Managing Cone Beam CT Dose in Paediatric Dental Imaging

Similar documents
Scientific Affairs. Scientific Affairs

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. What patients should know

Conversion coefficients for the estimation of effective dose in cone-beam CT

Comparison of adsorbed skin dose received by patients in cone beam computed tomography, spiral and conventional computed tomography scanninng

The Role of. 54 MARCH 2016 // dentaltown.com. radiology feature. by Deepak Gupta, MDS

Two-dimensional (2D) images have provided. Cone Beam Computed Tomography in Dentistry: What Dental Educators and Learners Should Know

STANDARD OF PRACTICE. Dental CT Scanners CONTENTS. April 2011

Awareness and Attitude among Dental Professional towards CBCT

International Journal of Radiation Research, January 2017

Johan Hartshorne 1. Summary. Introduction

Current status of diagnostic imaging in dental university hospitals in Japan

Downloaded from jdm.tums.ac.ir at 9:49 IRDT on Wednesday March 27th 2019 *** ****

SUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT

Head to new heights with your imaging SCANORA 3D

SEDENTEXCT: Guidelines and evidence-based use of CBCT

Kodak Dental Radiography Series. Radiation Safety in Dental Radiography. Dental

Patient dose assessment of CT perfusion scanning at the RSCH

Dosimetry of recently introduced CBCT Units for Oral and Maxillofacial Radiology

Summary at End of the Project, June 2011

Estimation of the effective dose of dental cone-beam computed tomography using personal computer-based Monte Carlo software

Current situation of cone beam computed tomography in dentomaxillofacial radiology education.

Margarete B McGuigan 1 Henry F Duncan 2 Keith Horner 3 RESEARCH AND SCIENCE 297

PHANTOM DOSIMETRY AND IMAGE QUALITY OF GALILEOS COMFORT PLUS CONE BEAM COMPUTED TOMOGRAPHY. Ross D. Hunter. Chapel Hill 2014

Estimated radiation risk of cancer from dental cone-beam computed tomography imaging in orthodontics patients

Multislice CT versus cone beam CT: dosimetric and image quality comparision.

) CBCT. . CRANEX 3D ( (CBCT) Cone beam computed tomography . : (CBVT) Cone beam volumetric tomography.

EFFECTIVE DOSE AND IMAGE QUALITY OF CBCT DIAGNOSTIC IMAGING SCANS. Cameron G. Walker DDS, PhD. Chapel Hill 2012

BRITISH SOCIETY OF DENTAL AND MAXILLOFACIAL RADIOLOGY CORE CURRICULA IN DENTAL RADIOGRAPHY AND RADIOLOGY FOR THE DENTAL TEAM

Dental Radiography STANDARDS & GUIDELINES TABLE OF CONTENTS. College of Dental Surgeons of BC

Radiation Safety & Determining Need for Radiographs

TEST GDP DCP. Dental Hygienists and Therapists. Radiography and Radiation Protection. Radiography and Radiation Protection IR(ME)R 2000

Facing the challenges: IAEA activities to protect the patients

Bone Densitometry Radiation dose: what you need to know

Dosimetry analysis of panoramic-imaging devices in differentsized

Policy Statement on Thyroid Shielding During Diagnostic Medical and Dental Radiology. American Thyroid Association

Prepublication Requirements

Effective dose from direct and indirect digital panoramic units

JCDP ABSTRACT INTRODUCTION /jp-journals

Dental Cone Beam CT. What is Dental Cone Beam CT?

ARTICLE IN PRESS. European Journal of Radiology xxx (2008) xxx xxx

RADIATION PROTECTION IN DENTISTRY AND ORAL AND MAXILLOFACIAL IMAGING

Planmeca Ultra Low Dose

IONISING RADIATION REGULATIONS 99

CBCT Specific Guidelines for South African Practice as Indicated by Current Literature:

Using cone beam technology in orthodontics

User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT)

AWARNESS TOWARDS RADIATION PROTECTION MEASURES AMONG DENTAL PRACTITIONERS IN COORG DISTRICT: A QUESTIONNAIRE STUDY

Highlights and Conclusions Thematic Area 4: Radiation of Patients

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

Image Diagnostic Technology Ltd 53 Windermere Road, London W5 4TJ Tel: +44 (0)

CS 9300 Family. The power of flexibility

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography

CT Scanning Protocol For V2R Guided Surgery Solutions

Ask EuroSafe Imaging Tips & Tricks. CT Working Group. On the use of Diagnostic Reference Levels in CT

Dental Hygiene Spring 2018 Summer 2014 Fall COURSE OUTLINE DHT 1032 Dental Radiography 2 Credit Hours

Entrance Surface Dose Measurement on the Thyroid Gland in Orthopantomography: The Need for Optimization

Chapter 10: Special Topics

A Radiology Portfolio: Today s Solutions for Successful Imaging 1. HISTORY

Assessment of Response of Dental Clinicians and Patients towards Different Imaging Modalities Used In Diagnostic Evaluation of Dental Implant Therapy.

DE UK. ultra low dose. your dental equipment sales, service and training partner. Planmeca

Extraoral Imaging. Chapter 42. Copyright 2018, Elsevier Inc. All Rights Reserved. 1

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

A Cone Beam Computed Tomography Study Among Dental Residents Based on a Questionnaire

Accuracy of digital periapical radiography and cone-beam computed tomography in detecting external root resorption

ISSN (Print) Original Research Article

Why radiation protection matters?

Effect of Spatial Position in the Field of View on Dimensional Changes in Cone Beam Computed Tomography

Radiation Protection- Cath lab

For more than half a century, 2-dimensional (2D)

Cone Beam Computed Tomography (CBCT) and Digital Radiographies Requested and Related Factors by Iranian General Dentists

Radiation protection in dentistry - Do we practice what we learn?

Radiation Protection Dosimetry Advance Access published September 13, 2006

Dosimetry of 3 CBCT Devices for Oral and Maxillofacial Radiology: CB Mercuray, NewTom 3G, and i-cat

Managing Radiation Risk in Pediatric CT Imaging

Course File 243 DDS Physics of Diagnostic Radiology and Oral and Maxillofacial Radiology

Comparison of the sensitivity for detecting foreign bodies among conventional plain radiography, computed tomography and ultrasonography

ENGLISH VGi - R ENG

Periapical radiography and cone beam computed tomography for assessment of the periapical bone defect 1 week and 12 months after root-end resection

Bwanga Phanzu. Degree of Master of Science in Dentistry by coursework and dissertation

Factors determining Radiation Protection Practices among Indian General Dental Practitioners: An Explorative Study

Ask EuroSafe Imaging. Tips & Tricks. Paediatric Imaging Working Group. Shielding in pediatric CT

Justification and optimization of dental panoramic tomography and lateral cephalometric radiography among Finnish children

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Patients, Staff, Public, Animals Evaluate RP needs related to emerging technologies. Produce recommendations and user-friendly guidance on RP.

DMFR UNCORRECTED PROOFS

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

New York Radiation Health and Safety (NYR)

Agenda: Dental Cone Beam Imaging

NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) DENTAL RADIOGRAPHY CURRICULUM REQUIREMENTS GENERAL REQUIREMENTS

International Journal of Current Medical and Pharmaceutical Research

Mastering Basic Skills Eases Conversion to Rectangular Collimation

ENGLISH. Distributed by: QR srl - Via Silvestrini, Verona Italy Tel

Ask EuroSafe Imaging Tips & Tricks. CT Working Group

Radiation Protection education of Radiology trainees

Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

The reduction methods of operator s radiation dose for portable dental X-ray machines

Dental Radiography Core Subject. Digital Radiography

Clinical details: Details of scan: CONE BEAM CT REPORT: Name: H. B. Gender: Reason for referral: Referred by:

Transcription:

Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Managing Cone Beam CT Dose in Paediatric Dental Imaging Raija Seuri (HUS Medical Imaging Center, FI) Cristina Almeida (Centro Hospitalar de Lisboa Central, PT) Theocharis Berris (University of Crete, GR)

Introduction Cone beam CT (CBCT) has been widely used in dentistry for over 10 years (Li, 2013). CBCT technology can provide multiple viewing angles and 3D reconstructions, which help in a more complete evaluation as compared to conventional dental imaging modalities (panoramic radiography, intraoral, etc.). The doses associated with CBCT span a considerably wide range, depending a lot on the equipment (Li, 2013; Ludlow et al., 2008; Pauwels, 2012). CBCT doses are generally lower than multidetector CT (MDCT) doses (Li, 2013). CBCT doses are generally higher than the doses from conventional dental radiography (Li, 2013).

Introduction Doses for dental exposures (In general. Not just for children) Radiographic technique Intraoral radiograph 0.3-21.6 Panoramic radiograph 2.7-38 Lateral cephalometric radiograph 2.2-14 Effective dose (µsv) CBCT 11-1025 (generally <300) CT (mandible) 250-1410 CT (mandible & maxilla) 430-860 Reproduced from: IAEA, 2017. Radiation Protection in Dental Radiology L09 Justification and appropriate use of dental radiology

Introduction Children are more sensitive to radiation than adults (Aps, 2013; IAEA, 2017). Children also have longer life expectancy than adults. This means that the potential long term effects (cancer) due to past irradiations have more time to develop and manifest. Dose to children is higher than dose to adults using the same exposure parameters. The same size of field-of-view (FOV) will cover a larger region in the case of a child. Adult Child Adapted from: IAEA, 2017 Radiation Protection in Dental Radiology L02 Special Considerations for Radiation Protection in Children

Introduction Taking into account that: The possibility of radiation induced stochastic effects (cancer) cannot be ruled out The probability of stochastic effects is cumulative The frequency of dental radiographic examinations is high in children (IAEA, 2017) Practitioners always need to keep radiation doses at the lowest reasonably achievable levels (ALARA Principle)

Justification and Optimisation Literature regarding the indications for CBCT use in dentomaxillofacial paediatric imaging remains limited (Oenning et al., 2017). Thus, the appropriate use of CBCT in paediatric dental imaging needs to be based on proper justification and optimisation of examinations. Applying the basic principles of radiation protection should suffice for safe CBCT use in children (Aps, 2013). Justification Optimisation

Justification Justification is related to the appropriate selection of an imaging technique in a given situation (IAEA, 2017). Any decision that alters the radiation exposure situation should do more good than harm. (ICRP 103, 2007). Justification of medical exposures is more stringent in children (IAEA, 2017). Because of the complexity and uncertainty of the process; The assessment of risk vs benefit should be performed at an individual patient basis and the benefit should clearly outweigh the risk (Aps, 2013; IAEA, 2017). Good Harm

Justification Justification should take dose into account. Patients should not be subjected to screening or routine imaging examinations just because they are new to a dental practice. Dentists should try to get as much relevant information as possible from previous examinations, patient history and clinical examination. Care should be taken to AVOID REPEAT SCANS. Adapted from: IAEA, 2017. Radiation Protection in Dental Radiology L09 Justification and appropriate use of dental radiology

Justification CBCT or MDCT? (If no other modality can be used. General information. Not only for children) CBCT advised when: High sharpness is needed when compared with MDCT (e.g. small anatomy/pathology) Only a localized region needs to be scanned (e.g. single tooth region); large amount of dose can be saved through horizontal collimation MDCT advised when: Soft tissue discrimination is needed Neurological symptoms Contrast agent needed MRI not available Not unequivocally clear for certain applications whether CBCT or MDCT provides better diagnostic image quality at the same dose Adapted from: IAEA, 2017. Radiation Protection in Dental Radiology L09 Justification and appropriate use of dental radiology Radiation Protection in Dental Radiology L09 Justification and appropriate use of dental radiology 9

Justification Further Information - Referral Criteria Cone beam CT for dental and maxillofacial radiology (Evidence-based guidelines) (EC RP 172, 2012). UK: Selection Criteria for Dental Radiography (Faculty of General Dental Practice). Other national guidelines. Adapted from: IAEA, 2017. Radiation Protection in Dental Radiology L09 Justification and appropriate use of dental radiology Radiation Protection in Dental Radiology L09 Justification and appropriate use of dental radiology 10

Optimisation Use child sized protocols which should be reviewed and optimised periodically (IAEA, 2017). A tube voltage of 90 kvp results in lowest doses for all sizes of patients (Pauwels et al, 2017). Use lowest possible mas keeping the image quality at clinically acceptable levels. Using lower mas leads to larger dose reductions than lower kvp. Dose reduction up to 50% was can be achieved by reducing mas for small head sizes (Pauwels et al., 2017; IAEA, 2017). Use the lowest resolution needed to get the desired clinical information (prefer larger voxel sizes than smaller ones) (Librizzi et al., 2011). Reduce field of view to the minimum possible.

Optimisation Shield the thyroid of patients. The thyroid gland seemed to receive four times more radiation in a 10-year-old than in an adolescent because of the anatomy of the patient (Theodorakou et al., 2012). Qu et al. (2012) reported dose reduction of approximately 50% to the thyroid when collar was used. Do not use thyroid shielding if region of interest is at the vertical level of the shielding (use scout image to verify) (IAEA, 2017; Hidalgo et al., 2015) Consider using lead goggles in case of imaging the orbita. Up to 67% dose reduction has been observed (Prins et al., 2011). Keep in mind that CBCT does not provide as good soft tissue differentiation as MDCT or MRI. Stitching of multiple scans may lead to higher doses. Horizontal stitching is discouraged (IAEA, 2017).

Optimisation Consider the artefacts before deciding to expose paediatric patients to CBCT. If you expect them to be very bad, consider alternate imaging methods (Aps, 2013). Use a quality assurance scheme for your equipment and processes. Perform patient dosimetry. Diagnostic reference levels (CBCT DRLs are not widely available yet); There is need for optimisation in CBCT (IAEA, 2017).

Conclusions Pediatric Dental CBCT dose can be reduced if appropriate, personalised justification and proper techniques are used. Dentists should avoid routine CBCT scans and always consider conventional radiographic imaging methods if possible.

References Aps, J.K.M., 2013. Cone beam computed tomography in paediatric dentistry: overview of recent literature. European Archives of Paediatric Dentistry, 14(3), pp.131-140. International Atomic Energy Agency (IAEA), 2107. Radiation Protection in Dental Radiology. Training material https://rpop.iaea.org/rpop/rpop/content/additionalresources/training/1_trainingmaterial/cardiolog y8520.htm. Oenning, A.C., Jacobs, R., Pauwels, R., Stratis, A., Hedesiu, M., Salmon, B. and DIMITRA Research Group, 2017. Cone-beam CT in paediatric dentistry: DIMITRA project position statement. Pediatric Radiology, pp.1-9. Li, G., 2013. Patient radiation dose and protection from cone-beam computed tomography. Imaging science in dentistry, 43(2), pp.63-69. Ludlow, J.B. and Ivanovic, M., 2008. Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology, 106(1), pp.106-114. Pauwels, R., Beinsberger, J., Collaert, B., Theodorakou, C., Rogers, J., Walker, A., Cockmartin, L., Bosmans, H., Jacobs, R., Bogaerts, R. and Horner, K., 2012. Effective dose range for dental cone beam computed tomography scanners. European journal of radiology, 81(2), pp.267-271. Pauwels et al. (2017) Determination of size-specific exposure settings in dental cone-beam CT. Eur Radiol. 2017;27:279-285.

References International Commission on Radiological Protection, 2007. ICRP publication 103. Ann. ICRP, 37(2.4), p.2. Hidalgo A et al. (2015) Effectiveness of thyroid gland shielding in dental CBCT using a paediatric anthropomorphic phantom. Dentomaxillofac Radiol. 44:20140285. Prins, R., Dauer, L.T., Colosi, D.C., Quinn, B., Kleiman, N.J., Bohle, G.C., Holohan, B., Al- Najjar, A., Fernandez, T., Bonvento, M. and Faber, R.D., 2011. Significant reduction in dental cone beam computed tomography (CBCT) eye dose through the use of leaded glasses. Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology, 112(4), pp.502-507. Librizzi, Z.T., Tadinada, A.S., Valiyaparambil, J.V., Lurie, A.G. and Mallya, S.M., 2011. Conebeam computed tomography to detect erosions of the temporomandibular joint: effect of field of view and voxel size on diagnostic efficacy and effective dose. American Journal of Orthodontics and Dentofacial Orthopedics, 140(1), pp.e25-e30. Qu, X.M., Li, G., Sanderink, G.C.H., Zhang, Z.Y. and Ma, X.C., 2012. Dose reduction of cone beam CT scanning for the entire oral and maxillofacial regions with thyroid collars. Dentomaxillofacial Radiology, 41(5), pp.373-378. Theodorakou, C., Walker, A., Horner, K., Pauwels, R., Bogaerts, R., Jacobs Dds, R. and SEDENTEXCT Project Consortium, 2012. Estimation of paediatric organ and effective doses from dental cone beam CT using anthropomorphic phantoms. The British Journal of Radiology, 85(1010), pp.153-160.