ATTUNE KNEE SYSTEM: SOFCAM CONTACT

Similar documents
ATTUNE Knee System: Stability in Total Knee Replacement

Fixed Bearing. stabilityinmotion

Rotating Platform. stabilityinmotion

ATTUNE Knee System: Cruciate Sacrificing Fixed Bearing

STABILITY & MOTION VS. THE PERSONA KNEE SYSTEM. The Zimmer Persona Knee

BALANCING THE PATIENT NEED FOR FREEDOM OF MOVEMENT AND JOINT STABILITY

Evolution. Medial-Pivot Knee System The Bi-Cruciate-Substituting Knee. Key Aspects

Contact stress at the post-cam mechanism in posterior-stabilised total knee arthroplasty

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing

Tibial Base Design Factors Affecting Tibial Coverage After Total Knee Arthroplasty: Symmetric Versus Asymmetric Bases

Frontal Plane Kinematics After Mobile- Bearing Total Knee Arthroplasty

TOTAL KNEE SYSTEM WHEN INNOVATION AND INTUITION ALIGN.

Estimating Total Knee Arthroplasty Joint Loads from Kinematics

BIOMECHANICAL MECHANISMS FOR DAMAGE: RETRIEVAL ANALYSIS AND COMPUTATIONAL WEAR PREDICTIONS IN TOTAL KNEE REPLACEMENTS

Rotating Platform System. Product Rationale

Advantages of Mobile-bearing Total Knee Arthroplasty

STIFFNESS AFTER TKA PRE, PER AND POST OPERATIVE CAUSING FACTORS

Rotaglide+ TM. Total Knee System Product overview

Think isometry Feel balance

Gold standard of a TKA. Conflicting goals? POLYETHYLENE WEAR THE SOLUTION: MOBILE BEARING KNEES. MOBILE BEARING A totally new approach (1977)

EXPERIENCE GPS FOR TOTAL KNEE ARTHROPLASTY DETERMINE YOUR OWN COURSE.

TOTAL KNEE ARTHROPLASTY (TKA)

Kinematics Analysis of Different Types of Prosthesis in Total Knee Arthroplasty with a Navigation System

Zimmer FuZion Instruments. Surgical Technique (Beta Version)

I lllll llllllll Ill lllll lllll lllll lllll lllll

Luminus Flex Total Knee System (Fixed Type) The Innovative Technology for Knee Replacement

EMPOWERING LIFE THROUGH NATURAL MOTION

Instrumentation Options

ATTUNE KNEE SYSTEM 25% 25% 25% 25% VALUE ANALYSIS BRIEF VALUE SUMMARY UNMET NEEDS IN TOTAL KNEE ARTHROPLASTY

Analysis of factors affecting range of motion after Total Knee Arthroplasty

Consistency in U2 Knee System over Time

BOOK CHAPTERS PUBLICATIONS

itotal Customized POSTERIOR-STABILIZED knee replacement system

Comparison of high-flex and conventional implants for bilateral total knee arthroplasty

Intramedullary Tibial Preparation

ConforMIS, Inc. 28 Crosby Drive Bedford, MA Phone: Fax:

The Cruciate Ligaments in Total Knee Arthroplasty

Stephen R Smith Northeast Nebraska Orthopaedics PC. Ligament Preserving Techniques in Total Knee Arthroplasty

Presented By Dr Vincent VG An MD BSc (Adv) MPhil Dr Murilo Leie MD Mr Joshua Twiggs BEng Dr Brett A Fritsch MBBS FRACS (Orth) FAOrthA.

LAMINA SPREADER SURGICAL TECHNIQUE

DATA & RESEARCH OVERVIEW

20 Kinematics of Mobile Bearing Total Knee Arthroplasty

Ju-Yeong Kwak, MD, Jae-Heon Jeong, MD*, Sang-Hak Lee, MD, Ho-Joong Jung, MD, Young-Bok Jung, MD

NexGen Legacy LPS-Flex Knee. Brochure

HIGH FLEXION IN CONTEMPORARY TOTAL KNEE DESIGN: A PRECURSOR OF UHMWPE DAMAGE? A FINITE ELEMENT STUDY

CLINICAL AND OPERATIVE APPROACH FOR TOTAL KNEE REPLACEMENT DR.VINMAIE ORTHOPAEDICS PG 2 ND YEAR

NexGen CR-Flex and LPS-Flex Knees. Design Rationale

Midterm results of cemented Press Fit Condylar Sigma total knee arthroplasty system

Extramedullary Tibial Preparation

TRUMATCH PERSONALIZED SOLUTIONS with the SIGMA High Performance Instruments

Bicruciate-Retaining or Medial Pivot Total Knee Prosthesis Pritchett 225 Fig. 3. The MP total knee prosthesis. Fig. 1. An anteroposterior radiograph o

Zimmer Segmental System

Surgical Technique. Hinge Disassembly and Rebuild Technique

RESECTION GUIDE SYSTEM. TRUMATCH Personalized Solutions Surgical Technique with ATTUNE Knee INTUITION Instruments

CLASSIFICATION OF MOBILE BEARING KNEE DESIGN: MOBILITY AND CONSTRAINT

Biomechanical Effects of Femoral Component Axial Rotation in Total Knee Arthroplasty (TKA)

Insert dissociation after fixed bearing PS constrained Genesis II total knee arthroplasty. A case series of nine patients

We carried out weight-bearing video radiological

and Screw-Home in a Mobile-Bearing Total Knee Arthroplasty

Knee Revision. Portfolio

Dora Street, Hurstville 160 Belmore Road, Randwick

JOINT RULER. Surgical Technique For Knee Joint JRReplacement

ANTERIOR REFERENCE NEXGEN COMPLETE KNEE SOLUTION. Multi-Reference 4-in-1 Femoral Instrumentation Anterior Reference Surgical Technique

Triathlon Knee System. Universal Baseplate Surgical Protocol

Biomechanics of the Knee. Valerie Nuñez SpR Frimley Park Hospital

Bone Preservation Stem

TKA Gap Planning. Supporting healthcare professionals

Newer designs in Knee replacement The good,bad & ugly! Dr VAIBHAV BAGARIA

ZIMMER NEXGEN CRUCIATE RETAINING (CR) AND LEGACY KNEE POSTERIOR STABILIZED (LPS) TRABECULAR METAL MONOBLOCK TIBIAS. Surgical Technique Addendum

why bicompartmental? A REVOLUTIONARY ALTERNATIVE TO TOTAL KNEE REPLACEMENTS

Persona. The Personalized Knee. Trabecular Metal Tibia. Surgical Technique

The information contained in this document is intended for healthcare professionals only.

The prosthetic implant of MBA Optetrak in the treatment of knee osteoarthritis: 5 years of experience

Masterclass. Tips and tricks for a successful outcome. E. Verhaven, M. Thaeter. September 15th, 2012, Brussels

The challenge. Knee replacement patients report lower levels of satisfaction versus hip replacement patients. Return to sports activity post surgery

Sasaki E 1,2, Otsuka H 2, Sasaki N 2, and Ishibashi Y 1

Range of Motion of Standard and High-Flexion Posterior Stabilized Total Knee Prostheses A PROSPECTIVE, RANDOMIZED STUDY

THE KNEE SOCIETY VIRTUAL FELLOWSHIP

ATTUNE KNEE SYSTEM 30% 30% 20% 20% VALUE ANALYSIS BRIEF VALUE SUMMARY BENEFITS OF TOTAL KNEE ARTHROPLASTY (TKA) UNMET NEEDS IN TKA

Unicondylar Knee Vs Total Knee Replacement: Is Less Better In the Middle Aged Athlete

Quality of Life. Quality of Motion.

Gap Balancing vs. Measured Resection Technique in Total Knee Arthroplasty

POSTERIOR REFERENCE NEXGEN COMPLETE KNEE SOLUTION. Multi-Reference 4-in-1 Femoral Instrumentation Posterior Reference Surgical Technique

Unicompartmental Knee Replacement

ACTIS. TOTAL HIP SYSTEM Approach Active Patients with Confidence DESIGN RATIONALE

The ability of isolated and combined ACL reconstruction and/or lateral monoloop tenodesis to restore intact knee laxity in the presence of isolated

Robotic-Arm Assisted Total Knee Arthroplasty Demonstrated Greater Accuracy to Plan Compared to Manual Technique

THE KNEE SOCIETY VIRTUAL FELLOWSHIP

PIN GUIDE SYSTEM SURGICAL TECHNIQUE. with the SIGMA High Performance Instruments System. This publication is not intended for distribution in the USA.

A Strain-Energy Model of Passive Knee Kinematics for the Study of Surgical Implantation Strategies

PARTIALS?1 SHOULD BE AT LEAST 20% DID YOU KNOW OF KNEE REPLACEMENTS

Triathlon Knee System Universal Baseplate Surgical Protocol

Continuing the Tradition. VerSys Heritage Hip System

ClassiQ. Scorpio. Anterior Referencing Surgical Protocol. Anterior Referencing. For use with Scorpio ClassiQ Instrument System

Knee kinematics after TKA depends on preoperative kinematics

A study of functional outcome after Primary Total Knee Arthroplasty in elderly patients

INSTRUMENTS. Design Rationale

Unusually High Rate of Early Failure of Tibial Component in ATTUNE Total Knee Arthroplasty System at Implant Cement Interface

Approach Patients with CONFIDENCE

Figure 3 Figure 4 Figure 5

Transcription:

ATTUNE KNEE SYSTEM: SOFCAM CONTACT Douglas A. Dennis, MD Medical Director at Porter Center for Joint Replacement Denver, Colorado Historically, sagittal plane instability following Total Knee Arthroplasty (TKA) due to insufficiency or resection of the posterior cruciate ligament resulted in the development of posterior cruciate substituting (PS) TKA designs. These devices incorporate a cam-spine mechanism to provide posterior stability and assure posterior femoral translation to enhance flexion following TKA. Many PS TKA designs have been developed which, while similar in concept, differ in many areas including the degree of flexion at which cam-spine engagement occurs, location, size and shape of the cam and spine, location of spine contact, freedom of axial rotation, and the amount of bone resection required to accommodate the intercondylar box of the femoral component. Numerous long term clinical reports of subjects implanted with PS TKA have demonstrated excellent results 1,2 as well as better weight bearing knee flexion compared to CR knees; however, despite these results, in vivo kinematic analyses still demonstrate wide variations in cam-spine mechanics. Some designs engage earlier providing kinematic control of anterior-posterior (A/P) translation during gait. In others, cam-spine engagement occurs later in flexion (65-100 degrees) and in some cases, cam-spine engagement never occurs throughout knee flexion. Figure 1. Photograph of a retrieved PS polyethylene insert demonstrating eccentric medial spine wear. Additionally, there are other considerations that must be taken into account. Cam-spine engagement may occur at flexion angles which differ from that reported by implant manufacturers due to varying patient kinematics or surgical placement which may cause cam-spine contact on the medial, lateral, or anterior aspect of the spine rather than centrally on the posterior aspect resulting in increased spine stresses and premature spine wear (Figure 1). 4 The velocity of the cam at the time of spine engagement varies and is affected by the condylar control of A/P femoral-tibial translation before engagement occurs. High engagement velocities result in elevated collision impact forces rather than a softer controlled pattern of cam-spine engagement. Some of these kinematic variances likely play a significant role in premature failures of PS TKA designs that have been reported due to excessive spine wear or fracture. 4-6

Anterior impingement can also occur with knee hyperextension or malpositioning of the femoral and/or tibial components. It is typically caused by the femoral component being placed in flexion and excessive posterior slope added to the tibia (Figure 2). Anterior impingement increases spine stresses and wear and adds additional stress to modular tibial insert locking mechanisms. 9 Cam-spine mechanisms must be robust to allow for surgical variation without the development of anterior spine impingement. Opponents of posterior cruciate substitution often object to the amount of bone removed from the intercondylar notch of the femoral component which affects available bone mass should later revision TKA become necessary. Historically, the shape of bone resection has been rectangular and the quantity of bone resection has not been proportional based on femoral component size. Figure 2. Photograph of a retrieved PS polyethylene insert demonstrating anterior spine impingement and subsequent wear. As a result, PS knees should be designed with the goal of reducing these issues by incorporating the following concepts. PS knees should have consistent cam-spine contact at a higher flexion angle than occurs during gait to lessen the frequency of cam-spine contact and enhance longevity of the cam-spine mechanism. Loads on the spine increase during progressive knee flexion; 7 therefore, it is desirable to have the location of camspine contact in deep flexion at the base of the spine where the strength is greatest. The geometry of the cam and spine must be forgiving of variances in axial rotation to avoid high eccentric spine stresses. The articular geometry of both the femoral component (J Curve) and tibial insert must produce A/P femoral-tibial translation patterns that assure the cam engages with the spine in a soft manner with minimized impact. And the amount of bone resection from the intercondylar notch should be minimized and proportional based on size of the femoral component. ATTUNE KNEE SYSTEM CAM-SPINE MECHANISM (SOFCAM CONTACT) Extensive research analyses using multiple analytical methodologies were performed to determine the ideal cam-spine mechanism for the ATTUNE Knee System posterior cruciate substituting TKA. Multiple design prototypes were developed as a part of the project in order to evaluate the parameters that affect the functional characteristics of the cam-spine engagement. Each design was evaluated for the factors mentioned above using computer modeling at the University of Denver and DePuy Synthes Joint Reconstruction, weight bearing cadaveric testing at the University of Kansas in a highly modified Oxford rig, and through multiple surgeon cadaveric and intra-operative evaluations. Ultimately, the SOFCAM Contact design best accomplished the design goals. Initially, the J Curve of the femoral component was determined. As is present in the native knee, a multi-radius concept was selected. Review of pre-existing multi-radius femoral component designs showed abrupt changes in the radius of curvature in most femoral components. 10 2

These abrupt changes result in a loss of control of A/P femoral-tibial translation and subsequent control of camspine engagement. To address this the ATTUNE GRADIUS Curve (gradually reducing femoral radius of curvature) was developed. The ATTUNE GRADIUS incorporates smooth transitions between radius changes which enhances the desired control of cam-spine engagement and lessens paradoxical anterior femoral translation seen in many current implants (Figure 3). The end result is that A/P femoral-tibial translation is initially controlled by a condylar geometry which gives rise to the femoral component translating slowly into engagement with the tibial spine, providing a smooth transition from condylar to cam-spine control. 7 This smooth (soft) engagement of the cam and spine occurs after the gait cycle (between 65 and 75 degrees of flexion) to eliminate excessive repetitive contact during each gait cycle which could lead to premature spine wear. A/P Femoral Translation during Flexion 5 A(+)/P Translation (mm) 2.5 0-2.5-5 -7.5-10 10 20 30 40 50 60 70 80 90 100 Multi-Radius Design ATTUNE GRADIUS Curve System Reducing Radius Design Knee Flexion (Deg) ATTUNE CR Fixed Bearing System Traditional Multi-Radius Design Point of Radii Transition in Multi-Radius Design Figure 3. Diagrams demonstrating the traditional multi-radius design versus the ATTUNE GRADIUS Curve design and the kinematic profiles of A/P translation of each demonstrating elimination of paradoxical anterior femoral translation in the ATTUNE Knee design. 10 The shape of the cam and spine has a significant impact on the contact and torsional stresses applied to the tibial spine and the tibial base fixation interface. With this in mind, the patented S-shape of the ATTUNE Knee cam and spine was created to provide a large contact area as the cam initially engages broadly on the spine (Figure 4A) and then quickly translates to the distal base of the tibial spine as the knee moves into deeper flexion (Figure 4B). The low contact position in high flexion directs the forces through the thickest portion of the tibial insert instead of through the upper spine which lessens spine stresses, the subsequent risk of spine fracture, and the danger of camspine dislocation. 7 The shape of the cam and spine also direct forces more distally, versus anteriorly, creating compressive forces at the tibia which are more protective of the tibial/bone interface. 8 Figure 4A: Cross-sectional diagram of the SOFCAM at 70 of flexion demonstrating broad levels of camspine contact. Figure 4B: Cross-sectional diagram of the SOFCAM at 145 of flexion demonstrating contact at the base of the spine where polyethylene strength is maximized. 3

Furthermore, several publications describe a spine that does not allow rotational freedom as a driver for tibial spine impingement and fracture. 6,9 With the ATTUNE Knee, the anterior surface and side of the tibial post, along with the corresponding surfaces on the femoral component, provide the necessary rotation and softened surfaces to avoid impingement should hyperextension or varus/valgus lift off occur. The shape of the posterior surfaces of the spine are designed to successfully manage forces transferred through the spine as well as loads transmitted to the tibial fixation interface by incorporating a smaller coronal radius, resulting in a curved shape. This curvature reduces edge loading of the femoral component against the tibial spine when the knee rotates in flexion (Figure 5). This design has been proven to reduce the contact forces on the tibial spine as compared to other knee designs. 7 It also helps to diffuse the torsional forces which are transferred to the modular insert locking mechanism and the tibial fixation interface during femoral rotation. Figure 5: Photograph demonstrating how the curved tibial spine allows freedom of rotation as knee flexion increases. Finally, the shape of the femoral intercondylar box has been designed to conserve bone by making an 18 angled cut (Figure 6). The amount of bone resected is proportional to the femoral component size, which can be beneficial in bone conservation for smaller patients with limited bone mass (Figure 7). The volume of bone removed is therefore appropriate for specific patient size. 18 Figure 6: Sagittal view of the ATTUNE System femoral component demonstrating the angular (18 ) intercondylar box which gives rise to less intercondylar bone removal. Figure 7: Diagram of the coronal cross-sectional geometry of the ATTUNE Knee femoral component demonstrating how the intercondylar box width, and subsequent intercondylar bone resection required, lessens with decreasing femoral component size. 4

SUMMARY In summary, SOFCAM Contact was developed after exhaustive testing to help address challenges with PS Knees. SOFCAM was designed with the goal of creating optimized cam-spine mechanics for a wide range of patient anatomies. The final components includes improvements on the cam and spine shape, location, and intercondylar box geometry to allow for slow engagement of the femoral component with the tibial spine, providing a smooth transition from condylar control to cam/spine control. Cam and spine shape, location, and intercondylar box geometry have been designed with the goal of creating optimized cam-spine mechanics for a wide range of patient anatomies. References 1. Meftah M, Ranawat AS, Ranawat CS. Ten-year follow-up of a rotatingplatform, posterior-stabilized total knee arthroplasty. J Bone Joint Surg Am. 2012 Mar 7;94(5):426-32. 2. Font-Rodriguez DE, Scuderi GR, Insall JN. Survivorship of cemented total knee arthroplasty. Clin Orthop Relat Res. 1997 Dec;(345):79-86. 3. Dennis DA, Komistek RD, Stiehl JB, Walker SA, Dennis KN. Range of motion after total knee arthroplasty: the effect of implant design and weight-bearing conditions. J Arthroplasty. 1998 Oct;13(7):748-52. 4. Puloski SK, McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB. Tibial post wear in posterior stabilized total knee arthroplasty. An unrecognized source of polyethylene debris. J Bone Joint Surg Am. 2001 Mar;83-A(3):390-7. 5. Furman BD, Lipman J, Kligman M, Wright TM, Haas SB. Tibial post wear in posterior-stabilized knee replacements is design-dependent. Clin Orthop Relat Res.2008 Nov;466(11):2650-5. 6. Chiu YS, Chen WM, Huang CK, Chiang CC, Chen TH. Fracture of the polyethylene tibial post in a NexGen posterior-stabilized knee prosthesis. J Arthroplasty.2004 Dec;19(8):1045-9. 7. Fitzpatric CK, Clary CW, Rullkoetter PJ. Post-Cam engagement During Dynamic Activity with Current Posterior-Stabilized TKR. Presentation number 1700. 18th European Society of Biomechanics. 2012. Lisbon Portugal. 8. Insall JN, Lachiewicz PF, Burstein AH. The Posterior Stabilized Condylar Prosthesis: A Modification of the Total Condylar Design, Two to Four Year Clinical Experience. J Bone Joint Surg Am. 1982; 64-A(9): 1317-1323. 9. Callaghan JJ, O Rourke MR, Goetz DD, Schmalzried TP, Campbell PA, Johnston RC. Tibial post impingement in posterior-stabilized total knee arthroplasty. Clin Orthop Relat Res. 2002 Nov;(404):83-8. 10. Fitzpatrick CK, Clary CW, Rullkoetter PJ. The Influence of Design on TKR Mechanics During Activities of Daily Living. Poster No. 2034. 58th Annual Meeting of the Orthopaedic Research Society. 2012, San Francisco, CA. www.depuysynthes.com DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN 46582 T. +1 (800) 366-8143 DePuy Synthes Joint Reconstruction, a division of DOI 2013 0612-54-513 EO 7/13