A more accurate method to estimate patient dose during body CT examinations with tube current modulation

Similar documents
FDG-18 PET/CT - radiation dose and dose-reduction strategy

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

Computed tomography for pulmonary embolism: scan assessment of a one-year cohort and estimated cancer risk associated with diagnostic irradiation.

Multislice CT versus cone beam CT: dosimetric and image quality comparision.

Scientific Exhibit Authors:

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T.

The effect of CT dose reduction on performance of a diagnostic task

The effect of CT dose reduction on performance of a diagnostic task

Cumulative radiation CT dose in young traumatic patients: a single centre 5 years retrospective review

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA

Quantifying Dual-energy computed tomography (DECT) in patients with renal calculi using a Toshiba Aquilion One Scanner.

64-MDCT imaging of the pancreas: Scan protocol optimisation by different scan delay regimes

Lifetime risk of radiation-induced cancer from screening mammography

Radiation induced DNA double strand breaks undergoing PET/CT examinations.

Purpose. Methods and Materials. Results

Uozu city/jp, Minatoku, Tokyo/JP Bones, Extremities, CT, Surgery, Physics, Artifacts, Image verification /ecr2014/C-0462

Artifact in Head CT Images Due to Air Bubbles in X-Ray Tube Oil

Postmortem Computed Tomography Finding of Lungs in Sudden Infant Death.

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction?

Scientific Exhibit Authors: V. Moustakas, E. Karallas, K. Koutsopoulos ; Rodos/GR, 2

AFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006).

Ultra-low dose CT of the acute abdomen: Spectrum of imaging findings

Death due to Hypothermia: Postmortem Forensic Computed Tomography.

Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study

High density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous?

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

An audit of radiation dose of 4D CT in a radiotherapy department

Standardizing mammographic breast compression: Pressure rather than force?

Influence of pulsed fluoroscopy and special radiation risk training on the radiation dose in pneumatic reduction of ileocoecal intussusceptions.

Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality.

Metal Artifact Reduction by Dual Energy CT

Digital chest radiography: collimation and dose reduction

Identification and numbering of lumbar vertebrae using various anatomical landmarks on MRI of lumbosacral spine

Institutional review board approval was obtained prior to the start of this study.

Application of international standards to diagnostic radiology dosimetry

Computed tomography and Modified RECIST criteria for assessment of response in malignant pleural mesothelioma

Slowly growing malignant nodules and rapidly growing benign nodules: Evaluation of the value of volume doubling time

The impact of SID and collimation on backscatter in radiography

Radiological features of Legionella Pneumophila Pneumonia

CT Fluoroscopy-Guided vs Multislice CT Biopsy ModeGuided Lung Biopies:a preliminary experience

Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks

Reliability of the pronator quadratus fat pad sign to predict the severity of distal radius fractures

B. CT protocols for the spine

Computed tomographic dacryocystography as compared with X-ray dacryocystography in patients with dacryostenosis

Diffuse high-attenuation within mediastinal lymph nodes on non-enhanced CT scan: Usefulness in the prediction of benignancy

Monophasic versus biphasic contrast application in CT of patients with head and neck tumour

Pulmonary changes induced by radiotherapy. HRCT findings

Alfred Health's CT quality control program

Imaging Features of Acute Pyelonephritis in Contrast Computed Tomography as Predictors of Need for Intervention

Imaging Features of Acute Pyelonephritis in Contrast Computed Tomography as Predictors of Need for Intervention

Is ascites a sensible predictive sign of peritoneal involvement in patients with ovarian carcinoma?: our experience with FDG-PET/CT

Low-dose computed tomography (CT) protocol in the screening of patients with social exposure to asbestos

Bolus administration of esmolol allows for safe and effective heart rate control during coronary computed tomography angiography

Application of CARE kv and SAFIRE in Contrast-Enhanced CT Examination on Thorax

A Randomized Controlled Study to Compare Image Quality between Fenestrated and Non-Fenestrated Intravenous Catheters for Cardiac MDCT

Emerging Referral Patterns for Whole-Body Diffusion Weighted Imaging (WB-DWI) in an Oncology Center

BI-RADS 3, 4 and 5 lesions on US: Five categories and their diagnostic efficacy and pitfalls in interpretation

Evaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray

Single cold nodule in Graves' disease: benign vs malignant

3D ultrasound applied to abdominal aortic aneurysm: preliminary evaluation of diameter measurement accuracy

How to obtain the waist circumference for retrospective studies - a prospective validation of CT images for the evaluation of the abdominal perimeter

Ethanol ablation of benign thyroid cysts and predominantly cystic thyroid nodules: factors that predict outcome.

Contrast agents, Abdomen, CT, Contrast agent-intravenous, Cancer /ecr2015/C-1760

Medical device adverse incident reporting in interventional radiology

Cognitive target MRI-TRUS fusion biopsies of MRI detected PIRADS 4 and 5 lesions

User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT)

Radiological Investigation of Renal Colic in an Emergency Department of a Teaching Hospital

Direct half value layer measurements in mammography - is near enough good enough?

Quantitative imaging of hepatic cirrhosis on abdominal CT images

Intra-abdominal abscesses radiology diagnostic

The "whirl sign". Diagnostic accuracy for intestinal volvulus.

3D cine PCA enables rapid and comprehensive hemodynamic assessment of the abdominal aorta

Cierny-Mader classification of chronic osteomyelitis: Preoperative evaluation with cross-sectional imaging

The Role of Radionuclide Lymphoscintigraphy in the Diagnosis of Lymphedema of the Extremities

Assessment of renal cell carcinoma by two PET tracer : dual-time-point C-11 methionine and F-18 fluorodeoxyglucose

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation

Diffusion-weighted MRI (DWI) "claw sign" is useful in differentiation of infectious from degenerative Modic I signal changes of the spine

Hyperechoic breast lesions can be malignant.

Trabecular bone analysis with tomosynthesis in diabetic patients: comparison with CT-based finite-element method

Computed Diffusion-Weighted Image in the Abdomen

Essure Permanent Birth Control Device: Radiological followup results at our center

Digital tomosynthesis in diagnosis of occult hip fractures

Seemingly isolated greater trochanter fractures do not exist

Medical Physics and Informatics Original Research

Computational simulation of 4D blood flow dynamics of the thoraco-abdominal aorta: prediction of long-term changes in aneurysm morphology

DEXA Scores and Bone Density Measured on Routine CT Scans

Our experience in the endovascular treatment of female varicocele

Abdominal fat distribution (subcutaneous vs. visceral abdominal fat compartments): correlation with gender, age, BMI and waist circumference

Resuscitation lateral cervical spine X-ray (LSCX): A useful mandatory screening tool in acute trauma?

Purpose. Methods and Materials

A pictorial review of normal anatomical appearences of Pericardial recesses on multislice Computed Tomography.

Sonographic and Mammographic Features of Phyllodes Tumours of the Breast: Correlation with Histological Grade

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Computed tomography for the detection of thumb base osteoarthritis, comparison with digital radiography.

Dual Energy CT: a new tool in evaluation of the urinary tract stones composition in clinical practice - initial study

Ultrasonic evaluation of superior mesenteric vein in cancer of the pancreatic head

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Breast ultrasound appearances after Mammotome vacuumassisted

Transcription:

A more accurate method to estimate patient dose during body CT examinations with tube current modulation Poster No.: C-0738 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Kawaguchi 1, Y. Matsunaga 2, K. Matsubara 3, S. Suzuki 1 ; 1 2 3 Toyoake/JP, Nagoya/JP, Kanazawa/JP Keywords: DOI: Radioprotection / Radiation dose, CT, Dosimetry, Dosimetric comparison 10.1594/ecr2014/C-0738 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. www.myesr.org Page 1 of 13

Aims and objectives BACKGROUND Tube current modulation (TCM) is a useful dose reducing technique, and substantial dose reduction is achievable without image quality deterioration (1-3). Modulation of tube current is based on the size, shape, and attenuation of the patient's scanned body part. TCM algorithms are based on angular modulation, z-axis modulation, or a combination of the two. It is important to accurately estimate the radiation dose that patients receive during CT procedures so that any potential risk can be determined and minimized. The ImPACT (Imaging Performance Assessment of CT Scanners, the scanner evaluation center of the United Kingdom National Health Service) CT patient dosimetry calculator offered by the ImPACT CT scanner evaluation group is a set of commercially available dose simulation tools in which Monte Carlo dose data sets are used to calculate normalized patient dose data from the irradiation of a mathematical phantom (4). This calculator is commonly used for organ doses; however, differences in patient size and variable tube current are not taken into account. Therefore, when CT is performed with TCM, it is possible that actual patient doses may vary from those estimated with the ImPACT calculator. Patient dose using TCM is important to account for the actual body stature and weight of the patient (5, 6). Previous study (7) reported that the best estimates of organ dose required raw projection data from the TCM function, however, organ (breast and lung) doses during chest CT were estimated from available data in the DICOM header of each CT image and the those from the raw projection data of the TCM function differed by less than 5%. PURPOSE We aimed to establish a more accurate and convenient method for estimating organ and effective doses on CT examinations with TCM using actual patient data from the DICOM header of each CT image and the dosimetry calculator. We also compared longitudinal variations in tube current values between iterative reconstruction (IR) and filtered back projection (FBP) reconstruction algorithms. Methods and materials CT Examinations CT scans were obtained using two helical scanners (Aquilion 64 and RXL, Toshiba Medical Systems, Tochigi, Japan) which have 64 and 16 channels, respectively. The Page 2 of 13

reconstruction algorithm of the Aquilion 64 was FBP and the reconstruction algorithm of the RXL was IR (AIDR 3D). Both CT examinations were performed using x, y, and z-axis TCM (Volume EC; Toshiba Medical Systems) in a clinical setting. For this retrospective study, we collected consecutive patient data from August to November 2012. In total, 100 consecutive patients were identified who had undergone plane body CT in CT scanners (Aquilion 64 and RXL). Fifty (25 women) of these patients were scanned using the Aquilion 64 with Volume EC (SD, 6; tube rotation speed, 0.5 seconds; maximal X-ray tube capacity, 550 ma; beam collimation, 32 mm; slice thickness, 10 mm; slice interval, 10 mm; beam pitch, 0.828; tube voltage, 120 kv) and 50 other patients (25 women) were scanned on the Aquilion RXL with Volume EC (SD, 6; tube rotation speed, 0.5 seconds; maximal X-ray tube capacity, 500 ma; beam collimation, 16 mm; slice thickness, 10 mm; slice interval, 10 mm; beam pitch, 0.938; tube voltage, 120 kv; reconstruction algorithm, AIDR 3D Mild). Additionally, a scout view was first obtained at 120 kv and 30 ma. The patients were examined in supine position. For each CT scan, the tube currents were acquired from DICOM tube current information on each image from the thyroid (the starting anatomical landmark) to the symphysis pubis (the ending anatomical landmark). Dose simulation The ImPACT CT patient dosimetry calculator (version 1.0.4) is a computer application in which Monte Carlo dose data sets are used to calculate normalized patient organ dose data from the irradiation of a mathematical phantom. The mathematical phantom has a fixed z-axis length from the head to pelvis. An Excel (Microsoft, Washington, USA) spreadsheet was utilized as a convenient user interface for estimating organ doses and effective dose using 23 NRPB Monte Carlo-generated scanner data sets. The ImPACT CT scanner evaluation group developed a method to match results from the original 23 scanner data sets to new CT scanners through the use of so-called "ImPACT factors;" these factors are based on CTDI air, CTDI center,and CTDI peripheral normalized to 100 mas measured in air and either a standard head or standard body CTDI dose phantom (8). The ImPACT spreadsheet can be used to determine organ doses and effective doses for a wide range of relevant examination parameters (9, 10) : scanner type, tube voltage, tube current, rotation time, head or body scan, and either scan width and increments for non-helical scans or beam pitch for helical scans. However, our CT scanners performed scanner matching that was selected from the 23 NRPB Monte Carlo-generated scanner data sets, because our CT scanners were not listed in the ImPACT scanner list. The CT scanner measurements CTDI air, CTDI center,and CTDI peripheral were normalized to 100 mas using a standard 100-mm pencil dosimeter (Model 9015, 10X5-3CT, Radcal Corp., Monrovia, CA, USA) and entered into the ImPACT spreadsheet along with the acquired scanner data values (Table 1). These simulations used the patient long axis z, with the value of z ranging from 0 to 75 cm, considering the anatomical organ regions (Figure 1). The conventional method Page 3 of 13

The conventional method is a fixed tube current simulation using a constant tube current value that is the average tube current over the entire scan length from the thyroid to symphysis pubis. This average is reported by the tube current data extracted from the DICOM header (the slice thickness and slice interval were 10 mm). The ImPACT spreadsheet was populated with the average tube current and calculated the organ doses and effective doses. Longitudinal variations method The longitudinal variations method used the tube current conversion factors to calculate organ doses, which are the tube current ratios over the entire scan length and each organ region. The longitudinal TCM function was obtained from each patient's actual CT images via the DICOM header. The DICOM header of each image contains a unique tube current value along with a corresponding table location. The tube current reported in the DICOM header is the average value of the tube current over the rotation used to reconstruct that image. Tube current and table location values for each image were read from the DICOM header of the image data. The average tube currents for 10 organs, including the thyroid, esophagus, lung, breast, liver, stomach, colon, bladder, gonads (the testes in men and the uterus in women instead of the ovaries, because the ovaries are often unclear on plane CT), and kidney were calculated from the DICOM images of each organ region from start to finish. The tube current conversion factors (F TCM ) were defined as Fig. 2: Equation is the tube current conversion factor. References: - Toyoake/JP The following formula was used to estimate organ doses taking TCM into account: Fig. 3: Equation is the organ dose taking TCM into account. References: - Toyoake/JP where D TCM is the organ dose taking TCM into account, and D T,R is the organ dose from the conventional method. The effective dose considering TCM (E TCM ) was calculated with the following equation: Page 4 of 13

Fig. 4: Equation is the effective dose considering TCM. References: - Toyoake/JP where w T is the tissue-weighting factor in ICRP publication 103 (11), H T is the equivalent dose, and W R is the radiation-weighting factor. F TCM, which represents the bone marrow, skin, bone surface, brain, and salivary glands, was assumed to be 1.0 because the tube current could not read at each region. The F TCM represents the remainder substituted kidney region because the kidney volume was large and the highest organ dose was in the remainder. Differences of F TCM s between Aquilion 64 (FBP) and RXL (IR) were statistically evaluated using student's t-test. And differences of effective dose between the conventional method and longitudinal variations method were statistically evaluated using paired t-test. Images for this section: Page 5 of 13

Page 6 of 13

Fig. 1: The scan ranged from 0 to 75 cm for body CT, as shown in the mathematical phantom in the ImPACT CT patient dosimetry calculator (version 1.0.4). Table 1: Scanning parameters for body CT Page 7 of 13

Results The mean age of the 100 patients was 63 years old (the age range was 28-88 years old). The average CTDI vol over the entire scan length limited to the region from the thyroid to the symphysis pubis is shown in Table 2. The range of average CTDI vol values over the entire scan length using the Aquilion 64 was 10.0-37.6 mgy (mean, 26.3 mgy) and the those of the Aquilion RXL were 5.6-18.3 mgy (mean, 10.5 mgy). Table 3 shows the results of the mean F TCM for 10 organs. The mean F AEC values for the breast using the Aquilion 64 (0.76) and RXL (0.73) demonstrated especially low F TCM values among the 10 organs. In contrast, the mean F TCM for the bladder using the Aquilion 64 (1.21) and RXL (1.20) demonstrated the highest F TCM values. These results indicated that the organ doses for bladder and breast estimated by the longitudinal variations method were 20% larger and 25% smaller than those estimated by the conventional method, respectively. Although there were no significant differences except for the esophagus, lung, and remainder, a significant difference was observed between the Aquilion 64 and RXL in the F TCM of all organs for the remainder (student's t-test, p = 0.046). The CVs for the thyroid and gonads were large using the Aquilion 64 (22% and 54%, respectively) and RXL (16% and 68%, respectively). Table 4 shows the results of the mean organ doses estimated by the longitudinal variations method. The mean organ doses for the testes using the Aquilion 64 and RXL were 4 and 1 mgy, respectively. These doses were especially low among the 10 organs. In contrast, the mean organ dose for the thyroid using the Aquilion 64 (50 mgy) and the RXL (21 mgy) were the highest observed doses. The range of organ dose ratios (RXL/64) was 0.28-0.45. The ratio calculated by the mean CTDI vol (RXL/64) was 0.399. The CVs for the testes were large using both the Aquilion 64 (91%) and RXL (130%). The CVs for the ovaries were low using the Aquilion 64 (18%) and RXL (30%). The range of effective doses using the Aquilion 64 with the longitudinal variations method was 11-45 msv (mean, 30 msv), while the Aquilion RXL values were 7-21 msv (mean, 13 msv) (Figure 2). The ratio calculated from the mean effective doses (RXL/64) was 0.42. Figure 2 shows the effective dose correlations between the conventional method and the longitudinal variations method on the two CT scanners with fitted regression lines. The correlation coefficients (R 2 ) were 0.986 and 0.987 for the Aquilion 64 and RXL, respectively. Although the differences of effective doses were relatively small (mean, 0.7 msv), there were significantly differences between them (paired t-test, p < 0.05). The difference in effective doses between the two methods was smaller than those of the organ doses. Mean F TCM values on the Aquilion 64 and RXL were not significantly different for the 10 organs except for the esophagus, lung, and remainder (Table 4). Moreover, in the ratios (RXL/64) of the organ doses, except for the testes, the mean CTDI vol and the mean effective doses were both approximately 0.4. This means Page 8 of 13

that the longitudinal variations in the tube current ratio over the entire scan length and in each organ region were almost the same for both the IR and FBP algorithms developed by the same manufacturer. Images for this section: Table 2: Average CTDIvol over the entire body CT scan Table 3: Estimated mean conversion factor, FTCM, from the tube current in the DICOM header Page 9 of 13

Table 4: Estimated mean organ doses from the longitudinal variations method Page 10 of 13

Fig. 5: The graph shows the relation between effective doses calculated by the conventional method and longitudinal variations method with a fitted regression line. Page 11 of 13

Conclusion The organ doses for bladder and breast on body CT with TCM estimated by the tube current ratios over the entire scan length and each organ regions (the longitudinal variations method) were 20% larger and 25% smaller than those estimated by the average tube current over the entire scan length (the conventional method), respectively. The difference of effective doses between the conventional method and longitudinal variations method was smaller than those of organ doses. The longitudinal variations in the tube current ratio over the entire scan length and in each organ region were almost the same with the IR and FBP reconstruction algorithms developed by the same manufacturer. Personal information References 1. Irie, T. and Inoue, H. Individual modulation of the tube current-seconds to achieve similar levels of image noise in contrast-enhanced abdominal CT. AJR Am J Roentgenol. 184, 1514-8 (2005). 2. Kalra, M. K., Maher, M. M., Toth, T. L., Schmidt, B., Westerman, B. L., Morgan, H. T. and Saini, S. Techniques and applications of automatic tube current modulation for CT. Radiology. 233, 649-57 (2004). 3. Kalra, M. K., Maher, M. M., Toth, T. L., Kamath, R. S., Halpern, E. F. and Saini, S. Comparison of Z-axis automatic tube current modulation technique with fixed tube current CT scanning of abdomen and pelvis. Radiology. 232, 347-53 (2004). 4. "ImPACT CTDosimetry. Imaging performance assessment of CT Scanners: a medical devices agency evaluation group. CT scanner matching data, tables of CTDI values in air, CTDIw, and phantom factor values," (ImPACT Internet home page (http:// www.impactscan.org); status: October 2011). 5. Schlattl, H., Zankl, M., Becker, J. and Hoeschen, C. Dose conversion coefficients for CT examinations of adults with automatic tube current modulation. Phys. Med. Biol. 55, 6243-61 (2010). 6. Israel, G. M., Cicchiello, L., Brink, J. and Huda, W. Patient size and radiation exposure in thoracic, pelvic, and abdominal CT examinations performed with automatic exposure control. AJR Am J Roentgenol. 195, 1342-6 (2010). Page 12 of 13

7. Khatonabadi, M., Zhang, D., Mathieu, K., Kim, H. J., Lu, P., Cody, D., Demarco, J. J., Cagnon, C. H. and McNitt-Gray, M. F. A comparison of methods to estimate organ doses in CT when utilizing approximations to the tube current modulation function. Med Phys. 39, 5212-28 (2012). 8. Jessen KA, Panzer W, Shrimpton PC, et al. European Guidelines on quality criteria for computed tomography. Brussels, Belgium: European Commission, 2000:EUR 16262. 9. American Association of Physicists in Medicine (AAPM). The measurement, reporting and management of radiation dose in CT: report of AAPM Task Group 23 of the Diagnostic Imaging Council CT Committee. College Park, MD: AAPM, 2007: report no. 96 10. McNitt-Gray, M. F. AAPM/RSNA Physics Tutorial for Residents: Topics in CT. Radiation dose in CT. Radiographics. 22, 1541-53 (2002). 11. ICRP Publication 103. International Commission on Radiological Protection. The 2007 Recommendations of the International Commission on Radiological Protection. Ann. ICRP. 2007. Page 13 of 13