Scottish Health Technologies Group

Similar documents
Portable Bladder Ultrasound. OHTAC Recommendation. Portable Bladder Ultrasound

They are updated regularly as new NICE guidance is published. To view the latest version of this NICE Pathway see:

In keeping with the Scottish Diabetes Group criteria, use should be restricted to those who:

Purpose. Policy Statement. Principles. Applicability. Responsibility

Incontinence in neurological disease

Advice Statement. Advice Statement November Advice for NHSScotland. Why is SHTG looking at this topic?

pat hways Medtech innovation briefing Published: 22 January 2016 nice.org.uk/guidance/mib50

Rezūm procedure for the Prostate

Prostate artery embolisation for benign prostatic hyperplasia

The Neurogenic Bladder and Telemedicine: urologic history and physical exam development with a. Objectives

OHTAC Recommendation

NHSScotland is required to consider the Scottish Health Technologies Group (SHTG) advice.

a guide to Reimbursement of Intermittent Catheters Know your options M2116N 04.08

Type of intervention Treatment. Economic study type Cost-effectiveness analysis.

Loss of Bladder Control

Interventional procedures guidance Published: 12 October 2016 nice.org.uk/guidance/ipg566

Quality standard Published: 18 September 2013 nice.org.uk/guidance/qs45

To instruct nursing to safely and accurately perform non-invasive measurement of urinary bladder volume using a portable ultrasonic bladder scanner.

Intermittent self-catheterisation

Incontinence: Risks, Causes and Care

Bulkamid. Patient Information. Obstetrics & Gynaecology Department

Diagnosis and Management of UTI s in Care Home Settings. To Dip or Not to Dip?

Suprapubic catheter insertion in the radiology department. Information for patients Urology

Lower Urinary Tract Symptoms K Kuruvilla Zachariah Associate Specialist

Cryotherapy for recurrent prostate cancer. Issue date: May 2005

Hong Kong College of Surgical Nursing

EAU GUIDELINES POCKET EDITION 3

Types of Intermittent Catheters and Access to Urological Supplies

Tied with pneumonia as the second most common type of healthcareassociated

PROTOCOL FOR BLADDER CARE MANAGEMENT DURING INTRAPARTUM AND POSTNATAL PERIOD

Urinary Catheters. Prevalence of Infections

Overview. Urology Dine and Learn: Erectile Dysfunction & Benign Prostatic Hyperplasia. Iain McAuley September 15, 2014

Clinical guideline Published: 23 May 2010 nice.org.uk/guidance/cg97

Placename CCG. Policies for the Commissioning of Healthcare

CAUTI CONFERENCE CAUTI Prevention and Appropriate Use of Indwelling Urinary Catheters in the Hospital Setting

Intermittent self-catheterisation (ISC) Information for patients Spinal Injuries

Urinary incontinence. Urology Department. Patient Information Leaflet

Intermittent Catheterisation

Interventional procedures guidance Published: 28 June 2017 nice.org.uk/guidance/ipg583

Diagnosis and treatment of overactive bladder (non-neurogenic) in adults: AUA/SUFU guideline.

FreshRN Podcast Season 4, Episode 6. All Things Urinary Catheters

Glossary of terms Urinary Incontinence

GUIDELINES ON URINARY INCONTINENCE

Intermittent Self-Catheterisation / Catheterisation Continence Advisory Service Community & Therapy Services North Lincolnshire

D-MANNOSE, CRANBERRY AND VITAMIN C (CYSTOMAN 100MG) ARE EFFECTIVE IN PREVENTING URINARY TRACT INFECTIONS IN MULTIPLE SCLEROSIS SUBJECTS.

Mrs Maya Harris Consultant Urologist INDWELLING CATHETERS: COMPLICATIONS AND SOLUTIONS

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE SCOPE

Indwelling urinary catheter

Medical technologies guidance Published: 10 September 2018 nice.org.uk/guidance/mtg38

Implementing Clean Intermittent Catheterisation (CIC)

Catheter-associated Urinary Tract Infection (CAUTI) Toolkit

Medical technologies guidance Published: 2 October 2018 nice.org.uk/guidance/mtg39

Advice Statement 005/18 April 2018

The inflow. Active Urologic Practice. Clinical Background and Urology Practice Overview

A Novel New Approach to Treatment of Catastrophic Urinary Dysfunction

No Catheter No CAUTI. The Urinary System. Urinary Catheters. External catheters. Intermittent catheters. Indwelling Urethral Catheters

Policy for Prostatism/Lower Urinary Tract Symptoms in men

UTI IN ELDERLY. Zeinab Naderpour

Ostomy and Urological Supply Categories are Not Well Suited for the Medicare Competitive Bidding Program March 22, 2018

Post operative voiding dysfunction and the Value of Urodynamics. Dr Salwan Al-Salihi Urogynaecologist Obstetrician and Gynaecologist

Loss of Bladder Control

Lower Urinary Tract Infection (UTI) in Males

Integrated Continence Service Policy. January SafeCare Council January Carol Giffin, Continence Advisor

Surveillance report Published: 7 July 2016 nice.org.uk. NICE All rights reserved.

BLADDERSCAN PRIME PLUS TM DEEP LEARNING

Technology appraisal guidance Published: 21 November 2018 nice.org.uk/guidance/ta546

Medical technologies guidance Published: 23 January 2019 nice.org.uk/guidance/mtg41

Complications Audit of Urological Issues in Spinal Cord Injury Evaluation Study (CAUSES)

Bladder issues were reported as one of the top three health priorities In the Ageing Well Survey, 2007.

Management of LUTS. Simon Woodhams February 2012

Sacral Nerve Stimulation (SNS) System Implantation

Flexible Cystoscopy. Patient Information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE SCOPE

Neurogenic bladder. Neurogenic bladder is a type of dysfunction of the bladder due to neurological disorder.

q7:480499_P0 6/5/09 10:23 AM Page 1 WHAT YOU SHOULD KNOW ABOUT YOUR DIAGNOSIS OF STRESS URINARY INCONTINENCE

Hydration Practices and Urinary Incontinence Care Planning

Managing urinary retention for a young man

Introduction. Spinal Cord Injury (SCI) Is it necessary to perform surveillance investigations for long term follow up of spinal cord injury patients?

Clinical practices in catheter

Describe the 2017 expectations of NPSG Effectively evaluate compliance with NPSG

User Appreciation of the VaPro Intermittent Catheter Characteristics and Packaging

Cost-effectiveness of urodynamic testing before surgery for women with pelvic organ prolapse and stress urinary incontinence Weber A M, Walters M D

Surgical repair of vaginal wall prolapse using mesh

INTRAPARTUM AND POSTNATAL BLADDER CARE

Preventing catheter-associated urinary tract infections: The CAUTI Bundle

Haemorrhoidal artery ligation

O3_A2_A_Scientific Evidence

Slide 1. Slide 2. Slide 3. Bladder Management and Preventing CAUTI. Bladder management CAUTI prevention standards

Discussions about having a Supra pubic catheter

MP A Prospective Evaluation of the Catheter Science M3 Mini Catheter for Patients with Prostatic Obstruction. Gaines W. Hammond Jr.

CONTINENCE MODULE 1 MIMIMUM STANDARDS FOR THE SPECIALIST ASSESSMENT & CONSERVATIVE MANAGEMENT OF FEMALE LOWER URINARY TRACT SYMPTOMS

Clinical Model for IC 5

GUIDELINES ON NEUROGENIC LOWER URINARY TRACT DYSFUNCTION

Incontinence. Anatomy The human body has two kidneys. The kidneys continuously filter the blood and make urine.

Continence Promotion in

Bladder Management. A guide for patients. Key points

Module 5 Management Of Urinary Incontinence

Diane K. Newman DNP, ANP-BC, PCB-PMD, FAAN

A randomized crossover study of silver-coated urinary catheters in hospitalized patients Karchmer T B, Giannetta E T, Muto C A, Strain B A, Farr B M

Transcription:

Scottish Health Technologies Group Advice Statement Portable bladder ultrasound scanners 8 March 2011 Advice Statement 002/11 Question: Can the use of a portable 3D ultrasound bladder scanner to measure bladder volume reduce the need for urinary catheterisation and hence catheter associated urinary tract infections? Advice: SHTG advise that to ensure evidence-based purchasing for portable bladder ultrasound scanners (PBUS), further research is required on their accuracy, suitability and cost effectiveness. Evidence Note 32 considered the following PBUS; BVI 3000, 6100, 6200 and 2500 ; Bardscan ; Sonosite 180 and ilook 15 ; HDI 4000 ; and BME-150A. Results reported show that these are of sufficient clinical accuracy in estimating post-void residual urine volume in most patients compared with sterile catheterisation. There is however conflicting evidence regarding their use in children aged less than 36 months and postnatal women. There is also insufficient evidence available to conclude that any of these models demonstrated superior clinical accuracy. There is some evidence that these devices are as effective as catheterisation, however cost effectiveness is unknown. No relevant economic evidence was identified, thus it was not possible to comment on the cost effectiveness of PBUS. Advice context: No part of this advice may be used without the whole of the advice being quoted in full. This advice represents the view of SHTG at the date noted. It is provided to inform NHS boards in Scotland when determining the place of health technologies for local use. The content of this Advice Statement was accurate and based upon the most up-to-date evidence available at the time of publication. Readers are asked to consider that new trials and technologies may have emerged since first publication and the evidence presented may no longer be current. SHTG Advice Statements will be reviewed on a 2-yearly basis. The evidence will be updated if requested by the clinical community, dependent on new published reports. This advice does not override the individual responsibility of health professionals to make decisions in the exercise of their clinical judgment in the circumstances of the individual patient, in consultation with the patient and/or guardian or carer. Chairman Scottish Health Technologies Group National Procurement File Name: 2011-002 PBUS v1.0 Version: 1.0 Page 1 of 7

SUMMARY OF EVIDENCE Portable bladder ultrasound scanners September 2010 Evidence Note 32 is available for download from the Healthcare Improvement Scotland website. Key points Portable bladder ultrasound scanners are of sufficient clinical accuracy in estimating PVR urine volume in most patients. Studies however report conflicting results regarding the use of the device in children aged less than 36 months and postnatal women. There was insufficient evidence available to conclude that any of the models of PBUS compared demonstrated superior clinical accuracy. Evidence suggests that adopting standard protocols, including staff training and identifying appropriate patients, for the use of PBUS is associated with reduced risk of unnecessary catheterisation, reduced UTIs and fewer adverse events as well as improved patient satisfaction, when compared with catheterisation. No relevant economic evidence was identified, thus it was not possible to comment on the cost effectiveness of PBUS. Further research is required on the accuracy and suitability of newer portable bladder ultrasound scanners so that NHS purchasing can be evidence based. Health technology description There are two methods for assessing post-void residual (PVR) urine volume: sterile catheterisation (direct measurement of urine volume) and bladder ultrasound (indirect estimation of urine volume) 1. Bladder ultrasound scanning offers a non-invasive alternative to catheterisation. The scanner is used, primarily as a diagnostic aid, to noninvasively identify incomplete bladder emptying and determine bladder volume 2,3 in adults and children with urinary problems of varying aetiologies, such as 4 : postoperative patients at risk of urinary retention (UR); patients with UTIs, urinary incontinence (UI), enlarged prostate, urethral stricture, neurogenic bladder and other lower urinary tract dysfunctions; and patients with conditions that interfere with voiding, such as spinal cord injuries, stroke and diabetes. Due to its non-invasive nature, the use of ultrasound technology may reduce the risk of urethral trauma and urinary infection associated with catheterisation. Stationary and portable ultrasound machines are available in both 3-dimensional (3D) and 2- dimensional (2D) versions. The PBUS offer potential benefits that the stationery machines do not due to their portability and ease of use 2,5 which permits the use of the device in community settings outside of the hospital environment. File Name: 2011-002 PBUS v1.0 Version: 1.0 Page 2 of 7

The portable, battery-powered, ultrasound scanner utilises automated technology to digitally register bladder volume, including PVR urine volume, while providing images of the bladder area 2,6-8. It consists of a base component with a display screen and a handheld ultrasound transducer (scan head) which is usually positioned on the patient s abdomen and pointed towards the bladder. The base component employs ultrasound technology, through electrical connection to the scan head, to automatically create an image of the bladder and calculate bladder and PVR measurements 2,9. Use and interpretation of results of PBUS can be complex, particularly in patients who are morbidly obese 2, have irregular bladder shapes or in postnatal women (immediately after childbirth) as the device records the volume of cystic structures within the pelvis 10 and so interpretation of results requires specific knowledge and high levels of clinical competence 11. Some authors have however suggested that although training is required to interpret results, only minimal technical skills are required to use the device 3. In the past, the BladderScan series, manufactured by Verathon Medical, were the most common PBUS 6,12. However, a number of alternatives have recently been launched on the United Kingdom (UK) market. These include the Bardscan, Cubescan and different Sonosite models. Some of these portable scanners, like the Bardscan and Cubescan provide real-time ultrasound imaging. There are however other scanners that provide only numeric bladder volumes with no image of the bladder shape. Epidemiology As detailed above, PBUS may find application in a number of bladder problems of varying aetiologies. Often the problem can manifest as either UI, UR or in women more commonly as recurrent UTIs. UI is defined as any involuntary urinary leakage 13 that may occur due to a number of abnormalities in the lower urinary tract function or as a result of other illnesses 14. Risk factors for UI include age, female gender, obesity, lower urinary tract symptoms, functional and cognitive impairment, pregnancy, childbirth, diet, smoking, genetics and family history 2,14. UI is under-diagnosed 14. The Leicestershire Medical Research Council (MRC) Incontinence Study 15 reported that 33.6% of individuals over 40 years of age have significant urinary symptoms. Of these, 6.2% found the condition bothersome while 2.4% found it both bothersome and socially disabling 15. Data from Information Statistics Division (ISD) Scotland estimate the prevalence of current UI in adults as 9% 16. UR is defined as the inability to pass urine despite persistent effort 17. The definition of UR, based on PVR urine volume, is influenced by the population and clinical condition of interest 2. There is no clear consensus regarding what constitutes a normal or abnormal residual volume; clinically significant volumes could vary between a lower limit of 50 ml and upper limit of 300 ml 2,14,18. UTIs are a common complication of urinary catheters that can extend hospital stays. It has been reported that approximately 40% of all hospital-acquired (nosocomial) infections can be attributed to UTIs and about 80% of UTIs (32% of nosocomial infections) are associated with both indwelling and intermittent catheterisation 2,19. Catheterisation is associated with a 5% risk of urinary infection in long-term care patients 5. CAUTI surveillance is not mandatory in Scotland. The Scottish National File Name: 2011-002 PBUS v1.0 Version: 1.0 Page 3 of 7

Healthcare Associated Infection Prevalence Survey 20, conducted between 2005 and 2006, estimated the percentage of urology patients with catheters as 40.3%. The greatest proportion of patients catheterised were identified in surgical, medical and elderly specialties 20. Clinical effectiveness The literature available for PBUS is variable in methodological quality (mostly nonexperimental designs involving small numbers of participants and varying requirements for staff training), thus limiting generalisability and the confidence which can be placed in any conclusions drawn. Accuracy of PBUS A National Institute for Health and Clinical Excellence (NICE) guideline 14, published in 2006, on the management of UI in women was identified. Based on three studies (with residual urine volumes cut-offs ranging between 50 200 ml) that assessed the accuracy of PBUS compared with catheterisation in men and women, the guideline concluded that the bladder scanner is less invasive with lower adverse effects. The sensitivity (range 67 95%) and specificity (range 63 99%) of ultrasound in detecting PVR volumes were found to be within clinically acceptable limits. However, one study showed a much lower sensitivity at residuals >100 ml. The findings led the guideline to recommend that a bladder scan should be used in preference to catheterisation in the measurement of PVR volumes, on the grounds of acceptability and reduced rate of adverse events. However, due to lack of evidence for what constitutes a clinically significant residual, this recommendation may be limited only to women with signs and symptoms suggestive of voiding dysfunction. Similar to the NICE guideline 14, there was general consensus amongst other secondary evidence 2,9,21-24 that although PBUS are not as accurate as catheterisation and require training, they offer acceptable levels of clinical utility and reduced risk of UTIs and improved patient satisfaction due to their non-invasive nature. Results of the studies suggest that the benefits resulting from the use of PBUS outweigh the risks and exceed the benefits of catheterisation. Based on a review of 17 non-randomised controlled trials (RCT) (prospective, observational studies with contemporaneous control), a health technology assessment (HTA) conducted in 2006 by the Ontario Medical Advisory Secretariat reported significant results regarding the negative health outcomes that were avoided due to the use of the PBUS 2. Catheterisations avoided ranged from 16 47% whilst UTIs were reduced by 38 72%. Subgroup analysis showed that PBUS are less accurate for women than men. Higher levels of accuracy were demonstrated in patients with spinal cord injuries compared with other acute care and rehabilitative patients.. There was variability in the type and model of PBUS used in the studies 2. A recent primary study 25 of 101 women with an indication for PVR measurement concluded that the BladderScan BVI 3000 is an accurate alternative to catheterisation. Results from this ultrasound scanner were highly reproducible and correlated significantly with catheterised volumes, with a mean difference of 12.9 ml (p<0.001) 25. NB: this literature 25 was not included in the identified secondary evidence. Accuracy of PBUS in specific population groups Additional primary literature studies which investigated the clinical utility and user satisfaction of PBUS in specific population groups were also identified. File Name: 2011-002 PBUS v1.0 Version: 1.0 Page 4 of 7

Assessment of residual urine in postpartum women Three observational studies 10,26,27, including one UK-based study 10 assessed the reliability of PBUS compared with catheterisation in postpartum women. The residual urine volume indicative of a positive result ranged from 100 ml to 400 ml. Two studies 26,27 concluded that the BladderScan BVI 3000 27 and BVI 6100 26 were reliable and non-invasive screening methods for detecting postpartum urinary retention after vaginal delivery and for reducing the number of unnecessary catheterisation, when used by trained nurses. Using a 400 ml threshold, one study 26 estimated the sensitivity and specificity of the BVI 6100 as 76% and 96% respectively. Despite obtaining clinically irrelevant differences in measured volumes, the UK-based study 10 reported that the BVI 3000 is not as accurate as urethral catheterisation for determining residual urine volume in the puerperium. Volumes measured by PBUS were overestimated (by 9 ml) and underestimated (by 33 ml) compared with volumes measured by standard real-time ultrasound and catheterisation respectively. Assessment of bladder volume in children Four randomised studies 28-31 and four non-randomised observational studies 32-35, published between 2005 and 2008, evaluated the accuracy of PBUS (different BVI models and Sonosite 180 ) in children in various settings. All four fairly large studies 28,30-32 investigating the clinical utility of the Sonosite 180 found that firstattempt urine collection success rates (collection of 2 2.5 ml) in children 36 months or younger were significantly improved by the bladder ultrasound device (range 92 100%) compared with the conventional use of catheters without imaging (range 67 78%). This reduction in unsuccessful urethral catheterisation was also deemed clinically meaningful. Conflicting findings however were reported in other studies 29,33-35, including one RCT 29, that utilised various BVI models in children. The studies 29,33-35 found that the portable bladder scanner was not reliable in assessing bladder volume in children aged under 36 months. Compared with catheterisation, the BVI 6200 and BVI 3000 underestimated bladder volumes in two studies 29,34 while the BVI 3000 failed to detect significant volumes in a small study with neonatal cases (n=10) 35. Furthermore, volumes were overestimated in a study comparing the BVI 2500 with standard ultrasound 33. Nonetheless, reliability was found to be within the acceptable limits in children older than 36 months 33,34 and those with bladder volumes greater than 20% of the expected bladder capacity for age 33. The authors of the studies concluded that although the PBUS could be used as a diagnostic aid for assessment of bladder filling (as there is currently no better alternative which reduces the risk of UTIs associated with unnecessary catheterisation), it needs to be used with caution in children less than 36 months 33-35, particularly neonates with complex problems 35, and should not replace current clinical assessment using catheters 29. Greater satisfaction and a higher likelihood of using bladder ultrasound compared with conventional catheterisation were reported by both caregivers and health care providers in a prospective RCT 36. Using a seven-point Likert scale (1= none, 7= a great deal ), caregivers in the catheterisation group rated the children s discomfort higher (4.4 versus 3.4; p=0.02) and were less satisfied (4.5 versus 6.4; p <0.0001) than those in the ultrasound group 36. However, another randomised study found no significant difference in caregiver satisfaction between the use of ultrasound and catheterisation in File Name: 2011-002 PBUS v1.0 Version: 1.0 Page 5 of 7

a paediatric population 31. It is likely that this study was underpowered to detect a difference in satisfaction scores 31. Comparative effectiveness of different portable bladder ultrasound scanners A UK study 12 investigated the accuracy of four commonly used portable bladder scanners (BVI 3000, BVI 6100, Sonosite ilook 15 and Bardscan ) in 28 healthy volunteers. Relative to the average of all four scanners, the ilook 15 and Bardscan under-predicted bladder volumes by 34 ml and 21 ml respectively, while the BVI 3000 and BVI 6100 over-predicted bladder volumes by 17 ml and 38 ml respectively (p<0.05). Volumes obtained from the ilook 15 and Bardscan were about 50 ml lower than both BVI models. Although not statistically significant (p=0.051), the BVI 3000 had a slightly greater error variability compared with the other scanners. The authors concluded that no scanner could be classed as being the most accurate. The BVI 6100 was however found to be the fastest and lightest scanner with a markedly less variable error 12. Another UK study 37 comparing the BVI 3000, Bardscan and another 3D-ultrasound system (HDI 4000 ), in a random sequence, concluded that the 3D-ultrasound system is the most accurate of all three scanners. Estimated volumes from all three ultrasound scanners were found to be significantly correlated to voided volumes (p <0.001). The Bardscan and BVI 3000 underestimated and overestimated respectively, large bladder volumes. Voided volumes were significantly underestimated by 21 ml by the Bardscan (p=0.008). This finding led the authors to also conclude that the Bardscan is not as accurate as the BVI 3000 despite having the advantage of producing realtime images and instantaneous volume calculation. Three studies 25,38,39 that compared BVI models with catheterisation 25,38 and other ultrasound devices 38,39 found the accuracy of the BVI scanners sufficient for clinical practice. In a population of 89 men and women with lower urinary tract symptoms, Choe et al 38 found no significant difference in the accuracy and precision in estimating residual volumes between the BME-150A (S & D Medicare Co., Seoul, Korea) and BVI 3000. Both devices were found to be clinically acceptable alternatives to urethral catheterisation. The portable 3D (BVI 2500 ) and stationary 2D (Vingmed CFM 800 ) ultrasound scanners also demonstrated sufficient accuracy for clinical practice in a study involving patients with permanent bladder catheters 39. One study 40 that assessed the accuracy of the BVI 6100 in measuring volumes 150 ml concluded that, compared with existing ultrasound methods, the 3D ultrasound device provides significantly greater accuracy for estimating lower bladder volume when appropriate patients are selected and examiner measurement error is reduced. Implementing portable bladder ultrasound scanning protocols Several authors 3-5,21,41,42 have recommended the development of protocols, in emergency hospitals and radiology departments, for the use of ultrasound scanners in the management of urinary retention. These should include training requirements, documentation, clinical indications for use and quality evaluation (results and outcomes tracking). Sparks et al 23 however posit that good quality research which documents the effectiveness of PBUS on clinical outcomes is required prior to adopting the use of the technology over catheterisation. A rapid response report by NHS National Patient Safety Agency 43 stated that ultrasound devices should be made available in acute and community hospitals, and File Name: 2011-002 PBUS v1.0 Version: 1.0 Page 6 of 7

training should be provided to staff in their use. Health care providers have also been advised to complement knowledge gained from training with clinical judgement when using PBUS to measure PVR volumes and total residual volumes 21. Successful adaptation of bladder scanning ultrasound programmes that provided staff training in facilities have been reported in a number of studies 8,24,44-47. These programmes resulted in easy identification of individuals at risk of developing UTIs 45, reduced unnecessary catheterisation 24,46,47 and reduced UTIs 46. Continuous quality assurance and additional training for staff demonstrating inaccurate measurements were however recommended 44. Safety Generally, there are no complications associated with PBUS. Some reported potential adverse effects include skin irritation, allergic reaction to the ultrasound gel and padding and pressure sore formation at the site where the sensor is placed 2. Cost effectiveness No relevant economic evaluations (based on or generalisable to UK costs) were identified. File Name: 2011-002 PBUS v1.0 Version: 1.0 Page 7 of 7