Australian Per Caput Doses from Diagnostic Imaging and Nuclear Medicine

Similar documents
Radiation exposure of the Italian population from diagnostic radiology and nuclear medicine examinations

Attention. Therefore, all the data and statements made in this presentation are preliminary and might change in the future.

Ionizing Radiation Exposure of the Population of the United States. David A. Schauer Executive Director

ROMANIA MEDICAL EXPOSURE IN 2016 IN ROMANIAN RADIOLOGICAL DEPARTMENTS. Olga GIRJOABA and Alexandra CUCU

ICRP Symposium on the International System of Radiological Protection

Introduction to RP154 methology for population dose estimates

Trends in examination frequency and population doses in Norway,

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Outline. NCRP Scientific Committee 6-2

People Exposed to More Radiation from Medical Exams

Diagnostic Reference Levels (DRLs)

3 rd International Symposium on the System of Radiological Protection Seoul, October John Harrison

Practice and Risk at Medical Facilities in Agency Operations

Managing Patient Dose in Computed Tomography (CT)

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February Radiation Safety

Population Dose from General X-ray and Nuclear Medicine: 2010

HEALTHFIRST 2011 RADIOLOGY PROGRAM CODE LIST

Dose Estimates for Nuclear Medicine Procedures: What are they? Where do they come from?

ADI Procedure Codes. August 2016 Revised April 2017 Page 1 of 7 ADI Procedure Codes

HIP RADIOLOGY PROGRAM CODE LISTS

Article Excerpts: Radiation protection concerns among staff performing Fluoroscopic procedures.

Radiation Protection- Cath lab

JHM-IRB Guidelines for Radiation Statements

EXPOSURE FROM DIAGNOSTIC NUCLEAR MEDICINE PROCEDURES. 14 Victor Babes Street, RO-6600, Iasi, Romania

Recent Progress in Radiation Dosimetry for Epidemiology and Radiological Protection. John Harrison ICRP Committee 2

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

Chapter 16 Worksheet Code It

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

JMSCR Vol 06 Issue 12 Page December 2018

MAGNETIC RESONANCE IMAGING (MRI) AND COMPUTED TOMOGRAPHY (CT) SCAN SITE OF CARE

Variation of Occupational Doses among Subspecialties in Diagnostic Radiology. A.N. Al-Haj, C.S. Lagarde, A.M. Lobriguito

Potential Communication Benefits by adoption of SI units in Radiation Safety

AIM 2014 CPT Radiology & Cardiac Codes Requiring Review

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

Radiology Codes Requiring Authorization*


Spine MRI and Spine CT Test Request Tip Sheet

Justification, Optimization and Communication in Pediatric CT Imaging: Recent Improvements and Persistent Challenges Designated Emphasis in Nuclear

THE UTILIZATION OF A DOSE MANAGEMENT SOLUTION TO LOWER RADIATION DOSES IN MEDICAL IMAGING

Interprovincial Out-Patient Rates Effective April 1, 2016

The activity. Suggested Answers

ALTERNATIVES TO THE EFFECTIVE DOSE FOR STOCHASTIC RISK ASSESSMENT IN MEDICAL IMAGING

Evening session 1 Stakeholder platform opportunity 18:30-20:00 IAEA International Atomic Energy Agency

Background Radiation in U.S. ~ msv/yr msv/yr ~0.02 ~0.02 msv msv/day /day (~2 m rem/day) mrem/day) NCRP 4

Early in Life. Noboru Takamura, M.D., Ph.D.

The New Jersey Radiographic Quality Assurance Program at 5 Years

X-rays How safe are they?

Spine MRI and Spine CT Test Request Tip Sheet

Radiation related cancer risk & benefit/risk assessment for screening procedures

Iranian physicians' knowledge about radiation dose, received by patients in diagnostic radiology

2010 Radiology Prior Authorization List for UnitedHealthcare s HealthChoice Members

MOLINA HEALTHCARE OF MICHIGAN PRIOR AUTHORIZATION / PRE-SERVICE REVIEW GUIDE IMAGING CODES REQUIRING PRIOR AUTHORIZATION EFFECTIVE 1/1/2014

Spine MRI and Spine CT Test Request Tip Sheet

The Fluoroscopic Technique for Monitoring Distraction of a Non-Invasive Lengthening Device in Early Onset Scoliosis

Why radiation protection matters?

2012 CPT Radiology Codes Requiring Review Blue Cross and Blue Shield of Louisiana

Radiology. General radiology department. X-ray

CPT CODES. Ph: (307) Fax: (307) CATSCAN IV Contrast: 87.00

Diagnostic Ionizing Radiation Exposure in a Population-Based Cohort of Children with Inflammatory Bowel Disease

Interprovincial Billing Out-Patient Rates Effective for Visits on or After April 1, 2017

UNMH Radiology Clinical Privileges. Name: Effective Dates: From To

Shielding Calculation: Radiographic Room. Jerry Williams

Radiation Doses in Radiology: Influence of Standards and Regulations

Population Collective Effective Dose from Nuclear Medicine procedures in Greece

Spine MRI and Spine CT Test Request Tip Sheet

High Tech Imaging Quick Reference Guide

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Certification Review. Module 28. Medical Coding. Radiology

Austin Radiological Association Ga-68 NETSPOT (Ga-68 dotatate)

Austin Radiological Association Nuclear Medicine Procedure PET SODIUM FLUORIDE BONE SCAN (F-18 NaF)

Current and Planned NCRP Activities

RADIOLOGIC TECHNOLOGY PROGRAM COURSE DESCRIPTIONS

Survey of patients CT radiation dose in Jiangsu Province

Radiation Dosimetry for CT Protocols

Ask EuroSafe Imaging Tips & Tricks. CT Working Group. CT in Pregnancy

Radiation Safety For Anesthesiologists. R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua

BlueAdvantage SM. & BlueChoice SM Radiology Prior Authorization Program Code List CPT /HCPS

JUSTIFICATION PROTOCOLS FOR CT SCANNING ALBURY WODONGA HEALTH WODONGA CAMPUS

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

Radiation Dose in X-Ray and CT Exams

Screening tests. When you need them and when you don t

Course specification

Interprovincial Billing Out-Patient Rates Effective for Visits on or After April 1, 2018

Description MRI, TMJ C T Head Without Contrast C T Head With Contrast C T Head Without & With Contrast

Radiation Dose Rates from Patients Administrated Radiopharmaceuticals Used for Brain Blood Flow Investigation.

2013 Coding Changes. Diagnostic Radiology. Nuclear Medicine

Trends in Patients Exposure Doses during Radiographic Examination in Japan

Ask EuroSafe Imaging Tips & Tricks. CT Working Group. On the use of Diagnostic Reference Levels in CT

Current and Planned Reports and Conferences of the National Council on Radiation Protection and Measurements

Diagnostic reference levels as a quality assurance tool

RADIOLOGY PROGRAM TABLE OF CONTENTS. OVERVIEW. . Assessment... and... Certification

Interprovincial Billing Out-Patient Rates Effective for Visits on or after September 1, 2017

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

HCTC/SKCTC Regional Radiography Program. IMG Course Descriptions and Student Learning Outcomes

screening; including image post processing CT, heart; without contrast material; with Requires authorization

73725x2 MRA Pelvis Runoff (to ankle) CTA Abdomen with & without CTA Cardiac Brain without 70551

Residents should be able to understand the role of imaging-guidance to facilitate common radiological procedures.

Clinical Programme. Diagnostic Imaging

Estimates of Risks LONG-TERM LOW DOSE EFFECTS OF IONIZING RADIATION

Transcription:

Australian Per Caput Doses from Diagnostic Imaging and Nuclear Medicine A.J.M Hayton, P.N Johnston, J Baldas, P.A Marks, K Edmonds, A.B Wallace Australian Radiation Protection and Nuclear Safety Agency, 619 Lower Plenty Road, Yallambie 3085, Australia The Australian use of CT increased strongly from its introduction in 1976 until 2010 with an associated increase in per caput dose from the use of radiation in medicine. Per caput doses from medical uses of radiation are now comparable to the natural background as the largest source of radiation exposure to the Australian population. The frequency of diagnostic imaging of diagnostic imaging procedures together with estimates of doses per procedure are reviewed based on data from the Australian national health insurance scheme, estimates of procedures funded by other mechanisms and best estimates of doses per procedure. Estimates of doses per procedure are informed by data from a national survey of diagnostic reference levels commenced in 2011. Key Words: Dose, Per Caput, Computed Tomography Introduction The total ionising radiation dose to the Australian population comes from two distinct sources, natural and man made. The dose from natural sources remains relatively constant at approximately 1.5 msv per year[1], however, the dose from man made sources is subject to greater variation and needs to be re-evaluated more often. Man made sources of ionising radiation include industrial sources, consumer products, nuclear fallout, and most importantly medical sources, in particular medical imaging. A 2009 report by the NCRP showed that medical sources contributed 48% of the total ionising radiation dose to the US public[2] and this percentage is expected to be similar in Australia. The ionising radiation dose to the Australian population as a result of medical imaging has been estimated twice in recent years. In 1999 it was estimated to be 0.8 msv per year[1] and in 2004 it was estimated to be 1.3 msv per year[3], showing an increase of 0.5 msv in just five years. This increase is largely due to the growth of Computed Tomography (CT), which was first introduced in Australia in 1976 and, with its superior image quality, has been gaining popularity every since. Between 2000 and 2010 the number of CT procedures performed in Australia grew by an average of 7.2% per year, more than four times greater than the average population growth over the same time period of 1.6% per year. In addition to the popularity of CT procedures there are also the relatively large associated patient doses. While a standard chest x-ray will result in a patient dose of approximately 0.01 msv, a standard chest CT will result in a patient dose of approximately 2.5 msv[4]. Considering the last dose estimate for medical imaging was made in 2004 and the rapid growth of CT procedures, it is time this dose estimate was updated.

Methods The ionising radiation dose to the Australian population as a result of medical imaging, both diagnostic imaging and nuclear medicine, was estimated for the year 2010. An individual estimate was made of the dose per caput from five different imaging modalities then these estimates were summed together to give a total dose for medical imaging. The five modalities were CT, Interventional procedures, Nuclear Medicine, General Radiography/Fluoroscopy and Mammography. The estimate for each modality was based on frequency data obtained from Medicare Australia[5] or BreastScreen Australia[6], with dose estimates obtained from the Dose Datamed study[4], an Australian and New Zealand Nuclear Medicine Survey[7] and Radiation Dose to Patients from Radiopharmaceuticals[8]. Frequency Data For all modalities except mammography, frequency data was obtained from Medicare Australia[5], the Australian national health insurance scheme. Medicare Australia records statistics in terms of item numbers, with each item number representing a specific procedure. The statistics for each item number are broken down in terms of geographical state, age and gender of the patient. For the mammography dose estimate frequency data from BreastScreen Australia[6] was used as these procedures are not captured by Medicare Australia. This data consisted of the total procedures recorded over a number of years only, no geographical state or age of patient information was obtained. Data Gap Medicare Australia does not capture all medical imaging procedures performed in Australia. Procedures that are not captured include public hospital patient radiology, military veterans, workplace accident referrals and transport accident referrals. A previous CT dose survey[3] conducted by the Australian Radiation Laboratory in the mid 90 s collected data on the percentage of procedures performed at practices that were not captured by Medicare Australia. A result of 24% was obtained indicating that Medicare Australia data represented only 76% of actual procedures performed for Computed Tomography. To account for this, a factor of 1.32 was applied to the frequency data for the four modalities using Medicare Australia data.

Dose Estimates Dose Datamed Study The Dose Datamed Study[4] involved a survey of a wide range of common radiological procedures in ten European countries. From the results of the survey a list of the Top 20 exams were outlined that were considered to make the largest contribution to population dose from medical imaging, see Table 1. For each of the Top 20 exams, a low, average and high exposure group was defined with an effective dose specified for each exam and exposure group. The low/high exposure groups being the average effective dose from the two countries with the lowest/highest doses and the average exposure group being the average effective dose from all ten participating countries. Table 1. Dose Datamed Top 20 Exams and mean effective doses[4].(colour added for this publication). Exam type Mean E per examination (msv) Average exposure group Higher exposure group Lower exposure group 1. Chest/Thorax 0.25 0.10 0.03 2. Cervical Spine 0.70 0.27 0.04 3. Thoracic Spine 2.00 1.00 0.40 4. Lumbar Spine 2.80 1.90 0.50 5. Mammography 0.40 0.33 0.25 6. Abdomen 1.80 1.50 0.50 7. Pelvis & Hip 1.35 0.90 0.45 8. Ba meal 15.00 7.70 2.60 9. Ba enema 12.50 8.60 6.40 10. Ba follow 24.50 10.00 4.40 11. IVU 3.50 4.00 2.60 12. Cardiac Angio. 11.25 9.10 5.30 All Angiography 8.60 9.20 7.30 13. CT head 2.40 2.00 1.60 14. CT neck 2.80 2.50 2.40 15. CT chest 8.20 8.00 6.60 16. CT spine 6.00 5.30 3.60 17. CT abdomen 13.50 12.00 10.20 18. CT pelvis 8.80 8.70 8.70 19. CT trunk 24.40 14.00 10.40 All CT 7.05 6.10 5.35 20. PTCA 17.00 14.00 13.151 All Interventional 15.35 10.70 6.50

Nuclear Medicine Survey The Dose Datamed list of Top 20 exams does not include any nuclear medicine procedures. This is because nuclear medicine procedures are much less common in European practice. Australian practice however, shares more similarity with US practice in that the frequency of nuclear medicine procedures is higher. The dose estimates for nuclear medicine were obtained from a joint Australian and New Zealand survey[7]and Radiation Dose to Patients from Radiopharmaceuticals[8]. Results Computed Tomography Seven of the Top 20 exams related to CT, see Table 1. The frequency of each exam was calculated by grouping the item numbers relevant to each exam and taking the sum of the procedures in each group for the calendar year 2010. The frequency of each exam type was multiplied by the dose per exam for each exposure level, see Table 2. The total dose from the seven exam types was then divided by the Australian population[9] in 2010 to obtain a dose per person value for each of the exposure groups, the highest being 1.11 msv per person per year, these are also shown in Table 2. Table 2. Frequency and dose data for CT exams. Dose Dose per procedure Datamed Frequency (msv) Total dose (msv) Category Low Average High Low Average High CT Head 736637 1.6 2.0 2.4 1178618 1473273 1767928 CT Neck 136644 2.4 2.5 2.8 327945 341609 382603 CT Chest 346375 6.6 8.0 8.2 2286072 2770997 2840272 CT Spine 357543 3.6 5.3 6.0 1287155 1894979 2145259 CT Abdo 24886 10.2 12.0 13.5 253837 298632 335960 CT Pelvis 24998 8.7 8.7 8.8 217484 217484 219984 CT Trunk 701628 10.4 14.0 24.4 7296926 9822785 17119711 Total Dose (msv) 12848038 16819759 24811716 Total Dose per person (msv) 0.58 0.76 1.11 Interventional Procedures Two of the Top 20 exams related to interventional procedures, Cardiac Angiography and Percutaneous Transluminal Coronary Angioplasty, (PTCA), see Table1. The total dose from these two procedures was calculated using the same method as described for CT and is shown in Table 3.The dose corresponding to the high exposure group was 0.12 msv per person per year.

Table 3. Frequency and dose data for Interventional procedures. Dose Datamed Category Frequency Dose per procedure Total dose (msv) (msv) Low Average High Low Average High Cardiac Angiography 204295 5.30 9.12 10.75 1082764 1863625 2196172 PTCA 31185 13.15 14.0 17.0 410083 436590 530145 Total Dose (msv) 1492847 2300215 2726317 Total Dose per person (msv) 0.07 0.10 0.12 Nuclear Medicine From the Medicare Australia item numbers eighteen categories based on body region were identified for nuclear medicine procedures. The frequency for each category was calculated using the same method as for CT. A dose estimate was made for each individual item number by multiplying the Most Common Activity (MCA) as obtained from the Australia and New Zealand joint survey[7] by the effective dose as obtained from Radiation Dose to Patients from Radiopharmaceuticals[8] in msv/mbq, see Equation 1. Effective Dose Most Common Activity ( MBq ) Absorbed Dose( msv / MBq ) Equation 1. Table 4 shows the frequency and total dose per category as well as the total dose per person per year of 0.11 msv. An effective dose per category could not be calculated due to their being a range of effective doses in each category.

Table 4. Frequency and total dose from nuclear medicine procedures. Body region Frequency Total Dose (msv) Heart 125108 1073706 Lung 22367 48634 Liver and Spleen 1090 5200 Gallbladder 7599 25838 GIT 8885 20022 Kidney 14543 29780 Urogenital 317 1648 Brain (CSF) 5415 20651 Bone (whole body) 143678 655173 Whole Body 6395 131106 Bone Marrow 154 224 Bone (limited) 70713 322450 Limited 2710 42641 Thyroid 33505 87113 Parthyroid 5669 35828 Adrenal 710 1846 Other 585 1050 Bone (infection) 342 2188 Total Dose (msv) 22505099 Total Dose per person (msv) 0.11 General Radiography/Fluoroscopy The remaining eleven of the Top 20 exams related to general radiography/fluoroscopy, see Table1. Mammography was included in this list but the Australian dose from mammography was calculated separately as its own modality. This was done because Medicare Australia captures only a small proportion of all breast screening procedures and also because the dose from mammography alone was sought. A number of high frequency item numbers relating to imaging of the spine were not compatible with the three spine exam types specified in Table 1, so to insure these procedures were included in the calculation an addition exam type Whole Spine was added to this list. A dose estimate for the Whole Spine exam was conservatively estimated by summing the dose from the Cervical, Thoracic and Lumbar Spine exam types. The total dose for the eleven procedures was then calculated using the same method as described for CT and is shown in Table 5. The dose corresponding to the high exposure group was 0.30 msv per person per year.

Table 5. Frequency and dose data for General Radiography/Fluoroscopy exams. Dose Datamed Dose per procedure Frequency Category (msv) Total dose (msv) Low Average High Low Average High Chest 2402223 0.03 0.12 0.21 72067 285865 504467 C Spine 359363 0.05 0.27 0.70 16171 98226 251554 T Spine 94187 0.70 0.70 0.70 65931 65931 65931 L Spine 456472 0.50 1.86 2.90 228236 849038 1323768 Whole Spine 429807 1.25 2.81 4.30 535109 1206037 1848168 Abdo 271832 0.50 1.53 1.80 135916 415902 489297 Pelvis 559889 0.45 0.89 1.35 251950 498301 755850 Ba meal 77999 2.65 7.71 15.05 206697 601371 1173882 Ba enema 2219 6.45 8.60 12.35 14312 19083 27404 Ba follow 5067 4.40 10.48 28.90 22297 53096 146450 IVU 3172 2.55 4.03 3.50 8088 12783 11102 Total Dose (msv) 1556774 4105632 6597873 Total Dose per person (msv) 0.07 0.18 0.30 Mammography Frequency data for mammography screening was obtained from BreastScreen Australia[6] in the form of total procedures for a series of financial years. From this data a frequency (total) for the calendar year of 2010 was estimated. The total dose for mammography was then calculated using the same method as described for CT and is shown in Table 6. The dose corresponding to the high exposure group was 0.03 msv per person per year. Table 6. Frequency and dose data for Mammography Dose Datamed Category Frequency Dose per procedures (msv) Total dose (msv) Low Average High Low Average High Mammography 1471000 0.25 0.33 0.40 367750 485430 588400 Total Dose (msv) 367750 485430 588400 Total Dose per person (msv) 0.02 0.02 0.03 Total Dose Estimate A summary of the dose estimates from each of the five modalities is shown in Table 7. The total dose per person per year varies from 0.9 msv for the low exposure group to 1.7 msv for the high exposure group.

Table 7. Summary of the dose per person for each exposure group and the collective total. Modality Dose (msv) Low Average High Computed Tomography 0.58 0.76 1.11 Interventional Procedures 0.07 0.10 0.12 Nuclear Medicine 0.11 General Radiography/Fluoroscopy 0.08 0.19 0.30 Mammography 0.02 0.02 0.03 Total Dose per person (msv) 0.9 1.2 1.7 The exposure group that best represents Australian practice was chosen by comparing CT doses from the mid 90 s survey[3] with those calculated using retrospective Medicare Australia frequency data and the CT doses from the Top 20, see Figure 1. Figure 1. The Australian Computed Tomography dose per person, a comparison with previous survey data. Figure 1 indicates that Australian CT doses are comparable with the High Exposure Group of the Dose Datamed study, and as such the dose to the Australian population from diagnostic imaging and nuclear medicine is estimated to be 1.7 msv per person per year, see Table 7.

Discussion There is uncertainty in both the frequency data and the dose estimates used. The uncertainty in the dose estimates used is thought to make the largest contribution to uncertainty in the overall dose but it is hard to quantify. The total dose of 1.7 msv was chosen based on a comparison with dose estimates made nearly twenty years ago. While it is possible to observe trends in procedure frequency, significant changes in technology mean that trends in dose are not directly proportional to frequency and hence cannot be estimated based on frequency alone. An Australian National Diagnostic Reference Level Survey is currently underway which will provide up to date dose data for a number of common Computed Tomography procedures which may lead to a recalculation of the dose from CT alone. It is intended that the National Diagnostic Reference Level Survey will be expanded in the future to cover interventional fluoroscopy, nuclear medicine, mammography and general radiography, which should provide more relevant dose data for all modalities. The uncertainty in the frequency data could be quantified by an uncertainty in the data gap between procedures captured by Medicare Australia and the total procedures that are actually performed. The estimate that 24% of procedures are not captured by Medicare Australia is currently the only one available in reference to the adult population. It is also acknowledged that 24% relates specifically to CT procedures and there is little justification, other than that it is the only percentage available, for its application to other modalities. That being said, it is consistent with the percentage reported in recent discussions with nuclear medicine practices. The decision to use 24% across all modalities calculated using frequency data obtained from Medicare Australia is considered to produce a conservative dose estimate. Conclusion The total ionising radiation dose to the Australian population as a result of diagnostic imaging and nuclear medicine was estimated to be 1.7 msv per person per year, This is considered to be a conservative estimate in relation to uncertainty in the frequency estimate. The uncertainty in the dose estimates is considered to be large but unknown. The results of current and future surveys of Australian doses from medical imaging will help refine this value.

References 1. Webb DV, Solomon SB, Thomson JEM. Background Radiation Levels and Medical Exposure Levels in Australia. Radiation Protection in Australasia 1999;16(1):7-14. 2. NCRP. Ionizing Radiation Exposure of the Population of the United States. NCRP Report 160. Bethesda, MD: National Council on Radiation Protection and Measurements (NCRP); 2009. 3. Wise KN, Thomson JEM. Changes in CT radiation doses in Australia from 1994 to 2002. The Radiographer 2004;51(2):81-5. 4. Commission E. European Guidance on Estimating Population Doses from Medical X-Ray Procedures. Radiation Protection No 154 2008. 5. Services AGDoH. Medicare Benefits Schedule. MBS Statistics 2010 [cited 2011; Available from: http://www.medicareaustralia.gov.au/provider/medicare/mbs.jsp#n10030 6. BreastScreen Australia monitoring report 2006-2007 and 2007-2008: Australian Institute of Health and Welfare; 2010. 7. Botros G, Smart RC, Towson JE. Diagnostic reference activities for nuclear medicine procedures in Austrlaia and New Zealand derived from the 2008 survey. ANZ Nuclear Medicine 2009;40(4):2-11. 8. ICRP. Radiation Dose to Patients from Radiopharmaceuticals - Addendum to ICRP 53. 1998. 9. Statistics ABo. Australian Demographic Statistics. [cited 2011; Available from: www.abs.gov.au/ausstats/abs@.nf/detailspage/3105.0.65.0012008?opendocument