Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS

Similar documents
Modified QuEChERS for HILIC LC/MS/MS Analysis of Nicotine and Its Metabolites in Fish

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods

Application Note. Author. Abstract. Introduction. Food Safety

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Application Note. Authors. Abstract. Food

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Mycotoxin Analysis in Infant Formula Using Captiva EMR Lipid Cleanup and LC/MS/MS

Mycotoxin Analysis in Peanut Butter Using Captiva EMR Lipid Cleanup and LC/MS/MS

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

GC-µECD Analysis and Confirmation of Contract Laboratory Protocol Pesticides in Olive Oil

Determination of 44 Pesticides in Foodstuffs by LC/MS/MS Application

Determination of Sulfonamide Antibiotics in Bovine Liver Using Agilent Bond Elut QuEChERS EN Kits by LC/MS/MS

Determination of Amphetamine and Derivatives in Urine

Validation Report 23 B

Determination of Aflatoxins in Food by LC/MS/MS. Application. Authors. Abstract. Experimental. Introduction. Food Safety

Analysis of Limonin in Citrus Juice Using QuEChERS and LC-MS/MS

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Analysis of mycotoxins in food matrices using the Agilent Ultivo Triple Quadrupole LC/MS

Author. Introduction. Small Molecule Pharmaceuticals & Generics

Phenomenex roq QuEChERS Kits. Overview. Spinach Facts

Development and Validation of Multiresidue Pesticide Methods at FDA/CFSAN

Multiclass Mycotoxin Analysis in Cheese Using Agilent Captiva EMR Lipid Cleanup and LC/MS/MS

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Application Note. Abstract. Authors. Pharmaceutical

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Quantitative Analysis of Amphetamine-Type Drugs by Extractive Benzoylation and LC/MS/MS. Application. Introduction. Authors. Abstract.

Author. Introduction. Abstract

Validation Report 8. EURL for Cereals and Feeding stuff National Food Institute Technical University of Denmark

Cannabinoid Quantitation Using an Agilent 6430 LC/MS/MS

Determination of Benzodiazepines in Urine by CE-MS/MS

Validation Report 20

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Determination of Amantadine Residues in Chicken by LCMS-8040

Rapid Screening and Quantitation of Pesticide Residues in Cannabis by Modified QuEChERS and LC-MS-MS

Authors. Abstract. Introduction. Environmental

Automated Sample Preparation for FAME Analysis in Edible Oils Using an Agilent 7696A Sample Prep WorkBench

Determination of Patulin in Apple Juice Using SPE and UHPLC-MS/MS Analysis

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

using the Agilent 7696A Sample Prep

Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System

Dienes Derivatization MaxSpec Kit

Determination of Chlorophenoxyacetic Acid and Other Acidic Herbicides Using a QuEChERS Sample Preparation Approach and LC-MS/MS Analysis

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Rapid and Robust Detection of THC and Its Metabolites in Blood

Analysis of HMF by HPLC

Determination of Eight Estrogens in Milk by UHPLC and the Agilent 6495 Triple Quadrupole Mass Spectrometer

Determination of Gamma-Hydroxy-Butyrate (GHB) in Biological Samples

Pesticide Residue Analysis in Whole Milk by QuEChERS and LC-MS/MS

A FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Agilent 6470 Triple Quadrupole LC/MS System and EMR-Lipid. The Combination for Confident Pesticide Quantitation. Christoph Mueller

Benefits of EMR Lipid Cleanup with Enhanced Post Treatment on Pesticides Analysis by GC/MS/MS

Neosolaniol. [Methods listed in the Feed Analysis Standards]

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

Authors. Abstract. Forensic Toxicology. Irina Dioumaeva, John M. Hughes Agilent Technologies, Inc.

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Identification of Imidacloprid Metabolites in Onions Using High Resolution Accurate Mass Spectrometry (LC/Q-TOF MS) and Accurate Mass Tools

CHROMATOGRAPHY. Think about food Care about safety. Vol. 4 Acrylamide.

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Agilent Sample Preparation The Pesticide Analysis Workflow

SPE and QuEChERS Method Development. Tina Chambers Christophe Deckers lent.com

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

Vitamin D3 and related compounds by ESI and APCI

QuEchERS Solid Phase Extraction

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Supporting Information

Improved Isolation and Analysis of Mycotoxins from Cereals, Beer and Wine

LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ

MALAYSIAN MALAYSIAN COCOA WORKSHOP ON THE SAFE USE OF PESTICIDES IN COCOA AND HARMONIZED

Direct Analysis of Folic Acid in Digestive Juices by LC/TOF-MS Application

Analyze Biogenic Amines in Salmon and Shrimp by Capillary Electrophoresis-Tandem MS

Rapid Analysis of Bisphenols A, B, and E in Baby Food and Infant Formula Using ACQUITY UPLC with the Xevo TQD

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

Analysis of Polar Pesticides Using a Mixed-Phase Mode HPLC Column

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Quantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Application of the QuEChERS extraction method for the analysis of pyrethrin and pyrethroid pesticides in fin and non-fin fish.

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

Agilent 6490 Triple Quadrupole LC/MS System with ifunnel Technology. Ultra sensitive quantitative performance

Determination of Bath Salts (Pyrovalerone Analogs) in Biological Samples

Modified QuEChERS Procedure for Analysis of Bisphenol A in Canned Food Products

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

Organophosphorus Residues in Olive Oil by GC/FPD with Agilent J&W DB-35ms Ultra Inert

Determination of Multi-Residue Tetracyclines and their Metabolites in Milk by High Performance Liquid Chromatography - Tandem Mass Spectrometry

Efficient Quantitative Analysis of THC and Metabolites in Human Plasma Using Agilent Captiva EMR Lipid and LC-MS/MS

LCMS-8030 Application Report. Steroids: Testosterone, Progesterone, Estradiol, Ethinylestradiol

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Integration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation

Transcription:

Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS Food Application Author Fadwa Al-Taher Institute for Food Safety and Health Illinois Institute of Technology Bedford Park, IL Abstract This application note describes the use of a quick, easy, cheap, effective, rugged, and safe (QuEChERS) sample preparation procedure. The QuEChERS method employs an extraction and cleanup protocol for the determination of acrylamide with 13 C 3 -acrylamide as the internal standard. The analyte and internal standard were separated on a reversed phase C-18 column (2.1 mm 150 mm, 3 µm) by isocratic elution and analyzed by a triple quadrupole mass spectrometer operating in focused positive electrospray ionization mode. The acrylamide recoveries ranged from 97% to 116% with a relative standard deviation of less than 5%. Introduction In 2002, researchers at the Swedish National Food Administration and Stockholm University reported finding acrylamide at levels up to 3 mg/kg in a wide range of potato and cereal-based products such as potato chips, french fries, roasted and baked potatoes, bread, breakfast cereals, and biscuits. Acrylamide forms in carbohydrate-rich foods that are subjected to high temperatures (>120 C) such as frying, baking and extrusion (1). It is recognized as a neurotoxin and a carcinogen in animals, and a possible carcinogen in humans by the International Agency for Research on Cancer (IARC) (2). High levels of acrylamide are found primarily in potato products, bakery and cereal products, and coffee. Acrylamide is generated from certain food products during heat treatment as a result of the Maillard reaction between certain amino acids and reducing sugars (1). Most laboratories use liquid chromatography/triple quadrupole mass spectrometry (LC/MS/MS) for the analysis of underivatized acrylamide in food samples. Liquid chromatography methods use water to extract acrylamide. A defatting step with hexane, toluene or cyclohexane is

used before or combined with the extraction step, and the aqueous phase is centrifuged (3). The cleanup steps often involve solid-phase extraction (SPE) such as that developed by the U.S. FDA (4). The QuEChERS method, which was originally designed for the analysis of pesticides in food was modified by Mastovska and Lehotay for the extraction of acrylamide (5). The method involves two major steps: extraction and dispersive SPE cleanup. Water extracts the sample, hexane defats the sample, and the MgSO 4 and NaCl induce separation of the water and acetonitrile layers, forcing the majority of acrylamide into the acetonitrile layer. An aliquot of the organic layer is cleaned up by dispersive solid-phase extraction (dspe) by using primary secondary amine (PSA) to remove organic acids and anhydrous MgSO 4 to reduce any water left in the extract. The final extract is then analyzed by liquid chromatography/ triple quadrupole mass spectrometry operating in positive ion multiple reaction monitoring (MRM) mode. This application note presents a method for the analysis of acrylamide in french fries with LC/MS/MS. The method includes sample preparation with Agilent Bond Elut QuEChERS Extraction kit for acrylamides (p/n 5982-5850) and Agilent Bond Elut AOAC Dispersive SPE 2 ml kit (p/n 5982-5022). Experimental Reagents and Chemicals All reagents were Optima or LC/MS grade. Acetonitrile, n- hexane, formic acid and water were purchased from Fisher Scientific (Hanover Park, IL, USA). Acrylamide (Figure 1) and 13 C 3 -acrylamide (internal standard) were purchased from Cambridge Isotope Laboratories (Andover, MA, USA). Acrylamide O NH 2 Standard Solutions The acrylamide standard stock solution (1 mg/ml) was prepared by dissolving 100 mg of the acrylamide in 100 ml acetonitrile and stored at 4 C. The internal standard (methacrylamide) stock solution (100, 000 µg/ml) was made up by pipetting 0.5 ml of the 1 mg/ml standard into 50 ml acetonitrile and stored at 4 C. All working solutions were prepared daily by serial dilution in acetonitrile. Equipment and Material The analysis was performed on an Agilent 1200 Series HPLC system equipped with an Agilent 6460 Triple Quadrupole LC/MS system (Agilent Technologies, Inc., CA, USA). Separation of the chemicals was performed on a reversed phase C18 column (2.1 mm 150 mm, 3 µm). The data was processed by Agilent MassHunter software. Extraction and cleanup were achieved with an Agilent Bond Elut QuEChERS Extraction kit for acrylamides (p/n 5982-5850) and an Agilent Bond Elut QuEChERS AOAC Dispersive SPE kit (p/n 5982-5022), Agilent Technologies (CA, USA). Instrument Conditions: Table 1. LC/MS/MS Conditions Column Reversed C-18 column, 2.1 mm 150 mm, 3 µm Column temperature 30 C Isocratic mode (%B) 2.5% methanol/97.5% of 0.1% formic acid Flow rate 0.2 ml/min Injection volume 10 µl Run time 7 min Post run time 3 min Mass Spectrometer Positive electrospray ionization mode with jet stream technology Capillary voltage 4000 Volts Nozzle voltage 500 V Sheath gas temperature 325 C at 5 L/min Drying gas temperature 350 C at 11 L/min Figure 1. Chemical structure of acrylamide. Sample preparation Frozen par-fried potato slices were purchased from a local store. 2

Extraction One gram of french fries was weighed into a 50 ml centrifuge tube from the Agilent Bond Elut QuEChERS Extraction kit. The internal standard ( 13 C 3 -acrylamide) was added at 500 ng/g. Hexane (5 ml) was added and the tube was vortexed. Water (10 ml) and acetonitrile (10 ml) were added followed by the Agilent Bond Elut QuEChERS extraction salt mixture for acrylamides (p/n 5982-5850). The sample tubes were shaken for 1 min vigorously, and centrifuged at 5000 rpm for 5 min. Dispersive SPE cleanup The hexane layer was discarded and 1 ml of the acetonitrile extract was transferred to a 2 ml Bond Elut QuEChERS AOAC Dispersive SPE tube (p/n 5982-5022). This tube contained 50 mg of PSA and 150 mg anhydrous MgSO 4. The tubes were vortexed for 30 sec and then centrifuged at 5000 rpm for 1 min. The supernatant was then placed in an autosampler vial for LC/MS/MS analysis. Results and Discussion Chromatographic analysis The separation of acrylamide and methacrylamide was obtained on a reversed phase C-18 column (2.1 x 150 mm, 3 µm) where an isocratic elution was used with 2.5% methanol/97.5% of 0.1% formic acid. The column temperature was set at 30 C, while the flow rate was maintained at 0.2 ml/min. Figure 3 shows a typical standard chromatogram of acrylamide and methacrylamide. 10 4 1.5 MRM transitions: TIC 12C-AA 72.0 55.2 1 12C-AA 72.0 26.7 Weigh 1g french fries in a 50 ml centrifuge tube Spike sample with 500 µl of 1000 ng/ml 13 C 3 -acrylamide 0.5 13C-AA 75.0 58.2 13C-AA 75.0 29.0 Add 5 ml hexane, vortex Add 10 ml water, 10 ml ACN, Bond Elut QuEChERS salt packet Shake vigorously for 1 min Figure 3. 1 2 3 4 5 6 Response vs. Acquisition Time (min) Chromatogram of the 10 ng/ml acrylamide standard and 500 ng/ml internal standard 13 C 3 -acrylamide. Centrifuge for 5 min at 5000 rpm Discard the hexane layer Transfer 1mL upper ACN layer to 2 ml microcentrifuge vial packed with 50 mg PSA and 150 mg MgSO 4 Vortex for 30 sec Centrifuge for 1 min at 5000 rpm Transfer 500 µl extract to an autosampler vial QuEChERS method Hexane was used to defat the extract. Addition of water was necessary to facilitate extraction of acrylamide from french fries. An Agilent Bond Elut QuEChERS salt packet (p/n 5982-5850) containing 4 g MgSO 4 and 0.5 g NaCl was used to extract acrylamide from a 1 g sample of french fries. Addition of salt allowed for the separation of the acetonitrile and water phase. Dispersive SPE was employed for sample cleanup (5). The QuEChERS protocol in this application note is simple. The long and tedious SPE process, which may cause low recoveries of acrylamide, is not necessary. Analyze by LC/MS/MS Figure 2. Flow chart for the QuEChERS sample preparation procedure. 3

Figure 4. Acrylamide calibration curve. Linearity A linear calibration curve was obtained by plotting the relative responses of analyte (peak area of analyte (acrylamide)/peak area of IS (methacrylamide) to the relative concentration of analyte (acrylamide)/concentration of IS ( 13 C 3 -acrylamide). The curve was generated by spiking acetonitrile with acrylamide at a concentration range of 0 1000 ng/ml (Figure 4). Excellent linearity was observed (r 2 >0.9999). Recovery and Reproducibility Acrylamide was spiked into 1:1 acetonitrile:water at two different fortification levels (50 and 100 ng/ml). The analysis was performed in triplicates at each level (n=3). Frozen parfried french fry samples containing 33 ng/ml acrylamide were also spiked at two different fortification levels (100 and 200 ng/ml) and blank corrections were made, since it was difficult to find blank french fry samples. A chromatogram of the spiked french fry sample is shown in Figure 5. The analysis was performed in triplicates at each level (n=3). Table 2 shows the recoveries and RSD values for acrylamide. 10 2 8 7 6 5 4 3 2 1 Figure 5. MRM transitions: TIC 12C-AA 72.0 55.2 12C-AA 72.0 26.7 13C-AA 75.0 58.2 13C-AA 75.0 29.0 1 2 3 4 5 6 Chromatogram of the spiked french fry extract. 4

Table 2. Recoveries and RSDs for Acrylamide in Spiked French Fry Samples and 1:1 Water:Acetonitrile (n=3) Matrix Concentration of acrylamide spike (ng/ml) %Recovery (n=3) %RSD (n=3) 1:1 ACN-water 50 116.6 4.07 1:1 ACN-water 100 114.06 4.85 French fries 100 97.14 (after blank correction) 5.04 French fries 200 97.50 (after blank correction) 2.55 Conclusion A simple and fast multiresidue method based on Agilent Bond Elut QuEChERS for acrylamide extraction and cleanup with LC/MS/MS analysis has been developed. High extraction yields were obtained with excellent RSD. Therefore, this method can be applied to monitor acrylamide in food samples. References 1. Al-Taher, F. 2009. Effect of heat-processed foods on acrylamide formation, In : Intentional and Unintentional Contaminants in Food and Feed. F. Al-Taher, L. Jackson, and J.W. DeVries, eds. Oxford University Press, New York, NY, pp.91-113. 2. IARC. Acrylamide. Some Industrial Chemicals: Monographs on the Evaluation of Carcinogenic Risks to Humans; International Agency for Research on Cancer: Lyon, France, 1994. 3. Wenzl, T.; Calle, B. Anklam, E. Analytical methods for the determination of acrylamide in food products: a review. Food Addit. Cont. 2003, 20, 885-902. 4. Roach, J.A.G.; Andrzewjewski, D.; Gay, M.L.; Nortrup, D.; Musser, S.M. Rugged LC/MS/MS survey analysis for acrylamide in foods, J. Agric. Food Chem. 2003. 51, 7547-7554. 5. Mastovska, K. and Lehotay, S.T. Rapid sample preparation method for LC/MS/MS or GC-MS analysis of acrylamide in various food matrices, J. Agric. Food Chem. 2006. 54, 7001-7008. 5

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2012 Printed in the USA January 6, 2012 5990-5940EN