

EPAG Sponsored Workshop on Abbreviated Impactor Measurement (AIM) and Efficient Data Analysis (EDA) Concepts in Inhaler Testing. Overview of AIM EDA

Understanding Regulatory Global Requirements for Nasal Drug Products. Julie D. Suman, Ph.D. April 8, 2016

Inhalation Therapy. Inhalation Therapy

International Pharmaceutical Aerosol Consortium on Regulation and Science

What Do Cascade Impaction Measurements Tell Us: In Vitro Aspects. Context of Presentation

COPLEY SCIENTIFIC. Quality Solutions for Inhaler Testing Edition. metered-dose inhalers dry powder inhalers nebulisers nasal sprays

2008 IPAC-RS Conference Doing the Right Thing Science, Quality and Patient Focus

Process Drift and it s Resolution in the Manufacture of Drug Products. MDI s and DPI s Metered Dose Inhalations and Dry Powder Inhalations

Inhalation devices, proper technique and cleaning

AEROSOL THERAPY: THE PRACTICALITIES

Delay Between Actuation and Shaking of a Hydrofluoroalkane Fluticasone Pressurized Metered-Dose Inhaler

RDD Europe 2009 Workshop

Adapting Pharmacopeial Testing for Pharmaceutical Equivalence Studies. Julie D. Suman, Ph.D. March 29, 2016

Drug/Device Combination Products: Bioequivalence

Predictive modeling of deposition, dissolution, absorption and systemic exposure

PREMIXES FOR MEDICATED FEEDING STUFFS FOR VETERINARY USE. Praeadmixta ad alimenta medicata ad usum veterinarium. Effervescent powders

Using an Inhaler and Nebulizer

IPAC-RS/UF Orlando Inhalation Conference March 20, S.T. Horhota 1, C.B. Verkleij 2, P.J.G. Cornelissen 2, L. Bour 3, A. Sharma 3, M.

SELECTING A DOSAGE FORM FOR DRUG DELIVERY TO THE LUNGS

TRAINING DEVICE OPPORTUNITIES AND SOLUTIONS FOR A CHANGING RESPIRATORY MARKET

COMPARISON OF THE RESPIRABLE FRACTION FROM THREE DIFERENT DPI DEVICES

Citation for published version (APA): Westerman, E. M. (2009). Studies on antibiotic aerosols for inhalation in cystic fibrosis s.n.

Generic Inhaled Medications

Case Study 1: Pharmaceutical Development of EXUBERA

ISO Process and Standards Under Development. ISO Principles

An update on inhalation devices

Testing inhalers. One of the longstanding challenges facing the

Metered Dose Inhalers (MDIs) In Vitro Measures to Confirm Patient Perceptions: HFA vs. CFC

Quality of Nasal and Inhalation Drug Products

innovative products. faster to market. reliably supplied.

Moving to More Realistic In Vitro Testing of OIDPs

Delivery of Iloprost Inhalation Solution With the HaloLite, Prodose, and I-neb Adaptive Aerosol Delivery Systems: An In Vitro Study

NEBULIZERS, METERED DOSE INHALERS, AND DRY POWDER INHALERS

PRACTICES OF COATING COLLECTION SURFACES OF CASCADE IMPACTORS: A SURVEY OF MEMBERS OF THE EUROPEAN PHARMACEUTICAL AEROSOL GROUP (EPAG)

COPD Device Workshop. Summary. Role of inhaler device in COPD. Why use inhaler device in COPD?

CAPSULE-BASED DRY POWDER INHALERS, AN OPTIMAL SOLUTION FOR DIFFERENT INSPIRATIONAL RATES

PPD S EXPERT HEMATOLOGY AND ONCOLOGY TEAM

Effect of Rise in Simulated Inspiratory Flow Rate and Carrier Particle Size on Powder Emptying From Dry Powder Inhalers

Anaesthetic and respiratory equipment Nebulizing systems and components

Beclometasone dipropionate (BDP) Prophylactic management of mild, moderate or severe asthma in adults

Considerations for Evaluation of Bioequivalence and Interchangeability of Orally Inhaled Products

FRI aerosol. deposition. A unique way to look at regional inhaled drug deposition.

Respiratory. Martin Jetzer DDL27 Edinburgh December 2016

Combining imaging techniques and CFD to model lung deposition in various age classes of the paediatric population

905 UNIFORMITY OF DOSAGE UNITS

SELECTING A DOSAGE FORM FOR DRUG DELIVERY TO THE LUNGS. Dr. Jay T. Holt, Ph.D., Sr. Director, Inhalation & Analytical

ISO INTERNATIONAL STANDARD. Anaesthetic and respiratory equipment Nebulizing systems and components

Appendix M: Device Technique

Appendix E: Device Technique

Achieving Optimal Particle Size Distribution in Inhalation Therapy

CLINICAL RELEVANCE OF IN VITRO PARTICLE SIZING DATA. Steve Newman, PhD Nottingham, UK November 2006

Pulmonary deposition of inhaled drugs

Perceptual reasons for resistance to change in the emergency department use of holding chambers for children with asthma

UPDATE ON THE GUIDELINE ON THE PHARMACEUTICAL QUALITY OF INHALATION AND NASAL PRODUCTS

General Chapter/Section: <232> Elemental Impurities - Limits Expert Committee(s): General Chapters Chemical Analysis No.

FRI aerosol. deposition. A unique way to look at regional inhaled drug deposition.

Latex Free. An affordable, easy to use, high density, small volume nebulizer with a breath enhanced design! Breath Enhanced High Density Jet Nebulizer

It is recommended that a mask and protective eyewear be worn when providing care to a patient with a cough

8/13/11. RSPT 1410 Humidity & Aerosol Therapy Part 3. Humidification Equipment. Aerosol Therapy

Invacare Aerosol Products. Stratos Aerosol Compressors Nebulizers Asthma Management

Device Change Management for Inhaled Products. Loy Britto, Ph.D. GlaxoSmithKline ISAM Congress Munich 2015

Use of Math Modelling to Understand Delivery of Biopharmaceutical Molecules to the Lung

Transcription:

As first AppeAred in Inhalation April 2014 www.inhalationmag.com Assessing the role of breathing simulators in OIP testing Exploring how the application of patient-representative inhalation profiles can improve the relevance of test data Mark Copley Copley Scientific Breathing simulators, instruments that generate an inhalation and/or exhalation profile that mimics that of a human subject, have become a routine feature of orally inhaled product (OIP) testing. Recently updated monographs for nebulizers 1,2 and a new draft USP monograph for the testing of pressurized metered dose inhalers (pmdis) with spacers and valved holding chambers (VHCs) 3 specifically call for the application of representative breathing profiles during testing. In addition to these specific protocols, researchers are employing breathing simulators to more robustly scope and fully understand OIP performance. 4-8 This article reviews the capabilities of breathing simulators and their application in nebulizer, dry powder inhaler (DPI) and pmdi testing, clearly differentiating between that which is specified by the pharmacopeias and that which remains optional but potentially valuable. The focus is delivered dose uniformity (DDU) testing and aerodynamic particle size distribution (APSD) measurement. Delivered dose and APSD are considered Critical Quality Attributes (CQAs) for OIPs, variables that define clinical efficacy and consequently drive the majority of in vitro testing. The impact of inhalation A case can be made for the application of breathing simulators for testing all types of OIPs, with the exception of pmdis in the absence of a spacer or a VHC. For nebulizers, dry powder inhalers and pmdis with either a spacer or a VHC, a patient s inhalation profile affects the efficiency of drug delivery. Nebulizers: Nebulizers continuously generate an aerosolized cloud of droplets for inhalation by the patient. The therapeutic dose received therefore depends directly on how effectively the repetitive, tidal breathing cycle draws the droplets into the lungs and on the duration of use. DPIs: The majority of DPIs are passive, which means that the only energy supplied for dispersion and aerosolization of the dose, and subsequent delivery, comes from the inhalation maneuver of the patient. For DPIs, the applied inhalation profile therefore influences the extent to which the dose disperses to a respirable particle size, the emptying of the device and the extent to which the drug is drawn into the lungs. pmdis with spacers/vhcs: In the absence of a spacer or VHC, efficient drug delivery with a pmdi relies on synchronizing device actuation with inhalation. However, beyond this, the drug delivery and aerosol generation process is propellant-driven and relatively insensitive to the breathing profile applied. The situation changes when a spacer or VHC (Figure 1) is used because they separate the pmdi from the patient. Both eliminate the need to precisely coordinate actuation and inhalation, capturing the dose emitted from the pmdi as a reservoir of dispersed particles that can be inhaled in much the same way as with a nebulizer. In doing so, they make pmdis suitable for a wider range of patients. However, such devices simultaneously make the efficiency of drug delivery more dependent on the breathing profile of the patient because retention of a portion of the dispersed aerosol is possible within the spacer/vhc.

Nebulizers: A regulatory requirement for representative breathing profiles New harmonized monographs for nebulizer testing Ph. Eur. 2.9.44 and USP <1601> 1,2 were adopted in January 2012 and August 2011 respectively. These monographs recommend that nebulizers, like other OIPs, are tested as combined products (formulation and device) and include well-defined breathing profiles for this testing that reflect guidance issued by the European Medicines Agency. 9 The new monographs reference four breathing profiles: adult, child, infant and neonate (Table 1). Breathing simulators that produce these profiles are now used routinely as part of delivered dose uniformity testing for nebulizers to measure two discrete parameters: active substance delivery rate and total active substance delivered. These parameters define the rate at which the drug will be inhaled by the patient and the total dose inhaled over a prescribed timeframe. Measurements of APSD for all nebulizers are carried out at a constant flow rate, as required for operation of the cascade impactors used for these measurements. The prescribed flow rate for the testing is 15 L/min, a figure broadly representative of the mid-tidal f low of a typical adult user. Nebulizers are essentially active devices so the droplet size they produce is minimally affected by the breathing profile of the patient. APSD measurement under these conditions is therefore deemed sufficiently representative of in-use performance. The Next Generation Impactor (NGI) has calibrated performance at 15 L/min and is particularly well-suited to nebulizer characterization, mostly because of the substantial capacity of the NGI collection cups for liquid droplets generated by nebulizers. This conclusion is reflected in the new monographs which provide useful guidance for its application. Figure 1 Using a spacer or a VHC (shown) eliminates the need to precisely coordinate pmdi actuation with inhalation, easing product use and increasing drug delivery efficiency. Each VHC has a oneway valve at the patient interface to prevent failed aerosol delivery in the event of exhalation through the device, whereas a spacer is simply a section of open tube. Figure 2 Air drawn through a DPI, by the patient s inhalation action, aerosolizes the powder bolus, ideally dispersing the drug particles to a suitable size (smaller than 5 microns) for deposition in the deep lung. Adapted with permission from M.J. Telko, et al. (September 2005) Dry powder inhaler formulation. Respiratory Care, Volume 50, Number 9, pages 1209-1227. 18 powder reservoir (e.g. turbuhaler) Blister disk (e.g.rotadisk) Blister strip (e.g.diskus) Capsule (e.g.rotahaler) formulation Metering dispersion passive/active Oropharyngeal/ deposition pulmonary delivery Dry powder inhalers: A broader scoping of product performance The passive nature of the vast majority of DPIs makes the mechanisms of drug delivery markedly different from those of other OIPs. Figure 2 illustrates these mechanisms, showing how the energy provided by the patient during inhalation influences both the size to which the particles are dispersed and the effectiveness with which the drug is drawn out of the device and into the lungs. This method of operation makes DPI performance more Table 1 Breathing Simulator Specification for Nebulizer Characterization Tests Adult Neonate Infant Child Total Volume 500 ml 25 ml 50 ml 155 ml Frequency 15 cycles/min 40 cycles/min 30 cycles/min 25 cycles/min Waveform sinusoidal sinusoidal sinusoidal sinusoidal I/E Ratio 1:1 1:3 1:3 1:2

sensitive to variability in the applied breathing profile than any other OIP. This conclusion is reflected in current pharmacopeial tests for DPIs, which are based on establishing product-specific test conditions. The first step in either DDU or APSD measurement is to determine the flow rate that generates a 4 kpa pressure drop over the DPI; a pressure drop deemed to be representative of that which a typical adult asthmatic or COPD patient will generate during use. This flow rate is dependent on the flow resistance of the device under test, with low resistance devices resulting in a higher flow rate than those that present higher resistance to flow (Figure 3). Using this derived flow rate (up to a limit of 100 L/min) the test duration is calculated on the basis of a total test volume, specified to represent the total inspiratory volume of the patient during use. The FDA recommends 2 L 10 and the pharmacopeias favor 4 L to better suit the size-fractioning mechanism of cascade impactors. 11 Both can be argued as representative of a typical adult patient, although the former is probably more representative of an asthmatic or COPD patient. This volume is used to define a square wave profile that is applied during both DDU and APSD measurement. This established method helps to ensure that the critical parameters that drive the aerosolization process during testing are broadly representative of actions that will occur during patient use, within the constraints of the test apparatus. However, it has limitations, most especially for the assessment of DPI performance for: weak patients or those with severely impaired lung function unable to generate a 4 kpa pressure drop during product use 12 healthy patients with a stronger inspiratory capability using a DPI for systemic treatment 13 any patient with a sub-optimal operating technique These limitations are becoming more problematic as the use of DPI technology is extended to, for example, pediatric and geriatric patients who do not have the lung capacity of a healthy adult and the delivery of systemic therapies such as insulin, antibiotics and vaccines to otherwise healthy patients with unimpaired lung function. Further - more, as the industry embraces Quality by Design and also seeks ways to show bioequivalence in the case of generic products, there is a broader requirement to fully scope and understand product performance. This requirement cannot be met completely using current testing methodology, which has evolved principally for the QC environment. Understanding the impact of any variability that may arise from, for example, differences in patient physiology or technique, is becoming increasingly important, prompting wider experimentation, often with alternative breathing profiles. 5,7,8 Figure 3 A product-specific, square-wave flow profile is applied during DPI testing, defined from the flow rate commensurate with a 4 kpa pressure drop across the device and an associated inspiratory volume (2 L or 4 L). flow rate (l/min) 150 100 4 l at 100 l/min 4 l at 60 l/min 50 0 0 1 2 3 4 time (seconds) With the existing pharmacopeial test set-up, the physiology of stronger or weaker patients can be assessed by varying the pressure drop used to determine test flow rate or by changing the test duration to investigate the impact of inspiratory capacity. However, air flow rate is essentially either on or off. Breathing simulators, in contrast, enable the application of variously shaped/dimensioned breathing profiles to investigate the ways performance is impacted by: a more realistic air flow acceleration rate, from zero to peak flow the shape of the breathing profile of specific patient groups (potentially obtained from clinical data) exhalation back into the device, i.e., incorrect operation As previously discussed, in the case of DPIs, both the aerodynamic size of the particles produced (APSD) and the dose delivered can be impacted by the breathing profile. Here, breathing simulators consequently have the potential to bring added value to APSD measurements, as well as to DDU testing. This raises the question of how to impose a defined breathing profile through the DPI while maintaining the constant flow rate through the cascade impactor that is required for its correct and calibrated operation. A mixing inlet is a device that can be used to ad - dress this issue (Figure 4). By decoupling the flow profile applied to the DPI from that which is applied to the impactor, in real-time, a mixing inlet facilitates the application of a breathing profile that reflects the conditions of interest, during DPI testing. 14 The accuracy of APSD measurements is maintained because the flow rate through the impactor remains constant. (See sidebar An optimized set-up for DPI testing? )

An optimized set-up for DPI testing? the test set-up in figure 4 illustrates the way bioequivalence testing can be improved by the combined application of three pieces of equipment that are routinely absent from the standard test set-up: a breathing simulator, an Alberta idealized throat (Ait) (in place of the standard Usp induction port) and a mixing inlet. the breathing simulator and mixing inlet together enable the application of a patient-relevant breathing profile through the dpi, while maintaining the constant flow rate through the impactor that is required for accurate Apsd measurement. this allows the robust demonstration of equivalent drug delivery performance across a range of conditions that represent the variability associated with the target user group. this flexibility to fully scope variability is far greater than with the standard pharmacopeial test set-up. Figure 4 A set-up for more patient-representative DPI testing, combining a breathing simulator, an Alberta Idealized Throat (AIT) and a mixing inlet. the Ait addresses the widely recognized limitations of the standard Usp induction port in accurately representing aerosol transport through the upper respiratory tract. part way between a human throat cast and the simple rightangled tubular design of the Usp induction port, the Ait produces data that are more representative of measured in vivo behavior, thereby supporting a more reliable demonstration of bioequivalence. 15,16 pmdis with valved holding chambers (VHC) and spacers: A new USP monograph The use of a spacer or VHC with a pmdi results in a reservoir of dispersed particles that can be inhaled in much the same way as with a nebulizer. It also allows for the introduction of a time delay between actuation of the MDI and inhalation of the aerosol cloud produced. A new USP monograph, currently in draft form, 3 highlights these facts and brings breathing simulators into the spotlight for the representative testing of this class of OIP. The approach outlined in the draft monograph draws on experience gained in Canada over the past 10 years, since the publication of an existing Health Canada standard defining clinically-appropriate performance test methods. 17 It closely echoes the strategy employed for nebulizers, providing breathing profiles for DDU testing, and suggesting that APSD measurement be carried out at a fixed flow rate, both of which are broadly representative of the target patient group. Defined breathing profiles are recommended for delivered dose measurement, with test conditions specified for adult and pediatric patients: neonate, infant and child. The draft monograph also includes a requirement, unique to pmdis with spacers and VHCs, to assess the impact on delivered dose when actuation and inhalation are not properly coordinated. This reflects the potential use scenarios by the patient: closely coordinate inhalation and device actuation, or device actuation during exhalation prior to inhalation. Testing at the extremes of fully coordinated and fully uncoordinated use, using breathing profiles programmed to start on inhalation or exhalation, helps to fully scope the impact of this source of variability. Looking ahead The availability of cost-efficient, reliable breathing simulators has encouraged their use throughout OIP testing. This technology eases compliance with new pharmacopeia monographs for nebulizers and with a new draft monograph for pmdis with spacers and VHCs. In addition, breathing simulators enable the more robust scoping of product performance for both innovator and generic submissions, especially for DPIs. By helping to quantify the likely impact of variability introduced by patient physiology and technique, they bring significant value, whether the aim is to demonstrate the suitability of a new product for a specific patient group or bioequivalence. References 1. Ph. Eur. Chapter 2.9.44 Preparations for Nebulisation: Characterisation. 2. USP <1601> Products for Nebulization: Characterization. 3. In-Process Revision: <1602> Spacers and Valved Holding Chambers Used with Inhalation Aerosols, USP Phar macopeial Forum 40(1), January 2014. 4. B. Olsson, et al. (2013) Validation of a General In Vitro Approach for Prediction of Total Lung Deposition in Healthy Adults for Pharmaceutical Inhalation Products. Journal of Aerosol Medicine and Pulmonary Drug Delivery, Volume 26, Number 6, pages 335-369. 5. D. Zanker, et al. (2012) In Vitro Dose Delivery Per for mance of Glycopyrronium Using Representative In spiratory Flow Profiles Derived from COPD Patients. Inamed Novartis, Poster presented at DDL23.

6. P.K.P. Burnell, et al. (1998) Design, Validation and Initial Testing of the Electronic Lung Device. Journal of Aerosol Science, Volume 29, Number 8, pages 1011-1025. 7. G. P. Martin, et al. (2006) Influence of Realistic In spiratory Flow Profiles on Fine Particle Fractions of Dry Powder Aerosol Formulations. Pharmaceutical Research, Volume 24, Number 2, pages 361-369. 8. B. Olsson, et al. (2010) Comparing Aerosol Size Dis tributions that Penetrate Mouth-Throat Models Under Realistic Inhalation Conditions. Respiratory Drug Delivery 2010, pages 225-234. 9. CPMP (2006) Guideline on the Pharmaceutical Quality of Inhalation and Nasal Products to be read in conjunction with CPMP (2004) Points to Consider on the Require ments for Clinical Documentation for Orally Inhaled Products (OIP). 10. U.S. Department of Health and Human Services Draft Guidance Document: Metered Dose Inhaler (MDI) and Dry Powder Inhaler (DPI) Drug Products. Published November 11, 1998. 11. H. Mohammed, et al. (September 2012) Effect of Sampling Volume on Dry Powder Inhaler (DPI)-Emitted Aerosol Aerodynamic Particle Size Distributions (APSDs) Measured by the Next-Generation Pharmaceutical Impactor (NGI) and the Andersen Eight-Stage Cascade Impactor (ACI). AAPS PharmSciTech, Volume 13, Number 3, pages 875-882. 12. M. Copley. (2010) Assessing Dry Powder Inhalers. White paper available for download at: http://www.copleyscientific.com/ documents/ww/assessing%20dry%20powder%20inhalers.pdf. 13. D. Harris. (September 2007) Testing Inhalers. Phar - maceutical Technology Europe, pages 29-35. 14. N. C. Miller, et al. (2000) Aerodynamic Sizing With Simulated Inhalation Profiles: Total Dose Capture And Measurement. Respiratory Drug Delivery VII, pages 191-196. 15. M. Copley, et al. (2011) Evaluating the Alberta Throat: An Innovation to Support the Acquisition of More Clinically Applicable Aerosol Aerodynamic Particle Size Distribution (APSD) Data in Oral Inhaled Product (OIP) Development. Inhalation, Volume 5, Number 4, pages 12-16. 16. A. Watkins, et al. (2012) Comparative Performance of Dry Powder Inhalers Assessed Using a Commercially Available Model of the Upper Human Respiratory Tract. Vectura Group plc. Poster presented at Respiratory Drug Delivery 2012. 17. CSA Group Standard, CAN/CSA-Z264.1-02 (R2011) Spacers and Holding Chambers for Use with Metered-Dose Inhalers. 18. M.J. Telko, et al. (September 2005) Dry Powder Inhaler Formulation. Respiratory Care, Volume 50, Number 9, pages 1209-1227. Mark Copley is Sales Director, Copley Scientific, Colwick Quays Business Park, Private Road No. 2, Colwick, Nottingham, UK NG4 2JY, Tel: +44 1159 616229, m.copley@copleyscientific.co.uk. Website: www.copleyscientific.com.

broadest inhalation capabilities. proven results. inhalation BROADEST DEVELOPMENT CAPABILITIES With our complete range of dosage forms and comprehensive, integrated services, we create customized solutions that ensure superior results. MORE INHALATION EXPERIENCE For over 20 years, our deep experience and effective program execution have accelerated inhalation drug development time to market. PROVEN TRACK RECORD Our global project and supply chain management capabilities and proven regulatory expertise ensure high quality service and on-time delivery. 2011 Catalent Pharma Solutions. All rights reserved. Breathe easier. Whatever your dosage form pmdi, DPI, nasal, solution/suspension, or nebulizer we are the catalyst for your success. With our broad range of services and deep industry experience, we create customized inhaled drug development and filling/manufacturing solutions that improve the value of your treatments from early stages to launch. And, our thorough understanding of regulatory requirements and submission protocols ensure ongoing, reliable supply. Catalent. More products. Better treatments. Reliably supplied. DEVELOPMENT DELIVERY SUPPLY Discover more solutions with Catalent. Call: + 1 888 SOLUTION (765 8846) Europe: 00800 88 55 6178 Email: solutions@catalent.com Visit: www.catalent.com