Calibration of two 90 Sr+ 90 Y dermatological applicators

Similar documents
Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source

COMPARING THE RESPONSES OF TLD 100, TLD 600, TLD 700 AND TLD 400 IN MIXED NEUTRON-GAMMA FIELDS

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration

CHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose

In vivo dosimetry with thermoluminescent. dosimeters in external photon beam. radiotherapy

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD

Evaluation of glasses containing cadmium for high dose dosimetry by the thermoluminescence technique

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

"Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 : Mn DOSIMETER FOR γ- IRRADIATION HIGH-DOSE DOSIMETRY *

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Mammography calibration qualities establishment in a Mo- Mo clinical system

SENSORIAL EVALUATION OF BRAZILLIAN HONEY

THERMOLUMINESCENT (TL) DOSIMETRY OF SLOW-NEUTRON FIELDS AT RADIOTHERAPY DOSE LEVEL

Sulfur status in long distance runners

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.

Good manufacturing practices (GMP) utilized on human blood irradiation process

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Implementation of personnel dosimetry systems using thermoluminescence dosimetry

BRACHYTHERAPY WITH 125-IODINE SOURCES: TRANSPORT AND RADIATION PROTECTION

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH A COMBINATION OF TLD ALBEDO AND SULPHUR ACTIVATION TECHNIQUES FOR FAST NEUTRON PERSONNEL DOSIMETRY

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Study of neutron-dna interaction at the IPEN BNCT Research facility ABSTRACT

THERMOLUMINESCENT DOSIMETRIC SYSTEM USED FOR CARACTERIZATION OF THE UNDERGROUND ENVIRONMENT

Dosimetric Characteristics of LiF:Mg,Cu,Na,Si Phosphor

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS

Male Breast Cancer-Dosimetry Evaluation in Teletherapy

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

Radiologic Units: What You Need to Know

PROSTATE BRACHYTHERAPY DOSIMETRY USING VOXEL PHANTOMS

Progress Report on Radiation Dosimetry at NPL

Applications in brachytherapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

The determination of timer error and its role in the administration of specified doses

Radiation Protection Services in R.Macedonia

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography

Luminescent materials for dosimetric applications

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

CALIBRATION OF A TLD ALBEDO INDIVIDUAL NEUTRON MONITOR

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Energy dependent response of Al 2 O 3 and its potential application in personal monitoring.

MAY NOT BE SAME AS THE PUBLISHED VERSION

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

The need for standardization of dosimetry in experimental radiation biology

The clinical use of OSLD

Infrastructure of the Brazilian Government in Radioprotection Education

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations

High-Dose Standardization Study For γ-rays radiation processing at NIM

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Development of radiation monitoring services for radiation workers in Saudi Arabia

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Process Control Methods in Radiation Technologies

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

METALS IN SOILS ADJACENT TO AVENUES OF HIGHLY DENSE TRAFFIC OF SÃO PAULO CITY, BRAZIL

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Progress in Reactor and Accelerator Based BNCT at Kyoto University Research Reactor Institute

Dosimetric Consideration in Diagnostic Radiology

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY

DOSE ASSESSMENT WITH PASSIVE PERSONAL DOSIMETERS EXPOSED TO X-RAY GENERATOR USING THE 241 Am CALIBRATION CURVE

First National Intercomparison of Individual Dosimetry of Photon Radiation in Ukraine

Radiation Dosimetry at the BNL High Flux Beam Reactor and Medical Research Reactor

EVALUATION OF THE ELUATE QUALITY OF 99 MO- 99M TC GENERATORS IN RECIFE, BRAZIL

EURADOS: Overview of actions and initiatives in radiation protection in medicine

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

IAEA SUBPROGRAMME ON DOSIMETRY AND MEDICAL RADIATION PHYSICS (DMRP) REPORT ON ACTIVITIES,

Protection of the contralateral breast during radiation therapy for breast cancer

9.7 Dosimetry Study on A p-n Junction Semiconductor Detector

Investigation of heavy trace elements in neoplastic and non-neoplastic human thyroid tissue: A study by proton induced X-ray emissions

Metrology Laboratory of Ionizing Radiation

Radiation Protection in BNCT Patients

Assessment of dosimetric parameters in mammography employing a homemade ionization chamber

STUDIES ON DOSIMETRIC TESTS APPLYING SOURCE IRRADIATING FORCE OF CS-137 FOR USING IN CHAMBERS FOR CALIBRATION AND TLD TYPE DOSIMETERS.

Strengthening Activity Measurement Quality in Radiation Protection from Metrological Science to Reliable End-user Application

Dosimetry Standards and Dissemination Systems for Radiation Processing in China

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Evaluation on Physical Properties of Irradiated Collards (Brassica oleracea L var.acephala)

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

DOSIMETRIC STUDY OF I-125 SEEDS USED IN PROSTATE BRACHYTHERAPY

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

A COMPARISON BETWEEN GEANT4 AND MCNPX ON THE DOSIMETRY OF THE 192IR MICROSELECTRON V2 HDR BRACHYTHERAPY

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ

Epithermal neutron beams from the 7 Li(p,n) reaction near the threshold for neutron capture therapy

Transcription:

Calibration of two 90 Sr+ 90 Y dermatological applicators Patrícia L. Antonio *, Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares, IPEN-CNEN/SP Av. Prof. Lineu Prestes 2242, 05508-000, São Paulo, Brazil Abstract. The 90 Sr+ 90 Y applicators need to be periodically calibrated, but in Brazil the service it not offered yet. The recommended method for the calibration of this kind of applicators is the use of extrapolation chambers. An alternative method for the calibration of clinical applicators is the use of thermoluminescent dosimeters. A dosimetric method of these applicators was already developed at Instituto de Pesquisas Energéticas e Nucleares (IPEN) and several types of thermoluminescent dosimeters were studied in previous works. The aim of this work was the application of this method to calibrate two dermatological applicators. Thin CaSO 4 :Dy pellets, with and without 10% of graphite were utilized. The reproducibility of these pellets was studied, and calibration curves were obtained using a standard applicator calibrated at the National Institute of Standards and Technology (NIST), USA. Both applicators showed similar results. The TL materials tested showed usefulness for dosimetry and calibration of this kind of applicators. KEYWORDS: Clinical applicators, beta radiation, beta dosimetry, thermoluminescent dosimeters 90 Sr+ 90 Y source, 1. Introduction Over the last five decades, 90 Sr+ 90 Y beta radiation sources have been commonly used in brachytherapy procedures, in the treatment of superficial injuries of eyes and skin. This kind of therapy can be used for post-operative treatment of pterygia, to reduce the rate of relapse, and to reduce the chances of formation of keloids, for example. These sources were developed by Friedell and collaborators in 1950 [1], and the procedure diffused quickly for the whole world. Approximately 10 different manufactures produced some hundred of sources, in the 1990 s decade [2]. 90 Sr+ 90 Y sources are very utilized due to their long half-life (28.8 years). The special sources used in brachytherapy are the clinical applicators. Since some years 90 Sr+ 90 Y dermatologic and ophthalmic applicators are not commercialized anymore, but many of them are still very utilized in Brazilian radiotherapy clinics. These applicators may be plane or concave; generally, the dermatologic applicators are plane and the ophthalmic applicators are concave. The 90 Sr+ 90 Y sources were delivered with usually calibration certificates of the manufacturers, but some of them were recalibrated by standard laboratories. The sources have not been periodically calibrated, because no such services are available in the Brazilian calibration laboratories yet. Moreover, it is difficult to remove these applicators from the radiotherapy clinics to send them for calibration, because this fact it would interrupt the treatments in progress. Recommendations as of the International Atomic Energy Agency [3] and De Almeida et al. [4] showed the need for periodic measurements and calibrations in secondary standard laboratories, hospitals and clinics that use these kinds of sources. The recommended method for calibration of clinical applicators in laboratory involves the use of special ionization chambers, called extrapolation chambers. These instruments allow variation of their sensitive volumes. These ionization chambers are metrological instruments that may be utilized for the detection of beta radiation and low energy X-rays. They are very useful for the determination of superficial absorbed dose rates of beta radiation sources, by the extrapolation technique. At the Calibration Laboratory of Instituto de Pesquisas Energéticas e Nucleares (IPEN), extrapolation chambers were developed for the calibration of dermatologic and ophthalmic applicators [5,6,7]. These * Presenting author, E-mail: patrilan@ipen.br 1

chambers are used only in laboratory, because they are not appropriate for field measurements, because they are not portable. The dosimetry of the clinical applicators can also be made using thermoluminescent dosimeters (TLD). In 1990, Reft et al. demonstrated already the ability to calibrate ophthalmic applicators of 90 Sr+ 90 Y, flat or curved, with the use of the phenomenon of thermoluminescence [8]. The best response for beta radiation was obtained for thin CaF 2 :Mn and CaSO 4 :Dy pellets. This method of calibration was also developed at IPEN in previous studies, using several types of thermoluminescent dosimeters (TLDs) [9]. Due to the low penetration power in the matter by beta radiation, the thin dosimeters are more appropriate for beta dosimetry [9]. For the calibration of a 90 Sr+ 90 Y concave applicator, studies by Oliveira and Caldas showed the usefulness of thin CaSO 4 :Dy pellets [10]. The Calibration Laboratory of IPEN has two 90 Sr+ 90 Y dermatological applicators (named A and B) of different activities. The aim of this work was the calibration of these applicators using thermoluminescent dosimeters against a dermatological applicator calibrated by National Institute of Standards and Technology (NIST), as reference. 2. Materials and Methods The calibration of applicators was realized using the method of thermoluminescent dosimetry. Initially, twenty thin CaSO 4 :Dy pellets were utilized: ten pellets with and ten pellets without 10% of graphite. Both the thin CaSO 4 :Dy pellets present dimensions of 6.0 mm in diameter and 0.2 in thickness. These pellets were produced by the Dosimetric Materials Laboratory of IPEN. A reproducibility study of TL response was performed and the result allowed the selection of the pellets in groups, with similar characteristics in relation to their TL sensitivity. One group of five CaSO 4 :Dy pellets and one group of five CaSO 4 :Dy + 10% of graphite pellets were chosen for the calibration curve using the NIST standard applicator and for the calibration of both dermatological applicators A and B. In this work, a 90 Sr+ 90 Y dermatological applicator calibrated by the primary standard laboratory of NIST, 90 Sr+ 90 Y Beta-Particle Source, Atlantic Research Corporation, Model B-1 S/N 233, was utilized to calibrate two 90 Sr+ 90 Y dermatological applicators: A, from Amersham, with nominal activity of 1.49 GBq and nominal average water absorbed-dose rate of 0.40 Gy/s (11/08/1968) and B, from Amersham, with unknown activity. The NIST calibration certificate presents a combined uncertainty, D W, of 6.0%. All three applicators were used in the same calibration conditions. Irradiations of the TL dosimeters using the NIST, A and B applicators were performed at a null sourcedetector distance. The applicators were fixed vertically by a support, inside of an acrylic box, and the pellets were placed on the source. The TL measurements were taken using a Harshaw Nuclear System, model 2000A/B,with a linear heating rate of 10ºC.s -1, and a constant flux of N 2 The light emission was integrated in the temperature interval of 50ºC to 200ºC. The thermal treatments for re-utilization of the TL pellets were made at 300ºC during 1 hour. 3. Results 3.1. Reproducibility of TL Response Initially, the reproducibility of the TL response of the samples was checked using ten pellets of each kind. The reproducibility was obtained after five series of irradiations measurements and thermal treatments, using 90 Sr+ 90 Y source, 1850 MBq (1981), at 11 cm, dose of 1 Gy, without the use of the flattening filter. The maximum percentual deviation obtained was 4.6% for the thin CaSO 4 :Dy + 10% of graphite pellets, and 4.3% for the thin CaSO 4 :Dy pellets. The associated uncertainties in this case were equal to 6.12% for CaSO 4 :Dy pellets and to 7.11% for CaSO 4 :Dy + 10% of graphite pellets. 2

3.2. Calibration Curves (NIST Applicator) The TL response as a function of absorbed dose was measured for six absorbed doses, in the interval of 5 Gy to 100 Gy. The measurement uncertainty obtained for the thin CaSO 4 :Dy pellets was equal to 11.6%, and for the thin CaSO 4 :Dy + 10% of graphite it was equal to 13.4%. For the calibration of the A and B applicators, three pellets of each kind were chosen, with similar TL characteristics. The calibration curves for the TL response, in function of the absorbed dose, are shown in Figures 1 and 2. These calibration curves show the usefulness of both kinds of TL dosimeters, thin CaSO 4 :Dy pellets and CaSO 4 :Dy + 10% of graphite, for calibration of dermatological applicators. 10 3 TL Response (a.u.) 10 2 10 1 10 0 10 0 10 1 10 2 10 3 Absorbed Dose (Gy) Figure 1: Calibration curve of the thin CaSO 4 :Dy pellets, for beta radiation, using the reference 90 Sr+ 90 Y NIST applicator. 10 2 TL Response (a.u.) 10 1 10 0 10-1 10 0 10 1 10 2 10 3 Absorbed Dose (Gy) Figure 2: Calibration curve of the thin CaSO 4 :Dy + 10% of graphite pellets, for beta radiation, using the reference 90 Sr+ 90 Y NIST applicator. 3

The calibration curves show a linear behaviour until 15 Gy. After this dose, the curves present supralinear characteristics. The fitting of the data was obtained using the Origin 7.5 software. 3.3. Calibration of Dermatological Applicators 3.3.1. Applicator A and B In the case of applicator A, all TL samples were irradiated four times, during 5, 10, 15 and 20 min. After each irradiation, the samples were evaluated and then thermally treated. The absorbed dose, relative to each time of irradiation was determined, through the calibration curve of the NIST applicator. The maximum percentual deviations obtained for the CaSO 4 :Dy (without and with 10% of graphite) pellets were 1.1% and 5.8%, respectively. In the case of the calibration of the dermatological applicator B, the samples were irradiated five times, during 5, 10 (twice), 15 and 20 min. The maximum percentual deviations obtained for the CaSO 4 :Dy (without and with 10% of graphite) pellets were 5.8% and 5.1%, respectively. The absorbed-dose rates were obtained from the calibration curves of the NIST applicator for both kinds of pellets. The results for applicators A and B are shown in Table 1. The difference between the results obtained for the applicator A was 4.6%, and for the applicator B it was only 0.41%. These results were considered satisfactory, because the NIST calibration certificate presents 6.0% as uncertainty for the absorbed dose rate of the standard applicator used in this work as the reference source. Table 1: Absorbed-dose rates obtained for the two dermatological applicator. Applicator A B Pellets Absorbed-Dose Rate (Gy/h) CaSO 4 :Dy 107.47 ± 17.01 CaSO 4 :Dy + 10% of graphite 102.79 ± 15.70 CaSO 4 :Dy 125.12 ± 19.24 CaSO 4 :Dy + 10% of graphite 124.61 ± 19.31 4. Conclusion The results obtained for the reproducibility study of the thin CaSO 4 :Dy pellets with and without graphite are within the expected values. The pellets have similar dosimetric characteristics, which is of great importance. From the calibration curve of both materials, obtained with a standard dermatological applicator calibrated by NIST, the calibration of the two dermatological applicators of IPEN was possible. The results obtained for applicator A could be compared with results of previous work and from its certificate, but for applicator B this was not possible, because its nominal activity was unknown. As the absorbed-dose rates and uncertainties obtained for the applicator A were satisfactory, it is possible to conclude that the values obtained for the applicator B are real, because the same calibration procedures were utilized. Therefore, the results of the calibration of the applicators A and B show that the TL materials present great usefulness for dosimetry and calibration of dermatological applicators. 4

Acknowledgments The authors would like to thank Dr. Letícia L. C. Rodrigues for providing the CaSO 4 :Dy pellets, and Comissão Nacional de Energia Nuclear (CNEN), Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq) and Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP), Brazil, for partial financial support. REFERENCES [1] FRIEDELL, H.L., THOMAS, C.I., KROHMER, J.S., Beta-ray application to the eye, with the description of an applicator utilizing 90 Sr and its clinical use, Am. J. Ophthalmol. 33 (1950) 525-535. [2] SOARES, C.G., Comparison of NIST and manufacturer calibrations of 90 Sr+ 90 Y ophthalmic applicators, Med. Phys. 22 (1995) 1487-1493. [3] INTERNATIONAL ATOMIC ENERGY AGENCY, Calibration of photon and beta ray sources used in brachytherapy, IAEA-TECDOC-1117, Vienna (2002). [4] DE ALMEIDA, C., DEWERD, L., JÄRVINEN, H., SOARES, C., Guidelines for the calibration of low energy photon sources and beta-ray brachytherapy sources, SSDL Newsletter 43 (2000) 4-25. [5] DIAS, S.K., CALDAS, L.V.E., Development of an extrapolation chamber for the calibration of beta-ray applicators, IEEE Trans. Nucl. Scie. 45 (1998) 1666-1669. [6] OLIVEIRA, M.L., CALDAS, L.V.E., A special mini-extrapolation chamber for calibration of 90 Sr+ 90 Y sources, Phys. Med. Biol. 50 (2005) 2929-2936. [7] OLIVEIRA, M.L., CALDAS, L.V.E., Performance of a prototype of an extrapolation minichamber in various radiation beams, Appl. Radiat. Isot. 65 (2007) 975-979. [8] REFT, C.S., KUCHNIR, F.T., ROSENBERG, I., MYRIANTHOPOULOS, L.C., Dosimetry of 90 Sr ophthalmic applicators, Med. Phys. 17 (1990) 641-646. [9] OLIVEIRA, M.L., CALDAS, L.V.E., Performance of different thermoluminescence dosemeters in 90 Sr+ 90 Y radiation fields, Radiat. Prot. Dosim. 111 (2004) 17-20. [10] OLIVEIRA, M.L., CALDAS, L.V.E, Performance of thin CaSO 4 :Dy pellets for calibration of a 90 Sr+ 90 Y source, Nucl. Instr. Meth. Phys. Res. A 580 (2007) 293-295. 5