Principles of intra-aortic balloon pump counterpulsation

Similar documents
Section 6 Intra Aortic Balloon Pump

IABP Timing & Fidelity. Pocket Reference Guide

CLINICAL SUPPORT SERVICES DEVELOPING AN IABP PRECEPTOR STRATEGY

The Role of Mechanical Circulatory Support in Cardiogenic Shock: When to Utilize

Useful? Definition of High-risk? Pre-OP/Intra-OP/Post-OP? Complication vs Benefit? Mortality? Morbidity?

Hemodynamic Monitoring and Circulatory Assist Devices

Rhondalyn C. McLean. 2 ND YEAR RESEARCH ELECTIVE RESIDENT S JOURNAL Volume VII, A. Study Purpose and Rationale

Cath Lab Essentials : LV Assist Devices for Hemodynamic Support (IABP, Impella, Tandem Heart, ECMO)

Indications of Coronary Angiography Dr. Shaheer K. George, M.D Faculty of Medicine, Mansoura University 2014

Assist Devices in STEMI- Intra-aortic Balloon Pump

Mechanics of Cath Lab Support Devices

ISPUB.COM. Intra Aortic Balloon Pump (IABP) Counterpulsation. P Oberwalder HISTORY PHYSIOLOGIC EFFECTS OF IABP THERAPY

Mechanics of Cath Lab Support Devices

Cardiac anaesthesia. Simon May

Counterpulsation. John N. Nanas, MD, PhD. Professor and Head, 3 rd Cardiology Dept, University of Athens, Athens, Greece

University of Florida Department of Surgery. CardioThoracic Surgery VA Learning Objectives

HOW TO PERFORM LEFT VENTRICULAR ASSISTANCE IN THE CATHLAB. Andreas Baumbach, MD FESC FRCP Bristol Heart Institute University Hospitals Bristol UK

Management of Cardiogenic Shock. Dr Stephen Pettit, Consultant Cardiologist

CSC Review Course: Intra Aortic Balloon Pump (IABP) Paula M Davis ARNP, MSN, CCRN-CSC

Percutaneous Mechanical Circulatory Support Devices

Experience with Intraaortic Balloon Counterpulsation

The use of Intra Aortic Baloon Pump in Patients undergoing Coronary Artery Bypass Grafting at the Aga Khan University Hospital, Karachi

Complications of Acute Myocardial Infarction

Intraaortic Balloon Counterpulsation- Supportive Data for a Role in Cardiogenic Shock ( Be Still My Friend )

Diastolic Augmentation with an External Pulsating Device To Treat Cardiogenic Shock

THE journal OF EXTRA-CORPOREAL TECHNOLOGY ABSTRACT

Echocardiography as a diagnostic and management tool in medical emergencies

Rationale for Prophylactic Support During Percutaneous Coronary Intervention

Epidemiology of Heart Failure in Adults

Cardiogenic Shock. Carlos Cafri,, MD

Ventricular Assisting Devices in the Cathlab. Unrestricted

Impedance Cardiography (ICG) Method, Technology and Validity

Bridging With Percutaneous Devices: Tandem Heart and Impella

CARDIOGENIC SHOCK. Antonio Pesenti. Università degli Studi di Milano Bicocca Azienda Ospedaliera San Gerardo Monza (MI)

Interventional procedures guidance Published: 26 September 2014 nice.org.uk/guidance/ipg504

Adult Echocardiography Examination Content Outline

Cardiovascular Nursing Practice: A Comprehensive Resource Manual and Study Guide for Clinical Nurses 2 nd Edition

PROBLEM SET 2. Assigned: February 10, 2004 Due: February 19, 2004

National Imaging Associates, Inc. Clinical guidelines CARDIAC CATHETERIZATION -LEFT HEART CATHETERIZATION. Original Date: October 2015 Page 1 of 5

CVICU EXAM. Mrs. Jennings is a 71-year-old post-op CABG x5 with an IABP in her left femoral artery

Detailed Order Request Checklists for Cardiology

Index. K Knobology, TTE artifact, image resolution, ultrasound, 14

Percutaneous Mechanical Circulatory Support for Cardiogenic Shock. 24 th Annual San Diego Heart Failure Symposium Ryan R Reeves, MD FSCAI

Pre-operative usage of IABP for patients for by pass surgery

The Cardiac Cycle Clive M. Baumgarten, Ph.D.

DECLARATION OF CONFLICT OF INTEREST

Cardiogenic Shock Protocol

EACTS Adult Cardiac Database

Cardiothoracic Fellow Expectations Division of Cardiac Anesthesia, Beth Israel Deaconess Medical Center

To ECMO Or Not To ECMO Challenges of venous arterial ECMO. Dr Emily Granger St Vincent s Hospital Darlinghurst NSW

Ejection across stenotic aortic valve requires a systolic pressure gradient between the LV and aorta. This places a pressure load on the LV.

Hemodynamic Monitoring in the CCU. Edward G. Hamaty Jr., D.O. FACCP, FACOI

Aortic Dissection Causes of Death

Acute heart failure: ECMO Cardiology & Vascular Medicine 2012

AllinaHealthSystem 1

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018

Cardiovascular Physiology

Relax and Learn At the Farm 2012

Disclosures. Objectives 10/11/17. Short Term Mechanical Circulatory Support for Advanced Cardiogenic Shock. I have no disclosures to report

Shock Quiz! By Clare Di Bona

Intra-operative Echocardiography: When to Go Back on Pump

Ejection across stenotic aortic valve requires a systolic pressure gradient between the LV and aorta. This places a pressure load on the LV.

Mechanical circulatory support in cardiogenic shock The Cardiologist s view ACCA Masterclass 2017

ECMO and VAD implantation

Cardiac Cycle MCQ. Professor of Cardiovascular Physiology. Cairo University 2007

A case of post myocardial infarction ventricular septal rupture CHRISTOFOROS KOBOROZOS, MD

Οξύ στεφανιαίο σύνδρομο και καρδιογενής καταπληξία. Επεμβατική προσέγγιση. Σωτήριος Πατσιλινάκος Κωνσταντοπούλειο Γ.Ν. Ν. Ιωνίας

(For items 1-12, each question specifies mark one or mark all that apply.)

Circulatory Support: From IABP to LVAD

Echo assessment of patients with an ECMO device

Cath Lab Essentials: Basic Hemodynamics for the Cath Lab and ICU

PHYSIOLOGY MeQ'S (Morgan) All the following statements related to blood volume are correct except for: 5 A. Blood volume is about 5 litres. B.

Inflation and deflation timing of the AutoCAT 2 WAVE intra-aortic balloon pump using the autopilot mode in a clinical setting

Emergency surgery in acute coronary syndrome

ORIGINAL ARTICLE. Alexander M. Bernhardt a, *, Theo M.M.H. De By b, Hermann Reichenspurner a and Tobias Deuse a. Abstract INTRODUCTION

Debate in Management of native COA; Balloon Versus Surgery

AORTIC DISSECTIONS Current Management. TOMAS D. MARTIN, MD, LAT Professor, TCV Surgery Director UF Health Aortic Disease Center University of Florida

External Counterpulsation

The development of cardiogenic shock portends an extremely poor prognosis. Cardiogenic Shock: A Lethal Complication of Acute Myocardial Infarction

Mechanical Cardiac Support in Acute Heart Failure. Michael Felker, MD, MHS Associate Professor of Medicine Director of Heart Failure Research

Καθετηριασμός δεξιάς κοιλίας. Σ. Χατζημιλτιάδης Καθηγητής Καρδιολογίας ΑΠΘ

Clinical Indications for Echocardiography

Review of Cardiac Mechanics & Pharmacology 10/23/2016. Brent Dunworth, CRNA, MSN, MBA 1. Learning Objectives

Cite this article as:

Vascular complications of intra-aortic balloon insertion in patients undergoing coronary reavscularization: analysis of 911 cases

Yes No Unknown. Major Infection Information

8th Emirates Cardiac Society Congress in collaboration with ACC Middle East Conference Dubai: October Acute Coronary Syndromes

How to do Primary Angioplasty. - Patients with Cardiogenic Shock

Assessing Cardiac Risk in Noncardiac Surgery. Murali Sivarajan, M.D. Professor University of Washington Seattle, Washington

Cardiovascular System

12/19/2017. Learning Objectives. Mechanical Circulatory Support. Mechanical Aids to External Massage. Noninvasive Mechanical Support Devices

Recovering Hearts. Saving Lives.

Modern Left Ventricular Assist Devices (LVAD) : An Intro, Complications, and Emergencies

FUNDAMENTALS OF HEMODYNAMICS, VASOACTIVE DRUGS AND IABP IN THE FAILING HEART

Part 3a. Physiology: the cardiovascular system

7. Echocardiography Appropriate Use Criteria (by Indication)

CMS Limitations Guide - Cardiovascular Services

Acute dissections of the descending thoracic aorta (Debakey

Cardiovascular Physiology. Heart Physiology. Introduction. The heart. Electrophysiology of the heart

1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER

Transcription:

Principles of intra-aortic balloon pump counterpulsation Murli Krishna MBBS FRCA FFPMRCA Kai Zacharowski MD PhD FRCA Key points The primary goal of intraaortic balloon pump (IABP) treatment is to increase myocardial oxygen supply and decrease myocardial oxygen demand. Decreased urine output after the insertion of IABP can occur because of juxtarenal balloon positioning. Haemolysis from mechanical damage to red blood cells can reduce the haematocrit by up to 5%. Suboptimal timing of inflation and deflation of the balloon produces haemodynamic instability. An IABP is thrombogenic; always anticoagulate the patient. Never switch the balloon off while in situ. Murli Krishna MBBS FRCA FFPMRCA Consultant in Anaesthetics & Pain Medicine Frenchay Hospital Bristol BS16 1LE, UK E-mail: drmkrish@gmail.com Kai Zacharowski MD PhD FRCA Chair of Cardiovascular Anaesthesia and Critical Care Consultant in Anaesthesia and Critical Care Department of Anaesthesia Bristol Royal Infirmary Bristol BS2 8HW, UK Tel: þ44 117 928 2301/2365 Fax: þ44 117 926 8674 E-mail: kai.zacharowski@bristol.ac.uk (for correspondence) Intra-aortic balloon pump (IABP) remains the most widely used circulatory assist device in critically ill patients with cardiac disease. The National Centre of Health Statistics estimated that IABP was used in 42 000 patients in the USA in 2002. Advances in technology, including percutaneous insertion, smaller diameter catheters, sheathless insertion techniques, and enhanced automation, have permitted the use of counterpulsation in a variety of settings, with greater efficacy and safety. History Kantrowitz 1 described augmentation of coronary blood flow by retardation of the arterial pressure pulse in animal models in 1952. In 1958, Harken 2 suggested the removal of some of the blood volume via the femoral artery during systole and replacing it rapidly in diastole as a treatment for left ventricular (LV) failure, so called diastolic augmentation. Four years later, Moulopoulos and colleagues 3 developed an experimental prototype of an IABP whose inflation and deflation were timed to the cardiac cycle. In 1968, Kantrowitz 1 reported improved systemic arterial pressure and urine output with the use of an IABP in two subjects with cardiogenic shock, one of who survived to hospital discharge. Percutaneous IABs in sizes 8.5 9.5 French (rather than 15 French used earlier) were introduced in 1979, and shortly after this, Bergman and colleagues 4 described the first percutaneous insertion of IABP. The first prefolded IAB was developed in 1986. Basic principles of counterpulsation Counterpulsation is a term that describes balloon inflation in diastole and deflation in early systole. Balloon inflation causes volume displacement of blood within the aorta, both proximally and distally. This leads to a potential increase in coronary blood flow and potential improvements in systemic perfusion by augmentation of the intrinsic Windkessel effect, whereby potential energy stored in the aortic root during systole is converted to kinetic energy with the elastic recoil of the aortic root. Physiological effects of IABP therapy doi:10.1093/bjaceaccp/mkn051 24 Continuing Education in Anaesthesia, Critical Care & Pain Volume 9 Number 1 2009 & The Board of Management and Trustees of the British Journal of Anaesthesia [2009]. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org The primary goal of IABP treatment is to improve the ventricular performance of the failing heart by facilitating an increase in myocardial oxygen supply and a decrease in myocardial oxygen demand. The overall haemodynamic effects of IABP therapy are summarized in Table 1. Although these effects are predominately associated with enhancement of LV performance, IABP may also have favourable effects on right ventricular (RV) function by complex mechanisms including accentuation of RV myocardial blood flow, unloading the left ventricle causing reduction in left atrial and pulmonary vascular pressures and RV afterload. 5 IABP inflates at the onset of diastole, thereby increasing diastolic pressure and deflates just before systole, thus reducing LV afterload. The magnitude of these effects depends upon: (i) Balloon volume: the amount of blood displaced is proportional to the volume of the balloon. (ii) Heart rate: LV and aortic diastolic filling times are inversely proportional to heart rate; shorter diastolic time produces lesser balloon augmentation per unit time. (iii) Aortic compliance: as aortic compliance increases (or SVR decreases), the magnitude of diastolic augmentation decreases. Myocardial oxygen supply and demand Inflation of IAB during diastole increases the pressure difference between aorta and left

Table 1 Summary of haemodynamic effects of IABP therapy Aorta Left ventricle Heart Blood flow ventricle, the so-called diastolic pressure time index (DPTI). The haemodynamic consequence of this is an increase in coronary blood flow and, therefore, myocardial oxygen supply. Myocardial oxygen demand is directly related to the area under the LV systolic pressure curve, termed as tension time index (TTI). Balloon deflation during systole causes a reduction in the LV afterload, thereby decreasing TTI. Thus, the ratio of oxygen supply (DPTI) to oxygen demand (TTI), known as the endocardial viability ratio (EVR), should increase if the IABP is working optimally. This can be evidenced by a decrease in coronary sinus lactate. Coronary perfusion According to the Hagen Poiseuille principle, flow through a tube is directly proportional to the pressure difference across it and the fourth power of the radius while being inversely proportional to the length of the tube and the viscosity of fluid flowing through it. Hence, in patients with severe coronary artery disease in whom autoregulation is perceived to be absent, coronary blood flow is directly related to diastolic perfusion pressure. Therefore, IABP should theoretically improve coronary flow in these patients. Renal function Renal blood flow can increase up to 25%, secondary to increase in cardiac output. Decrease in urine output after insertion of IABP should raise the suspicion of juxta-renal balloon positioning. Haematological effects The haemoglobin levels and the haematocrit often decrease by up to 5% because of haemolysis from mechanical damage to the red blood cells. Thrombocytopenia can result from mechanical damage to the platelets, heparin administration, or both. 6 Indications #systolic pressure, "diastolic pressure #systolic pressure, #end-diastolic pressure, #volume, #wall tension #afterload, #preload, "cardiac output "! coronary blood flow Over the years, indications for the use of IABP have developed in clinical practice and are summarized along with contraindications in Table 2. Acute myocardial infarction IABP is aimed at achieving haemodynamic stability until a definitive course of treatment or recovery occurs. By decreasing myocardial work and SVR, intracardiac shunting, mitral regurgitation, or both (if present) are reduced while coronary perfusion is enhanced. Table 2 Indications and contraindications for the use of IABP therapy Indications Acute myocardial infarction Refractory LV failure Cardiogenic shock Refractory ventricular arrhythmias Acute MR and VSD Cardiomyopathies Catheterization and angioplasty Sepsis 9 Refractory unstable angina Infants and children with complex cardiac anomalies 10 Cardiac surgery Weaning from cardiopulmonary bypass Contraindications Absolute Relative Aortic regurgitation Uncontrolled sepsis Aortic dissection Abdominal aortic aneurysm Chronic end-stage heart disease with no Tachyarrhythmias anticipation of recovery Aortic stents Severe peripheral vascular disease Major arterial reconstruction surgery Severe mitral regurgitation secondary to papillary muscle dysfunction or rupture after myocardial infarction can lead to significant haemodynamic instability. This can initially be managed by IABP, pending definitive surgery. Ventricular arrhythmias IABP is also effective in stabilizing patients with refractory ventricular ectopy after myocardial infarction by increasing the coronary perfusion pressure, reducing ischaemia and trans-myocardial wall stress, and maintaining adequate systemic perfusion. Cardiogenic shock This is life-threatening complication of acute myocardial infarction, is characterized by low cardiac output, hypotension unresponsive to fluid administration, elevated filling pressures and tissue hypoperfusion leading to oliguria, hyperlactaemia, and altered mental status. IABP therapy is considered to be a class I indication (ACC/AHA guidelines) for the management of cardiogenic shock not rapidly reversed by pharmacological therapy. 7 Unstable angina Unstable angina refractory to drug treatment is an indication for IABP. These patients are at increased risk of developing acute myocardial infarction and death. By improving the haemodynamic condition of these patients, IABP can facilitate further percutaneous interventions or bridge the patient to surgery. Refractory ventricular failure IABP has a role in managing patients with refractory ventricular failure outside the setting of acute myocardial infarction, such as those with cardiomyopathy or severe myocardial damage associated with viral myocarditis. This can aid the progression to more definitive treatments such as ventricular assist device or cardiac transplantation. Continuing Education in Anaesthesia, Critical Care & Pain j Volume 9 Number 1 2009 25

Cardiac surgery IABP is used for stabilization of patients with acute myocardial infarction referred for urgent cardiac surgery. IABP support is often initiated in the cardiac catheterization laboratory and continued through the perioperative period. Elective placement is considered in high-risk patients such as those with significant left main stem disease, severe LV dysfunction (ejection fraction,30%), congestive heart failure, cardiomyopathy, chronic renal failure, or cerebrovascular disease. Weaning from cardiopulmonary bypass may be difficult in cases where aortic cross-clamping is prolonged, revascularization is only partially achieved, or preexisting myocardial dysfunction is present. Separation from cardiopulmonary bypass may be marked by hypotension and a low cardiac index despite the administration of inotropic drugs. The use of IABP in this setting decreases LV resistance, increases cardiac output, and increases coronary and systemic perfusion, facilitating the patient s weaning from cardiopulmonary bypass. Contraindications The contraindications to IABP are summarized in Table 2. It is contraindicated in patients with aortic regurgitation because it worsens the magnitude of regurgitation. IABP insertion should not be attempted in case of suspected or known aortic dissection because inadvertent balloon placement in the false lumen may result in extension of the dissection or even aortic rupture. Similarly, aortic rupture can occur if IABP is inserted in patients with sizable abdominal aortic aneurysms. Patients with end-stage cardiac disease should not be considered for IABP unless as a bridge to ventricular assist device or cardiac transplantation. IABP device placement should be avoided in patients with severe peripheral vascular disease. Percutaneous femoral IABP device insertion is contraindicated in the presence of bilateral femoral popliteal bypass grafts. Uncontrolled sepsis and bleeding diathesis are relative contraindications to the placement of IABP device. Technique of insertion and operation The IABP device has two major components: (i) a double-lumen 8.0 9.5 French catheter with a 25 50 ml balloon attached at its distal end; and (ii) a console with a pump to drive the balloon. The balloon is made of polyethylene and is inflated with gas driven by the pump. Helium is often used because its low density facilitates rapid transfer of gas from console to the balloon. It is also easily absorbed into the blood stream in case of rupture of the balloon. Before insertion, the appropriate balloon size is selected on the basis of the patient s height (as supplied by Datascope, for a patient,152 cm in height, a balloon volume of 25 cc is appropriate; for height between 152 and 163 cm, balloon volume 34 cc; for height 164 183 cm, balloon volume 40 cc, and for height.183 cm, balloon volume 50 cc). Smaller balloons are available for paediatric use. The diameter of the balloon, when fully expanded, Table 3 Complications associated with IABP Transient loss of peripheral pulse Limb ischaemia Thromboembolism Compartment syndrome 11 Aortic dissection Local vascular injury false aneurysm, haematoma, bleeding from the wound Infection Balloon rupture (can cause gas embolus) Balloon entrapment Haematological changes, for example thrombocytopenia, haemolysis Malpositioning causing cerebral or renal compromise Cardiac tamponade should not exceed 80 90% of the diameter of the patient s descending thoracic aorta. The IABP catheter is inserted percutaneously into the femoral artery through an introducer sheath using the modified Seldinger technique. Alternative routes of access include subclavian, axillary, brachial, or iliac arteries. The catheter can also be inserted surgically using a transthoracic or translumbar approach, but this is associated with an increased periprocedural mortality. 8 Once vascular access is obtained, the balloon catheter is inserted and advanced, usually under fluoroscopic guidance, into the descending thoracic aorta, with its tip 2 to 3 cm distal to the origin of the left subclavian artery (at the level of the carina). Intraoperatively, balloon placement can be ascertained using transoesophageal echocardiography. The outer lumen of the catheter is used for delivery of gas to the balloon and the inner lumen can be used for monitoring systemic arterial pressure. Complications associated with IABP are summarized in Table 3. The console is programmed to identify a trigger for balloon inflation and deflation. The most commonly used triggers are the ECG waveform and the systemic arterial pressure waveform. The balloon inflates with the onset of diastole, which corresponds with the middle of the T-wave. The balloon deflates at the onset of LV systole and this corresponds to the peak of the R-wave. Poor ECG quality, electrical interference, and cardiac arrhythmias can result in erratic balloon inflation. The balloon is set to inflate after the aortic valve closure (which corresponds to the dicrotic notch on the arterial waveform) and deflate immediately before the opening of the aortic valve (which corresponds to the point just before the upstroke on the arterial pressure waveform). IABP timing refers to inflation and deflation of the IAB in relation to the cardiac cycle. The cardiac cycle is monitored by continuous display of the arterial pressure waveform. As the balloon inflates at the onset of diastole, a sharp and deep V is observed at the dicrotic notch (Fig. 1). Balloon inflation causes augmentation of diastolic pressure and a second peak is observed. This peak is referred to as diastolic augmentation. Diastolic augmentation is ideally higher than the patient s systolic pressure except when reduced stroke volume causes a relative decrease in augmentation. Depending upon the patient s 26 Continuing Education in Anaesthesia, Critical Care & Pain j Volume 9 Number 1 2009

Fig 1 One complete cardiac cycle and the corresponding waveform of the IABP during inflation and deflation. Reproduced with permission from Datascopew. haemodynamic status, the balloon is programmed to assist every beat (1:1) or less often (1:2, 1:4, or 1:8). With haemodynamic improvement, the device can be weaned to less frequent cycling before complete removal. However, the device should never be left unused in situ to prevent thrombosis. Suboptimal timing of inflation and deflation of the balloon will result in haemodynamic instability (Fig. 2A D): Examples of this include: (i) Early inflation: inflation of the IAB before aortic valve closure (Fig. 2A). (ii) Late inflation: inflation of the IAB markedly after closure of the aortic valve (Fig. 2B). (iii) Early deflation: premature deflation of the IAB during the diastolic phase (Fig. 2C). (iv) Late deflation: deflation of the IAB after the onset of systole (Fig. 2D). Weaning from IABP should be considered when the inotropic requirements are minimal, thus allowing increased inotropic support if needed. Weaning is achieved gradually (over 6 12 h) reducing the ratio of augmented to non-augmented beats from 1:1 to 1:2 or less and/or decreasing the balloon volume. The balloon should never be turned off in situ except when the patient is anticoagulated because of the risk of thrombus formation on the balloon. Patient care should be carried out with three primary goals in mind: (i) evaluation in terms of haemodynamic status, systemic perfusion, and relief of cardiac symptoms; (ii) observation for early signs of complications including limb ischaemia, balloon malpositioning, thrombus formation, bleeding, and infection; (iii) ensuring proper functioning of IABP, including correct timing, consistent triggering, and troubleshooting of alarms. Continuing Education in Anaesthesia, Critical Care & Pain j Volume 9 Number 1 2009 27

Fig 2 (A) Waveform characteristics: inflation of IAB before dicrotic notch; diastolic augmentation encroaches onto systole, may be unable to distinguish. Physiological effects: potential premature closure of the aortic valve; potential increase in LVEDV and LVEDP or PCWP; increased LV wall stress or afterload; aortic regurgitation; increased MVO 2 demand. (B) Waveform characteristics: inflation of IAB after the dicrotic notch; absence of sharp V. Physiological effects: suboptimal coronary artery perfusion. (C) Waveform characteristics: deflation of IAB is seen as a sharp decrease after diastolic augmentation; suboptimal diastolic augmentation; assisted aortic end-diastolic pressure may be equal to or less than the unassisted aortic end-diastolic pressure; assisted systolic pressure may increase. Physiological effects: suboptimal coronary perfusion; potential for retrograde coronary and carotid blood flow; suboptimal afterload reduction; increased MVO 2 demand. (D) Waveform characteristics: assisted aortic end-diastolic pressure may be equal to the unassisted aortic end-diastolic pressure; rate of increase of assisted systole is prolonged; diastolic augmentation may appear widened. Physiological effects: afterload reduction is essentially absent; increased MVO 2 consumption because of the left ventricle ejecting against a greater resistance and a prolonged isovolumetric contraction phase; IAB may impede LV ejection and increase the afterload. Reproduced with permission from Datascopew. References 1. Kantrowitz A. Experimental augmentation of coronary flow by retardation of the arterial pressure pulse. Surgery 1953; 34: 678 87 2. Harken DE. The surgical treatment of acquired valvular disease. Circulation 1958; 18: 1 6 3. Moulopoulos SD, Topaz SR, Kolff WJ. Extracorporeal assistance to the circulation and intraaortic balloon pumping. Trans Am Soc Artif Intern Organs 1962; 8: 85 9 4. Bergman HE, Casarella WJ. Percutaneous intra-aortic balloon pumping: initial clinical experience. Ann Thorac Surg 1980; 29: 153 5 5. Miller RD. Miller s anaesthesia. In: Nyhan D, Johns RA eds. Anesthesia for Cardiac Surgery. Elsevier, 1991/2007 6. Walls JT, Boley TM, Curtis JJ, Silver D. Heparin induced thrombocytopenia in patients undergoing intra-aortic balloon pumping after open heart surgery. ASAIO J 1992; 38: M574 6 7. Ryan TJ, Antman EM, Brooks NH et al. 1999 update: ACC/AHA Guidelines for the Management of Patients with Acute Myocardial Infarction: Executive Summary and Recommendations: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee on Management of Acute Myocardial Infarction). Circulation 1999; 100: 1016 30 8. Arafa OE, Geiran OR, Svennevig JL. Transthoracic intra-aortic balloon pump in open heart operations: techniques and outcome. Scand Cardiovasc J 2001; 35: 40 4 9. Mercer D, Doris P, Salerno TA. Intra-aortic balloon counterpulsation in septic shock. Can J Surg 1981; 24: 643 5 10. Pinkney KA, Minich LL, Tani LY et al. Current results with intraaortic balloon pumping in infants and children. Ann Thorac Surg 2002; 73: 887 91 11. Velez CA, Kahn J. Compartment syndrome from balloon pump. Catheter Cardiovasc Interv 2000; 51: 217 9 Please see multiple choice questions 21 25 28 Continuing Education in Anaesthesia, Critical Care & Pain j Volume 9 Number 1 2009