SEPARATION OF BRANCHED PFOS ISOMERS BY UPLC WITH MS/MS DETECTION

Similar documents
Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

John Haselden 1, Gordon Dear 1, Jennifer H. Granger 2, and Robert S. Plumb 2. 1GlaxoSmithKline, Ware, UK; 2 Waters Corporation, Milford, MA, USA

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Method Development for the Analysis of Endogenous Steroids Using Convergence Chromatography with Mass Spectrometric Detection

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

Meeting Challenging Requirements for the Quantitation of Regulated Growth Promoters Dexamethasone and Betamethasone in Liver and Milk

Matrix Factor Determination with the Waters Regulated Bioanalysis System Solution

INTRODUCTION CH 3 CH CH 3 3. C 37 H 48 N 6 O 5 S 2, molecular weight Figure 1. The Xevo QTof MS System.

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

Profiling Flavonoid Isomers in Highly Complex Citrus Juice Samples Using UPLC Ion Mobility Time-of-Flight Mass Spectrometry

Analysis of the Non-Ionic Surfactant Triton-X Using UltraPerformance Convergence Chromatography (UPC 2 ) with MS and UV Detection

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

PosterREPRINT RAPID, SELECTIVE SCREENING OF URINE SAMPLES FOR GLUCURONIDES BY LC/MS/MS INTRODUCTION ABSTRACT

[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System

Paul Silcock 1, Anna Karrman 2, and Bert van Bavel 2. Aim. Introduction. Sample preparation

Reducing Sample Volume and Increasing Sensitivity for the Quantification of Human Insulin and 5 Analogs in Human Plasma Using ionkey/ms

Improving Selectivity in Quantitative Analysis Using MS 3 on a Hybrid Quadrupole-Linear Ion Trap Mass Spectrometer

CHALLENGES ASSOCIATED WITH THE ANALYSIS OF MONO SUBSTITUTED POLYFLUORINATED PHOSPHATE ESTERS BY LCMS

Simultaneous Analysis of Intact Human Insulin and Five Analogs in Human Plasma Using μelution SPE and a CORTECS UPLC Column

Direct Analysis of Urinary Opioids and Metabolites by Mixed-Mode µelution SPE Combined with UPLC/MS/MS for Forensic Toxicology

Identification & Confirmation of Structurally Related Degradation Products of Simvastatin

Chemical Analysis Business Operations Waters Corporation Milford MA

Uptake and Metabolism of Phthalate Esters by Edible Plants

Extraction of 25-hydroxy Vitamin D from Serum Using ISOLUTE. PLD+ Prior to LC-MS/MS Analysis

Dry eye disease commonly known as atopic keratoconjunctivitis is an autoimmune disease of

Rapid Analysis of Bisphenols A, B, and E in Baby Food and Infant Formula Using ACQUITY UPLC with the Xevo TQD

Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Detection of Cotinine and 3- hydroxycotine in Smokers Urine

[ APPLICATION NOTE ] APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

The use of mobile-device data to better estimate dynamic population size for wastewater-based epidemiology

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

A RAPID AND SENSITIVE UPLC/UV/MS METHOD FOR SIMVASTATIN AND LOVA S TAT IN IN SU P P O RT O F C L E A NING VA L I DAT IO N S T U DIES

Jody Dunstan, 1 Antonietta Gledhill, 1 Ailsa Hall, 2 Patrick Miller, 2 Christian Ramp 3. Waters Corporation, Manchester, UK 2

Jose Castro-Perez, Henry Shion, Kate Yu, John Shockcor, Emma Marsden-Edwards, Jeff Goshawk Waters Corporation, Milford, MA, U.S. and Manchester, UK

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.

A UPLC/MS Approach for the Diagnosis of Inborn Errors of Metabolism

CHARACTERIZATION AND DETECTION OF OLIVE OIL ADULTERATIONS USING CHEMOMETRICS

The Application of UPLC/MS E for the Analysis of Bile Acids in Biological Fluids

Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS

Improved method for the quantification of lysophospholipids including enol ether

Supplementary Information

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

UPLC/MS/MS Method for the Routine Quantification of Regulated and Non-Regulated Lipophilic Marine Biotoxins in Shellfish

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION

PosterREPRINT SIMULTANEOUS QUANTIFICATION OF PSYCHOTHERAPEUTIC DRUGS IN HUMAN PLASMA BY TANDEM MASS SPECTROMETRY

Determination of Amantadine Residues in Chicken by LCMS-8040

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

Automated Lipid Identification Using UPLC/HDMS E in Combination with SimLipid

Sulfate Radical-Mediated Degradation of Sulfadiazine by CuFeO 2 Rhombohedral Crystal-Catalyzed Peroxymonosulfate: Synergistic Effects and Mechanisms

Supporting Information

[ APPLICATION NOTE ] Profiling Mono and Disaccharides in Milk and Infant Formula Using the ACQUITY Arc System and ACQUITY QDa Detector

Comparison of a UPLC Method across Multiple UHPLC Systems

Supporting information

Dienes Derivatization MaxSpec Kit

LCMS-8030 Application Report. Steroids: Testosterone, Progesterone, Estradiol, Ethinylestradiol

SWATH Acquisition Enables the Ultra-Fast and Accurate Determination of Novel Synthetic Opioids

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Neosolaniol. [Methods listed in the Feed Analysis Standards]

Impurity Identification using a Quadrupole - Time of Flight Mass Spectrometer QTOF

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Quantification of Budesonide Using UPLC and Xevo TQ-S

Analysis of N-Linked Glycans from Coagulation Factor IX, Recombinant and Plasma Derived, Using HILIC UPLC/FLR/QTof MS

Latest Innovations in LC/MS/MS from Waters for Metabolism and Bioanalytical Applications

A High Sensitivity UPLC/MS/MS Method for the Analysis of Clopidogrel and Clopidogrel Carboxylic Acid Metabolite in Human K 2 EDTA Plasma

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

[ application note note ] ]

PosterREPRINT EVALUATION OF A RF-ONLY STACKED RING BASED COLLISION CELL WITH AXIAL FIELD FOR THE LC-MS-MS ANALYSIS OF THE METABOLITES OF RABEPRAZOLE

Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS

The Comparison of High Resolution MS with Triple Quadrupole MS for the Analysis of Oligonucleotides

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Analysis of Food Sugars in Various Matrices Using UPLC with Refractive Index (RI) Detection

Analysis of Pesticides (II) Metribuzin & their metabolites in Rice Jun Yonekubo, Nihon Waters, JAPAN

Sample Concentration and Analysis of Human Hormones in Drinking Water

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

Fast and simultaneous analysis of ethanol metabolites and barbiturates using the QTRAP 4500 LC-MS/MS system

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

Fast Separation of Triacylglycerols in Oils using UltraPerformance Convergence Chromatography (UPC 2 )

Don t miss a thing on your peptide mapping journey How to get full coverage peptide maps using high resolution accurate mass spectrometry

[ APPLICATION NOTE ] A Generic Kit-Based Approach for LC-MS/MS Quantification of Urinary Albumin for Clinical Research APPLICATION BENEFITS

ApplicationNOTE A MULTI-RESIDUE LC/MS/MS METHOD FOR THE DETERMINATION OF 81 PESTICIDE RESIDUES IN FRUIT AND VEGETABLES: PART 1, METHOD OVERVIEW

LC/MS/MS Analysis of Metabolites of Synthetic Cannabinoids JWH-018 and JWH-073 in Urine

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Designer Cannabinoids

Comparison of Biotage Extrahera vs. Manual Sample Processing Using a Vacuum Manifold

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer

Transcription:

SEPARATION OF BRANCHED PFOS ISOMERS BY UPLC WITH MS/MS DETECTION Marian Twohig 1, Nicholas Ellor 1, Keith Worrall 2, Tim Jenkins 2, Gordon Kearney 2 1 Waters Corporation, 1 Cummings Center, Beverly, MA 1915, USA; 2 Waters Corporation, Atlas Park, Simonsway, Manchester M22 5PP, UK SUMMARY EXPERIMENTAL DESCRIPTION Waters ACQUITY Ultra Performance LC (UPLC ) is shown to achieve rapid separation of perfluoroalkane sulphonate homologs and isomers. Tandem quadrupole MS/MS detection provides rich spectral information to facilitate compositional analysis and structural elucidation of perflourinated alkane sulphonates. Sensitive quantitation of perfluorinated octyl sulphonate (PFOS) isomers is demonstrated using the mass spectrometer in the highly selective multiple reaction monitoring (MRM) mode. INTRODUCTION Perfluorooctane sulfonate (PFOS) is a fully fluorinated anion that can be formed by degradation from a large group of related substances, close to 1 in total, referred to as PFOS-related substances. These substances are used in a wide variety of applications including the paper and packaging industries, the textiles/upholstery industries and as a surfactant. Extensive biological monitoring of PFOS and other related perfluorinated compounds in recent years has revealed PFOS as a global pollutant 1 and perfluorinated compounds have even been found in An LC/MS/MS method of analysis using negative ion electrospray is described. Separations were performed using a Waters ACQUITY Ultra Performance LC system. Sample components were separated using a Waters ACQUITY UPLC column and gradient elution. All mass spectra were acquired using a Waters Quattro Premier XE tandem quadrupole mass spectrometer with ZSpray interface. When operated in negative ion electrospray mode, the mass spectra acquired provided useful structural information for analysis of isomer composition and proved suitable for sensitive quantitative analysis. In the first instance, UPLC/MS/MS was shown to provide fast and efficient separation of perfluorinated alkyl sulphonates in the homologous series containing C 4 to C 9. Rapid separation of the PFOS isomer group was achieved and structural information for isomers was obtained from analysis of product ion spectra. Subsequent analysis targeted the PFOS isomer group demonstrating the quantitative potential of the UPLC/MS/MS approach. remote areas of the Arctic. 2 PFOS has been detected at parts per billion levels in human serum and the livers of fish, birds and marine mammals. 2,3,4 Toxicity tests in rodents have raised concerns about potential developmental, reproductive and systemic effects of PFOS. It is for this reason that sensitive detection of all perfluorinated compounds including PFOS and its branched isomers is of great importance. Waters ACQUITY UPLC and Quattro Premier XE Mass Spectrometer.

EXPERIMENTAL CONDITIONS LC conditions - Compositional Analysis System: ACQUITY Ultra Performance LC Column: ACQUITY UPLC BEH Shield RP18 2.1 x 1 mm, 1.7 μm. Column temp: 5 o C Mobile phase: A: 1 mm aqueous ammonium acetate B: 1 mm ammonium acetate in 8/2 v/v methanol/acetontirile Flow rate: 6 ml min -1 Gradient: Time (mins) B. 47 5. 55 6. 55 6.1 47 MS Conditions MS System: Quattro Premier XE Ion Mode: Negative ion ESI Cone Voltage: 2-5 V Collision Energy: 25-4 ev Full scan data: m/z 1 8 Product ion scan: m/z 65-51 (precursor m/z ) MRM data: See Table 1 below Software Data was acquired and processed with Waters MassLynx Software. LC conditions Total PFOS Quantitative Analysis As for compositional analysis except: Column: Waters ACQUITY UPLC BEH C8 2.1 x 5 mm, 1.7 μm. Gradient: Time (mins) B. 47 1. 9 1.2 9 1.3 47 Channel Precursor Ions m/z Product Ions m/z Dwell Time (s) Cone (V) Collision Energy (ev) 1 8.1 5 5 2.1 5 4 3 13.1 5 4 4 169.1 5 4 Table 1. MRM conditions for monitoring of PFOS.

RESULTS AND DISCUSSION Compositional Analysis Analytical methods which provide good chromatographic separations of perfluorinated alkane sulphonates, when coupled with mass spectroscopic detection, allow for the characterization and/or quantitation of complex mixtures. The chromatographic separation of a Technical Grade sample of PFOS in under seven minutes using Waters ACQUITY UPLC Technology in shown in Figure 1 below. MS data were collected in full scan mode. Extracted ion chromatograms (XICs) corresponding to the homologous series [C n F n+1 SO 3 ] - from n=4 to n=9 are shown in Figure 2 below. This shows that the Technical Grade product contains material of different chain lengths. The number of isomers for each homolog is seen to increase with n as expected. Figure 1. Total Ion Chromatogram of Technical Grade PFOS Sample. n=4 n=5 n=6 n=7 n=8 n=9 TIC Figure 2. Extracted ion chromatograms from sample of Technical Grade PFOS.

Further experiments focused upon the group of isomers associated with PFOS (n=8 homolog from Figure 2). A product ion scan experiment monitoring product ions of m/z was conducted. Below, in Figure 3, are the XICs for product ions m/z 8 and m/z 13 (red trace). The peaks observed in the product ion spectra from the precursor m/z generally fit the series [C n F 2n+1 ]- for peaks from m/z 119 to m/z 419; or [C n F 2n SO 3 ] -, which is characterized by peaks from m/z 8 to m/z 33. The chromatograms show different relative intensity profiles depending upon which product ions are monitored. This indicates that the relative intensities of product ions in the spectra of PFOS isomers are different. These differences are the result of structural variations between isomers. Mass spectrometry can therefore provide useful information for structural elucidation. To illustrate this further the spectra of several isomer groups labelled in the chromatograms are shown in Figure 5 overleaf. Further investigations revealed a minor eighth peak with common fragments at Rt 3.71 min and there are clearly further unresolved components within the labelled isomer groups. Spectral information was obtained over the range m/z 65-51 with the use of tandem quadrupole MS/MS enabling structurally significant low mass fragments to be observed, which would not be seen on a conventional ion trap mass spectrometer due to low mass cut-off. Previous work 5 has postulated that the relative abundance of such series may be an indicator of degree of branching in the isomers, which is an important structural characteristic. In particular, the [C n F2 n+1 ]- ions are indicative of cleavage in the a- position of a CF 3 branched chain and the [C n F 2n SO 3 ] - series of the linear alkyl sulphonate portion of the molecule (Figure 4). F 3 C CF 2 m/z 369 [C n F 2n+1 ] - F 3 C m/z 13 [C n F 2n SO 3 ] - CF 2 CF 2 CF 2 CF 2 SO 3 - Figure 4. Ions resulting from alpha-cleavage of branched PFOS. CF 1 3 4 7 5 1 2 6 2. 3. 4. 5. 6. 7. Time Figure 3. Extracted ion chromatograms for m/z 8 and m/z 13 (red trace).

1 8 13 119 18 219 23 28 1 1 13 169 261 28 2 1 8 13 119 18 23 219 28 3 1 8 13 119 261 28 4 1 8 83 169 23 33 5 1 1 8 219 98 119 147 211 419 83 119 13 169 23 8 1 12 14 16 18 2 22 24 26 28 3 32 34 36 38 4 42 44 46 48 5 6 7 m/z Figure 5. Product ion spectra corresponding to peaks in the UPLC separation of PFOS isomers shown in Figure 3. The mixture analyzed is a complex one and the spectra recorded show the characteristic fragments associated with PFOS. Further work with branched model compounds would facilitate the assignment of exact structures. Nevertheless, it is clear that MS/MS can produce useful spectral information containing structurally significant information. Further optimization of chromatographic separation and MS/MS fragmentation is proposed in order to enable better characterization of complex PFOS mixtures. Investigation of simpler systems containing specific isomers will enable better characterisation of the perfluorinated alkyl compounds. The obvious differences in response factor also have an implication for accurate quantitation of PFOS mixtures. Quantitative Analysis In addition to determining the structural composition of PFOS it also important to offer sensitive quantitative methods for the analysis of the material in a variety of environmental matrices. To improve quantitation, a BEH C8 column chemistry was used to produce a rapid separation of PFOS as a single chromatographic peak. The mass spectrometer was operated in MRM mode using the conditions described in the method section. MRM use offers the ultimate combination of selectivity and sensitivity as demonstrated by the chromatogram shown in Figure 6. This shows the trace for the MRM tranisition m/z > 8. The calibration curve for PFOS over the range.1-1 ng/ml is also shown (Figure 7).

1.98.3.4.5.6.7.8.9 1. 1.1 1.2 1.3 1.4 Time Figure 6. PFOS standard at.5 ng/ml. Compound name: PFOS ii Correlation coefficient: r =.9646, r^2 =.9292 Calibration curve: 279.193 * x +.871978 Response type: External Std, Area Curve type: Linear, Origin: Exclude, Weighting: 1/x, Axis trans: None 25 2 Response 15 1 5 ng/ml 1 2 3 4 5 6 7 8 9 1 Figure 7. PFOS calibration curve.

CONCLUSIONS REFERENCES The combination of UPLC with tandem quadrupole mass spectrometry has considerable potential for the compositional analysis and quantitative determination of perfluorinated alkyl compounds. Rapid separation of perfluorinated alkane sulphonate homologs can be achieved in under seven minutes. Tandem quadrupole mass spectral data provides useful information for determining the structural composition of PFOS isomers. The use of MRM enables highly selective detection of PFOS at low levels. 1. Giesy J.P., Kannan K., Environ. Sci. Technol. 35 (21) 1339. 2. Bossi R, Riget FF, Dietz, Sonne C, Fauser P, Dam M,Vorkamp K. Environ Pollut. 25 Jul; 136 (2):323-9. 3. Olsen G.W., Church, T.R., Miller J.P., Burris J.M., Hansen K.J., Lundberg J.K., Armitage J.B., Herron R.M., Medhdizadehkashi Z., Nobiletti J.B., O Neill M.E., Mandel J.H., Zobel L.H., Environmental Health Perspectives. Vol 111, No.16, Dec 23. 4. Koichi I, Okado F, Ito R, Kawaguchi M, Okanouchi N, Nakazawa H., Journal of Chromatography B, 81 (24) 49-56. 5. Langlois I, Oehme M; Organohalogen Compounds, Volume 66 (24) 3973-3978. ACKNOWLEDGEMENT The authors would like to thank Wellington Laboratories (Guelph, Ontario, Canada) for the provision of perfluorinated standards. Waters, ACQUITY Ultra Performance LC, ACQUITY UPLC, UPLC, Quattro Premier, ZSpray, and MassLynx are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 26 Waters Corporation Produced in the U.S.A. May 26 721694EN SE-PDF