Analysis of Food Sugars in Various Matrices Using UPLC with Refractive Index (RI) Detection

Similar documents
[ APPLICATION NOTE ] Profiling Mono and Disaccharides in Milk and Infant Formula Using the ACQUITY Arc System and ACQUITY QDa Detector

Comparison of a UPLC Method across Multiple UHPLC Systems

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Mark E. Benvenuti, Gareth Cleland, and Jennifer Burgess Waters Corporation, Milford, MA, USA INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS

USP Method Transfer of Ziprasidone HCl from HPLC to UPLC

Rapid Quantitative Analysis of Cannabidiol from Consumer Products Using UltraPerformance Convergence Chromatography

Application Note. Authors: C. Ledesma, M. Gibert, J.R. Gibert Ingenieria Analitica S.L. Extracts from various food products

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Quantification of Budesonide Using UPLC and Xevo TQ-S

[APPLICATION NOTE] DETERMINATION OF AMINO ACIDS IN BEERS USING THE UPLC AMINO ACID ANALYSIS SOLUTION

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis

[ APPLICATION NOTE ] The Separation of 8 -THC, 9 -THC, and Their Enantiomers by UPC 2 Using Trefoil Chiral Columns INTRODUCTION APPLICATION BENEFITS

Matrix Factor Determination with the Waters Regulated Bioanalysis System Solution

Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

Meeting Challenging Requirements for the Quantitation of Regulated Growth Promoters Dexamethasone and Betamethasone in Liver and Milk

Rapid Analysis of Bisphenols A, B, and E in Baby Food and Infant Formula Using ACQUITY UPLC with the Xevo TQD

Profiling Flavonoid Isomers in Highly Complex Citrus Juice Samples Using UPLC Ion Mobility Time-of-Flight Mass Spectrometry

Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

APPLICATIONS TN Determination of Sugars in Animal Feed using HPLC-ELSD

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

HC-75 Calcium Form. 305 x 7.8 mm HC-75 Calcium Form (P/N 79436)

A High Sensitivity UPLC/MS/MS Method for the Analysis of Clopidogrel and Clopidogrel Carboxylic Acid Metabolite in Human K 2 EDTA Plasma

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinElmer Altus UPLC System with PDA Detection

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

GlycoWorks Sample Preparation Consumables

[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

Confident, lower-pressure analysis of carbohydrates, alcohols, and organic acids. Agilent Hi-Plex Ligand-Exchange HPLC Columns

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

Analysis of the Non-Ionic Surfactant Triton-X Using UltraPerformance Convergence Chromatography (UPC 2 ) with MS and UV Detection

Fast Separation of Triacylglycerols in Oils using UltraPerformance Convergence Chromatography (UPC 2 )

CHARACTERIZATION AND DETECTION OF OLIVE OIL ADULTERATIONS USING CHEMOMETRICS

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

[Application Note] TOXICOLOGY SCREENING BY UPLC/PDA IN COMBINATION WITH AN EXTENSIVE COMPOUND LIBRARY

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Waters Amide Column Technology For Food Analysis

Analysis of Common Sweeteners and Additives in Beverages with the PerkinElmer Flexar FX-15 System Equipped with a PDA Detector

Determination of Carbohydrates in Kombucha Using HPAE-PAD

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

HPLC Analysis of Sugars

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

John P. McCauley and Rui Chen Waters Corporation, Milford, MA, USA INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEY WORDS

The Application of UPLC/MS E for the Analysis of Bile Acids in Biological Fluids

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION

Challenges in Developing an Ultra-Sensitive Bioanalytical Method for Ethinylestradiol in Human Plasma

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

Reducing Sample Volume and Increasing Sensitivity for the Quantification of Human Insulin and 5 Analogs in Human Plasma Using ionkey/ms

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.

Quantitative Analysis of Carbohydrates and Artificial Sweeteners in Food Samples Using GFC- MS with APCI Interface and Post-column Reagent Addition

LC-Based Lipidomics Analysis on QTRAP Instruments

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

A RAPID AND SENSITIVE UPLC/UV/MS METHOD FOR SIMVASTATIN AND LOVA S TAT IN IN SU P P O RT O F C L E A NING VA L I DAT IO N S T U DIES

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

Method Development for the Analysis of Endogenous Steroids Using Convergence Chromatography with Mass Spectrometric Detection

Determination of Patulin in Apple Juice Using SPE and UHPLC-MS/MS Analysis

A Metabolomics Approach to Profile Novel Chemical Markers for Identification and Authentification of Terminalia Species

ANALYSIS OF STARCH USING THE ALLIANCE GPCV2000 SYSTEM

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

SACCHARIDES (Liquid Chromatography)

METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE

Application Note. Sensitive analysis of the lactose content of lactose-free labelled products using HPAEC-PAD. Summary.

A New Stability-Indicating and Validated RP-HPLC Method for the Estimation of Liraglutide in Bulk and Pharmaceutical Dosage Forms

Analysis of Counterfeit Antidiabetic Drugs by UHPLC with the Agilent 1220 Infinity Mobile LC

Analysis of Cannabinoids in Cannabis by UHPLC Using PDA Detection

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

[ APPLICATION NOTE ] UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research INTRODUCTION APPLICATION BENEFITS

Jose Castro-Perez, Henry Shion, Kate Yu, John Shockcor, Emma Marsden-Edwards, Jeff Goshawk Waters Corporation, Milford, MA, U.S. and Manchester, UK

A UPLC/MS Approach for the Diagnosis of Inborn Errors of Metabolism

Dosage of sugars in wine by HPLC (Resolution Oeno 526/2016)

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

Direct Analysis of Urinary Opioids and Metabolites by Mixed-Mode µelution SPE Combined with UPLC/MS/MS for Forensic Toxicology

Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS

Thermo Scientific Dionex CarboPac MA1 Column

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Application Note. Author. Abstract. Introduction. Food Safety

Iodine HPLC Assay. Catalog Number: IOD34-H Tests For Research Use Only. Not for use in diagnostic procedures.

INTRODUCTION CH 3 CH CH 3 3. C 37 H 48 N 6 O 5 S 2, molecular weight Figure 1. The Xevo QTof MS System.

ApplicationNOTE A MULTI-RESIDUE LC/MS/MS METHOD FOR THE DETERMINATION OF 81 PESTICIDE RESIDUES IN FRUIT AND VEGETABLES: PART 1, METHOD OVERVIEW

Hypersil BDS and Hypersil Classical HPLC Columns

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

[ CARE AND USE MANUAL ] GlycoWorks Single Use Sample Preparation Kit CONTENTS

Journal of Chemical and Pharmaceutical Research, 2013, 5(1): Research Article

Analysis of HMF by HPLC

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

John Haselden 1, Gordon Dear 1, Jennifer H. Granger 2, and Robert S. Plumb 2. 1GlaxoSmithKline, Ware, UK; 2 Waters Corporation, Milford, MA, USA

3-Acetyldeoxynivalenol. 15-Acetyldeoxynivalenol

VITAMIN B1 & B6 VITAMIN B1&B6 COMBI-DETERMINATION IN EDTA- OF HEPARINE WHOLE BLOOD. HPLC & UPLC Method

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

Separation of Saccharides Using TSKgel Amide-80, a Packing Material for High-performance Normal Phase Partition Chromatography (2) Table of Contents

Improving Food Process Control Using the

Comprehensive Study of SLE as a Sample. Preparation Tool for Bioanalysis

Transcription:

Analysis of Food Sugars in Various Matrices Using UPLC with Refractive Index (RI) Detection Mark E. Benvenuti Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Using UPLC with RI detection enables the detection of the food sugars fructose, glucose, sucrose, maltose, and lactose in a variety of matrices resulting in the following benefits: Run time of less than 3.5 minutes increases throughput. Isocratic elution eliminates the need for column re-equilibration between injections. Method allows the use of robust UPLC BEH Amide column chemistry for sugar separation. INTRODUCTION Sugars are found in a variety of food matrices as either naturally occurring or artificially added. Fructose, glucose, and sucrose are important constituents of various fruit juices. Maltose is found in products derived from corn and grain products. Lactose, also known as milk sugar, exists in dairy products. This set of sugars is known as the five food sugars. Analysis of these sugars is important for quality control purposes, or to determine authenticity or adulteration of food products. Due to the lack of chromophores on these molecules, typical LC methods use evaporative light scattering (ELS), electrochemical, or refractive index (RI) detection. In many laboratories, RI is the detection method of choice because it is simple to use and needs no equilibration between injections. RI provides excellent repeatability and shows a linear response for quantification. There is no need for a clean nitrogen supply, which provides an economic benefit and is ideal for a manufacturing facility. In laboratories where RI detection is the preferred detection method for these samples, a UPLC-compatible version was required to take advantage of the benefits of sub-2-µm particle technology. This application note shows the efficacy of the Waters ACQUITY UPLC H-Class System with the ACQUITY RI Detector to analyze sugars in several different food matrices with a run time of less than 3.5 minutes. WATERS SOLUTIONS ACQUITY UPLC H-Class System ACQUITY Refractive Index Detector ACQUITY UPLC BEH Amide Column Chemistry KEY WORDS Food sugars, BEH Amide column chemistry Figure 1. ACQUITY UPLC H-Class System with RI Detector. 1

EXPERIMENTAL LC conditions System: ACQUITY UPLC H-Class Run time: 3.5 min Column: ACQUITY BEH Amide 2.1 x 5 mm, 1.7µm Temp.: 85 C Mobile phase: 77:23 acetone/water with.5% triethylamine (TEA) Flow rate:.15 ml/min Injection volume: 1. µl Detection: ACQUITY Refractive Index Detector Data acquisition: 2 points/s Time constant: Normal Flow cell: 4 C Standard preparation A stock solution of fructose, glucose, sucrose, maltose, and lactose was prepared by dissolving ~1 gram of each of these sugars in water and diluting to 1 ml with same. From this stock, six dilutions were made in 5:5 acetonitrile/water to create working standards with concentrations listed in Table 1. The fructose high standard was ~14 ppm greater than that in the other sugars due to the amount weighed for the stock, extending the curve for this analyte compared to the other sugars. Sample preparation Individual samples of orange, white grape, and pineapple juices were centrifuged at 4 rpm for 3 min to remove solids. The supernatants were diluted 1:5 with 5:5 acetonitrile/water, and injected. A corn syrup sample was prepared by dissolving 1.24 g syrup in water, and further diluting to 5 ml. 5 µl of this was diluted with the same amount of 5:5 acetonitrile/water, then injected. A sample of low fat milk was diluted 1:1 with 5:5 acetonitrile/water, and centrifuged at 4 rpm for 3 min to remove the precipitated lipids. A portion of the supernatant was collected for injection. Std Fructose Glucose Sucrose Maltose Lactose 1. 655. 5125. 511. 565. 5165. 2. 262. 25. 244. 226. 266. 3. 131. 125. 122. 113. 133. 4. 65.5 512.5 511. 56.5 516.5 5. 52.4 41. 48.8 45.2 413.2 6. 26.2 25. 24.4 22.6 26.6 Table 1. Concentrations of the mixed sugar standards in ppm (mg/kg). 2

RESULTS AND DISCUSSION The resulting chromatograms from injections of each of the six standards are shown in Figure 2. Note that all target analytes elute in less than 3.5 minutes. Figure 3 shows the linear calibration curves for each of the five food sugars where R 2 values are greater than.999 for all compounds. 5 48. 46. 44. 42. 4 38. 36. 34. 32. 3 28. 26. 24. 22. 2 18. 16. 14. 12. 1 8. 6. 4. 2. -2. -4. -6. 1 2 3 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2. 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 3. 3.1 3.2 3.3 3.4 3.5 4 5 1- Fructose 2- Glucose 3- Sucrose 4- Maltose 5- Lactose Figure 2. Overlay of the chromatograms from the six standards of differing concentrations. 2 1 2 1 Fructose R 2 =.9992 Glucose R 2 =.99951 2 1 Sucrose R 2 =.99992 1 Maltose R 2 =.99991 2 1 Lactose R 2 =.99988 1 2 3 4 5 6 7 8 9 Amount Figure 3. Calibration curves for the five food sugars. 3

To assess the utility of the method for the quantification of sugars in fruit juices, three fruit juice samples were analyzed, and the resulting chromatograms are shown in Figure 4. Differences in sugar content and ratios for each juice were quite apparent. The ratio of fructose, glucose, and sucrose was ~1:1:2, while the white grape juice contains fructose and glucose, but negligible sucrose. This is consistent with the literature. 1 To assess the retention time and amount repeatability, each juice was injected three times. The resulting data are shown in Table 2. The %RSD for retention time and amount for all analytes was less than.11 and 1.25, respectively. 12 1 8 6 4 1 2 3 Orange 1- Fructose 22.9 g/l 2- Glucose 21.4 g/l 3- Sucrose 53.4 g/l 2 12 1 8 1 2 White grape 1- Fructose 79.5 g/l 2- Glucose 77.2 g/l 6 4 2 12 1 8 6 4 1 2 3 Pineapple 1- Fructose 32.3 g/l 2- Glucose 34.6 g/l 3- Sucrose 44.6 g/l 2 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2. 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 3. 3.1 3.2 3.3 3.4 3.5 Figure 4. Chromatograms of orange, white grape, and pineapple juices. Sugar Orange White grape Pineapple RT Amount RT Amount RT Amount Fructose.14 1.218.48.365.66.371 Glucose.38.827.35.659.49.321 Sucrose.35.588 N/A N/A.43.429 Table 2. Reproducibility data (%RSD) for retention time and amount for three injections of each fruit juice. 4

The chromatograms for corn syrup and low fat milk are shown in Figures 5 and 6. These samples show very different profiles from those expected for fruit juices. The supermarket corn syrup sample used here consists primarily of glucose and maltose, as shown in Figure 5. Unlike high fructose corn syrups, where cornstarch has been enzymatically converted to glucose and then to fructose, this sample still has high levels of maltose. Lactose is a food sugar found in dairy products and is present in the chromatogram of the milk sample in Figure 6. 1 2 9.5 9. 8.5 8. 7.5 2- Glucose 154.4 g/l 4- Maltose 14.4 g/l 7. 6.5 6. 5.5 5. 4.5 4 4. 3.5 3. 2.5 2. 1.5 1..5 -.5 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2. 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 3. 3.1 3.2 3.3 3.4 3.5 Figure 5. Chromatogram of corn syrup showing calculated concentration of the two sugars; glucose and maltose. 21. 2 19. 18. 17. 16. 5 5- Lactose 47.9 g/l 15. 14. 13. 12. 11. 1 9. 8. 7. 6. 5. 4. 3. 2. 1. -1. -2. 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2. 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 3. 3.1 3.2 3.3 3.4 3.5 Figure 6. Chromatogram of low fat milk showing the presence of lactose. 5

The use of UPLC Technology enables highly efficient, rapid sugar separations in less than 3.5 minutes. The BEH Amide column chemistry allows for the use of basic modifiers and higher column temperatures to improve peak shape and quantification. The ACQUITY Refractive Index Detector s low internal volume delivers low dispersion that is compatible with narrow UPLC peak volumes, while still delivering stable baseline performance for reliable quantitative results, evident in the sugar analysis. Reference 1. Sanz et al. Inositols and Carbohydrates in Different Fresh Fruit Juices. Food Chemistry. 24; 87: 326. CONCLUSIONS This application note demonstrates a rapid isocratic method for the analysis of food sugars. Waters ACQUITY UPLC H-Class System with RI detection provides the following benefits: Rapid analysis time with no column equilibration necessary A reduction in turnaround time resulting in increased throughput Low eluent flow rate reduces solvent cost and waste disposal Waters, ACQUITY UPLC, ACQUITY, and UPLC are registered trademarks of Waters Corporation. The Science of What s Possible is a trademark of Waters Corporation. All other trademarks are the property of their respective owners. 213 Waters Corporation. Produced in the U.S.A. May 213 724677EN LL-PDF Waters Corporation 34 Maple Street Milford, MA 1757 U.S.A. T: 1 58 478 2 F: 1 58 872 199 www.waters.com