ANALYSIS OF BENZODIAZEPINES IN WHOLE BLOOD, PLASMA AND URINE BY RADIOIMMUNOASSAY

Similar documents
Validation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens

Determination of Benzodiazepines in Urine by CE-MS/MS

A Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

Using Liquid Chromatography Tandem Mass Spectrometry Urine Drug Testing to Identify Licit and Illicit Drug-Use in a Community-based Patient Population

Offline and online sample extraction for the quantification of benzodiazepines in human plasma or serum for clinical research

Drug Profiles of Apprehended Drivers in Victoria

NOT FOR HUMAN CONSUMPTION

ORAL FLUID AS A CHEMICAL TEST FOR THE DRE PROGRAM : HISTORY, THE FUTURE, AND PRACTICAL CONSIDERATIONS

BENZODIAZEPINE FINDINGS IN BLOOD AND URINE BY GAS CHROMATOGRAPHY AND IMMUNOASSAY

HUMAN LEPTIN RIA KIT 250 TUBES (Cat. # HL-81K)

INTERPATH LABORATORY, INC. TEST UPDATES


Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

RAT LEPTIN RIA KIT 250 TUBES (Cat. # RL-83K)

1/27/ New Release, Quest Diagnostics Nichols Institute, Valencia

Clonazepam diazepam dose equivalent

Improving Benzodiazepine Immunoassay Sensitivity by Rapid Glucuronide Hydrolysis Technology

Overview. Introduction. Experimental. Cliquid Software for Routine LC/MS/MS Analysis

Cozart DDS Drug Detection System On-site Saliva Drug Testing

Automated Targeted screening of Benzodiazepines in urine using LDTD-MS/MS at 400 samples per hour rate

EliKine Free Thyroxine (ft4) ELISA Kit

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

Mectest Opiate Radioimmunoassay (MOR)

References Required document for Laboratory Accreditation by the College of American Pathologists.

Clonazepam temazepam clonazepam temazepam compared temazepam clonazepam klonopin temazepam Clonazepam vs Temazepam Clonazepam Temazepam Temazepam

DKK-1 ELISA, Cat.No. BI For the quantitative determination of DKK-1 in human serum

Buprenorphine DOUBLE ANTIBODY. For Veterinary Use Only

Clonazepam to lorazepam conversion

a very short half-life, but its main metabolite, which

1. Carrying Personal Medications into Thailand A. Medications containing narcotic drugs of category 2

Agilent Clinical LC/MS: Review of Established LC-MS QQQ Clinical Research Applications. dr. Jan Srbek, HPST. Page 1

Glucagon Kit. Procedure for the Radioimmunoassay of Human Plasma Glucagon TABLE OF CONTENTS. For In Vitro Diagnostic Use I. INTRODUCTION...

Test Definition: PCDSO Pain Clinic Drug Screen, Urine

diazepam lorazepam converting converting Diazepam Valium Lorazepam Ativan lorazepam diazepam conversion Valium Lorazepam Ativan conversion lorazepam

Rat cholesterol ELISA Kit

Instructions for Use. IA-2 Antibody RIA. for the Quantitative Determination of Antibodies against Protein-Tyrosine-Phosphatase (IA2) in Serum

Test Definition: PDSOX Pain Clinic Drug Screen, Chain of Custody, Urine

Librium to ativan conversion

DrugSmartCup & DrugSmartDip Accuracy Report as produced by Ameditech, Inc.

Fast and easy separation of 23 drugs of abuse. including high, stable resolution of isobaric opioids from human urine by UHPLC-MS/MS

Human TSH ELISA Kit. User Manual

Prevalence of Drugs Used in Cases of Alleged Sexual Assault

VGKC-Autoantibody Assay RIA

Standard Operating Procedure

DRINK AND DRUG-RELATED DRIVING

Identification and Quantification of 22 Benzodiazepines in Postmortem Fluids and Tissues using UPLC/MS/MS

Cortisol Assays. The Good, The Bad and The Indifferent. David Ducroq. Cardiff and Vale ulhb WEQAS. Unit 6, Parc Tŷ Glas. Llanishen.

ab Human Cathepsin D SimpleStep ELISA Kit

PROCEDURE REF NO SABP/EXECUTIVE BOARD/0017

CoQ10(Coenzyme Q10) ELISA Kit

Rapid Hydrolysis of Benzodiazepines in Urine. Alicia Zook 1 and Crystal Xander B.S. 2. Cedar Crest College, Allentown, PA 1

Validation of a Benzodiazepine and Z-Drug Method Using an Agilent 6430 LC/MS/MS

September HCMC Toxicology Transition: Additional information and Frequently Asked Questions

DOF. Scheme Description. Drugs in Oral Fluid Scheme

ab HIV-1 p24 SimpleStep ELISA Kit

What Your Drug Test Really Means. Krista Beiermann, RN, OHS Occupational Health Services, Columbus Hospital

Rat Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA kit

Ativan to klonopin conversion

Ms. Holly Golightly (Audrey Hepburn), Breakfast at Tiffany s, 1961

New Test Announcements

Benzodiazepines. Benzodiazepines

Simultaneous Determination of Prescription and Designer Benzodiazepines in Urine and Blood by SPE and LC-MS/MS

AssayMax Human Aldose Reductase ELISA Kit

Key words: Collagen synthesis - N-Terminal peptide of type III procollagen - Tumor marker - Liver cancer - Liver cirrhosis

LC/MS/MS Analysis of Gabapentin, and Opiates. Applications. Presentation Outline. Polarity, MW and Volatility LC/MS Triple Quad Components

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer

Standard Operating Procedure Screening Test for Cocaine Metabolite in Urine

STANDARD OPERATING PROCEDURE TITLE: TOTAL IGE EIA (HYTEC 288)

LDL (Human) ELISA Kit

HbA1c (Human) ELISA Kit

Lispro Insulin. 250 Tubes. Cat. # LPI-16K

Supplementary Information. Sonorensin: A new bacteriocin with potential of an anti-biofilm agent and a food

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Stability of thyroxine and triiodothyronine in

Note: During 30 minute incubation; proceed thru appropriate sections below (e.g. sections II, III and V).

GLUCAGON-LIKE PEPTIDE-1 (ACTIVE) RIA KIT 125 TUBES (Cat. # GLP1A-35HK)

Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System

Preparation and Validation of Double Antibody Radioimmunoassay for Thyroglobulin (Tg) using Balb/C Mice as Host Animals

Reduction in the Use of Night Sedation to Reduce the Risk of Falls

Canine Thyroid Stimulating Hormone, TSH ELISA Kit

Falls most commonly seen in RACFs are due to tripping, slipping and stumbling (21.6%). Falling down stairs is relatively uncommon in

European Guidelines for Workplace Drug Testing in UrineandOral fluid

Human IL-2. Pre-Coated ELISA Kit

Preparation of prostate specific antigen standards for immunoradiometric assay

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit

EXOTESTTM. ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids

Scottish Blues- Benzodiazepines and Drug Related Deaths in Scotland

Urine Drug Testing. Methadone/Buprenorphine 101 Workshop. Ron Joe, MD, DABAM December 10, 2016

BETA-GLUCURONIDASE PRODUCT LINE INNOVATION THROUGH CHEMISTRY. FORENSICS

The Drug Testing Process. Employer or Practice

See external label 2 C-8 C Σ=96 tests Cat # 6101Z. Cortisol. Cat # 6101Z

Urea Nitrogen (BUN) detection Kit

Exendin-4 (Exenatide) ELISA Kit

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

Analyze Drugs of Abuse with Agilent J&W Ultimate Plus Tubing in an Inert Flow Path

Why does ativan not show up in urine tests

Tetracycline Plate Kit (for Honey) PN 52254B

A guide to reducing or stopping mental health medication. Notes for prescribers

Strategies in Managing Opioid and Benzodiazepine Co-Prescribing

Transcription:

ANALYSIS OF BENZODIAZEPINES IN WHOLE BLOOD, PLASMA AND URINE BY 125-1 RADIOIMMUNOASSAY C. W.Hand. Mary Rutterford, R. F. Smith and R. A. Moore DPC-European Research Institute, Witney, Oxfordshire 0X8 6AN Running title: Benzodiazepines RIA Correspondence to: Dr. R.A. Moore DPC European Research Institute 31 Station Lane Witney Oxfordshire 0X8 6AN SUMMARY Two benzodiazepine RIA kits (DPC Double Antibody) formulated for use with either serum or urine samples respectively were used to investigate benzodiazepine concentrations in whole blood, plasma and urine. Both kits use oxazepam calibrators ranging from 10 to 1000 ng/ml in an appropriate matrix and have broad specificity for the drug class. Fifty-three paired samples of haemolysed whole blood and plasma were analysed with the serum kit: 26 pairs were from drug-free volunteers and all blood and plasma samples gave undetectable results. Twenty-seven paired samples were from patients receiving benzodiazepines and of these 26 plasma and 25 blood samples were positive. The average plasma/blood ratios for samples from patients taking a single of benzodiazepine were: for diazepam (5, 10 or 20 mg) 2.1, n=7; for temazepam (10, 20, 40 or 60 mg) 2.3, n=12; and for nitrazepam (5, 10 and 15 mg) 2.0, n=3. The urine kit was compared to another commercially available assay (Roche Abuscreen) using 41 positive (patient urines) and 39 negative (volunteer) samples. There was 100% agreement between the two methods relative to a 100 ng/ml oxazepam cut-off. The DPC kit was also used to detect benzodiazepines in urine after a single, low (1 mg) therapeutic dose of either flunitrazepam (n=5) and lorazepam (n=l). After both drugs, benzodiazepines were detectable for at least 12 hours after the dose. Key Words: benzodiazepines, radioimmunoassay, blood INTRODUCTION Two RIA kits formulated for the detection of benzodiazepines in urine or serum were used to investigate concentrations of benzodiazepines in urine, plasma and whole blood from patients taking benzodiazepines or from drug-free volunteers. The serum kit was used to compare the concentration of benzodiazepines in paired samples of whole blood and plasma in patients taking benzodiazepines as part of multi-drug therapy for the relief of chronic pain. Control samples of paired whole blood and plasma were analysed from drug-free volunteers. 893

The urine kit was compared to another commercially available RIA for benzodiazepines and was also used to examine the concentrations of benzodiazepines detectable in urine after low doses of lorazepam and flunitrazepam. METHODS Radioimmunoassay Both RIA kits were used as supplied by the manufacturer (EURO/DPC Ltd, Witney UK). The assays follow basically similar procedures with the exception of sample size. The RIA for urine uses samples of 50 ul; the serum (plasma & whole blood) assay uses a 25 ul sample. Standard, control or sample were mixed with 100 ul ^251-benzodiazepine and 100 ul antiserum solution. After a 60 minute incubation at room temperature bound and free radioactivity were separated by the addition of 1 ml of cold precipitating solution (donkey anti-sheep/goat in PEG/saline) and centrifugation. Tubes were decanted using a foam rack and the pellets counted for 1 minute in a multi-head gamma counter (DPC Gambyt). Sample concentrations were calculated automatically from a logit-log transformation of the standard data. Specificity Cross-reactivity of both RIA kits was assessed with 18 different benzodiazepines at concentrations of 50, 100, 500, 1000 and 10000 ng/ml in drug-free urine or serum. Percentage cross-reactivity was calculated on a weight for weight basis relative to the kits oxazepam calibrators. Fifty-five unrelated drugs were also tested for cross-reaction at a concentration of 100 ug/ml Method Comparison: Urine RIA The DPC Urine Benzodiazepine RIA was compared to a qualitative RIA for benzodiazepines (Roche Abuscreen). The qualitative procedure uses 100 ng/ml oxazepam as a positive/ negative cut-off. Eighty urine samples were analysed for the presence of benzodiazepines by the both RIA procedures using a 100 ng/ml oxazepam cut-off with both the Roche and DPC kits. Forty-one samples were from patients known to be taking benzodiazepines; 39 samples were from drug-free volunteers. Low Dose Benzodiazepines in Urine Urine samples were collected from patients taking a single dose of flunitrazepam (1 mg; N=5) or lorazepam (1 mg; N=l). Urine was collected for 24 hours after the dose and samples were analysed with the DPC Urine Benzodiazepine RIA using the quantitative method. Paired Plasma and Haemolysed Whole Blood Twenty-seven blood samples were collected into lithium heparin tubes from 894

patients receiving various benzodiazepines (on a regular basis) as part of multi-drug therapy for the relief of chronic pain. Blood samples were also collected from 26 patients or healthy volunteers who declared they were not taking benzodiazepines. Samples were split into two: one aliquot was centrifuged and plasma collected; the other was retained as whole blood and subjected to three cycles of freeze/thawing to cause haemolysis of red cells. Plasma and haemolysed whole blood samples were then analysed using the DPC Double Antibody Serum Benzodiazepines RIA. RESULTS Specificity The RIA has a wide class-specific cross-reactivity for the benzodiazepines (Tables 1 & 2). Fifty-five other non-related compounds were not detectable even at 100 ug/ml. TABLE 1 Specificity: Urine Benzodiazepine RIA TABLE 2 Specificity: Serum Benzodiazepine RIA Method Comparison: Urine RIA There was 100% agreement between the two methods in classifying the 80 urine samples as positive or negative relative to the 100 ng/ml oxazepam cut-off. Both kits correctly identified the 39 drug-free volunteers as negative and the 41 samples from patients taking benzodiazepines as positive. Low Dose Benzodiazepines Samples gave detectable amounts of benzodiazepines (ie above the lowest calibrator: greater than 10 ng/ml oxazepam equivalents) for up to 12 hours in each of the six patients after the administration of either 1 mg lorazepam (N=l) or 1 mg flunitrazepam (N=5). Patient 1 taking flunitrazepam had a concentration of 34 ng/ml oxazepam equivalents at 14 hours post dose; similar figures for the other fours patients taking flunitrazepam were: 56 ng/ml at 13 hours; 13 ng/ml at 14 hours; 31 ng/ml at 14.5 hours and 58 ng/ml at 12 hours. The single patient taking lorazepam had a urinary benzodiazepine concentration of 25 ng/ml at 13 hours post dose. Paired Whole Blood & Plasma All 26 paired blood and plasma samples from volunteers and patients not taking benzodiazepines gave undetectable levels of benzodiazepines. The RIA counts were similar for these samples to the counts of the zero calibrator which demonstrates a lack of matrix effect on the assay. Benzodiazepines were detectable in 26 of the plasma samples and 25 whole blood samples from the 27 patients taking benzodiazepines (Table 3). 895

TART.F. 3 Paired Plasma & Whole Blood Patient Benzo Dose Time Plasma Blood Ratio ID -diazepines (mg) since (ng/ml)(ng/ml) P/B last dose (hr:min) 1 Diazepam 10 59:10 430 136 3.2 2 Temazepam 10 14:25 1168 500 2.3 3 Temazepam 20 13:55 1956 988 2.0 5 Temazepam 10 13:20 1048 836 1.3 6 Temazepam 10 19:00 289 168 1.7 7 Temazepam 20 11:00 96 71 1.4 *8 Tem/Diaz 10 >100 42 14 3.0 10 Nitrazepam 15 11:30 3578 1491 2.4 13 Temazepam 20 18:00 2028 1169 1.7 14 Temazepam 60 13:45 6544 2538 2.6 18 Triazolam 0.06 18:00 <10 <10-20 Nitrazepam 10 11:45 120 49 2.4 24 Nitrazepam 5 >100 13 <10-25 Diazepam 10 >100 56 28 2.0 29 Temazepam 10 18:30 and Diazepam 2 4:30 1306 682 *31 Tem/Diaz 10 >100 19 15 1.3 34 Temazepam 60 14:30 47500 10360 4.6 37 Temazepam 40 18:15 2058 720 2.9 38 Diazepam 20 13:15 1100 527 2.1 39 Nitrazepam 10 13:40 112 98 1.4 40 Diazepam 10 13:50 812 450 1.8 41 Temazepam 10 12:30 869 330 2.6 47 Diazepam 10 13:15 1026 672 1.5 50 Temazepam 20 12:10 2154 804 2.7 54 Temazepam 20 9:45 2552 1242 2.1 56 Diazepam 5 12:10 299 149 2.0 75 Diazepam 10 12:00 1010 560 1.8 * patients 8 and 31 were in patients prescribed benzodiazepines for use "as required" but none had been taken for greater than 100 hours. The average plasma/blood ratios for samples from patients taking a single of benzodiazepine were: for diazepam (5, 10 or 20 mg) 2.1, n=7; for temazepam (10, 20, 40 or 60 mg) 2.3, n=12; and for nitrazepam (5, 10 or 15 mg) 2.0, n=3. Table 3 gives actual values for the 27 paired samples from patients taking benzodiazepines. DISCUSSION Both kits (serum and urine) were broadly cross-reactive within the benzodiazepine class and were able to satisfactorily detect the presence of various benzodiazepines in urine, plasma or haemolysed whole blood. 896

The ratio of benzodiazepine concentrations in plasma to whole blood was generally of the order of 2:1. One paired sample of blood and plasma and one separate blood sample in the paired whole blood/plasma study had undetectable levels of benzodiazepines. The paired sample was from a patient who had taken only 0.06 mg triazolam (1/2 tablet) 18 hours previously and this together with the low cross-reactivity of the antibody to triazolam gave an undetectable result. The other negative whole blood had a paired plasma sample containing 13 ng/ml oxazepam equivalents and was from a patient who had taken 5 mg nitrazepam more than 100 hours before the sample was taken. The urine kit was able to classify 80 samples correctly as positive or negative for benzodiazepines and was also able to detect the presence of benzodaizepines in urine for up to 12 hours after a low therapeutic dose of lorazepam or flunitrazepam. 897