Relationship between glucose meter error and glycemic control efficacy

Similar documents
Using Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues

Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES

It s Just a Waived Glucose, Isn t It?

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

GLYCEMIC CONTROL SURVEY

Disclosures. Glycemic Control in the Intensive Care Unit. Objectives. Hyperglycemia. Hyperglycemia. History. No disclosures

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.

ACCORDING to the Centers for Disease Control and

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A comparative study between capillary and venous blood glucose levels of type 2 diabetes mellitus patients in intensive care units

Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter

Today s Goals 10/6/2017. New Frontiers in Diabetes Technology. Disclosures

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

Point-of-Care Blood Glucose Meter Accuracy in the Hospital Setting

Point of Care Testing (POCT)

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

In - Hospital Diabetes Care. A review and personal experience

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS

EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System

SAMPLE. Glucose Monitoring in Settings Without Laboratory Support

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

Accuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:

What Should Be the Therapeutic Glycemic Target in Intensive Care Units?

BLOOD GLUCOSE MONITORING: CLINICAL NEED AND TECHNOLOGY

A Children s Bedtime Story

Quick Reference Guide

Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies

Aina Blood Monitoring System

Quick Reference Guide

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes. pump. pump

Method Precision and Frequent Causes of Errors Observed in Point-of-Care Glucose Testing. A Proficiency Testing Program Perspective

INTENSIVE INSULIN THERAPY: A Long History of Conflicting Data.

Performance of three portable blood glucose monitoring devices used in a veterinary application

Control of Blood Glucose in the ICU: Reconciling the Conflicting Data

Quick Reference Guide

Quality Control of Self-Monitoring of Blood Glucose: Why and How?

Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond)

Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the Effects of Bias, Precision, and Rate of Glucose Change

PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS)

Advances in Diabetes Care Technologies

Analytical Performance in the The Test Evaluation Cycle

Early Goal-Directed Therapy

Interference and Point-of-Care Testing Devices


Controversies in Hospital Medicine: Critical Care. Vasopressors, Steroids, and Insulin Therapy

Parenteral Nutrition The Sweet and Sour Truth. From: Division of Endocrinology, Diabetes and Bone Disease Icahn School of Medicine at Mount Sinai

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System

Implementing Glucose Control in 2009 and Beyond: Changes in Patterns and Perceptions

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes Individuals: With type 1 diabetes

Diabetes Management: Current High Tech Innovations

Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim

123 Are You Providing Evidence-Based Diabetes Care? - Martin

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.

Accuracy of Point-of-Care Blood Glucose Measurements in Critically Ill Patients in Shock

April Dear (Editor):

MQ Comparison of glucometers using whole blood

NICO /14 Nipro Diagnostics, Inc.

MQ Comparison of glucometers using whole blood

The performance of flash glucose monitoring in critically ill patients with diabetes

Ultra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

The performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.

VANDERBILT UNIVERSITY MEDICAL CENTER MULTIDISCIPLINARY SURGICAL CRITICAL CARE TRAUMA INTENSIVE CARE UNIT GLYCEMIC CONTROL PROTOCOL

Effects of Measurement Frequency on Analytical Quality Required for Glucose Measurements in Intensive Care Units: Assessments by Simulation Models

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013

Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration

METHOD VALIDATION: WHY, HOW AND WHEN?

Objectives. Glucose Monitoring Today. Tracking Glucose ISO Standards. Continuous Glucose Monitoring & Diabetes Management 3/7/2015

Update in Critical Care Medicine

FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols

AACE Module on Patient Safety in Inpatient Diabetes Care

Glucose Management in the ICU: The Role of the Pharmacist

Significant Insulin Dose Errors May Occur if Blood Glucose Results Are Obtained from Miscoded Meters

Subject Index. Breastfeeding, self-monitoring of blood glucose 56

GLYCEMIC CONTROL IN NEUROCRITICAL CARE PATIENTS

Journal Club ICU

How Should Glucose Meters. Be Evaluated. For Critical Care

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System

Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase

Point of Care Testing Everything old is new again

Conflicts of Interest

The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary Blood Glucose Values

MQ Comparison of glucometers using heparin whole blood

Intensive Insulin in the Intensive Care Unit

Phlebotomy Top Gun. Drawing Blood From a Patient Receiving Intravenous Fluids. HOT TOPIC / Phlebotomy Top Gun, Intravenous Fluids HOT TOPIC / 2018

Clinical Chemistry and Clinical Toxicology Devices Panel of the Medical Devices Advisory

Performance of the Medtronic Sentrino continuous glucose management (CGM) system in the cardiac intensive care unit

The Ever Changing World of Sepsis Management. Laura Evans MD MSc Medical Director of Critical Care Bellevue Hospital

Peripartum and Postpartum Management of Diabetes

Vishwanath Pattan Endocrinology Wyoming Medical Center

GB/T Translated English of Chinese Standard: GB/T

Αναγκαιότητα και τρόπος ρύθμισης του διαβήτη στους νοσηλευόμενους ασθενείς

Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation Is Part of the Assessment

Transcription:

Relationship between glucose meter error and glycemic control efficacy Brad S. Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Department of Laboratory Medicine and Pathology Mayo Clinic Rochester, MN

Learning objectives List regulatory and clinical issues related to use of glucose meters for critically ill hospitalized patients Weigh the benefits of glycemic control vs. the risks of hospital-acquired hyoglycemia Discuss the impact of glucose meter accuracy on glycemic control effectiveness Review various recommendations for glucose meter accuracy 2

Glucose meters in the hospital Multiple uses for glucose meters in hospital o Dose subcutaneous insulin for diabetic mildly ill patients Same accuracy requirements as home use o Screen for neonatal hypoglycemia o Screen for hypoglycemia or hyperglycemia in hospitalized patients o Manage intravenous insulin for critically ill patients on glycemic control Hourly glucose measurement, hourly IV insulin adjustment Narrower insulin dosing ranges, more opportunity for dosing errors 3

Glycemic control vs. hypoglycemia Van den Berghe 2001 1500 ICU patients randomized into two groups: o Conventional treatment: maintain glucose 180-200 mg/dl, insulin infusion if glucose > 215 mg/dl o Intensive insulin therapy: Intravenous insulin if glucose > 110 mg/dl, maintain glucose 80-110 mg/dl Primary findings: o Among patients in ICU > 5 days, mortality reduced 30% in intensive insulin group o Bloodstream infections, acute renal failure, RBC transfusions, polyneuropathy all reduced 40-50% in intensive insulin group o Increased rate of hypoglycemia in intensive group (6x, 5% of intensive group )

Glycemic control vs. hypoglycemia Leuven II (NEJM 2006) o Repeat of study in medical ICU o TGC only effective in patients with > 3 d ICU stay o Hypoglycemia significant limitation, increased mortality for patients < 3 d in ICU o 6-fold increased rate of hypoglycemia (18.7%) o Glucose meters instead of ABG Subsequent studies o Mixed outcome results (more negative than positive) o Glucose targets varied o Average 5-fold increase in rate of hypoglycemia o Leuven I used arterial blood gas glucose o Most other studies used glucose meters or methods/sample types differed by location

Glycemic control vs. hypoglycemia Single episode of severe hypoglycemia (< 40 mg/dl) associated with increased mortality o OR 2.3 X for death (Krinsley, 2007) In same population patients glycemic control reduced mortality Sensitivity analysis performed to determine how much SH would offset TGC o 4X increase in SH (from 2.3% to 9.2%) predicted to completely offset survival benefit of glycemic control o Could glucose meter inaccuracy be leading to hypoglycemia?

Technologic limitations of glucose meters Number of factors influence relationship of glucose meter to true (usually lab plasma) glucose o Whole blood vs. plasma (conversion factor) o Sample type (capillary vs. venous catheter vs. arterial catheter) Physiologic and technologic limitations o Interferences (medications, po2, others) 7

Technologic limitations of glucose meters Whole blood vs. plasma glucose o Whole blood glucose 15% lower than plasma glucose o US Vendors now calibrate reagents to express plasma-equivalent units

Technologic limitations of glucose meters Conversion of WB to plasma equiv glucose o Function of water content of plasma (PW), water content of red cells (RW), and percent red cells in WB (Hematocrit) o Vendors used agreed upon standards for one conversion factor o Does patient acuity impact validity of PW, RW and Hct assumptions? Lyon ME and Lyon AW Clin Biochem 2011;44:412-7

Technologic limitations of glucose meters Conversion of WB to plasma equiv glucose o Compared PW, RW, Hct values among oupatients, inpatients, and adult ICU patients o Adult ICU patient mean and distribution PW, RW, and Hct values differed markedly from assumptions o Lower Hct and higher PW in adult ICU patients predicted to result in 8.3% of results with > 10% error at value of 10 mm (180 mg/dl) Lyon ME and Lyon AW Clin Biochem 2011;44:412-7

Bias (%) Technologic limitations of glucose meters Hematocrit interference 20.0 Meter A 20.0 Meter B 10.0 10.0 0.0 0.0-10.0-10.0-20.0-30.0 y = 0.0079x - 0.67 r 2 = 0.0001-20.0-30.0 y = -0.74x + 29.80 r 2 = 0.4573-40.0-40.0-50.0 20 25 30 35 40 45 50 55 Hematocrit (%) -50.0 20 25 30 35 40 45 50 55 Hematocrit (%) > 10% overestimation at low Hct > 10% underestimation at high Hct Karon et al Diabetes Tech Ther 2008;10:111-20.

Technologic limitations of glucose meters Capillary vs. arterial/venous glucose Impact of BP, edema and shock, tissue perfusion o Blood pressure: Shock (systolic BP less than 80 mm Hg) associated with falsely decreased or increased capillary glucose measurement Accuracy of capillary WB at low and high glucose o Khan et al Arch Pathol Lab Med 2006;130:1527-32 o Kanji et al Crit Care Med 2005;33:2778-85 Technologic vs. physiologic limitations of capillary sampling largely unknown

Technologic limitations of glucose meters Venous catheter WB glucose in critically ill Overestimates venous plasma glucose o Cook et al Am J Crit Care 2009;18:65-75 o Shearer et al Am J Crit Care 2009;18:224-30 o Karon et al Am J Clin Pathol 2007;127:919-26 Bias with venous catheter samples differs by meter technology o Karon et al, Diabetes Technol Ther 2009:11:819-25 Arterial catheter whole blood best available sample for glucose meter monitoring Assess meter technology with venous catheter whole blood if that will be primary sample type

Technologic limitations of glucose meters Interference studies, ascorbic acid Karon et al Diabetes Tech Ther 2008;10:111-20.

Glucose meters in hospital Error and outliers with WB glucose Condition Sample type Shock, hypotension, dehydration, edema Capillary Hematocrit effect Failure to let alcohol dry Underdosing strips PW or RW effect Medication interference ph, O2 or CO2 tension Use of expired or incorrectly stored strips Temperature extremes Incorrect calibration info Improper/incorrect disinfection Operator error/untrained operators All Capillary Capillary, All All, CVC > art line? All All All All All All All

Glucose meter regulatory issues timeline March 2010 o FDA public forum on glucose meter accuracy o Consensus that 2003 ISO 15197 not appropriate for ICU glucose meter use (95% results within ± 15 mg/dl for glucose < 75 mg/dl, ± 20% for glucose 75 mg/dl) o Debate about whether separate home and hospital, or home/hospital/icu criteria needed o FDA announced new criteria forthcoming

Glucose meter regulatory issues timeline 2011 NACB guidelines on glucose meter accuracy o 95% of glucose meter results within ± 15 mg/dl at glucose < 100 mg/dl ± 15% at glucose 100 mg/dl November 2012, AccuChek Inform II FDA approval o No draft guidance on required accuracy o Limitation statement: the performance of this meter has not been evaluated on critically ill patients o FDA notes limitation statement to be added to all approved hospital use glucose meters o FDA opinion is that critical care use constitutes off label use of device

Glucose meter regulatory issues timeline January 2013 CLSI POCT12-A3 guidelines on glucose meter accuracy o 95% of glucose meter results within ± 12 mg/dl at glucose < 100 mg/dl ± 12.5% at glucose 100 mg/dl 98% within 2003 ISO 15197 guidelines 2013 ISO 15197 revision o 95% of glucose meter results within ± 15 mg/dl at glucose < 100 mg/dl ± 15% at glucose 100 mg/dl use of Parkes Error grid (99% zones A and B)

Glucose meter regulatory issues timeline Sept 2014 o StatStrip receives FDA approval for all hospitalized patients Nov 2014 Venous and arterial whole blood only (neonates) o CMS memo to state surveyors, use meters according to intended use and limitation statement, other use off-label Oct 2016 Makes critical care use for most meters high complexity Validation requirements in specific patient population Personnel requirements (4 yr degree, transcripts) o FDA final guidance for glucose meter manufacturers Home use: slightly more stringent but similar to ISO 15197 Hospital use: similar to CLSI POCT12A-3

Glucose meters in the hospital Will improving glucose meter accuracy and reducing interferences and outliers lead to better patient outcomes during glycemic control in the ICU? 20

Variables impacting glycemic control outcome Elements of glycemic control protocol that may impact patient outcome o Glucose target range o Sophistication of dosing algorithm (point to point vs trending) o System to prompt glucose measurement (manual vs. IT system) o System to relate gluc conc to insulin dose (paper vs. electronic) o Accuracy of glucose monitoring device Hematocrit, bias and precision, medication interference o Competency of staff performing measurement

Variables impacting glycemic control outcome TGC protocols associated with 5 X increase incidence of hypoglycemia Absolute rates of hypoglycemia vary widely between TGC studies depending on target and protocol o 0.34% (Stamford Hospital) o 18.7 % (Leuven II) Does the glucose meter accuracy have anything to do with glycemic control outcomes or rate hypoglycemia?

Mayo glucose meter accuracy study Can newer glucose meter technologies achieve 12-15% total error when fresh whole blood samples are tested on critically ill patients after cardiovascular surgery? If so, because bias or imprecision is reduced? Where are we at today, how did we get there (reducing bias or reducing imprecision) Does reducing glucose meter error improve efficacy of glycemic control in the cardiovascular ICU? Does it matter?

Mayo glucose meter accuracy study At Mayo Rochester StatStrip replaced AccuChek Inform 10/2012 Assess impact on accuracy and precision of glucose measurements in ICU Accuracy when routine clinical samples tested at bedside Retrospective study with Inform and StatStrip Precision with fresh arterial whole blood from critically ill patients

Mayo glucose meter accuracy study Precision (prospective study) AccuChek Inform I (20 ICU patients with 5x measurement at the bedside) o CV of 2.0% at an average glucose value of 142 mg/dl (7.89 mm) StatStrip (20 ICU patients with 5x measurement at the bedside) o CV of 2.7% at an average glucose value of 140 mg/dl (7.78 mm) Both meters precise when fresh whole blood tested at bedside

Mayo glucose meter accuracy study Accuracy (retrospective study) o Over 3 month period, 1602 Inform whole blood glucose measurements performed within 5 minutes of drawing serum glucose (Roche Hexokinase) o Over separate 3 month period, 1093 StatStrip whole blood glucose performed within 5 minutes of serum glucose

Mayo glucose meter accuracy study Median bias 11 mg/dl (0.61 mm) Median (IQR) % bias 9 (4 to 14) %

Mayo glucose meter accuracy study Median bias 1 mg/dl (0.06 mm) Median (IQR) % bias 1 (-3 to 5) % 28

Mayo glucose meter accuracy study Inform (n=1602) Percent within 10% lab 55% 89% StatStrip (n=1093) Percent with 20% lab 92% 98% % within 12.5%/12.5 mg/dl (CLSI POCT12-A3) serum 69% 95% By reducing bias, reduced TEa from ~20% 12.5%

Is StatStrip accurate in different ICU settings? Prospective accuracy study across 5 ICUs o 2 Netherlands, 1 Belgium, 2 US sites o Surgical, medical, burn patients o 1815 paired measurements from 1698 patients o 96.1% met CLSI POCT12-A3 criteria o 99% zone A Parkes Error Grid, 100% zones A/B o 99.1% (223/225) concordance in characterizing hypoglycemia (glucose < 70 mg/dl) o DuBois et al, Crit Care Med 2017;45:567-71.

Impact of insulin dosing errors on glycemic control in ICU Impact on patient outcome o ICU/hospital mortality o Hospital morbidity (infections, transfusions, renal failure) o Requires randomized trial > 1000 patients Impact on glycemic control efficacy o Glycemic variability o Time within target range o Incidence hypo and hyperglycemia o Requires 50-150 patients per study arm

Impact of insulin dosing errors on glycemic control in ICU Why measure glycemic control efficacy? o Hypoglycemia important outcome o Hyperglycemia is what is being avoided o Glycemic variability More variability = more hypo and hyperglycemia Increased variability (extreme highs and lows) may alone decrease survival in ICU o time in target range, hypo and hyperglycemia, variability = better protocol o Can reducing meter error alone lead to a better protocol? 32

Study design Given improved accuracy of meter in ICU o ~20% 12.5% TEa Can we measure impact on glycemic control efficacy? Retrospective review patients post cardiovascular surgery placed on glycemic control in CVS ICU o 12-24 consecutive (30-120 min) glucose values on insulin drip o Period 1 (70 patients monitored with AccuChek Inform) o Period 2 (70 patients monitored with StatStrip) o No change infusion protocol, testing personnel, etc

Study design Measures glycemic variability o Standard deviation (SD) o Continuous overall net glycemic action (CONGA) o Percent values in target range (110-150 mg/dl) o Incidences of hypoglycemia and hyperglycemia Patient demographics Period 1 (6-11/2012) Period 2 (8/13-2/14) P value Mean ± SD age (range) 68 ± 12 (28-92) 65 ± 12 (29-86) 0.22 Gender 39 M/ 31 F 42 M/ 28 F 0.61 Diabetes 35 ND/ 35 T2DM 35 ND/ 35 T2DM Median (range) number glucose values 22 (12-24) 21 (12-24) 0.16

Results Glycemic variability and time within target range Overall results (non-diabetic and T2DM) Median (IQR) glucose (mg/dl) Period 1 (n=70) Period 2 (n=70) P value 141 (126, 156) mg/dl 136 (125, 148) mg/dl 0.005 Median (IQR) standard deviation (SD) Median (IQR) CONGA Median (IQR) percent values in target range (%) 21.6 (16.9, 26.3) mg/dl 19.4 (16.0, 24.2) mg/dl 13.7 (12.4, 19.1) mg/dl 13.5 (10.9, 17.3) mg/dl < 0.0001 < 0.0001 66.7 (50, 74.2) % 74.5 (58.5, 86.7) % 0.002 Glycemic variability decreased and time in target range increased with improved meter accuracy

Results Glycemic variability and time within target range Non-diabetic patients only Median (IQR) standard deviation (SD) Median (IQR) CONGA Median (IQR) time in target range (%) Period 1 (n=35) Period 2 (n=35) P value 18.7 (16.3, 25.6) mg/dl 18.3 (13.3, 21.6) mg/dl 15.4 (12.4, 19.9) mg/dl 13.5 (10.2, 19.0) mg/dl 0.004 0.04 68.8 (61.9, 79.2) % 73.7 (62.5, 87.5) % 0.10 Glycemic variability (SD and CONGA) decreased ~ 20% No significant change in time in target range

Results Glycemic variability and time within target range Type 2 diabetes only Median (IQR) standard deviation (SD) Median (IQR) CONGA Median (IQR) time in target range (%) Period 1 (n=35) Period 2 (n=35) P value 22.4 (17.7, 28.0) mg/dl 21.4 (18.3, 27.5) mg/dl 13.6 (12.3, 18.3) mg/dl 13.5 (11.7, 15.2) mg/dl <0.0001 <0.0001 61.9 (46.7, 72.7) % 78.3 (54.2, 85.7) % 0.006 ~ 40% decrease in glycemic variability (SD and CONGA) ~25% increase in time in target range Bigger impact on patients with Type 2 diabetes

Results Incidence of hypo and hyperglycemia Hypoglycemia (< 70 mg/dl, 3.89 mm) 1 patient, 1 value Period 1 0 patients, 0 values Period 2 Hyperglycemia (> 200 mg/dl, 11.11 mm) 26 patients (7 non-diabetic and 19 T2DM), Period 1 6 patients (1 non-diabetic and 5 T2DM), Period 2

Pediatric burn patients Similar before and after retrospective study design o 63 patients monitored with Inform 1 o 59 patients monitored with StatStrip o Glycemic target 80-130 mg/dl (lower) Mean bias 7.4 ± 13.5 (Inform 1) vs. -1.7 ± 6.9 mg/dl (StatStrip) Glycemic control improved with StatStrip (CONGA, CV, MAGE, MODD) Time to therapeutic range 13.1 5.7 hours Time in range 57.9 85.2% Tran et al, Pediatr Crit Care Med 2016;17:e406-12 39

Conclusions Glucose meter use in the hospital o Capillary sampling and hematocrit effects major issues o Technology can address hematocrit effects o Capillary sampling limitations remain largely undefined 40

Conclusions Glucose meter use in the hospital o Often done on non-diabetic patients o Tighter glucose ranges, more opportunities to translate glucose measure error into insulin dosing error o Sources of error (hematocrit, medication interferences, sample type differences) more pronounced effects Newer glucose meter technologies reduce error of glucose measurement when used at the bedside on critically ill patients Evidence emerging that improving glucose meter performance (reducing error) will improve efficacy of glycemic control 41

Questions? 42