INDIRECT ANKLE, BRACHIAL AND DIRECT INTRA -ARTERIAL BLOOD PRESSURE MEASUREMENTS

Similar documents
Assessment of techniques for measurement of blood pressure in infants and children

An Indian Journal FULL PAPER ABSTRACT KEYWORDS. Trade Science Inc.

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

AFFORDABLE TECHNOLOGY

Lecture #3 - Blood Pressure Recording Procedure. Equipment

2. Measure a subject's blood pressure and heart rate both at rest and during exercise.

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine

Protocol 4: Measuring Blood Pressure

Section 03: Pre Exercise Evaluations and Risk Factor Assessment

Comparison of blood pressure measured at the arm, ankle and calf

Webinar Logistics. Now What Do I Need To Do? Blood Pressure in Kids January NC Department of Health and Human Services Division of Public Health

Chapter 12 - Vital_Signs_and_Monitoring_Devices

Blood Pressure Competency Training

PART I: HEART ANATOMY

External Oscillatory Blood Pressure - EOBPTM

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS

Clinical Research Coordinator Skills Program

Ambulatory Services Orientation & Skill Review Skill Performance Checklist: Vital Signs. Name: Date:

AN INTRODUCTION TO DOPPLER. Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust.

Data Collection Worksheet

SECTION 16: BLOOD PRESSURE MEASUREMENT

A GUIDE TO OUR NIBP TECHNOLOGY

Intra-Arterial Blood Pressure Monitoring

PLANK 1 Direct Care Staff Trained in Accurate BP Measurement

Managing Blood Pressure: It Takes a Team

Competency Title: Caring for a patient with an arterial line

Blood Pressure Measurement by Pulse Oxymetric Method and Comparison with Conventional Technique

Blood Pressure and Exercise Lab

Vital Signs and Oxygen Administration

Measuring body temperature, blood pressure, pulse, respiratory rate and oxygen saturation

Biomedical Instrumentation E. Blood Pressure

1 Ankle Brachial Pressure Index

Human Cardiovascular Physiology: Blood Pressure and Pulse Determinations

Vital Signs. Temperature. Pulse. Respirations. Blood Pressure. O2 Saturation

An Automated System for Blood Pressure Determination During Exercise

to use the proper technique often lead to the wrong diagnosis, which may indirect measurement ofblood pressure. 240k

GUIDELINES FOR THE MEASUREMENT OF ANKLE BRACHIAL PRESSURE INDEX USING DOPPLER ULTRASOUND

Bio& 242, Unit 3/ Lab 4 Blood Vessels, Lymphatic System and Blood Pressure G. Blevins/ G. Brady Summer 2009

SMALL GROUP SESSION 18A January 17th or January 19th. Groups 1-12: VS and Chest Exam and Harvey Stethophone Session

science-u.org What affects blood Blood Pressure Directions You Will Need BEST FOR GRADES 5-8 ESTIMATED TIME Minutes

High Blood Pressure Blood Pressure Monitoring at Home -- F...

How To Measure Vital Signs

Chapter 29 1/8/2018. Vital Signs. Measuring and Reporting Vital Signs. Key Terms

MEASUREMENT OF BLOOD PRESSURE. Department of Clinical Physiology, The Hospital for Sick Children, Great Ormond Street, London, W.C.

Arteriosonde, and mercury sphygmomanometer in

Crucial Signs כל הזכויות שמורות למד"א מרחב ירושלים

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

LAB: Blood Pressure Measurable Indicator of the Health of the Circulatory System!

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure

WISEWOMAN UPDATE. December The Importance of Accurate Blood Pressure Measurements

in anaesthetised companion animals

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

ECG, Blood Pressure, and Exercise Lab

Physiology of the Circulatory System modified from

2. Need for serial arterial blood gas determinations. 2. Anticipation of the initiation of thrombolytic therapy

5.2. MEASUREMENT OF BLOOD PRESSURE

A DEVICE FOR THE INDIRECT RECORDING OF BLOOD PRESSURE

Welcome to the S410. Blood Pressure System. Instruction Manual

Blood Pressure Protocol Training

AMERICAN ASSOCIATION OF SLEEP TECHNOLOGISTS TECHNICAL GUIDELINE FOR PATIENT ASSESSMENT AND VITAL SIGNS MEASUREMENT AND DOCUMENTATION

Therefore MAP=CO x TPR = HR x SV x TPR

Blood Pressure Laboratory

Lab #10 Physiology of the Circulatory System

Vital Signs Monitoring Clinical Skills

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Vital Signs. Provide information about body function Include: Changes may be the first sign of disease Accuracy is imperative!

1 Non-invasive measurement of arterial pressure

VITAL SIGNS AND PHYSICAL EXAMINATION OF THE CHEST Eve Bargmann, MD 9/22/03

Chapter 15: Measuring Height, Weight, and Vital Signs. Copyright 2012 Wolters Kluwer Health Lippincott Williams & Wilkins

SMALL GROUP SESSION 13 December 1 st or December 3 rd

ASSESSING THE VASCULAR STATUS OF THE FEET FOR PATIENTS WITH DIABETES

VITAL SIGNS. Academic Success Center

Topic: Baseline Vitals and Sample History Company Drill

Physiology lecture 15 Hemodynamic

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018

LAB 9: Metabolic Rates

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Importance Of Taking Manual Blood Pressure In Both Arms

BP-600 Noninvasive Blood Pressure Sensor

BLOOD PRESSURE ASSESSMENT

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License

CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL

Vital Signs Monitor Series (VSM) Outline Train the Trainer

Critical Care Nursing Program August to November, 2015 Full-time Lesson A13 Pumping and Perfusion III Basic Hemodynamic Monitoring

Pancreas: digestive enzymes Liver: bile-helps to digest fat, stored in gallblader; breaks down toxins, makes some parts of blood

Recommendations for Blood Pressure Measurement in Humans and Experimental Ani...

Lesson 16 BLOOD PRESSURE

Non-invasive examination

Chapter 21. Assisting With Assessment. Elsevier items and derived items 2014, 2010 by Mosby, an imprint of Elsevier Inc. All rights reserved.

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE

HAEMODYNAMIC IN THE CATH LABORATORY INTRO TO BASICS

Cardiovascular Physiology Laboratory Clive M. Baumgarten, Ph.D.

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

Transcription:

VOLUME 26, NO. 3 JUNE 1985 INDIRECT ANKLE, BRACHIAL AND DIRECT INTRA -ARTERIAL BLOOD PRESSURE MEASUREMENTS C K Koay M K Chin SYNOPSIS The purpose of this study was to compare the accuracy of indirect ankle and brachial blood pressure measurements in a group of neurosurgical patients. Intraradial cannulation was used for direct arterial pressure measurement. The correlations of the systolic and diastolic pressure between the indirect ankle and direct intra -arterial blood pressure was good. The coefficient of correlation for systolic and diastolic pressure were 0.95 and 0.81 respectively. The correlation between the indirect brachial blood pressure and direct infra -arterial blood pressure was also good. The coefficient of correlation for systolic and diastolic pressures were 0.94 and 0.87 respectively. It was concluded that the indirect ankle blood pressure measurement was a good method of assessing the blood pressures and should be used when the upper limb measurements were not possible. INTRODUCTION Department of Anaesthesia Tan Took Seng Hospital Moulmein Road Singapore 1130 C K Koay, MBBS, M Med (Arises), FFARACS Registrar M K Chin, MB Bch, BAO (UNI), FFARACS Read and Consultant In clinical practice, the commonest method of measuring blood pressure is by means of auscultation using a sphynomanometer at the upper limbs. However, there are occasions when this is not possible. Such situations are encountered in patients with bilateral upper limb injuries or bilateral extensive burns of the upper limbs. It is usually difficult to take the brachial blood pressure in surgical patients when the operations are carried out around the head, neck and shoulder regions. Intra -arterial blood pressure monitoring is invasive, technically more difficult and requires more sophisticated equipments. When it is not possible to use the upper limbs or when direct arterial blood pressure monitoring is not warranted, the lower limbs become the alternative site for blood pressure assessment. The purpose of the study is to evaluate the accuracy of ankle blood pressure measurement in relation to indirect brachial blood pressure and direct intra -arterial blood pressure monitoring. 283

SINGAPORE MEDICAL JOURNAL MATERIAL AND METHODS This study was conducted over a 2 month period in the Neurosurgical Theatre of Tan Tock Seng Hospital. 15 consecutive patients given general anaesthesia by one of the authors (Koay) were included. Patients with known coarctation of aorta or vascular disease of the lower limbs were excluded from the study. There were 10 males and 5 females. Their ages ranged from 15 years to 84 years with a mean of 45 years. All the patients had continuous intra -arterial blood pressure monitoring indicated for the neurosurgical procedure. Cannulation of the radial artery was done by means of a gauge 23 or 27 cannula with a 60 inch pressure tubing which was connected to a continuous flush system and a strain gauge transducer. The recording machine used was Physio-Control VSM 1. The machine was zeroed and calibrated. Readings were taken with the transducer at the level of mid-axillary line with the patient in the supine position. The intra - arterial blood pressure was recorded for each brachial and ankle readings. For brachial BP measurement, the standard cuff (ACCOSON 22 cm long and 14 cm wide) was used for all patients. The cuff bladder was applied at the inner aspect of the arm with its lower margin about 2 cm above the antecubital fossa. Similar cuff was used for ankle blood pressure measurement. The cuff bladder was applied on the medial aspect of the lower limb with its lower border just above the medial malleolus. The brachial artery and dorsalis pedis artery were used for the upper limb and lower limb pressure measurement respectively. The systolic pressure was recorded when the korotkoff sound was first heard and the diastolic pressure was taken when the sound ceased to be tapping in quality. RESULTS The results of this study were shown in Tables I and II. The correlation between ankle and intra -arterial blood pressures was shown in Figures 1 and 2. Analysis of the data gave a regression line of Y, _ 1.05x, - 11 (r = 0.95) when X, was the intra -arterial systolic pressure and Y, was the ankle systolic pressure. As for diastolic pressure, the regression equation was Y2 = 0.64x2 + 39 (r = 0.81) where X2 and Y2 were the intra -arterial and ankle diastolic pressures respectively. Similary, the correlation between the direct intra - arterial and indirect brachial pressure measurements was shown in Figures 3 and 4. For the systolic pressure, analysis of the data gave the following regression line of Y3 = 0.85x3 + 13.7 (r = 0.94) and X, and Y3 represent the intra -arterial and bracnlai systolic pressures respectively. The regression equation for the diastolic pressure was Y4 = 0.56x4 + 40 (r = 0.87) where X4 and Y4 were the intra -arterial diastolic and brachial diastolic pressure respectively. The coefficient of correlation is denoted by r in each case. TABLE 1: DIRECT INTRA -ARTERIAL AND INDIRECT ANKLE BLOOD PRESSURE IN STUDIED PATIENTS Patients Intra -arterial blood pressure (mm Hg) Ankle blood pressure (mm Hg) Systolic Diastolic Systolic Diastolic 1 146 86 140 90 2 108 70 120 3 117 70 118 90 4 114 79 90 5 147 95 130 6 150 96 134 7 138 90 130 90 8 182 82 180 90 9 158 80 15B 90 10 167 80 160 85 11 129 64 125 70 12 114 67 85 13 170 102 180 110 14 165 106 170 110 15 105 66 80 284

VOLUME 26, NO. 3 JUNE 1985 TABLE II: DIRECT INTRAARTERIAL AND INDIRECT BRACHIAL BLOOD PRESSURE IN STUDIED PATIENTS Patients Intra -arterial blood pressure (mm Hg) Brachial blood pressure (mm Hg) Systolic Diastolic Systolic Diastolic 1 146 86 130 90 2 120 84 115 90 3 130 86 120 90 4 90 57 105 70 5 119 68 110 80 6 118 70 120 80 7 150 90 140 85 8 192 87 180 80 9 83 40 80 60 10 144 76 140 90 11 120 63 120 80 12 151 78 160 90 13 130 80 110 80 14 123 85 110 90 15 110 60 110 70 FIGURE 1: THE CORRELATION BETWEEN INTRA - ARTERIAL AND INDIRECT ANKLE SYSTOLIC BLOOD PRESSURES. TIE CORPEIATION RENTE1 DIRECT INRA-ARTEPEAL MD INDIRECT 190 MILLE SYSTOLIC BLOM PRESSURES Y 170 160 p ñ 150 tao 130 120 too 130 120 130 lw 160 160 170 DIRECT IIRM-MTEBIAL SYSTOLIC BLOOD PRESSURE RRW01 285

SINGAPORE MEDICAL JOURNAL FIGURE 2: THE CORRELATION BETWEEN INTRA ARTERIAL AND INDIRECT ANKLE DIASTOLIC BLOOD PRESSURES 120. TRE CORRELATION BETWEEN DIRECT INTRA -ARTERIAL AND INDIRECT ANKLE DIASTOLIC BLOOD PRESSURES 110. 90 80-70 50 50 60 70 80 90 110 120 DIRECT INTRA -ARTERIAL DIASTOLIC BLOOD PRESSURE (mmhg) FIGURE 3: THE CORRELATION BETWEEN INTRARTERIAL AND INDIRECT BRACHIAL SYSTOLIC BLOOD PRESSURES. THE CORRELATION BETWEEN DIRECT INTRA -ARTERIA[. AND INDIRECT BRACHIAL SYSTOLIC BLOOD PRESSURES 160 140 O' 120-80 A 80 120 140 160 180 200 DIRECT INTRA -ARTERIAL SYSTOLIC BLOOD PRESSURE (RmBg) 286

VOLUME 26, NO. 3 JUNE 1385 FIGURE 4: THE CORRELATION BETWEEN INTRA - ARTERIAL AND INDIRECT BRACHIAL DIASTOLIC BLOOD PRESSURES. THE CORRELATION BETWEEN DIRECT INTRA -ARTERIAL AND INDIRECT BRACHIAL DIASTOLIC BLOOD PRESSURES 90 80 70 60 50 40 40 50 60 70 80 90 DIRECT INTRA -ARTERIA- DIASTOLIC BLOOD PRESSURE (mmhg/ DISCUSSION It has been shown by many workers that indirect blood pressure measurements are imprecise and that both systolic and diastolic measurements could be in error by up to +20 and -30 mmhg (1, 2). Observers also differ widely in their estimations of blood pressure. Errors in indirect blood pressure measurements can be due to improper application of cuffs (3). The deflated cuff should be applied with the bladder bag over the area of the limb where the artery runs; ie. over the inner aspect of the upper arm for a brachial blood pressure measurement. Inappropriate cuff size can also give rise to inaccurate measurements. Various recommendations have been made regarding the choice of correct length or width of cuff in relation to the limb length or circumference. The American Heart Association (1967) recommends a cuff width 20 per cent greater than the diameter of the limb on which it is to be used. For children, the association recommends that the cuff should cover two-thirds length of the upper arm. However, recent studies suggest that length of the bladder cuff should completely encircle the limb so that the intra -bladder pressure is transmitted with least decay to the underlying compressible artery (4, 5). Park, Kawabori and Guntheroth suggested that the practice of selecting the cuff size based on the length of the arm should be discontinued and that the standard recommended by the American Heart Association Subcommittee for adults should be adopted for children as well (6, 7). The different interpretation of korotkoff sounds also gives rise to variation in indirect blood pressure measurement. In this article, the onset of tapping sounds heard through the stethoscope is taken as systolic pressure and cessation of the tapping sound or muffling of the sounds is the sign for end of diastole (Phave IV). Interpretation of phase IV diastolic point is a common cause of difficulty (8). The same criterias were also used during the indirect blood pressure measurement on the lower limbs. Blood pressure measurement in the thigh is iniconvenient for routine theatre use because a prone position is required and besides that, a much larger cuff is needed. This study has shown that both indirect brachial and ankle blood pressure measurements correlates well with intra -arterial pressure readings. In fact, both the systolic and diastolic of either the brachial or ankle blood pressure have a coefficient of correlation of more than 0.8 with the respective intra -arterial pressure. It might be due to the fact that in most cases. the circumference of the arm is about the same size as that of the ankle. The dorsalis pedis being a more superficial artery than the brachial artery, its pulsation is then more easily felt in normal circumstances. The disadvantage of taking measurement around the ankle is that under certain circumstances, the pulse becomes unrecordable. Dorsalrs pedis artery being a smaller and a more superficial artery than the brachial artery, is more 287

SINGAPORE MEDICAL JOURNAL susceptible to vasocontriction in severe cold, stress or shock. In conclusion, the ankle blood pressure measurement can serve as an alternative in some situations when the routine method of brachial blood pressure monitoring becomes impossible or inconvenient. It is adequate for following changes in blood pressure in patients who are not in shock or criticaly ill in the theatre. ACKNOWLEDGEMENTS We would like to thank the nursing staff for their assistance during this study and Miss Mallika Murugaiah for typing this manuscript. REFERENCES 1. Runciman WB, Rutten AJ, Ilsley AH: An evaluation of blood pressure monitoring. Anaesth Intens Care 1981. 9:314-25. 2. Van Bergen FH, Weatherhead DS, Treloar AE, Dobkin' AB, Buckley JJ: Comparison of indirect and direct methods of measuring arterial blood pressure. Circulation 1954; 10:481-91. 3. Rushmer RF: Cardiovascular Dynamics, 3rd ed. Philadelphia, W B Saunders Co 1970:156. 4. Steinfeld L, Alexander H, Cohen ML: Updating sphygmomanometry. Am J Cardiol 1974; 33:107-10. 5. King GE: Errors in clinical measurement of blood, pressure in obesity. Clin Sci 1967; 32:223-37. 6. Park MK, Kawabori I, Guntheroth WG: Need for an improved standard for blood pressure cuff size. Clin Pediatr 1976; 15:784-6. 7. Kirkendal WM, Burton AC, Epstein FH, Freis ED: Recommendations for human blood pressure determination sphygmomanometers. Circulation 1967: XXXVI:980-8. 8. Rose G: Standardization of observers in blood pressure measurement. Lancet 1965; 1:673-4. 288