Clinical Study McGrath Video Laryngoscope May Take a Longer Intubation Time Than Macintosh Laryngoscope

Similar documents
The Superiority Of Mcgrath Videolaryngoscope After Failed Conventional Laryngoscopy

INFORMED CONSENT FOR publication of the present

NiravKotak 1, Abhishek Kesarwani 2 1

In 2011 I received an unrestricted lecture honorarium from Ambu I have been loaned equipment by Intavent Direct (Teleflex) and Aircraft Medical As an

Research Article Comparison of Direct and Indirect Laryngoscopes in Vomitus and Hematemesis Settings: A Randomized Simulation Trial

The reasons 13/11/ Cost 2. Availability 3. Comparison 4. Complications 5. Knowledge. Pulmonary and critical care medicine (PCCM) fellows.

British Journal of Anaesthesia 100 (4): (2008) doi: /bja/aen002 Advance Access publication January 31, 2008

Comparison of laryngoscopy and intubating conditions using kings vision laryngoscope and C-MAC video laryngoscope

Comparison of McGrath Series 5 video laryngoscope with Macintosh laryngoscope: A prospective, randomised trial in patients with normal airways

ISPUB.COM. The Video-Intubating Laryngoscope. M Weiss THE LARYNGOSCOPE INTRODUCTION TECHNICAL DESCRIPTION

A comparison of the Glidescope R to the McGrath R videolaryngoscope in patients

Omiros Chalkeidis*, Georgios Kotsovolis, Apostolos Kalakonas, Maria Filippidou, Christos Triantafyllou, Dimitris Vaikos, Epaminondas Koutsioumpas

The Glidescope Ò system: a clinical assessment of performance

Video Laryngoscopy and its future in Airway Management

Published online 2015 December 5. Research Article

Original Contributions

British Journal of Anaesthesia 102 (1): (2009) doi: /bja/aen342

Is the Airtraq optical laryngoscope effective in tracheal intubation by novice personnel?

Evaluation of the Airtraq Ò and Macintosh laryngoscopes in patients at increased risk for difficult tracheal intubation*

Al-Qasmi et al. Al-Qasmi et al. Trials 2013, 14:298

Cronicon ANAESTHESIA. Khaled Mohammed Elnaghy 1 and Javier Yuste 2 * Abstract. Keywords: Intubation; Video laryngoscopes; GlideScope; LMA CTrach

Hemodynamic response to laryngoscopy in ischemic heart disease: Macintosh blade versus GlideScope videolaryngoscope

LEVITAN S FIBREOPTIC STYLET: BEYOND BARRIERS. - Our Perspective.

C-MAC videolaryngoscope compared with direct laryngoscopy for rapid sequence intubation in an emergency department A randomised clinical trial

Comparison between C-MAC video-laryngoscope and Macintosh direct laryngoscope during cervical spine immobilization

Study of Comparison of Standard And English Macintosh Laryngoscope Blades in Patients Undergoing Surgery Under General Anaesthesia.

Original Article. Summary. Introduction

A comparison between the GlideScope Ò Video Laryngoscope and direct laryngoscope in paediatric patients with difficult airways a pilot study

Dr.Bharghavi.M 2 nd year post graduate Dept of Anaesthesia

Hong Kong Journal of Emergency Medicine. KL Tsui, CY Hung, CW Kam. t p= p=0.004

Comparison. F Agro immobilization. to DL Trauma/SCI. GS experiences. to other. GS experiences. GS experiences. to DL.

Clinical Comparison of Lightwand-guided versus Conventional Endotracheal Intubation

British Journal of Anaesthesia 101 (6): (2008) doi: /bja/aen288 Advance Access publication October 3, 2008

Comparison of the Airtraq to the Bonfils Fibroscope for Endotracheal Intubation in a Simulated Difficult Airway

Randomized controlled trial of the A.P. Advance, McGrath, and Macintosh laryngoscopes in normal and difficult intubation scenarios: a manikin study

British Journal of Anaesthesia 102 (4): (2009) doi: /bja/aep013 Advance Access publication February 20, 2009

Z.U.M.J.Vol.19; N.5; September; 2013

Airway Anatomy. Soft palate. Hard palate. Nasopharynx. Tongue. Oropharynx. Hypopharynx. Thyroid cartilage

Optimising tracheal intubation success rate using the Airtraq laryngoscope

The LMA CTrach TM, a new laryngeal mask airway for endotracheal intubation under vision: evaluation in 100 patients

Samal et al: Comparison of the Macintosh and Airtraq Laryngoscope

RESPIRATION AND THE AIRWAY Comparison of the performance of four laryngoscopes in a high-fidelity simulator using normal and difficult airway

LARYNGOSCOPE IN MYASTHENIA GRAVIS PATIENTS

Recent Advances in Airway Management HA Convention 2014

Accident and Emergency Medicine Academic Unit, The Chinese University of Hong Kong, Shatin, Hong Kong

Endotracheal intubation with airtraq W versus storz W videolaryngoscope in children younger than two years - a randomized pilot-study

Medicine. McGrath Series 5 videolaryngoscope vs Airtraq DL videolaryngoscope for double-lumen tube intubation A randomized trial

Displacement of the epiglottis during intubation with the Pentax-AWS Airway Scope. Suzuki, Akihiro ; Katsumi, Norifumi ; Honda, Takashi ; Sasakawa,

Comparison of the Berman Intubating Airway and the Williams Airway Intubator for fibreoptic orotracheal intubation in anaesthetised patients.

Case Report Complete Obstruction of Endotracheal Tube in an Infant with a Retropharyngeal and Anterior Mediastinal Abscess

Comparative Evaluation of Performance of Videolaryngoscope vs Fastrach Intubating Laryngeal Mask Airway

This is a repository copy of Videolaryngoscopy versus direct laryngoscopy for tracheal intubation in children (excluding neonates).

Laryngoscopy: Time to broaden our horizon.

Complications following the use of the Combitube, tracheal tube and laryngeal mask airway

DIFFICULT tracheal intubation is a cause of severe

ENDOTRACHEAL INTUBATION POLICY

How do you use a bougie as an airway adjunct for endotracheal intubation?

Case Report Long-Term Outcomes of Balloon Dilation for Acquired Subglottic Stenosis in Children

Video Laryngoscopes Novelty, Mandatory or Morbidity. Dr Brent May FANZCA MBBS Specialist Anaesthetist Retrieval Consultant - Adult Retrieval Victoria

COMPARISON OF DIRECT VERSUS INDIRECT LARYNGOSCOPY FOR INTUBATION IN THE PREHOSPITAL SETTING: A SYSTEMATIC REVIEW AND META-ANALYSIS

Diagnosis & Management of the Difficult Airway

Research Article Comparison of GlideScope Videolaryngoscopy to Direct Laryngoscopy for Intubation of a Pediatric Simulator by Novice Physicians

N. Jagannathan 1, *, J. Hajduk 1, L. Sohn 1, A. Huang 1, A. Sawardekar 1, B. Albers 1, S. Bienia 2 and G. S. De Oliveira 3.

Anesthetic challenges when elective case becomes emergent

World Journal of Research and Review (WJRR) ISSN: , Volume-5, Issue-4, October 2017 Pages

Airway Scope for Emergency lntubations: Usefulness of a New Video-laryngoscope

PREDICTORS OF DIFFICULT INTUBATION: STUDY IN KASHMIRI POPULATION

Airway/Breathing. Chapter 5

Educational Session: Evaluation and Management of the Difficult Airway

Comparison of two methods of gum elastic bougie aided endotracheal intubation using Airtraq video laryngoscope

NECK CIRCUMFERENCE TO THYROMENTAL DISTANCE RATIO IN COMPARISON WITH MALLAMPATI TEST: A STUDY TO PREDICT DIFFICULT INTUBATION

Research Article Clinical Outcome of a Novel Anti-CD6 Biologic Itolizumab in Patients of Psoriasis with Comorbid Conditions

A clinical evaluation of the Bonfils Intubation Fibrescope*

Hemodynamic responses to orotracheal intubation with a video laryngoscope. in infants: a comparison study

Current Anaesthesia & Critical Care

Other methods for maintaining the airway (not definitive airway as still unprotected):

A study to rank predictors of Preoperative Airway Assessment on the basis of accuracy of prediction while performing Laryngoscopy in Obese patients

Haemodynamic response to orotracheal intubation: direct laryngoscopy versus fiberoptic bronchoscopy

This interdisciplinary clinical support document provides guidelines for the safe establishment of an artificial airway.

Research Article Opioid Use Is Not Associated with Incomplete Wireless Capsule Endoscopy for Inpatient or Outpatient Procedures

Research Article Predictions of the Length of Lumbar Puncture Needles

ORIGINAL ABSTRACT. Introduction Endotracheal intubation may be challenging, SIGNA VITAE 2011; 6(2): LAURENCE C. TORSHER

DIFFICULT AIRWAY MANAGMENT. Dr.N.SANTHOSH KUMAR MD ANESTHESIA (2 nd Yr)

Haemodynamic response to endotrachial intubation: direct versus video laryngoscopy

A systematic review of the role of videolaryngoscopy in successful orotracheal intubation

Difficult Airway Management during Anesthesia: A Review of the Incidence and Solutions

Pre anaesthetic evaluation of difficult tracheal intubation

A Systematic Review of Evidence on the Effectiveness of Video Laryngoscopes in Acute Care Facilities

Comparison of Ease of Insertion and Hemodynamic Response to Lma with Propofol and Thiopentone.

Original Article. Tracheal intubation with a camera embedded in the tube tip (Vivasight TM ) Summary. Methods

Video laryngoscopy in paediatric anaesthesia in South Africa

Advanced Airway Management. University of Colorado Medical School Rural Track

Video Laryngoscope: A Review of the Literature

Case Report Bilateral Distal Femoral Nailing in a Rare Symmetrical Periprosthetic Knee Fracture

Correspondence should be addressed to Alicia McMaster;

Difficult Airway. Department of Anesthesiology University of Colorado Health Sciences Center (prepared by Brenda A. Bucklin, M.D.)

Author s response to reviews

The era of video laryngoscopy has entered its second decade. Video Laryngoscopy: A Debate

The implication of using dominant hand to perform laryngoscopy: an analysis of the laryngoscopic view and blade-tooth distance

Research Article Balloon Dilatation of Pediatric Subglottic Laryngeal Stenosis during the Artificial Apneic Pause: Experience in 5 Children

Transcription:

Anesthesiology Volume 2015, Article ID 901903, 4 pages http://dx.doi.org/10.1155/2015/901903 Clinical Study McGrath Video Laryngoscope May Take a Longer Intubation Time Than Macintosh Laryngoscope Prerana N. Shah and Kaveri Das Department of Anaesthesiology, Seth G. S. Medical College & KEM Hospital, Parel, Mumbai, Maharashtra 400012, India Correspondence should be addressed to Prerana N. Shah; pps@kem.edu Received 1 August 2015; Revised 27 October 2015; Accepted 8 November 2015 Academic Editor: Young-Chang P. Arai Copyright 2015 P. N. Shah and K. Das. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Background. Video laryngoscopes provide better view and can improve ease of intubation compared with standard laryngoscopes. Methods. A prospective randomized study was done on 60 patients, 18 to 65 years old, comparing McGrath video laryngoscope and Macintosh laryngoscope. The aim was to compare the ease, efficacy, and usability of them during routine airway management. The primary endpoint was duration of intubation and the secondary endpoints were Cormack and Lehane grade of laryngoscopic view, number of intubation attempts, and incidence of complications. Results. There was an increase in total duration of intubation with McGrathvideolaryngoscopewith42.9± 19.5secondscomparedtoMacintoshlaryngoscopewith17.9± 4.6seconds.InMacintosh group, 73.3% had grade I, 20% had grade II, and 6.7% had grade III Cormack Lehane view, while in McGrath group, 83.3% had grade I, 13.3% had grade II, and 3.3% had grade III. In McGrath group, 6 patients (20%) required more than 120 seconds to get intubated and only 73.3% were intubated in 1 attempt, while patients in Macintosh group had 100% successful intubation in 1 attempt. Pharyngeal trauma was seen with McGrath videolaryngoscopy. Conclusion. Duration of laryngoscopy, intubation, and total duration of intubation were significantly higher in McGrath group than in Macintosh group. McGrath group required a higher number of intubation attempts. 1. Introduction Difficulty with tracheal intubation, especially in patients with unanticipated difficult airway, remains a frequent cause of anaesthesia related morbidity and mortality. Tracheal intubation can be attempted if laryngeal structures especially thecordsaredistinctlyseen.itisusuallydoneusinglaryngoscopes with different types of blades, namely, Macintosh, Miller, McCoy, and so forth. Laryngoscopes are devices typically consisting of a blade attached to a handle with a light source that allows examination of the upper airway through the mouth, often for the purposes of intubation. New video laryngoscopes like McGrath, Storz, and so forth may help in intubation by giving a better view of the airway or improve intubation success by reducing the number of attempts of intubation and thereby decrease stress related to intubation [1]. The McGrath video laryngoscope is a videobased system for tracheal intubation which uses a video camera embedded into a camera stick [2]. The resulting video image is displayed on a small colour liquid crystal display (LCD) which is attached to the handle of the device [3]. The unit is used in much the same manner as an ordinary laryngoscope with the exception that once it has been introduced into the mouth by a few centimetres, one s attention should be directed to the LCD display. Thus, it helps to have a better view of the glottis and makes intubation easier by better manipulationofthetubethroughthelarynx[2].theaimof our study was to compare the ease, efficacy, and the usability of the McGrath video laryngoscope with those of standard Macintosh laryngoscope during routine airway management. The primary endpoint was duration of intubation [4] and the secondary endpoints were Cormack and Lehane grade of laryngoscopic view, number of intubation attempts (Table 3), andincidenceofcomplications(table4)likeliptrauma,teeth trauma, and pharyngeal trauma.

2 Anesthesiology Enrollment Assessed for eligibility (n =60) Randomized (n =60) Excluded (n =0) (i) Not meeting inclusion criteria (n =0) (ii) Declined to participate (n =0) (iii) Other reasons (n =0) Allocated to intervention (n =30) (i) Received allocated intervention ( n=30) (ii) Did not receive allocated intervention (give reasons) (n =0) Allocation Allocated to intervention (n =30) (i) Received allocated intervention ( n=30) (ii) Did not receive allocated intervention (give reasons) (n =0) Lost to follow-up (give reasons) (n =0) Discontinued intervention (give reasons) (n =0) Analysis Excluded from analysis (give reasons) (n =0) Excluded from analysis (give reasons) (n =0) Figure 1: CONSORT 2010 flow diagram. 2. Material and Methods The study was approved by the Ethics Committee for Research on Human Subjects, Seth G. S. Medical College & KEM Hospital, Parel, Mumbai 400012 (address: Department of Pharmacology & Therapeutics, 1st Floor, College Building, Seth G. S. Medical College & KEM Hospital, Parel, Mumbai 400012) (Chairperson: Dr. Kamal Hazari, on 28 December 2011) (Ethical Committee EC/156/2011). This study was conducted with written informed consent from the study subjects. Theprospectivestudywasdoneintheagegroup18to 65 years undergoing elective surgical procedures needing general anaesthesia and allocated randomly using sealed envelope technique to undergo laryngoscopy using either McGrath video laryngoscope or Macintosh laryngoscope. From previous studies done by Walker et al. [5], the difference in intubation time in two groups was 17 seconds with standard deviationof19seconds.assumingalphaof5%andpowerof study set at 90%, the sample size obtained was 26 per group. Accounting for dropouts final sample size was 30 per group (Figure 1). Standard general anaesthesia technique using agents like Midazolam 0.02 mg/kg, Pentazocine 0.6 mg/kg, Propofol 2 mg/kg, and Rocuronium 1 mg/kg muscle was used. Patients refusing to consent for study and patients intubated prior to surgery were excluded from the study. After achieving muscle relaxation, laryngoscopy was done by the same anaesthesiologistwhowasfamiliarwithbothtypesoflaryngoscopesand had experience of at least 50 intubations with McGrath laryngoscope. Either McGrath video laryngoscope or Macintosh laryngoscope was used followed by tracheal intubation using suitable size endotracheal tube. In Macintosh laryngoscopy, intubation was done by direct visualization of the larynx [5]. In McGrath laryngoscopy, after the scope was introduced a few cm into the mouth, attention was shifted to the LCD screen attached to the scope for viewing the glottis. Once the glottis was visible, an appropriate sized endotracheal tube was guided with the help of view seen on the LCD screen. The number of attempts needed was noted. Total duration of intubation (T) was sum of T1 and T2 where T1 was duration of laryngoscopy and T2 was duration of intubation. For McGrath videolaryngoscopy, the time taken from removal of face mask and insertion of laryngoscope blade between the teeth to indirect view of the glottis seen on the LCD screen was T1, while for Macintosh laryngoscopy, it was the time taken from removal of face mask and insertion of laryngoscope blade between the teeth to visualization of vocal cords and noting the Cormack and Lehane grading. For McGrath videolaryngoscopy, the time taken from indirectviewofthevocalcordsseenonthelcdscreenand passing the endotracheal tube in the trachea to confirmation

Anesthesiology 3 Table 1: Duration of laryngoscopy and intubation. Group (mean ± SD) Macintosh McGrath P value Duration of laryngoscopy (T1) 9.3 ± 3.33 15.6 ± 6.05 0.0001 (seconds) Duration of intubation (T2) 8.6 ± 2.7 28.0 ± 18.5 0.0001 (seconds) Total duration of intubation (T) (T1 + T2)(seconds) 17.9 ± 4.6 42.9 ± 19.5 0.0001 of its tracheal placement was T2. For Macintosh laryngoscope, the time taken from visualization of vocal cords and passing the endotracheal tube in the trachea to confirmation of its tracheal placement was T2. Cormack and Lehane grade of laryngoscopic view [6] and number of intubation attempts were noted. When tracheal intubation failed at the first attempt but succeeded at the second attempt, the sum of the times taken for the first time and the successful attempt was noted. Failure to intubate was considered if there was oesophageal intubation, if there was inability to place tracheal tube into the trachea within 120 seconds, and if more than three attempts were required. Incidence of complications like trauma to lips, pharynx, and teeth was noted and appropriately managed. Student s t-test was used for between-group comparisons. CategoricaldatawascomparedusingChi-squaretestor Fisher s exact test. 3. Results The demographic data with respect to age, gender, weight, height, and ASA grading was comparable. Both Macintosh group and McGrath group had 28 patients (93.3%) belonging to Mallampati grade I and 2 patients (6.67%) belonging to grade II. T1 was 9.33±3.33 seconds in Macintosh group and 15.6± 6.05 seconds in McGrath group. T2 was 8.6 ± 2.7 seconds in Macintosh group and 28.0 ± 18.5 seconds in McGrath group required. Both T1 and T2 were statistically significant using unpaired t-test. Total duration of intubation (T = T1 + T2) in Macintosh group was 17.9 ± 4.6 seconds and for McGrath group it was 42.9 ± 19.5 seconds. This is shown in Table 1. In Macintosh group, 73.3% had grade I, 20% had grade II, and 6.7% had grade III Cormack Lehane view, while in McGrath group, 83.3% had grade I, 13.3% had grade II, and 3.3% had grade III. This was not significant using Chi-square test which was applied and shown in Table 2. All patients in Macintosh group were intubated in the first attempt. In McGrath group, 22 patients (73.3%) were intubated in 1 attempt, 2 patients (6.6%) in 2 attempts, and 6 patients (20%) in 3 attempts. This was statistically significant using Pearson Chi-square test. McGrath group showed successful intubation of 24 patients (80%) with 6 patients (20%) requiring more than 120 seconds to intubate which Cormack Lehane grade of view (%) Table 2: Cormack Lehane grade of view. Grade Macintosh McGrath P value 1 22 (73.3%) 25 (83.3%) 2 6 (20%) 4 (13.3%) 3 2 (6.67%) 1 (3.3%) 4 0 (0%) 0 (0%) Table 3: Intubation attempts. 0.629 Attempts Macintosh McGrath P value 1 30 (100%) 22 (73.3%) Number of intubation 0.009 attempts % 2 0 (0%) 2 (6.6%) 3 0 (0%) 6 (20%) Failure to intubate 0 (0%) 6 (20%) 0.024 Table 4: Complications. Complications Macintosh McGrath Number (%) Number (%) Lip trauma (%) 0 (0%) 0 (0%) Teeth trauma (%) 0 (0%) 0 (0%) Pharyngeal trauma (%) 0 (0%) 3 (10%) was considered as intubation failure. This was statistically significant using Fisher s exact test. Complicationslikepharyngealtraumawereslightlymore in McGrath videolaryngoscopy than in Macintosh. There were no cases of lip or teeth trauma. 4. Discussion Direct laryngoscopy using Macintosh laryngoscope has been used for laryngoscopy and intubation since 1943. Video laryngoscopes provide better laryngoscopic view on a video monitor and can hence potentially improve ease of intubation. In this study of 60 patients, intubating condition comparing the ease, efficacy, and the usability of the McGrath video laryngoscope with those of standard Macintosh during routine airway management was done. OurstudyhadnosignificantdifferencesinCormackand Lehane grading. Laryngoscopy and intubation time required in our study was less in Macintosh group. The number of attempts required to intubate was more in McGrath group than Macintosh group. Haemodynamics like pulse, blood pressure, and oxygen saturation were well maintained and were comparable in both groups. Duration for laryngoscopy and intubation in our study was comparable with the various studies done by Shippey et al. [3], Walker et al. [5], Piepho et al. [7], and Taylor et al. [8]. Shippey et al. found that the median time to intubate the trachea with the McGrath video laryngoscope by experienced anaesthetists was 24.7 seconds [3] and Walker et al. [5]

4 Anesthesiology found a median intubation time of 47 seconds with inexperienced anaesthetic trainees. The mean time to intubate using the McGrath video laryngoscope by paramedics in the study made by Piepho et al. [7] was 48.2 seconds whereas in the studymadebytayloretal.itwas37.2seconds[8].inour study, second-year trainee did all the intubations. The increased duration of laryngoscopy and intubation in our study can be attributed to the fact that the video laryngoscope is new and was used by second-year trainee, thus probably requiring a greater learning curve, whereas the Macintosh laryngoscope was being used routinely. However, the study done by Taylor et al. [8] showed that McGrath video laryngoscope had better glottic view with more successful intubation thanmacintoshlaryngoscopewhenusedinpatientsinsimulated difficult airway requiring manual in-line stabilization. However, studies done by Noppens et al. [9], Piepho et al. [7], and Taylor et al. [8] showed that glottis view was better in McGrath video laryngoscope. The higher number of attempts required to intubate in McGrath group was also comparable to the study done by Walker et al. [5]. Piepho et al. [7] showed that McGrath video laryngoscope had similar rates of successful intubation like in Macintosh. Noppens et al. [9] showed that they had more successful intubation with McGrath video laryngoscope than Macintosh. Although video laryngoscopes provide a good view of the larynx, they may not guarantee an easy tracheal intubation and may prolong the time required for successful intubation [3, 9], as seen in our study. While the angle of the McGrath and other video laryngoscope blades relative to the axis of the trachea improves the indirect view of the laryngeal inlet, it is necessary to redirect the tracheal tube more anteriorly, which canmakeitmoredifficulttoachievesuccessfulintubationof the trachea. It is also often necessary to bend the styletted tracheal tube to a greater degree than with direct vision laryngoscopes, which can hinder stylet removal and increase the time of intubation [8]. In spite of prolonged duration of intubation in McGrath group, hemodynamic changes were comparable to those in Macintoshgroupandwerenotstatisticallysignificant. Complications like pharyngeal trauma were also slightly more with McGrath than Macintosh which can be explained by the nonfamiliarity of the video laryngoscope. This could also be attributed to difference in vision. While Macintosh provided a direct vision of the interior of the oropharynx and larynx, the McGrath video laryngoscope showed an indirect view on the display and the hand-eye coordination for firsttime users could lead to trauma. Thus, it was seen that demographic characteristics, view of larynx, and haemodynamic response and complications were comparable between the two groups. But duration of laryngoscopy, intubation, and total duration of intubation were significantly higher in McGrathgroupthaninMacintoshgroup.Also,McGrath group required a higher number of intubation attempts. However, videolaryngoscopy gives a display that can be seen by others and can be used for teaching especially to novices. The limitation of the study was that the study was done by a single anaesthesiologist with a relatively more experience with Macintosh than with McGrath. 5. Conclusion Although there was no significant difference in Cormack Lehanegradeofvieworhaemodynamicresponselikepulse rate, blood pressure, oxygen saturation, and incidence of complications in both groups, McGrath video laryngoscope needed longer time and more attempts as compared to Macintosh laryngoscope. Disclosure All the authors attest the fact that they have read the paper, attest the validity and legitimacy of the data and its interpretation, and agree on its publication. Conflict of Interests The authors report no conflict of interests regarding the publication of this paper. Authors Contribution PreranaN.Shahwasresponsibleforconductingthestudy, study design, patient recruitment, data collection, data analysis, and writing. Kaveri Das helped design and conduct the study and helped in patient recruitment, data collection, data analysis, and writing. References [1] B. Shippey, D. Ray, and D. McKeown, Case series: the McGrath videolaryngoscope an initial clinical evaluation, Canadian Anesthesia,vol.54,no.4,pp.307 313,2007. [2] Aircraft Medical, McGrath Brochure, 2014, http://www.aircraftmedical.com/index.php?option=com content&view= article&id=263&itemid=180, http://www.aircraftmedical.com. [3] B.Shippey,D.Ray,andD.McKeown, UseoftheMcGrathvideo laryngoscope in the management of difficult and failed tracheal intubation, British Anaesthesia, vol.100,no.1,pp. 116 119, 2008. [4] D.C.Ray,C.Billington,P.K.Kearnsetal., Acomparisonof McGrath and Macintosh laryngoscopes in novice users: a manikin study, Anaesthesia,vol.64,no.11,pp.1207 1210,2009. [5] L. Walker, W. Brampton, M. Halai et al., Randomized controlled trial of intubation with the McGrath Series 5 video laryngoscope by inexperienced anaesthetists, British Anaesthesia,vol.103,no.3,pp.440 445,2009. [6] R.S.CormackandJ.Lehane, Difficulttrachealintubationin obstetrics, Anaesthesia, vol. 39, no. 11, pp. 1105 1111, 1984. [7] T.Piepho,K.Weinert,F.M.Heid,C.Werner,andR.R.Noppens, Comparison of the McGrath Series 5 and GlideScope Ranger with the Macintosh laryngoscope by paramedics, Scandinavian Trauma, Resuscitation and Emergency Medicine, vol. 19, no. 1, article 4, 2011. [8] A.M.Taylor,M.Peck,S.Launcelottetal., TheMcGrathSeries 5 Series 5 videolaryngoscope vs the Macintosh laryngoscope: a randomised, controlled trial in patients with a simulated difficult airway, Anaesthesia,vol.68,no.2,pp.142 147,2013. [9] R. R. Noppens, S. Möbus, F. Heid, I. Schmidtmann, C. Werner, and T. Piepho, Evaluation of the McGrath series 5 videolaryngoscope after failed direct laryngoscopy, Anaesthesia, vol. 65, no.7,pp.716 720,2010.

MEDIATORS of INFLAMMATION The Scientific World Journal Gastroenterology Research and Practice Diabetes Research International Endocrinology Immunology Research Disease Markers Submit your manuscripts at BioMed Research International PPAR Research Obesity Ophthalmology Evidence-Based Complementary and Alternative Medicine Stem Cells International Oncology Parkinson s Disease Computational and Mathematical Methods in Medicine AIDS Behavioural Neurology Research and Treatment Oxidative Medicine and Cellular Longevity