A practical guide to using pulmonary artery catheters

Similar documents
Overview of CVADs. Type of device commonly used. Dwell time Flushing requirement Associated complications. lumens

Technique. Technique. Technique. Monitoring 1. Local anesthetic? Aseptic technique Hyper-extend (if radial)

Vascular access in practice: best practice update

This is a provisional PDF comprising this cover note and the manuscript as it was upon acceptance for publication.

Hemodynamic Monitoring and Circulatory Assist Devices

Children s Acute Transport Service

Hemodynamic Monitoring

PVDOMICS: Right Heart Catheterization Training

Central Line Care and Management

Central Venous Line Insertion

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018

Cath Lab Essentials: Basic Hemodynamics for the Cath Lab and ICU

High Impact Intervention Peripheral intravenous cannula care bundle

Successful IV Starts Revised February 2014

Central venous access devices for children with lysosomal storage disorders

PULMONARY ARTERY PRESSURE

Peel-Apart Percutaneous Introducer Kits for

The University of Toledo Medical Center and its Medical Staff

Per-Q-Cath* PICC Catheters with Excalibur Introducer* System

Pulmonary Artery Catheter Helpful Hints 2017

Curraheen, Co. Cork. Guidelines on the Management and Care of Central Venous Access Devices

Port Design. Page 1. Port Placement, Removal, and Management. Selecting a Vascular Access Device. Thomas M. Vesely, MD

ThruPort systems ProPlege peripheral retrograde cardioplegia device

Arterial Line Insertion Pre Reading

Home Health Foundation, Inc. To create more permanent IV access for patients undergoing long term IV therapy.

If viewing a printed copy of this policy, please note it could be expired. Got to to view current policies.

BPG 03: Continuous Renal Replacement Therapy (CRRT)

Pulmonary Artery Catheter Insertion (Assist) and Pressure Monitoring

Vascular access device selection & placement. Alisa Seangleulur, MD Anesthesiology Department, Faculty of Medicine, Thammasat University

Dr. prakruthi Dept. of anaesthesiology, Rrmch, bangalore

Competency Title: Caring for a patient with an arterial line

CARDIOVASCULAR MONITORING. Prof. Yasser Mostafa Kadah

PATIENT CARE PLAN FOR CARE OF PERIPHERAL MIDLINE. Manufacturers specific recommendations should be noted and adhered to by individual practitioners.

2. Need for serial arterial blood gas determinations. 2. Anticipation of the initiation of thrombolytic therapy

Sterile Technique & IJ/Femoral Return Demonstration

Research Article Can we predict the Position of Central Venous Catheter Tip Following Cannulation of Internal Jugular Vein?

SOUTHERN HEALTH & SOCIAL CARE TRUST. Children & Young People s Directorate Procedure/Guidelines/Protocol Checklist & Version Control Sheet

ompanionport Speciality Medical Devices For The Veterinary Community Surgical Suggestions

Section 6 Intra Aortic Balloon Pump

Prevent Bloodstream Infections by Using Appropriate Devices

Venepuncture and Cannulation. Louise Smith Clinical Nurse Specialist

RECOMMENDED INSTRUCTIONS FOR USE

Lifecath Twin & Dualyse/Trilyse Permanent and Temporary Renal Catheters

Procedures/Risks:central venous catheter

Document No. BMB/IFU/40 Rev No. & Date 00 & 15/11/2017 Issue No & Date 01 & 15/11/2017

CHEST DRAIN PROTOCOL

Peripherally Inserted Central Catheter & Midline Placement with ECG Confirmation of Tip Placement

Cardiac Catheterization Lab Procedures

Definition- study of blood flow Haemodynamic monitoring refers to monitoring of blood in the cardiovascular system Uses Is NB in the critically ill

Patient guide: pfm Nit-Occlud PDA coil occlusion system. Catheter occlusion of. Patent Ductus Arteriosus. with the

Ultrasound Guided Vascular Access. 7/25/2016

Relax and Learn At the Farm 2012

CENTRAL VENOUS ACCESS DEVICES. BETHANY COLTON

IV therapy. By: Susan Mberenga, RN, MSN. Copyright 2016, 2013, 2010, 2006, 2002 by Saunders, an imprint of Elsevier Inc.

Optimize your hemodialysis procedural bundle.

Heart catheterization for adults with congenital heart disease

BACHELOR OF SCIENCE IN CARDIO VASCULAR TECHNOLOGY

PROTOCOL FOR THE INSERTION OF NON TUNNELLED DIALYSIS CENTRAL VENOUS CATHETERS

If viewing a printed copy of this policy, please note it could be expired. Got to to view current policies.

Peripherally Inserted Central Catheter (PICC) Booklet

2. Indications Infusion of hyperosmolar medication, e.g. TPN. Administration of vasoactive/irritant drugs.

Appendix E: Overview of Vascular

Preventing CLABSI & CAUTI Preventive Measures for Central Line Associated Bloodstream Infection & Catheter Associated UTI

You have a what, inside you?

IB TOPIC 6.2 THE BLOOD SYSTEM

You have a what, inside you?

Mary Lou Garey MSN EMT-P MedFlight of Ohio

Guidelines for the Care and Maintenance of Intravenous Access Devices in Paediatric Patients

Central Venous Catheter Insertion: Assisting

FLEXIC ATH LTD. Peripherally Inserted. Instructions n For Use.

IV Fluids. Nursing B23. Objectives. Serum Osmolality

Emergency clamp should always be readily available in case of accidental catheter fracture

Education for Self Administration of Intravenous Therapy HOME IV THERAPY PICC. Portacath

HOSPITAL INFECTION CONTROL POLICY. Insertion and maintenance of peripheral intravenous cannulae

Central Venous Access Devices. Stephanie Cunningham Amy Waters

ER REBOA Catheter. Instructions for Use

North West London Trauma Network. Management of Chest Drains

The CardioMEMS HF system (Abbott Vascular,

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

Cardiac Catheterization, Coronary Angiogram, and Percutaneous Coronary Intervention (PCI)

FUNDAMENTALS OF HEMODYNAMICS, VASOACTIVE DRUGS AND IABP IN THE FAILING HEART

IV Fluids Nursing B23 Objectives Serum Osmolality 275 to 295 Isotonic

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

The Heart. Happy Friday! #takeoutyournotes #testnotgradedyet

Transcatheter Aortic Valve Implantation Procedure (TAVI)

Vascu-PICC WITH CUFF PERIPHERALLY INSERTED CENTRAL VEIN ACCESS CATHETER INSTRUCTIONS FOR USE

Core procedures assessment form

The Power of Purple* Polyurethane PICC. Patient Guide. Access Systems

Understanding your HICKMAN CATHETER

#4 Cardiovascular I The Heart

Information for patients undergoing Angiography (Angiogram) or Arteriography (Arteriogram) Patient Information

Computerized PAC Waveform Interpretation

All bedside percutaneously placed tracheostomies

Background & Indications Probe Selection

BASINGSTOKE AND NORTH HAMPSHIRE NHS FOUNDATION TRUST. Clinical Policy for Peripheral Venous Cannula Insertion and Management (Adults)

THE CARE AND MAINTENANCE OF TOTALLY IMPLANTED VENOUS ACCESS DEVICES (PORTS)

NOTTINGHAM CHILDREN S HOSPITAL Nursing Guideline. P07 Care and Management of Children with a Totally Implantable Venous Access Device (Portacath).

37 1 The Circulatory System

Central Venous Catheter Care and Maintenance (includes catheter troubleshooting guide)

Transcription:

A practical guide to using pulmonary artery catheters Scales K, Collie E (7) A practical guide to using pulmonary artery catheters. Nursing Standard. 21, 43, 42-48. Date of acceptance: May 4 7. Summary This article reviews the physiological principles underlying the use of the pulmonary artery (PA) catheter. The procedure for PA catheter insertion, PA pressure monitoring, PA catheter removal and complications associated with the PA catheter are discussed, as are nursing management of the catheter and nursing care of the patient. Authors Katie Scales is consultant nurse critical care, Eileen Collie is sister, critical care outreach, Charing Cross Hospital, Hammersmith Hospitals NHS Trust, London. Email: kscales@hhnt.nhs.uk Keywords Cardiovascular system and disorders; Coronary care; Haemodynamic monitoring; Pulmonary artery catheters These keywords are based on the subject headings from the British Nursing Index. This article has been subject to double-blind review. For author and research article guidelines visit the Nursing Standard home page at www.nursing-standard.co.uk. For related articles visit our online archive and search using the keywords. THE PULMONARY ARTERY (PA) flotation catheter is a specialised central venous catheter that is inserted into the right side of the heart and then floated into the PA (Scales 1999). These catheters are also known as flow-directed balloontipped PA catheters (Morton et al 5) and are more commonly referred to as PA catheters. There are several designs of PA catheter ranging from a simple catheter that measures PA pressure to complex multiple lumen catheters that also measure cardiac output and other variables. PA catheters are usually inserted in critical care environments such as the intensive therapy unit (ITU), coronary care unit (CCU), anaesthetic room or cardiac catheter laboratory where full haemodynamic monitoring is available. PA catheters are inserted to (Morton et al 5): Assess right ventricular function. Directly measure PA pressure. Indirectly measure left heart pressure by PA occlusion. Calculate cardiac output and other variables. Use the information gained to aid diagnosis and to assess the effect of cardiovascular interventions. Pulmonary artery catheters PA catheters were developed in the early half of the th century. Bradley, a British physician, first reported the use of a PA catheter in clinical practice in 1964. Branthwaite and Bradley (1968) described the clinical use of the PA catheter to measure cardiac output by the thermodilution method. Branthwaite and Bradley used a rigid, thermistor-tipped catheter placed in the PA under fluoroscopy and a second injection catheter placed in the right atrium (Steel and Bihari ). A breakthrough came when Swan and Ganz developed the multiple lumen, balloon-tipped catheter that could be placed without the use of fluoroscopy (Swan et al 197). In recent years, alternative, less invasive, technologies have been developed to measure cardiac output, for example, the oesophageal Doppler, and doubts have been raised about the benefits and safety of PA catheters (Connors et al 1996). A recent randomised controlled trial found no clear evidence of benefit or harm from the use of PA catheters (Harvey et al 5). The clinical decision to use a PA catheter should be made on a case-by-case basis weighing up the risks and the benefits for each individual patient (Reade and Angus 6). A PA flotation catheter has an inflatable balloon incorporated into the distal tip of the catheter, which allows the catheter to float within the bloodstream. The PA catheter is inserted through a central vein. When the tip of the catheter is in the right atrium the balloon is inflated and the catheter floats through the tricuspid valve into the right ventricle. The catheter then passes through the pulmonary valve and enters the PA. It is further advanced until the balloon is wedged in a small branch of the PA (Morton et al 5). The wedged balloon isolates a segment of the pulmonary bed and blood flow in that area ceases (Gomez and Palazzo 1998). 42 july 4 :: vol 21 no 43 :: 7 NURSING STANDARD

The pulmonary arteries divide to form pulmonary capillaries and these converge to form pulmonary veins which ultimately drain into the left atrium. Once the catheter is wedged there is effectively a static column of blood from the end of the catheter through the pulmonary capillary bed to the left atrium (Gomez and Palazzo 1998). When the mitral valve opens, the column of blood extends to the left ventricle. The distal lumen of the catheter records the pressure in this isolated segment of the pulmonary bed. The pressure recorded is an indirect measurement of the pressure in the left atrium because the pressure from the left side of the heart is transmitted through the capillary bed back to the distal tip of the catheter (Figure 1). The expression pulmonary capillary wedge pressure (PCWP) is sometimes used in clinical practice to describe the pressure recording obtained when the balloon is inflated. This is a misnomer and the terminology is misleading (Weed 1991). The pulmonary capillaries are not wedged, the downstream pressure in the occluded segment of PA equilibrates with pulmonary venous pressure (Wiedemann 1986, Gomez and Palazzo 1998) which is influenced by left atrial and left ventricular pressure. This technique was described by Swan et al (197) and a device to catheterise the right side of the heart was commercially marketed as a Swan Ganz catheter. Many devices are available for catheterisation of the right heart and the generic name for this type of device is the PA flotation catheter; it should only be referred to as a Swan Ganz catheter if the branded commercial product is specifically used. There are several designs of PA catheter (Scales 1999): Simple: a balloon at the tip and a distal lumen through which pulmonary pressure is measured. Intermediate: a balloon, a distal lumen and a proximal lumen opening into the right atrium through which central venous pressure can be measured. Complex: a balloon, distal and proximal lumens, plus a thermistor and electronic connections for cardiac output studies. Complex catheters can be used to measure PA pressure, pulmonary occlusion pressure, cardiac output and mixed venous oxygenation. Additional variables such as stroke volume, systemic vascular resistance (SVR), pulmonary vascular resistance (PVR), coronary perfusion pressure, oxygen delivery and oxygen uptake can also be calculated (Steel and Bihari ). These variables can only be calculated if the catheter is connected to a cardiac output computer or to a NURSING STANDARD FIGURE 1 Pulmonary artery (PA) catheter with inflated balloon wedged in a branch of the PA Pulmonary artery PA catheter FIGURE 2 Pulmonary artery catheter Cardiac output connection cm markers Balloon Balloon lumen cardiac output module that can be inserted into the patient s haemodynamic monitoring system. The complexity of the catheter selected will depend on the amount of haemodynamic information that is needed and the patient s clinical condition (Morton et al 5). After measuring the PA occlusion pressure (or PA wedge pressure) the balloon must be deflated to restore blood flow to the isolated segment of the pulmonary bed. Because the PA catheter sits within the heart it is subject to the rhythmical contractions produced by the systolic and diastolic phases of the cardiac cycle, consequently the position of the catheter tip can move. The catheter may need to be repositioned to obtain a wedged pressure. To prevent contamination of the catheter and to allow it to be advanced and withdrawn the catheter is enclosed in a plastic sleeve (Figure 2). Insertion procedure Static column of blood from distal tip of catheter to left atrium Left atrium The internal jugular is usually the preferred site for PA catheter insertion as the subclavian vein is associated with a higher risk of pneumothorax and the femoral vein is a difficult site from which to float the catheter (Gomez and Palazzo 1998). Plastic protective sleeve Locking device july 4 :: vol 21 no 43 :: 7 43

BOX 1 Equipment for pulmonary artery (PA) catheter insertion 1. Absorbent sheet to place beneath the patient. 2. Dressing pack or procedure pack to provide a sterile field. 3. Large, sterile fenestrated drape. 4. Sterile gown and gloves. 5. 2% chlorhexidine gluconate in 7% isopropyl alcohol. 6. Lidocaine 1%, ml syringe, blue/orange needle. 7. Introducer sheath. 8. PA catheter kit. 9. Sodium chloride.9% for injection.. Suture or StatLock 11. Large, moisture-permeable transparent dressing. FIGURE 3 Two-part introducer sheath with self-sealing diaphragm or haemostasis valve Sheath Sheath FIGURE 4 Side arm/port Side arm/port The heart showing normal pressure values with pulmonary artery (PA) catheter in place PA catheter Flexible section Right atrium 5 Flexible section Suture point Connection Right ventricle Haemostasis valve 25 5 Aorta Haemostasis valve self-sealing 1 8 Pulmonary artery 25 Left atrium 8 Left ventricle 1 8 The use of two-dimensional ultrasound is recommended for the placement of central venous catheters in the internal jugular vein (National Institute for Clinical Excellence 2). The procedure should be explained to the patient and informed consent obtained wherever possible (Scales 1999). The environment is prepared to ensure that the electrocardiogram (ECG) monitor can be easily observed and resuscitation equipment is readily available. Once the equipment has been assembled (Box 1), the patient is positioned flat or head down in the Trendelenburg position, which increases the pressure within the vein and reduces the risk of air embolism during the insertion procedure (Scales 1999). The catheter should be inserted under strict asepsis following the epic2 guidelines (Pratt et al 7): Full surgical scrub, sterile gown and gloves. Skin antisepsis using 2% chlorhexidine gluconate (unless the patient is allergic to chlorhexidine). Clean the skin for a full minute and allow to air dry. Full barrier precautions using large, sterile drapes. To insert a PA catheter a wide bore percutaneous introducer sheath must first be placed in a central vein. The introducer sheath is often a two-part device composed of a relatively short, wide bore catheter to which a self-sealing diaphragm with side arm is attached (Figure 3). Once the introducer sheath is in place the clinician flushes all the lumens of the PA catheter with.9% sodium chloride to expel air and to reduce the risk of air embolism. The balloon at the tip of the catheter is tested to ensure that it inflates correctly and has no leaks (Morton et al 5). A self-limiting syringe is provided in the PA catheter kit, which prevents the balloon from being over inflated: the volume limit of the balloon is 1.5ml. The dedicated syringe should be retained and used to inflate the balloon once the catheter is in situ in the PA. Having checked the balloon the air is withdrawn and the sterile plastic sleeve is placed over the catheter. The catheter is then fed through the self-sealing diaphragm of the introducer sheath and the plastic sleeve connected to the introducer sheath to protect the catheter. The port for the distal lumen of the catheter is connected to transduced haemodynamic monitoring so that the pressure waveforms can be displayed as the catheter passes through the chambers of the right heart and into the PA. To ensure accuracy of the recordings the haemodynamic monitoring system must be zeroed to atmospheric pressure and levelled with the patient s right atrium (McGhee and Bridges 2). 44 july 4 :: vol 21 no 43 :: 7 NURSING STANDARD

To interpret the waveforms as the PA catheter is directed through the heart, the nurse should be familiar with normal cardiac anatomy and physiology including the normal pressures within the heart and arteries (Figure 4). With the balloon deflated the PA catheter is advanced through the superior vena cava until it enters the right atrium. A right atrial pressure wave (or central venous pressure) will be displayed on the monitor (Figure 5a). Once the tip of the catheter is in the right atrium the balloon is inflated and the catheter gently advanced. The air-filled balloon makes the tip of the catheter lightweight and buoyant and the catheter is carried along in the blood flow into the right ventricle. The ECG monitor should be observed as dysrhythmias may occur as the catheter passes through the tricuspid valve. Atrial and ventricular ectopics and salvos of ventricular tachycardia are not uncommon and usually resolve without treatment when catheter manipulation ceases (Iberti et al 1985). The position of the catheter within the right ventricle is confirmed by the change in pressure which is recorded on the monitor (Figure 5b). The balloon remains inflated and the catheter is further advanced. With gentle manipulation the catheter will turn and float upwards through the pulmonary valve and into the PA. The position in the PA is confirmed by the change in pressure seen on the monitor (Figure 5c). Once PA pressure has been recorded the catheter is further advanced with the balloon inflated. The PA divides and the branches become progressively smaller. The PA catheter is advanced until the inflated balloon becomes wedged in a branch of the PA. This occludes the artery and prevents further blood flow past the catheter. The distal tip of the catheter is now exposed to the pressure in the pulmonary vascular bed, which is continuous with the left side of the heart. The pressure recorded is the pulmonary occlusion pressure or pulmonary wedge pressure and is an indirect measurement of the pressure in the left side of the heart. The wedged position is confirmed by the change in pressure on the monitor (Figure 5d). After the wedged pressure has been recorded the balloon is deflated and the waveform on the monitor is checked to ensure that the PA pressure is displayed (Morton et al 5), indicating that blood flow has been restored to the section of the pulmonary bed that had been occluded. The pressure changes as the catheter is advanced from the right atrium, through the right ventricle, into the PA and finally wedged can be seen in Figure 6. Once the correct position of the catheter has been confirmed the device should be secured. Traditionally sutures are used to secure central NURSING STANDARD FIGURE 5 Individual pressure waveforms seen as the pulmonary artery (PA) catheter is advanced through the heart a) b) c) d) - Right atrial pressure waveform - - Right ventricular pressure waveform Pulmonary artery pressure waveform - Pulmonary occlusion or wedged pressure waveform PA catheter venous catheters but there is growing evidence to suggest that sutures contribute to catheterrelated bloodstream infection (Maki and Crnich 3). Sutureless devices such as StatLock have been shown to reduce catheter-related bloodstream infection (Crnich and Maki 2). The entry site should be covered with a sterile transparent dressing to increase security and make inspection of the entry site easier (Royal College of Nursing (RCN) 5). The proximal end of the plastic sheath should be locked onto the PA catheter to ensure that the catheter does not become dislodged and to maintain the sterility of the catheter within the sheath. The nurse should note the position of the PA catheter by reading the catheter depth july 4 :: vol 21 no 43 :: 7 45

FIGURE 6 markings at the entrance to the percutaneous sheath, for example, cm and confirming that this position produces an unwedged PA waveform on the monitor. The catheter length should be recorded in the nursing notes. The nurse s role during the insertion of the PA catheter is to ensure that the sterile technique is maintained, to observe the haemodynamic waveforms on the monitor, to record the pressures as the catheter passes through the heart and into the PA and to monitor the ECG for dysrhythmias during the procedure (Morton et al 5). The nurse will also need to give the patient psychological support and reassurance. Cannulation of the neck veins usually requires the placement of sterile towels over the patient s face which may be difficult for patients with claustrophobia and head down positioning may be uncomfortable for patients who experience breathlessness (Scales 1999). The position of the catheter is usually confirmed on chest X-ray once the catheter has been secured. The distal lumen of the PA catheter provides continuous real-time measurement of PA pressure. This lumen is rarely used for drug or fluid administration but is used to obtain mixed venous blood samples to assess oxygen consumption. The proximal lumen of the PA catheter can be used for central venous pressure monitoring and drug or fluid administration. If the catheter is a cardiac output catheter, the proximal lumen will also be used to administer the injectate required for thermodilution cardiac output assessment. Each lumen is clearly labelled. Recording pulmonary artery and wedge pressures PA pressure should be monitored and recorded with all other observations. PA wedge pressure Pressure changes as the catheter is advanced from right atrium to right ventricle to pulmonary artery to wedged position - Right atrium Right ventricle Pulmonary artery Wedge should also be recorded if the nurse has been trained to do so. Before recording the PA wedge pressure the nurse should ensure that the transducers have been zeroed to atmosphere and levelled to the patient s right atrium (Morton et al 5). The position of the patient should be consistent with previous recordings. The PA balloon should be inflated gently while observing the monitor. When the waveform changes from the pulsatile PA pressure with clear systolic and diastolic values to a waveform that is similar to a central venous pressure waveform, the PA catheter is considered to be wedged successfully. The mean value is recorded. The balloon is deflated and the nurse must check to ensure that a pulsatile PA pressure wave is again displayed on the monitor. There is usually a gate device or a tap on the balloon lumen of the PA catheter and it is standard practice to close the lumen after the balloon has been allowed to deflate. This prevents accidental inflation of the balloon. If the nurse has not been trained to perform wedge pressures, this haemodynamic indicator will only be recorded intermittently when a member of the medical team performs the procedure. Complications of pulmonary artery catheter insertion Complications of PA catheter insertion include arterial puncture, pneumothorax, air embolism and infection, which are general complications of central venous catheter insertion (Scales 1999). The post-insertion chest X-ray will reveal or exclude the presence of a pneumothorax. Dysrhythmias are commonly associated with PA catheter insertion because the catheter passes through the heart. While most dysrhythmias resolve spontaneously, 1-3% of dysrhythmias require treatment (Boyd et al 1983). PA rupture is an important and often fatal complication and is associated with over inflation of the PA balloon (Hardy et al 1993), but can also be caused by erosion of the wall of the PA (Gomez and Palazzo 1998). PA rupture is more common in patients with pulmonary hypertension, patients on anticoagulants and patients aged over 6 years (Hardy et al 1993). The PA catheter can become accidentally wedged, without balloon inflation, due to the forward movement of the catheter into a narrower section of artery. Distal (forward) migration of the PA catheter occurs in approximately 1% of patients and may result in pulmonary infarction if not recognised (Boyd et al 1983). This may also occur when patients are repositioned. If a wedged waveform can be achieved using less than 1ml of air to inflate the balloon it is likely that the catheter has migrated 46 july 4 :: vol 21 no 43 :: 7 NURSING STANDARD

from the correct position (Morton et al 5). The nurse should routinely observe the PA waveform on the monitor, particularly when the patient is repositioned, and should be able to recognise the wedged trace and take action to rectify the problem (Scales 1999). In some ITUs nurses who have been trained to use PA catheters will manipulate the catheter into a safe position, in other units this will be a medical responsibility. Pulmonary artery catheter removal When the PA catheter is no longer required it should be removed (RCN 5). Removal of the PA catheter is similar to removal of other central venous catheters, however, the catheter is significantly longer and less easy to handle. It is necessary to place the catheter on a surface when it has been removed such as a disposable, absorbent waterproof sheet (Scales 1999). Before removal of the PA catheter the nurse must check that the balloon is completely deflated. Significant damage could occur to the pulmonary and tricuspid valves if the catheter is withdrawn with the balloon inflated (Scales 1999). The catheter is withdrawn through the right ventricle and it is important to remember that the ventricle contains papillary muscles which are connected to the tricuspid valve by chordae tendineae. There are reports of PA catheters becoming entangled in the chordae tendineae leading to chordae rupture (Arnaout et al 1) and of catheters becoming knotted around the chordae (Tremblay et al 1992, Ismail et al 1998). If resistance is felt during the removal of a PA catheter the procedure should be stopped and medical advice sought. Ideally the introducer sheath should be removed at the same time as the PA catheter. Introducer sheaths are wide-bore devices that carry a significant risk of air embolism and haemorrhage and should not be left in situ as routine vascular access (Darovic 2). The procedure for removal of a PA catheter is outlined in Box 2. Dysrhythmias may occur as the PA catheter is removed. Atrial ectopics, ventricular ectopics and small runs of ventricular tachycardia are not uncommon. If dysrhythmias occur it is prudent to continue to remove the catheter rather than stopping, which may risk leaving the catheter tip positioned in a valve, thus causing further and more severe dysrhythmias (Scales 1999). If dysrhythmias persist after the catheter has been removed from the heart, medical assistance should be sought immediately. If catheter-related bloodstream infection is suspected the tip of the catheter should be sent to microbiology to be cultured. If this is required the NURSING STANDARD BOX 2 Removal of pulmonary artery (PA) catheter Equipment Non-sterile gloves. Stitch cutter. 2% chlorhexidine gluconate in 7% isopropyl alcohol. Sterile gauze. Transparent dressing. Disposable, absorbent waterproof sheet. Procedure 1. Explain the procedure to the patient. 2. Position patient supine or in Trendelenburg if tolerated. 3. Position electrocardiogram (ECG) monitor where it can be easily observed. 4. Put on non-sterile gloves. 5. Remove transparent dressing, clean site with 2% chlorhexidine gluconate. 6. Check the PA catheter balloon is fully deflated. 7. Discontinue PA monitoring, close distal lumen with cap or three-way tap. 8. Transfer any ongoing infusions to alternative venous access. 9. Remove sutures or open StatLock.. Place sterile gauze over the catheter entry site and apply light pressure. 11. Gently withdraw catheter while observing the ECG for dysrhythmias. 12. Place catheter on disposable, absorbent waterproof sheet. 13. Apply firm pressure to the puncture site for at least two minutes or longer if necessary to achieve haemostasis. 14. Immediately apply transparent air-tight dressing ensuring a good seal. Leave dressing in place for 72 hours, replace if wet or dirty (Scales 1999, RCN 5) and check the dressing regularly for bleeding and to ensure that the dressing is intact. catheter must be placed on a sterile surface when it has been removed. The tip should be cut off using sterile scissors and the tip placed in a sterile specimen container which should be labelled with the patient s details. Conclusion PA catheters have provided valuable haemodynamic information for more than years, however, they are not without july 4 :: vol 21 no 43 :: 7 47

complications and the benefit of the information obtained should be balanced against the risks of their use. Alternative methods of cardiac output assessment have been developed and many units prefer to use less invasive methods such as the oesophageal Doppler to measure cardiac output, although most units have at least two methods of cardiac output measurement at their disposal (Esdaile and Raobaikady 5). Management of the PA catheter and performance of haemodynamic measurements are important aspects of the care of critically ill patients. Interpretation of PA occlusion pressure requires experience and an understanding of the clinical situations that can induce errors in interpretation (Steel and Bihari ). To expand practice to include PA catheter management it is essential that nurses have a thorough knowledge of cardiac anatomy and physiology as well as a practical understanding of haemodynamic monitoring to care safely for the patient NS References Arnaout S, Diab K, Al-Kutoubi A, Jamaleddine G (1) Rupture of the chordae of the tricuspid valve after knotting of the pulmonary artery catheter. Chest. 1, 5, 1742-1744. Boyd KD, Thomas SJ, Gold J, Boyd AD (1983) A prospective study of complications of pulmonary artery catheterizations in 5 consecutive patients. Chest. 84, 3, 245-249. Bradley RD (1964) Diagnostic right-heart catheterisation with miniature catheters in severely ill patients. The Lancet. 284, 7366, 941-942. Branthwaite MA, Bradley RD (1968) Measurement of cardiac output by thermal dilution in man. Journal of Applied Physiology. 24, 3, 434-438. Connors AF Jr, Speroff T, Dawson NV et al (1996) The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. Journal of the American Medical Association. 276, 11, 889-897. Crnich CJ, Maki DG (2) The promise of novel technology for the prevention of intravascular device-related bloodstream infection. I. Pathogenesis and short-term devices. Clinical Infectious Diseases. 34, 9, 1232-1242. Darovic GO (2) Percutaneous sheath introducer systems used for venous access. Critical Care Nurse. 22, 1, 74-77. Esdaile B, Raobaikady R (5) Survey of cardiac output monitoring in intensive care units in England and Wales. 25th International Symposium on Intensive Care and Emergency Medicine. March 21-25, Belgium, Brussels. Gomez CM, Palazzo MG (1998) Pulmonary artery catheterization in anaesthesia and intensive care. British Journal of Anaesthesia. 81, 6, 945-956. Hardy JF, Morissette M, Taillefer J, Vauclair R (1993) Pathophysiology of rupture of the pulmonary artery by pulmonary artery balloon-tipped catheters. Anesthesia and Analgesia. 62,, 925-9. Harvey S, Harrison DA, Singer M et al (5) Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. The Lancet. 366, 9484, 472-477. Iberti TJ, Benjamin E, Gruppi L, Raskin JM (1985) Ventricular arrhythmias during pulmonary artery catheterization in the intensive care unit. Prospective study. American Journal of Medicine. 78, 3, 451-454. Ismail KM, Deckmyn TJ, Vandermeersch E, Stockx L (1998) Nonsurgical extraction of intracardiac double-knotted pulmonary artery catheter. Journal of Clinical Anesthesia., 2, 16-162. Maki DG, Crnich CJ (3) Line sepsis in the ICU: prevention, diagnosis, and management. Seminars in Respiratory and Critical Care Medicine. 24, 1, 23-36. McGhee BH, Bridges EJ (2) Monitoring arterial blood pressure: what you may not know. Critical Care Nurse. 22, 2, 6-73. Morton PG, Fontaine DK, Hudak CM, Gallo BM (5) Critical Care Nursing: a Holistic Approach. Eighth edition. Lippincott Williams and Wilkins, Philadelphia PA. National Institute for Clinical Excellence (2) Guidance on the Use of Ultrasound Locating Devices for Placing Central Venous Catheters. Technology Appraisal Guidance No. 49. NICE, London. Pratt RJ, Pellowe CM, Wilson JA et al (7) epic2: national evidence-based guidelines for preventing healthcare-associated infections in NHS hospitals in England. Journal of Hospital Infection. 65, Suppl 1, S1-S64. Reade MC, Angus DC (6) PAC-Man: game over for the pulmonary artery catheter? Critical Care., 1, 3. Royal College of Nursing (5) Standards for Infusion Therapy. RCN, London. Scales K (1999) Vascular access in the acute care setting. In Dougherty L, Lamb J (Eds) Intravenous Therapy in Nursing Practice. Churchill Livingstone, London, 261-299. Steel A, Bihari D () Pulmonary artery catheterisation: a problem swept under the carpet? Critical Care and Resuscitation. 2, 2, 146-149. Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D (197) Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter. New England Journal of Medicine. 283, 9, 447-451. Tremblay N, Taillefer J, Hardy JF (1992) Successful non-surgical extraction of a knotted pulmonary artery catheter trapped in the right ventricle. Canadian Journal of Anaesthesia. 39, 3, 293-295. Weed HG (1991) Pulmonary capillary wedge pressure not the pressure in the pulmonary capillaries. Chest., 4, 1138-11. Wiedemann HP (1986) Wedge pressure in pulmonary veno-occlusive disease. New England Journal of Medicine. 315, 19, 1233. 48 july 4 :: vol 21 no 43 :: 7 NURSING STANDARD