Lower Cutoffs for LC-MS/MS Urine Drug Testing Indicates Better Patient Compliance

Similar documents
A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography

Validation of an Automated Method to Remove

PROFESSIONALISM AND COMMENTARY. Optimizing Urine Drug Testing for Monitoring Medication Compliance in Pain Management

3703 Camino del Rio South 100-A San Diego, CA, Phone Fax CLIA# 05D years

Using Liquid Chromatography Tandem Mass Spectrometry Urine Drug Testing to Identify Licit and Illicit Drug-Use in a Community-based Patient Population

A Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS

LCMS-8050 Drugs of Abuse: 113 Analytes with Polarity Switching

Illicit Drug Use Correlates with Negative Urine Drug Test Results for Prescribed Hydrocodone, Oxycodone, and Morphine

Rapid and Sensitive Analysis of a 93-Compound Forensic Panel in Urine using the QTRAP /Triple Quad 4500 LC- MS/MS System

nextgen precision Test Report

1/27/ New Release, Quest Diagnostics Nichols Institute, Valencia

September HCMC Toxicology Transition: Additional information and Frequently Asked Questions

The Drug Testing Process. Employer or Practice

Direct Analysis of Urinary Opioids and Metabolites by Mixed-Mode µelution SPE Combined with UPLC/MS/MS for Forensic Toxicology

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

Fully Automated Online Sample Preparation and LC-MS/MS Analysis of Drugs of Abuse in Oral Fluids

Improving Benzodiazepine Immunoassay Sensitivity by Rapid Glucuronide Hydrolysis Technology

One Source Toxicology Laboratory, 1213 Genoa Red Bluff, Pasadena, Texas 77504

Interpretation of Pain Management Testing Results Using Case Examples

Drug Adherence Assessment Report

Modernizing the Forensic Lab with LC-MS/MS Technology

Validation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens

Laboratory Testing to Support Pain Management: Methods, Concepts and Case Studies

Fast and easy separation of 23 drugs of abuse. including high, stable resolution of isobaric opioids from human urine by UHPLC-MS/MS

Rapid LC/TOF MS for Analytical Screening of Drugs in the Clinical Research Lab

Ultra-Fast Forensic Toxicological Screening and Quantitation under 3 Minutes using SCIEX X500R QTOF System and SCIEX OS 1.

METHAMPHETAMINE & AMPHETAMINE

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Medical Affairs Policy

Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System

Rapid Hydrolysis of Benzodiazepines in Urine. Alicia Zook 1 and Crystal Xander B.S. 2. Cedar Crest College, Allentown, PA 1

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Immunoassay-Based Drug Tests Are Inadequately Sensitive for Medication Compliance Monitoring in Patients Treated for Chronic Pain

Agilent Clinical LC/MS: Review of Established LC-MS QQQ Clinical Research Applications. dr. Jan Srbek, HPST. Page 1

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

Comparison of Different Whole Blood Sample Pretreatment Methods for Targeted Analysis of Basic Drugs

Validation of a Benzodiazepine and Z-Drug Method Using an Agilent 6430 LC/MS/MS

3703 Camino del Rio South 100-A San Diego, CA, Phone Fax CLIA# 05D Director: David J.

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

[ APPLICATION NOTE ] APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

Overview. Introduction. Experimental. Cliquid Software for Routine LC/MS/MS Analysis

PAIN PANEL. Research and Solutions

Pain Management and Compliance Toxicology. Greg Jellick, MSFS, D-ABFT-FT Technical Director Quality Toxicology San Antonio, TX

APPLICATION TN Rapid Analysis of a 48 Drug Research Panel in Less Than 4 Minutes Using Kinetex Biphenyl Core-Shell Technology HPLC/UHPLC Columns

executive summary Specifically, this paper provides: 2 precision diagnostics Statistics and trends on prescription drug abuse, misuse, and diversion

URINE DRUG TOXICOLOGY

Automated Targeted screening of Benzodiazepines in urine using LDTD-MS/MS at 400 samples per hour rate

Clinical Policy: Outpatient Testing for Drugs of Abuse Reference Number: PA.CP.MP.50

E XCEL LENCE JUST GOT BETTER UCT FORENSICS CLEAN SCREEN XCEL SPE COLUMNS

APPLICATIONS TN Analytical Method

Pain Medication Management Program Monitors Patient Compliance

LC Application Note. Dangerous driver?

NEW. Kinetex. Biphenyl TMS TMS TMS TMS. Selectivity that a C18 just can t give you!

Pain Medication Management Program Supports Patient Outcomes and Adherence

Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations

Evaluation of the Impact of Expanding ELISA Screening in DUID Investigations. Aileen Lu*, Karen S. Scott, Aya Chan-Hosokawa, and Barry K.

A FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ

3/8/2018. Reasons for Doing UDT. UDT: A Tool in Risk Assessment. Faculty/Presenter Disclosure. Urine Drug Testing in Chronic Pain Management

Urine Drug Screening: The Essentials of Interpretation

One of These Tests is Not Like the Other: Comparative effectiveness, cost-effectiveness and utilization guidance in pain management testing

Urine Drug Testing PracticeNotes Clinical Guide

Drug Screening: Things You Need to Know

Detection of Neonatal Drug Exposure Using Umbilical Cord Tissue and Liquid Chromatography Time-of-Flight Mass Spectrometry

Fast and simultaneous analysis of ethanol metabolites and barbiturates using the QTRAP 4500 LC-MS/MS system

Simultaneous Determination of Prescription and Designer Benzodiazepines in Urine and Blood by SPE and LC-MS/MS

Analysis and interpretation of drug testing results from patients on chronic pain therapy: a clinical laboratory perspective

Drugs of Abuse in Oral Fluids: Automated SPE Extraction and LC/MS/MS Determination using a Robotic Autosampler

Toxicology Testing Is it a Good Fit for Your Laboratory? Rebecca Kenner, BA, MT, DLM(ASCP) Leigh Ann Smith, BS, MLS(ASCP), CLS

Laboratory Service Report

Urine Drug Testing. Methadone/Buprenorphine 101 Workshop. Ron Joe, MD, DABAM December 10, 2016

A New Highly Specific Buprenorphine Immunoassay for Monitoring Buprenorphine Compliance and Abuse

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

Substance abuse is a significant problem in the United

Screening by immunoassay and confirmation & quantitation by GC-MS of buprenorphine and norbuprenorphine in urine, whole blood and serum

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Pain Management Drug Testing: A Laboratory Perspective

Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer

Community Drug Early Warning System (CDEWS-3): Washington, DC - Site 3 of 4

3703 Camino del Rio South 100-A San Diego, CA, Phone Fax CLIA# 05D years

Powering Efficient Behavioral Health Care Services

Evaluation of the Integrated E-Z Split Key Cup II for Rapid Detection of Twelve Drug Classes in Urine

Oral Fluid Drugs of Abuse Testing

Urine Drug Testing to Monitor Opioid Use In Managing Chronic Pain

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

BETA-GLUCURONIDASE PRODUCT LINE INNOVATION THROUGH CHEMISTRY. FORENSICS

EDUCATIONAL COMMENTARY METHADONE

Jonathan Rochon, Pier-Luc Plante, Serge Auger, Réal Paquin, Jacques Corbeil, Pierre Picard

Workflow for Screening and Quantification of the SAMHSA (NIDA) Panel in Urine Using UHPLC-TOF

Alcohol. Ethanol Highlands Parkway, Suite 100 Smyrna, GA 30082

Applications of High Resolution Mass Spectrometry in Forensic Toxicology. Patrick Kyle, PhD.

ORAL FLUID AS A CHEMICAL TEST FOR THE DRE PROGRAM : HISTORY, THE FUTURE, AND PRACTICAL CONSIDERATIONS

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Proof of Concept for Automated SPE/HPLC/MS/MS Methods to Replace Traditional Immunoassay with MS Confirmation of Driving Under the Influence Samples

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

Determination of Selected Illegal Drugs and its Important Metabolites in Waste Water by Large Volume Direct Injection HPLC-MS/MS

Physician s Reference for Urine and Blood Drug Testing and Interpretation

Disclosures. You're in Control or Urine Control Clinical Pearls of Drug Testing Case Studies. 9/20/17

Linking Opioid Treatment in Primary Care. Roxanne Lewin M.D.

Abstract. Introduction

Transcription:

Pain Physician 2017; 20:E1107-E1113 ISSN 2150-1149 Retrospective Study Lower Cutoffs for LC-MS/MS Urine Drug Testing Indicates Better Patient Compliance Kevin Krock, PhD, Amadeo Pesce, PhD, Dennis Ritz, MS, JD, Richard Thomas, PhD, Agnes Cua, PhD, Ryan Rogers, BS, Phil Lipnick, BS, and Kristen Kilbourn, BS From: Precision Diagnostics, San Diego, CA Address Correspondence: Kevin Krock, PhD Precision Diagnostics 4215 Sorrento Valley Blvd San Diego, CA 92121 Email: kevin.krock@precisiondxlab.com Disclaimer: There was no external funding in the preparation of this manuscript. Conflict of interest: Each author certifies that he or she, or a member of his or her immediate family, has no commercial association (i.e., consultancies, stock ownership, equity interest, patent/licensing arrangements, etc.) that might pose a conflict of interest in connection with the submitted manuscript. Manuscript received: 12-02-2016 Accepted for publication: 06-13-2017 Free full manuscript: Background: Urine drug testing is used by health care providers to determine a patient s compliance to their prescribed regimen and to detect non-prescribed medications and illicit drugs. However, the cutoff levels used by clinical labs are often arbitrarily set and may not reflect the urine drug concentrations of compliant patients. Objectives: Our aim was to test the hypothesis that commonly used cutoffs for many prescribed and illicit drugs were set too high, and methods using these cutoffs may yield a considerable number of false-negative results. The goals of this study were to outline the way to analyze patient results and estimate a more appropriate cutoff, develop and validate a high sensitivity analytical method capable of quantitating drugs and metabolites at lower than the commonly used cutoffs, and determine the number of true positive results that would have been missed when using the common cutoffs. Study Design: This was a retrospective study of urine specimens submitted for urine drug testing as part of the monitoring of prescription drug compliance described in chronic opioid therapy treatment guidelines. Setting: The study was set in a clinical toxicology laboratory, using specimens submitted for routine analysis by health care providers in the normal course of business. Methods: Lognormal distributions of test results were generated and fitted with a trendline to estimate the required cutoff level necessary to capture the normal distributions of each drug for the patient population study. A validated laboratory derived liquid chromatography tandem mass spectrometry (LC-MS/MS) analysis capable of achieving the required cutoff levels was developed for each drug and/or metabolite. Results: The study shows that a lognormal distribution of patient urine test results fitted with a trendline is appropriate for estimating the required cutoff levels needed to assess medication adherence. The study showed a wide variation in the false-negative rate, ranging from 1.5% to 94.3% across a range of prescribed and illicit drugs. Limitations: The patient specimens were largely sourced from patients in either a long-term pain management program or in treatment for substance use disorder in the US. These specimens may not be representative of patients in other types of treatment or in countries with different approaches to these issues. Conclusions: The high-sensitivity method reduces false-negative results which could negatively impact patient care. Clinicians using less sensitive methods for detecting and quantifying drugs and metabolites in urine should exercise caution in assessing patient adherence using and changing the treatment plan based on those results. Key words: Urine drug testing, patient adherence, clinical toxicology, immunoassay, LC-MS, definitive drug testing, REMS, negative test results, false negative Pain Physician 2017; 20:E1107-E1113

Pain Physician: November/December 2017: 20: E1107-E1113 It has been well-established that urine drug testing is an important component of treatment plans for patients on chronic opioid therapy (1-6). Much has been written about false-positive results (7,8), however in the clinical setting, false-negative results are just as important (8). Physicians monitoring patients on chronic opioid therapy presume that the cutoffs provided by their laboratory are like those reference intervals for specific analytes to establish possible health and disease states of a patient (9-11). Nevertheless, there are no such standards to test for compliance in patients being monitored for illicit drugs, pain, anxiety, or other scheduled medications. Most laboratories use cutoffs to determine patient compliance that have been established by the manufacturers of immunoassay reagents or the sensitivity of liquid chromatography tandem mass spectrometry (LC-MS/MS) instrumentation (12-15). In part, these cutoffs reflect those established for workplace testing or availability from immunoassay manufacturers. Complicating a simple interpretation of the drug test is the estimation that half of all medications are not taken as prescribed (16), and patients on chronic opioid therapy often take these medications depending on how they feel (17). This work was done to test whether clinically validated lower cutoffs would better describe patient compliance and detect illicit drug use. We have described an algorithm using the frequency distribution of urine drug concentrations and suggest that this may be used to estimate appropriate cutoffs (18-20). By analogy to the usual clinical chemistry analytes, we chose to estimate if this mathematical model could estimate the lower cutoff. The problem in establishing such cutoffs is that pain patients take varying amounts of medication depending on how they feel or are prescribed. Because of this wide range of concentration values, transformation of the data to convert it to an approximation of a Gaussian curve is used. It has been posited that 5 requirements must be met to identify appropriate cutoffs for pain management drugs and their metabolites. First, an analytical procedure that covers the wide dynamic range of urine excretion of these drugs and their metabolites must be developed and validated (12,13,21). Second, a large database of test results on this patient population must be obtained (22). Third, the urine samples must be clinically validated using creatinine and other validation tests (23). Fourth, mathematical models must be used to calculate the cutoffs (18-20). Finally, the results must be consolidated to provide the best estimates (18-20). Our laboratory recently validated a more sensitive quantitative method for monitoring these drugs and metabolites. We compared the results of the highsensitivity method to what would be reported using our previous cutoff levels, which in most cases reflected those commonly used by other laboratories to calculate a false-negative rate. Methods This is a retrospective study derived from specimens sent for urine drug testing from pain physician offices and rehabilitation centers. This study was approved by Aspire IRB (Santee, CA). A high-sensitivity LC-MS/MS method was developed and validated, capable of quantitating low concentrations of 71 drugs and metabolites, which is important for monitoring medication adherence and substance use disorder. A 20-XR series binary pump system (Shimadzu, Kyoto, Japan), well-plate autosampler, and temperature controlled column oven was paired with a 6500 triple quadrupole mass spectrometer (Sciex, Framingham, MA). LC-MS grade water was obtained from an ultrapure water system (Sartorius, Bohemia, NY), LC-MS grade methanol was obtained from EMD Millipore (EMD Millipore, Billerica, MA), and LC-MS grade formic acid was obtained from Covachem (Covachem, Loves Park, IL). A Kinetex phenyl-hexyl column (Phenomenex, Torrance, CA) with dimensions of 50 x 4.6 mm and 2.6 µm particle size was used for chromatographic separation. The binary pump system delivered a mixture of the mobile phases at the proportion and flow rate that is displayed in Table 1. Mobile phase A (MPA) was 0.1% formic acid in LC-MS grade water and mobile phase B (MPB) was LC-MS grade methanol containing 0.1% formic acid. The mass spectrometers used the following settings common to all analytes: curtain gas of 35 L/min, collision gas of 10 L/min, positive mode IonSpray voltage of 2500 V, source temperature of 450 C, ion source gas 1 of 60 L/min, and ion source gas 2 of 50 L/min. Two transitions for each analyte were optimized for declustering potential collision cell energy and exit potential. Analytes and internal standards were obtained from Cerilliant (Cerilliant, Round Rock, TX). Four-point calibration curves were prepared from the cutoff level to 30 times the cutoff level for each analyte. Two quality control samples were analyzed with each batch of specimens to ensure acceptability of results. E1108

Lower Cutoffs for LC-MS/MS Urine Drug Testing Indicates Better Patient Compliance Table 1. The false-negative rate of analytes across drug classes. Substance Category n High-Sensitivity Cutoff (ng/ml) Low-Sensitivity Cutoff (ng/ml) False-Negative Rate Amitriptyline Antidepressants 431 10 100 14.6% Nortriptyline Antidepressants 456 10 100 23.9% Venlafaxine Antidepressants 397 2 50 9.8% 9-Hydroxyrisperidone Antipsychotic 583 5 50 24.7% Quetiapine Antipsychotic 1155 5 50 9.7% Norquetiapine Antipsychotic 1142 25 50 2.5% Clonazepam Benzodiazapines 1343 5 50 94.3% 7-Aminoclonazepam Benzodiazapines 2016 5 50 30.1% Alprazolam Benzodiazapines 3378 5 50 39.1% Alpha-Hydroxyalprazolam Benzodiazapines 3736 5 50 29.0% Lorazepam Benzodiazapines 1300 10 50 19.5% Nordiazepam Benzodiazapines 5473 5 50 14.6% Oxazepam Benzodiazapines 3315 10 50 22.4% Temazepam Benzodiazapines 2559 10 50 21.0% Benzoylecgonine Illicit Drugs 4252 5 50 44.9% 6-Monoacetylmorphine Illicit Drugs 2183 5 25 19.3% Carisoprodol Muscle Relaxants 301 10 100 36.9% Cyclobenzaprine Muscle Relaxants 1193 5 100 44.8% Hydrocodone Natural Opiates 5667 5 50 32.9% Norhydrocodone Natural Opiates 5366 10 50 24.1% Hydromorphone Natural Opiates 9444 5 50 37.4% 6-Beta Naltrexol Opioid Inverse Agonist 305 10 50 4.9% Naloxone Opioid Inverse Agonist 2472 10 100 22.0% Zolpidem Sedatives 464 1 50 81.0% Buprenorphine Semi-Synthetic Opiates 15142 5 10 3.3% Norbuprenorphine Semi-Synthetic Opiates 10194 5 10 1.5% Oxycodone Semi-Synthetic Opiates 1241 10 50 10.6% Noroxycodone Semi-Synthetic Opiates 4488 25 50 4.7% Oxymorphone Semi-Synthetic Opiates 2973 10 50 11.9% Duloxetine SNRI/SSRI 373 10 50 29.0% Norfluoxetine SNRI/SSRI 802 10 50 34.5% Paroxetine SNRI/SSRI 124 5 50 29.8% Sertraline SNRI/SSRI 555 10 50 7.2% Amphetamine Stimulants 3166 25 100 15.0% Fentanyl Synthetic Opiates 1418 1 5 31.7% Norfentanyl Synthetic Opiates 2246 2 5 10.2% Tapentadol Synthetic Opiates 119 2 50 18.5% Tramadol Synthetic Opiates 1548 25 50 6.1% The samples were prepared by a dilute and shoot method described briefly here. A Freedom EVO (Tecan, Maennedorf, Switzerland) was used to add 120 µl of urine specimen to a 96-well plate. Then, 30 µl of β-glucuronidase solution (IMCS, Columbia, SC) and 30 µl of 2 mm ammonium acetate buffer (Fisher Scientific, Waltham, MA) in water were added. The plates were incubated at 60 C for 30 minutes. Finally, 300 E1109

Pain Physician: November/December 2017: 20: E1107-E1113 µl of methanolic internal standard solution and 500 µl of water were added to each specimen. Five µl of prepared specimen was injected onto the analytical column for LC-MS/MS analysis. The results were analyzed using ASCENT software (Indigo BioAutomation, Inc., Indianapolis, IN). A 4-point calibration curve was used with a linear fit and one/x weighting. Calibrator acceptability was within ± 20% of the expected concentration with an R 2 value of greater than 0.98. The area ratio of the analyte to a deuterated internal standard was used to account for ion suppression. All analytes had a signal-to-noise calculation of greater than 10 at the lower limit of quantitation. The precision and accuracy of the assay was evaluated over 5 days for both intraday and interday variability, and all analytes were within 20% CV. Recovery was determined to be within ± 20% for all analytes. An estimate of the high-sensitivity cutoff levels required were determined by plotting previous patient data on a semi-logarithmic plot and using the intersection of an applied trend line and x-axis for each drug or metabolite in Microsoft Excel (Microsoft Corp., Redmond, WA). The lower limit of quantitation (LOQ) was set at the higher level of the trendline intersection and where S/N 10. Results The Sciex 6500 LC-MS/MS is one of the most sensitive instruments available to the clinical laboratory. Its dynamic range is about 10 5. Compared to other instruments by the same manufacturer, it has a signal-to-noise and LOQ improvement of up to 5-fold over the widest range of compounds and about a 20-fold increase in detector dynamic range. The result is that the limit of quantitation can be set to be considerably lower than many other laboratories performing the same drug test. These lower limits of quantitation, often termed cutoffs, were used in this study (please see Table 1). The lognormal distribution of hydromorphone (Fig. 1) shows the typical truncated bell curve found for most drugs analyzed with greater than 100 positive patient results during the study period. The trendline indicated a more appropriate cutoff level of 10 ng/ ml. The analytical system was validated to 5 ng/ml for hydromorphone, so this lower cutoff value was used instead. A retrospective analysis of 83,205 anonymized patient results, of which 46,717 had at least one positive analyte, was performed by plotting the concentrations on a lognormal distribution graph for compounds with greater than 100 positive results. The number of positive results with concentrations below the typical industry cutoff level (4,5,14), but above the high-sensitivity cutoff level, were determined. The ratio of these results to the number of low-sensitivity positives were used to calculate the false-negative rate. The distribution of hydromorphone (Fig. 2) using the high-sensitivity method shows a bell curve that extends further down the normal distribution of results. This demonstrates that a greater proportion of the positive population is being captured by the high-sensitivity LC-MS/MS method. The low-sensitivity method with a cutoff level of 50 ng/ml has a 37.4% false-negative rate. The rates were calculated for all analytes with sufficient positive results and are displayed in Table 1, along with the cutoff levels used for the calculation. Discussion Our observations are not surprising. Previous urine drug testing work has shown that using lower cutoffs increases the number of positive results (24-26). The central problem is figuring out the most objective method of determining the optimal cutoff. We suggest using frequency distributions after log transformation offers a solution. This work shows that the lower cutoffs enabled by a more sensitive LC-MS/MS analysis allow good estimates of medication cutoffs from their observed distributions. We show that these cutoffs can readily be derived from patients data. Visual inspection of the frequency distribution can be used to establish whether the cutoff is appropriate. Laboratory-derived cutoffs offer better clinical insights than arbitrary administrative cutoffs currently accepted by health care providers. These data show the common cutoff levels are insufficient to adequately detect and quantitate most common prescription medications and illicit substances. This can result in false-negative reports being sent to the treating health care provider. Therefore, falsenegative results can lead the health care provider to doubt a patient s adherence to the prescribed regimen and reduce or not prescribe needed pain medications. Also, drug diversion may be incorrectly suspected when cutoff levels are inappropriately set. Our analysis of the distributions of the high-sensitivity urine drug concentrations show that a method with even greater sensitivity should be investigated for fentanyl, norfentanyl, cyclobenzaprine, oxazepam, temazepam, and zolpidem. Our data set was insufficiently large to determine cutoff levels for less com- E1110

Lower Cutoffs for LC-MS/MS Urine Drug Testing Indicates Better Patient Compliance Fig. 1. A plot of hydromorphone concentrations in urine. The lognormal distribution of positive patient samples shows a truncated bell curve at low concentrations when using the typical LC-MS/MS cutoff for hydromorphone. It shows that falsenegative results are being reported due to incorrectly set cutoff levels. Fig. 2. A plot of the same hydromorphone data using a lower cutoff. The lognormal distribution of positive patient samples show a more complete bell curve at low concentrations when using a high-sensitivity LC-MS/MS analysis, validated to 5 ng/ml. E1111

Pain Physician: November/December 2017: 20: E1107-E1113 monly prescribed medications. Caution should be used when evaluating adherence based on urine drug test results for antidepressants and antipsychotics. A high-sensitivity method was not investigated for the natural opiates morphine and codeine due to concerns about incidental ingestion from food sources. The cutoff levels for methadone and its metabolite EDDP were not lowered because a full distribution was observed at typical cutoff levels of 50 ng/ml. It is important to consider that the distribution data comes from a patient population that is not representative of the world at large, but is indicative of patients who are frequently tested in the American health system. The population presented in this paper is likely over-representative of patients with chronic pain who are prescribed opioid analgesics and patients with substance use disorder. Acknowledgments Author contributions: Drs. Kevin M. Krock, Amadeo J. Pesce, Richard J. Thomas, Agnes S. Cua, and Mr. Dennis Ritz had full access to all of the data in the study and take responsibility for the integrity of the data. The accuracy of the data analysis was the responsibility of Ryan H. Rogers, Phil Lipnick, and Kristen R. Kilbourn. Dr. Kevin Krock, Dr. Agnes S. Cua, and Mr. Dennis Ritz designed the study protocol. Dr. Kevin Krock and Dr. Amadeo Pesce managed the literature searches and summaries of previous related work and wrote the first draft of the manuscript. Dr. Kevin Krock provided revision for intellectual content and final approval of the manuscript. We also would like to thank the editorial board of Pain Physician for review of the manuscript. References 1. Trescot AM, Helm S, Hansen H, Benyamin, R, Glaser SE, Adlaka R, Patel S, Manchikanti L. Opioids in the management of chronic non-cancer pain: An update of American Society of the Interventional Pain Physicians (ASIPP) Guidelines. Pain Physician 2008; 11(2 Suppl):S5-S62. 2. Manchikanti L, Atluri S, Trescot AM, Giordano J. Monitoring opioid adherence in chronic pain patients: Tools, techniques, and utility. Pain Physician 2008; 11(2 Suppl):S155-S180. 3. Christo PJ, Manchikanti L, Ruan X, Bottros M, Hansen H, Solanki DR, Jordan AE, Colson J. Urine drug testing in chronic pain. Pain Physician 2011; 14:123-43. 4. Owen GT, Burton AW, Schade CM, Passik S. Urine drug testing: Current recommendations and best practices. Pain Physician 2012; 15(3 Suppl):ES119-ES133. 5. Substance Abuse and Mental Health Services Administration (SAMH- SA). Clinical Drug Testing in Primary Care. Technical Assistance Publication (TAP) 32. HHS Publication No. (SMA) 12-4668. Rockville, MD: Substance Abuse and Mental Health Services Administration, 2012. http://store. samhsa.gov/product/tap-32-clinical-drug-testing-in-primary-care/ SMA12-4668. 6. Heit HA, Gourlay DL. Philosophy of urine drug testing in pain management. www.prescriberesponsibly.com/articles/ urine-drug-testing. 7. Saitman A, Park HD, Fitzgerald RL. Falsepositive interferences of common urine drug screen immunoassays: A review. J Anal Toxicol 2014; 38:387-396. 8. Reisfield GM, Goldberger BA, Bertholf RL. False-positive and false-negative test results in clinical urine drug testing. Bioanalysis 2009; 1:937-952. 9. Clinical Laboratory and Standards Institute. Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline. 3rd ed. EP28-A3c. Wayne, PA, 2008. 10. Horn PS, Pesce AJ. Reference Intervals: A User s Guide. AACC Press, Washington, DC, 2005. 11. Horn PS. Reference intervals and clinical decision limits. In: Kaplan LA, Pesce AJ (eds). Clinical Chemistry: Theory, Analysis, Correlation. 5th ed. Mosby, St. Louis 2010, pp 439-455. 12. Cone EJ, Caplan YH, Black DL, Robert T, Moser F. Urine drug testing of chronic pain patients: Licit and illicit drug patterns. J Anal Toxicol 2008; 32:530-543. 13. Heltsley R, Zichterman A, Black DL, Cawthon B, Robert T, Moser F, Caplan YH, Cone EJ. Urine drug testing of chronic pain patients II. Prevalence patterns of prescription opiates and metabolites. J Anal Toxicol 2010; 34:32-38. 14. White RM, Black ML. Pain Management Testing Reference. AACC Press, Washington, DC 2008, pp 204-208. 15. Mikel C, Almazan P, West R, Crews B, Latyshev S, Pesce AJ, West C. LC-MS/ MS extends the range of drug analysis in pain patients. Ther Drug Monit 2009; 31:746-748. 16. Chew C, Kim S; Food and Drug Administration. Transcript: Medication adherence. www.fda.gov/drugs/resourcesforyou/healthprofessionals/ucm470169. htm. 17. Gourlay D, Heit HA. The art and science of urine drug testing. Clin J Pain 2010; 26:358. 18. Pesce AJ, West C, West R, Crews B, Mikel C, Almazan P, Latyshev S, Rosenthal M, Horn P. Reference intervals: A novel approach to detecting drug abuse in a pain patient population. J Opioid Manag 2010; 6:341-350. 19. Pesce AJ, West C, West R, Crews B, Mikel C, Rosenthal M, Almazan P, Latyshev S. Determination of medication cutoff values in a pain patient population. J Opioid Manag 2011; 7:117-122. 20. West R, Pesce AJ, Crews B, Mikel C, Rosenthal M, Almazan P, Latyshev S, West C. Determination of illicit drug cutoff values in a pain patient population. Clin Chim Acta 2011; 412:1589-1593. 21. Pesce AJ, West C, West R, Latyshev S, Masters-Moore D, Friel P, Hughes JM, Almazan P, Gonzales E, Mikel C. Analytical considerations when monitoring pain medications by LC-MS/MS. J Anal Bioanal Tech 2012; S5:003. E1112

Lower Cutoffs for LC-MS/MS Urine Drug Testing Indicates Better Patient Compliance 22. Pesce AJ, Gonzales E, Almazan P, Mikel C, Latyshev S, West C, Strickland J. Medication and illicit substance use analyzed using liquid chromatography tandem mass spectrometry (LC-MS/MS) in a pain population. J Anal Bioanal Tech 2012; 3:135. 23. Wu AHB. Urine adulteration before testing for drugs of abuse. In: Kwong TC, Magnani B, Rosano TG, Shaw LM (eds). The Clinical Toxicology Laboratory: Contemporary Practice of Poisoning Evaluation. AACC Press, Washington, DC 2001, pp 157-171. 24. Wingert WE. Lowering cutoffs for initial and confirmation testing for cocaine and marijuana: Large-scale study of effects on the rates of drug-positive results. Clin Chem 1997; 43:100-103. 25. Hattab EM, Goldberger BA, Johannsen LM, Kindland PW, Ticino F, Chronister CW, Bertholf RL. Modification of screening immunoassays to detect subthreshold concentrations of cocaine, cannabinoids, and opiates in urine: Use for detecting maternal and neonatal drug exposures. Ann Clin Lab Sci 2000; 30:85-91. 26. West R, Pesce AJ, West C, Crews B, Mikel C, Almazan P, Rosenthal M, Latyshev S. Comparison of clonazepam compliance by measurement of urinary concentration by immunoassay and LC-MS/MS in a pain management population. Pain Physician 2010; 13:71-78. E1113