Survey of patients CT radiation dose in Jiangsu Province

Similar documents
Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

Computed tomography Acceptance testing and dose measurements

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Accounting for Imaging Dose

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Ask EuroSafe Imaging Tips & Tricks. CT Working Group. On the use of Diagnostic Reference Levels in CT

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Assessment of effective dose in paediatric CT examinations

Assessment of Patient Dose from CT Examinations in Khorasan, Iran

The relationship of the mean glandular dose with compressed breast thickness in mammography

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan

Quality Control and Patient Dosimetry on line for Computed Tomography

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

A multicentric study on patient dose in multislice CT

Introduction and Background

Regional diagnostic reference levels and collective effective doses from CT scanners in India

Why is CT Dose of Interest?

Managing Radiation Risk in Pediatric CT Imaging

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Assessing the standard dose to standard patients for x-ray investigations

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

CT Quality Control Manual FAQs

Chief Radiographer TEI Clinical Associate 2016

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

Cumulative radiation CT dose in young traumatic patients: a single centre 5 years retrospective review

Managing Patient Dose in Computed Tomography (CT)

A Standardised Approach to Optimisation

Radiation Dosimetry for CT Protocols

Radiation Protection Dosimetry Advance Access published April 22, 2008

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Patient Radiation Doses from Adult and Pediatric CT

8/1/2017. Financial Disclosures. Dose Tracking at MGH. How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center

Radiation Dose Reduction Strategies in Coronary CT Angiography

Scientific Exhibit Authors:

Trends in Patients Exposure Doses during Radiographic Examination in Japan

State of the art and future development for standardized estimation of organ doses in CT

Medical Physics and Informatics Original Research

Patient doses from X-ray computed tomography examinations by a single-array detector unit: Axial versus spiral mode

EXPOSURE FROM DIAGNOSTIC NUCLEAR MEDICINE PROCEDURES. 14 Victor Babes Street, RO-6600, Iasi, Romania

FDG-18 PET/CT - radiation dose and dose-reduction strategy

Adopting DAP as a dose metric in CT

CT Radiation Risks and Dose Reduction

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

Patient Doses in Chest CT Examinations: Comparison of Various CT Scanners

Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD)

Patient / Organ Dose in CT

Fetal Dose Calculations and Impact on Patient Care

Bone Densitometry Radiation dose: what you need to know

Radiation dose optimization in CT planning of corrective scoliosis surgery. A phantom study.

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations

Low Dose Era in Cardiac CT

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Prediction of Organ and Effective dose with known mas and kvp for dose Optimisation Protocol and Recommendations in CT

JMSCR Vol 06 Issue 12 Page December 2018

Tracking Doses in the Pediatric Population

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

Current status of diagnostic imaging in dental university hospitals in Japan

ICRP Symposium on the International System of Radiological Protection

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

International Journal of Radiation Research, January 2017

CT Fluoroscopy-Guided vs Multislice CT Biopsy ModeGuided Lung Biopies:a preliminary experience

Xiaohuan Pan 1,2 *, Xinguan Yang 1,2 *, Jingxu Li 1,2, Xiao Dong 1,2, Jianxing He 2,3, Yubao Guan 1,2. Original Article

Estimating Patient Radiation Dose from Computed Tomography

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

Doses from CT examinations to children suffering from hydrocephalus

Patient dose in routine X-ray examinations in Yazd state

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Estimation of the number of total and pediatric CT procedures based. on a nationwide survey in Japan

Journal of Radiology and Imaging

Radiation related cancer risk & benefit/risk assessment for screening procedures

Justification of the use of CT for individual health assessment of asymptomatic people: the Chinese experience

Trends in examination frequency and population doses in Norway,

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

PNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY

Variation of Occupational Doses among Subspecialties in Diagnostic Radiology. A.N. Al-Haj, C.S. Lagarde, A.M. Lobriguito

An audit of radiation dose of 4D CT in a radiotherapy department

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

National Cancer Institute

How to Develop CT Protocols for Children

Optimizing radiation dose by varying age at pediatric temporal bone CT

Application of international standards to diagnostic radiology dosimetry

IMAGE GENTLY HOW CAN YOU HELP?

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Combined Anatomical and Functional Imaging with Revolution * CT

Original Article Thoracic Imaging

Transcription:

Original Article Page 1 of 6 Survey of patients CT radiation dose in Jiangsu Province Yuanyuan Zhou 1, Chunyong Yang 1, Xingjiang Cao 1, Xiang Du 1, Ningle Yu 1, Xianfeng Zhou 2, Baoli Zhu 1, Jin Wang 1 1 Institute of Occupational Disease Prevention, Jiangsu Provincial Center for Disease Prevention and Control, Nanjing 210009, China; 2 Jiangsu Academy of Safety Science and Technology, Nanjing 210042, China Contribution: (I) Conception and design: J Wang; (II) Administrative support: B Zhu; (III) Provision of study materials or patients: Y Zhou, C Yang, X Zhou, N Yu; (IV) Collection and assembly of data: Y Zhou, C Yang; (V) Data analysis and interpretation: Y Zhou; (VI) Manuscript writing: All authors; (VII) Final approval of manuscript: All authors. Correspondence to: Jin Wang. Jiangsu Provincial Center for Disease Prevention and Control, 172 Jiangsu Road, Nanjing 210009, China. Email: jin.wang@163.com. Background: Through investigating 245 adult patients CT scanning conditions and the radiation dose, to analysis the patients radiation dose level of CT examination in Jiangsu Province. Methods: We investigated 245 patients CT scanning parameters from eight different hospitals of five cities in Jiangsu Province, collected and measured the CT dose index (CTDI), weighted CT dose index (CTDIw) and dose-length-product (DLP) by using standard adult dose model under the same scanning conditions, then estimated the effective doses(e) with transformation coefficient. Results: The CTDIw values of patients head, chest and abdomen scanning were lower than the national reference level. The third quartile (75%) DLP of 245 patients head and chest scanning were close to IAEA s reference level, and the abdomen scanning value was lower than IAEA and other foreign research data. Conclusions: At present the CT scanning doses level of Jiangsu Province is lower than the national standard. It is conducive to the realization of the optimization of medical radiation protection with reasonable optimization of CT scanning parameters and improving the level of medical radiation dose reference. Keywords: Computed tomography (CT); CT dose index (CTDI); dose length product (DLP); effective dose; reference level of the medical radiation Received: 10 October 2016; Accepted: 16 November 2016; Published: 06 January 2017. doi: 10.21037/jphe.2016.12.10 View this article at: http://dx.doi.org/10.21037/jphe.2016.12.10 Introduction Computed tomography (CT) is an important diagnostic imaging modality that plays a key role in medical diagnosis. With increasingly widespread clinical application of CT imaging technology, the patient exposures of CT examinations has shown a steady increase (1). According to the UNSCEAR, CT contribution to collective effective dose was estimated to account for 34% of the total (2). Therefore, it is useful for researching the CT examinations of dose distribution caused and there have been many related investigations and reports about analyzing the radiation dose levels of different types scanning in the worldwide (3,4). The Chinese national technical standard file called GBZ165-2012 (5) has released in 2012, it provides a diagnostics reference level of CT examination on radiological protection requirements with different types of patients and patient s different parts. This study through the investigation and analyzing 245 cases of adult patients with CT examination of radiation dose in Jiangsu province, and to compare it with diagnostic reference level (DRL) provided by the newly national standard. Methods Object For this CT dose survey, samples were selected by the

Page 2 of 6 Journal of Public Health and Emergency, 2017 multi-stage cluster sampling method. First, five cities (Nanjing, Zhenjiang, Xuzhou, Suzhou and Yancheng) were selected randomly according to the economic situation in Jiangsu Province, then choosing eight hospitals with different sizes in those cities and considering different types of CT equipment in proportion. Finally a total of 245 cases of adult patients with CT scan parameters as the research object in those hospitals, at the same time the patients sex, weight and age, CT equipment type, scanning area and scanning mode will be recorded. Equipment The equipment consists of the following models, Philips offerings six row helical CT, Philips offerings 128 row helical CT, GE Lightspeed 16 row helical CT, GE Lightspeed 16 row helical CT Pro, GE CT325 spiral CT, GE Hispeed spiral CT, Siemens Somatom Definition since 16 row helical CT, Siemens Somatom Definition since 64 row helical CT, Siemens Somatom Definition since 128 row helical CT, Siemens Somatom Definition as spiral CT, Siemens Somatom Definition. And testing instruments such as Swedish Barracuda X-ray, CT quality control detector, TM160 and TM164 CT dose phantom are also be needed. Testing instruments are in the calibration period of validity. Parameter measurement Putting the dose model on the scanning field center, the cylindrical phantom axis and the vertical scanning level, respectively, placing CT dose probe in the center of the dose model and 1 cm from the surface 3, 6, 9, 12 points in the direction of the hole. The effective sensitive center of probe is located in the middle of the scanning plane position, and the rest of the hole filled with organic glass rod. Choose the actual examination of scanning sequence, and at the same working voltage kvp, current mas parameters such as scanning, then measure and record the CT dose index, according to the following formula to calculate respectively CTDIw and DLP. CTDIw =1/3 CTDI 100 (center) + 2/3 CTDI l00 (peripheral), CTDI 100 (center) is a position in the center of the die body measurement CTDI 100 ; CTDI 100 (peripheral) for the die bodies around four different positions (10 mm) under the surface of the body measurement of average. DLP (cm), mgy = CTDI vol (mgy) scanning length (cm); CTDI vol = CTDIw/pitch type, CTDI vol was volumetric CT dose index, for a particular scan scheme under the radiation dose. DLP reflects a specific scan of the total absorbed dose. Effective dose estimation Using the conversion factor (6) by the DLP estimating effective dose based on the report ICRP102, then through the following formula for estimating effective dose (E) of the different parts of the scanning. E = K*DLP, and K is a type of experience weighted factor with the units msv cm 1 mgy 1, related to the parts of the body. Statistical processing In this article, the relevant data are written in the form of x±s, data analysis by SPSS 13.0. Results Main CT scanning parameters used in adult patients in Jiangsu Province Table 1 summarizes the four types of inspection of Jiangsu Province, a total of 245 cases of adult patients with main parameters used in the CT scanning. As can be seen from Table 1, the change of the tube voltage range is small between different kinds of examination and scanning way, average about 120 kv, the range in kv [110 140], the head and spine scanning time is 1.3 times longer than that of the chest and abdomen. The chest average scanning length has a maximum value (220.78 mm), is nearly 2.8 times that of the lumbar spine (79.98 mm), but between the chest and abdomen scanning pitch have no obvious difference. Adult patients with weighted CTDIw CT dose index value of CT examination in Jiangsu Province Weighted CT dose index called CTDIw is an important characterization of dose in patients with CT examination. Table 2 summarizes the adult patients with weighted CT dose index value of CT examination in Jiangsu Province. As can be seen from Table 2, different type checking CTDIw mean have obvious difference, the head scan has a maximum average scanning length value of 44.54 mgy, which is 2.6 times than the chest scan value (17.31 mgy). When checking with the same scanning type, CTDIw value range also have bigger difference, the lumbar CTDIw minimum value range, the ratio of the maximum and minimum value is 3.25, and the chest check CTDIw value range is the largest, the ratio of the maximum and the minimum of 12.6 times. Table 3 shows the national standards in the diagnosis of the reference level with CT examination, so

Journal of Public Health and Emergency, 2017 Page 3 of 6 Table 1 Main CT scanning parameters used in adult patients in Jiangsu Province Examination type Scanning way Cases Tube voltage (kv) Current time product (mas) Scan length (mm) Pitch Range Mean ± SD Range Mean ± SD Range Mean ± SD Range Mean ± SD Head Axis 70 110 140 120.57±6.99 150 435 256.43±94.43 48 140 110.8±25.59 Abdomen Spiral 69 110 140 119.85±6.06 82 270 186.20±65.88 90 700 114.29±205.72 0.75-1.38 1.08±0.28 Chest Axis 46 110 120 118.70±3.41 43 270 191.91±62.86 108 320 220.78±59.31 1-1.75 1.43±0.22 Spine Spiral 60 120 140 124.31±8.31 100 400 245.83±74.84 30 136.8 79.98±27.15 SD, standard deviation. Table 2 Adult patients with weighted CTDIw of CT examination in Jiangsu Province Examination type Scanning way Cases Range CTDIw (mgy) Head Axis 70 19.36 66.9 44.54±12.78 Abdomen Spiral 69 4.15 35.13 21.90±8.36 Chest Spiral 46 2.4 30.34 17.31±8.16 Spine Axis 60 16.48 53.71 31.75±10.34 CT, computed tomography; CTDIw, CT dose index. Table 3 Reference level with CT examination in adult patients Examination type CTDIw (mgy) Brain 50 Lumbar 35 Abdomen 25 The value is measured by the value of the rotation axis of a water phantom deduced, the length of phantom is 15 cm, the diameter is 16 cm (head) and 30 cm (lumbar spine and abdomen). CT, computed tomography; CTDIw, CT dose index. it is obvious the dose level of adult patients with different types of CT scan in Jiangsu are lower than the national standard. The DLP values of adult patients CT examination in Jiangsu Province The DLP value is derived from the product of the scanning length is a representational characteristic parameter which can appear the CT examination s diagnostic radiology reference (guidance) levels currently. Table 4 summarizes the adult patients the DLP values of four types CT examination in Jiangsu Province. As you can see in Table 4, there were significant differences in values of DLP from different kinds of examination, and the head scan average DLP is nearly 2 times than the chest and lumbar spine. Even if the same types of inspection, different client to scan the DLP value also have obvious difference, the head scanning maximum DLP value is times 4.2 larger than the minimum value, and abdominal scan DLP ratio as high as 17.5 times the maximum and the minimum. With the IAEA s reference level in Table 5, compared to the corresponding diagnostic reference levels in Jiangsu s adult patients with head scan, DLP s third quartile (75%) value was slightly higher than the IAEA data, and the scan of the chest and abdomen third quartile (75%) values are lower than the IAEA s; Compared to the research, the third quartile (75%) value of the head of Jiangsu s patients scanning was a value of 65% of the patients in UK, chest check for 39% of the UK, but abdominal examination s values of the two countries are basically equal. Compared with Europe, the head DLP s third quartile (75%) value of Jiangsu inspection is 1.8 times that of Europe, chest is 1.4 times as much, and the abdomen to check for 74% of European research data. Figure 1 shows intuitive adult patients DLP values of different types of CT scan in Jiangsu and the comparison of different studies of diagnostic reference levels DRL results.

Page 4 of 6 Journal of Public Health and Emergency, 2017 Table 4 DLP values of adult patients CT examination in Jiangsu Province Examination type Scanning way Cases DLP (mgy cm) Range The third quartile (75%) value of DLP (mgy cm) Brain Axis 70 493.16±198.65 222.62 936.63 601.52 Abdomen Spiral 69 408.96±250.44 82.9 1,452.23 536.51 Chest Spiral 46 256.93±127.92 67.86 511.38 365.25 Lumbar Axis 60 253.25±134.71 98.96 580.05 290.89 DLP, dose-length-product; CT, computed tomography. Table 5 Adult patients diagnostic reference levels in different studies DRL (mgy cm) Examination type IAEA reach data (Tsapaki etc., 2006) UK s DRL of MDCT (Shrimpton etc., 2005) Europe s DRL of MDCT (Bongartz etc., 2004) Jiangsu s third quartile (75%) value of DLP (mgy cm) Brain 527 930 337 601.52 Chest 447 940 267 365.25 Abdomen 696 560 724 536.51 These data was from ten representative research centers of six countries, DLP (mgy cm) was considered as a dose parameter in this table, and the listed data was the third quartile (75%) value of the average. DRL, diagnostic reference level; MDCT, multi-detector computed tomography; DLP, dose-length-product. 1000 900 800 700 600 500 400 300 200 100 0 Brain Chest Abdomen IAEA reach data UK s DRL of MDCT Europe s DRL of MDCT Jiangsu's third quartile (75%) value of DLP (mgy cm) Figure 1 The comparison of different studies of diagnostic reference levels DRL results. DRL, diagnostic reference level; MDCT, multi-detector computed tomography. Adult patients effective dose of CT examination in Jiangsu Province Table 6 lists the adult patients with effective dose of CT examination in Jiangsu province (msv), and the comparative analysis with the relevant study in the UK, as you can see from the Table 6, adult patients with different types of effective dose of CT examination in our province values were much lower than the UK and the United States related research data. Results Reasonable optimization of CT scan parameters is the key to reduce the patient radiation dose Radiation dose of CT examination subjects mainly depends on the scanning parameters, such as tube voltage, tube current and so on. So without affecting the quality of CT images, the reasonable optimization of CT scanning parameters, as far as possible to make the client doses of radiation can be reasonably reach the lowest level is the direction (9). Scanning parameters, such as in the previous summary analysis, when the average tube voltage is 120 kv, scanning the mean length of 110 mm, the head scanning tube current time product 256 mas and abdominal tube current product 186 mas differences, cause head scan s CTDIw value is 2 times of the abdomen s. When the pipe average voltage set at 120 kv, and current product of time average 190 mas, it is found that because of the abdominal scan length is 114 mm and chest scan length is 220 mm in length of different, then the abdominal dose product of DLP is 1.6 times of the chest. These gaps of CTDIw or DLP values in patients not only depends on the individual differences and diagnosis of actual need, has also been different hospitals, different radiation workers by reason of personal habits and the

Journal of Public Health and Emergency, 2017 Page 5 of 6 Table 6 The comparison of CT examination effective dose in Jiangsu province (msv) with the UK and the United States Examination type Scanning way Conversion factor (7) (msv mgy 1 cm 1 ) Range Effective dose UK (6) United States (8) Brain Axis 0.0021 0.46 1.97 1.04±0.42 2 2 Abdomen Spiral 0.015 1.24 21.78 6.13±3.76 10 10 Chest Spiral 0.014 0.95 7.15 3.59±1.79 8 7 CT, computed tomography. technical problems in the choice in a CT scan parameters (tube voltage, tube current, pitch and scanning length, etc.) effects. All in all, there are many influence factors in client dose, we should keep in does not affect the diagnosis quality conditions, select individual scan scheme to achieve the effect of the patients with lower dose. Constantly update perfect reference level dose of CT examination in order to reduce patient doses to achieve protection optimization This study through the investigation of 245 cases of CT scan parameters Jiangsu Province, using the standardization of the CT die body and long ionization chamber to measure the CT dose index CTDIw and DLP values which from CTDI again obtained at various points. Through comparative analysis of adult patients head weighted with CT examination in Jiangsu, lumbar and abdominal CT dose index CTDIw is lower than the national standard. Compared with foreign research data, the head CT scan DLP s third quartile (75%) value in Jiangsu was less than in Britain, higher than that of the IAEA and the European research data, the chest CT scan DLP s third quartile (75%) value was higher than that of Europe, below the IAEA and the UK, and the abdominal CT scan DLP value was the minimum of four. It is concerned that these results may be related to Chinese adult subjects weight and waist are lower than western adults, so this actual value standard may be lower than the display. Therefore, to establish scientific and perfect diagnostic radiology medical reference (guidance) level, calculation is the important measure of promoting medical radiation protection optimization. Chinese current national standard GBZ165-2012 gives the CTDIw values in adult patients with CT scan head, chest and abdomen, for other parts such as lumbar is lack of guidance levels, we can be on the basis of the current standard complement dose reference level of different type checking, and reduce the existing can achieve the minimum level of reasonable guidance to further promote protection optimization. Weighted CT dose index CTDIw, dose of product of the length of the DLP, and the effective dose all can be well on surface characterization of the dose of CT examination in the patients, but effective dose can t be directly measured, instead of calculation from the measured dose conversion factors and the DLP value is multiplied. In this study, our province adult subjects head, chest and abdominal CT scan of the DLP average 493.16, 256.93 and 408.96 mgy cm in Jiangsu, so the estimation of the head, chest and abdominal CT scan to client effective doses were 1.04, 3.59 and 6.13 msv, which were far lower than effective dose of CT examination in the UK and the US. By comparing the hint we can find an actual dose conversion factors based on China s population through further experimental study. This study by adopting CT equipment management system, under the condition of each patient s actual scanning using CT model measurement of CT dose index and then the method to estimate how radiation dose belong to the fixed mold test. The limitation of the method is that cannot accurately reflect the actual situation of the individual client illuminated. When real-time automatic exposure control technology is used, the fixed die bodies of the limitations of the test will be more obvious. Therefore, we will be having a fixed die body test method, on the basis of the search for more convenient and practical method to detect radiation dose of CT client. The next study step we will acquire real patients scan parameters such as CTDIw, CTDIvol and DLP value directly from CT image equipment management system and patient dose information, then regular summary and analysis of these data, comparing with the DRL, finally find the cause of the wave data and take corrective action. We can imagine that the result of the guarantee of CT examination will be in a state protection optimization in Jiangsu, China. Acknowledgements Funding: This research was supported by Projects of Health

Page 6 of 6 Journal of Public Health and Emergency, 2017 industry research special funds (Grant No. 201002009). Footnote Conflicts of Interest: The authors have no conflicts of interest to declare. References 1. United Nations Scientific Committee on the Effects of Atomic Radiation. Levels and trends of exposure in diagnostic radiology, Report to the General Assembly with Scientific Annexes. Vienna: United Nations, 2010. 2. United Nations Scientific Committee on the Effects of Atomic Radiation. Sources and effects of ionizing radiation medical radiation exposures, annexe D. UNSCEAR 2000 Report to the General Assembly with annexes. New York: United Nations, 2000. 3. Kritsaneepaiboon S, Trinavarat P, Visrutaratna P. Survey of pediatric MDCT radiation dose from university hospitals in Thailand: a preliminary for national dose survey. Acta Radiol 2012;53:820-6. 4. Pantos I, Thalassinou S, Argentos S, et al. Adult patient radiation doses from non-cardiac CT examinations: a review of published results. Br J Radiol 2011;84:293-303. 5. Ministry of Health of the People s Republic of China. GBZ165-2012 Radiological protection requirements for X-ray computed tomography. Beijing: People s Medical Publishing House. 6. Shrimpton PC, Hillier MC, Lewis MA, et al. National survey of doses from CT in the UK: 2003. Br J Radiol 2006;79:968-80. 7. International Commission on Radiological Protection. Managing patient dose in multi-detector computed tomography (MDCT). ICRP Publication 102. Oxford: Pergamon Press, 2007. 8. Chen J, Moir D. An estimation of the annual effective dose to the Canadian population from medical CT examinations. J Radiol Prot 2010;30:131-7. 9. Li S, Li Q, Zhang Y, et al. Research of CT examination radiation dose of 1200 tumor patients. Journal of International Oncology 2014;41:302-5. doi: 10.21037/jphe.2016.12.10 Cite this article as: Zhou Y, Yang C, Cao X, Du X, Yu N, Zhou X, Zhu B, Wang J. Survey of patients CT radiation dose in Jiangsu Province..