Insert dissociation after fixed bearing PS constrained Genesis II total knee arthroplasty. A case series of nine patients

Similar documents
H.P. Teng, Y.J. Chou, L.C. Lin, and C.Y. Wong Under general or spinal anesthesia, the knee was flexed gently. In the cases of limited ROM, gentle and

Kinematics Analysis of Different Types of Prosthesis in Total Knee Arthroplasty with a Navigation System

Clinical Results of Genesis-I Total Knee Arthroplasty

Intramedullary Tibial Preparation

Comparison of high-flex and conventional implants for bilateral total knee arthroplasty

Extramedullary Tibial Preparation

Surgical Technique. VISIONAIRE FastPak Instruments for the LEGION Total Knee System

CLINICAL AND OPERATIVE APPROACH FOR TOTAL KNEE REPLACEMENT DR.VINMAIE ORTHOPAEDICS PG 2 ND YEAR

DePuy Attune CR and Attune PS. Contents Recorded Usage in NJR Patient and Procedure Details Revision and Survivorship APPENDIX A Component List

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing

Evolution. Medial-Pivot Knee System The Bi-Cruciate-Substituting Knee. Key Aspects

DePuy Attune CR and Attune PS. Contents Recorded Usage in NJR Patient and Procedure Details Revision and Survivorship APPENDIX A Component List

ATTUNE KNEE SYSTEM: SOFCAM CONTACT

Surgical Technique. VISIONAIRE Disposable Instruments for the LEGION Total Knee System

Sasaki E 1,2, Otsuka H 2, Sasaki N 2, and Ishibashi Y 1

Retrospective Study of Patellar Tracking in an Anatomical, Motion Guided Total Knee Design. Adam I. Harris, M.D. & Michelle Ammerman

Clinical evaluation of 292 Genesis II posterior stabilized high-flexion total knee arthroplasty: range of motion and predictors

Knee Revision. Portfolio

DePuy Attune CR and Attune PS

Why precision is powerful

AOANJRR Automated Industry Report Depuy Synthes Australia Attune PS Total Knee Data Period: 1 September August 2018

Automated Industry Report 824 Depuy Synthes Australia Attune PS Total Knee

TOTAL KNEE ARTHROPLASTY (TKA)

Unicompartmental Knee Replacement

Diagnosis and Management of Knee Conditions. Jenny Love / Lynn Robertson AFLAR Oct 2009

Complications of Total Knee Arthroplasty

A study of functional outcome after Primary Total Knee Arthroplasty in elderly patients

Zimmer FuZion Instruments. Surgical Technique (Beta Version)

TOTAL KNEE ARTHROPLASTY SYSTEM

JOINT RULER. Surgical Technique For Knee Joint JRReplacement

Analysis of factors affecting range of motion after Total Knee Arthroplasty

Case Report Meniscal Bearing Dislocation of Unicompartmental Knee Arthroplasty with Faint Symptom

Revolution. Unicompartmental Knee System

Automated Industry Report 823 Depuy Synthes Australia Attune CR Total Knee

Case report. Open Access. Abstract

CONSENSUS ORTHOPEDICS INC. CONSENSUS KNEE SYSTEM

Advanced instrumentation for surgical simplicity

Comprehensive Solutions

Stephen R Smith Northeast Nebraska Orthopaedics PC. Ligament Preserving Techniques in Total Knee Arthroplasty

LEGION Total Knee System. Clinical Heritage

CLINICAL PRESENTATION AND RADIOLOGY QUIZ QUESTION

NexGen Legacy LPS-Flex Knee. Brochure

Constrained Posterior Stabilized (CPS)

PRIMARY. ConforMIS itotal G2 XE and itotal G2 (Bicondylar tray)

Dora Street, Hurstville 160 Belmore Road, Randwick

Important notice: the device(s) can be prescribed and implanted only by a doctor legally authorized to perform this type of surgery.

Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs: Report of Two Cases

Ankylosis due to heterotopic ossification following primary total knee arthroplasty

LEGION. Total Knee System. Clinical Heritage

Knee kinematics after TKA depends on preoperative kinematics

Unicondylar Knee Vs Total Knee Replacement: Is Less Better In the Middle Aged Athlete

Knee arthroplasty: What radiologists should know.

Constrained Posterior Stabilized (CPS) Surgical Technique

Persona. The Personalized Knee. Trabecular Metal Tibia. Surgical Technique

Total Knee Original System Primary Surgical Technique

Extra-articular deformities in TKA

Enhanced Stability Constrained Liners. Design Rationale Surgical Technique

Tibial Base Design Factors Affecting Tibial Coverage After Total Knee Arthroplasty: Symmetric Versus Asymmetric Bases

Metal block augmentation for bone defects of the medial tibia during primary total knee arthroplasty

Augmented Glenoid Component for Bone Deficiency in Shoulder Arthroplasty

THE KNEE SOCIETY VIRTUAL FELLOWSHIP

ESC. Enhanced Stability Liners. Design Rationale & Surgical Technique

Title of the article: the epidemiology of revision knee arthroplasty at a tertiary care center in india.

Condylar constrained system in primary total knee replacement: our experience and literature review

Correlation between Knee Kinematics and Patello-femoral Contact Pressure in Total Knee Arthroplasty

Gold standard of a TKA. Conflicting goals? POLYETHYLENE WEAR THE SOLUTION: MOBILE BEARING KNEES. MOBILE BEARING A totally new approach (1977)

Scorpio NRG PS (cementless)/series 7000 (cementless) Total Knee Investigation

Surgical Technique Final Trial Reduction and Component Implantation of

Coronal Instability after Total Knee Arthroplasty using Bi-Surface type 3 : An Averaged 10 Year Follow-up Study

Saiph Knee System. Technical Dossier

ACL Athletic Career. ACL Rupture - Warning Features Intensive pain Immediate swelling Locking Feel a Pop Dead leg Cannot continue to play

Clinical Results of Lateral Unicompartmental Knee Arthroplasty: Minimum 2-Year Follow-up

Functional Outcome of Uni-Knee Arthroplasty in Asians with six-year Follow-up

Stefan Rahm MD University Hospital Balgrist

SEVERE VARUS AND VALGUS DEFORMITIES TREATED BY TOTAL KNEE ARTHROPLASTY

For Commercial products, please refer to the following policy: Preauthorization via Web-Based Tool for Procedures

Total Knee Arthroplasty in a Patient with an Ankylosing Knee after Previous Patellectomy

Effects of Posteromedial Vertical Capsulotomy on the Medial Extension Gap in Cruciate-retaining Total Knee Arthroplasty

Duraloc CONSTRAINED LINER

EARLY CLINICAL RESULTS OF PRIMARY CEMENTLESS TOTAL KNEE ARTHROPLASTY

BIOMECHANICAL MECHANISMS FOR DAMAGE: RETRIEVAL ANALYSIS AND COMPUTATIONAL WEAR PREDICTIONS IN TOTAL KNEE REPLACEMENTS

ATTUNE Knee System: Cruciate Sacrificing Fixed Bearing

Contemporary Advances in Hip, Knee, and Shoulder Arthroplasty April 11, 2015

Bicruciate-Retaining or Medial Pivot Total Knee Prosthesis Pritchett 225 Fig. 3. The MP total knee prosthesis. Fig. 1. An anteroposterior radiograph o

Knee Replacement Complications

Complications of Medial Unicompartmental Knee Arthroplasty

Appendix E-1 (Figures and Tables) Fig. E-1

Triathlon Knee System

Anterior knee pain following total knee replacement caused by isolated Paget's disease of patella

ATTUNE Knee System: Stability in Total Knee Replacement

Case Report Total Knee Arthroplasty in a Patient with Bilateral Congenital Dislocation of the Patella Treated with a Different Method in Each Knee

Biomechanics of. Knee Replacement. Mujda Hakime, Paul Malcolm

Mid-term Results of Total Knee Arthroplasty Using PFC Sigma RP-F

Mako. Total Knee with Triathlon. Surgical protocol

Bipolar Radial Head System

Consistency in U2 Knee System over Time

THE KNEE SOCIETY VIRTUAL FELLOWSHIP

The effect of polyethylene creep on tibial insert locking screw loosening and back-out in prosthetic knee joints

NexGen CR-Flex and LPS-Flex Knees. Design Rationale

National Joint Replacement Registry. Lay Summary 2015 Annual Report Hip and Knee Replacement

Transcription:

Acta Orthop. Belg., 2015, 81, 747-751 ORIGINAL STUDY Insert dissociation after fixed bearing PS constrained Genesis II total knee arthroplasty. A case series of nine patients Timothy Voskuijl, Thijs A. Nijenhuis, Gijs G. Van Hellemondt, Jon H. M. Goosen From Department of Orthopaedic Surgery, Sint Maartenskliniek, Nijmegen, The Netherlands Dissociation of the polyethylene insert after fixed bearing posterior stabilized Genesis II total knee arthroplasty has been rarely described. We present a case series of nine patients with a dissociation of the insert within a period of two years after surgery. Revision surgery was performed in all patients. In this report we discuss clinical presentation, patient characteristics and possible etiologies for tibial insert dissociation seen in the presented cases. In conclusion, tibial insert dissociation does t lead to a uniform clinical presentation. Therefore, in this point of view regular physical examination and imaging after TKA regardless the presence of symptoms seems to be indicated. Keywords : insert dissociation ; Genesis II ; polyethylene ; total knee arthroplasty. Introduction Tibial insert dissociation has been listed by The Knee Society as one of 22 complications and adverse events important for reporting outcomes of Total Knee Arthroplasty (TKA) (3). This complication is rare since the insert is primarily loaded with a compressive force and there is a relative absence of tensile stress on the insert-baseplate locking mechanism (2). In mobile bearing TKA systems a higher incidence of dissociation of the insert is observed than fixed bearing TKA (8,9). In fixed bearing TKA, it is nearly always associated with cruciate retaining (CR) type prostheses (8). Main reasons for dissociation of a fixed bearing insert after TKA described in previous literature include a) misplacement of the insert, b) inadequate ligament balancing (eg resulting from trauma) causing the locking mechanisms to fail due to excessive polyethylene wear, and c) impingement of the insert on soft tissue or osseous structures in flexion (5,8,9). Furthermore, the use of a posterior stabilized (PS) insert with post can contribute to polyethylene insert dissociation due to strong lift-off forces during high flexion damaging the post of the insert and eventually dissociating it (5,8). The Genesis II prosthesis (Smith and Nephew, Memphis, TN) having both CR and PS designs has an asymmetrical tibial baseplate for maximal coverage of the asymmetrical resected tibial surface and an asymmetrical femoral component with a posterolateral femoral condyle thicker than the posteromedial condyle for filling of the trapezoidal n Timothy Voskuijl, MD. n Thijs A. Nijenhuis, MD. n Gijs G. van Hellemondt, MD. n Jon H.M. Goosen, MD, PhD. Department of Orthopaedic Surgery, Sint Maartenskliniek, Nijmegen, The Netherlands. Correspondence : J. H. M. Goosen, MD, PhD, Sint Maartenskliniek Nijmegen, Department of Orthopaedic Surgery, Knee Reconstruction Unit, PO Box 9011, 6500 GM Nijmegen, The Netherlands. E-mail : j.goosen@maartenskliniek.nl 2015, Acta Orthopædica Belgica. No benefits or funds were received in support of this study. The Acta authors Orthopædica report Belgica, conflict Vol. of 81 interests. - 4-2015 voskuijl-.indd 747 18/01/16 13:19

748 t. voskuijl, t. a. nijenhuis, g. g. van hellemondt, j. h. m. goosen flexion space without externally rotating (6). Using revision for any reason as endpoint, the mean five years survival rate of this prosthesis as primary TKA is 97.0%. Most important device-related causes underlying failure of this prosthesis are polyethylene wear, aseptic loosening, instability, malalignment or malposition (1). Dissociation of the polyethylene insert from the tibial baseplate in the high flexion PS Genesis II TKA is described in a couple recent case reports (5,8,9). We present a series of nine cases in which dissociation of the insert after fixed bearing posterior stabilized Genesis II TKA was seen. Material and methods From January 2012 to December 2013 885 fixed bearing PS constrained Genesis II total knee arthroplasties were performed in our clinic by eight different orthopaedic surgeons. During these two years nine patients were diagsed with dissociation of the polyethylene insert from tibial baseplate after this type TKA. Since this complication is rarely described in literature, we performed a retrospective analysis on patients records. Acquired data included demographic data (sex, age, length, weight), patient history and clinical presentation during insert dissociation. Furthermore we reviewed radiological images made after TKA, and reports of the revision surgery for insert dissociation. Possible aetiologies for insert dissociation described in earlier literature were compared to the cases in our series. Results Insert dissociation was diagsed in nine patients after PS constrained Genesis II TKA. Of these nine patients, five were female ; the mean BMI was 30.5 ; mean age during primary TKA was 63.8 years (Table I). Indication for the primary TKA in all patients was osteoarthritis, with two of the patients having prior surgery in the affected knee. None of the patients had rheumatoid arthritis. In all patients the components were fixed with cement, and five patients had their patella resurfaced during primary TKA. In six patients dissociation of the polyethylene insert followed primary surgery for TKA ; in one patient insert dissociation was diagsed after manipulating the knee under narcosis for arthrofibrosis after primary TKA ; two patients had insert dissociation after preceding insert renewal, of which one was performed because of instability and the other because of suspicion of early infection. Surgeries prior to dissociation were performed by four different orthopaedic surgeons. Mean duration between diagsis of dissociated insert and latest surgery was 145 days. A delay in diagsing the dissociation of the insert and thus in revision surgery was seen in four patients. The dissociation was missed on earlier X-rays in three patients (with one patient having two earlier X-rays) by both orthopedic surgeon and radiologist and in one patient by the treating orthopedic surgeon (Table I). Table I. Patient characteristics and diagsis of dissociated insert M/F BMI age latest surgery prior to diagsis of dissociation T (days) latest surgery diagsis of dissociation insert dissociation missed on earlier X-ray resulting delay in surgery Patient 1 F 31.9 44.4 liner exchange (instability) 47 by radiologist only Patient 2 F 31.6 59.1 total knee arthroplasty 49 by radiologist only Patient 3 M 28.6 77.6 total knee arthroplasty 30 t applicable Patient 4 F 29.1 72.9 total knee arthroplasty 133 t applicable Patient 5 M 42.3 63.3 total knee arthroplasty 89 by both R and O 6 weeks Patient 6 M 26.6 58.7 liner exchange (infection) 348 by O only 10 months Patient 7 M 28.4 68.2 total knee arthroplasty 433 by both R and O 8 months (twice) Patient 8 F 27.6 56.1 manipulation (arthrofibrosis) 125 by both R and O 8 weeks Patient 9 F 28.0 74.2 total knee arthroplasty 50 t applicable M = Male; F = Female; BMI = Body Mass Index; T = time; R = radiologist; O = orthopaedic surgeon. voskuijl-.indd 748 18/01/16 13:19

fixed bearing ps constrained genesis ii total knee arthroplasty 749 Trauma prior to insert dissociation was documented for ne of the patients. Documented patient history revealed one patient ticed a clicking sensation, one patient complained about instability in extension and one patient ticed a decreased range of motion. Three patients complained about pain (one of them specifically about anterior pain). Three patients had complaints at all. Physical examination revealed an extensor lag in three patients and instability was found in ather three patients. All patients were diagsed anterior dislocation of the polyethylene insert on plain X-rays. On the post-operative X-ray of two patients an osteophyte was seen on the dorsal side of the distal femur. During revision surgery, all inserts were indeed dislocated anteriorly. A possible physical cause for the dislocation was seen in 5 patients. In one operation a deformation of the postero-lateral lip of the dislocated insert was observed. Wear of the metal components by polyethylene displacement and potentially metal on metal contact were t documented in operation reports (Table II, Fig. 1). The dislocated inserts were removed and replaced, with three new inserts having a 2 mm increase in thickness. One of these was placed in a patient who already had a prior insert renewal for suspicion on early infection. During the revision surgery, tissue samples were taken for culturing in one patient, but these showed bacterial growth. One of the revision surgeries was complicated by an infection, for which a re-operation was performed. None of the new inserts had a recurrent dislocation. Can you discuss if there was wear of the metal components by poly displacement and potentially metal on metal contact. See line 150-152. Please explain if any recidive was observed after poly exchange except the case mentioned above here. Already stated in line 169? Discussion Worldwide, most common complications leading to revision surgery after TKA are aseptic loosening and periprosthetic joint infection (6). Only a couple recent case reports describe dissociation of the Table II. Clinical presentation of insert dissociation history physical examination (range of motion ; stability) Patient 1 pain (anteriorly) ROM t described ; clear anterior luxation Patient 2 complaints extensor lag ; stable Patient 3 decreased ROM extensor lag (15 degrees) ; stability t described Patient 4 clicking sensation extensor lag (5 degrees) ; stability t described Patient 5 complaints extensor lag ; slight instability Patient 6 complaints extensor lag ; stable Patient 7 pain ; instability in extension extensor lag ; stability t described Patient 8 pain extensor lag (10 degrees) ; evident subluxation in flexion Patient 9 t documented ROM and stability both t documented radiology (osteophyte dorsal distal femur) yes yes intra-operative findings damage / deformation of the insert t documented some syvitis and hydrops t documented t documented ROM = range of motion. voskuijl-.indd 749 18/01/16 13:19

750 t. voskuijl, t. a. nijenhuis, g. g. van hellemondt, j. h. m. goosen B A Fig. 1. Intra-operative findings. A : Anterior dissociation of the insert. B : deformation of the locking mechanism (posterolateral dovetail). polyethylene insert from the tibial baseplate after fixed bearing high flexion PS Genesis II TKA (8,9,5). Despite this complication is commonly labelled rare, we presented a case series of nine tibial insert dissociations within two years in our clinic. Trauma and failure of the locking mechanism are usually presented as a cause for tibial insert dissociation (5). Trauma may result in ligament laxity, likely causing polyethylene wear which in turn might result in late dissociation of the insert (2,8). However, in ne of our cases a history of trauma was documented. Nevertheless, one knee was manipulated because of arthrofibrosis before the insert dislocated. Locking mechanisms regularly used in TKA can be categorized as linear, peripheral or central capture mechanisms. The design of the capture mechanism and the multi-axial dynamic loading of the components influence magnitude and direction of relative motion between insert and tibial tray (10). If locking mechanisms fail, the polyethylene insert tends to lift-off anteriorly of the tibial baseplate during flexion because of the downward force on the posterior half of the insert (2,10). Anterior dislocation is then presumably caused by the femoral cam engaging the tibial post in flexion, thereby producing an anteriorly directed force (4). Repetitive and minute anterior lift-off may cause wear of the snap-fit locking mechanism and eventually failure (2). Reasonably, higher BMI results in greater downward force during flexion, which in turn increases the risk of anterior lift-off and failure of the locking mechanism. All patients in our series had BMI over 25 (overweight), with three of them being obese (BMI > 30). Previous liner exchange may also cause weakening of the locking mechanism, resulting in higher chances of recurrent dislocation (9). Two of the cases we presented actually had preceding liner exchange before dislocation. Adequate placement of the insert remains critical for proper functioning of the locking mechanism. As presented in an earlier case report, the surgical exposure using the mini subvastus approach is limited because of the extensor mechanism and misplacement of the insert or soft tissue impingent could occur (5). Osseous impingement of an osteophyte at the dorsal distal femur on the posterior insert in flexion may lead to gradual posterior lift-off of the insert from the baseplate (8,9). In retrospect, the postoperative X-ray of two of our cases showed an osteophyte at the dorsal side of the distal femur, possibly causing impingement leading to dissociation. In four patients a delay in diagsing the dissociation and thus revision surgery was seen since voskuijl-.indd 750 18/01/16 13:19

fixed bearing ps constrained genesis ii total knee arthroplasty 751 A Fig. 2. Insert dissociation after Total Knee Arthroplasty may be missed on standard lateral view with extended knee (A) while lateral view of the flexed knee (B) may reveal insert dissociation more evidently. dissociation was missed on standard postoperative X-rays with anteroposterior and lateral view of the extended knee. However, a lateral view with the knee flexed 90 degrees eventually did expose the dissociation (Fig. 2). Previous case reports suggest the PS Genesis II design may be prone to dissociation due to the combination of the shallow anterior tab snap-fit locking mechanism, facilitating transition from initial liftoff to complete dissociation of the insert from tibial component, and thin dovetail lips susceptible to damage (5,8). In one of our cases deformation of the posterolateral lip of the insert was found (Fig. 1). None of the operation reports mentioned a damaged tibial post. However, it is believed that strong liftoff forces during high flexion can damage the post of the insert, eventually leading to dissociation of the insert. One case report describing recurrent dislocation of PS insert, states after conversion to a n-posted deep dished insert recurrent dissociation occurred (5). Although it has been assumed that high flexion PS inserts may increase the risk of component wear, and so increase the risk on insert dissociation, compared to standard PS inserts, analyses indicate the high flexion design of the Genesis II does t harmfully affect outcomes of TKA (1). B Although previous case reports present insert dissociation diagsed because of mechanical malfunctioning, pain or clunking sounds, t all patients in our series presented with a malfunctioning knee prosthesis. In fact, two of them were completely satisfied with their prosthesis and had disturbing abrmalities during physical examination. This indicates the urge for both thorough physical examination and imaging after TKA regardless the presence of symptoms indicating a complication. Postoperative X-ray with lateral view of the knee flexed 90 degrees may be considered. A systematic approach could prevent a delay in needed revision surgery for insert exchange. References 1. Bhandari M, Pascale W, Sprague S, Pascale V. The Genesis II in primary total knee replacement : a systematic literature review of clinical outcomes. Knee 2012 ; 19 (1) : 8-13. 2. Chen CE, Juhn RJ, Ko JY. Dissociation of polyethylene insert from the tibial baseplate following revision total knee arthroplasty. J Arthroplasty 2011 ; 26 (2) : 339.e11-3. 3. Healy WL et al. Complications of total knee arthroplasty : standardized list and definitions of the Knee Society. Clin Orthop Relat Res 2013 ; 471 (1) : 215-20. 4. Hepinstall MS, Rodriguez JA. Polyethylene subluxation : a radiographic sign of locking mechanism failure after modular total knee arthroplasty. J Arthroplasty 2011 ; 26 (1) : 98-102. 5. In Y, Sur YJ, Won HY, Moon YS. Recurrent dissociation of the tibial insert after mini-subvastus posterior-stabilized total knee arthroplasty : a case report. Knee 2011 ; 18 (6) : 461-3. 6. Kasahara Y, Majima T, Kimura S, Nishiike O, Uchida J. What are the causes of revision total knee arthroplasty in Japan? Clin Orthop Relat Res 2013 ; 471 (5) : 1533-8. 7. Laskin RS, Davis J. Total knee replacement using the Genesis II prosthesis : a 5-year follow up study of the first 100 consecutive cases. Knee 2005 ; 12 (3) : 163-7. 8. Lee DH, Lee TG, Park SJ, Han SB. Spontaneous late dissociation of the tibial insert after high-flex posteriorstabilized Genesis II total knee arthroplasty. J Arthroplasty 2013 ; 28 (2) : 374.e13-6. 9. Rutten SG, Janssen RP. Spontaneous late dislocation of the high flexion tibial insert after Genesis II total knee arthroplasty. A case report. Knee 2009 ; 16 (5) : 409-11. 10. Thienpont E. Failure of tibial polyethylene insert locking mechanism in posterior stabilized arthroplasty. Knee Surg Sports Traumatol Arthrosc 2013 ; 21 (12) : 2685-8. voskuijl-.indd 751 18/01/16 13:19