Carotid and radial pulse feature analysis with EMFi sensor

Similar documents
Comparison of local pulse wave velocity values acquired with EMFi sensor

Medical Engineering & Physics

Clinical application of Arterial stiffness. pulse wave analysis pulse wave velocity

The Cardiac Cycle Clive M. Baumgarten, Ph.D.

Which method is better to measure arterial stiffness; augmentation index, pulse wave velocity, carotid distensibility? 전북의대내과 김원호

Lab #3: Electrocardiogram (ECG / EKG)

37 1 The Circulatory System

A Comparative Study of Methods of Measurement of Peripheral Pulse Waveform

Cardiovascular Diseases Detecting via Pulse Analysis

Cardiovascular Physiology

Impedance Cardiography (ICG) Method, Technology and Validity

Cardiovascular System

CRITICAL THINKING QUESTIONS AND ANSWERS AND CYCLE 2 LAB EXAM TEMPLATE. There are two main mechanisms that work in conjunction to return the blood

Cardiovascular System Notes: Heart Disease & Disorders

Mechanisms of heart failure with normal EF Arterial stiffness and ventricular-arterial coupling. What is the pathophysiology at presentation?

Arterial Line Waveform Interpretation

Lab #3: Electrocardiogram (ECG / EKG)

Chapter 2 The Human Cardiovascular System

IB TOPIC 6.2 THE BLOOD SYSTEM

Seven years follow up of ballistocardiogaphy

Study and Design of a Shannon-Energy-Envelope based Phonocardiogram Peak Spacing Analysis for Estimating Arrhythmic Heart-Beat

IB TOPIC 6.2 THE BLOOD SYSTEM

Topic 6: Human Physiology

Cardiovascular System Notes: Physiology of the Heart

Arterial Age and Shift Work

2. The heart sounds are produced by a summed series of mechanical events, as follows:

Blood Pressure Laboratory

Chapter 08. Health Screening and Risk Classification

Heart Rate, Blood Pressure, and Exercise. Evaluation copy

Coronary artery disease (CAD) risk factors

Blood Vessels and Our Pulse

Health Science 20 Circulatory System Notes

ANALYSIS OF BALLISTOCARDIOGRAM WITH MULTIWAVELETS IN EVALUATION OF CARDIAC FITNESS

Pearson's Comprehensive Medical Assisting Administrative and Clinical Competencies

Biomedical Instrumentation E. Blood Pressure

The Heart and Cardiovascular System

CARDIAC CYCLE CONTENTS. Divisions of cardiac cycle 11/13/13. Definition. Badri Paudel GMC

Blood Pressure Determination Using Analysis of Biosignals

Cardiac Cycle MCQ. Professor of Cardiovascular Physiology. Cairo University 2007

LATEST TRENDS on SYSTEMS (Volume I) New Seismocardiographic Measuring System with Separate QRS Detection. M. Stork 1, Z. Trefny 2

Cardiovascular System

Heart Rate and Blood Pressure as Vital Signs

Cardiovascular System

The Circulatory System (p )

Estimation of Pulse Arrival Time Using Impedance Plethysmogram from Body Composition Scales

Lab 16. The Cardiovascular System Heart and Blood Vessels. Laboratory Objectives

PHYSIOLOGY MeQ'S (Morgan) All the following statements related to blood volume are correct except for: 5 A. Blood volume is about 5 litres. B.

Test-Retest Reproducibility of the Wideband External Pulse Device

The reproducibility of central aortic blood pressure measurements in healthy subjects using applanation tonometry and sphygmocardiography

Preprocessing PPG and ECG Signals to Estimate Blood Pressure Based on Noninvasive Wearable Device

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time

MR Advance Techniques. Cardiac Imaging. Class IV

IABP Timing & Fidelity. Pocket Reference Guide

Chp. 5 The cardiovascular system. What are the function of the cardiovascular system? Arteries and arterioles:

Central pressures and prediction of cardiovascular events in erectile dysfunction patients

Acute Effect of Resistance Exercise on Vascular and Cognitive Function

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS

HTA ET DIALYSE DR ALAIN GUERIN

IS PVR THE RIGHT METRIC FOR RV AFTERLOAD?

TODAY S TOPIC Blood Pressure & Pulse Wave Measurement Combined in One Procedure Re-classification of Risk Patients

Chapter 9, Part 2. Cardiocirculatory Adjustments to Exercise

Introduction of diverse devices to measure arterial stiffness

End of chapter exercises

Anatomy Review: The Heart Graphics are used with permission of A.D.A.M. Software, Inc. and Benjamin/Cummings Publishing Co.

Heart Abnormality Detection Technique using PPG Signal

2.02 Understand the functions and disorders of the circulatory system

Chapter 16: Circulation

The Heart. Happy Friday! #takeoutyournotes #testnotgradedyet

Dynamic measures of arterial stiffness in a rodent model

Local Pulse Wave Velocity Estimation in the Carotids Using Dynamic MR Sequences

Keun-Ho Kim and Sang-Suk Lee* Department of Oriental Biomedical Engineering, Sangji University, Wonju , Korea

Chapter 14. The Cardiovascular System

PHONOCARDIOGRAPHY (PCG)

d) Cardiovascular System Higher Human Biology

Section 5.1 The heart and heart disease

What is the mechanism of the audible carotid bruit? How does one calculate the velocity of blood flow?

BIOL 219 Spring Chapters 14&15 Cardiovascular System

Blood Pressure Estimation Using Photoplethysmography (PPG)

Special Lecture 10/28/2012

Objectives of the Heart

The ancient Babylonians, Egyptians, Indians and Chinese believed the heart was the centre of thinking and emotions

Results of geometric ballistocardiography

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

Analysis of Human Cardiovascular System using Equivalent Electronic System

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work

PART I: HEART ANATOMY

Central hemodynamics and prediction of cardiovascular events in patients with erectile dysfunction

Circulatory system of mammals

Impedance Cardiography (ICG) Application of ICG for Hypertension Management

Circulatory System Notes

Ultimate Solution. for Healthy Life. HANBYUL Overview

The Heart 1 of 34 Boardworks Ltd 2012

Mr. Epithelium s Anatomy and Physiology Test SSSS

Reducing the risk of CVD

Performance Enhancement. Cardiovascular/Respiratory Systems and Athletic Performance

PULSE WAVE VARIABILITY WITHIN TWO SHORT-TERM MEASUREMENTS. David Korpas, Jan Halek

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Physical Fitness Biology Concepts of Biology 13.1

Major Function of the Cardiovascular System. Transportation. Structures of the Cardiovascular System. Heart - muscular pump

Transcription:

Carotid and radial pulse feature analysis with EMFi sensor Jarmo Alametsä 1, Ari Palomäki 2, Jari Viik 1 1 Tampere University of Technology, Tampere, Finland, 2 Kanta Häme Central Hospital, Hämeenlinna, Finland Jarmo Alametsä, Tampere University of Technology, Tampere, Finland, FINLAND. Email: jarmo.alametsa@tut.fi. Abstract The purpose of this work is to show the potential of Electromechanical Film (EMFi) sensor in vascular elasticity studies when pulse wave features from carotid pulse (CP) and radial pulse are studied. ECG, seat ballistocardiogram (BCG) and pulse signals from the limbs and CP were recorded from 48 working aged men in sitting position. Duration and amplitudes of the signal components from the ballistic seat signal, CP and radial pulse according to R wave of the ECG were studied. Several calculated parameters used to obtain vasculature stiffness information were compared with Bland Altman (BA) plots and with Pearson correlation in order to study, whether CP and radial pulse give consistent information about vascular elasticity. Results from the BA plots and Pearson correlation show that elastic information obtained from the CP and radial pulse signals clearly differ from each other. The elasticity changes along the arterial tree seen in local pulse signals reflect also to the form of the seat BCG signal. Keywords: EMFi, BCG, carotid and radial pulse 14.5.2012 FinJeHeW 2012;4(2) 80

Introduction Ballistocardiography (BCG) is a non invasive method for cardiac and respiratory evaluation and it reflects closely the strength of myocardial contraction revealing the condition of the heart [1]. When the heart pumps blood to the pulmonary arteries and ascending aorta, through aortic arch to the peripheral circulation, recoil of opposite direction is applied to the body and its force and direction is changing according to the cardiac cycle. The carotid pulse (CP) is a pressure signal recorded from carotid artery when it passes near the surface of the body in the neck. It indicates the variations in arterial blood pressure and volume with each heart beat. As the recording place is located very near the heart, the CP signal resembles the morphology of the pressure signal at the root of aorta [2]. The CP rises abruptly with the ejection of blood from left ventricle to ascending aorta reaching a peak called percussion wave (P). The following secondary wave is called as a tidal wave (T), caused by a reflected pulse returning from the lower body travelling back to the heart. Dicrotic notch (D) is caused by a closure of aortic valve following a brief transient increase in the aortic pressure, the dicrotic wave (DW) [2] (Fig. 1). The carotid pulse supplements the BCG data by giving the onset of the ejection phase and can give information about the coordination of the cardiac activity. In this paper a Mobile Physiological Signal Measurement Station has been used as a device, which enables the recording of seat BCG and local pulse signals from different locations of the body [3] with EMFi sensors. Pulse wave and contour analysis are used to determine arterial stiffness and vascular distension. The main goal of this study is to show, that EMFi sensor can produce elastic information from the vasculature. Another objective is to study, if consistent elastic information is obtained from a CP and wrist pulse signal measured with EMFi sensor. As arterial properties have an influence on the amplification of the pulse wave, differences in elasticity are expected to found even between CP and wrist pulse signals. As arterial tree have different elasticity values, changes in these have an influence to the BCG signal also. Methods The EMFi [4] sensor is basically a thin biaxially oriented plastic film coated with electrically conductive layers which are permanently polarized. Changes in the pressure acting on the film generate a charge on its electrically conductive surfaces and this charge can be measured as a current or a voltage signal. It can convert mechanical energy to electrical and vice versa. Thus the EMFi acts as a sensitive movement sensor suitable for BCG recordings. Signals from EMFi sensors were recorded with the Mobile Physiological Signal Measurement Station [3] into a notebook computer with a data acquisition card (Daqcard 6036E) and the recordings were made into ASCII format. In the Mobile Physiological Signal Measurement Station an active Butterworth 8. degree low pass filter was used, where the cut off frequency was 256 Hz. In chair recordings the EMFi sensor (42 cm x 36 cm) was beneath the measured person. Seven EMFi sensor strips (15 cm x 2 cm) were attached to the limbs and on the neck near carotid artery. ECG trace was used as reference in detecting features from the pressure related signals from EMFi sensors. The form of the pulse wave is defined by the properties of the heart beat, blood pressure (BP) and the properties of the vessel wall. Age and BP affect to the aortic size in adults and also to the form of the pulse wave. The crosssectional area of the aorta has been shown to double between the ages of 20 to 80 having a considerable effect to the aortic elastic properties in different locations of the aorta [5]. 14.5.2012 FinJeHeW 2012;4(2) 81

In pulse contour evaluation the carotid pulse obtained from healthy persons and from patients having arterial atherosclerosis show recognizable, marked contour differences being compatible with the clinical condition of the subjects. These contours change along with the clinical signs of the disease and show characteristic abnormal features. Atherosclerosis contributes the mechanical properties of the arterial wall which becomes rigid and loses its elasticity [6]. Elasticity index e (amplitude ratio: a/b; reflecting arterial elasticity), and dicrotic index d (amplitude ratio c/a in %; reflecting the amount of peripheral resistance) [6] are presented in Figure 2. Measurements In this study we present pulse feature analysis obtained from CP, radial pulse and seat BCG. The recordings were made from 48 middle aged men (age 41 65 years) in a sitting position measured with EMFi sensors. The medical history of the measured persons relating to smoking, hypertension, atherosclerotic disease and diabetes was known (Table 1). All the measurements lasted about 3 min and the used sampling frequency was 500 Hz. Just before the measurements the blood pressure and the pulse were measured with Omron M5 I blood pressure monitor device. Table 1. Population characteristics with the mean value ± standard deviation and the number of patients. Variable All Number of patients (n) 48 Age (years) 58±5.3 Height (cm) 177.8±6.2 Weight (kg) W0 82.6±12.9 Mean systolic blood pressure (mmhg) a) W0 136±17 Mean diastolic blood pressure (mmhg) a) W0 86±9 Mean heart rate (bpm) b) W0 61±9 Smoking status (present/previous/never) (n) 8/24/16 Hypertension (n) 20 Diabetes mellitus (n) 8 Atherosclerotic disease (CHD, PAD, CVD) (n) 28 W, week; CHD, coronary heart disease; PAD, peripheral artery disease; CVD, cerebrovascular disease. a) Values from the Omron M5 I blood pressure monitor device. b) Mean values from the ECG s R R interval (bpm=beats/min). The BCG recordings were carried out in the morning between 7 and 10.30. Eating and drinking were not allowed 12 hours before the measurement. Informed written consent was obtained from all participants. The study was approved by the ethics committee of Kanta Häme Hospital district. The measurements were made in Linnan Klinikka, Hämeenlinna, Finland. Signals were first band pass filtered (0.5 30 Hz FIR, 700 taps, time delay corrected), down sampled into 100 Hz and the analysis was done with 0.5s window length. The index of the R point was detected first by differentiating (2 points), squaring and integrating (5 points) and by taking the maximum from the ECG signal. The I slope from the BCG was detected by local minimum method and then the J slope was detected by local maximum using the index of the I point as a starting point. Other slopes in CP and wrist pulse signal were detected at the same way. Matlab software was used in previous calculations. 14.5.2012 FinJeHeW 2012;4(2) 82

The R wave of ECG was used as a reference in detecting the slopes from BCG, carotid and radial pulse signals. The P point in the neck CP signal was chosen for the measurement for the reason that the end point of ejection is not seen so clearly as an exact end point in the CP signal, but it is overlapped by a reflected pulse returning from the lower body. Sitting Radial pulse Ao Bo To Do T DWo Carotid pulse P D DW O H J L N Seat BCG K M ECG R I T P S Q 8.5 9 9.5 10 10.5 11 11.5 12 Time (s) Figure 1. Signals recorded in sitting position (Case 48). Pulse waves from the right side of the neck (CP) and from the right wrist (radial) are labelled as mentioned in this study. As seat BCG signal corresponds to the total recoil of the heart mediated via ejected blood burst from the heart seen as pulse, the labels of the main components of the seat BCG signal are also attached. Four channels of seven recorded ones are presented [3]. In order to study the contraction of the left ventricle, temporal values, like ejection time (TO D) from the carotid artery signal and (TAo Do) from the radial pulse signal were extracted (Fig. 1). As the origin of the pulse waves and seat BCG trace is the movement of blood inside large arteries due to the pumping action of the heart seen also as recoil traces in seat BCG signal, the ejection time TH K from seat BCG signal was included into the study. THI is the time of the rapid ejection phase of the systole. THK is the period of the blood expulsion phase from the ventricles. TRH is the measure of the isometric tension phase of the ventricles. TRK is the duration of the mechanical systole of the heart [1]. Bland Altman plot (BA) and Pearson correlation coefficient were determined to estimate similarity of the measurements of CP and radial pulse as well as CP and seat BCG. All statistical analyses were performed by using the SPSS 17.0 program (SPSS Inc.). 14.5.2012 FinJeHeW 2012;4(2) 83

Results Detection of the signal components from BCG, CP and radial pulse signals with minimum and maximum method by using a moving detecting window was successful. Calculated elasticity values from CP and radial pulse signal showed distinctive differences judged by the BA plots and very weak correlation between CP and radial pulse (Figs. 3 5). No close relation exists between ejection time determined from CP and seat BCG signal (Fig. 6). Quite modest Pearson correlation values indicate differences between signals. Figure 2. Indexes e and d can be defined from aortic and radial pulse waveforms. They depend on the relative amplitude and reflected (shaded in the picture) pressure waves summing to produce the overall waveform. Redrawn from [7]; CP and wrist pulse labelled as is in this study. Figure 3. Bland Altman plot from the ejection time calculated from the CP and radial pulse signal. T O D and T Ao Do are the corresponding ejection times in seconds (Figs. 1 2). Pearson correlation coefficient 0.312. 14.5.2012 FinJeHeW 2012;4(2) 84

Figure 4. BA plot from the index e (Figs. 1 2) from the CP and correspondingly from the radial pulse (n=43). In [6] the index e value 1.27 was considered as healthy and with value 0.58 having an arterial atherosclerosis. Pearson correlation coefficient 0.329. Figure 5. BA plot from the index d from the CP and correspondingly from the radial pulse (n=43). In [6] the index d value 0.59 was considered as healthy and with value 1.23 having an arterial atherosclerosis. Pearson correlation coefficient 0.30. 14.5.2012 FinJeHeW 2012;4(2) 85

Figure 6. BA plot from the ejection time in seconds between the seat BCG and CP. Pearson correlation coefficient 0,041. Discussion Our main result is that carotid pulse features clearly differ from that of radial pulse and seat BCG. It is known, that the contour of the pulse waveform and systolic BP differ between central and peripheral arteries [7]. When measured from two different locations of the arterial tree, the form of the pulse wave is influenced by the local reflections, BP and measurement position. The local elastic information from radial and carotid pulse differ from each other being also different between different subjects. When measuring local arterial stiffness, it is important to describe measurements as done, because the measurement location in proportion to arterial tree has its own effect to the measurement results. We have earlier shown that signal components of BCG are repeatable in consecutive recordings as well as reproducible in longer time recording intervals [8]. Spreading of the obtained values in BA plots from the CP radial pulse ejection time and CP between seat BCG may also indicate possible inadequate foot point detection of the pulse wave. In this study, the foot of the pulse wave was detected by local minimum method. Minimum method in foot point detection may fail, as some fluctuation in the signal may happen before the main ejection of the heart placing the detected point before the actual start of ejection. Therefore, the intersecting tangent algorithm would be worth of study. Another source of error may be the inadequate coupling of the EMFi strip sensor to the underlying skin producing small amplitude signal. The time domain properties of blood vessels change when blood pressure rises and this has an influence on the ballistocardiographic waveforms and to the elasticity of the veins. Usually, ballistocardiographic amplitude decreases while arterial pressure increases [1]. Some small differences in BCG waveforms between different people can be explained different anatomical causes; as the blood pressure values differ with different people, so does 14.5.2012 FinJeHeW 2012;4(2) 86

the contraction ability of the heart and the elasticity of the veins. In Figure 6 the differences in ejection time between CP and seat BCG can be explained differences in the elasticity of the arterial tree seen as a response of the ejection of the heart. Differences in the density of the elastic fibers and thus in the elastic modulus between different parts of the aorta has been discovered: the highest elastic modulus was found from the thoracic section of the aorta having also the highest density of elastin fibers [9]. The augmentation index (AIx) has been used widely in describing arterial and radial elasticity. For carotid pulse it is the amplitude difference between a and b divided by amplitude b. For radial pulse AIx is defined as b/a (Fig. 2). Aortic AIx has been used to evaluate central systolic BP by the means of radial to aortic transfer function. Central systolic BP differs from brachial BP due to peripheral amplification [7]. As AIx depends on higher frequency information of the pulse wave, but systolic BP not, this approach in the evaluation of central systolic BP has been questioned [7]. AIx is influenced by the arrival time of the pulse wave reflection, thus depending on magnitude and timing of the reflected waves [10,11] and heart rate [12] and it is proposed, that the properties of peripheral muscular arteries and arterioles affect to AIx by changing the amplitude of the reflected wave [10]. Since the validity of AIx as a measurement of arterial elasticity has been questioned [7,10,13,14] and as AIx definition for CP and radial pulse differed, we did not calculate this determination. As Reflection Index (RI) is defined as b/a being inverse with index e (Fig. 2), it was omitted from this study. This study shows that elastic information is obtainable from both carotid pulse and radial pulse signals with EMFi sensors. However, the information obtained from carotid pulse clearly differs from that from radial pulse and seat BCG. Hence in scientific publications (and even in their titles) the methodology of the measurement of local arterial elasticity should be conclusively described. Acknowledgments This study was financially supported by the Finnish Cultural Foundation, the Wihuri Foundation as well as by the Ministry of Social Affairs and Health in Finland through grants from Kanta Häme Central Hospital. We thank Markku Järvinen MD, Kalevi Oksanen MD and the personnel of Linnan Klinikka for help in coordination. We also thank measured persons for participating in this study. References [1] Weissler AM. The Ballistocardiographic waveforms, in Noninvasive Cardiology Monographs. NY: Grune & Stratton Inc.; 1974. pp. 55 125. [2] Rangayyan RM. A Case study Approach to Solve Problems in Biomedical Signal Analysis. The IEEE Press, Piscataway, NJ; 2000. Version: March 13, p. 31 [3] Alametsä J, Värri, A, Viik J, Hyttinen J, Palomäki A. Ballistocardiogaphic studies with acceleration and electromechanical film sensors. Medical Engineering & Physics. 2009(31):1154 1165. [4] Kirjavainen K. Electromechanical film and procedure for manufacturing same. U.S. Patent no. 4654546. Manufacturer of EMFi: Emfitech Ltd, Vaajakoski, Finland; 1987. http://www.emfit.com [5] Towfiq B, et al. Effect of age and blood pressure on aortic size and stroke distance. Br Heart J 1986;55:560 568. 14.5.2012 FinJeHeW 2012;4(2) 87

[6] Chlebus H, Early Diagnosis of Arterial Atherosclerosis by Means of Resonance Electrosphygmography, in Bibliotheca Cardiologica 30: Hemodynamic Stress and Relief of the Heart, G. Juznic, Editor. 1973, Karger S. Basel. p. 9 25. [7] Millasseau SC, et al. Pressure Wave Reflection Assessed From the Peripheral Pulse Is a Transfer Function Necessary? Hypertension 2003;41: 1016 1020. [8] Alametsä J, Palomäki A, Viik J. Short and Longer Term Repeatability of Ballistocardiography in Sitting Position with EMFi sensor, Medical & Biological Engineering & Computing 2011;49:881 889. [9] Moriwaki T, et al. Variations in local elastic modulus along the length of the aorta as observed by use of a scanning haptic microscope (SHM). Journal of Artificial Organs, 2011;14(4):276 283. [10] Shimizu M, Kazuomi, K. Role of the augmentation index in hypertension. Ther Adv Cardiovasc Dis. 2008;2:25:25 35. [11] Segers P, et al. Amplification of the Pressure Pulse in the Upper Limb in Healthy, Middle Aged Men and Women. Hypertension 2009;54:414 420. [12] Kohler M, et al. Endothelial Function and Arterial Stiffness in Minimally Symptomatic Obstructive Sleep Apnea. Am J Respir Crit Care Med. 2008;178:984 988. [13] Laurent S, et al. Expert consensus document on arterial stiffness: methodological issues and clinical applications. European Heart Journal, 2006 27, pp. 2588 2605. [14] Namasivayam M, et al. Aortic Augmentation Index and Aging: Mathematical Resolution of a Physiological Dilemma? Hypertension 2010, 56:e9 e10. 14.5.2012 FinJeHeW 2012;4(2) 88