British Society for Dermatological Surgery (BSDS) & British Heart Rhythm Society (BHRS) Guidance on Implanted Devices & Dermatological Surgery

Similar documents
Peri-operative management of pacemakers and implantable cardiac defibrillators

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields

BRITISH HEART RHYTHM SOCIETY GUIDELINES FOR THE MANAGEMENT OF PATIENTS WITH CARDIAC IMPLANTABLE ELECTRONIC DEVICES (CIEDs) AROUND THE TIME OF SURGERY

NORTH OF ENGLAND CARDIOVASCULAR NETWORK (NECVN)

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

Cardiac Arrest Survivors and Implantable Defibrillator Recipients:

Emergency Department Management of Patients with Implantable Cardioverter Defibrillators

Deactivating the shock function of an implantable cardioverter defibrillator (ICD) towards the end of life

Procedure for the Reprogramming of Pacemakers (PPMs) Prior to Elective and Emergency Surgery

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Implantable Cardioverter Defibrillator

Use of Automated External Defibrillators (AED s) Frequently Asked Questions

Direct Current Cardioversion (DCCV)

Cardiac Arrest Survivors and Implantable Defibrillator Recipients:

Division 2, Surgical and Anaesthetics Directorates All surgical and anaesthetics staff Patients with an implanted spinal cord stimulator For:

Date of Meeting: Ratified Date: 11/03/2010

NHS. Implantable cardioverter defibrillators (ICDs) for arrhythmias. National Institute for Health and Clinical Excellence. Issue date: January 2006

09/05/2018. True/False: All pacemakers have defibrillation capabilities.

Pediatric pacemakers & ICDs:

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death

2009 CPT Codes for Cardiac Device Monitoring

Different indications for pacemaker implantation are the following:

Perioperative Management of Cardiac Implantable Devices

Electrosurgery (diathermy and coblation) for tonsillectomy

Implantation of an Implantable Cardioverter Defibrillator (ICD)

About atrial fibrillation (AFib) Atrial Fibrillation (AFib) What is AFib? What s the danger? Who gets AFib?

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

Pacemaker and ICD Interrogation

HEART OF THE MATTER: cardiac issues in safe endoscopy & sedation

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul,

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

9/5/17. Anesthetic considerations for patients with implanted devices for treating chronic pain and more. Objectives:

Implantation of a Cardiac Resynchronisation Therapy-Defibrillator (CRT-D)

ICD Your quick guide. British Heart Foundation 2017, a registered charity in England & Wales (225971) and Scotland (SC039426).

Quality Standards for the Implantation of Cardiac Rhythm Management Devices. Pan- London Arrhythmia Project Group. Version 3 (18 th July 2011)

What to expect when having a pacemaker implantation

Abnormal Heart Rhythms (Arrhythmias)

Specialised Services Policy Position PP151

The protection you need - without touching your heart

Living With Your Implantable Cardioverter Defibrillator (ICD)

Cardiac Resynchronisation Therapy Defibrillator (CRT-D)

Management of RT patients with implanted cardiac devices: From recommendation to implementation

ICD Implantation Patient Information

MEDICAL POLICY Cardiac Event Monitors/ Cardiac Event Detection

Supplemental Material

Preoperative Evaluation: Patients with Cardiac Disease

Cardiac Resynchronization Therapy with Defibrillation (CRT-D)

BHRS Prep course Pub style Quiz NOT A

STANDARD OPERATING PROCEDURE (SOP)

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

Automated external defibrillators

Direct Current (DC) Cardioversion

Cardiac Resynchronisation Therapy Patient Information

Sacral Nerve Stimulation. Patient Information Leaflet

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only.

Atrial Fibrillation. Damage to your heart caused by a heart attack or rheumatic heart disease

Guidance on competencies for Spinal Cord Stimulation Reviewed

UNDERSTANDING ELECTROPHYSIOLOGY STUDIES

What is an. arrhythmia?

Cardiac Perioperative Risk Assessment American Heart Association Guidelines

Arrhythmia Care in the DGH What Still Needs to be Done? Dr. Sundeep Puri Consultant Cardiologist

Atrial fibrillation. Understanding NICE guidance

NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE

Model 5392 EPG Temporary Pacer

1. The 2010 AHA Guidelines for CPR recommended BLS sequence of steps are:

BHRS Certification 2016: An Overview

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.

MEDICAL DEVICES. our advice on. automated external. defibrillators

Implantable Cardioverter-Defibrillators (ICD)

Implantable Loop Recorder

Sphenopalatine ganglion (SPG) stimulation for the treatment of cluster headaches

Title Deactivation of Implantable Cardioverter Defibrillators (ICD) towards the end of life Guidelines

Review guidance for patients on long-term amiodarone treatment

Elective DC Cardioversion

Sudden death from abnormal heart rhythm: Am I at risk?

Cardiac physiology tests. Walrus outpatients

Version Effective date Changes Prepared By CPR + AED

Heart Failure. Symptoms and Treatments. FloridaHospital.com

Cardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand:

Everyone Needs a Device Christopher L. Fellows, MD, FACC, FHRS Virginia Mason Medical Center Seattle, Wa.

Review of Pacemakers and ICD Therapy: Overview and Patient Management

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

Insertion of a Haemodialysis Catheter

Yes, out of sight out of mind.

Keep It Pumping. Talking to your doctors

12 th International Winter Arrhythmia School Collingwood, Ontario February 7, Dr. Derek Yung. MD, FRCPC

Dr. Guy Buchanan MBBS, FANZCA, FFPMANZCA Specialist Pain Medicine Physician & Anaesthetist Associate - Victoria Pain Specialists

Need to Know: Implantable Devices. Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia

Implantable Cardioverter-Defibrillators (ICD)

Ambulatory Cardiac Monitors and Outpatient Telemetry Corporate Medical Policy

Electrocardiography for Healthcare Professionals

Diagnosing Arrhythmias

Joint Working Group to produce guidance on delivering an Endovascular Aneurysm Repair (EVAR) Service.

Accessing Battery Information. Status symbol. Figure 1. Summary Dialog screen. Figure 2. Battery Status screen of a TELIGEN ICD 1

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

A Patient EDUCATION HANDBOOK ON ICD. A Patient Education Handbook on ICD 1

Sudden Cardiac Death in Youth Athletes. Dublin City Schools Athletic Health Care

How to Perform CPR on a Cardiac Arrest Victim (Adult Victim Only)

Implantable Cardioverter Defibrillators (ICDs) How ICDs Treat Fast Heart Rhythms Before, During, and After Implantation Living Well with an ICD

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2

Transcription:

British Society for Dermatological Surgery (BSDS) & British Heart Rhythm Society (BHRS) Guidance on Implanted Devices & Dermatological Surgery Contents Page 1. Introduction 2. Summary & PPM 3. ICD & ILR 4. Cardiac physiology reports & bookings 5. References & Table: Types of electrosurgery with relevance to commonly used Dermatological equipment 6. Flowchart Glossary: Permanent Pacemakers (PPM), Implantable Cardiac Defibrillators (ICD), Implantable Loop Recorders (ILR), Electromagnetic Interference (EMI) Introduction: At present there is no specific national safety guidance available for clinicians who perform cutaneous surgery on patients with cardiac rhythm devices. The lower risk of skin surgery procedures and the electrosurgical devices that are typically used means some of the generic guidance is not so applicable to local anaesthetic skin surgery and may have significant resource implications. This guidance has therefore been developed to complement and clarify the detailed advice published by the BHRS 1 and MHRA 2 that was devised primarily for the general theatre and endoscopy setting, rather than the outpatient procedure room or clinic setting in which skin surgery is often performed. Skin surgery procedures are some of the commonest performed in the UK. A typical Dermatology department conducts well over 2000 procedures per year, and found 2% of these patients had implanted cardiac devices. 3 Therefore any advice can have significant ramifications for patient pathways and appropriate use of resources, as well as safety. We reviewed the published literature, contacted the Medicines & Healthcare products Regulatory Agency (MHRA), and conducted a national survey of members of the BSDS to determine if electromagnetic interference had been reported in practice in the context of skin surgery. We found no reported or published instances of any adverse events. Nevertheless definite risks still exist. It is possible that adverse events have occurred but either not been noticed immediately or not been reported. Precautions must still be taken. Page 1 of 6

Disclaimer: July 2017 v1.8 Review date: 2020 This is a guide only and clinical judgement should ultimately determine the degree of risk, particularly for complex patients. Advice from a multidisciplinary team may also be helpful. Limitations of Guidance: This document has been prepared on behalf of the BSDS and is based on the most relevant data and expert opinion available when the document was prepared. A full systematic literature review has not been conducted. It is recognised that under certain conditions it may be necessary to deviate from the guidelines and that the results of future studies may require some of the recommendations herein to be changed. Failure to adhere to these guidelines should not necessarily be considered negligent, nor should adherence to these recommendations constitute a defence against a claim of negligence. Limiting the review to English language references was a pragmatic decision but the authors recognise this may exclude some important information published in other languages. Summary of key points (also see flowchart) 1. Monitor the patient clinically and with pulse oximeter whilst using bipolar electrosurgery in short bursts >5cm away from a PPM. 2. In addition to (1) switch off any ICD immediately pre-op and back on immediately post-op. 3. Booking clerk or clinician to contact cardiac physiology pre-op: a. to request device report (see below) b. to arrange visit to theatre immediately pre-and post-op if ICD c. to request cardiac physiology have someone on standby on the procedure date if the surgical site is within 5 cm of the pacemaker box d. to allow an opportunity for ILR data to be downloaded pre-op if required. 4. If a PPM is under regular routine review by cardiac physiology then no additional pre- or postop checks are usually required. 5. If the patient has an ILR record the procedure in the patient s event diary Background Electrosurgery passes electricity through tissue to cause coagulation, damage, or cutting. It can interfere with the usual running of some implanted devices by EMI. Queries about these devices may cause concern, or there may be a risk to the safety of the patient or staff. Pacemakers (PPM) All pacemakers Pacemakers may regulate heart rhythm in a variety of ways, but of most relevance is when the patient is pacemaker dependent i.e. they have no natural pacemaker of their own, or their heart rate without the pacemaker is too low (leading to any of: lightheadedness, fainting, shortness of breath, chest pain, weakness, fatigue, collapse). In some cases the device could mistake the electrosurgery current for heart activity and stop pacing temporarily. This effect is unlikely if using bipolar electrosurgery, particularly if away from the device. Also short pulses can be used to minimise the effect of this Page 2 of 6

problem even if it does occur, as the device should resume pacing immediately after the electrosurgery pulse stops. If the patient s pulse remains steady through the electrosurgery pulse, it is obviously not having this effect in any case. Although theoretically possible, any reprogramming of the device by electrosurgery EMI is not considered a real risk. Pacemakers within 5cm of surgical site If the surgical site is within 5 cm of the pacemaker box (although the risk is still low with bipolar) the pacing team should be informed of the date that the patient is coming in to have someone on standby. In the unlikely case that the bipolar diathermy does inhibit the pacemaker (i.e. if the patient becomes symptomatic during electrosurgery use or if some inhibition is noted via pulse oximeter) the team will come and re-programme the device to an asynchronous mode for the duration of the procedure. Defibrillators (ICD) ICD often look and feel similar to a pacemaker, and are usually in the same site. There is a greater danger as these devices are constantly sensing for odd electrical activity which might indicate an imminent cardiac arrest. They then administer a shock directly to the heart via internal electrodes. Again the device could mistake the electrosurgery current for heart activity and shock the patient inappropriately. These therefore need to be switched off immediately before, and back on immediately after, procedures involving electrosurgery. The cardiac physiology technicians will do this in the operating theatre by prior arrangement. The device should be off for the minimum time possible i.e. only the period when electrosurgery is needed, not during LA infiltration, or recovery. It is critical that the WHO checklist is followed to prevent devices from inadvertently being left switched off. When booking surgery it is important to ask the patient, study the case-notes, and contact the cardiac physiology department that looks after the device, to ensure what you may presume to be a pacemaker is not also a defibrillator. The booking team should also obtain from the cardiac physiology team a summary of every device and the reason it was implanted. Loop Recorders (ILR) These are devices implanted usually in a similar site to record a constant running ECG whilst in place, when cardiac arrhythmias are suspected but have been difficult to prove (e.g. syncope, falls, palpitations). They are not a risk to patient or staff as they will merely show some extra electrical activity from the electrosurgery in their recording data, and neither are the devices at particular risk from the electrosurgery itself. The patient keeps a diary of any symptoms that they feel during the time the device is implanted, so they should merely add a note saying they had a procedure with electrosurgery at that time, to help the technicians to interpret the findings correctly. The follow-up centre should be informed that a procedure is planned, to give them the opportunity to download the stored data from the recorder beforehand. Other devices Other implanted pumps and brain stimulators etc may also be present. Be vigilant and seek advice from the relevant department if found. Intra-operative monitoring Continuous ECG using a cardiac monitor has previously been suggested, but is not usually helpful in the out-patient skin surgery setting. The monitor itself readily picks up EMI from the electrosurgery which obscures the patient s heart rhythm at the very moment the surgeon wishes to see the effect. Also with the surgeon busy, no anaesthetist, and not always a scrub nurse, the monitor may not actually be checked during the procedure, even if attached. Although it might be useful for the resuscitation team to view the aberrant rhythm quickly if an ICD patient suffered cardiac arrest whilst their device was switched off, this is extremely unlikely if the proper precautions are taken. Also many Page 3 of 6

out-patient or day case surgery units may have an Automated External Defibrillator (AED) on the crash trolley, the use of which does not require viewing the heart rhythm (and the cardiac monitor leads may actually get in the way). The cardiac monitor is also slow to fit and prone to lead connection problems, which can delay surgery. Instead a pulse oximeter can be used for heart rate monitoring, is quick and unobtrusive to fit, often can be monitored audibly by the surgeon themselves if necessary, and is not susceptible to EMI from electrosurgery. Device reports from cardiac physiology team A summary of every device and the reason it was implanted is needed: To identify ICD+PPM devices To ensure device status and indication known (especially if patient from out of area or moved since implantation) To ensure device is not nearing end of life/battery change To give the cardiac physiology team an opportunity to arrange a device check pre-op if required To give clinician definitive information on the indication for the device (this will not usually affect their peri-operative management, but gives an awareness of level of risk) Organisation of bookings Booking clinician to ask patient where are you normally seen for your PPM/ICD? when completing the electronic surgical booking form (to help obtain accurate device/indication report - may be another hospital) Booking admin team to contact cardiac physiology to obtain device/indication report prior to dating, and give to booking clinician for comment 1 week minimum notice needed for cardiac physiology team for routine ICD patient bookings (otherwise by special arrangement if possible) Call cardiac physiology team at end of case prior to ICD patient to allow time to attend department (and update them during surgery if procedure duration changes) Authors Dr Adam Bray, Consultant Dermatologist, Dermatological & Mohs Surgeon. BSDS Guidance Lead. adam.bray@uhbristol.nhs.uk Mr Gareth Davies, Lead Cardiac Devices Physiologist, Bristol Heart Institute. Dr Beth Wright, Speciality Registrar, Dermatology. University Hospitals Bristol NHS Foundation Trust. Acknowledgements: Many thanks to the British Heart Rhythm Society (BHRS) Council, the Executive Committee of The British Society for Dermatological Surgery (BSDS), and Mr Simon Page 4 of 6

Holmes Senior Medical Device Specialist Device Safety and Surveillance MHRA, for their contributions to this guidance. References: 1. British Heart Rhythm Society Guidelines for the management of patients with cardiac implantable electronic devices (CEIDs) around the time of surgery. Thomas H, Turley A, Plummer C, on behalf of BHRS Council January 2016 http://www.bhrs.com/guidelines-information Accessed on: 7 th May 2017 2. Guidelines for the perioperative management of patients with implantable pacemakers or implantable cardioverter defibrillators, where the use of surgical diathermy/electrocautery is anticipated. Medicines & Healthcare products Regulatory Agency (MHRA) 2006 3. Bristol Dermatology Centre. Personal correspondence. Types of electrosurgery with relevance to commonly used Dermatological equipment Type of Electrosurgery Equipment comments Description Monopolar single electrode Conmed Hyfrecator with normal handle probe attachment Current passes from probe to patient and disperses through the body (possible as it is only a small amount of current and table is insulated) Bipolar CoaComp M unit Conmed Hyfrecator with bipolar forceps attachment Current passes from 1 side of forceps to the other through a small area of the patient s tissue Monopolar with return plate Unusual in Dermatological surgery, commoner in main theatres Current passes from probe through patient to a separate plate (earth electrode) connected to machine Electrocautery Largely replaced by other electrosurgery machines but some older units may still be used or battery powered e.g. Bovie Using current to heat a metal probe (like a soldering iron) to cause coagulation or damage to tissue Page 5 of 6

Page 6 of 6 July 2017 v1.8 Review date: 2020