Improved Tissue Phantoms for Experimental Validation of Microwave Breast Cancer Detection

Similar documents
EFFECTS OF FIBROGLANDULAR TISSUE DISTRIBU- TION ON DATA-INDEPENDENT BEAMFORMING AL- GORITHMS

EFFECTS OF DIELECTRIC HETEROGENEITY IN THE PERFORMANCE OF BREAST TUMOUR CLASSIFIERS

Time-Domain Microwave Breast Cancer Detection: Extensive System Testing with Phantoms

UWB Microwave Breast Cancer Detection Using SAR

CONFOCAL MICROWAVE IMAGING FOR BREAST TUMOR DETECTION: A STUDY OF RESOLUTION AND DETECTION ABILITY

Electrical impedance characterization of normal and cancerous human hepatic tissue

MICROWAVE IMAGING FOR THE DETECTION AND LOCALIZATION OF BREAST CANCER USING ARTIFICIAL NEURAL NETWORK

Radar Based Breast Cancer Detection Using SAR

Indirect Holography Breast Cancer Detection System (IH-BCDS) Using VORD

Exploring Joint Tissues with Microwave Imaging

BREAST CANCER DETECTION BASED ON DIFFEREN- TIAL ULTRAWIDEBAND MICROWAVE RADAR

Development of Breast Tissue Phantoms for Enhanced Terahertz Imaging Utilizing Microdiamond and Nano-Onion Particles

SPIKING NEURAL NETWORKS FOR BREAST CANCER CLASSIFICATION IN A DIELECTRICALLY HETEROGE- NEOUS BREAST

3D EXPERIMENTAL DETECTION AND DISCRIMINA- TION OF MALIGNANT AND BENIGN BREAST TUMOR USING NN-BASED UWB IMAGING SYSTEM

BREAST CANCER DIAGNOSIS USING MICROWAVE AND HYBRID IMAGING METHODS

Design and analysis of Microstrip Patch Antenna for Lung Tumor

Microwave breast imaging with a monostatic radar-based system: a study of application to patients

NUMERICAL MODELLING FOR ULTRA WIDEBAND RADAR BREAST CANCER DETECTION AND CLASSI- FICATION. Informatics, National University of Ireland Galway, Ireland

An Inexpensive and Easy Simulation Model of Ocular Ultrasound That Mimics Normal Anatomy as Well as Abnormal Ophthalmologic Conditions

Visualizing transplanted muscle flaps using minimally invasive multielectrode bioimpedance spectroscopy

The Doppler Dot - September 2017 Product Design Specifications

Ingredient Listing Qty. Unit NDC # Supplier

Breast Cancer Detection Using Flexible Microstrip Antenna

RECONSTRUCTION OF EXTREMELY DENSE BREAST COMPOSITION UTILIZING INVERSE SCATTERING TECHNIQUE INTEGRATED WITH FREQUENCY-HOPPING APPROACH

SUPPORT VECTOR MACHINES FOR THE CLASSIFICA- TION OF EARLY-STAGE BREAST CANCER BASED ON RADAR TARGET SIGNATURES

(a) Find a 95% confidence interval on the difference in mean active concentrations for the two catalysts.

Multi-Modality Phantom Development

13. SUPPOSITORY Suppository Bases. The active substance is prepared in a suitable bases. An ideal suppository bases should carry:

Universal Plunger Loc Attachment

A reliable, practical and easy-to-use Determination of Soil Conductivity and Salt Content with TRIME Probes

Report. Novozymes Meat protein extract in chopped ham Jakob Søltoft-Jensen. Summary. Background


TISSUE DIELECTRIC CONSTANT (TDC) AS AN INDEX OF LOCALIZED ARM SKIN WATER: DIFFERENCES BETWEEN MEASURING PROBES AND GENDERS

Different types of starches (corn, potato, waxy corn, amylomaize and pregelatinized

10 th International Junior Science Olympiad, Pune, India. Experimental Tasks. Examination Rules: Task: C Tomato (6 marks for this task)

Vessels by Design: Basic Vessel Anatomy. Student Information Page 3A

1. INTRODUCTION ABSTRACT

The Aerosol Dispensers Directive. Labelling Requirements

limbsandthings.com Advanced Female Catheterisation Trainer User Guide For more skills training products visit Limbs & Things Ltd.

Respiratory System. Organization of the Respiratory System

Breast Pedicle Protector

Patient Dosimetry in Mammography and Tomosynthesis:

Teaching Presentation. Lesson plan- Blueberry and Walnut Clusters. Anamarie Orgera 12/5/2011. J. Frank

50 Shades of Teflon Part 1

Determination of the antioxidant activity by the Rancimat method

A tissue-equivalent phantom series for mammography dosimetry

Dielectric characterization of healthy and malignant colon tissues in the GHz frequency band

Ship the Chip. Food packages have four main functions besides providing food containers.

Fluoxetine Hydrochloride 5.6 mg Oral Chewable Treats (Solid Suspension, 60 x 1 ml Chewable Treats)

A novel equivalent circuit model for gap-connected cells

Molecular Gastronomy Lab Lesson Plan: Spherification and Olive Oil Powder For an advanced high school Chemistry class

DANECHURN PRODUCED IN DENMARK 2 xxxxxxxxxxxxxxxxxxxxxxx Danechurn Spreadable

Breast Pedicle Protector

6.02 Uniformity of Dosage Units

Compressor Nebulizer System

Meshes. Meshes. Non-absorbable meshes. Absorbable meshes

Chapter 9 The body and its movement

limbsandthings.com Advanced Catheterisation Trainer User Guide For more skills training products visit Limbs & Things Ltd.

Everyday Life in the Research World

Reference to primary publication Properties of alginates in ZWP 4/2014 Xantalgin Crono - Introduction

Adding a New Dimension to Osseointegration. Thinner. Faster. Stronger.

Brief Communication. Variation in Dielectric Properties Due to Pathological Changes in Human Liver. Hazards, Public Health England, Didcot, UK

DNA Extraction. Teachers Guide Introduction:

Assessment of Adipose Tissue from Whole Body 3T MRI Scans

Title: The Effect of Flaxseed Meal on the Texture and Water Activity of Brownies

The right impact on taste and texture CONDITIONING COATING BROCHURE

Measurement of Tibial Translation in Dogs with Anterior Cruciate Ligament Rupture

Heart Abnormality Detection Technique using PPG Signal

ISO 1567 INTERNATIONAL STANDARD. Dentistry Denture base polymers. Art dentaire Polymères pour base de prothèses dentaires. Third edition

THE DEVELOPMENT AND MANUFACTURE OF FIXED- ULTRASONIC INSPECTION REFERENCE REFLECTORS AND TRANSDUCERS FOR COMPRESSOR BLADE DOVETAILS

2. Identify and distinguish between homogeneous and heterogeneous mixtures.

Oral Specimen Collection Device

Understanding Your Pap Test Results

Male Reproductive System

An Introduction to Nanosomes from

Methimazole 5 mg Oral Chewable Treats (Solid Suspension, 60 x 1 ml Treats) FIN F v2. Ingredient Listing Qty. Unit NDC # Supplier

COMMISSION DIRECTIVE 97/48/ EC. of 29 July 1997

A NEW STRATEGY OF DRUG DELIVERY

Biomechanical Elbow Model. Rochester Institute of Technology, Rochester, NY, USA

Mammography. The Lebanese Society of Obstetrics and Gynecology. Women s health promotion series

RESEARCH REVIEW Diagnostic Performance of a Novel Device for Real-Time Margin Assessment in Lumpectomy Specimens

Acrylic thinner, standard, fast, slow and extra slow

COMPARISONS OF NEEDLE INSERTION IN BRACHYTHERAPY PROTOCOLS USING A SOFT TISSUE MODEL

MODELING OF THE CARDIOVASCULAR SYSTEM AND ITS CONTROL MECHANISMS FOR THE EXERCISE SCENARIO

THE PROBLEM JOB OFFER. Oxygen, not always an ally. Department Head, Food Conservation Research Unit

HiPer Western Blotting Teaching Kit

Lipovisor is a unique medical device developed and tested by a team of medical specialists and technicians. It is able to detect and display

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

2. Identify and distinguish between homogeneous and heterogeneous mixtures.

Tissues 10/21/2016. Epithelial Tissue

Lesson 9A Tissues in Animals

ANNEX I SUMMARY OF PRODUCT CHARACTERISTICS

April-May, Diabetes - the Medical Perspective Diabetes and Food Recipes to Try Menu Suggestions

Improved Inspection of Composite Wind Turbine Blades with Accessible Advanced Ultrasonic Phased Array Technology

Dr.N.Damodharan Professor and head Department of pharmaceutics SRM college of pharmacy

The product must be only used by specifically trained personnel and under the relevant application by proven techniques.

NPG-151. Characteristics Unit Conditioning Typical Values SPEC Test Method. Volume resistivity MΩ-cm C-96/35/ x

Glaucoma What You Should Know

Transcription:

Improved Tissue Phantoms for Experimental Validation of Microwave Breast Cancer Detection Emily Porter, Jules Fakhoury, Razvan Oprisor, Mark Coates and Milica Popović Department of Electrical and Computer Engineering, McGill University 3480 University Street, Montreal, QC, Canada {emily.porter, jules.fakhoury, razvan.oprisor}@mail.mcgill.ca {mark.coates, milica.popovich}@mcgill.ca Abstract This paper presents a methodology for constructing realistic breast phantoms for use in microwave detection of breast cancer. Phantoms for fat, skin, gland and tumour are made from a combination of everyday chemicals. The dielectric permittivity and conductivity of the phantoms are measured at microwave frequencies. Finally, we describe a procedure to combine the four phantom tissue types into a single breast phantom that is sufficiently realistic, both in terms of dielectric properties and physiological structure, to be used in experiments designed to assess the performance of microwave breast cancer detection techniques. Index Terms Breast cancer detection, microwave imaging, tissue phantoms. I. INTRODUCTION Microwave imaging of the breast, intensely researched in the past decade, assumes a difference in the dielectric properties between healthy breast tissue and tumour tissue at microwave frequencies [1]. Realistic breast models, also called phantoms, which mimic the dielectric properties of actual human breasts, are needed in order to test and further develop microwave imaging technologies. Recently, two studies have been conducted which characterized the microwave dielectric properties of healthy and malignant breast tissue [2], [3]. These provided the most comprehensive set of measurements of the permittivity and conductivity of different breast tissues available to date and, therefore, make a good starting point for construction of tissue phantoms. In [4], an oil-in-gelatin mixture is used as a phantom material and it is shown that with varying concentrations, materials with a large range of dielectric properties can be obtained. In this paper, we present a method for constructing breast phantoms with dielectric properties that mimic those of actual breast tissue. For the phantoms to be as realistic as possible, four different tissue-mimicking models are made: fat, skin, glands and tumour. Here, we build on the procedure suggested for preparing tissue-mimicking materials in [4] by identifying the specific concentrations of chemicals that lead to phantoms with dielectric properties that closely fit those of each of the four tissue types. Multiple phantoms are made using these sets of chemical concentrations and the permittivity and conductivity of each is measured from 200 MHz to 6 GHz. Finally, we propose techniques that allow us to incorporate the four phantom types into one stable, hemispherical breast model. The resulting breast phantom consists of a 2-mm layer of skin enclosing a mixture of adipose tissue, various sized glands, as well as (possibly) one or more tumours of different sizes. II. METHODOLOGY The construction of the breast phantoms involved three main steps: construction of the individual tissue phantoms from the chemical mixtures, measurements of their respective dielectric properties, and, finally, merging of the four different tissue types into a single complete breast model. A. Step 1: Mixture of Tissue Phantoms The first step in the construction process is to make four phantom types (fat, skin, gland and tumour) that have electrical properties in the microwave range similar to those of human breast tissue. In addition, the choice of materials was driven by a few other advantageous factors: the materials should be low in cost, easy to obtain (off-the-shelf) and stable over long periods of storage time. The phantoms are made from a mixture of gelatin and oil (50% safflower oil and 50% kerosene) [4]. The complete list of ingredients and the amounts or concentrations used for each hemispherical (radius = 6.5 cm) phantom type are given in Table I. TABLE I LIST OF INGREDIENTS AND THE AMOUNT USED FOR EACH TISSUE PHANTOM Amount Used Fat Gland Skin Tumour p-toluic acid (g) 0. 133 0.253 0.294 0.346 n-propanol (ml) 6.96 12.71 28.69 17.00 deionized water (ml) 132.7 241.9 279.5 328.0 200 Bloom gelatin (g) 24.32 43.27 50.02 58.67 Formaldehyde (37% by weight) (g) 1.53 2.74 3.33 3.72 oil (ml) 265.6 141.5 98.6 38.4 Ultra Ivory detergent (ml) 12.00 6.79 5.86 2.00

The chemicals are mixed using the same technique as has been suggested in [4]. Once the chemicals are mixed, it can take 1-2 days for the phantoms to solidify. After this time, they can be removed from the container they were allowed to set in (in the case of our work, a breast-shaped bowl) and will keep their shape. However, to ensure maximum shelf-life, the phantoms should be protected from moisture loss by thin plastic wraps or by placing them in air-tight containers. If they are not properly stored, the phantoms can become dehydrated, with significant change of dielectric properties. B. Step 2: Measurement of Phantom s Dielectric Properties Once the phantoms were allowed to solidify, the dielectric properties were tested. The measurements for the conductivity and relative permittivity for each phantom type were made at room temperature using an HP 85070B dielectric probe attached to a vector network analyser. The probe was held flat on the surface of the phantom. A minimum of ten measurements were taken from different locations on each individual tissue phantom, and averaged. In some cases, extra measurements were taken from inside the phantoms to ensure the mixtures homogeneity. The experimental result showing the relative permittivity variation with frequency for each of the four phantoms is given in Figure 1. The results for the gland and fat phantoms are shown in Figure 1a, along with the median of measurements on actual gland and fat tissue that were conducted in [5]. Figure 1b shows the permittivity for skin and tumour phantoms as well as the median from measurements of actual skin tissue (from [4], [6]) and actual malignant tissue (from [3]). Fig. 1b Relative permittivity versus frequency for tumour and skin tissue phantoms. Also shown is the median of the measurements found in [3], [4] and [6] for actual tumour and skin tissue. Similarly, the measured conductivity values at each frequency for the various phantoms are shown in Figure 2. As with the relative permittivity, first the results for gland and fat phantoms are plotted in Figure 2a. These are accompanied by the median conductivity found in measurements done on actual gland and fat tissue in [5]. Finally, Figure 2b shows the conductivity of the tumour and skin phantoms and the median of measurements of actual tumour and skin tissues [3], [4], [6]. Fig. 2a Conductivity versus frequency for gland and fat tissue phantoms. Also shown is the median of the measurements conducted in [5] on actual gland and fat tissue. Fig. 1a Relative permittivity versus frequency for gland and fat tissue phantoms. Also shown is the median of the measurements conducted in [5] on actual gland and fat tissue.

Fig. 2b Conductivity versus frequency for skin and tumour tissue phantoms. Also shown is the median of the measurements found in [3], [4] and [6] on actual skin and tumour tissue. It is observed that the tumour phantom has, at worst, a deviation of 9 units of relative permittivity from the median of the measurements on actual malignant tissue reported in [3]. The gland phantom has a relative permittivity up to 15 units lower than the median of the gland tissue measured in [5], and the fat phantom s is up to 10 units lower than is measured in [5]. The skin phantom is the most closely matched to its corresponding measurements on actual skin tissue, here the phantom is 5 units away from the median relative permittivity given in [4], [6]. It may seem that the phantom s relative permittivities are not well matched to actual tissue types, but in fact for the tumour and gland phantoms, the relative permittivity is still within the range of values measured on actual tissue [5]. For instance, at 6 GHz, the gland phantom has a relative permittivity of 26, while the median measured value is just over 40 [5]. However, gland tissue measured at 6 GHz has a relative permittivity of anywhere from 20 to 35 [5], so clearly the gland phantom relative permittivity resides within this interval. Thus for the skin, tumour and gland phantom, the difference from the desired values is tolerable, and certainly so when compared with the intrinsic variation of breast tissue properties within the breast and between different patients [3]. It is the fat phantom s relative permittivity which the least representative of actual fat tissue measurements; this will be discussed later. We also note the differences in conductivity between the phantoms and the median of actual tissue samples of 0.5 S/m, 2 S/m, 0.9 S/m and 1 S/m for fat [5], gland [5], tumour [3], and skin [4], [6], respectively. When discussing the variations in dielectric properties between the tissue phantoms and actual tissue, there are two important points that must be mentioned. As stated above, tissue properties can vary widely between patients [3], so it is not necessary for tissue models to be an exact match to the measured quantities. Also, with the exception of the actual skin tissue, all of the measurements referred to have been those conducted on tissue first removed from the body [5], [3]. Recently, a study was published which concludes that in measurements conducted on excised tissue, the dielectric properties tend to decrease as compared to those of living tissue still in the body [7]. Ideally, our breast model should mimic the properties of living tissue that has not been removed, since a breast cancer detection system should be non-invasive. However for the purpose of phantom construction, it is more important to note that the variation in microwave dielectric properties for tissues is large and at present not definitively known, but the tissue phantoms should still approximate them as closely as possible. The above results and discussion has focused on freshly made phantoms. Measurements were also made on a fat phantom that was six months old. The same mixture was used for both of the two fat phantoms. A comparison of the relative permittivity of the fresh fat and the older fat phantoms is shown in Figure 3. It is seen that the relative permittivity follows the same trend for both phantoms, however the old fat phantom has a relative permittivity about four points lower than the fresh fat. The conductivity, similarly, is lower in the older phantom. This brings the old fat phantom s relative permittivity and conductivity into the range expected for actual fat tissue by the studies in [3] and [5]. This observation guides our current work, which focuses on introducing a higher ratio of oil to gelatin in the fat phantom mixture in order to acquire the dielectric property range close to the one in the reported tissue measurement. For most tests, however, the fresh fat phantom presented here is acceptable since it still leads to a significant contrast between tumour and fat. Here, we note that the contrast between tumour and fresh fat phantoms is 4:1 for relative permittivity and 5.5:1 for conductivity. The dielectric contrast between tumour and old fat is, correspondingly, 4.5:1 and 6:1. Fig. 3 Relative permittivity vs. frequency for freshly mixed and 6-month old fat-mimicking phantoms.

Figure 4 shows the average relative permittivity and conductivity of the fresh fat with the standard deviation of the measurements shown as error bars. Similarly, Figures 5, 6 and 7 show the variations in relative permittivity and conductivity versus frequency for the gland, tumour and skin-mimicking phantoms, respectively. It is noted that the variation in measurements does not have a negative effect on the overall results as the contrast between the tumour and fat would vary only slightly even with worst-case deviation. Fig. 4 Relative permittivity and conductivity vs. frequency for fresh fatmimicking Fig. 7 Relative permittivity and conductivity vs. frequency for skinmimicking C. Step 3: Building a Complete Breast Model Fig. 5 Relative permittivity and conductivity vs. frequency for glandmimicking Fig. 6 Relative permittivity and conductivity vs. frequency for tumourmimicking The four materials mimicking the breast tissues need to be merged into a model that, geometrically, matches the breast anatomy as realistically as possible. First, the 2-mm thick skin phantom is made by pouring the liquid skin mixture between two hemispherical bowls of strategically chosen radii. After the liquid mixture has had sufficient time to harden, the bowls can be separated. The practical challenges with this technique are: 1) difficulty in avoiding the formation of air pockets within the liquid when it is compressed between the two bowls and 2) the separation of the two containers can cause the skin phantom to tear. However, we found that compressing the liquid phantom very slowly between the two bowls avoids creating air pockets. As well, by pulling the bowls apart by applying equal pressure on all sides of the bowl it easy to separate them without tearing. Once the skin has set, the fat phantom can be poured directly into it. If the breast model is to include one or more tumours, a batch of tumour mixture should be made ahead of time and allowed to solidify. Thereafter, the tumour phantom is easily cut into the desired shape and size. The fatmimicking phantom can be poured into the hemispherical skin phantom in successive layers, enabling one or more tumour models to be placed at any location within the fat.

Incorporating the glands into the model is more challenging. Glands have approximately conical shape, with the narrower part of the cone being closer to the skin surface [8], [9]. To structurally mimic the glands within the phantoms, cylindrical glass rods were anchored in the fat before it solidified. Once the fat phantom hardens, the rods can be removed and the remaining hole can be carved out into a conical shape. At this point, the liquid gland mixture can be poured directly into the conical hole. This technique permits multiple gland models, and, further, for tumours to be placed within the glands. Figure 8 shows an example of a resulting breast phantom with skin, fat, glands and tumour. The phantom has a radius of 6.5 cm, which is a realistic size. Two different gland structures are included. The smaller structure is all gland tissue, while the larger one contains a 1.5-cm tumour. ACKNOWLEDGMENT The authors would like to thank Mr. Jules Gauthier of the Centre PolyGrames at the Université de Montréal for his laboratory help with the tissue measurements. We are grateful to Mr. Donald Pavlasek of the Electrical & Computer Engineering Department at McGill University for the construction of the moulds needed for the phantoms. Last but not least, the authors would like to acknowledge the support of the following funding agencies: Natural Sciences and Engineering Research Council of Canada (NSERC), Fonds de recherche sur la nature et les technologies (FQRNT), and Partnerships for Research on Microelectronics, Photonics and Telecommunications (PROMPT). Fig. 8 Photograph of the complete breast phantom including skin, fat, glands and tumour. III. CONCLUSIONS & FUTURE WORK In this paper, we have presented a method for constructing tissue phantoms with dielectric properties similar to those recently reported for breast tissue. Our procedure makes significant advances in physiological representation compared to the simplified models used to date, allowing for the inclusion of multiple conical-shaped glands of varying size and the placement of tumours within these glands. Four phantoms were made with characteristics that approximate the permittivity and conductivity of actual breast fat, gland, skin and tumour tissue. Measurements were performed on each phantom type to verify that the conductivity and permittivity are appropriate. Finally, a complete breast model was assembled which incorporated the four different tissue phantoms in a distribution pattern representative of the human breast anatomy. For parametric studies, the contrast between healthy and malignant tissues can be controlled by changing the concentration of oil in the tumour phantom. This overall phantom offers a comprehensive breast model that is suitable for validation experiments on microwave breast cancer detection. REFERENCES [1] E.C. Fear, Microwave Imaging of the Breast, Technology in Cancer Research and Treatment, Vol. 4, no.1, pp. 69-82, 2005. [2] M. Lazebnik, L. McCartney, D. Popovic, C. Watkins, M. Lindstrom, J. Harter, S. Sewall, A. Magliocco, J. Booske, M. Okoniewski and S. Hagness, A large-scale study of the ultrawideband microwave dielectric properties of normal breast tissue obtained from reduction surgeries, Phys. Med. Biol., Vol. 52, pp. 2637-2656, 2007. [3] M. Lazebnik, D. Popovic, L. McCartney, C. Watkins, M. Lindstrom, J. Harter, S. Sewall, T. Ogilvie, A. Magliocco, T. Breslin, W. Temple, D. Mew, J. Booske, M. Okoniewski, and S. Hagness, A large-scale study of the ultrawideband microwave dielectric properties of normal, benign and malignant breast tissues obtained from cancer surgeries, Phys. Med. Biol., Vol. 52, pp. 6093-6115, 2007. [4] M. Lazebnik, E. Madsen, G. Frank and S. Hagness, Tissue-mimicking phantom materials for narrowband and ultrawideband microwave applications, Phys. Med. Biol., Vol. 50, pp. 4245-4258, 2005. [5] E. Zastrow, S. Davis, M. Lazebnik, F. Kelcz, and S. Hagness, Development of Anatomically Realistic Numerical Breast Phantoms With Accurate Dielectric Properties for Modeling Microwave Interactions With the Human Breast, IEEE Transactions on Biomedical Engineering, Vol. 55, no.12, pp. 2792-2800, 2008. [6] S. Gabriel, R.W. Lau and C. Gabriel, The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz, Phys. Med. Biol., Vol. 41, pp. 2251-2269, 1996. [7] R. J. Halter, T. Zhou, P.M. Meaney, A. Hartov, R.J. Barth Jr, K.M. Rosenkranz, W.A. Wells, C.A. Kogel, A. Borsic, E.J. Rizzo and K.D. Paulsen, The correlation of in vivo and ex vivo tissue dielectric properties to validate electromagnetic breast imaging: initial clinical experience, Physiol. Meas., Vol. 30, pp. S121-S136, 2009. [8] (2009) Canadian Cancer Society, What is breast cancer? [Online]. Available: http://www.cancer.ca/ [9] A.C. Guyton, J.E. Hall, Textbook of Medical Physiology, 11 th ed., Saunders, 2005.