Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Similar documents
An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan

Assessment of effective dose in paediatric CT examinations

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Accounting for Imaging Dose

Radiation Dosimetry for CT Protocols

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

Skyscan 1076 in vivo scanning: X-ray dosimetry

Comparison of organ doses estimations in radiology with PCXMC application based on MIRD phantoms and CALDose-X application based on voxel phantoms

Computed tomography Acceptance testing and dose measurements

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Medical Physics and Informatics Original Research

CT Dosimetry in the Clinical Environment: Methods and Analysis

Estimating Patient Radiation Dose from Computed Tomography

Dosimetric Consideration in Diagnostic Radiology

Austin Radiological Association Nuclear Medicine Procedure PET SODIUM FLUORIDE BONE SCAN (F-18 NaF)

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

CT Radiation Dosimetry Study Using Monte Carlo Simulation and. Computational Anthropomorphic Phantoms

Machine Learning Powered Automatic Organ Classification for Patient Specific Organ Dose Estimation

Radiation Dose Reduction Strategies in Coronary CT Angiography

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

Why is CT Dose of Interest?

Assessment of Patient Dose from CT Examinations in Khorasan, Iran

Calculation of Normalised Organ and Effective Doses to Adult Reference Computational Phantoms from Contemporary Computed Tomography Scanners

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

CT Radiation Risks and Dose Reduction

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

Austin Radiological Association Ga-68 NETSPOT (Ga-68 dotatate)

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

FDG-18 PET/CT - radiation dose and dose-reduction strategy

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

Survey of patients CT radiation dose in Jiangsu Province

X-Ray & CT Physics / Clinical CT

GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT

A Real-Time Monte Carlo System for Internal Dose Evaluation Using an Anthropomorphic Phantom with Different Shapes of Tumors Inserted

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

State of the art and future development for standardized estimation of organ doses in CT

Impact of ICRP-89 Based Models on Dose Estimates for Radiopharmaceuticals and CT Exams. Stabin MG, Kost SD, Clark JH, Pickens DR, Price RR, Carver DE

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial

National Cancer Institute

Quality Control and Patient Dosimetry on line for Computed Tomography

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations

Patient doses from X-ray computed tomography examinations by a single-array detector unit: Axial versus spiral mode

Radiation exposure of the Yazd population from medical conventional X-ray examinations

FAX, A FEMALE ADULT VOXEL PHANTOM FOR RADIATION PROTECTION DOSIMETRY

Multislice CT versus cone beam CT: dosimetric and image quality comparision.

Adopting DAP as a dose metric in CT

Introduction and Background

ABSTRACT. Objectives: To investigate the radiation dose received by patients undergoing routine plain x-ray

Tracking Doses in the Pediatric Population

Dosimetry of recently introduced CBCT Units for Oral and Maxillofacial Radiology

Chief Radiographer TEI Clinical Associate 2016

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

Doses from CT examinations to children suffering from hydrocephalus

Bone Densitometry Radiation dose: what you need to know

Patient / Organ Dose in CT

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology

A multicentric study on patient dose in multislice CT

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates

A Standardised Approach to Optimisation

The impact of SID and collimation on backscatter in radiography

ORGAN DOSES FOR PATIENTS UNDERGOING COMPUTED TOMOGRAPHY EXAMINATIONS: VALIDATION OF MONTE CARLO CALCULATIONS USING ANTHROPOMORPHIC PHANTOMS

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers

Patient Radiation Doses from Adult and Pediatric CT

Assessment of radiation risk to pediatric patients undergoing conventional X-ray examinations

Combined Anatomical and Functional Imaging with Revolution * CT

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

National collection of local diagnostic reference levels in Norway and their role in optimization of X-ray examinations

Determination of the conversion factor for the estimation of effective dose in lungs, urography and cardiac procedures

EFFECTIVENESS OF PROTECTIVE PATIENT EQUIPMENT FOR CT: AN ANTHROPOMORPHIC PHANTOM STUDY Barbara Ott 1, *, Anja Stüssi 2 and Roberto Mini 1

Fetal Dose Calculations and Impact on Patient Care

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

International Association of Scientific Innovation and Research (IASIR) (An Association Unifying the Sciences, Engineering, and Applied Research)

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography

Low-dose CT Lung Cancer Screening Guidelines for Pulmonary Nodules Management Version 2

EFFECTIVE DOSE ESTIMATION AND CANCER RISK ASSESSMENT DUE TO SOME SELECTED CT EXAMINATIONS AT SWEDEN GHANA MEDICAL CENTRE (SGMC)

Whole-body biodistribution and radiation dosimetry estimates for the β-amyloid radioligand [ 11 C]MeS-IMPY in non-human primates

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Impact of ICRP-89 Based Models on Dose Estimates for Radiopharmaceuticals and CT Exams

Imaging doses from the Elekta Synergy X-ray cone beam CT system

Amira K. Brown, Ph.D. Molecular Imaging Branch, NIMH Bldg. 1 Rm. B3-10

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Equivalent dose to organs and tissues in Hysterosalpingography calculated with the FAX (Female Adult voxel) phantom

Managing Patient Dose in Computed Tomography (CT)

Application of 3D Printing to Molecular Radiotherapy Phantoms. Nick Calvert Nuclear Medicine Group The Christie NHS Foundation Trust, Manchester

Estimation of Effective Doses for Radiation Cancer Risks on ISS, Lunar and Mars Missions with Space Radiation Measurements

Radiation Dose Assessment and Risk Estimation During Extracorporeal Shock Wave Lithotripsy

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Transcription:

Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P. Ghafarian 4 1 Department of Physics, Amir Kabir University & TPP Co., Tehran, Iran 2 Department of Nuclear Engineering, Shahid Beheshti University, Tehran, Iran 3 Department of Medical Physics & RCSTIM, TUMS, Tehran, Iran 4 Department of Medical Physics, IUMS, Tehran, Iran ABSTRACT Background: CT is a diagnostic imaging modality giving higher patient dose in comparison with other radiological procedures, so the calculation of organ dose in CT exams is very important. While methods to calculate the effective dose have been established (ICRP 26 and ICRP 60), they depend heavily on the ability to estimate the dose to radiosensitive organs from the CT procedure. However, determining the radiation dose to these organs is problematic, direct measurement is not possible and comparing the dose as functions of scan protocol such as ma is very difficult. One of the most powerful tools for measuring the organ dose is Monte Carlo simulation. Materials and Methods: Today the predominant method for assessment of organ absorbed dose is the application of conversion coefficients established by the use of Monte Carlo simulations. One of the most famous dose calculation software is CTDOSE, which we have used it for calculation of organ dose. In this work we measured the relationship between the ma, KV and scanner type with the equivalent organ dose and effective dose in mathematically standard phantom (Hermaphrodite 170cm/70Kg) in an abdomen-pelvis CT exam by Monte Carlo method. For this measurement we increased the ma in steps of 10 ma and plot curves for organ dose as a function of ma for different KV setting. Results: As expected, with increasing ma, patient organ dose increased, but the simulation results showed that the slope of organ dose as a function of ma increased with KV increasing. By increasing KV from 120 to 140 the increase in slope of curves representing patient organ dose versus ma for different scanner types show almost similar behavior whereas the slope of the corresponding curves in scanners which equipped xenon detectors was almost 22% more than the slope of scanners equipped with scintillation detectors. Conclusion: Our research showed that regarding equivalent dose the system incorporating scintillation detector has a superior performance. Incorporating such software in various CT scanners, marketed by different vendors, will offer the ability to get a print out of patient organ dose in any examination according to the imaging parameters used for imaging any part of the body. Iran. J. Radiat. Res., 2004; 1(4): 187-194 CT Dosimetry INTRODUCTION I Corresponding author: Dr. M.R. Ay, Department of Physics, Amir Kabir University & TPP Co. E-mail: Farshid.ay@tppgems.com n X-ray diagnostics, only parts of the human body are exposed during an examination, and due to the relatively low energy of photons, the energy deposition is very inhomogeneous. Furthermore, different kinds of tissues and organs have different sensitivities for radiation. In order to take these circumstances into account, the effective dose, E, was given by ICRP (ICRP 1996): (1)

M.R. Ay et al. Where w T is the tissue weighting factor, and H T is the equivalent dose to organ or tissue T. The effective dose is the bottom line for statements about patient doses and the associate risks also for computed tomography (CT). However, the ways to find necessary knowledge about organ doses from CT examinations are not straight forward. A number of measurement methods have been reported that use a variety of ways to describe or characterize the radiation delivery by CT, and these are quite different from methods and procedures normally employed in conventional X-ray diagnostics. A fundamental dosimetric quantity is the computed tomography dose index, CTDI, defined by: (2) Where T is the thickness of the tomographic slice, and D(z) is the distribution of absorbed dose along a line parallel to the axis of rotation, designated the z axis. The CTDI may be assessed free in air or in phantoms, and the measurement may be done with TLDs or ionization chambers (Olerud 1999). The Food Drug Administration (FDA) in the US recommends doing the measurements in the center and periphery of cylindrical PMMA phantoms of 16 cm and 32 cm diameter, respectively. Because of the scattered radiation in the phantom, the total integration length must be defined. According to FDA, the dose is to be integrated over 14 slices thickness, which implies that the total integration length depends on the slice thickness. This approach was adopted by the IEC in 1994, but is not very practical, so the predominant method is now to apply a fixed integration length of 100 mm for all measurements (IEC 1998). There is, however, still some confusions concerning the definition and interpretation of various quantities found for single and multi slice CT scanner. This is basically caused by the definition of the nominal tomographic section thickness, T, and the number of tomographic section, N. As suggested by Cynthia and Zink (Cynthia et al. 1999), there is a need for the definition of a new quantity; the total nominal tomographic sections thickness, T T, representing the total nominal active detector width. The CTDI 100 would then be defined as: (3) where D(z) is the dose profile along a line z perpendicular to the tomographic plane and T T is the total nominal tomographic section thickness, which in turn is defined as the width of the sensitivity profile taken at the center of a tomographic section. For multi slice equipments, this width is the sum of the sensitivity profile of all active detectors. The weighted and normalized n CTDI w was defined as: (4) Where CTDI 100 is measured in the center and periphery (1 cm under the surface) of a 16 cm (head) and 32 cm (trunk) phantom, respectively; division normalizes the weighted quantity with the current time product per slice, C (mas). The actual CTDI w is obtained by multiplying with the mas value used in the clinic. Correspondingly, the dose length product DLP, in units of mgy.cm, for a complete CT examination was defined for conventional axial CT and for helical CT, respectively, as: (5) Where T is the slice thickness, N the number of slice, A the tube current, and t the total acquisition time. Relationship between Different Dose values CTDI is a parameter that gives information about the dose due to a particular set of acquisition parameters. It is useful to allow dose comparisons between protocols for a given CT 188 Iran. J. Radiat. Res.; Vol. 1, No. 4, March 2004

Measurement of organ dose by Monte Carlo system, and can be used to calculate the dose length product. However, it should be used with caution when comparing patient doses from two different scanners. DLP can be used as an indicator of the radiation dose delivered throughout the examination. The Effective Dose can be derived from the DLP and the normalized effective dose factors for each of the organs irradiated during the examination. This parameter is the most important one in making risk benefit-decisions, because it gives an estimation of patient dose that can be related to biological risk (GE 2001). The relationship between these parameters is shown in figures 1 and 2. Figure 1. Relationship between different dose values. Figure 2. Relationship between dose and scan area. Factors influencing patient dose in CT It is often the case that an improvement in image quality is obtained at the expense of a higher radiation dose to the patient. Conversely, when reducing the dose, the quantity of X-rays reaching to the detector decreases as well. The parameters, which affected in patient dose divided in two different categories; Equipmentrelated factors and Application-related factors. The Equipment-related factors are: waveform of the generator, range of tube current settings, beam filtration, beam shaper, focus-axis distance, slice collimation, detector array, scan geometry, partial fan beam scanners and scan angle. The Application-related factors are: exposure time, object diameter, slice thickness, pitch factor, number of slices, reconstruction filter, window width, and matrix size and filed of view (Nagel 2000). MATERIALS AND METHODS Today the predominant method for assessment of organ absorbed dose is the application of conversion coefficients established by the use of Monte Carlo simulations. This technique is used in most fields of medical radiation physics. The Monte Carlo method, in this context, is a computational model in which physical quantities are calculated by simulating the transport of X-ray photons. Early attempts to model the shape of a human being and its internal organs in order to calculate absorbed radiation doses were made by Snyder et al. (1996) and Koblinger (1972). The earliest version of Monte Carlo software for calculations of absorbed dose and effective dose in CT is CTDOSE, which has been written by Le Heron JC, from National Radiation Laboratory, Christchurch, New Zealand (Heron 1993). Iran. J. Radiat. Res.; Vol. 1, No. 4, March 2004 189

M.R. Ay et al. CTDOSE Software In this work, we used commercially available Monte Carlo software CTDOSE (Heron 1993) for calculation of organ dose. The CT-Dose calculation program uses a Monte Carlo simulation routine to estimate dose distribution and consequently the effective dose in a mathematically standard Hermaphrodite phantom (170cm/70Kg) from a given CT-procedure and a given CT-scanner type, as well as the dose-length-product from the Measurement Conditions One of the most popular CT exam is abdomen-pelvis scan, the routine protocol for this exam is: slice thickness 10 mm, table feed per slice 10 mm, scan start position 13.5 cm and scan end position 52.9 cm (figure 3). For measurement of organ absorbed dose in this examination as a function of ma, we increased the ma in steps of 10 and measured the absorbed organ and effective dose for specific kvp and scanner type; on the other hand for measurement of effect of kvp and scanner type (Geometry and detection system) in organ absorbed dose we repeated increasing the ma and measured organ dose with different kvp and scanner type. Figure 3. Hermaphrodite phantom CT-procedure. The CTDOSE used normalized organ dose data sets, together with measured values of freein-air axial dose for particular models of scanner, and details of the clinical technique for each examination type (Jones et al. 1991, 1993). This software required the following input parameters: scanned volume (in terms of baseline in the phantom and number of slices), slice width, couch increment, effective mas, kvp and CT dose index per mas (CTDI). RESULTS ma The absorbed dose is directly proportional to the tube current (ma). The main point in this study is calculation of the slope of increasing dose via ma; this slope could be helpful to estimate the absorb dose caused by increasing ma. Using CTDOSE Monte Carlo package we acquired the equivalent dose in different organs and effective dose in an abdomen-pelvis exam as a function of ma (30 ma up to 200 ma in steps of 10 ma) at 120 kvp in GE CTi Scanner (figures 4 and 5). The slope of increasing equivalent dose for some organs is calculated in table 1. 190 Iran. J. Radiat. Res.; Vol. 1, No. 4, March 2004

Measurement of organ dose by Monte Carlo kvp Dose increased with an increase in kvp. To obtain the relation between equivalent dose and kvp, CTDOSE Monte Carlo package was used. In figures 6 and 7, the equivalent dose and effective dose was calculated as a function of kvp in 130 ma tube current in GE CTi scanner for an abdomen-pelvis exam. mgy mgy 30 25 20 15 10 5 0 Liver Equivalent Dose in 120 KVp Pelvis Spine Testes Ovaries Red Blood Marrow Stomac Wall Uteres 0 50 100 150 200 250 ma Figure 4. Equivalent dose as a function of ma. 30 Liver Equivalent Dose in CTi scanner @130 ma Pelvis 25 Spine Ovaries Red Bone Marrow 20 Stomac Wall Uteries 15 10 msv msv 10 Effective Dose in 120 KVp 9 8 7 6 5 4 3 2 1 0 0 50 100 150 200 250 ma Figure 5. Effective dose as a function of ma. 10 Effective Dose in CTi scanner @ 130 ma 8 6 4 5 2 0 90 100 110 120 130 140 150 KVp Figure 6. Equivalent dose as a function of kvp. 0 90 100 110 120 130 140 150 KVp Figure 7. Effective dose as a function of kvp Iran. J. Radiat. Res.; Vol. 1, No. 4, March 2004 191

M.R. Ay et al. Table 1. Slope of increasing equivalent dose in terms of ma and kvp. Organ Slope of Increasing Equivalent Dose mgy/kvp @ 130 ma in GE CTi Scanner Slope of Increasing Equivalent Dose mgy/ma @ 120 kvp in GE CTi Scanner Lungs 0.18250 0.06000 Stomach Wall 0.29250 0.09470 Urinary Bladder Wall 0.03850 0.01200 Breasts 0.16250 0.06058 Liver 0.27500 0.09000 Esophagus 0.09500 0.03152 Thyroid 0.00200 0.00032 Skin 0.06750 0.02500 Bone Surface 0.13250 0.05470 Red bone marrow 0.11750 0.03882 Testes (Gonads) 0.00340 0.00090 Ovaries (Gonads) 0.16250 0.06000 LLI Wall (Colon) 0.10750 0.03411 Muscle 0.08750 0.03047 Adrenals 0.27000 0.07176 Brain 0.00001 0.00003 Small Intestine 0.26250 0.08470 ULI Wall 0.26250 0.08470 Kidneys 0.30000 0.10529 Pancreas 0.27750 0.09000 Spleen 0.26250 0.09058 Thymus 0.06000 0.01700 Uterus 0.15000 0.05000 Pelvis 0.35000 0.13529 Spine 0.29250 0.11529 Skull Cranium 0.00052 0.00004 Skull Facial 0.00232 0.00046 Rib Cage 0.37500 0.16000 Clavicles 0.01475 0.00388 Eye Lenses 0.00052 0.00001 Gall Bladder Wall 0.28750 0.10058 Heart 0.25500 0.09058 192 Iran. J. Radiat. Res.; Vol. 1, No. 4, March 2004

Measurement of organ dose by Monte Carlo Scanner Type One of the most important techniques in CT dose reduction is increasing the detection efficiency. Nowadays the scintillation detectors are specifically developed for CT application. The major advantages are excellent properties such as high absorption efficiency (99%) and stability. The increase of detection efficiency allows better performance in low signal conditions such as those encountered using low dose protocols in highly attenuating patient regions. To obtain the effect of detection efficiency in dose reduction we compared two scanner Figure 8. Equivalent dose in different Scanners systems with different detection systems, which had fairly the same geometry (GE Prospeed with Xenon detector and GE CTi with scintillator detector). We calculated the equivalent organ and effective doses with the same techniques for both scanners. According to our Monte Carlo simulation the effective dose for abdomen-pelvis exam in system which was equipped with scintillator detector in the same technique, (120 kvp and 130 ma) was 15% less than the Xenon detector (7.3 msv in Prospeed and 6.2 msv in Cti) system. Besides the equivalent organ dose in CTi scanner was less than Prospeed scanner (figure 8). DISCUSSION In this research we simulated the variation of equivalent and effective doses for an abdomenpelvis exam as a function of kvp (from 100 to 140) and ma (from 30 to 200) for CTi scanner at 130 ma and 120 kvp respectively. The equivalent dose with the above mentioned parameters were simulated for different organs (figures 4 and 6). In both figures we noticed that the maximum equivalent dose and sharpest slope variation were for pelvis whereas the minimum equivalent dose with lowest slope variation was related to red bone marrow. The equivalent dose for stomach wall and liver is almost equal to each other in both figures 4 and 6. The effective dose for the above mentioned scanner at 130 ma increased from 3.7 msv at 100 kvp to 9.2 msv at 140 kvp and when kvp was kept constant at 120 kvp the effective dose increased from 1.4 msv at 30 ma to 9.5 msv at 200 ma (Jones 1991). Finally, the equivalent dose for red bone marrow at 120 kvp as a function of ma using two different types of scanners was simulated namely: one with Xenon detector and the other with scintillation detector (figure 8). Regarding equivalent dose the system incorporating Iran. J. Radiat. Res.; Vol. 1, No. 4, March 2004 193

M.R. Ay et al. scintillation detector had a superior performance (GE 2002). REFERENCES Cynthia H.C. and Zink F.E. (1999). Performance evaluation of a multi-slice CT scanner. Med. Phys., 28: 2223-2230. EUR report 16262 (1999). European guidelines on quality criteria for computed tomography. Office of Official Publications of the European Communications, Brussels. GE Medical Systems (2001). Dose in computed tomography, basic, challenges and solutions, Technical Report, 71113-BE, France. GE Medical Systems (2002). Solid state CT detector material, Technical Report, 97-5175, Milwaukee. Heron J.C. (1993). CTDOSE Monte Carlo Package, Christchurch, New Zealand: National Radiation Laboratory. ICRP (1996). International Commission on Radiological Protection and Safety in Medicine, Publication 73, Pergamon Press, Oxford. IEC (1998). Particular requirements for the safety of x-ray equipment for computed tomography. IEC 60601-2-44 Ed.2/CDV. Jones D.G. and Shrimpton P.C. (1993). Normalized organ doses for X-ray CT calculated using Monte Carlo techniques. NRPB Software Report 250, Chilton, NRPB. Jones D.G. and Shrimpton P.C. (1991). Survey of CT practice in the UK. Part 3: Normalized organ doses calculated using Monte Carlo techniques. Chilton, NRPB-R250, London. Nagel H.D. (2000). Radiation Exposure in Computed Tomography. COCIR, 2031-2042 Olerud H.M. (1999). CT- Dose Survey Report, Section of Dosimetry and Medical Applications. Norwegian Radiation Protection Authority, P.O. 55, No-1363. 194 Iran. J. Radiat. Res.; Vol. 1, No. 4, March 2004