Analysis of mycotoxins in food matrices using the Agilent Ultivo Triple Quadrupole LC/MS

Similar documents
Mycotoxin Analysis in Peanut Butter Using Captiva EMR Lipid Cleanup and LC/MS/MS

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Mycotoxin Analysis in Infant Formula Using Captiva EMR Lipid Cleanup and LC/MS/MS

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods

Modified QuEChERS for HILIC LC/MS/MS Analysis of Nicotine and Its Metabolites in Fish

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis

Multiclass Mycotoxin Analysis in Cheese Using Agilent Captiva EMR Lipid Cleanup and LC/MS/MS

Ultivo Applications Terri Sosienski LC/MS Applications Scientist. June 14, 2017

Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

Application Note. Author. Abstract. Introduction. Food Safety

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

Analysis of Vitamins Using an SFC/UHPLC Hybrid System with a Triple Quadrupole LC/MS for Quantification

2D-LC as an Automated Desalting Tool for MSD Analysis

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement

Agilent 6470 Triple Quadrupole LC/MS System and EMR-Lipid. The Combination for Confident Pesticide Quantitation. Christoph Mueller

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

Determination of Eight Estrogens in Milk by UHPLC and the Agilent 6495 Triple Quadrupole Mass Spectrometer

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Improved Isolation and Analysis of Mycotoxins from Cereals, Beer and Wine

Quantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine-

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Neosolaniol. [Methods listed in the Feed Analysis Standards]

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Determination of Benzodiazepines in Urine by CE-MS/MS

LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ

Determination of Aflatoxins in Food by LC/MS/MS. Application. Authors. Abstract. Experimental. Introduction. Food Safety

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

Extended Mass Range Triple Quadrupole for Routine Analysis of High Mass-to-charge Peptide Ions

Efficient Quantitative Analysis of THC and Metabolites in Human Plasma Using Agilent Captiva EMR Lipid and LC-MS/MS

Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC

Accurate quantification of regulated mycotoxins by UHPLC-MS/MS and screening for 200+ mycotoxins in food and feed.

A FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ

Application Note. Authors. Abstract. Food

Determination of Amantadine Residues in Chicken by LCMS-8040

Determination of Fusarium Mycotoxins in Wheat, Maize and Animal Feed Using an Online TurboFlow and Orbitrap LC/MS Method

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Development and Validation of an UPLC-MS/MS Method for Quantification of Mycotoxins in Tobacco and Smokeless Tobacco Products

Determination of Patulin in Apple Juice Using SPE and UHPLC-MS/MS Analysis

Chiral Screening for SFC and UHPLC with the Agilent 1260 Infinity II SFC/UHPLC Hybrid System

Determination of Amphetamine and Derivatives in Urine

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Protein Precipitation for Biological Fluid Samples Using Agilent Captiva EMR Lipid 96-Well Plates

The New 6495 Triple Quadrupole LC/MS

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Author. Introduction. Small Molecule Pharmaceuticals & Generics

Determination of Sulfonamide Antibiotics in Bovine Liver Using Agilent Bond Elut QuEChERS EN Kits by LC/MS/MS

Improving Coverage of the Plasma Lipidome Using Iterative MS/MS Data Acquisition Combined with Lipid Annotator Software and 6546 LC/Q-TOF

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Determination of Isoflavones in Soybean by LC/MS/MS

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

Performance Characteristics of the Agilent 1290 Infinity II Multicolumn Thermostat

For Client Review Only. All Rights Reserved. Advanstar Communications Inc Scarlett Biselli, Lutz Hartig, Heiner Wegner, and Christian Hummert

Fast quantitative Forensic Analysis of THC and its Metabolites in Biological Samples using Captiva EMR- Lipid and LC/MSMS

Validation Report 23 B

Quantitative Determination of Drugs of Abuse in Human Whole Blood by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Two-Dimensional LC/MS/MS to Reduce Ion Suppression in the Determination of Cannabinoids in Blood Plasma

Authors. Abstract. Introduction. Environmental

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Analysis of Limonin in Citrus Juice Using QuEChERS and LC-MS/MS

Accurate Quantitation of Regulated Mycotoxins by UHPLC/MS/MS

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Determination of Multi-Residue Tetracyclines and their Metabolites in Milk by High Performance Liquid Chromatography - Tandem Mass Spectrometry

Rapid Screening and Quantitation of Pesticide Residues in Cannabis by Modified QuEChERS and LC-MS-MS

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

Analysis of Per- and Polyfluoroalkyl Substances (PFASs) in Biological Fluid Using a Novel Lipid Removing Sorbent and LC-MS/MS

Agilent 6460 LC-QQQ Highly Sensitive and Robust Analysis for Lipophilic Marine Toxins in Shellfish

Agilent 6410 Triple Quadrupole LC/MS. Sensitivity, Reliability, Value

Determination of Copper in Green Olives using ICP-OES

Determination of Chlorophenoxyacetic Acid and Other Acidic Herbicides Using a QuEChERS Sample Preparation Approach and LC-MS/MS Analysis

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography

Author. Introduction. Abstract

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Validation Report 8. EURL for Cereals and Feeding stuff National Food Institute Technical University of Denmark

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Rapid and Robust Detection of THC and Its Metabolites in Blood

Fat-Soluble Vitamins Analysis on an Agilent ZORBAX Eclipse PAH Polymeric C18 Bonded Column

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

Determination of B-vitamins in Energy Drinks by CE/MS/MS

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts

Authors. Abstract. Forensic Toxicology. Irina Dioumaeva, John M. Hughes Agilent Technologies, Inc.

MALAYSIAN MALAYSIAN COCOA WORKSHOP ON THE SAFE USE OF PESTICIDES IN COCOA AND HARMONIZED

Dienes Derivatization MaxSpec Kit

Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

Agilent 6490 Triple Quadrupole LC/MS System with ifunnel Technology. Ultra sensitive quantitative performance

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Direct Analysis of Folic Acid in Digestive Juices by LC/TOF-MS Application

Pesticide Residue Analysis in Whole Milk by QuEChERS and LC-MS/MS

HPLC to UHPLC Transfer of USP Method for Amlodipine Besylate Using the Agilent 1290 Infinity II LC

Transcription:

Application Note Analysis of mycotoxins in food matrices using the Agilent Ultivo Triple Quadrupole LC/MS Figure 1. Agilent Ultivo Integrated into LC stack. Authors Theresa Sosienski 1, Dan Hui Dorothy Yang 1, Mark Sartain 1, Christian Hegmanns, and Joni Stevens 1 Agilent Technologies, Inc. Santa Clara, CA Agilent Technologies, Inc. Waldbronn, Germany Abstract This Application Note demonstrates a sensitive and precise method for analyzing 1 mycotoxins in corn and peanut matrices and five mycotoxins in black pepper matrix on the Agilent Ultivo triple quadrupole mass spectrometer. The Ultivo LC/MS was designed to save laboratory space, while maintaining the performance required for high-throughput analyses. All mycotoxins could be quantified below the maximum levels defined by the European Union Commission Regulations (EC) No. 1881/6 and No. 15/1 in each matrix. Excellent method precision was achieved on the Ultivo system, with relative standard deviations (%RSD) of <1 % at the lowest level of quantitation. The combination of the cleanup of matrix interferences, chromatography, and the newly developed triple quadrupole allows for sensitive and precise detection of mycotoxins.

Introduction Mycotoxins are produced by fungi that can grow on various crops. At certain levels and combinations, mycotoxins can be harmful to humans and livestock through consumption of contaminated crops; therefore, mycotoxin levels are regulated in foods to minimize the risk of ingestion 1. To ensure that harmful levels of mycotoxins do not enter the food supply, regulatory agencies have established maximum levels (MLs), ranging from.1 ppb in some baby foods to ppm levels in some animal feeds. It is important to accurately quantify the mycotoxin contents below the ML across various food matrices, as each matrix composition poses different challenges. The Agilent Ultivo Triple Quadrupole LC/MS is designed to address many of the challenges faced by labs performing environmental and food safety analyses. Innovative technologies within the Ultivo allowed us to achieve a reduced overall footprint, while conserving the performance found in traditional systems (Figure 1). Innovations such as the Cyclone Ion Guide, Vortex Collision Cell, and the Hyperbolic Quads maximize quantitative performance in a small package. These innovations enhance instrument reliability and robustness, resulting in greater uptime. VacShield and easy change of detector assembly reduce the time and expertise required for system maintenance, making it easier for the nonexpert MS user to operate and maintain. The Agilent MassHunter software suite simplifies data acquisition, method setup, data analysis, and reporting. This software leads to faster acquisition-to-reporting time, increasing lab productivity and confidence in results. This Application Note demonstrates the sensitive and precise quantification of up to 1 regulated mycotoxin compounds in three commonly regulated food commodities using the novel Ultivo triple quad LC/MS. Experimental Reagents and chemicals All reagents used in this application were HPLC or LC/MS grade. Acetonitrile was purchased from Honeywell (Morristown, NJ, USA), and ultrapure water was sourced from a Milli-Q Integral system with a LC-Pak Polisher and a. µm point-of-use membrane filter cartridge (EMD Millipore, Billerica, MA, USA). Formic acid (FA) and ammonium formate were purchased from Fluka (Sigma Aldrich Corp., St. Louis, MO, USA), and ammonium fluoride was purchased from Aldrich (Sigma-Aldrich Corp., St. Louis, MO, USA). Chemical standards were purchased from Sigma-Aldrich or Cayman Chemical. Sample preparation Corn flour, peanuts, and ground black pepper were obtained from local grocery stores. Five grams of corn flour, 5 g of ground peanuts (ground in a blender, and frozen until analysis), or g of ground black pepper were weighed into 5 ml polypropylene tubes and extracted using 1 ml of acetonitrile and 1 ml of ultrapure H O with. % FA and shaken using a rotary shaker (Heidolph Hei-MIX Multi reax, 545-1 ) for minutes. Agilent QuEChERS EN Extraction Salts (p/n 598-565) were added to the extract, and the tubes were shaken for a further minutes, then centrifuged for 5 minutes at 4,5 rpm (Sorvall Heraeus, Labofuge 4K). Black pepper extract was subjected to an additional cleanup step using an Agilent universal dispersive SPE kit (p/n 598-9). Then, 1.6 ml of all extracts were diluted with.4 ml of ultrapure water, and cleaned using Agilent Captiva EMR Lipid cartridges (p/n 519-1). Spiked black pepper extracts were diluted :7 extract/ultrapure water before analysis. Figure presents the flowchart of sample preparation. Instrumentation Agilent 19 Infinity II UHPLC Agilent 19 Infinity high speed pump (G71A) Agilent 19 Infinity II multisampler with cooler (G7167B) Agilent 19 Infinity II multicolumn thermostat (G7116B) Agilent Ultivo Triple Quadrupole LC/MS system Agilent Jet Stream electrospray ionization source Method Table 1 summarizes the Agilent 19 Infinity II UHPLC conditions. Table summarizes the Ultivo Triple Quadrupole parameters and Agilent Jet Stream ESI source parameters. Analysis was carried out with positive ionization and dynamic multiple reaction monitoring (dmrm). Data were evaluated using the Agilent MassHunter Quantitative Analysis Software B.9 with the Quant-My-Way feature.

Weigh 5 g (corn or peanuts) or g (black pepper) into a 5-mL centrifuge tube with two porcelain homogenizers. Add 1 ml water (. % FA) and 1 ml acetonitrile. Shake for minutes on a vortex shaker. Add Agilent QuEChERS EN extraction kit salts. Shake for minutes on a vortex shaker. Black pepper only Add 6 ml of EN extract to a dspe Universal kit 15-mL centrifuge tube. Centrifuge at 4,5 rpm for 5 minutes. Shake for minutes on a vortex shaker. Dilute 1.6 ml of extract with.4 ml of ultrapure water. Centrifuge at 4,5 rpm for 5 minutes. Pass through an Agilent Captiva EMR Lipid cartridge under constant vacuum, one drop every 5 seconds. Dilute. ml of extract with.7 ml of ultrapure water. Figure. Sample preparation procedure using Agilent Captiva EMR Lipid for the analysis of mycotoxins in corn, peanuts, and black pepper. Table 1. Agilent 19 Infinity II UHPLC parameters. Parameter Value Column Agilent Eclipse Plus C18. 15 mm, 1.8 µm (p/n 959759-) Column temperature 45 C Injection volume Mobile phase Flow rate Gradient Stop time Post time µl Corn, peanut; 1 µl black pepper A).5 mm Ammonium fluoride + 5 mm ammonium formate +.1 % formic acid in water B).5 mm Ammonium fluoride + 5 mm ammonium formate +.1 % formic acid in methanol.45 ml/min Time %B.5 7.5 1 9. 1 9.1 9.1 minutes 1.9 minutes Table. Agilent Ultivo Triple Quadrupole and Agilent Jet Stream source parameters. Parameter Value Drying gas temperature 5 C Drying gas flow 8 L/min Sheath gas temperature 5 C Sheath gas flow Nebulizer pressure 1 L/min psi Capillary voltage, V (+) Nozzle voltage V (+) Cycle time 5 ms

Results and Discussion Mycotoxin regulatory maximum levels Mycotoxins are regulated at many different concentrations depending on the specific mycotoxin, the matrix being analyzed, the regulatory agency, and the targeted consumer. This study used the European Union regulations No. 1881/6 and No. 15/1 as a benchmark. This procedure was done because the EU generally had the lowest maximum level (ML) for mycotoxins compared to other regional regulatory agencies. The nonregulated compounds were either not included in the matrix, or estimated based on current levels of concern established in Table 4. Table. Transitions for mycotoxin detection in dmrm mode. Compound Precursor (m/z) Product (m/z) RT (min) RT Window (min) Fragmentor (V) CE (V) Polarity Aflatoxin B1 (AB1) 1. 85.1 5.9 1 19 Positive Aflatoxin B1 (AB1) 1. 41.1 5.9 1 19 4 Positive Aflatoxin B (AB) 15.1 87.1 5.7 1 19 4 Positive Aflatoxin B (AB) 15.1 59.1 5.7 1 19 8 Positive Aflatoxin G1 (AG1) 9.1 11.1 5. 1 18 Positive Aflatoxin G1 (AG1) 9.1 4.1 5. 1 18 8 Positive Aflatoxin G (AG) 1.1 1.1 5.1 1 19 4 Positive Aflatoxin G (AG) 1.1 115 5.1 1 19 8 Positive Deoxynivalenol (DON) 97.1 49.. 1 1 4 Positive Deoxynivalenol (DON) 97.1 77. 1 1 8 Positive Fumonisin B1 (FB1) 7.4 5. 6.6 1 4 6 Positive Fumonisin B1 (FB1) 7.4 4. 6.6 1 4 4 Positive Fumonisin B (FB) 76.4 6. 7.4 1 4 6 Positive Fumonisin B (FB) 76.4 18.4 7.4 1 4 4 Positive Fumonisin B (FB) 76.4 6. 7. 1 4 6 Positive Fumonisin B (FB) 76.4 18.4 7. 1 4 6 Positive HT- Toxin (HT-) 44. 6. 6.8 1 1 4 Positive HT- Toxin (HT-) 44. 15.1 6.8 1 1 4 Positive Ochratoxin A (OTA) 44.1 9 7.6 1 14 Positive Ochratoxin A (OTA) 44.1 1 7.6 1 14 6 Positive T- Toxin (T-) 484. 15. 7. 1 14 1 Positive T- Toxin (T-) 484. 185.1 7. 1 14 4 Positive Zearalenone (ZEA) 19. 1. 7.7 1 1 4 Positive Zearalenone (ZEA) 19. 185.1 7.7 1 1 4 Positive Table 4. MLs for mycotoxins in this study. EU reg No. 1881/6 and No. 15/1 are used for reference. All assigned MLs in this study are equal to or lower than the EU ML. Mycotoxin Corn (ppb) European Union ML for mycotoxins, Peanut (ppb) Black pepper (ppb) Assigned ML used for this study Corn and peanut (ppb) Black pepper (ppb) Aflatoxin B1 5 5 Aflatoxin B 5 Aflatoxin G1 Sum of aflatoxins: Sum of aflatoxins: Sum of aflatoxins: 4 ppb 4 ppb 1 ppb 5 Aflatoxin G 5 Ochratoxin A n/a 15 15 Fumonisin B1 Sum of B1 and B: n/a 5 Not included n/a Fumonisin B 1, ppb n/a 5 Not included Fumonisin B n/a n/a n/a 5 Not included Deoxynivalenol 75 n/a n/a 75 Not included Zearalenone 1 n/a n/a 1 Not included T- Toxin n/a n/a n/a 1 Not included HT- Toxin n/a n/a n/a 5 Not included 4

Method recovery The extraction procedure for this method was simple and effective. The Agilent Captiva EMR Lipid Kit was a quick step that adequately separated matrix components from compounds of interest. This method demonstrated good recovery for all compounds in each matrix, with recoveries ranging 6 11 % at ML for each compound (Table 5). Method sensitivity Figure shows the excellent signal response of the 1 mycotoxins analyzed in this method, displaying the chromatogram at ML in corn matrix. Mycotoxin standards at levels ranging from 1/ of the assigned ML to 1 times the assigned ML were evaluated. Eight concentration levels were spiked into matrix extracts for limit of quantitation (LOQ) and %RSD analysis. Enhanced signal response allowed most of the mycotoxins in each matrix to have a quantitation limit at 1/ the assigned ML. Figure 4 shows that all the mycotoxins had LOQs at least 1/5 the assigned ML. LOQs were defined as having four out of six replicate injections with accuracy of 8 1 %, and a signal to-noise ratio (S/N) greater than 1. Table 5. Recovery (%) of each mycotoxin studied in corn, peanut, and black pepper matrix at ML. Fumonisin compounds were evaluated at ½ ML. Mycotoxin Corn Mycotoxin recovery (%) Peanut Black pepper Aflatoxin B1 17 15 9 Aflatoxin B 11 19 97 Aflatoxin G1 19 18 1 Aflatoxin G 11 96 14 Ochratoxin A 8 8 19 Fumonisin B1 6 65 Fumonisin B 67 77 Fumonisin B 9 61 Deoxynivalenol 111 7 Zearalenone 98 9 T- Toxin 15 14 HT- Toxin 18 1 Counts (%) 1 1.1 1. T- 1 1.9 OTA.8.7 1.6.5 FB 7.6 7.7.4. FB1 HT- FB..1 DON ZEA AG AG1 AB AB1.5..5 4. 4.5 5. 5.5 6. 6.5 7. 7.5 Acquisition time (min) Figure. Composite dmrm chromatogram of mycotoxins in corn matrix at assigned ML. No. of mycotoxins 14 1 1 8 6 4 1 1 1 1 1 1 1/5 ML 1/1 ML 1/ ML Corn Peanut Black pepper Matrix Figure 4. LOQ for all mycotoxins studied in each matrix. All could be accurately quantified at a fraction of the assigned ML. 5

Method precision and linearity Excellent precision was achieved for all compounds studied in each matrix. Figure 5 shows the %RSD of six replicate injections of mycotoxins at their LOQ in each matrix. All compounds have a %RSD less than 1 % in each matrix, with most of the compounds having a %RSD less than 5 %. The exceptional precision in this method is highlighted in the close-up view of the six replicate injections of aflatoxin B1 at ppt in corn and peanut matrix, and 5 ppt in black pepper matrix, corresponding to 1x lower than the ML (Figure 6). No. of mycotoxins 14 1 1 8 6 4 1 9 RSD 5 1 % RSD 5 % Corn Peanut Black pepper Matrix Figure 5. The mycotoxins show excellent precision in all three matrices, with all %RSDs <1 % at LOQ. 1 1 1 Corn Peanut 1. RSD % = 5.7 Avg. S/N = 71.9 1. RSD % =.4 Avg. S/N = 169 1. Black Pepper RSD % = 5.65 Avg. S/N = 5.6.9.9.9.8.8.8.7.7.7 Counts (%).6.5.4 Counts (%).6.5.4 Counts (%).6.5.4.......1.1.1 -.1 -.1 -.1 Figure 6. Excellent precision demonstrated for six replicate injections of aflatoxin B1 at 1/1 ML ( ppt or 5 ppt) in all matrices. 6

All compounds demonstrated outstanding linearity. Six to eight calibration levels were analyzed for each compound, ranging from 1/ to 1x the ML, resulting in R values.99 for all compounds in all matrices. Figure 7 shows examples of excellent linearity for six selected compounds in peanut matrix. 1 4 1.6 1.4 1. 1..8.6.4. Aflatoxin B1.1 ppb R =.999 1 4. 1.8 1.6 1.4 1. 1..8.6.4. Deoxynivalenol.75 75 ppb R =.999 1 4 1. 1.1 1..9.8.7.6.5.4...1.1-1 4 6 5 4 1 4 6 8 1 1 14 16 18 Aflatoxin G.1 ppb R =.999 4 6 8 1 1 14 16 18 HT- Toxin 5 5, ppb R =.999 1 4 1.8 1.6 1.4 1. 1..8.6.4. 1 5.5 5. 4.5 4..5..5. 1.5 1..5 1 4 5 6 7 8 Aflatoxin G1.1 ppb R =.999 4 6 8 1 1 14 16 18 Ochratoxin A.15 ppb R =.999 5 1, 1,5,,5,,5 4, 4,5 5, 4 6 8 1 1 14 16 18 4 6 8 Figure 7. Calibration curves for selected compounds in peanut matrix. Linear fit and no weighting used. Arrows indicate ML. 7

Conclusions The Agilent Ultivo Triple Quadrupole LC/MS is an exceptionally innovative mass spectrometer that minimizes laboratory workspace needs. It also reduces maintenance challenges, creating a productive work environment for high throughput laboratories. Ultivo is a small but powerful tool enabling the accurate and sensitive detection of commonly regulated mycotoxins in various food matrices well below established MLs. Agilent MassHunter software provides an easy-to-use, all inclusive tool for acquiring and reporting LC/MS data. References 1. Bennett, J. W.; Klich, M. Mycotoxins. Clinical Microbiology Reviews, 16(), 497-516... Commission Regulation (EC) No 1881/6. Setting maximum levels for certain contaminants in foodstuffs. Official Journal of the European Union, 19 December 6, L 64/5-4.. Commission Regulation (EU) No 15/1. Amending Regulation (EC) No 1881/6 setting maximum levels for certain contaminants in foodstuffs as regards ochratoxin A. Official Journal of the European Union, 5 February 1, L 5/7-8. 4. Determination of Mycotoxins in Peanuts with Enhanced Matrix Removal Lipid by LC/MS/MS. Agilent Technologies Application Note, publication number 5991 781. www.agilent.com/chem This information is subject to change without notice. Agilent Technologies, Inc. 18 Printed in the USA, February 14, 18 5991-896EN