Low-dose Three-dimensional CT of Paranasal Sinuses

Similar documents
Original Research THE USE OF REFORMATTED CONE BEAM CT IMAGES IN ASSESSING MID-FACE TRAUMA, WITH A FOCUS ON THE ORBITAL FLOOR FRACTURES

Relationship of the Optic Nerve to the Posterior Paranasal Sinuses: A CT Anatomic Study

Why is CT Dose of Interest?

JMSCR Vol 05 Issue 09 Page September 2017

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

National Imaging Associates, Inc. Clinical guidelines/considerations SINUS & MAXILLOFACIAL AREA CT 70486, 70487, 70488

"The prevalence and CT detection of Onodi cell types."

Imaging of the Paranasal Sinuses

Extranodal Natural Killer/T-Cell Lymphoma Nasal Type: Detection by Computed Tomography Features

HI-Res Extremity Sensation 16

Hi RES Extremity - (04/18/2011) CTDI: ~13 mgy per acquisition Used for evaluation of: Ankle Elbow Hand Wrist Foot /Calcaneous Toes Fingers

X-Ray & CT Physics / Clinical CT

Radiological anatomy of frontal sinus By drtbalu

Prevertebral Soft Tissue Measurements on Lateral Roentgenogram of Cervical Spine in Chinese

Pediatric CT Protocols (18 years old or less)

HEAD & NECK IMAGING. Iranian Journal of Radiology September; 10(3): Published Online 2013 August 30.

Image quality and dose reduction in sinus computed tomography using iterative reconstruction: a cadaver study*

The diagnostic value of Computed Tomography in evaluation of maxillofacial Trauma

The Computed Tomography Examination

Translating Protocols Across Patient Size: Babies to Bariatric

Managing Radiation Risk in Pediatric CT Imaging

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Review Article The Prevalence of Concha Bullosa and Nasal Septal Deviation and Their Relationship to Maxillary Sinusitis by Volumetric Tomography

HEAD AND NECK IMAGING. James Chen (MS IV)

STRUCTURED EDUCATION REQUIREMENTS EFFECTIVE: JANUARY 1, 2016

CT Head Dose Reduction Using Spiral Scanning Protocol

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

B. CT protocols for the spine

STANDARD OF PRACTICE. Dental CT Scanners CONTENTS. April 2011

Practical CT and MRI Anthony J. Fischetti, DVM, MS, DACVR Department Head of Diagnostic Imaging The Animal Medical Center, New York OBJECTIVE:

ORIGINAL ARTICLE RELATIONSHIP OF CONCHA BULLOSA WITH OSTEOMEATAL UNIT BLOCKAGE. TOMOGRAPHIC STUDY IN 200 PATIENTS.

2

CT Radiation Risks and Dose Reduction

Computed Tomography (CT) - Sinuses

Tips and Tricks in Ventral Skull Base Dissection Narayanan Janakiram, Dharambir S. Sethi, Onkar K. Deshmukh, and Arvindh K.

FOR CMS (MEDICARE) MEMBERS ONLY NATIONAL COVERAGE DETERMINATION (NCD) FOR MAGNETIC RESONANCE IMAGING:

CT SCAN PROTOCOL. Shoulder

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling

International Journal of Current Medical and Pharmaceutical Research

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Patient dose assessment of CT perfusion scanning at the RSCH

Variation in frontal cells in relation to chronic frontal sinusitis

Bone and CT Scans Are Complementary for Diagnoses of Bone Metastases in Breast Cancer When PET Scans Findings Are Equivocal: A Case Report

COMPUTED TOMOGRAPHY COURSE

CT examination is a high-radiation-dose imaging technique

International Journal of Biological & Medical Research

Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose?

The Relation between Anatomical Variations of Osteomeatal Complex & Nasal Structures and Chronic Sinusitis by Computed Tomography

CT of the Head, Spine, and Cerebral Vessels

3D high-resolution MR imaging can provide reliable information

Fetal Dose Calculations and Impact on Patient Care

JMSCR Vol 06 Issue 10 Page October 2018

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

Dental Cone Beam CT. What is Dental Cone Beam CT?

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

Reliability of computed tomography scans in the diagnosis of chronic rhinosinusitis

Anatomical and Functional MRI of the Pancreas

Proceeding of the SEVC Southern European Veterinary Conference

Study of Mastoid Air Cells Diseases using Spiral CT

Nonfunctioning Islet Cell Tumors of the Pancreas: Computed Tomography Findings

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami

Catalog Addendum

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Lung Perfusion Analysis New Pathways in Lung Imaging. Case Study Brochure PLA 309 Hospital

Computerised tomographic profile of ethmoid roof on basis of keros classification among ethnic Kashmiri s

MRI and CT of the CNS

ORIGINAL ARTICLE. Utku Aydil, Ali Koksal, Tuncay Ozcelik, Nuri Ozgirgin

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

DIGITAL PANORAMIC RADIOGRAPHY IN DETERMINING THE PREVALENCE OF HALLER S CELLS: A RETROSPECTIVE RADIOGRAPHIC STUDY

Three-Dimensional Volumetric Display of the Nasal Ostiomeatal Channels and Paranasal Sinuses

UNIVERSITY OF MEDICINE AND PHARMACY GR. T. POPA - IASI FACULTY OF DENTAL MEDICINE

Keywords: Mandible, Maxilla, Teeth, Cone Beam CT, Implants, Tumor and tumor-like conditions

Image Diagnostic Technology Ltd IDT Ireland, 15 Market Street, Kinsale, Co. Cork, Ireland P17 XN65 Tel: Mob: IRL:

Clinical Assessment of the Diagnostic Value of Facial Radiography in Facial Trauma Patients at the Emergency Department

Alessandro Albonico Philips


Low Dose Era in Cardiac CT

Computed tomography Acceptance testing and dose measurements

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Accounting for Imaging Dose

The Low Sensitivity of Fluid-Attenuated Inversion-Recovery MR in the Detection of Multiple Sclerosis of the Spinal Cord

PACS: ERGONOMIC CONSIDERATIONS 1

ORIGINAL ARTICLE. Chronic Sphenoid Sinusitis Revisited

Mucocele of paranasal sinuses

Computed tomography of the chest: I. Basic principles

Head to new heights with your imaging SCANORA 3D

Ethan M. Braunstein, M.D. 1, Steven A. Goldstein, Ph.D. 2, Janet Ku, M.S. 2, Patrick Smith, M.D. 2, and Larry S. Matthews, M.D. 2

Breast Tomosynthesis. What is breast tomosynthesis?

Article. Reference. Comparison of organ doses and image quality between CT and flat panel XperCT scans in wrist and inner ear examinations

Neuroradiology MR Protocols

Chronic Frontal Rhinosinusitis: Diagnosis and Management

Managing Patient Dose in Computed Tomography (CT)

Endoscopic Management Of A Giant Ethmoid Mucocele

CT angiography techniques. Boot camp

Optimizing radiation dose by varying age at pediatric temporal bone CT

Facial Soft Tissue Thickness in North-West of Iran

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Gender differences in CT calcium scoring: A phantom study

Transcription:

Chin J Radiol 2004; 29: 7-13 7 Low-dose Three-dimensional CT of Paranasal Sinuses LIANG-KUANG CHEN 1,2,3 YUH-FENG TSAI 1 BO-LIN LIU 1 CHIN-SI LIN 1,3 CHIEN-LI LU 1,3 AN-CHENG SHIAU 4 CHONG-HONG TSAI 4 MING-CHONG TSENG 1,3 KUN-MU LU 1,3 CHENG-TAU SU 1,2 MIN-SZU YAO 5 Department of Diagnostic Radiology 1, Department of Radiation Oncology 4, Shin-Kong Wu Ho-Su Memorial Hospital School of Medicine 2, Fu Jen Catholic University Department of Radiotechnology 3, Yuanpei Institute of Science and Technology College Department of Diagnostic Radiology 5, Taipei Medical University-Municipal Wan Fang Hospital The goal of this study was to develop an optimal radiologic techniques for low-dose 3D CT of paranasal sinus to enhance the image quality in a reasonable absorbed x-ray dose for the patients. The ultimate aim is to provide higher quality images for physicians to identify pathological changes from the image, to make preoperative planning, to evaluate the risk of the operation of paranasal sinus and to minimize the occurrence of intra-operative complications. We analyzed spatial resolution and noise signals generated by various groups of mas by using the dual technique of Plexiglas phantom and water phantom. In addition, Rando phantom was used to measure CTDI (computed tomography dose index) in the same scan area in a function of different slice thicknesses and spacings. The concept of MSAD (Multiple- scan average dose) was applied to calculate patient s absorbed dose. From July 2001 to January 2002, 19 patients were enrolled in the study and underwent CT examinations of paranasal sinuses. Among these patients, 19 were male and 5 female. Conventional 2D CT had been generally used with satisfactory results, especially for patients with chronic paransal sinusitis. However, 3D CT not only provides a better anatomical information of the paranasal sinuses but also follows the ALARA principle. Reprint requests to: Dr. Liang-Kuang Chen Department of Diagnostic Radiology, Shin Kong Wu Ho-Su Memorial Hospital. No. 95, Wen Chang Road, Taipei 111, Taiwan, R.O.C. The acceptable image quality could be achieved under the imaging parameters of 90 mas and 77 mas for axial and coronal sections, respectively. Three-dimension images could further be analyzed with SSD (shaded surface display), MPR (multiplanar reformats), and MIP (maximum intensity projection) to provide more information for both radiologists and clinicians. Key words: 3D CT; ALARA; Paranasal sinuses; Spatial resolution Computed tomography has been widely used to image paranasal sinuses, depicting anatomical pathologies in related diseases, such as chronic paranasal sinusitis, tumor, face trauma, etc. However, possible body damage caused by radiation dose is often neglected, when patients undergo axial and coronal slice scanning. When post-contrast scanning is required, the exposure dosage will be higher. The study is to reconstruct the 3D paranasal sinus images with low-dose CT scanning to reduce the radiation dose in the body while provide better anatomic images for radiologists and otorhinolaryngologist, etc. Surgeons can make better preoperative plans and risk evaluations and further reduce side effects. In the meanwhile, patients receive lower doses of x-ray exposure than with traditional axial and coronal CT examination. Less artifacts were generated with this technique when patient wears dentures. 3D images can be viewed from different angles and thus, provides more anatomical information. METHODS AND MATERIALS The tomoscanner is a Siemens PLUS4 A. The choices of the phantom and procedure were based on the image characteristics. The directions were as follows:

8 Paranasal sinuses Materials 1. Plexiglas phantom: It was built of six groups of lead thread couples. The bar size is used for the determination of spatial resolution. The spacing between the groups are, 0.5, 0.6, 0.8, 1, 1.3 and 1.6 mm respectively. 2. Water phantom: Imaging was developed by homogenized water to measure noise signals. 3. Rando phantom: It was a phantom emulating actual body exterior but contained true bone inside. It was used to measure the computed tomography dose index (CTDI). Methods Plexiglas phantom experiment 1. Under the conditions of 219 mas and 171 mas for the axial and coronal sections respectively, plexiglas phantom scan was performed at 140 kvp, 3 mm slice thickness and 4 mm spacing. That was the conventional technique used for paranasal sinus scanning. 2. Under the conditions of 219, 194, 166, 141, 116, 90 and 65 mas for axial scan and 171, 159, 146, 111, 94, 77 and 60 mas for coronal sections respectively, plexiglas phantom scan was performed at 140 kvp, 2 mm slice thickness and 4 mm spacing. 3. The bar sizes (mm), namely, spatial resolution was analyzed under conventional fashion or different mas scan conditions. (see Fig. 1). Water phantom experiment 1. Under the condition of 219 mas and 171 mas for axial and coronal sections respectively, Water phantom scan was performed at 140 kvp, 3 mm slice thickness and 4 mm spacing. That was the conventional technique used for paranasal sinus scanning. 2. Under the conditions of 219, 194, 166, 141, 116, 90 and 65 mas for axial scan and 171, 159, 146, 111, 94, 77 and 60 mas for coronal sections respectively, water phantom scan was performed at 140 kvp, 2 mm slice thickness and 4 mm spacing. 3. The regions of interest (ROIs) were marked out at four places in equal distance;then, the noise signals generated in various groups of mas were determined by the given software (see Fig. 2). Rando phantom experiment The images of Rando phantom were processed with the Plus 4 software to calculate the CTDI under various scanning parameters, respectively. CTDI conversed to MSAD After scanning the axis and coronal section, the CTDI is conversed to the MSAD using the following formula: MSAD = CTDI T/BI where BI refers to table feeding (mm),and T means slice thickness [3]. Clinical application A group data of lowest MSAD, with spatial resolution and noise signals approaching those in conventional techniques, were chosen to image patients. The 3D image was obtained by reconstructing the selected thin cut slices containing suspected lesions and was presented to radiologists and otorhinolaryngologists for lesion identification and analysis. From July 2001 Figure 1. Spatial resolution of lead bar size (mm). Figure 2. The regions of interest at four equidistances.

Paranasal sinuses 9 to January 2002, 19 patients underwent 3D CT examinations of paranasal sinuses. The 2D scan image and 3D reconstructed image were presented to radiologists and otorhinolaryngologist who did not know factual scanning conditions. The images of 19 patients were analyzed in one session.the diseases of these 19 patients included 11 chronic paranasal sinusitis, 1 rhinopolyps, 2 mucocele, 1 fracture of orbital floor, 1 fistula between nonnasality and nasal sinuses and 3 tumor extruded nasal cavity and sinuses. RESULTS We scanned Plexiglas phantom, water phantom and Rando phantom with CT under different conditions. The data for the axial section are shown in Table 1. When the mas value was reduced from 219mAs to 90mAs, spatial resolution degraded to 0.8 from 0.6. The noise signals rose from 4.2 to 8.12. If it was further reduced to 65 mas, spatial resolution was degraded to 1 mm, and noise signals further rose to 9.2. In this circumstance, the image quality significantly differed from that was obtained in conventional imaging parameter. The MSAD would be higher when we scan an equidistance in 39 thinner sections with >141mAs than that in 20 thicker sections with 219 mas. Therefore, this condition is in appropriate for thin slice scan. Because of the low reduction rate of MSAD at 90 mas, the group of scanning parameters at 90 mas was chosen for axial scan. The data for the coronal section are shown in Table 2. When the mas value was reduced 77 mas, there was no obvious change in spatial resolution as compared with the original 171 mas (all are 0.8), but noise signals rose from 4.9 to 8.9. For a value of 60 mas, spatial resolution degraded to 1 mm and noise signals rose to 10, seriously affecting image quality. If we applied over 111 mas to scan an equidistance of Table 1. The relation between spatial resolution, noise signals, CTDI, MSAD and section during axial scanning under different mas, slice thicknesses and spacings condition Axial Sections mas 219 219 194 166 141 116 90 65 Spatial resolution (mm) 0.60 0.60 0.80 0.80 0.80 0.80 0.80 1.00 Noise signals (standard error) Region 1 of interest 4.20 5.40 5.50 5.50 6.50 7.80 8.00 9.00 Region 2 of interest 4.50 4.50 5.40 5.50 7.10 6.60 8.00 10.00 Region 3 of interest 4.00 5.80 5.40 5.30 6.90 7.50 8.20 10.00 Region 4 of interest 4.10 5.30 5.30 5.70 7.90 7.30 8.30 9.00 Total 16.80 21.00 21.60 22.00 28.40 29.20 32.50 38.00 Average 4.20 5.25 5.40 5.60 7.10 7.30 8.12 9.20 Slice thickness/ spacing 3\4 2\2 2\2 2\2 2\2 2\2 2\2 2\2 CTDI (mgy) 32.00 41.00 36.20 31.10 26.40 21.60 16.80 12.10 MSAD (mgy) 23.60 41.00 36.20 31.10 26.40 21.60 16.80 12.10 Section 20.00 39.00 39.00 39.00 39.00 39.00 39.00 39.00 Note: data in the first column are the original conditions used previously. data in the seventh column is the data applied in the clinical practice after the experiment. Table 2. The relation between spatial resolution, noise signals, CTDI, MSAD and section during coronal scanning under different mas, slice thicknesses and spacing conditions Coronal Sections mas 171 171 159 146 111 94 77 60 Spatial resolution (mm) 0.80 0.80 0.80 0.80 0.80 0.80 0.80 1.00 Noise signals (standard error) Region 1 of interest 4.80 6.10 6.30 6.80 7.50 7.40 9.80 10.00 Region 2 of interest 5.20 5.60 5.70 6.80 7.80 7.10 9.40 9.10 Region 3 of interest 5.00 5.40 5.30 7.40 7.50 8.10 8.00 9.50 Region 4 of interest 4.50 5.60 6.20 6.80 7.70 8.20 8.40 11.00 Total 19.80 22.70 23.50 28.10 30.50 30.80 35.60 40.00 Average 4.90 5.60 5.90 7.00 7.60 7.70 8.90 10.00 Slice thickness/ spacing 3\4 2\2 2\2 2\2 2\2 2\2 2\2 2\2 CTDI (mgy) 24.60 32.00 29.70 27.30 20.80 17.60 14.00 12.00 MSAD (mgy) 18.50 32.00 29.70 27.30 20.80 17.60 14.00 12.00 Section 20.00 39.00 39.00 39.00 39.00 39.00 39.00 39.00 Note: data in the first column are the original conditions used previously. data in the seventh column is the data applied in the clinical practice after the experiment.

10 Paranasal sinuses 39 cuts, the MSAD is instead higher than that with171 mas and 20 thicker slice. Thus, this condition also is inproper for thin slice scan. Because its reduction rate of MSAD is 22.2% at 77 mas, the imaging parameters at 77 mas is suitable for coronal section scan. With a fixed kvp value and image algorithms filter, at 77 mas and 90 mas, 2 mm slice thickness and 2 mm spacing, we perform scanning at the axial and coronal sections. Comparing the images obtained in conventional fashion, all physicians agreed the image quality degraded as the imaging parameters declined from171 mas to 90 mas, but the change couldn t interfere with the interpretations(see Fig. 3). They all agreed 3D image could display more diagnostic details about adjacent anatomic structures, and facilitate the otorhinolaryngologist more information to make preoperative planning and to evaluate the risk of the operation. DISCUSSION The objective of this study was to reasonably reduce mas in CT scanning so that we can obtain three sets of thin cut slices for reconstruction image without increased patient s dose. As shown in Table 1.and table2., the reduction of mas would decrease the output of X-ray. Two effects occur: 1. it displayed the detail of CT radiograph resulted in increasing spatial resolution. 2. X-ray photons that reach the detector would have wide distribution and increase the noise signals. After the different ROI calculations, the effect not only occurred in those of specific gray scales, even the images with some statically variance, which would not affect the ability of the analytical method. According to the different image characteristics, low-dose 3D CT of paranasal sinuses could be reconstructed into 3D images in three algorithyms of pattern, including SSD (shaded surface display), MPR (multiplanar reformats) and MIP (maximum intensity projection). SSD could achieve the delicate visualization of the bony injuries. The details of bone could be displayed more clearly. The 3D images of parenchyma and bone tissue can be contrasted out and be viewed in different angles. We can adjust the threshold values to remove the metal artifacts resulting from dentures. (see Fig. 4, 5). MIP images could provide the reconstructed image with more hierarchy than the original image scanned under conventional conditions. The correlation of parenchyma, bone tissue and the focus could be easily shown by contrast. The images also could be visualized in different angles (see Fig. 6). MPR images could depictt the images on the basis of standard coronal scans with additional axial and sagittal scans if necessary. More information was generated so that the clinicians could anticipate the depth and angles when performing endoscopic examinations.(see Fig. 7-9). Figure 3. The original image under coronal section with low dose (77 mas) and the fistula between left nonnasality and paranasal sinuses (marked with arrow). Figure 4. SSD (shaded surface display) shows the reconstructed coronal image of bone contrast and the fistula between left nonnasality and paranasal sinuses (marked with arrow).

Paranasal sinuses 11 In addition, exposure dose of radiation to patients could be reduced. According to the literature, radiation damage had been proved when paranasal sinus was imaged with both axial and coronal scanning or 3D CT scanning. In general, the exposure dose of eyeballs in conventional scanning was 0.5 and 2 Gy in 3D CT scanning, respectively; about 1/3 of the dose was absorbed by the eyeballs. If the absorbed dose over 5 Gy, the cataract may occur [1, 2, 3]. Therefore, the mas is reduce from 171 mas to 90 mas, while the kvp and image algorithms were kept constant, there was no significant effect on the image quality and the interpretation of the pathological region and regional anatomy. CONCLUSION Comparison of the results with other reports in the literature yielded the following useful information. Figure 7. MPR shows the reconstructed arrowy image of bone contrast and the fistula between left nonnasality and paranasal sinuses (marked with arrow). Figure 5. SSD shows the reconstructed coronal image of parenchyma contrast and the fistula between left nonnasality and paranasal sinuses (marked with arrow). Figure 8. MPR (multiplanar reformats) shows the reconstructed arrowy image and the mucocele (marked with arrow). Figure 6. MIP (maximum intensity projection) shows the reconstructed coronal image of parenchyma contrast and the fistula between left nonnasality and paranasal sinuses (marked with arrow). Figure 9. MPR shows the reconstructed axial image of parenchyma contrast and the fistula between left nonnasality and paranasal sinuses (marked with arrow).

12 Paranasal sinuses Because scanning parameters and sampling counts are different and the imaging perception may be subjective to radiologists and clinicians, the results of every study aren t consistent. According to the ALARA principle, that is, lowest absorbed dose with acceptable images quality is appreciated. This study showed that the original image quality was acceptable for diagnosis under the scan condition of 90 mas and 77 mas for axial and coronal sections respectively. Three dimensional images not only offered lower exposure to radiation than that of conventional CT, but also could provide further information for radiologists and clinicians in preoperative planning, evaluation of surgical risk for paranasal sinus and minimizing the risk of intra-operative complications. REFERENCES 1. Damilakis J, Prassopoulos P, Mazonakis M, Bizakis J, Papadaki E, Gourtsoyiannis N. Tailored low dose threedimensional CT of paranasal sinuses. Clin Imaging 1998; 22: 235-239 2. Maclennan A.C. Radiation dose to the lens from coronal CT scanning of the sinuses. Radiology 1995; 50: 265-267 3. Gholkar A, Gillespie E, Isherwood I. Dynamic low dose three dimensional computed tomography: a preliminary study. Br J Radiol 1988; 61: 1095-1099 4. Zinreich J, Kennedy D, Rosenbaum A, Gayler B, Kumar A, Stammberger H. Paranasal sinuses CT Image for endoscopic surgery. Radiology 1987; 163: 769-775 5. Zhang S. The measure of the patient s absorbed dose during the CT scans. Graduation thesis of radiology faculty of Zhongtai Medical Technic Institute. 2000; 1: 10-11 6. Roberts DN, Hampal S, East CA, Lloyd GAS. The diagnosis of inflammatory sinonasal disease. J Laryngol Otol 1995; 109: 27-30 7. Katzberg RA, Zonneveld FW, Fukuta K. A decade of clinical three-dimensional imaging: A review. Invest Radiol 1994; 29: 574-589 8. Marmolya G, Wieson EJ, Yagan R. Paranasal sinuses: low-dose CT. Radiology 1991; 181: 689-691 9. Sillers MJ, Kuhn FA, Vickery CL. Radiation exposure in paranasal sinus imaging. Otolaryngol Head Neck Surg 1995; 112:248-251 10. Som PM. CT of the paranasal sinuses. Neuroradiology 1985; 27: 189-201

Paranasal sinuses 13