MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015

Similar documents
ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program

ACR Accreditation Update in MRI

ACR MRI Accreditation: Medical Physicist Role in the Application Process

MRI FAQs. After reading these documents and checking your protocols, you can apply online here:

Nuts and Bolts of MRI Physics Tests or Selling Preparation H. Robert A. Bell, PhD. Learning Objectives

ACR Breast MRI Accreditation Program - DRAFT

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

AAPM Annual Meeting. ACR Accreditation Update in CT

ACR Accreditation Toolkit for Validation Site Surveys

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

Hands-On Ultrasound Physics and Quality Control Workshop. Outline

CT Quality Control Manual FAQs

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

Herlev radiation oncology team explains what MRI can bring

Update on MQSA and Mammography Accreditation

Imaging Rotation. University of Michigan Department of Radiation Oncology Division of Radiation Physics. Resident:

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

MRI to fit your planning. Philips Panorama HFO Oncology Configuration

CT Low Dose Lung Cancer Screening. Part I. Journey to LDCT LCS Program

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

Preparing for Medical Physics Components of the ABR Core Examination

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA

PET/MR. Are You Ready?? Derek Lee, BS, CNMT, PET

5/28/2015. The need for MRI in radiotherapy. Multiparametric MRI reflects a more complete picture of the tumor biology

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

History of MQSA ACR creates testing procedures Oct Mammography Quality Standards Act Dec 1993 Interim Regulations Oct All mammog

RECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE

8/2/2018. Working on MRI Simulator for MP who are NOT trained in MRI. Background What does trained in MRI mean?

The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions (Revised: July 31, 2017)

Mammography Quality Control: A Refresher

Department of Radiology

UCLA UCLA UCLA 7/10/2015. The need for MRI in radiotherapy. Multiparametric MRI reflects a more complete picture of the tumor biology

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

X-Ray & CT Physics / Clinical CT

Outline. QA/QC of Ultrasound Imagers: Basic Physics, Procedures and Experiences. Frame Rate Limitation. US Imaging Range Equation.

Ultrasound Accreditation and Regulatory Compliance from A to Z

Debbie Childs RDMS, RVT Sonographer Murphy Medical Center Murphy, NC

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

Steven Tate Nicholas

MQSA Physicist Qualifications. Jon J. Erickson, Ph.D., DABR

3/22/2017. MR Protocol Review Clinical Opportunities for Physicists. AAPM Newsletter March/April Technical to Clinical Transition - How?

Tissue-engineered medical products Evaluation of anisotropic structure of articular cartilage using DT (Diffusion Tensor)-MR Imaging

MR Advance Techniques. Vascular Imaging. Class II

icat Precise Policy and Procedure Manual

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

Steven LaFontaine Position: Certification: State of California Approved Physicist List: ACR Accreditation Programs Qualified For: Education:

Liver Fat Quantification

354 Korean J Radiol 9(4), August 2008

Radiologic Imaging Magnetic Resonance Imaging (MRI)

Image Fusion, Contouring, and Margins in SRS

Sasa Mutic a) Department of Radiation Oncology, Washington University School of Medicine, St. Louis, Missouri 63110

Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS

A quality control program for MR-guided focused ultrasound ablation therapy

PINPOINTING RADIATION THERAPY WITH THE PRECISION OF MR.

Acknowledgements. QA Concerns in MR Brachytherapy. Learning Objectives. Traditional T&O. Traditional Summary. Dose calculation 3/7/2015

BioMatrix Tuners: CoilShim

CT Imaging at the Point-of-Care

Unrivaled, End-to-End

Prepublication Requirements

ph fax

Medical Diagnostic Imaging

Q-Sense fmri in the MRI environment

Quality assurance in external radiotherapy

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

Current Ultrasound Quality Control Recommendations and Techniques

EORTC Member Facility Questionnaire

Overview of TG225 - Medical Physics Practice Guideline #1

The American College of Radiology Digital Mammography QC Manual: Frequently Asked Questions (Revised 12/12/2018; new and updated items in red)

UK Biobank. Imaging modality Cardiovascular Magnetic Resonance (CMR) Version th Oct 2015

Expanding therapy options for

Please follow all instructions carefully.

Process Improvement: High Quality Breast MRI

Cochlear Implants. Medical Procedures. for MED EL Implant Systems. AW33290_2.0 (English US)

Typical PET Image. Elevated uptake of FDG (related to metabolism) Lung cancer example: But where exactly is it located?

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

Routine Quality Assurance Cookbook

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRAINING PHANTOMS

Visualization strategies for major white matter tracts identified by diffusion tensor imaging for intraoperative use

8/2/2017. Acknowledgement. Disclaimer. How to manage radiotherapy patients with CIED from initial consult to treatment: TG203 recommendations

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

Magnetic Resonance Angiography

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)

FieldStrength. Achieva 3.0T enables cutting-edge applications, best-in-class MSK images

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

CARDIAC MRI. Cardiovascular Disease. Cardiovascular Disease. Cardiovascular Disease. Overview

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

Fetal Dose Calculations and Impact on Patient Care

Q. Where can I find out if my state currently requires stereotactic breast biopsy accreditation?

Nuclear Medicine and PET. D. J. McMahon rev cewood

Full ultrasound breast volumes. Faster scans. Streamlined workflow. ACUSON S2000 Automated Breast Volume Scanner. Answers for life.

Multicenter Trial, Phase I Assessment of 2-D FFDM Versus Combo of 2-D and 3-D Tomosynthesis in Breast Cancer Screening

Quality Control Program Of Real Time Medical Ultrasound Machines In Sudan

KNEE ALIGNMENT SYSTEM (KAS) MRI Protocol

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)

Clinical Applications

Transcription:

MR QA/QC for MRgRT Rick Layman, PhD, DABR Department of Radiology July 13, 2015 The Ohio State University Comprehensive Cancer Center Arthur G. James Cancer Hospital and Richard J. Solove Research Institute

Quality Assurance and Control Quality begins with proper equipment selection W.R. Hendee The Selection and Performance of Radiologic Equipment, Baltimore, MD: Williams and Wilkins, 1985, p. 460. 2

Quality Assurance and Control Quality begins with proper equipment selection W.R. Hendee The Selection and Performance of Radiologic Equipment, Baltimore, MD: Williams and Wilkins, 1985, p. 460. Consult with Diagnostic Radiologist and Physicist Work collaboratively and develop synergies 3

Agenda MR Siting Considerations Imaging Quality Control MR Simulation Specific QC 4

Agenda MR Siting Considerations Imaging Quality Control MR Simulation Specific QC 5

Siting Considerations New Comprehensive Cancer Center December 2014 Opening 275 private in-patient beds Radiation Oncology located on 2 nd floor 7 LINACS with OBI and CBCT CT Simulator PET/CT Simulator Brachytherapy Suite MR Simulator 6

7 Siting Considerations

8 Siting Considerations

MRI Zones Zone 1: free accessible Zone 2: interface Zone 3: restricted area Zone 4: MR magnet room MR Personnel Non-MR personnel: patients, visitors, staff Level 1: passed minimal safety and education training, Zone 3-4 Level 2: extensively trained, gatekeeper of Zone 4 9

Siting Requirements Interfere with Magnet Source of Interference 1.5T X/Y and Z Axis 3.0T X/Y and Z axis Steel reinforcement 4-2 4-2 Water cooling unit, chiller 13-1 13-1 Transport devices up to 440 lbs 17-5 / 21-40 19-8 / 22-11 Vehicles up to 2,000 lbs 18-5 / 24-8 21-3 / 26-2 Elevators, trucks up to 10,000 lbs 20-5 / 29-7 22-11 / 31-2 AC transformers less than 100 KVA 39-5 / 26-2 39-4 / 26-2 AC cables, motors less than 100 AMPS 9-10 / 6-6 9-10 / 6-6 Note: Example specifications applicable to Siemens Skyra (3T) and Aera (1.5T), Siemens Healthcare, Erlangen, Germany 10

Siting Considerations (Magnetic Fringe Fields) Interfere with Object Devices Field Strength 1.5T X/Y and Z axis 3.0T X/Y and Z axis 1.5T Small motors, watches, cameras, credit card 3.0 mt 6-1 / 9-2 6-11 / 10-6 Computers, magnetic disk, processors 1.0 mt 7-3 / 11-6 7-7 / 13-2 Cardiac pacemakers, x-ray tubes, insulin pumps 0.5 mt 8-3 / 13-2 8-7 / 15-2 Color monitors, CT scanner 0.15 mt 9-9 / 16-1 11-2 / 20-1 LINAC 0.1 mt 10-4 / 17-1 12-6 / 22-4 X-ray image intensifier, gamma camera, PET/cyclotron 0.05 mt 13-1 / 22-3 16 1 / 26-11 Note: Example specifications applicable to Siemens Skyra (3T) and Aera (1.5T), Siemens Healthcare, Erlangen, Germany 11

Siting Considerations (Magnetic Fringe Fields) Interfere with Object Devices Field Strength 1.5T X/Y and Z axis 3.0T X/Y and Z axis 1.5T Small motors, watches, cameras, credit card 3.0 mt 6-1 / 9-2 6-11 / 10-6 Computers, magnetic disk, processors 1.0 mt 7-3 / 11-6 7-7 / 13-2 Cardiac pacemakers, x-ray tubes, insulin pumps 0.5 mt 8-3 / 13-2 8-7 / 15-2 Color monitors, CT scanner 0.15 mt 9-9 / 16-1 11-2 / 20-1 LINAC 0.1 mt 10-4 / 17-1 12-6 / 22-4 X-ray image intensifier, gamma camera, PET/cyclotron 0.05 mt 13-1 / 22-3 16 1 / 26-11 Note: Example specifications applicable to Siemens Skyra (3T) and Aera (1.5T), Siemens Healthcare, Erlangen, Germany 12

13 MR Artifact

14 MR Artifact

QC Example - CF Change Over Time 140.00 Central Frequency (ppm) 120.00 100.00 80.00 60.00 40.00 20.00 0.00 5/28/2005 10/10/2006 2/22/2008 7/6/2009 11/18/2010 4/1/2012 8/14/2013 12/27/2014

QC Example - CF Change Over Time 140.00 Central Frequency (ppm) 120.00 100.00 80.00 60.00 40.00 20.00 HDX Upgrade? Cold Head Service 0.00 5/28/2005 10/10/2006 2/22/2008 7/6/2009 11/18/2010 4/1/2012 8/14/2013 12/27/2014

Agenda MR Siting Considerations Imaging Quality Control MR Simulation Specific QC 17

Updates and Changes to ACR MRI QC Manual Interslice RF interference removed Magnetic homogeneity and percent image uniformity procedure changes 87.5% for systems up to 1.5T 82% for 3T systems Signal ghosting added 2.5% Low-contrast detection 9 rows total for systems up to 1.5T 37 rows for 3T systems Assessment of MR safety program 18

ACR MRI Phantom Filled with nickel chloride and sodium chloride solution (10 mm NiCl 2 and 75 mm NaCl) Length 148 mm Diameter 190 mm J.M. Specialty Parts Inc ACR MRI Phantom Must be capable of providing tests substantially equivalent to the ACR phantoms and after they have been approved by a QMP or MR scientist. 19

QMP Responsibilities Commission testing Picture goes here 20

QMP Responsibilities Picture goes here Commission testing Annual testing ACR, 12 months not to exceed 14 months JC, 12 months not to exceed 13 months Repeat appropriate testing after major repair or upgrade 21

QMP Responsibilities Picture goes here Commission testing Annual testing ACR, 12 months not to exceed 14 months JC, 12 months not to exceed 13 months Repeat appropriate testing after major repair or upgrade Establish Quality Control program Baseline measurements and action limits Establishment of new baseline values if needed after major repair Review QC records at least annually 22

QMP Responsibilities Picture goes here Commission testing Annual testing ACR, 12 months not to exceed 14 months JC, 12 months not to exceed 13 months Repeat appropriate testing after major repair or upgrade Establish Quality Control program Baseline measurements and action limits Establishment of new baseline values if needed after major repair Review QC records at least annually Assessment of MR safety program 23

QMP Annual Tests 24 Setup and table position accuracy Center frequency Transmitter gain or attenuation Geometric accuracy High-contrast spatial resolution Low-contrast detectability Artifact evaluation Film printer QC (if applicable) Visual checklist Magnetic field homogeneity Slice position and thickness accuracy Performance testing for coils used clinically SNR Percent image uniformity Percent signal ghosting Soft-copy (monitor) QC

MRI QC Technologist s Responsibilities Picture goes here Setup and table position accuracy Center frequency Transmitter gain or attenuation Geometric accuracy measurements High-contrast spatial resolution Low-contrast detectability Artifact evaluation Film printer QC (if applicable) Visual checklist All Performed Weekly 25

Agenda MR Siting Considerations Imaging Quality Control MR Simulation Specific QC 26

MR Simulation Specific Tests Daily Gantry lasers with center of image plane 27

MR Simulation Specific Tests Daily Gantry lasers with center of image plane Weekly Gantry lasers with respect to imaging plane Lateral wall lasers with respect to gantry lasers and scan plane Wall lasers with respect to the imaging plane Ceiling laser with respect to the imaging plane Orientation of MR tabletop with respect to imaging plane Table vertical and longitudinal motion RT tolerance is typically ±1mm while most manufacturer s specify ±2mm Table indexing and position Scan localization 28

29 MR Simulator Acceptance Testing Results

MR Distortion Analysis Commercially available head phantom (body phantom prototype) Spatial resolution of 0.1 mm Acquisition Technique 3D OEM gradient distortion correction is turned ON 3D T1 weighted sequence with 1 mm 3 isotropic voxels (GE-FLASH, Philips-Fast Field Echo, Siemens-VIBE) 2 NEX TE ~ 4 ms TR ~ 9 ms Flip angle ~ 10 o Pixel BW ~ 120 Hz Percent sampling 100% Percent phase FOV 100% 30

MR Head Distortion Phantom Testing Uncorrected Coronal Sagittal Corrected Dimension 14x13x11 cm 3 Spatial resolution 0.1 mm Images courtesy of Lanchun Lu, PhD Coronal Sagittal 31

MR Body Distortion Phantom Testing Uncorrected Coronal Corrected Coronal Sagittal Sagittal 37 cm diameter 32 cm length 1562 mineral oil points Spatial resolution <1 mm 3 isotropic voxel Images courtesy of Lanchun Lu, PhD 32

Future Considerations New phantom design with improved tolerances Deformable co-registration Multi-modal fusion and co-registration 33

Which of the following items is least likely to impact the performance of the MR system? A. C-arm in an adjacent room B. External lasers C. Forklift operated in area behind the MR D. Power injector E. Patient monitoring device within the MR room 46% 19% 19% 10% 7% A. B. C. D. E.

Which of the following items is least likely to impact the performance of the MR system? ANSWER: A C-arm in an adjacent room The performance of the c-arm could be impacted by the MRI but the distance of the c-arm would be too great to cause any interference with the MR system Reference: All OEM specification cut sheets. Example, https://ftp.siemensmedical.com/pp-cutsheets/mr/skyra/cutsheet-10024.pdf

Based on the current technologies available for evaluation and equipment performance, what is the achievable MR laser accuracy? 40% A. 4.0 mm B. 3.0 mm C. 2.0 mm D. 1.0 mm E. <1.0 mm 3% 23% 22% 12% A. B. C. D. E.

Based on the current technologies available for evaluation and equipment performance, what is the achievable MR laser accuracy? Answer: C 2.0 mm The manufacturer tolerance is 2.0 mm accuracy. Reference: OEM equipment manuals, ECRI, MD Buyline

Additional References ACR website (www.acr.org) MRI Accreditation Program Requirements Breast MRI Accreditation Program Requirements Phantom Test Guidance for the ACR MRI Accreditation Program ACR-AAPM Technical Standard for Diagnostic Medical Physics Performance Monitoring of Magnetic Resonance Imaging (MRI) Equipment AAPM Report No 100: Acceptance Testing and Quality Assurance Procedures for Magnetic Resonance Imaging Facilities Kanal E, Barkovich AJ, Bell C, et al. ACR guidance document on MR safe practices: 2013. Journal of Magnetic Resonance Imaging. 2013; 37(3): 501-530. Gilk T, Kanal E. Interrelating sentinel event alert #38 with the ACR guidance document on MR safe practices: 2013. An MRI accreditation safety review tool. Journal of Magnetic Resonance Imaging. 2013; 37(3):531-543. 38