Urine/CSF Albumin Reagent. Urine/CSF Albumin Calibrator

Similar documents
QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

Gentian Canine CRP Immunoassay Application Note for scil VitroVet*

Cardiac Assessment Controls

Third Party Controls for Your Dade Behring Systems

For In Vitro Diagnostic Use

Bio-Rad Laboratories. Cardiac Assessment Controls. Value Assigned for Clinical Laboratory and Point of Care Test Systems

liquicolor (AMP Buffer, IFCC) Design Verification

LIPASE liquicolor. Design Verification. Multipurpose Reagent

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:

Cardiac Assessment Controls A critical element of reliable cardiac testing

Quick Reference Guide. on Turbodyne SC

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

1 PROTOCOL. Comparison Study Summary

The Adiponectin Turbidimetric Immunoassay Reagent Kit

ARK Methotrexate Assay Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 4600 Chemistry System

SCIEX Vitamin D 200M Assay for the Topaz System

Clinical Chemistry Specific Proteins Presentation

Beckman Coulter introduces Sentinel Pancreatic Amylase reagent and calibrator on AU Clinical Chemistry Analyzers

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

DRI METHAQUALONE APPLICATION BECKMAN COULTER AU5800

CEDIA OPIATE APPLICATION BECKMAN COULTER AU5800

HbA1c3 PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

Product Correction Urgent Immediate Action Required

DIAZYME PRODUCT LIST LIST INNOVATIONS IN CLINICAL DIAGNOSTICS

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

human Total Cathepsin B Catalog Number: DY2176

RealLine HIV quantitative Str-Format

DIAZYME PRODUCT LIST LIST INNOVATIONS IN CLINICAL DIAGNOSTICS

C-Peptide I and II (Rat) ELISA

Urgent Field Corrective Action

HumaTex CRP. Design Verification. Contents

Technical Bulletin. Performance Characteristics of the BN ProSpecä System. By Liana Frances Romero, BS, MT(ASCP)

READ HIGHLIGHTED CHANGES

Smart Diagnostics for Point-of-Care!

BLOOD IS COMPLEX ANALYZING IT SHOULDN T BE DxH 500 * *Not available for sale in the U.S.

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey

Rat C-Peptide EIA. Cat. No. YII-YK010-EX FOR LABORATORY USE ONLY

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

ab ORAC Assay Kit

NEPHROCHECK Liquid Control Kit Package Insert

DIAZYME PRODUCT LIST LIST INNOVATIONS IN CLINICAL DIAGNOSTICS

ASSAYED URINE CONTROL - LEVEL 2 (URN ASY CONTROL 2)

For in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date

GC & LC Application Note

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Date. Measurements of cocaine and cocaine metabolites are used in the diagnosis and treatment of cocaine use or overdose.

Aina Blood Monitoring System

INSTRUCTIONS FOR USE

GSI Canine IL-5 ELISA Kit-2 Plates DataSheet

91154 PRIME-XV T Cell CDM Liquid 1 L Additional package sizes are available at request

Original Article. Evaluation of two new high-sensitivity methods for C-reactive protein. Introduction. Abstract

HbA1c PRINCIPLE REF ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

Rat Glicentin EIA FOR RESEARCH USE ONLY. <Distributed by> DF Kasumigaseki Place, 3-6-7, Kasumigaseki, Chiyoda-ku Tokyo Japan

Daily Quality Control Calibration Verification / Linearity. Quality control made easy. Product Information

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

ASSAYED URINE CONTROL - LEVEL 3 (URN ASY CONTROL 3)

New York State Department of Health. Notice of Voluntary Recall of Influenza A (H1N1) 2009 Monovalent Vaccine

AmpliX HBV Quantitative

S150 KEEP Analytical Methods. American Journal of Kidney Diseases, Vol 55, No 3, Suppl 2, 2010:pp S150-S153

Commutability studies undertaken by the LNE : the case of lipid and lipoprotein testing. Vincent Delatour, PhD

Evidence Based Commutability: Bias 2 Study. Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA

Nori Rabbit IL-2 ELISA Kit DataSheet

Chapter 2. High-sensitivity C-reactive protein methods examined

Performance Guide. icheck Iodine The test kit to measure iodine in salt

Rat Leptin ELISA FOR LABORATORY USE ONLY YANAIHARA INSTITUTE INC AWAKURA, FUJINOMIYA - SHI SHIZUOKA, JAPAN

The superior solution for disease state diagnosis. IMMAGE 800 Special Chemistry System

YK052 Mouse Leptin ELISA

INSULIN MOUSE SERUM KITS

Mouse GLP-2 EIA FOR LABORATORY USE ONLY

1. Calibra - H - Store at 2-8 ºC.

Thyroxine (T4) Human ELISA Kit

Product information: Human Apolipoprotein B kit 10,000 tests 1. ASSAY DESCRIPTION AND INTENDED USE 2. PROTOCOL AT GLANCE

Evaluation Report. Eurolyser Phenobarbital test kit (VT0211) on. solo Analyser

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2

GLP-2 (Rat) ELISA. For the quantitative determination of glucagon-like peptide 2 (GLP-2) in rat serum and plasma

PRODUCTS FOR EXPORT 2018

SERUM GLUCAGON KITS PROTOCOL

Total Cost of Ownership (TCO): An evidence-based approach to compare laboratory equipment

RealLine HCV quantitative Str-Format

Urgent field safety notice

RealLine Mycoplasma genitalium Str-Format

Documentation, Codebook, and Frequencies

Comparison of Biotage Extrahera vs. Manual Sample Processing Using a Vacuum Manifold

Rat Leptin-HS ELISA FOR LABORATORY USE ONLY YANAIHARA INSTITUTE INC AWAKURA, FUJINOMIYA - SHI SHIZUOKA, JAPAN

NEPHROCHECK Calibration Verification Kit Package Insert

Abstract. Introduction

I. Introduction. II. Characteristics

ab Lipoxygenase Inhibitor Screening Assay Kit

FOR LABORATORY USE ONLY

The NEPHROCHECK Test System Training Program. The NEPHROCHECK Training Program US: Astute Medical, Inc PN Rev D 2014/09/16

Transcription:

December 2014 PRODUCT ANNOUNCEMENT AU Clinical Chemistry Systems AU400/400 e, AU480, AU640/640 e, AU680, AU2700, AU5400, and AU5800 Product Urine/CSF Albumin Reagent Urine/CSF Albumin Reagent Urine/CSF Albumin Calibrator Kit REF B38858 B46435 B38859 Dear Beckman Coulter AU Customer: Beckman Coulter, Inc. is pleased to announce the release of a new Urine/CSF Albumin reagent in addition to a new Urine/CSF Albumin calibrator for use on the Beckman Coulter AU clinical chemistry analyzers. The new Beckman Coulter Urine/CSF Albumin reagent and calibrator will replace the current Microalbumin reagent, REF OSR6x67, and calibrator, REF ODR3024. The reason for this transition is to harmonize reagents between Beckman Coulter AU and Synchron platforms plus the added features of extended upper analytical range and increased prozone tolerance. As recommended by the National Kidney Disease Education Program - International Federation of Clinical Chemistry (IFCC) Working Group on the standardization of albumin in urine, the name of the assay will now change from Microalbumin to Urine/CSF Albumin in order to reflect current practice. The new Urine/CSF Albumin reagent is a turbidimetric method for the quantitative determination of albumin concentration in human urine and cerebrospinal fluid (CSF). The new reagent kit consists of a 2-component Urine/CSF Albumin reagent and is available in 2 kit sizes. The tests per vial for each kit size on each AU analyser are outlined below. New Reagent REF Package Configuration Analyser Tests/ Vial Test/Kit Urine/CSF Albumin B38858 R1: 4x32.6 ml R2: 4x4.4 ml AU400/400 e /480/640/ 640 e /680 129 516 AU2700/5400/5800 203 812 B58824-AA

New Reagent REF Package Configuration Analyser Tests/ Vial Test/Kit Urine/CSF Albumin B46435 R1: 4x54.8 ml R2: 4x7.3 ml AU400/400 e /480/640/ 640 e /680 240 960 AU2700/5400/5800 366 1464 A new multi-level ready to use liquid calibrator is also available; New Calibrator REF Package Configuration Urine/CSF Albumin Calibrator Level 1, 2, 3, 4, 5 B38859 Levels 1-5: 2 ml Features of the new Urine/CSF Albumin assay include: 2-part ready to use liquid reagent for human Urine and CSF Albumin 60 day reagent on-board stability 60 day calibration stability Analytical range: Urine 7-450 mg/l, CSF 10-450 mg/l On-board dilution for samples exceeding the analytical range (1 in 10 dilution with deionized water or 0.9% saline, using automatic dilution function) Limit of Quantitation (LOQ): Urine 7 mg/l ( 0.7 mg/dl), CSF 10 mg/l ( 1.0 mg/dl) Prozone tolerance up to 20,000 mg/l using Data Check Parameters Reagent probe Contamination Avoidance Parameters have been developed for use with the new reagent. New sample probe Contamination Avoidance Parameters have also been developed for use when changing from Serum to Urine samples on the AU2700, AU5400, AU5800, AU480 and AU680. Recommended quality control materials for the new AU Urine/CSF Albumin assay include the Bio-Rad Liquichek Microalbumin and Spinal Fluid controls. The following marketed lots contain value assignment for the new AU Urine/CSF Albumin assay. The target values and ranges are available on the Bio-Rad website; www.myeinserts.com: Control REF Lots Bio-Rad Liquichek Microalbumin Controls Level 1 and Level 2 Bio-Rad Liquichek Spinal Fluid Controls Level 1 and Level 2 378, 379 751, 752 96660, 96670, 96680, 96690, 96700 55580, 55590, 55600, 55610, 55620 The Urine and CSF Method Comparison data for Urine/CSF Albumin reagent B46435/B38858 versus Microalbumin reagent OSR6x67 on an AU5800 analyzer is attached. The new Urine/CSF Albumin assay recovers higher than the current Microalbumin assay due to the difference in B58824-AA Page 2 of 5

calibrator traceability. The new assay is traceable to IFCC standard CRM 470 while the current assay is traceable to a purified Albumin Standard quantified using a theoretical extinction coefficient. The Urine and CSF Method comparison data for the Beckman Coulter Urine/CSF Albumin reagent B46435/B38858 versus an Albumin test on a Siemens BN ProSpec Nephelometer is also attached. For both Urine and CSF, the new Urine/CSF Albumin reagent on the AU5800 correlates extremely well with the Siemens nephelometry method. The Urine/CSF Albumin Instructions for Use (BLB38858) will be available in multiple languages on the Technical Documents page of the Beckman Coulter website through the following link: https://www.beckmancoulter.com/wsrportal/page/techdocsearch. Instrument Settings/ Parameters (BSB38858), Calibrator Value Assignment Sheets, and (Material) Safety Data Sheets (M)SDSs as well as Contamination Avoidance Parameters will also be available at the above link. Please share this information with your laboratory staff and retain this notification as part of your laboratory documentation. If you have any questions, please contact Beckman Coulter Technical Support at 1-800-854-3633 from the United States and Canada or +353 65 683 1170 from Europe or your local Beckman Coulter Representative. Thank you for your continued support of Beckman Coulter products and services. Sincerely, Kerri Breasley Senior Marketing Manager Chemistry Systems Enclosures: Method Comparison graphs for OSR6167 versus B46435/B38858 on AU5800 for CSF and Urine Method Comparison graphs for B46435/B38858 on AU5800 versus an Albumin test on a Siemens BN ProSpec Nephelometer for CSF and Urine 2014 Beckman Coulter, Inc. All rights reserved. Beckman Coulter, the Beckman Coulter logo, and AU are trademarks of Beckman Coulter, Inc. and are registered in the USPTO. Bio-Rad and Liquichek are trademarks of Bio-Rad Laboratories Ltd. Siemens and ProSpec are trademarks of Siemens. All other trademarks are the property of their respective owners. B58824-AA Page 3 of 5

Method Comparison of B46435/B38858 versus OSR6167 on an AU5800 analyser A. Urine samples B. CSF samples n: 117 Slope: 1.116 Intercept: 2.839 mg/l r: 0.9982 n: 49 Slope: 1.101 Intercept: -0.544 mg/l r: 0.9992 Figure 1. Method comparison graphs for OSR6167 vs. B46435 using A. Urine samples and B. CSF samples including Deming regression summary details. B58824-AA Page 4 of 5

Method Comparison of B46435/B38858 on an AU5800 versus an Albumin Urine/CSF test on a Siemens BN ProSpec Nephelometer A. Urine samples B. CSF samples n:121 Slope: 0.964 Intercept: -1.552 mg/l r: 0.9975 n: 51 Slope: 0.968 Intercept: -0.298mg/L r: 0.9983 Figure 2. Method comparison graphs for B46435 vs. an Albumin Urine/CSF test on a Siemens BN ProSpec Nephelometer method using A. Urine samples and B. CSF samples including Deming regression summary details. B58824-AA Page 5 of 5