In vitro Assessment of Clasps of Cobalt-Chromium and Nickel-titanium Alloys in Removable Prosthesis

Similar documents
Evaluation of Acetal Resin and Cobalt Chromium Clasp Deformation and Fatigue Resistance in Removable Partial Denture Clasps - An In Vitro Study

Comparison between Acetal Resin and Cobalt-Chromium. Removable Partial Denture Clasp Retention: An in vitro Study.

Stress distribution within circumferential clasp arms

Dynamic retentive force of a mandibular unilateral removable partial denture framework with a back-action clasp

Removable partial dentures

AESTHETIC DESIGNS IN REMOVABLE PARTIAL DENTURES

Acknowledgments Introduction p. 1 Objectives p. 1 Goals p. 2 History of Dental Materials p. 3 The Oral Environment p. 4 Characteristics of the Ideal

ABSTRACT INTRODUCTION /jp-journals

IJopRD INTRODUCTION ABSTRACT MATERIALS AND METHODS /jp-journals

Comparison of Flexural Strength of two Provisional Materials Used In Fixed Prosthodontics -An in Vitro Study.

Removable Partial Dentures

The Loading And Unloading Properties Of Various Arch Wires As A Function Of Cross Sectional Dimension And Inter Bracket Span Width System

INDIAN DENTAL JOURNAL

Construction of Removable Partial Denture

Palatal Vault Depth Influence on the Flexural Strength of Two Heat Cure Acrylic Denture Base Resins: An in vitro Study

Comparative analysis of load/deflection ratios of conventional and heat-activated rectangular NiTi wires

Comparison of Different Methods of Abutment Splinting and Attachments on Teeth Kinetics (Part I)

Novel Removable Keeper System for Magnetic Attachments on Overdenture Abutments

Is nickel titanium superior to multistranded stainless steel wire in aligning crowded lower anteriors? A comparative in-vivo study.

Three-dimensional finite element analysis of the application of attachment for obturator framework in unilateral maxillary defect

Prosthetic V. Removable dentures I.

FR Attachment Instructions Removable Partial Dentures

Experimental and Numerical Comparison of Stresses Level between Titanium and Novel Composite Single Tooth Implant

ACRYLIC REMOVABLE PARTIAL DENTURE(RPD)

Clasp design. Many RPD design principles are based. prosthetics PRACTICE

Rehabilitation of Resorbed Mandibular Ridge with Implant Supported Overdenture- A Clinical Report

Study to evaluate the Effect of Silane Treatment and Three Different Woven Fiber Reinforcement on Mechanical Properties of a Denture Base Resin

Supervisor at Dental Technology Lab.

THE LOAD EXERTED BY REMOVABLE PARTIAL DENTURE CLASPS CONFORMING TO AVERAGE TOOTH CURVATURES, IN CLINICALLY ENCOUNTERED UNDERCUTS.

Dual Path: A Concept to Improve the Esthetic Replacement of Missing Anterior Teeth with a Removable Partial Denture

Reliability of Orthopantomography and Cone-beam Computed Tomography in Presurgical Implant Planning: A Clinical Study

Effect of Thickness and Recycling on Transverse Strength of Relined Acrylic Resin Denture

An evaluation of the hardness of flexible Denture Base Resins

The key to exceptional craft lies in exceptional materials.

Double-casting method for fixed prosthodontics with functionally generated path

Retentive forces and fatigue resistance of thermoplastic resin clasps

IMPRESSION TECHNIQUE AND MATERIALS

DENT Advanced Topics in Removable Prosthodontics, Winter 2008

Lect. 14 Prosthodontics Dr. Osama

Comparison of Fracture Resistance of Cast Post and Core using Different Alloys: An Unexplored Frontier

Prosthetic Options in Implant Dentistry. Hakimeh Siadat, DDS, MSc Associate Professor

for RPDs made with Ultaire AKP (Aryl Ketone Polymer) Make METAL-FREE Possible with Ultaire AKP Solvay Dental 360 Ultaire AKP Reimbursement Guide 1

Treatment of an open bite case with 3M Clarity ADVANCED Ceramic Brackets and miniscrews.

Department of Prosthodontics and Periodontics, Piracicaba Dental School, Campinas State University, Piracicaba, SP, Brazil

Oral Rehabilitation with CAMLOG implants after loss of dentition due to an accident

3D RECONSTRUCTION AND SLM SURVEY FOR DENTAL IMPLANTS

RELATIONSHIPS MATTER. Innovation, Expertise and. Craftsmanship. in Laboratory. Products and. Services for the. Dental Profession

ISW for the treatment of moderate crowding dentition with unilateral second molar impaction

GC Manufacturing Europe

Implant and Tooth Supported Full-Mouth Rehabilitation with Hobo Twin-Stage Technique

SCREW LOOSENING has been reported as

Comparative study of mechanical properties of dental restorative materials and dental hard tissues in compressive loads

MDJ The effect of design on Removable Partial Dentures Vol.:11 No.:1 2014

Department of Prosthodontics, Faculty of Dentistry, Mahidol University, Bangkok, Thailand

Case study for quality control & quality assurance in metal additive manufacturing: A dental bridge

"Designing a fixed partial denture without a pontic"- Case report

Thakur H et al.applicability of various Mixed Dentition analysis among Sriganganagar School children

In Vitro Comparative Study of Retention of Maxillary Kennedy Class I Removable Partial Denture Retained with Four Different Designs of Esthetic Clasps

Frequency of the Various Classes of Removable Partial Dentures and Selection of Major Connectors and Direct/Indirect Retainers

International Journal of Medical Research & Health Sciences

Evaluation Of Complication Rate And Patient Satisfaction With Removable Denture

SURVEYING OF REMOVABLE PARITAL DENTURES FEB, 11, 2015

The Reinforced Removable Retainer

INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY

The costs and benefits of prosthodontic interventions:

Dowel restorations Treatment with a post and core

product catalog Dentures Partial Dentures Implant Dentures Mouthguards

Selection and arrangement of teeth in rpd

Assessment of Root Resorption and Root Shape by Periapical and Panoramic Radiographs: A Comparative Study

Shear Bond Strength of Acrylic Teeth to Acrylic Denture Base after Different Surface Conditioning Methods

Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A Three-Dimensional Finite Element Analysis

CAUSE OF TECHNICAL FAILURES OF CONICAL CROWN-RETAINED DENTURE (CCRD): A CLINICAL REPORT

Chapter 12. Prosthodontics

Load-Deflection Characteristics of Superelastic Nickel-Titanium Wires

Government of Karnataka PARA MEDICAL BOARD

Unless otherwise noted, the content of this course material is licensed under a Creative Commons Attribution 3.0 License.

A Study to evaluate Some of the Esthetic Factors determining Attractive Smile

An Obturator With A Soft Touch: A Case Report

An Anterior Tooth Size Comparison in Unilateral and Bilateral Congenitally Absent Maxillary Lateral Incisors

Frame designed partial prosthesis for the rehabilitation of a maxillectomy patient

Surveying. 3rd year / College of Dentistry/University of Baghdad ( ) Page 1

DIAGNOSTIC/PREVENTIVE SERVICES

A Survey on the Current Trends in the Use of Different Core Buildup Materials and Luting Agents in General Dental Practice

DL 313 Removable Partial Dentures II

Assessment of level of satisfaction and problems in patients treated with fixed partial denture in Alkharj City (Kingdom of Saudi Arabia)

IJCRR Vol 05 issue 01 Section: Healthcare Category: Case Study Received on: 19/09/12 Revised on: 16/10/12 Accepted on: 09/11/12

REMOVABLE PROSTHODONTICS

The new NobelProcera system for clinical success: The next level of CAD/CAM dentistry Authors_ Hans Geiselhöringer & Dr Stefan Holst, Germany

Department of Orthodontics, Faculty of Dentistry, Mahidol University.

Original Article MATERIALS AND METHODS INTRODUCTION

Wirobond C+ Processing information for CAD/CAM-produced restorations. Partners in Progress

The load/deflection characteristics of thermally activated orthodontic archwires

Overdenture: A Way of Preventive Prosthodontics

Implant Restorations: A Step-By-Step Guide

Prosthodontic Rehabilitation of a Partially Edentulous Hemiglossectomy Patient: A Clinical Report

Telescopic overdenture - A case report

A clinical comparison of three aligning archwires in terms of alignment efficiency: A prospective clinical trial

Case Report INTRODUCTION CASE REPORT. Sneha Gada 1, Deepak Nallaswamy 2, Ashish R. Jain 2 * ABSTRACT

Reinforcing Heat-cured Poly-methyl-methacrylate Resins using Fibers of Glass, Polyaramid, and Nylon: An in vitro Study

Transverse Strength of Reinforced Denture Base Resin with Metal Wire and E-Glass Fibers

Transcription:

ORIGINAL RESEARCH Cobalt-Chromium 10.5005/jp-journals-10024-1836 and Nickel-titanium Clasps In vitro Assessment of Clasps of Cobalt-Chromium and Nickel-titanium Alloys in Removable Prosthesis 1 Mohammed Zaheer Kola, 2 Deepti Raghav, 3 Prince Kumar, 4 Fawaz Alqahtani 5 Mohammed S Murayshed, 6 Tushar Vithal Bhagat ABSTRACT Background: It has been since a long time that the use of basemetal alloys has significantly increased in the area of prosthetic science. One of the factors contributing for the increasing use of base-metal alloys is the increasing and high cost factor of noble metals. Although numerous materials have been tried since the recent past, Cobalt-Chromium (CoCr) alloys still form the most commonly used ones because of the numerous advantages they offer. Titanium alloys are also significantly being used in this field because of their high resistance strength and high corrosion resistance. Therefore, we aimed to evaluate the flexure strength and fatigue resistance of clasps made up of CoCr alloys and Nickel-titanium (NiTi) alloys. Materials and methods: We evaluated the retentive forces of CoCr and NiTi clasps two different sizes engaging retentive undercuts of different depths (0.25 and 0.50 mm). Mandibluar second premolar was prepared to frame the prosthesis for missing mandibular first and second molar. A total of 50 clasps were included in the study out of which 30 were NiTi clasps and 30 CoCr clasps. We evaluated the loading of the force at the rate of half Newton(N)/second at the central part of the specimen until fracturing of the specimen occurred and finally measurement of the modulus of elasticity was also done. Statistical analysis was carried out; unpaired t test was used for evaluating the level of significance. Results: The mean load needed for CoCr clasps was 0.7450 and for NiTi clasp was 0.6140 Kgf for producing a deflection of 0.25 mm. As far as flexibility is concerned, more value was seen in NiTi 1,4-6 Department of Prosthetic Dental Sciences, College of Dentistry, Prince Sattam Bin Abdulaziz University, Al-Kharj Kingdom of Saudi Arabia 2,3 Department of Prosthodontics, Shree Bankey Bihari Dental College, Ghaziabad, Uttar Pradesh, India Corresponding Author: Prince Kumar, Senior Lecturer Department of Prosthodontics, Shree Bankey Bihari Dental College, Ghaziabad, Uttar Pradesh, India, e-mail: princekumaronline@gmail.com group than CoCr group. For deflecting the clasp up to 0.50 mm, the mean load needed for CoCr clasps was 1.4102 and for NiTi clasp was 0.8260 Kgf. The results were statistically significant. While measuring the flexibility, more value (p < 0.05) was seen in NiTi group than in CoCr group. NiTi alloy clasps had mean flexural strength of 1640 MPa and modulus of elasticity of 32 GPa. Conclusion: For producing deflection of 0.25 and 0.50 mm, the mean loading force was higher for NiTi alloys. Although being increasingly used these days, NiTi alloys have limited use as compared with CoCr alloys due to high cost of titanium and certain other limitations. Keywords: Cobalt-chromium, Flexure strength, Nickel-titanium. How to cite this article: Kola MZ, Raghav D, Kumar P, Alqahtani F, Murayshed MS, Bhagat TV. In vitro Assessment of Clasps of Cobalt-Chromium and Nickel-titanium Alloys in Removable Prosthesis. J Contemp Dent Pract 2016;17(3):253-257. Source of support: Nil Conflict of interest: None INTRODUCTION It has been since 1980 that the base-metal alloy s application in removable and fixed dentures in the field of prosthetic dentistry has increased. High cost factor of noble metals contributes to base metal s increasing use. Although numerous alloys have been tried and used in the past, cobalt-chromium (CoCr) alloys are one of the best known. High strength and resistance of currently used CoCr alloys has evolved because of Elwood Haynes who did large work in this field and later identified and showed high increase in its strength by addition of molybdenum (Mo) and tungsten (W). 1 Cobalt-chromium alloys have numerous qualities, including high resistance and strength, compatibility to oral tissues, and reduction of overall weight of the substance. Construction of orthopedic parts for prosthetic purpose forms the main biomedical use of CoCr alloys. The Journal of Contemporary Dental Practice, March 2016;17(3):253-257 253

Mohammed Zaheer Kola et al Current dental use of these alloys along with NiCr alloys is for the fabrication of removable partial dentures. 2-4 Titanium alloys also have significant use in dental prosthesis. The properties responsible are its magnificent resistance strength to corrosion and compatibility to biological tissues. Shape memory and high elastic features are some of the beneficial properties being possessed by titanium-nickel (TiNi) alloys. Titanium-nickel alloys have a unique property of rebounding from elastic deformation and remaining constant till the proportional limit. 5-7 The stresses acting on the clasps of a removable partial denture should be as low as till the limit to avoid permanent deformation of the clasp. While during continuous insertion and removal of the clasp of RPD, some amount of fatigue occurs and this varies depending upon the material or alloy used. 8 Therefore, we compared the strength of clasps or removable partial dentures of NiTi and CoCr alloys. Therefore, the aim of the study was to evaluate the fatigue resistance of NiTi clasps when subjected to repeated constant deflections as compared with CoCr. MATERIALS AND METHODS In this in vitro study, we evaluated the retentive forces of two different clasp materials in two different sizes engaging retentive undercuts of different depths. Missing mandibular first and second molar was used for the study in which preparation of mandibular second premolar was done for surveying the complete metal crown. A total of 50 clasps were included in the study out of which 30 were NiTi clasps and 30 CoCr clasps. Investment of standard half round 0.8 mm wax pattern was done followed by CoCr casting. Finishing and polishing of the final clasp was done for evaluation. Fabrication of dies of size approximately 19.2 mm 3 was done followed by fabrication and adaptation of clasps was done. Casting was done using sprue former followed by flasking. All the NiTi clasps and CoCr clasps were grouped into four groups depending upon the amount of deflection produced (Table 1). Instron machine was used for measurement of the fatigue test. While carrying the fatigue test, two end points were assessed. Either 10 6 cycles or permanent deformation, which ever occurred first, was taken as the point of measurement. Statistical analysis was carried out; unpaired t test was used for evaluating the level of significance. Digital Vernier caliper was used for assessment of the flexure strength with ±0.01 mm accuracy. Universal testing machine was employed for carrying the bending test. Loading of the force at the rate of half Newton (N)/ second was done at the central part of the specimen until fracturing of the specimen occurred and finally measurement of the modulus of elasticity was done. 11 The underlying equational formula was used for evaluating the flexural strength: Flexural strength = 3FL/2BH 2 MPa 12 F: Max. load in N, L: Distance between the supports in mm, B; width of the specimen in mm, H: Height of the specimen in mm. The underlying equational formula was used for evaluating the flexural modulus: Flexural Modulus = 3F 1 L 3 /4BH 3 d MPa 12 F 1 : Load in N, d: Deflection in mm RESULTS For deflecting the clasp up to 0.25 mm, the mean load needed for CoCr clasps was 0.7450 and for NiTi clasp was 0.6140 Kgf (Table 2, Graph 1). The results were statistically significant. Table 2: Load needed for producing a 0.25 mm deflection in the clasp and the clasp s flexibility Load needed for producing a 0.25 mm deflection in the clasp (Kgf) Clasp s flexibility (mm/kgf) S: Significant Study group Mean SD t (unpaired t test) p-value Group I 0.7450 0.0764 18.758 0.0016 s Group III 0.6140 0.0622 Group I 0.4020 0.0360 17.956 0.0024 s Group III 0.6024 0.0812 Table 1: Nickel-titanium and Cobalt-chromium clasps divided into four study groups on the basis of amount of deflection produced Groups Clasp material No. of specimens (n) I CoCr 15 0.25 II CoCr 15 0.50 III NiTi 15 0.25 IV NiTi 15 0.50 Amount of deflection produced (mm) Graph 1: The mean loading force and SD for producing deflection of 0.25 mm in CoCr alloys (Group I) and NiTi alloys (Group III) 254

Cobalt-Chromium and Nickel-titanium Clasps As far as flexibility is concerned, more value (p < 0.05) was seen in NiTi group (0.6024 mm/kgf) than in CoCr (0.4020 mm/kgf) group (Table 2, Graph 2). For deflecting the clasp up to 0.50 mm, the mean load needed for CoCr clasps was 1.4102 and for NiTi clasp was 0.8260 Kgf (Table 3, Graph 3). The results were statistically significant. Graph 4: The mean flexibility and SD for producing deflection of 0.5 mm in CoCr alloys (Group II) and NiTi alloys (Group IV) Table 4: Flexure strength and modulus of elasticity of nickel-chromium Flexure strength Modulus of elasticity Mean value 1640 MPa 32 GPa Graph 2: The mean flexibility and SD for producing deflection of 0.25 mm in CoCr alloys (Group I) and NiTi alloys (Group III) Table 3: Load needed for producing a 0.50 mm deflection in the clasp and the clasp s flexibility Load needed for producing a 0.50 mm deflection in the clasp (Kgf) Clasp s flexibility (mm/kgf) S: Significant Study group Mean SD t (unpaired t test) p-value Group II 1.4102 0.0764 16.465 0.0023 S Group IV 0.8260 0.0622 Group II 0.3868 0.0154 19.485 0.0018 S Group IV 0.5980 0.0784 Fig. 1: Cobalt-chromium clasp While measuring the flexibility, more value (p-value < 0.05) was seen in NiTi group (0.5980 mm/kgf) than in CoCr (0.3868 mm/kgf) group (Table 3, Graph 4). The mean flexural strength of NiTi alloy clasps is 1640 MPa and modulus of elasticity is 32 GPa (Table 4). Cobalt-chromium clasps are shown in Figure 1 and casting machine used in the present study is shown in Figure 2. DISCUSSION Graph 3: The mean loading force and SD for producing deflection of 0.5 mm in CoCr alloys (Group II) and NiTi alloys (Group IV) Adequate flexibility is a feature that removable partial denture must possess for the success and retention of prosthesis. At the same time, abutment teeth should also be relieved from undue load and forces of the clasps occurring during the functional process of the prosthesis. The Journal of Contemporary Dental Practice, March 2016;17(3):253-257 255

Mohammed Zaheer Kola et al Fig. 2: Casting machine On the other side, too flexible clasps are also not successful due to inadequate retention provided by them when used for shallow undercuts. 9 In the present study, we evaluated the flexibility of the NiTi and CoCr clasps at 0.25 and 0.50 mm deflection by load. Selection of 0.25 mm of clasp was done, as they are most routinely used undercuts for CoCr clasps. For larger undercuts, patients demand placing of clasps more toward the gingival region because of esthetic reasons. As the CoCr clasps are stiffer in nature, it is difficult to use these alloy clasps for larger undercuts. 9 Shape memory and good elastic characteristics are some of the useful properties possessed by NiTi alloys. The alloy has a constant stress value up to proportional limit and the value remains constant even when the alloys rebounds from elastic deformation. 7 Fatigue limit of CoCr alloys has been reported in the range from 40 to 70K psi, in various studies. 10 Therefore, 10 6 loading cycles were used in the recent study. The amount of loading force that will produce a deflection of 0.25 and 0.5 mm is approximately equal to the retentive strength of the material for the given two undercuts. For producing equal amount of deflection (0.25 and 0.50 mm) in both the groups, the mean amount of load required was significantly higher for NiTi alloys. As far as flexibility is concerned, clasps made of NiTi alloys had higher flexibility than CoCr alloys indicating the higher retentive strength of NiTi alloys (Tables 2 and 3). Kim et al also compared the clasp retention of NiTi clasps with conventional CoCr clasps and concluded that even though for all the claps included in their study, there was minimal or very less alteration seen in the values of retentive forces in the NiTi clasps as compared with the conventional clasps, even after repeated cycling procedures. 12 On evaluating the modulus of elasticity, we found that NiTi alloys (elastic modulus; 32 GPa) had higher flexibility than CoCr alloy (elastic modulus; 190 GPa) clasps (Table 4). We also conducted fatigue test and found that only CoCr alloys of 0.50 mm deflection showed deformation, whereas CoCr alloy clasps of 0.25 mm deflection and NiTi alloy class of 0.25 and 0.50 mm deflection did not produce any deformation. Savitha et al evaluated the flexural strength and fatigue resistance of CoCr alloys as compared with acetal resin alloys and concluded that acetal resin clasps may provide less retention than cobalt chromium based alloys. 13 Because of their ability of evaluating the clinical stresses occurring, these tests of flexibility are of great significance. By applying a loading force at the rate of half Newton/second at the flexure specimen s center, measurement of the elasticity modulus and amount of maximum load that can be sustained will be done. From the above tests, it was found that mean flexural strength of NiTi was 1640 MPa and elasticity modulus was 32 GPa. Bates 14 in their study demonstrated that 0.15 mm deflection would be shown by a 10 mm long Cr-Co clasp nearby to their proportional limit. Morris et al also evaluated the permanent deformations in clasps tips and concluded that permanent deformation of the tip of the clasp by more than 0.25 mm may be considered as significant. 15 However, our each group had relatively small number of patients (n = 15). A further study with larger study groups is required for improving the quality of the clasps and the prognosis of the treatment. CONCLUSION Keeping in mind the shortcomings of the present study, it can be concluded that for producing deflection of 0.25 and 0.50 mm, the mean loading force was higher for NiTi alloys. Nickel-titanium alloys are increasingly becoming popular these days but due to high cost of titanium and certain other limitations, their use is still very limited in actual practice as compared to CoCr alloys. REFERENCES 1. Crook P. Corrosion of cobalt based alloys. In: Davis J, editor. Corrosion. Materials Park, OH: ASM International; 1987. p. 1. 657-700. 2. Evans EJ, Thomas IT. The in vitro toxicity of cobalt-chrome molybdenum alloy and its constituent metals. Biomaterials 1986 Jan; 7(1):25-29. 256

Cobalt-Chromium and Nickel-titanium Clasps 3. Craig RG, Hanks CT. Reaction of fibroblasts to various dental casting alloys. J Oral Pathol 1988 Aug;17(7):341-347. 4. Marti A. Cobalt-base alloys used in bone surgery. Injury 2000 Dec;31 Suppl 4:18-21. 5. Castleman LS, Motzkin SM, Alicandri FP, Bonawit VL. Biocompatibility of nitinol alloy as an implant material. J Biomed Mater Res 1976 Sep;10(5):695-731. 6. Speck KM, Fraker AC. Anodic polarization behavior of TiNi and Ti-6Al-4V in simulated physiological solutions. J Dent Res 1980 Oct;59(10):1590-1595. 7. Miura F, Mogi M, Ohura Y, Hamanaka H. The super-elastic property of the Japanese NiTi alloy wire for use in orthodontics. Am J Orthod Dentofacial Orthop 1986 Jul;90(1):1-10. 8. Vallittu PK, Kokkonen M. Deflection fatigue of cobaltchromium, titanium, and gold alloy cast denture clasp. J Prosthet Dent 1995 Oct;74(4):412-419. 9. Mcgiveney GP, Carr AB. Mccracken s removable partial prosthodontics. 10th ed. Elsevier Science; 1999. p. 258-265. 10. Gapido CG, Kobayashi H, Miyakawa O, Kohno S. Fatigue resistance of cast occlusal rests using CoCr And Ag-Pd-Cu-Au alloys. J Prosthet Dent 2003 Sep;90(3):261-269. 11. Anusavice KJ. Phillips science of dental materials. 11th ed. St. Louis: Elsevier; 2003. 12. Kim D, Park C, Yi Y, Cho L. Comparison of cast TiNi alloy clasp retention with conventional removable partial denture clasps. J Prosthet Dent 2004 Apr;91(4):374-382. 13. Savitha PN, Lekha KP, Nadiger RK. Fatigue resistance and flexural behavior of acetal resin and chrome cobalt removable partial denture clasp: an in vitro study. Eur J Prosth 2015; 3(3):71-76. 14. Bates JF. The mechanical properties of the cobalt-chromium alloys and their relation to partial denture design. Br Dent J 1965 Nov 2;119(9):389-396. 15. Morris HF, Asgar K, Brudvik JS, Winkler S, Roberts EP. Stress-relaxation testing. Part IV: Clasp pattern dimensions and their influence on clasp behavior. J Prosthet Dent 1983 Sep;50(3):319-326. The Journal of Contemporary Dental Practice, March 2016;17(3):253-257 257