QUICK REFERENCE GUIDE. The XCaliber Meta-Diaphyseal Fixator

Similar documents
QUICK REFERENCE GUIDE. The XCaliber Articulated Ankle Fixator. By Dr. S. Berki, Dr. V. Caiaffa, Dr. F. Lavini and Dr. M. Manca

QUICK REFERENCE GUIDE. The PreFix Fixator (92000 Series) ALWAYS INNOVATING

QUICK REFERENCE GUIDE. Arthrodesis. Joint Fusion. By Dr. S. Agostini and Dr. F. Lavini

QUICK REFERENCE GUIDE. Arthrodiatasis. Articulated Joint Distraction

QUICK REFERENCE GUIDE. The Orthofix Femoral Nailing System. By Prof. Dr. D. Pennig

A. XCaliber External Fixator (91000 Series)

QUICK REFERENCE GUIDE. The Pennig Dynamic Wrist Fixator. Part A: Trans-articular application

OPERATIVE TECHNIQUE. Knee Hinge (LRS Advanced System)

N. Pennig Dynamic Wrist Fixator (37000 Series)

DFS STANDARD FIXATOR DFS ANKLE CLAMP DFS T-CLAMP

Orthopedic Bone Nail System - Distal Femoral Nail Surgical Technique Manual

Technique Guide. Adjustable Distal Radius Fixator. Part of the Synthes External Fixation Systems.

3. Insert Tocar Sleeves Insert the NCB tissue protection sleeve assembly 1.6 to 10mm through a skin incision (Fig. 38).

MiniRail System. Part B: Foot Applications. By Dr. B. Magnan, Dr. E. Rodriguez and Dr. G. Vito

OPERATIVE TECHNIQUE. The PreFix 2 Fixator

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

OPERATIVE TECHNIQUE GALAXY FIXATION SHOULDER

Monotube Triax. External Fixation System. External Fixation. Operative Technique

Surgical Technique. Anterolateral and Medial Distal Tibia Locking Plates

NCB Distal Femur System. Surgical Technique

Monolateral External Fixation System for Trauma and Orthopaedics

LCP Distal Tibia Plate

Knee Surgical Technique

OPERATIVE TECHNIQUE. The Ring Fixator System. The Sheffield Ring Fixator - Limb Reconstruction. and Complex Trauma. By Prof. M.

Knee spanning solutions

MEFiSTO. Monolateral External Fixation System for Trauma and Orthopaedics.

Pre-Operative Planning. Positioning of the Patient

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

For Distal Femur Fractures. 95º Condylar Plate. Quick Reference Chart

3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM

SMV Scientific Bone Plate and Screw System Surgical Technique

Pediatric LCP Plate System. For osteotomies and fracture fixation of the proximal and distal femur.

3. PATIENT POSITIONING & FRACTURE REDUCTION 3 8. DISTAL GUIDED LOCKING FOR PROXIMAL NAIL PROXIMAL LOCKING FOR LONG NAIL 13

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments

Surgical Technique. Targeter Systems Overview

Technique Guide. The Distraction Osteogenesis Ring System. Nonarticular tibia frame.

LCP Distal Humerus Plates

OPERATIVE TECHNIQUE. The Pennig Dynamic Wrist Fixator. Fractures of the Distal Radius and Radial Osteotomies. By Prof. Dr. D.

Elbow Hinge Fixator. Guided Flexion/Extension for Unstable Elbow Fractures.

Surgical Technique. 3.5mm and 4.5mm Lateral Proximal Tibia Locking Plates

Zimmer MIS Periarticular 3.5mm Proximal Tibial Locking Plate

3.5 mm LCP Extra-articular Distal Humerus Plate

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

System. Humeral Nail. Surgical Technique

28 Surgical Technique

3.5 mm Locking Attachment Plate

Medical Devices HUMERUS. WRIStX

humerus InSafeLOCK Nail

OPERATING MANUAL AND TECHNIQUE GUIDE FOR TITANIUM FEMORAL AND TIBIAL NAILING SYSTEMS

Large Distractor Femur

OSS Orthopedic Salvage System Compress Device. Surgical Technique

PediLoc 3.5mm and 4.5mm Contour Femur Plate Surgical Technique

VA-LCP Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Patient Guide. Intramedullary Skeletal Kinetic Distractor For Tibial and Femoral Lengthening

Small Fragment Plating System. Securing optimal fixation through locked and compression plating technology

PediLoc Tibia Surgical Technique

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

Low Bend Distal Tibia Plates

DFS Hip Distractor. Surgical Technique

Biomet Large Cannulated Screw System

VA-LCP Condylar Plate 4.5/5.0. Part of the Synthes Variable Angle Periarticular Plating System.

Double Engine Orthopedic Bone Nail System Universal Humeral Nail

Zimmer MIS Periarticular Distal Femoral Locking Plate

OPERATIVE TECHNIQUE. Galaxy UNYCO Diaphyseal Tibia Sterile Kit. Key contributors: S. Nayagam, MD T. Bégué, MD W.T. Gordon, MD

Olecranon Locking Plate II

PROXIMAL TIBIAL PLATE

Sirus Antegrade Femoral Nail System Surgical Technique

Periarticular Aiming Arm Instruments for LCP Proximal Tibial Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large).

Access Pelvic Fixator

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

PediLoc 3.5mm and 4.5mm Bowed Femur Plate Surgical Technique

Small Fragment Plating System

3.5 mm LCP Hook Plate

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

OPERATIVE TECHNIQUE GALAXY FIXATION SYSTEM UPPER EXTREMITIES

Femur. Monoaxial Locking Plate System. Operative Technique. Distal Lateral Femur Universal Holes Targeting Instrumentation.

Surgical Technique. CONQUEST FN Femoral Neck Fracture System

LUMBAR POSTERIOR MINIMALLY INVASIVE SYSTEM. Surgical Technique

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

3.0/3.5/4.0/4.5/6.5/7.0/7.3. Cannulated Screws. Surgical Technique

pediatric orthopedic implants fixator

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Cannulated Pediatric Osteotomy System (CAPOS)

Surgical Technique. Distal Humerus Locking Plate

Mini External Fixator.

AxSOS. Locking Plate System. Operative Technique. Small Fragment Basic Fragment

QUICK REFERENCE GUIDE. MiniRail System. Part B: Foot Applications. By Dr. B. Magnan, Dr. E. Rodriguez and Dr. G. Vito ALWAYS INNOVATING

Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area.

GREENS SURGICALS. Redefining Excellence INSTRUMENT SYSTEM PREPARED BY: DR. VINAY KUMAR

Transcription:

17 The XCaliber Meta-Diaphyseal Fixator

GENERAL POINTS The XCaliber Fixator is made of radiolucent material for unobstructed X-ray visualization. The metallic bolts and the cam and bush of each ball-joint, are the only radio-opaque components. Because it is radiolucent and made of a composite material, the ball-joint deforms after repeated tightening. It can be adjusted on the patient if repositioning of the fracture is required, but will not be strong enough for use on a second patient. In addition the joint is sealed and cannot be dismantled for cleaning. The XCaliber Fixator is strictly single patient use. Ball-joints move 22 in any direction Detachable T-clamp with converging screws Straight clamp Cams locked from either side. Torque wrench not required. EQUIPMENT REQUIRED 99-91215 XCaliber Meta-Diaphyseal Kit with Dynamic Compression/ Distraction Unit, sterile 19200 XCaliber Manipulation Forceps 99-91038 Supplementary Screw Holder (Bar and Clamps), sterile 90037 Supplementary Screw Holder Standard Instrumentation for Screw Insertion or 99-91600UE XCaliber Meta-Diaphyseal Complete Kit, sterile Consisting of: 1x91150 Universal T Wrench 1x91017 Universal Allen Wrench 1x11106 Drill Guide 3.2x40 mm 1x11104 Drill Guide 4.8x40 mm 1x91000 XCaliber Meta-Diaphyseal Fixator 3x11102 Screw Guide L 60 mm 1x91015 Dynamic Compression/Distraction Unit 1x99-611560 XCaliber Osteotite Screw 150/60, sterile 1x99-611550 XCaliber Osteotite Screw 150/50, sterile 1x99-611540 XCaliber Osteotite Screw 150/40, sterile 3x99-611530 XCaliber Osteotite Screw 150/30, sterile 1x1101101 Cannulated Drill Bit 3.2x200 mm 2x11014 X-Wire without olive 1.5x250 mm 1x1100101 Drill Bit 4.8x180 mm 19200 XCaliber Manipulation Forceps 99-91038 Supplementary Screw Holder (Bar and Clamps), sterile 90037 Supplementary Screw Holder STERILE R CAUTION: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. Contents sterile unless package opened or damaged; Do not use if package is opened or damaged. 1

DIAPHYSEAL FRACTURES TIBIA Reduce the fracture as anatomically as possible, emphasizing rotational correction. A stable, good reduction on a fracture table with skeletal traction is recommended. Screw Insertion Insert screws into the shortest or most difficult segment first, from the anterior or antero-medial aspect of the tibia. Make a 15-20 mm incision so that the skin around each screw is not too tight. The underlying tissues also require broad dissection down to the bone. Insert a screw guide perpendicular to the longitudinal axis of the bone. Use a trocar to locate the midline by palpation. Keep the screw guide in contact with the cortex by gentle pressure, withdraw the trocar, and tap the screw guide lightly to anchor its distal end. Note: If the placement of this screw is critical because it is close to a joint, the position can first be checked by inserting a 2 mm Kirschner wire. A screw guide and drill guide can be centred over the wire and the bone drilled with the 3.2 mm cannulated drill bit. Insert the correct drill guide into the screw guide. Use screw guides for every screw insertion to minimize soft tissue trauma. When using the standard 5-6 mm bone screws, we recommend a 3.2 mm drill bit for cancellous bone, and a 4.8 mm drill bit for cortical bone. If self-drilling XCaliber bone screws are being used, we recommend that they are used self-drilling in cancellous bone. However, in diaphyseal bone, pre-drilling is recommended; use a 4.8 mm drill bit through a drill guide when the bone is hard; when the bone quality is poor, or in the metaphyseal region where the cortex is thin, a 3.2 mm drill bit should be used. 2

Drill at 500-600 rpm through the first cortex with the correct drill bit and drill stop, checking that the drill bit is at right angles to the bone. The force applied to the drill should be firm and the drilling time as short as possible to avoid thermal damage. Stop at the second cortex, offset the stop collar by 5 mm, and continue through the bone. Ensure that the drill bit completely penetrates the second cortex. If a preliminary K-wire has been used, a cannulated drill bit can be used over the wire. Note: Kirschner wires which are used for this purpose and cannulated drill bits should NEVER be reused. Remove the drill bit and drill guide, keeping pressure on the handle of the screw guide. Insert the selected screw and turn it with the T-wrench until it reaches the second cortex. A further 5 or 6 half turns are then normally required to ensure that about 2 mm of the screw protrudes beyond the second cortex. Diaphyseal bone screws should always be inserted in the centre of the bone axis, to avoid weakening it. In all cases the surgeon should be mindful of the amount of torque required to insert the screw. If it seems tighter than usual, it is safer to remove the screw and clean it, and drill the hole again with a 4.8 mm drill bit, even if it has already been used. Warning! As the thread is tapered, repositioning the screw by turning counterclockwise will loosen the bone-screw interface. If the self-drilling XCaliber Bone Screws are being used, please refer to the Insertion Technique PG 20. Central body locking nut Fixator Application If desired, the T-clamp may be removed for this application. Leave the screw guide in position and apply the XCaliber Fixator with the central body locking nut and cams loosened. Before screw insertion into the second bone fragment, adjust the fixator body to the correct length, making sure that it is neither completely closed nor fully open. This will allow for final reduction. Check that the central body locking nut faces away from the bone so that it can be tightened, and that it does not obstruct views of the fracture site. The fixator body should be parallel to the bone axis. Insert a screw guide into the second clamp, as far as possible from the first for maximal stability; make the incision, and insert it down to the bone as before. Tighten the clamp locking screws with the Allen wrench to ensure that the clamps grip the screw guides. Repeat the screw insertion procedure as before. 3

Place the remaining screw guides in the inner screw seats of the clamps down to the bone; tighten the clamp locking screws to ensure that the screw guides are parallel, and insert the bone screws as before. The number of screws in each clamp is determined by the stability of the fracture, the weight of the patient, the bonefixator distance and the quality of the bone. In adults we generally recommend that three screws are used each side of the fracture. In this case, a third central screw is inserted after the two outer screws. Partially loosen the clamp locking screws and remove the screw guides. Position the XCaliber Fixator at least 2 cm from the skin to allow for post-operative swelling and cleaning. Align the body parallel to the major axis of the segment being treated. Fully tighten the clamp locking screws. Obtain final reduction with manipulation forceps. Place hooks of forceps over bone screws. Tighten the nut to lock the forceps in place. Insert Allen wrench through hole in handle if greater torque is required. Accurate reduction is aided by the fact that the fixator is radiolucent, allowing unobstructed views on image intensification. Hold the reduction in a good position, while an assistant PARTIALLY tightens the cams with the Allen wrench. Tighten the central body locking nut. Check reduction and lock the cams definitively. Note: Final locking of the ball-joints is achieved with the Allen wrench; a torque wrench is not required. The cams can be locked from either side of the clamp. They should be turned towards the thicker section of the coloured insert until tightly closed, and the cam is at least 50% of the way across the recess. 4

SUPPLEMENTARY SCREWS A fracture will be held in a more stable position if the nearest bone screws are applied fairly close to the fracture margin, and if these distances are equal on both sides of the fracture. A minimum of 2 cm is recommended between the fracture and the nearest screw. A supplementary screw holder is supplied to achieve this. A screw should be inserted into the longer bone segment at an equal distance from the fracture as the nearest screw in the short segment, using a screw guide. This screw is attached to the fixator either with a clamp over the fixator body, or over the nearest convenient bone screw. A 6 mm Allen wrench should be used to tighten the supplementary screw holder clamps. A supplementary screw can also be used to stabilize a third fragment. This screw should be removed before the fixator is dynamized. When a supplementary screw is used, it is normally sufficient to use 2 screws in the clamp in positions 1 and 3. COMPRESSION-DISTRACTION UNIT To achieve compression or distraction, remove the plastic covers from the pins on the fixator body, attach the compressiondistraction (C-D) unit and lock it in position with the Allen wrench. With the central body locking nut loosened, use the Allen wrench to turn the central element of the compression-distraction unit either clockwise or counterclockwise, (one full turn clockwise = 1 mm compression; one full turn counterclockwise = 1 mm distraction). Compression is never recommended in a fresh fracture. Release any skin tension around the screws by extending the skin incision. The dynamic C-D unit is designed to allow early dynamization of the fracture, by cyclic micromovement on weightbearing, up to 2 mm at the fracture site. Attach the unit to the fixator pins and lock it firmly in position. Loosen the body locking nut. Note: For full dynamization, remove the C-D unit and leave the body locking nut undone. This is normally recommended at 4 weeks, but should be delayed in cases of bone deficit or unstable fractures. Healing time is shortened and non-union rate decreased if full dynamization is carried out at the correct time. Dynamic 5

FEMUR Reduce the fracture, with particular attention to correction of rotational displacement. Apply the fixator to the lateral aspect of the femur. Insert 6 screws, 3 in each clamp, beginning with the shortest segment. Position the second group of screws so that the body of the fixator is open at least 1 cm. Screws should never be less than 2 cm from the fracture line. Use a supplementary screw if indicated as for the tibia. Obtain final reduction and lock the fixator as above. Note: When unusually high loading conditions are likely, such as weightbearing with a femoral application or when the patient is very heavy, before the ball-joints are locked the fixator body should be aligned so that the body locking nut is at 90 degrees to the plane of the bone screws. In addition for increased stability the compression-distraction unit may be applied to the fixator body and locked into place. HUMERUS Reduce the fracture, with particular attention to correction of rotational displacement. Apply the fixator to the lateral aspect of the bone. Insert 4 screws, 2 in each clamp. Insert the most distal screw first, 1 cm proximal to the lateral epicondyle. The second distal screw should be inserted in the second seat of the clamp. To avoid risk of damage to the radial nerve, use an open procedure. Position the proximal screws to allow the body of the fixator to be open a minimum of 1 cm. Reduce and lock as above. 1 2 METAPHYSEAL FRACTURES TIBIA: PROXIMAL METAPHYSEAL FRACTURES Antero-Medial Application Insert the most anterior proximal screw first freehand, 2 cm distal to the knee joint. OsteoTite (HA-Coated) bone screws may be useful in this application. Position the screw within its screw guide in the proximal seat of the straight clamp. When using the T-clamp, the first screw should always be inserted in this screw seat. 6

The T-clamp allows for either parallel or convergent positioning of the proximal screws. The latter is achieved by rotating the T-clamp to find the most favourable position for the remaining screw(s). Rotation of the T- clamp is locked by tightening the plastic screw until the Allen wrench slips in the hexagon in the screw head. To loosen this screw, insert the end of a 3 mm Allen key into one of the holes in the edge of the screw head, and turn the screw counterclockwise. Insert the second screw. A third screw may be used. Screws may be inserted in a straight or triangular configuration, depending on the shape of the proximal bone fragment. Three screws should always be used in adults, spaced as far apart as possible. Note: When the convergent mode is used, make sure that the fixator is placed at the correct distance from the bone before inserting the second screw, as the fixator will not slide along convergent screws. Insert the diaphyseal screws, again checking that the fixator body is not completely closed. Lock the fixator as above. Anterior Application Insert the first screw in the proximal seat of the straight clamp and the second screw so that it converges slightly with the first, in the screw seat at the other end of the clamp. Note: OsteoTite (HA-Coated) bone screws may be useful in this application. 7

Insert the diaphyseal screws, ensuring that the fixator body is not completely closed. Lock the fixator as above. TIBIA: DISTAL METAPHYSEAL FRACTURES Insert the most posterior distal screw first, freehand, immediately anterior to the posterior border of the medial malleolus. Insert it with a screw guide into the most distal seat of the straight clamp. Insert the second distal screw in one of the seats of the T-clamp. A triangular screw configuration will be the most stable if there is room for the screws. Note: OsteoTite (HA-Coated) bone screws may be useful in this application. If the placement of this screw is critical because it is close to a joint, the position can first be checked by inserting a 2 mm Kirschner wire. A screw guide and drill guide can be centred over the wire and the bone drilled with the 3.2 mm cannulated drill bit. Insert the diaphyseal screws, checking that the fixator body is not fully closed. Lock the fixator as above. 8

FEMUR: DISTAL METAPHYSEAL FRACTURES Apply the XCaliber Fixator laterally, inserting the most anterior distal screw first, about 1 cm behind the anterior edge of the lateral condyle. With this screw and its screw guide in the most distal seat of the straight clamp, insert the second distal screw in one of the seats of the T-clamp. Three screws should be used distally, in a triangular configuration if possible. Note: OsteoTite (HA-Coated) bone screws may be useful in this application. If the placement of this screw is critical because it is close to a joint, the position can first be checked by inserting a 2 mm Kirschner wire. A screw guide and drill guide can be centred over the wire and the bone drilled with the 3.2 mm cannulated drill bit. Insert the diaphyseal screws, checking that the fixator body is open at least 1 cm. Lock the fixator as above. The Orthofix Quality System has been certified to be in compliance with the requirements of: Medical Devices Directive 93/42/EEC, Annex II - (Full Quality System) as amended in 2007/47/EC International Standards ISO 13485 / ISO 9001 for external fixator devices, implants for osteosynthesis and related instruments.! See Orthofix External Fixation System instruction leaflet (PQ EXF) prior to use. 9

Orthofix wishes to thank S. Berki, MD Department of General, Trauma and Hand Surgery, University and County Hospital, Szentes, Hungary V. Caiaffa, MD Orthopaedic Department, SS. Annunziata Hospital, Taranto, Italy F. Lavini, MD Institute of Clinical Orthopaedics and Traumatology, University of Verona, Verona, Italy M. Manca, MD Orthopaedic Department, Versilia Hospital, Viareggio, Italy for their invaluable help in the preparation of this Quick Reference Guide and their contribution to the design and refinement of the equipment described.

Manufactured by: ORTHOFIX Srl Via Delle Nazioni 9 37012 Bussolengo (Verona) Italy Your Distributor is: Telephone +39-0456719000 Fax +39-0456719380 0123 Deformity Correction I Trauma I Pediatrics I Bone Growth Stimulation www.orthofix.com PG 170 E0 E 04/10