Biomedical Instrumentation

Similar documents
Biomedical Instrumentation

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

APPLICATIONS OBJECTIVES PRE-LECTURE

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

High Voltage Safety. Dangers of Electricity. W. Maes. Department of Marine Engineering Antwerp Maritime Academy HV, 2015

VALLIAMMAI ENGINEERING COLLEGE

Dry body skin has a ( higher / lower ) resistance to the flow of the electric current than the internal organs of the body.

08-Electrical Safety. EE570 Energy Utilization & Conservation Professor Henry Louie

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

Principles of Electrical Currents. HuP 272

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

ISO INTERNATIONAL STANDARD. Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers

Biomedical Instrumentation

EC6001 MEDICAL ELECTRONICS

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

Implants for surgery Active implantable medical devices. Part 7: Particular requirements for cochlear implant systems

Pacing Lead Implant Testing. Document Identifier

EE 4BD4 Lecture 20. Therapeutic Stimulation

Figure 2. Normal ECG tracing. Table 1.

Electrodes (for TransAeris System)

8/28/18. Anesthetic considerations for patients with implanted devices for treating chronic pain and more. Objectives:

EI2311 BIOMEDICAL INSTRUMENTATION

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

ISO INTERNATIONAL STANDARD. Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers

Mini Pulse Electronic Stimulator

ProSim 3 and 2. Vital Signs Simulators

ProSim 3 and 2 Vital Signs Simulators

The HeartCheck TM PEN Handheld ECG Is your heartbeat slow, fast, or irregular? Are you at risk? Put your heart health in your own hands

MultiStim SWITCH and MultiStim SENSOR Setting the trend in nerve stimulation

EXTERNAL TEMPORARY PACEMAKERS

The Normal Electrocardiogram

ELECTRICAL INJURY 9/21/2015 I HAVE NO DISCLOSURES WE HAVE OBTAINED APPROVAL FOR USE OF IDENTIFIABLE PATIENT PHOTOS

CSA Z32 TESTING GUIDELINE AND PROCEDURES

CORONARY ARTERIES HEART

This leaflet is intended for patients with permanent pacemakers and their parents.

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields

The Automated Defibrillator: A Biomedical Engineering Success Story. Dr. James A. Smith

Advanced ICD Concepts

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Neurostimulation Systems

ELECTROCARDIOGRAPHY (ECG)

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

Pediatric pacemakers & ICDs:

Compatible with the following products TENS. Replacement Pads. User Manual. Last revised: V Last revised: V

WHOLE BODY VIBRATION THERAPY

The Electrocardiogram

Function of Cell Membrane

Model 5392 EPG Temporary Pacer

C J Mieny, U Mennen: Principles of Surgical Patient Care - Volume I. Chapter 2: Intraoperative Care

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

TIP CARD. Model Temporary External Pacemaker

MAGPRO. Versatility in Magnetic Stimulation. For clinical and research use

STUDY REPORT ON THE EFFECT OF RADIO WAVES ON MEDICAL DEVICES

Cardiac Cycle. Each heartbeat is called a cardiac cycle. First the two atria contract at the same time.

Chapter 23 Unit 28. Therapeutic Modalities

This document is a preview generated by EVS

How Do We Sense, Think, and Move? -- Lab #11 Bioelectronics Measuring Electrical Properties of the Body

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

OWNER'S MANUAL OxyTrainer

Lotus transducer Instructions for use

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Chapter 3 Biological measurement 3.1 Nerve conduction

The Do s and Don ts of Pressure Transducers

Contents. Helpline No. UK/ North Ireland Rep. Ireland Web Support Model Number: 12005

Yes, out of sight out of mind.

SLEEP IMPROVING WRISTBAND. Item No Owner s Guide

LABORATORY INVESTIGATION

ECG SENSOR ML84M USER S GUIDE. CENTRE FOR MICROCOMPUTER APPLICATIONS

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

Biomedical Engineering in Root Cause Analysis Example: Assessing Infant Apnea-Related Deaths

NOVIN medical engineering Co. Designer and manufacturer of physiotherapy equipment

SaeboStim Micro. Saebo INSTRUCTION MANUAL

Peri-operative management of pacemakers and implantable cardiac defibrillators

Advisa MRI and Ensura MRI SureScan pacing systems

After Your Pacemaker Generator Change

Patient Safety During Cardiac Catheterization

Introduction to Physical Agents Part II: Principles of Heat for Thermotherapy

Module 1: Fundamentals of Patient Monitoring. Contents. Introduction. Why is monitoring in MR different? Exercise 1_1

Contra-Indications, Warnings, Cautions & Precautions

PD233: Design of Biomedical Devices and Systems

Electrocardiography Biomedical Engineering Kaj-Åge Henneberg

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

ECG Check Wireless 1- Lead Cardiac Monitor. Rx Use

Analysis of Human Cardiovascular System using Equivalent Electronic System

Safety Issue of Static Magnetic Field (1)

Effects of myocardial infarction on catheter defibrillation threshold

PROFESSIONAL RECOVERY FOR SERIOUS ATHLETES. Vibrating 4 Speed Rechargeable Foam Roller Manual

V4, V5 and V6 follow the 5 th intercostal space and are NOT horizontal as indicated in the image. Page 1 of 7

VENTRICULAR DEFIBRILLATOR

RN-BC, MS, CCRN, FAHA

PATIENT PROGRAMMER 37642

Muscle Stimulation. Page 1

AspireSR. Essential information about your new device

Defibrillator. BiomedGuy

Your Subcutaneous Implantable Cardiac Defibrillator (ICD)

Transcription:

University of Zagreb Faculty of Electrical Engineering and Computing Biomedical Instrumentation Safety of electrical medical devices prof.dr.sc. Ratko Magjarević

Medical technology Achievements Significant improvement of health care in all medical specialties Reducing the number of seriously ill people Reduction of mortality Consequences increased complexity of medical devices and their applications increased incidence of adverse effects of implementation of MT in the U.S. approx. 10,000 cases of patient injury due to the application of MT

Medical technology Causes of unintended consequences: improper handling insufficient or improper staff practice (training) in handling with MT insufficient experience example, medical staff rarely read manuals for the users (User Manual) equipment malfunction Demand for MT developers developement of safe devices (failure-safe)

Danger to patients In a clinical environment patient is exposed to various risks, more than a typical workplace or at home frequent invasive (blood) operations - penetration through skin or mucous membranes presence of potentially hazardous chemicals and substances - anesthetics, medicines, medical gases sources of infection - particularly "hospital infection" various sources of energy that penetrate into or through the patient: current, voltage, ionizing and non-ionizing radiation, sound and ultrasound, electric and magnetic field, UV radiation, lasers, microwave radiation, mechanical stress, etc.

Safe MT development Electrical MT security is considered: physiological effects of electricity possibility (risk) of failures and their consequences methods of patients and staff protection standards describing electrical safety electrical safety testing modes Understanding the possible dangers and risks Implementation in achieving security

Physiological effects of electricity Body (tissue) becoming a part of an electrical circuit The amount (amplitude) of electricity often depends on the ratio of voltage present and the sum of all (serially connected) impedance Usually, the highest impedance is the impedance of the skin The consequence of current flow: nerve and/or muscle tissue stimulation heating of tissues (a result of tissue resistance) burns

Pulsed current I t curves for sensory and motor responses

Physiological effects of current

Electricity grid Unintended consequences are usually a result of human involvement in the circuit, in some way connecting to the grid (inattention, failure) Several levels of action perception contractions (let-go) paralysis (respiratory), pain, fatigue ventricular fibrillation tetanic contraction burns and physical injuries

Consequences of dangerous voltage Fibrilation Convulsion Consequences depende on the voltage current relation, but also on individual susceptibility Expected values Intact skin, R K = 2-5 kω Damaged skin - injuries, surgical procedures, R O = 100 Ω 1 kω

Consequences of dangerous voltage Focus on the worst possible scenario, for example, a person holding two conductors at different potentials in his hands resulting muscle spasms, inability to release captured conductors In case of intact skin, R K = 2-5 kω, cramps can occur even at a voltage of few dozen ma In case of damaged skin - injuries, surgical procedures, R O = 100 Ω 1 kω, fibrillation may occur at voltages of several volts and current of few dozen ma

Excitability dependence of the pulse waveforms Monopolar vs. bipolar pulses, sinusoidal current Extrapolation of sinusoidal current What is the duration of the half-cycle sine current to achieve maximum excitability?

Safe voltages - AC Security measures in designing defining voltage values that may not appear on the conductive parts of instruments or equipment coming into contact with the patients skin or user for alternating current: safety extra low voltage, SELV of 50V, 50/60Hz medical safety extra low voltage, MSELV of 25V, 50/60Hz

Safe voltages - DC Security measures in designing defining voltage values that may not appear on the conductive parts of instruments or equipment coming into contact with the patients skin or user for direct current: safety extra low voltage, SELV of 120V, medical safety extra low voltage, MSELV of 60V, 50/60Hz

Ventricular fibrillation In cases of shock due to unwanted connection to power grid voltage, current passing through the heart is considered the most dangerous effect and can cause ventricular fibrillation (ventricles) with fatal consequences Ventricular fibrillation is completely irregular, asynchronous heartbeat; no synchronism in ventricular contractions, blood flow stops and if such heart work does not stop, it leads to death within a few minutes

Microshock vs. macroshock The consequences of current passage through the tissue depend on the contact point with the source of voltage Even greater power will not necessarily cause fatal consequences, eg. fibrillation, when going through the limbs

Microshock vs. macroshock Macroshock is caused by current passing through the body through the skin, and the current that can cause harmful effects is relatively high and significantly depends on the point of contact Microshock is caused by the passage of relatively low current, but the source of electricity was brought directly into contact with the heart, for example, during cardiac catheterization (catheter in the heart is a diagnostic procedure). Electricity sufficient to induce fibrillation is of n x 10μA. Another point of contact can be anywhere on the body (eg limbs). Current of 10 μa is considered safe limit to prevent microshock

Protective grounding Generally, patients are only occasionally and accidentally exposed to risk of getting in touching with devices whose conductive parts may be energized In hospitals, especially in intensive care units, patients are deliberately connected to the diagnostic and/or therapeutic electrical devices/equipment particularly careful with the isolation of any conductive parts connected to the heart or its vicinity from all other conductive parts all conductive parts in the vicinity of the patient must be connected to a single point grounding (eg, metal bed, cupboard, etc.) periodic testing of the grounding impedance must be provided Tolerable difference in potential between the grounded conductive parts in clinical areas: in the general - of 500 mv intensive care and other critical cases- of 40 mv

Power in clinical environments Under normal conditions, increasing patient safety is provided by: formation of isolated parts of the grid installation of automatic monitoring of impedance grounding (line isolation monitor)

Power failure In case of grid power failure, in critical areas in hospitals power is automatically provided within 10s after stoppage Critical components are: lighting alarms equipment in intensive care

Equipment safe during the first failure The principle of redundancy can be applied to the safety of medical devices and equipment (KEU) For example, it is required that electrical medical equipment is safe even in case of a breakdown on the conductive contact with patient Grounding impedance must be sufficiently low so that the patient may not get in touch with any dangerous voltage

Leakage currents Occur between conductors that are not in direct contact (touch) and are at different potentials For devices that are powered from the grid, leakage currents are: capacitive character (spreading capacity) operative character (current through the insulation, dust, humidity...)

Leakage currents pathways Pathways of leakage current through the patient a) Intact ground lead b) Broken protective ground lead, patien connected to ground through a catheter c) Broken protective ground lead, patien connected to ground by an electrode or unintentianally.

Leakage currents

Conductive paths to / from the heart In clinical settings there are current flow pathways that cause heart microshock: The external pacemaker connected to epicardial (heart area) or endocardial (inside heart) electrodes Intracardial ECG measuring electrodes Catheters for invasive heart pressure measuring taking blood samples injection of contrast or medication into the heart

Regulations, standards and recommendations Regulations are subordinate legislation which are obligatory and are related to the characteristics of devices and equipment that must be assured Standards stipulate features that must be assured, but their appliance is voluntary. Contain instructions on how to accomplish / verify compliance with the standard Recommendations (usually manufacturers) stipulate how to ensure the safety and prescribe manner and frequency of testing safety of devices

Standards 60601 Various standards institutes prescribe standards for electromedical devices and equipment These standards are mostly aligned, due to the global market The fundamental norm in Europe, IEC 60601-1, Medical Electrical Equipment, Part 1:General requirements for basic safety and essential performances These codes are followed by the standard 60601-2- xxx, Particular requirements for...

Design Recommendations for Safety Reliable grounding for equipment Reduction of leakage current Double insulated equipment Operation at low voltages Driven right leg circuit Current limiters Electrical isolationisolated heart connectors

Safety checking Electrical Safety Analysis: general for certain types of electrical medical devices and equipment Demo practice: "General" analyzer according to 60601-1 Defibrillator analyzing operation

Determination of device categorie

Mind the costs, but satisfy safety requirements! Price/performance

Literature J. Webster: Medical Instrumentation, Chapter 14, Electrical Safety