EVOS MINI with IM Nailing

Similar documents
Knee spanning solutions

Cost and Time Considerations: Are Minifragment Plates Worth It? Disclosure. More Disclosures. Are minifragment plates worth it? it depends!

Increasing surgical freedom Restoring patient function

Surgical Technique. Anterolateral and Medial Distal Tibia Locking Plates

INTERTAN Nails Geared for Stability

Pediatric Tibia Fractures Key Points. Christopher Iobst, MD

Knee Surgical Technique

Surgical Technique. Fibula Rod System

Practical Reduction Techniques: Diaphyseal Reduction. Philip Wolinsky University of California at Davis Medical Center

Surgical Technique. Targeter Systems Overview

Surgical Technique. 3.5mm and 4.5mm Lateral Proximal Tibia Locking Plates

Fibula Rod System. Lateral Malleolus Fracture Indications:

Surgical Technique. Distal Humerus Locking Plate

ROTATIONAL PILON FRACTURES

Distal Femur Fractures in The Elderly The Ideal Construct

Trauma fixation choices chart your fracture

Surgical Technique International Version

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

D-RAD SMART PACK Plating System. A series of case studies

GREENS SURGICALS. Redefining Excellence INSTRUMENT SYSTEM PREPARED BY: DR. VINAY KUMAR

CURRENT TREATMENT OPTIONS

Intramedullary Tibial Preparation

Femur. Monoaxial Locking Plate System. Operative Technique. Distal Lateral Femur Universal Holes Targeting Instrumentation.

Zimmer ITST Intertrochanteric/ Subtrochanteric Fixation System. Abbreviated Surgical Technique

Surgical Technique International Version

Locking Ankle Plating System. Surgical Technique

Specification Summary

7/23/2018 DESCRIBING THE FRACTURE. Pattern Open vs closed Location BASIC PRINCIPLES OF FRACTURE MANAGEMENT. Anjan R. Shah MD July 21, 2018.

Screw Placement and Types Matter

Surgical Technique. Calcaneal Locking Plate

Surgical Technique. Olecranon Locking Plate

PediLoc 3.5mm and 4.5mm Contour Femur Plate Surgical Technique

Extramedullary Tibial Preparation

Fractures of the tibia shaft treated with locked intramedullary nail Retrospective clinical and radiographic assesment

Minimally Invasive Plating of Fractures:

Zimmer Small Fragment Universal Locking System. Surgical Technique

Technique Manual Technique Manual Rev. A 01/11/2011

Zimmer MIS Periarticular Distal Femoral Locking Plate

PediLoc 3.5mm and 4.5mm Bowed Femur Plate Surgical Technique

TRIGEN Hindfoot Fusion Nail Surgical Technique

Surgical Technique. Lower Extremity Plates and Straight Plates

No disclosures relevant to this topic Acknowledgement: some clinical pictures were obtained from the OTA fracture lecture series and AO fracture

Surgical Technique. CONQUEST FN Femoral Neck Fracture System

Tibial Shaft Fractures

Distal Cut First Femoral Preparation

CASE REPORT. Bone transport utilizing the PRECICE Intramedullary Nail for an infected nonunion in the distal femur

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

Surgical Technique. Clavicle Locking Plate

Angular Malalignment in Subtrochanteric and Proximal Shaft Femur Fractures after Intramedullary Nailing using SIGN Nails

Pocket Guide. Version 4.1: Fracture reduction and deformity correction software

Lengthening & Deformity correction with. Fixator Assisted Nailing

LCP Distal Tibia Plate

Zimmer MIS Periarticular 3.5mm Proximal Tibial Locking Plate

5/31/2018. Ipsilateral Femoral Neck And Shaft Fractures. Ipsilateral Neck-Shaft Fractures Introduction. Ipsilateral Neck-Shaft Fractures Introduction

PediLoc Tibia Surgical Technique

Pilon fractures. Pat Yoon, MD Minneapolis Veterans Affairs Medical Center Associate Professor, University of Minnesota

Zimmer Patient Specific Instruments. Surgical Technique for Gender Solutions Natural-Knee Flex System

Techique. Results. Discussion. Materials & Methods. Vol. 2 - Year 1 - December 2005

Surgical Technique. Proximal Humerus Locking Plate

Polyax Distal Femoral Locked Plating System. Surgical Technique

Surgical Technique. VISIONAIRE FastPak Instruments for the LEGION Total Knee System

Index. B Backslap technique depth assessment, 82, 83 diaphysis distal trocar, 82 83

Principles Starting Point Trajectory L/A/R Stable Construct. DISCLOSURES Hassan R. Mir, MD, MBA, FACS 5/16/2017

CASE REPORT. Antegrade tibia lengthening with the PRECICE Limb Lengthening technology

Principles of intramedullary nailing. Management for ORP

DARCO. Bow 2 Plate SURGIC AL TECHNIQUE

Surgical Technique. *smith&nephew. Supracondylar Plates

IMPORTANT MEDICAL INFORMATION Advanced Orthopaedic Solutions INTRAMEDULLARY NAILS Warnings and Precautions (SINGLE USE ONLY)

A New Benchmark in Tibial Fixation

OptiLock Periarticular Plating System For Distal Tibial Fractures. Surgical Technique

FIBULAR & SYNDESMOSIS MALUNIONS

Ankle Fractures in the Elderly: How to Deal with Poor Bone Quality

FIRST DESCRIBED RETROGRADE TECHNIQUE USING MEDIAL EPICONDYLE 6/5/2018. RETROGRADE vs. ANTEGRADE FEMORAL NAILING

PediNail Pediatric Femoral Nail

The TRIGEN SURESHOT system has my back.

Distal Femoral Locked Plating System. Product Rationale & Surgical Technique

Small Fragment Plating System

Operasjonsteknikk. Tibia

Proximal Tibial Locked Plating System. Surgical Technique

THE RING BUTTERFLY FRAGMENT

Patient Specific Instruments. Surgical Technique

Case Report. Antegrade Femur Lengthening with the PRECICE Limb Lengthening Technology

Surgery-Ortho. Fractures of the tibia and fibula. Management. Treatment of low energy fractures. Fifth stage. Lec-6 د.

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Percutaneous Reduction Of Femoral Fractures During Intramedullary Fixation: A Technical Tip

PROXIMAL TIBIAL PLATE

Surgical Technique International Version. Clavicle Locking Plate

OPERATIVE TECHNIQUE GALAXY FIXATION SHOULDER

Femur Condylar Plate System Procedural Steps.

Small Fragment Plating System. Securing optimal fixation through locked and compression plating technology

INVISION Total Ankle Replacement System with PROPHECY Preoperative Navigation Revision of a Failed Agility Total Ankle Replacement

BICEPTOR Tenodesis System

AcUMEDr. FoREARM ROD SYSTEM

Clinical. Solutions. Synthes Solutions. Foot and Ankle.

Zimmer Periarticular Proximal Humeral Locking Plate

Adult Posttraumatic Reconstruction Using a Magnetic Internal Lengthening Nail

tep Ankle Fracture Plating System

DISCLOSURES Hassan R. Mir, MD, MBA, FACS

Intramedullary Nailing: History & Rationale

Intramedullary Rodding of Distal Tibial Shaft Fractures with Intra Articular Extension

Transcription:

Case Series Dr. John A. Scolaro EVOS MINI with IM Nailing A series of studies

Introduction Intramedullary nailing has become the standard for many long bone fractures. Fracture reduction prior to nail insertion is of paramount importance. Multiple closed, percutaneous and open techniques have been described to achieve fracture reduction. Traditional reduction techniques in IM Nailing still play a major role in fracture reduction when nailing. However, each technique has its weakness. Bumps and rolls provide support but do not secure the fracture site and can be unstable. Percutaneous clamps are less effective in the presence of fracture comminution and achieving the correct clamp vector to reduce the fracture is often times difficult. Placement of clamps in muscular or obese patients can be challenging and potentially place unnecessary pressure on the soft tissue envelope. Finally, applied clamps can loosen during the course of nailing. External fixators/distractors provide length and traction, but are less effective in controlling alignment and rotation. Pins must be kept out of the path of the nail and this technique is unfamiliar to some surgeons.

In some instances, limited open fracture reduction and stabilization with mini-fragment implants has a role in long bone intramedullary nailing. These implants can be especially beneficial in: Segmental fractures. Comminuted fractures. Proximal or distal fractures subject to deforming muscle forces. Open fractures providing direct access to fracture site. Closed fractures unable to be reduced by closed or percutaneous methods. When using this technique: Soft tissue stripping should be minimized in order to achieve reduction and apply the plate. Only the necessary soft tissue is cleared from the cortical edges to obtain a reduction read. Screws (especially unicortical) are placed with sharp drills and under bulb saline irrigation in order to avoid thermal necrosis. In the setting of an open fracture, traumatic soft-tissue injury is used when possible for fracture visualization. The plate is almost always removed when an open wound is present at the fracture site but is left on at the surgeons discretion in closed injuries when a surgical incision is made through non-compromised skin/soft tissue. When a surgical procedure has time constraints (poly trauma/unstable patient/etc.) reliable and expeditious reduction via mini fragment plates prior to intramedullary fixation may assist the surgeon. Mini fragment plates can be very beneficial in obtaining and maintaining reduction during IM Nailing. They can provide reduction of both simple and complex fractures.

Case 1: Femur Patient information 50-year-old male Segmental femur fracture with comminuted proximal and simple transverse distal fracture. Case information This patient presented after he sustained multiple gunshot wounds to the right lower extremity and abdomen (Figures 1 2). An injury to the profunda femoris artery was initially identified. The patient was initially placed in a large JET-X spanning external fixator on the day of injury to allow for vascular intervention (Figure 3). Challenges The deforming muscles forces of the proximal femur presented a problem with the sub-trochanteric fracture. There was difficulty in obtaining a reliable cortical read with regard to length, alignment and rotation. The transverse distal diaphyseal fracture was at the junction of femoral diaphysis and the wider metaphyseal region and required reduction prior to nail passage. Figure 1 Figure 2 Figure 3

Solution Performed ORIF of the distal femoral segmental fracture. Utilized a modified clamp to obtain fracture compression, and utilized a 2.7mm Strength Plate to maintain and reduce the distal fracture (Figure 4). Unicortical non-locking screws were used to avoid disrupting the path of the nail. This technique converted the segmental femur fracture into a subtrochanteric fracture. EVOS MINI Implants EVOS MINI 8 hole 2.7mm Strength Plate 6 x unicortical 2.7mm Non-Locking Screws Figure 4 Figure 5 The proximal ex-fix Schantz pin was left in place to help obtain correct starting point for nail (Figure 5). The guide wire was passed, the canal reamed and the nail placed with the plate still in place (Figures 6 7). Limb length and rotation was based on fluoroscopy of contralateral limb. The distal 2.7mm strength plate was modified in-situ to allow for two interlocks to be placed in the nail (Figure 8). IM Nail implants used: TRIGEN FAN Nail Figure 6 Figure 7 Figure 8

Case 2: Tibia Patient information 45-year-old male Type IIIA open segmental tibia fracture Case information This patient was struck by a motor vehicle traveling at unknown speed. The patient was intubated and sedated in the trauma bay. He was found to have a 3mm open wound about the proximal tibia and a 5cm open wound about the distal tibia with exposed bone. The patient also sustained multiple blunt thoracoabdominal injuries. Challenges The segmental tibia fracture consisted of a simple oblique fracture proximally and an oblique fracture with butterfly segment distally. There was also an associated comminuted fibula fracture (Figure 9 10). Figure 9 Figure 10

Solution The proximal and distal fractures were anatomically reduced using two 2.4mm Strength Plates to obtain length, alignment and rotation (Figure 11). These plates also maintained reduction during nail placement. EVOS MINI Implants 2 x EVOS MINI 6 hole 2.4mm Strength Plate 8 x 2.4mm Unicortical Non-Locking Screws Figure 11 The wounds were debrided and irrigated. A reamed and statically locked tibial nail was placed through the suprapatellar approach. The mini fragment plates were removed upon completion of the nail procedure. Meticulous layered soft tissue closure of both wounds (Figures 12 13). Other implants used: TRIGEN META-NAIL Tibial Nail Figure 12 Figure 13

Case 3: Tibia Patient information 43-year-old male Closed segmental right tibia and fibula fracture Case information This patient sustained a 9 foot fall from a ladder. The patient had immediate pain and deformity in the right leg with inability to bear weight (Figures 14 15). Challenges The proximal third tibia fracture presented a potential valgus and procurvatum deformity during nail placement. The proximal fracture was a closed oblique fracture with posterior extension while the distal fracture was a closed diaphyseal fracture with a butterfly fragment. Figure 14 Figure 15

Solution The proximal fracture was anatomically reduced with a 2.7mm Strength Plate. This aided in maintaining anatomic fracture reduction during nail placement. EVOS MINI Implants EVOS MINI 5 hole 2.7mm Strength Plate 4 x 2.7mm Unicortical Non-Locking Screws A small surgical incision was made along the posterior medial proximal tibia. The fracture edges were exposed, and a clamp was applied to obtain an anatomic reduction of the fracture. A 2.7mm Strength Plate with Unicortical Non-Locking Screws was applied along the posterior medial surface of tibia to maintain reduction during reaming and subsequent nail passage. Other implants used: TRIGEN META-NAIL Tibial Nail A reamed and statically locked tibial nail was placed through the suprapatellar approach. The plate was left in to support fixation (Figures 16 17). Figure 16 Figure 17

Case 4: Tibia Patient information 44-year-old female Closed segmental tibia fracture with distal intra-articular extension Case information This patient presented with an isolated closed fracture of the right lower extremity after being struck by a car at high speeds. The patient was found to have a segmental tibial fracture. A CT scan was obtained to evaluate the distal tibial plafond due to findings on the plain radiographs. Challenges The patient sustained a comminuted segmental tibia fracture with a closed transverse fracture proximally and a closed comminuted diaphyseal fracture distally (Figures 18 19). A CT scan also demonstrated intraarticular extension with a simple minimally displaced coronal plane fracture and sagittal plane medial malleolus fracture (Figures 20 22). An ipsilateral comminuted fibula fracture was also present. Figure 20 Figure 21 Figure 18 Figure 19 Figure 22

Solution The distal tibia joint surface was addressed first. Percutaneous clamps were used to reduce the fracture. Three 2.7mm Non-locking Screws were placed using lag technique to secure the fragments. EVOS MINI Implants 3 x 2.7mm Non-Locking Screws with Washers (placed using lag technique after percutaneous clamp reduction) EVOS MINI 6 hole 2.7mm Flex Plate with Non-Locking Screws A small posterior medial incision was made to access to the proximal tibia fracture. The fracture was initially reduced using a clamp and a 2.7mm Flex Plate with 4 posteriorly directed bicortical screws was subsequently placed to aid in maintaining length, alignment and rotation. A reamed, statically locked tibial nail was placed through the suprapatellar approach (Figures 23 24). Other implants used: TRIGEN META-NAIL Tibial Nail Figure 23 Figure 24

Case study author What specific details about the EVOS MINI Plating System do you like? Option of Strength or Flex Plate Pre-contoured specialty plates (e.g. hook) Multiple plate options at 2.0, 2.4 and 2.7mm sizes Long mini fragment screw sizes, especially in 2.4mm and 2.7mm sizes Surgeon quote The EVOS minifragment set provides the surgeon with multiple fixation options to manage both simple and complex fractures; it can be a valuable tool in gaining and maintaining reduction while treating long bone fractures with intramedullary nails. John A. Scolaro, MD Dr. Scolaro is an Assistant Professor of Orthopaedic Surgery at the University of California, Irvine Medical Center in Orange, California. Dr. Scolaro is a fellowship trained orthopaedic traumatologist who specializes in acute orthopaedic trauma as well as the post traumatic reconstruction of fracture malunions and nonunions. This series of studies is provided for information and educational purposes only. It is not intended to serve as medical advice. It is the responsibility of the treating physician to determine and utilize the appropriate products and techniques according to their own clinical judgment for each of their patients. For detailed product information, including indications for use, contraindications, effects, precautions and warnings, please consult the product s Instructions for Use (IFU) prior to use. Smith & Nephew, Inc. 1450 Brooks Road Memphis, TN 38116 USA www.smith-nephew.com Telephone: 1-901-396-2121 Information: 1-800-821-5700 Orders/Inquiries: 1-800-238-7538 Trademark of Smith & Nephew. 2015 Smith & Nephew. 01279 V1 10/15