ANNEX III: CONDOM TESTING SPECIFICATIONS

ANNEX III: MALE CONDOM INSPECTION, SAMPLING AND TESTING SPECIFICATIONS

SECTION VIII: ANNEX D: MALE CONDOM INSPECTION, SAMPLING AND TESTING SPECIFICATIONS PRE-SHIPMENT TESTING

SECTION VIII: ANNEX A- FEMALE CONDOM INSPECTION, SAMPLING AND TESTING SPECIFICATIONS PREQUALIFICATION TESTING

INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD. Male condoms Requirements and test methods for condoms made from synthetic materials

WHO/UNFPA Specification for Male Latex Condoms

Provläsningsexemplar / Preview INTERNATIONAL STANDARD. Rubber condoms - Part 1: Requirements

Chapter 5: Checking and Maintaining Ultrasound Equipment

?Cl-\-e~+ J:l 0 t q;its -'{

Quality Analysis of Male Latex Condoms Available in Private and Public Facilities in Lusaka, Zambia A Comparative Study

Kondomer av naturgummilatex för män Natural rubber latex male condoms

Part 1: General requirements

An ISO 9001:2008 Certified Company

The Aerosol Dispensers Directive. Labelling Requirements

LUBRICANT, WATER BASED, NON-IRRITANT, FOR USE DURING INTERCOURSE

Technique Guide. SynPOR Porous Polyethylene Implants. For craniofacial and orbital augmentation and reconstruction.

Tel. : / Fax :

Operational Guideline

LIQUID (DYE) PENETRANT INSPECTION (LPI)

Nephrostomy Tube Care

INSTRUCTIONS FOR USE AND CARE

TRAINING PROGRAM CUPID FEMALE CONDOM

The Method of Manufacturing Nonmetallic Test- Blocks on Different Sensitivity Classes

3000 Series Articulator System

See Instructions on Other Side. Guide. Sizing. 25 mm. Small. Medium. 29 mm. Intermediate. 32 mm. 36 mm. Large. X-Large. 41 mm

The Mechanics of the Dental Anesthetic Cartridge

Preventing Sexually Transmitted Infections Storytelling Activity

Figure 2. Pinion with original cutting tool marks. Polishing on right side is caused by misalignment. Figure 3. Severe gear profile wear.

Scope This policy applies to all personnel and departments that clean, prepare and/or sterilize items intended for patient care use.

Patient Self-Bandaging Leg and Individual Toe Application Guide

CORESTA RECOMMENDED METHOD N 22

Urinary Catheter Passport

5. LIQUID PENETRANT TESTING (PT) : LEVEL I ( 4 DAYS )

Mechanical Calibration

Manufacturer s information

Bureau of Laboratory Quality Standards

Advanced Catheterisation Trainer User Guide

Patient/Carer instructions for the administration of Subcutaneous Cytarabine

Instruction Manual Aerogen, Inc. Part No. AG-AL1010 Rev. C

University College Hospital at Westmoreland Street. Urinary sheaths. Urology Directorate

SKF bearings with Solid Oil the third lubrication choice

Technical Procedure for Physical Match Analysis

limbsandthings.com Advanced Catheterisation Trainer User Guide For more skills training products visit Limbs & Things Ltd.

Stülpa Instructions for dressing retention with tubular bandages

M102 New TOMOKO. Introduction/ Table of Contents 1 Purpose 2 Content 3 Precautions concerning handling 4 Usage 5-7 Storage method 8.

Acknowledgement goes to WHO for use of their hand washing technique

Products for a Fast and Reliable Fill

SYNPOR POROUS POLYETHYLENE IMPLANTS. For craniofacial and orbital augmentation and reconstruction

LIQUID PENETRANT EXAMINATION PROCEDURE

Education for Self Administration of Intravenous Therapy HOME IV THERAPY PICC. Portacath

Sterile hypodermic needles for single use Requirements and test methods

REVISION OF MONOGRAPH ON TABLETS. Tablets

CHEMISTRY CHANGES EVERYTHING

SOUTHERN AFRICAN INSTITUTE OF WELDING. Non Destructive Testing (NDT) Course Syllabus & Programme

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide

Policy x.xxx. Issued: Artificial Airways and Airway Care. ABC Home Medical Company Policy & Procedure Manual. A. Tracheostomy Tubes ( trach tubes)

About Your PleurX Drainage Catheter

6. Gear Measurement And Testing

The Omicron vaporizer is the world s first and only non-flame powered

Engineering Materials

Guidelines for the Qualification and Certification of Penetrant Testing Personnel

BS EN ISO & 2: 2004 Type 5 Protective Clothing

BONE GRAFTING. Product Catalog

Instructions for Use. Welcome!

UNICLIP DIRECTIONS FOR USE REF C226U - 0,8 MM / 1 MM FOR DENTAL USE ONLY. FISDR / F EN / 11 / updated 06/2016 1/7

Yasargil Permanent Aneurysm Clips

Wear sleeves. General

Skills workshop: Screening tests for syphilis

with' some material which has enough body' to act as a splint,' EVERY surgeon has his own pet method of dressing skin

Learning Objectives. We will look at: pumps infusion sets injection sites programming/testing

WHO SPECIFICATIONS AND EVALUATIONS FOR PUBLIC HEALTH PESTICIDES DELTAMETHRIN LONG-LASTING (COATED ONTO FILAMENTS) INSECTICIDAL NET

Compressor Nebulizer. Guidebook

The right impact on taste and texture CONDITIONING COATING BROCHURE

Stoma Care Policy and Procedures

V2 June Pump Mishaps

Dentistry Pre-capsulated dental amalgam

WHO SPECIFICATIONS AND EVALUATIONS FOR PUBLIC HEALTH PESTICIDES

NTS ULTRASONICS PTY LTD (N

Fibrlok II 2529 Universal Optical Fiber Splice

Adjustable Ab Board G

limbsandthings.com Advanced Female Catheterisation Trainer User Guide For more skills training products visit Limbs & Things Ltd.

Home Care for Your Nephrostomy Catheter - The James

WHO SPECIFICATIONS AND EVALUATIONS FOR PUBLIC HEALTH PESTICIDES DELTAMETHRIN LONG-LASTING (COATED ONTO FILAMENTS) INSECTICIDAL NET

CLICK SEAL SYSTEM. Thermo Scientific Samco Clicktainer Vial. Unique click seal system providing sample protection and user safety

Radiochromic film dosimetry in water phantoms

User s Guide 2004 Horizon Fitness Products Designed & Engineered in the U.S.A. Made in China

A patient guide to administration of subcutaneous immunoglobulin replacement therapy - using manual push technique

ProUltra ULTRASONIC SURGICAL TIPS

ISO 1567 INTERNATIONAL STANDARD. Dentistry Denture base polymers. Art dentaire Polymères pour base de prothèses dentaires. Third edition

Fluoxetine Hydrochloride 5.6 mg Oral Chewable Treats (Solid Suspension, 60 x 1 ml Chewable Treats)

MICAFUNGIN MIXING

How to Use ENBREL : Vial Adapter Method

The nomenclature describing the orthosis is basically the same as that suggested for the molded ankle-foot orthosis (2), but the anterior shank enclos

HOME GYM HG2011 ASSEMBLY INSTRUCTIONS

VistaCam ix Proof interchangeable head

Maintain your veins.

SKF SPEEDI-SLEEVE. and large diameter wear sleeves

Grey Bruce Health Network EVIDENCE-BASED CARE PROGRAM PATIENT EDUCATION BOOKLET TURP (TRANS URETHRAL RESECTION OF THE PROSTATE) PATHWAY

Transcription:

THIS IS UNFPA Male Latex Condom Testing Guidance UNICEF UNFPA WHO Joint Meeting September 2017 David Hill

2 Condom tests in ISO 4074:2015 The following tests are required in ISO 4074:2015:- Biocompatibility Microbial contamination Quantity of lubricant Dimensions (length, width, thickness) Bursting volume and pressure Stability and shelf life Freedom from holes Visible defects Package integrity. Packaging and labelling

3 Condom tests in the WHO/UNFPA Male Latex Condom Guidelines 2013 In addition, the UNFPA guidelines also recommend the following are measured:- Water-extractable proteins N-nitrosamines Note: the Male Latex Condom Guidelines are being revised.

4 Testing laboratories This presentation is aimed at testing that can be carried out by condom manufacturers. Assessment of biocompatibility, microbial contamination, and the analysis of N-nitrosamines and water-extractable proteins will probably have to be outsourced. Specialist testing laboratories, accredited to ISO 17025, will have the procedures, equipment, skills and experience to carry out most of these tests.

5 General points to note. Follow the specified sample sizes. Randomize the sampling. Never use scissors or sharp instruments to open the condom pack (except possibly for lubricant determination). Ensure that the calibration of instruments is up to date. Ensure that the operators are adequately trained.

6 Determination of total lubricant Two methods are specified in ISO 4074, both using a washing procedure. Propan-2-ol or a 5% solution of potassium laurate used. Points to note: Ensure the washing process does not remove any identification on the condom and foil. Note that propan-2-ol is flammable. Ensure that constant weight is achieved. The mass loss will include any powder.

7 Dimensions - length Length is measured by hanging the condom on a calibrated rod. Points to note: Lubricated condoms are sticky: it is better to kill the tack by powdering, with or without removal of the lubricant. Ensure that the bead of the condom is at eye-level to avoid parallax errors. Measure the length at several points and record the minimum length to the nearest millimetre. Automated measurement equipment is available.

8 Dimensions - width Width is measured by hanging the condom over the edge of a ruler. Points to note: Lubricated condoms are sticky: it is better to kill the tack by powdering, with or without removal of the lubricant. Ensure that the condom hangs freely with the axis of the condom perpendicular to the ruler. Measure the width to the nearest 0.5 millimetre. Automated measurement equipment is available.

9 Dimensions - thickness Can use a micrometer gauge or the mass of a test piece from the condom. Points to note: Lubricant is removed before the measurement. Measure at three points (near the closed end, the open end and the centre of the condom). With textured condoms measured with a micrometer, measure a smooth region as close as possible to the specified points, or use the mass method. Condom cut open to measure as single-wall thickness.

10 Dimensions - thickness Micrometer foot pressure to be 22 ± 5 kpa. Ensure the shaft of the micrometer is clean. Ensure the foot is parallel to the platen.

11 Bursting volume and pressure Bursting volume and pressure measure important properties of a condom. Several possible causes of variability. Loading of the condom. Incorrect inflation length. Condom slippage during inflation. Correct calibration of test equipment Note that any condom showing an obvious leak must be removed and replaced.

12 Loading the condom The condom may be unrolled beforehand, or unrolled over the supporting rod/mandrel. The length to be inflated is 150 ± 3 mm., so Do not stretch the condom over the rod/mandrel this will shorten the tested length, and give lower burst volumes and higher pressures. Do not hang the condom loosely over the rod/mandrel - this will increase the tested length and give higher volumes, lower pressures.

13 Check the equipment Described in Annex O of ISO 4074. Check at appropriate intervals, and include: Confirmation of the correct inflation length. Confirmation of no slippage of the condom. Confirmation of no leaks in the system. Confirmation of the calibration of the pressure gauge. Confirmation of the calibration of airflow rate. Confirmation of no sharp edges on the clamping mechanism.

14 Freedom from holes Like bursting volume and pressure, a critical property of condoms. Can be assessed by filling the condom with water and rolling over absorbent paper ( hang and roll ). Can also be assessed by looking for electrical leakage through the condom film. Includes an assessment of visibly open seals. The ASTM D3492 method ( hang and squeeze ) may also be used.

15 Freedom from holes - water leak test Inspect the condom packages for visibly open seals. Unroll the condoms and check for visible defects. Fill the condom with 300 cm 3 of water (more water for extra large condoms). Observe for leaks. Remove the condoms, carefully twist the open end to seal and roll on absorbent paper.

16 Water leak test Points to note. Wipe the condom before rolling to remove any water that may have splashed. Press the hand onto the condom so that the hand is 25 to 35 mm. above the paper. Roll through at least two full revolutions. Repeat, applying pressure to other parts of the condom. Hold the condom vertically, and press and rotate the closed end onto the paper.

17 Water leak test things to avoid Do not Roll the condom only a short distance. Ensure that the whole condom surface contacts the paper. Compress the condom too much. The hand must be 25 to 35 mm. above the paper. Stretch the condom. Use absorbent paper held together with adhesive tape. Squeeze the condom too much if using the ASTM test see the standard for illustrations.

18 Water leak testing No! Yes!

19 Freedom from holes electrical test Inspect the condom packages for visibly open seals. Unroll the condoms and check for visible defects. Load onto the test machine and fill with 300 cm 3 of electrolyte (more electrolyte for extra large condoms). Apply a voltage of 10V dc and lower the condom into a bath of electrolyte. Check for current flow.

20 Water leak test Points to note. Any condom failing the electrical test must be confirmed as holed using the hang and roll test. Ensure that the electrolyte level in the bath covers the condoms up to 25 mm. from the open end. Ensure the equipment is properly calibrated. Verify the equipment frequently, using condoms holed before testing.

21 Visible defects Include: Adhesion of the internal surface of the condom. Permanent creases or pleats. Broken or missing beads. Thin areas of the film. Delamination. Blisters or bubbles. Large (>1 mm) embedded particles or coagulum. Cracking of the film. The WHO/UNFPA guidelines contain a list and description of critical visible defects.

22 Package seal integrity Improved test being developed. Current test submerges the condom packs under water and holds them at a reduced pressure of 20 ± 5 kpa for 60 ± 5 seconds. The packs are opened after the test to confirm the absence of leaks. A low foaming surfactant may be added to the water to improve the wetting of the packs. A dye may be added to the water to improve the detection of any water in the pack.

23 Package seal integrity points to note. Ensure the correct level of vacuum is used. The packs must be submerged at least 25 mm below the surface. Start observing the packs for escaping bubbles as soon as the vacuum is applied. Open the packs after the test to confirm leakage or no leakage. Do not overload the test chamber. Vacuums are potentially dangerous use suitable protective equipment.

quality value choice www.unfpaprocurement.org