Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa

Similar documents
Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Quality assurance in external radiotherapy

Characteristics Analysis of High Energy External Radiotherapy Beams in Water

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

[Setawati et. al., Vol.4 (Iss10): October, 2017] ISSN: DOI: /zenodo

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

The Journey of Cyberknife Commissioning

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable Phantom

Detector comparison for small field output factor measurements with flattening filter free photon beams

Eric E. Klein, Ph.D. Chair of TG-142

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn

Imaging of Radiation Dose Using Cherenkov Light

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

INTRODUCTION. Material and Methods

Spinal Cord Doses in Palliative Lung Radiotherapy Schedules

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy

The determination of timer error and its role in the administration of specified doses

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Energy dependent response of Al 2 O 3 and its potential application in personal monitoring.

Diode calibration for dose determination in total body irradiation

Comparison of Percentage Depth Dose (PDD) and Post Time Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams Using MRI Technique

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

8/3/2016. Low-Z linear accelerator targets. Disclosures. Low-Z linear accelerator targets

A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

Amendment No. 2. Item No. 2 (Rfx/ Event number )

MD 4300: Introduction to Medical Dosimetry (3semester hour) Course Syllabus for Fall 2009

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

Bureau of Laboratory Quality Standards

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Protection of the contralateral breast during radiation therapy for breast cancer

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Ali S. Meigooni, Kai Dou, Navid J. Meigooni, Michael Gnaster, Shahid Awan, Sharifeh Dini, and Ellis L. Johnson

Commissioning and Radiobiology of the INTRABEAM System

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison

Radiochromic film dosimetry in water phantoms

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

Measurement of Absorbed Doses in Anatomical Phantoms

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Dosimetry comparison of orthovoltage x-ray and 137 Cs irradiation of the murine bone marrow compartment Matthew Belley

Application(s) of Alanine

Accurate Accurate boost or Simply Accuboost

Ion recombination corrections of ionization chambers in flattening filter-free photon radiation

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

Chapters from Clinical Oncology

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Activity report from JCOG physics group

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Transcription:

International Journal of Sciences: Basic and Applied Research (IJSBAR) ISSN 237-4531 (Print & Online) http://gssrr.org/index.php?journal=journalofbasicandapplied --------------------------------------------------------------------------------------------------------------------------- Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa Awan Pelawi a, Herty Afrina Sianturi b, Martha Rianna c, Kerista Sebayang d, Nasruddin Noer e a,b,c,d,e Department of Physics, Faculty Mathematics and Natural Sciences, University of Sumatera Utara, Medan, Indonesia a Radiology Installation Dr. Pringadi General Hospital, Medan, Indonesia a Email: awan.pelawi@yahoo.com Abstract Patient dose in the Linac Electa and Siemens determined with Source Surface Distance (SSD) technique using Precentage Depth Dose (PDD) method and Source Axis Distance (SAD) technique using Tissue Phantom Ratio (TPR) method. The study was conducted by measuring the PDD and calculating the TPR to compare the calculation of the beam quality index on the Linear plane of photon energy at a measurement of 6 MV. PDD is done with a 1 cm SSD technique at a depth of cm to 25 cm. The calculation of TPR is done at SAD position of 1 cm at depth of 1 cm and 2 cm with standard field 1 cm x 1 cm. The result of Linac aircraft exposure with big 6 MV photon energy using Tissue Phantom Ratio (TPR) method is greater.98%, according to the research result obtained.66% from TPR Linac Siemens Aircraft, while Linac Elekta.67% BATAN standards are allowed at 3%). Keywords: PDD; TPR; SSD; Photon; Quality File Index. ------------------------------------------------------------------------ * Corresponding author. 1

1. Introduction Linear accelerator (Linac) is designed to produce photons and electrons. Electron accelerator or Linac has become a standard treatment machine in radiotherapy centers [7]. This tool is used to irradiate cancer inside and on the body surface [4]. The existence of Linac aircraft is very beneficial and gives hope to cancer patients to recover. To calculate the dose at a given point in the patient determined some things that describe the effect of the dose so it is necessary to see the basic parameters such as energy of the beam, field size, depth of the desired calculation point and the surface distance of the source file. In using a dosimetry prototype the photon beam in the clinic must first consider the quality of the file [3]. For megavoltage the quality of the beam is determined by the concept of the penetration ability of the beam in water. By comparing the ability of different types of radiation energy to the pattern of the penetration of the beam in the water can be distinguished the quality value of each beam, the deeper the penetration of the beam in the water shows the higher the file quality. There are 2 set-ups for determining the dose measurements in the patient's Source Surface Distance (SSD) technique using the Precentage Depth Dose (PDD) method and Source Axis Distance (SAD) technique using Tissue Phantom Ratio (TPR) method [5]. Analysis of the depth dose characteristics may help the selection of appropriate files for radiotherapy treatment when various energy beams are available. This is a dosimetric parameter to confirm the number of doses of measured values in increasing accuracy in radiotherapy treatment [2]. A modified formula for defining TPR from photon rays but less effective in clinical applications [1] and small field ratios in TPR data generation for Elekta Agility 6 MV photon rays that result in similarities in the bulid up area to a depth of 2 cm and the meeting point at depth of 1 cm [8]. Previous research by Laksono and his colleagues [6] has measured and compared the PDD and TPR methods to determine the beam quality index on a 6 MV photon linear energy plane. The measurement of PDD method is done on 1 cm SSD and TPR method on SAD 1 cm. The quality of the files can be quantized so that an index or constant of various measurable radiation energies and parameters is called the file quality index which can be determined from the PDD or TPR value [9]. Sianturi and his colleagues [1] reported that the value of tube voltage deviation (kvp) and tube time (mas) to meet the tolerance limit specified by PERKA BAPETEN No 9, 211 to four X-ray devices in hospitals at Medan give only onedevice does not exceed the limited permitted by PERKA BAPETEN No. 9,211. As per the BAPETEN No. 9 Year 211 and Western Australia Standard, that the HVL for each voltage on the X-ray plane The general radiography tested in this study is still within the permitted minimum range. In relation to the accuracy of dosage TRS-398 (Technical Reports Series No.398) has provided practical guidance on the measurement of high energy photon radiation beam on medical accelerator to determine the file quality index using SSD technique with PDD method but theoretically using SAD technique with TPR method. 2. Materials and methods The study was conducted in two Medan City Hospital. Materials and tools used in this study are Siemens M5782 aircraft, Linac Elekta Pricise 151614 aircraft and External Radiation dosimetry (Radiation Measurement Instruments). Measurement of field area is done on surface 1 cm x 1 cm with 1 cm SSD. Then do irradiation with a fixed SSD 1 cm by lowering the chamber from the surface with a depth of cm to a depth of 25 cm. Afterwards, the measurements of 6 MV Siemens and Elekta TPR were measured on 1 cm 1 cm surface with 1 cm SSD, then the chamber value was lowered to a depth of 25 cm from the surface with 1 11

cm SAD, perform irradiation by lowering the water level in phantom from the depth of chamber 25 cm from surface to cm with SAD remain 1 cm. 3. Results and Discussions 1. PDD on Photon Energy 6 MV Siemens From the measurement results of PDD 6 MV photon energy on the Siemens M5782 aircraft. Figure 1 shows the depth distribution of the dose of the PDD method on the Siemens 6 MV photon energy. Dmax at 6 MV Siemens photon energy occurs at a depth of 15 mm. Then the percentage of dose distribution will decrease gradually as the radiation on the medium will give its energy to the medium it passes. 12 1 8 6 4 2 Doses Percentages (%) PDD 6 MV Siemens PDD (%) 15 3 45 6 75 9 115 13 145 16 175 19 25 22 235 25 Figure 1: Graph of Siemens PDD on Photon Energy 6 MV 2. PDD on Photon Energy 6 MV Electa From the measurement result of PDD of 6 MV photon at Linac Electa Pricise 151614 plane. Figure 2 shows the depth distribution of dose of PDD method on Electa 6 MV photon energy. Dmax at 6 MV Electa photon energy occurs at a depth of 16 mm. Then the percentage of the dose distribution will drop gradually as the radiation on the medium will give its energy to the medium. Then the percentage of dose distribution will decrease gradually as the radiation on the medium will give its energy to the medium it passes. Prosentase dosis (%) 15 1 5 PDD 6 MV Electa PDD (%) 3 6 9 12 15 18 21 24 Depth (mm) Figure 2: Graph of Electa PDD on Photon Energy 6 MV 12

From the measurement results of PDD energy of 6 MV and 1 MV photons on Linac Siemens M5782 and Electa Pricise 151614 aircraft. Figure 3 below shows the depth distribution of dosage of PDD method on the energy of 6 MV Siemens and 6 MV Elekta photons. In the build up region the distribution of doses is almost identical in the line graphs almost coincidental until the maximum dose (Dmax) is just far apart. Dmax at 6 MV Siemens photon energy occurs at a depth of 15 mm while at 6 MV Elekta photon energy occurs at a depth of 16 mm, Then the percentage of dose distribution will decrease gradually as radiation on the medium will give its energy to the medium it passes. From the second Linac plane at the same energy of 6 MV there is a difference of () x 1% = 93.75%, this result in accordance with the standard BATAN allowed 3%. Doses Percentages (%) 12 1 8 6 4 2 PDD 6 MV Siemens Vs Electa PDD (%) Siemens PDD (%) Elekta 15 3 45 6 75 9 115 13 145 16 175 19 25 22 235 25 Depth (mm) Figure 3: Graph of PDD Siemens and Elekta on Photon Energy 6 MV 4. Conclusions 6 MV photon energy PDD method, Dmax Linac Elekta aircraft is larger than.94% compared to Dmax Linac Siemens aircraft, according to research results Dmax Plane Linac Siemens is at a depth of 15 mm while Elekta is at a depth of 16 mm. At 6 MV photon energy Tissue Phantom Ratio (TPR) method, TPR Linac Elekta plane.98% larger, according to the research results obtained TPR Plane Linac Siemens obtained.66% while Linac Elekta aircraft.67% References [1] Alam M. J., Rabbani Z., Hussain A., Baig A. K. V., 27, A modified formula for defining tissue phantom ratio of photon beams, Bangladesh Medical Research Council. 33: 92-97 [2] Buzdar S. A., Rao A., Aalia N., 29, An Analyis f Depth Dose Charateristik of Photon in Water, Journal Ayub Med Coll Abbottabad, The Islamia University, Bahawalpur, Pakistan. 21(4) [3] Chen C., 27, Principles and requirements of external beam dosimetry, Elselvier, Department of Therapeutic Radiology, Yale University School of Medicine, New Haven, Connecticut, USA. S2 S21 [4] Khan F. M., 23, Title : Physics of Radiation Therapy, The 3rd Edition, Lippincott Williams & 13

Wilkins, Philadelphia, USA(136-137,164) [5] Kriˇzan P., Rihard H., 28, Teleradiography Dose Calculation, Faculty of Mathematics and Physics, University of Ljubljana. [6] Laksono, M., Susila Wardaya, Wahyu Setia B., and Eko Hidayanto. Comparative Study Method of Percentage Depth Dose (Pdd) and Tissue Phantom Ratio (Tpr) for Determining Quality of Energy File Index 6 Mv And 1 Mv on Linac Aircraft. Periodical Physics ISSN: 141-9662 Vol. 18, No. 1, January 215, pp. 1-8 [7] Podgorsak E.B., 25, Radiation Oncology Physics: A Handbook for Teachers and Students, International Atomic Energy Agency, Technical Editor, Vienna. Thirth Edition.(179-191) [8] Richmond N., Robert B., 214, A Comparison of Small-Field Tissue Phantom Ratio Data Generation Methods for an Elekta Agility 6 MV Photon Beam, Elsevier, Medical Physics Department, The James Cook University Hospital, Middlesbrough, United Kingdom.6-63 [9] Sardari D., Maleki R., Samavat H., Esmaeeli A., 29, Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code, Elselvier, Department of Medical Physics, Hamadan University of Medical Science, Hamadan, Iran. Reports of practical oncology and radiotherapy 15 (21)64 68 [1] Sianturi, HertyAfrina., Martha Rianna., TimbangenSembiring., Marhaposan Situmorang.,218. Measurement and Analysis of Output Radiation Dose on X-Ray Device over 1 Years at Hospitals in Medan City.J. Aceh Phy. Soc. Vol. 7, No. 1 pp.1-5, 218 14