Identification and Quantitation of Microcystins by Targeted Full-Scan LC-MS/MS

Similar documents
Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

Impurity Profiling of Carbamazepine by HPLC/UV

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

Increased Identification Coverage and Throughput for Complex Lipidomes

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

LC-MS/MS Method for the Determination of Tenofovir from Plasma

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Integrated Targeted Quantitation Method for Insulin and its Therapeutic Analogs

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

Using Multiple Mass Defect Filters and Higher Energy Collisional Dissociation on an LTQ Orbitrap XL for Fast, Sensitive and Accurate Metabolite ID

Enhancing Sequence Coverage in Proteomics Studies by Using a Combination of Proteolytic Enzymes

Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection

Use of a Tandem Mass Spectrometry Research Method for the Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Determination of Amantadine Residues in Chicken by LCMS-8040

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

PIF: Precursor Ion Fingerprinting Searching for a Structurally Diagnostic Fragment Using Combined Targeted and Data Dependent MS n

Dynamic Analysis of HIV-Human Protein-Protein Interactions During Infection

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode

Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC

Sample Concentration and Analysis of Human Hormones in Drinking Water

IC-MS Environmental Applications - Water Testing. Application Notebook

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, San Jose, CA, USA

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT.

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

Quantitation of Protein Phosphorylation Using Multiple Reaction Monitoring

[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection

Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer

Characterization of an Unknown Compound Using the LTQ Orbitrap

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Agilent 6490 Triple Quadrupole LC/MS System with ifunnel Technology. Ultra sensitive quantitative performance

Thermo Scientific LipidSearch Software for Lipidomics Workflows. Automated Identification and Relative. Quantitation of Lipids by LC/MS

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Analysis of Limonin in Citrus Juice Using QuEChERS and LC-MS/MS

Essential Lipidomics Experiments using the LTQ Orbitrap Hybrid Mass Spectrometer

PTM Discovery Method for Automated Identification and Sequencing of Phosphopeptides Using the Q TRAP LC/MS/MS System

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method

Abstract. Introduction

Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer

Application Note. Author. Abstract. Introduction. Food Safety

Testing robustness: Immunosuppressant drugs in blood with a TSQ Quantis MS for clinical research

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples

Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

A Fully Integrated Workflow for LC-MS/MS Analysis of Labeled and Native N-Linked Glycans Released From Proteins

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Authors. Abstract. Introduction. Environmental

Mercury Speciation Determinations in Asian Dietary Supplements

Robust and Fast Analysis of Tobacco-Specific Nitrosamines by LC-MS/MS

Fast, Robust and Reliable Method for the Identification and Quantitation of Sildenafil Residue in Honey using LC-MS/MS

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Supporting information

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Determination of Fusarium Mycotoxins in Wheat, Maize and Animal Feed Using an Online TurboFlow and Orbitrap LC/MS Method

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine-

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Key Advantages of Comprehensive Cannabis Analysis

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Screening and Speciation of Raw and Processed Meat Products

Journal of Chemical and Pharmaceutical Research

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

Determination of Tigecycline in a Cell Lysate

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

MALAYSIAN MALAYSIAN COCOA WORKSHOP ON THE SAFE USE OF PESTICIDES IN COCOA AND HARMONIZED

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Designer Cannabinoids

FT-Raman Surface Mapping of Remineralized Artificial Dental Caries

Transcription:

Identification and Quantitation of Microcystins by Targeted Full-Scan LC-MS/MS Terry Zhang, Reiko Kiyonami, Leo Wang and Guifeng Jiang Thermo Fisher Scientific, San Jose, CA, USA Application Note 569 Key Words Velos Pro, UltiMate, Water Analysis, Cyanobacteria, Microcystin Goal Develop a simple and sensitive LC-MS method for definitive identification and quantitation of microcystins in water. Introduction Cyanobacteria, commonly referred to as blue-green algae, are photosynthetic prokaryotes that occur naturally in surface waters. They contribute significantly to primary production and nutrient cycling. Eutrophic, warm and low turbulent conditions in freshwater bodies typically promote the dominance of cyanobacteria within phytoplankton communities. Excessive proliferation of cyanobacteria leads to blooms that disrupt ecosystems, adversely affect the taste and odor of water, and increase water treatment costs. Blooms of toxic cyanobacteria species in surface drinking water sources and recreational waters threaten human health. Gastrointestinal illness, skin irritation, and death following renal dialysis have been attributed to acute cyanotoxin exposure. Chronic exposure can cause liver damage and may be associated with primary liver cancer. 1 The incidence and severity of cyanobacterial blooms are increasing globally, underscoring the importance of cyanotoxin monitoring. The most commonly encountered cyanotoxins are the microcystins, a group of hepatotoxic cyclic heptapeptides produced by various genera of cyanobacteria, including Microcystis, Planktothrix, and Anabaena. The chemical structure of a microcystin, depicted in Figure 1, is characterized by the presence of the amino acid 3-amino- 9-methoxy-2,6,8-trimethyl-10-phenyl-deca-4,6-dienoic acid (Adda), which modulates the biological activity of these toxins, and N-methyldehydroalanine (Mdha). Microcystin nomenclature is based on the L-amino acids present at two positions (X and Y in Figure 1) in the molecule. Over 80 structural variants are known, differentiated by the two variable L-amino acids as well as by chain modifications. The inhibition of serine/threonine protein phosphatases type 1 and 2A is considered the major mechanism of microcystin toxicity. Microcystin-LR, one of the most prevalent and potent microcystins, is designated as possibly carcinogenic to humans by the International Agency for Research on Cancer (IARC). 2 The potential risk of chronic exposure to microcystins in drinking water supplies prompted the World Health Organization (WHO) to issue a provisional guideline of 1 μg/l as the maximum concentration of total microcystin-lr (free plus cell-bound) in drinking water. 3 Many national and regional governments have since adopted this guideline value directly or have established slightly modified variants. Figure 1. The chemical structure of MC-LR contains leucine (L) and arginine (R) at positions X and Y, respectively. Microcystin nomenclature is based on the L-amino acids present at these two positions.

2 A toxic cyanobacterial bloom usually consists of multiple microcystin congeners in varying concentrations. Several techniques for the analysis of microcystins have been developed. Mouse bioassays, protein phosphatase inhibition assays, and enzyme-linked immunosorbent assays (ELISA) are effective for rapid screening but lack specificity. Reversed-phase high-performance liquid chromatography (HPLC) with ultraviolet (UV) detection is the most common approach used for the separation, detection and quantitation of microcystins. An ISO method for microcystin analysis by HPLC-UV has been validated for MC-RR, MC-YR and MC-LR. 4 However, UV detection is susceptible to interferences from water matrices and requires sample cleanup and concentration to achieve desirable detection limits. Furthermore, UV-based methods do not provide unequivocal identification of known microcystins nor enable identification of unexpected variants. Liquid chromatography in combination with multi-stage mass spectrometry (LC-MS n ) enables structural characterization and unambiguous identification of trace levels of microcystins. LC-MS/MS in multiple reaction monitoring (MRM) acquisition mode allows highly selective and sensitive quantitation and confirmation of target microcystins, but this approach requires extensive compound-dependent parameter optimization and cannot be used to detect unexpected toxins. Full-scan MS/MS approaches obviate the need for compound optimization and enable determination of all microcystins present in a sample. The Thermo Scientific Velos Pro dual-pressure linear ion trap mass spectrometer delivers sensitivity and speed for qualitative and quantitative applications. High-quality full-scan MS n spectra enable confident structural elucidation and identification. Rapid scanning and fast cycle times generate more scans across chromatographic peaks for robust quantitation and allow the acquisition of more MS n spectra in shorter chromatographic runs. A wide dynamic range of up to six orders of magnitude facilitates identification and quantitation of low-abundance compounds in complex matrices. Complementary fragmentation techniques may be performed in parallel to enable more MS n information to be obtained from a single sample. In this application note, we describe a simple and sensitive targeted full-scan LC-MS/MS method for the identification and quantitation of the microcystins MC-RR, MC-YR, and MC-LR using the Velos Pro ion trap mass spectrometer coupled to a Thermo Scientific Dionex UltiMate 3000 x2 Dual RSLC system. Experimental Sample Preparation MC-RR, MC-YR and MC-LR standards were purchased from Sigma-Aldrich. A stock solution of a mixture of these three microcystins was prepared at a concentration of 5 µg/ml. Calibration solutions, with concentrations of 0.025 µg/l to 50 µg/l, were prepared by serial dilution of the stock solution. LC-MS/MS Analysis A 50 µl sample was injected on a Thermo Scientific Acclaim 120 guard cartridge with 150 L/min, washed for two minutes to waste and then eluted onto a Thermo Scientific PepMap100 analytical column for separation. LC-MS/MS analysis was performed on an UltiMate 3000 x2 Dual RSLC system coupled to an Velos Pro mass spectrometer. LC Parameters Guard cartridge: Acclaim 120 C18 (10 x 3.0 mm i.d., 5.0 µm particle size, 120 Å pore size) Analytical column: Acclaim PepMap100 C18 (150 x 1.0 mm i.d., 3.0 µm particle size, 100 Å pore size) Mobile Phase A: Water containing 0.1% formic acid Mobile Phase B: Acetonitrile containing 0.1% formic acid Column temperature: 40 C Sample injection volume: 50 µl Flow rate: 150 µl/min Gradient: Table 1 Table 1: LC Gradient Time % A % B 0.1 98 2 1.5 98 2 2.0 80 20 3.0 60 40 7.4 40 60 7.5 2 98 7.9 2 98 MS Parameters Ionization mode: Positive electrospray ionization (ESI) Collision energy: 35% Isolation window: 2 Targeted full-scan MC-RR [M+2H] 2+ at m/z 520 [m/z 150-1100] MS/MS: MC-YR [M+H] + at m/z 1045 [m/z 285-1100] MC-LR [M+H] + at m/z 995 [m/z 285-1100]

Results and Discussion 3 Structural Identification and Confirmation Figure 2 shows extracted ion chromatograms and MS/MS spectra obtained from full-scan LC-MS/MS analysis of a mixture containing MC-RR, MC-YR and MC-LR at concentrations of 0.5 µg/l. MC-RR, MC-YR and MC-LR eluted at 5.62, 6.85, and 6.93 minutes, respectively. The MS/MS spectrum of MC-RR was generated by collisioninduced dissociation (CID) of the [M+2H] 2+ ion and is characterized by major fragment ions at m/z 505, 452 and 887, which correspond to [M+2H-CO] 2+, [M+2H-C 9 H 10 O] 2+ and [M+H-C 9 H 10 O-NH 3 ] +, respectively (C 9 H 10 O is a fragment of the Adda residue). The closely eluting compounds MC-YR and MC-LR are easily distinguished by their MS/MS spectra. The MS/MS spectrum of MC-YR, generated by CID of the [M+H] + ion, contains major fragment ions at m/z 1017, 599, and 916, which correspond to [M+H-CO] +, [Arg+Adda+Glu + H] +, and [Arg+Adda+Glu+Mdha+Ala+Tyr + H] +, respectively. The CID MS/MS spectrum of the [M+H] + ion of MC-LR is characterized by major fragment ions at m/z 967, corresponding to [M+H-CO] + ; m/z 599, corresponding to [Arg+Adda+(Glu or MeAsp) + H] + ; m/z 866, corresponding to [Ala+Adda+Arg+(Glu or MeAsp) +Leu+Mdha + H] + ; and m/z 553, corresponding to [Ala+Arg+(Glu or MeAsp) +Leu+Mdha+ H] +. MC-RR MC-YR MC-LR Figure 2. Extracted ion chromatograms and MS/MS spectra for MC-RR, MC-YR and MC-LR at concentrations of 0.5 µg/l

4 Quantitative Analysis The high scan speeds and fast analytical cycle time of the Velos Pro mass spectrometer enabled higher numbers of analytical scans across chromatographic peaks for optimal quantitative reliability (Figure 3). Excellent linearity in detector response was observed over the range of 0.05-50 µg/l for all three microcystins. Calibration curves for MC-RR, MC-YR and MC-LR are shown in Figure 4, with coefficients of determination of 0.9986, 0.9994, and 0.9994, respectively. The lowest detectable amount (LOD) of 0.025 µg/l and quantifiable amount (LOQ) of 0.05 µg/l were achieved for each microcystin. Both QC samples, at levels of 0.5 and 5 µg/l, achieved quantitation accuracy better than 94% for all three microcystins. Signal-to-noise ratios of >25 with automatic ICIS algorithm integration in Thermo Scientific Xcalibur software were obtained for MC-LR at the LOQ (Figure 5), demonstrating that this LC-MS/MS method can be used to determine MC-LR at concentrations well below the WHO s recommended guideline level of 1 µg/l. Method reproducibility was investigated by analyzing five replicate injections of each analyte. Peak area RSDs for MC-LR and MC-YR were less than 7% and 11%, respectively, over the entire linear dynamic range (Table 2). For MC-RR, peak area RSDs over the range 0.10-50 µg/l were under 6%; at the LOQ, the peak area RSD was 16% (Table 2). Retention time precisions were 0.3% RSD or less over the entire dynamic range (Figure 6) for all three microcystins. Tap water, filtered water and surface pond water were analyzed using this method. No microcystins were in any of the three water sources. Figure 4. Calibration curves for quantitation of MC-RR, MC-YR and MC-LR Figure 3. High scan speeds and fast cycle times enable more than 20 data points to be acquired across the MC-LR chromatographic peak.

5 Figure 5. For MC-LR at the LOQ (0.05 µg/l), S/N > 25 and peak area RSD = 6.91% 0.025 µg/l RT 0.29% RSD 0.1 µg/l 0.2 µg/l 0.5 µg/l 1 µg/l 5 µg/l 10 µg/l 50 µg/l Figure 6. High retention-time precision (< 0.3% RSD) over a wide linear dynamic range

Table 2. Peak area precision (from five replicate injections) for LC-MS/MS assay of MC-RR, MC-YR and MC-LR Levels µg/l MC-RR MC-YR MC-LR 0.05 16.01 10.5 6.91 0.10 2.82 5.88 3.97 0.20 3.54 5.25 4.89 0.50 4.86 8.54 3.03 1.00 5.84 1.76 4.25 5.00 2.28 2.13 2.47 10.00 4.54 1.30 1.31 50.00 2.40 1.76 2.66 Conclusion A simple, sensitive and robust LC-MS method for quantitative determination of microcystins was developed. Targeted full-scan MS/MS analysis using the LTQ Velos Pro linear ion trap mass spectrometer provided excellent selectivity and sensitivity for the identification and quantitation of MC-RR, MC-YR and MC-LR across a wide linear dynamic range. The LOD and LOQ were 0.025 µg/l and 0.05 µg/l, respectively. The LOQ was significantly lower than the provisional guideline value established by the WHO for MC-LR concentrations in drinking water. Assays performed in full-scan MS/MS mode enable compound confirmation and quantitation without the need for compound-dependent parameter optimization. The method was used to analyze tap, filtered and surface pond water. No microcystins were detected from these three water sources. References 1. Toxic cyanobacteria in water: A guide to their public health consequences, monitoring and management; Chorus I, Bartram J, eds. Published by E & FN Spon, London, on behalf of the World Health Organization, Geneva. 1999 2. Ingested Nitrate and Nitrite and Cyanobacterial Peptide Toxins. IARC Monogr Eval Carcinog Risks Hum, 2010, 94:1 477. 3. WHO Guidelines for Drinking-water Quality, 3rd Ed., Vol. 1, Recommendations; WHO, Geneva. 2004 4. ISO Water Quality: Determination of microcystins - method using solid phase extraction (SPE) and high performance liquid chromatography (HPLC) with ultraviolet (UV) detection. ISO, Geneva, Switzerland (ISO 20179:2005). Application Note 569 www.thermofisher.com 2016 Thermo Fisher Scientific Inc. All rights reserved. Sigma-Aldrich is a registered trademark of Sigma-Aldrich Biotechnology L.P. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa-Other +27 11 570 1840 Australia +61 3 9757 4300 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland/Norway/Sweden +46 8 556 468 00 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9434 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Russia/CIS +43 1 333 50 34 0 South Africa +27 11 570 1840 Spain +34 914 845 965 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752 AN63631_E 07/16S