Simple Strategies for GC/MS Success. Rachael Simon Agilent Technologies Product Manager, Gas Chromatography Supplies

Similar documents
Analyze Drugs of Abuse with Agilent J&W Ultimate Plus Tubing in an Inert Flow Path

Analysis of Omega 3 and Omega 6 FAMEs in Fish Oil and Animal Fat Using an Agilent J&W DB-FATWAX Ultra Inert GC Column

MS/MS Library Creation of Q-TOF LC/MS Data for MassHunter PCDL Manager

Gas Chromatography Syringe Selection Guide

Sensitive, Accurate and Precise. MS Solutions Through SGE Products. Diamond MS Syringes evol XR MEPS SilTite FingerTite GC Capillary Columns

Agilent 1260 Infinity Analytical SFC System

Choose for Your Instrument. GC Supplies - Shimadzu Instrument Quick Pick Guide

General Single Ion Calibration. Pete 14-May-09

Restek PAL SPME Fibers cat.#

Thermo Finnigan. 6-2 LCQ Advantage Getting Started-

A comparison study of the analysis of volatile organic acids and fatty acids

Determination of Allergens in Fragrance Products Using Agilent Deconvolution Reporting Software Application Brief

Forensic Analysis of Blood Alcohol Concentration

Premium Chemical Free Iron/Sulphur Filter Installation Manual

Analysis of several common. organic acids in tobacco leaf, snus, and moist snuff

Analysis of short chain organic acids in tobacco and in some flavored e-liquids

Improving the Analysis of 37 Fatty Acid Methyl Esters

Analysis of Organic Acids and Alcohols Using the Agilent J&W DB-624UI Ultra Inert GC Column

The Measure of Excellence. GC Autosampler Syringes. CTC GC PAL Autosampler Syringes (PAL is a registed trademark of CTC Analytics AG, Switzerland)

Shimadzu s Guide to Vial Selection. Not all vials are the same

More accurate results.

HPLC. Varian Dynamax Preparative HPLC Columns

Results of Using Re-washed Vials and Closures

Instructions for Use Enbrel (en-brel) (etanercept) for injection, for subcutaneous use Multiple-dose Vial

PLEASE READ THIS USER MANUAL BEFORE USE

California TOF/QTOF Users Meetings: MS Settings. Jim Lau, Agilent Technologies.

Toxicology Screening of Whole Blood Extracts Using GC/Triple Quadrupole/MS

Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual

QSTAR Operation. May 3, Bob Seward

Independent Column Temperature Control Using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds

Highly Repeatable Ultra Low Detection of Estradiol Using Triple Quadrupole GC/MS in NCI Mode

Lipidomic Analysis by UPLC-QTOF MS

Super Clean TM. Gas Filters BIG TRAP INLINE FILTERS. Ideal for Central Purifying Solutions Largest Big Trap Available High Pressure Stainless Steel

max airless Operating Manual for Dispenser Read this manual thoroughly before operating the Labmax airless!

NovoPen Echo User guide

How to Use ENBREL : Vial Adapter Method

MAKE SMART INJECTION CHOICES INJECTION PROVIDERS GUIDE FOR SAFE INJECTIONS

PLEASE READ THESE INSTRUCTIONS BEFORE USE

Separation of 37 Fatty Acid Methyl Esters Utilizing a High-Efficiency 10 m Capillary GC Column with Optimization in Three Carrier Gases

INTEGRAL-BONNET NEEDLE VALVES H-300 SERIES

PEEK SS Columns & Guard Columns

The Evolution GC-MS/MS Triple Quadrupole System

Savillex Technical Note C-Flow 700d PFA Concentric Nebulizer for ICP-OES

Instructions for Use. HUMALOG KwikPen. insulin lispro injection (rdna origin) 100 units/ml, 3 ml pen

Organophosphorus Pesticides Analysis Using an Agilent J&W DB-5ms Ultra Inert Capillary GC Column. Application Note. Environmental. Abstract.

Savillex Technical Note

GlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System

DO NOT TRANSFER TO A SYRINGE SEVERE OVERDOSE CAN RESULT

CORESTA Recommended Method No. 84

Cap Clip Rubber Seal Plunger Pen Body Dose Window

MEDIUM-FLOW PNEUMOTACH TRANSDUCER

INSTRUCTIONS FOR USE

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Instructions for Use. BASAGLAR KwikPen. insulin glargine injection (100 units/ml, 3 ml pen)

Detection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Ordering Information You may contact us via phone, fax, , or through our website.

Rapidly Analyze a Wide Range of Glycol Ethers by GC-MS Using the New Rxi -1301Sil MS Column

Single Quadrupole Compact Mass Spectrometer for Liquid Sample

Installation and Maintenance Instructions

Adventure 3 Plus Service Manual

Alignment and Calibration of the APEX Instrument. March 30, Chemical Crystallography Laboratory

Savillex Technical Note

Innovations in Large Volume Injection

Elemental Scientific Semiconductor Catalog for Thermo ELEMENT2

Ordering Information You may contact us via phone, fax, , or through our website.

LC-MS/MS Method for the Determination of Tenofovir from Plasma

User s instructions The Flash Series. FL-CIC Completely-in-canal

Agilent 1220 Infinity LC System Performance Specifications

DIRECT EXTRACTION OF BENZODIAZEPINE METABOLITE WITH SUPERCRITICAL FLUID FROM WHOLE BLOOD

GlucCell TM SYSTEM USER S GUIDE ver 2.3 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System

HUMULIN 70/30 KwikPen

INSTRUCTIONS FOR PREPARING AND GIVING AN INJECTION OF ENBREL POWDER

Agilent s LC/MS Portfolio and Applications Examples

Identification. Eartips. Battery Door. 2-Position Switch. Eartip. Removal Cord. Flexible Neck Cord Protective Case. Filter Tool and Extra ACCU Filters

Instructions for Use. For use with. 10 mg vial

The challenging extraction of non-polar contaminants out of a non-polar vegetable oil sample

Rapid and Robust Detection of THC and Its Metabolites in Blood

Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution

For use only with Lilly 3 ml insulin cartridges (100 units/ml). Do not use other brands of insulin cartridges.

BROEN UNIFLEX FITTINGS FOR NON-BURNING 2.0 GASES / VACUUM

Successful Splicing Made Easy

Fizzy Extraction of Volatile and Semivolatile Compounds into the Gas Phase

Dip Station G

Sample Preparation Techniques for Biological Matrices: Finding the right balance to achieve optimal results

For use only with Lilly 3 ml insulin cartridges (100 units/ml). Do not use other brands of insulin cartridges.

Hamilton Syringes. LT N LTN RN TLL TLL with slots. Hamilton Syringe Types Microliter. Hamilton Point Styles

BIOLOGY 1101 LAB 1: OSMOSIS & DIFFUSION. READING: Please read pages & in your text prior to lab.

#20 Separation of Alcohols Using Gas Chromatography

Extraction of 11-nor-9-carboxy-tetrahydrocannabinol from Hydrolyzed Urine by ISOLUTE. SLE+ Prior to GC/MS Analysis

Wear sleeves. General

LEARNING TO USE YOUR FIRST HEARING AID

STANDARD OPERATING PROCEDURE: RIGAKU RU 300 XRD POWDER SAMPLES

Supplementary Information

Application. Authors. Abstract. Introduction. Forensics

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Rapid Analysis of 37 FAMEs with the Agilent 8860 Gas Chromatograph

Pure 312 Nx, Pure 13 Nx

OWNER S MANUAL. BODYCRAFT 7699 Green Meadows Dr., Lewis Center, Ohio Tel:

ProVari Owner s Manual

Transcription:

Simple Strategies for GC/MS Success Rachael Simon Agilent Technologies Product Manager, Gas Chromatography Supplies 1

Time robbers in the GC/MS lab Looking for leaks Frequent leaks affect sensitivity, damage column, shift peak retention times Locating leaks can be tedious Unplanned maintenance Over time, deposits on MS source impair performance Closer inspection of routine tune reports can proactively identify need for source maintenance Sample reruns/reworks Enabling greater sensitivity helps prevent poor peak shape and contamination Page 2

GC/MS Time Robber #1: Leaks Page 3

Where to look: common leak sites Page 4

Where to look: common leak sites Page 4

Where to look: common leak sites Page 4

Where to look: common leak sites Page 4

Hunting for leaks: the analyzer side plate Tip 1: Ensure seal integrity at the side plate o-ring Factors that disturb seal integrity Sticking/dislodging upon opening Aging and wear Physical obstructions: dust, lint, metal fines How is it fixed and prevented? Using gloves, gently manipulate o-ring within groove to ensure uniform surface Visually inspect seal integrity prior to pump down Replace worn or damaged o-rings Wipe o-ring perimeter with lint-free tissue Stop using thumbscrews Image taken with analyzer door removed for better visibility Agilent GCMS Side plate o-ring P/N 0905-1442 link Can create metal fines with excessive use Only use during shipment/relocation Ensure MS is well-ventilated Periodically vacuum beneath outer covers Page 8

Page 9 Hunting for leaks: the analyzer side plate Don t let this happen to you!

Hunting for leaks: the vent valve Tip 2: Forgetting to retighten- or overtightening- the vent valve How does this happen? Easily overlooked Urge to overtighten fittings Why is it bad? Overtightening compresses o-ring Prevents leak-free seal formation How is it fixed? After venting, retighten vent valve before maintenance Tighten gently to avoid cross-threading If leaks are persistent, replace vent valve o-ring Agilent MSD Vent Valve o-ring P/N 0905-1014 link Page 10

Hunting for leaks: transfer line interface Tip 3: Tight, but not too tight Better yet, self-tightening Why is it bad? Over tightening damages transfer line threads, column Audible squealing overtightening Self-tightening column nut installed How is it fixed? Stop overtightening Stripped transfer line threads require replacement by service engineer 400 injections later Just tight enough- use MS tune to check Switch to self-tightening MS column nuts Spring-driven piston continuously presses against ferrule Easily installed, no tools Benefits- longer column lifetime, lower column bleed Available for Agilent, Perkin Elmer, Shimadzu, Thermo For Agilent MS transfer line P/N 5190-5233 Page 11

A critical connection A proper connection between the column and MS is critical to achieving a leak free seal and getting good chromatography. Focus on your ferrules Even experts get distracted. Try to take an extra moment to ensure you grab the right ferrule from others you might have on hand. Double check the diameter Make sure the ferrule diameter accommodates your column s ID. Using a 320um ferrule on a 250um column will result in a leak. Just tight enough! Overtightening damages fittings and your column. Periodically retighten brass column nuts, or switch to self-tightening. The Facts on Ferrules Recycling is great for the environment, but not so good for ferrules- their ability to make a leak free seal declines with age and wear. Never reuse ferrules. To MS Right side up Installing the ferrule the wrong way is a common cause of leaks. The widest part of the ferrule should oriented toward the MS. Short selfies and brass barrels Self-tightening column nuts need short ferrules. Brass column nuts require barrel-shaped ferrules. Use this mnemonic device to help you remember. Graphite doesn t go Pure graphite ferrules are a no-go with GC/MS- it s too soft, permeable, and has a tendency to flake or crumble. Always use Vespel /graphite ferrules for GC/MS applications. Page 12

Hunting for leaks: weapon of choice Tip: Arm yourself with the best or in a pinch, what s nearby A good leak detector is a mass spectrometrist s best friend High sensitivity Recommended for leak detection in gas plumbing and fittings If you don t have a leak detector, use electronics duster instead Hold can upright (don t spray liquid!) Spray short bursts around likely leak suspects Spray just enough Live tune scanning for ions to pinpoint leak Typical Electronic Duster Components and Ions 1,1- difluoroethane 1,1,1- trifluoroethane 1,1,1,2- tetrafluoroethane m/z 51,65 m/z 83 m/z 101 Agilent G338B Leak Detector link Page 13

Hunt for leaks in manual tune mode From MS tuning interface In Agilent s MassHunter Acquisition, Vacuum/Tune Control window 1. Select Parameters > Manual Tune 2. Click Scan tab on right-hand side 3. Set low m/z = 10, high m/z = 100 amu 4. Ensure PFTBA valve closed 5. Click Scan 6. Spray common leak sites, watch scan window for large duster ion spikes 7. Exit manual tune screen, do not save parameters 8. Return to acquisition Page 14

GC/MS Time Robber #2: Taking tune reports for granted Page 15

Tuning tip #1: Tune routinely to monitor MS performance Frequency depends on how often MS used Recommend at least weekly Provides comparison points over time If method calibration/response factors are a concern, use QuickTune Adjusts mass axis, peak width, and EM voltage only No adjustment of mass abundances Page 16

Tuning tip #2: Tune before AND after changes to GCMS Provides reference point in case something goes wrong Compare tune pdf files over time and look for differences Use these to troubleshoot, what changed from one tune to the next? Examples: Repair or preventative maintenance After cleaning MS source (changed abundance) Inlet maintenance (ensure no leaks) New column installation/column trim Change gas filter or cylinders Quickly catch contaminated gas tank Filament switch Vacation or prolonged period of downtime On Agilent 5977 and newer models, pdf copies are automatically created Default pathway: D:\MassHunter\GCMS\1\5977\PDF Page 17

Tuning tip #3: Create a maintenance GC method and cool MS tune file Create cool.u tune file Load your normal tune file Adjust quad and source temps 100 C Save as cool.u Create a Maintenance.m GC method Lower heated zones (set points 100 C, ideally cool enough to enable handling) Oven, inlets, mass spec transfer line, etc. Do NOT turn off carrier gas flows Specify cool.u tune file in MS section of method Load Maintenance.m GC method before venting MS, or even program into end of sequence for convenient cool down Page 18

Tuning tip #4: Monitor tunes for signs of a dirty source Tune ion peaks sharp and symmetric? 502 ion mass usually declines first Peak widths (Pk50) are usually ~ 0.6 amu Peak Shape Consistent primary tune ion abundances? Atune, HES Atune: m/z 69 ~450,000 600,000 Etune: m/z 219 ~450,000 600,000 Consistency is key Relative abundances within accepted values? 502 ion relative to m/z 69 > 3% Consistency is key EM voltage consistent from last tune? As source grows dirtier, EM voltage increases Max voltage is 3000V Consistency is key Is background free from excessive peaks? Excessive peaks can indicate source contamination Isotopic ratios within expected values? 502 isotopic ratio ~ 10.1% (±20%) Page 19

Case Study Monitoring tune reports over time Day 1 Peak Shape Peak shape: smooth and symmetric Etune ion abundance: m/z 219,526K Relative ion abundance: m/z 502, 7.2% EM voltage: 1711 # Background peaks: 282 Isotopic ratio: m/z 502, 10.1% Page 20

Case study Monitoring tune reports over time Day 2 Peak Shape Peak shape: smooth and symmetric EM voltage: 1985 # Background peaks: 181 Isotopic ratio: m/z 502, 10.8% Friday s Tune 219 Abund.- 526K Rel. Abund.- 7.2% EM Volts- 1711 #Bkgrd.- 282 Iso Ratio- 10.1% Page 21

Case study Monitoring tune reports over time Day 3 Peak Shape Peak shape: 502 slightly jagged Etune ion abundance: m/z 219, 79K Relative ion abundance: m/z 502, 3.6% EM voltage: 2175 # Background peaks: 266 Isotopic ratio: m/z 502, 8.2% Friday s Tune Monday s Tune 219 Abund.- Tune not 526K looking great. 219 Abund.- Cleaned 553K source, reran Rel. Abund.- tune and 7.2% all looked Rel. good. Abund.- Started 7.7% sequence. EM Volts- 1711 EM Volts- 1985 #Bkgrd.- 282 #Bkgrd.- 181 Iso Ratio- 10.1% Iso Ratio- 10.8% Page 22

Case study Monitoring tune reports over time Day 3 Peak Shape Peak shape: 502 slightly jagged Etune ion abundance: m/z 219, 79K Relative ion abundance: m/z 502, 3.6% EM voltage: 2175 # Background peaks: 266 Isotopic ratio: m/z 502, 8.2% Tune not looking great. Cleaned source, reran tune and all looked good. Started sequence. Tune passed. Started sequence. Page 23

What could have happened Day 4 Peak Shape Peak shape: really jagged EM voltage: 3000 # Background peaks: 628 Isotopic ratio: m/z 502, 6.2% Tuesday s Tune 219 Abund.- 79K Rel. Abund.- 3.6% EM Volts- 2175 #Bkgrd.- 266 Iso Ratio- 8.2% Page 24

GC/MS Time Robber #3: Poor Sensitivity Page 25

Improving sensitivity tip #1 Stop the bleeding Signal to noise relationship Excessive bleed noise Reduces spectral strength and quality Background & Bleed Sensitivity Potential sources of bleed or excessive background in GC/MS: Analytical column Leaks bleed Look for bleed ions in mass spectral background Problematic consumables Poor quality Overtightening Not changing often enough Bad gases Poor quality Lack of filtration, or not changing filter often enough Page 26

Improving sensitivity First aid and preventative medicine for column bleed Choose the right column Inert columns and liners are essential for trace GC/MS Better bleed levels Greater sensitivity for active analytes More runs before maintenance, source cleaning Choose the right dimensions Semivolatiles by GC/MS Blue trace: Agilent Inert Flow Path column and consumables Black trace: Competitor column and consumables Less column bleed, even at 320 C Thick films + GC/MS Treat your column right Oxygen is toxic to your column Column: Agilent JW DB-UI 8270D, 20m x 0.18mm x 0.36µm, P/N 121-9723 Liner: Agilent Ultra Inert single taper with wool, P/N 5190-2293 For detailed information, see Agilent application note 5991-2745 Maintain gas flow when installed in GC Cap column when not in use Prevent leaks (self-tightening column nuts!) Use and filter high purity gases Page 27

Improving sensitivity tip #2 Take care of your consumables Removed during PM Don t overtighten your septum nut Compresses septa chunks in flow path bonus peaks Replace routinely to prevent coring Choose high quality vials and caps Poorly constructed vial septa siloxanes bleed peaks Low quality vial leach contaminants into sample Choose the right cap/septa for your solvent Page 28

Reduced response Improving sensitivity tip #3 Bad gas stinks Decreased retention Oxygen in carrier gas is detrimental to GC, resulting in: Reduced response Elevated background Irreversible column damage Impaired electron multiplier function Premature filament, liner lifetime Excessive source maintenance Use UHP carrier gases 99.9995% or greater Use high quality, GC/MS gas filters GasClean gas filters remove oxygen and water Monitor gas filter indicators ~ 5% O 2 Replace when needed Correct any leaks present GC/MS filter Agilent P/N CP17973 Page 29

Improving sensitivity tip #4 Check yourself before you wreck your chromatography Lowest recommended ion to start a scan range Scanning masses <50 problematic due to common carrier gas contaminants: Water, m/z 18 Nitrogen, m/z 28 Oxygen, m/z 32 Carbon dioxide, m/z 44 Minimum recommended number of points, scans across a peak for quant Recommended number of SIM ions searched for a given compound Page 30

Thank you! Questions? Page 31 4/10/2018

Contact Agilent Chemistries and Supplies Technical Support 1-800-227-9770 Option 3, Option 3: Option 1 for GC/GCMS Columns and Supplies Option 2 for LC/LCMS Columns and Supplies Option 3 for Sample Preparation, Filtration and QuEChERS Option 4 for Spectroscopy Supplies gc-column-support@agilent.com lc-column-support@agilent.com spp-support@agilent.com spectro-supplies-support@agilent.com